FD-7000 User Manual
FD-7000 User Manual
User's Guide
MDD
93/42/EEC
May 1999
8200590 Quint-7000 User's Guide
Manufactured by:
FINNDENT OY
Kenkäpellontie
Teollisuustie 5 4
FIN-07230 MONNINKYLÄ
FIN-04600 MÄNTSÄLÄ
Tel. +358 19 5221500
Tel.
Fax +358 (0)207
+358 43 5115
19 643457
Fax. +358 (0)207 43 5116
Email:info@finndent.com
I
II
Contents
1. Introduction ...................................................................................................... 1
Scope of this manual .......................................................................................................... 1
Associated documents ....................................................................................................... 1
Warnings and precautions .................................................................................................. 2
III
4. Cleaning, disinfecting and sterilising the unit ............................................. 23
Suction system ............................................................................................................ 23
Suction hoses .............................................................................................................. 23
Instruments and suction nozzles................................................................................... 24
Unit surfaces ............................................................................................................... 24
5. Maintenance ................................................................................................... 27
Oil collector ............................................................................................................... 27
Tightening instrument holders ..................................................................................... 27
Instrument water adjustment ........................................................................................ 28
6. Service ............................................................................................................. 29
IV
1. Introduction
2 User's Manual
2. The main parts
Dental light
Instrument arm
Instrument tray
Water bottle
(Option)
Foot control
NOTE
With the dental units Quint 7000 BI, 7000 B2 and 7000 E2
the cuspidor is attached to the floor box via a metal tube.
Metal tube
4 User's Manual
Dentist's Control Panel
Lifting handle
Micromotor spray switch
6 User's Manual
3. Using the dental unit
On
Off
Note
If the dental chair is connected to the power supply
via the dental unit (i.e. the dental chair does not
have its own power supply), the dental chair will
automatically come on when the unit is switched
on.
Note that the chair's ON/OFF switch must be in the
ON position for this to happen.
Cuspidor
Handle
8 User's Manual
If the instrument bridge "drifts" out of position, the
friction of the arm joints needs to be increased.
Contact your service technician for assistance.
Handle
Instrument cart
The Micromotor
10 User's Manual
Activating the spray
Press the right-hand foot switch once to switch the
instrument spray on. When the spray is on the
green M light on the dentist's display will come on.
Press the right-hand foot switch a second time to
switch the spray off. The green light will go out.
Chip blower
Double press and hold down the foot lever. Chip
blow will start. To stop chip blow remove your foot
from the foot switch.
Chip blower
Double press and hold down the foot lever. Chip
blow will start. To stop chip blow remove your foot
from the foot switch.
12 User's Manual
The Scaler
14 User's Manual
The Polymerisation light (Satelec)
Lift the polymerisation light from its holder and
press the button on the side of the instrument. The
light will come on. The light will automatically go
out after 30 seconds. During the time the light is on
you will hear a "beep" every ten seconds. If you
wish to turn the light out before the 30 seconds
elapses, press the button a second time.
Activation switch
Air
Water
Heater off
Heater on
16 User's Manual
Dentist's control panel
Water bottle
Tray holder
The tray holder is attached to a tray holder arm
which is attached to the underside of the instru-
ment bridge on either the left- or the right-hand
side. The arm is angled allowing the tray to be
positioned either high or low depending on which
way round the arm is attached.
To position the tray holder loosen the two tray arm
knobs, move the tray holder to the required posi-
tion and then tighten the two knobs.
Instrument bridge
Tray holder arm
Attaching knobs
Tray holder
18 User's Manual
The Suction Arm
Locking handle
Warning
Do not leave the suction arm under the chair
backrest as the arm could be damaged if you lower
the chair.
Suction nozzle
Decrease
Knob
Suction tip
To attach the suction tip push it into the opening at
the end of the suction nozzle.
The Syringe
Press the green button to activate the water spray.
Press the blue button to activate the air. Press
both buttons together to activate the air/water
Spray spray.
Air
Water
20 User's Manual
Suction arm controls
On/off switch
Intensity switch
22 User's Manual
4. Cleaning, disinfecting and sterilising the unit
Suction system
For information on how to clean the suction system,
refer to the documentation supplied with the
system
Suction hoses
After every patient allow one glass of water to be
sucked through each suction hose.
Connector
Suction hoses
Caution
Switch the dental unit off before you remove any
instruments. Press the syringe air and water
buttons to relieve any pressure before removing the
syringe.
Unit surfaces
Unit surfaces must cleaned and disinfected at least
once a day with a suitable cleaning and disinfecting
agents. Make sure that you prepare and use them
in accordance with the manufactures instructions.
Do not uses abrasive cleaning agents.
Instrument bridge
Instrument support
24 User's Manual
The rollers and holders at the end of the whip arms
can be removed so that the instrument tubes can be
cleaned.
To remove the roller and holder slide the holder up
slightly until the roller can be removed, and then
remove the holder.
Whip arms
Flat
Slot
Instrument holder
Tilt forward Lift and remove
26 User's Manual
5. Maintenance
Oil collector
The cotton pad in the turbine oil collector must be
replace monthly or more frequently if free oil starts
to appear. The oil collector on the underside of the
instrument bridge. To remove and empty the oil
collector unscrew it.
Instrument bridge
Oil collector
Tightening iinstrument
nstrument holders
On units that are fitted with hanging tube
instruments the instrument holders can tilted to
make it easier to remove and replace instruments.
If a holder becomes loose it can be tightened with
screws on the side of the holder.
Instrument holder
Screws
28 User's Manual
6. Service
Every six months a FINNDENT approved service
technician must carry out the following inspection
and service tasks:
• check the output pressures of the air and water
control valves
• check the air pressure of the instruments
• check to see if the filters in the air and water
control valves need replacing
• check the operation of the instruments and if
necessary replace and instrument O-rings or
seals
Caution
Switch the dental unit off before you remove
any instruments. Press the syringe air and
water buttons to relieve any pressure before
removing the syringe.
Micromotor drills
Kavo Intramatic Lux - Models 190, 192 and 196
Turbine drills
Kavo - 630 630B, 640B and 650B
Air micromotors
Kavo - all models
Light probes
Kavo - Dia-Lux 2300L
30 User's Manual
Tartar scalers
Satelec Suprasson SP 4055
Amdent 930
Polymerisation lights
Lysta LC-85
Satelec Acta
Syringes
Quint 3-way syringe
Water heaters
SIE-Dental - water heating element 50/100W, 24V
Amalgam separators
Cattani - Suction separator
Dental lights
ALLOWED OPERATING
CONDITIONS:
- Temperature: 15oC - 40oC
- Relative humidity: 45% - 75%
ELECTRICAL FEATURES:
- Power supply: 230 V, 50 Hz
- Maximum power
consumption: 1200 W (includes power consumption of
possible accessories connected to the unit)
- Main fuse: 6.3 AT.
- Primary fuse of
transformer: 2AT
COMPRESSED AIR
CONNECTION: 5.5 bar - 8 bar
MANUFACTURER:
FINNDENT OY
TEOLLISUUSTIE 5
07230 MONNINKYLÄ
FINLAND
TEL. +358 19 - 5221500
FAX +358 19 - 643457
32 User's Manual
Symbols and abbreviations
v Voltage
A Ampere
W Watt
~ Alternating current
FI Approved by Fimko
34 User's Manual
Quint 7000 Dental Unit 35
36 User's Manual