Professional Documents
Culture Documents
BORU
BORU
C
Title: Performance Qualification Protocol For Superheated Water Sterilizer
model HWSM15010
Doc. No.: Equipment ID: Page:
SS-PQ-001/SV W1501A 1 / 28
Qualification Document
Signature/Date
Production Department
Signature/Date
R& D Department
Signature/Date
Approved by QA Manger
Signature/Date
Authorized By QA Director
1. Objective..................................................................................................................................... 4
2. Scope______________________________________________________________________________________4
3. Concept of the Equipment Qualification___________________________________________________________5
4. Validation Team members______________________________________________________________________5
5. Responsibility________________________________________________________________________________5
6. Training of Personnel__________________________________________________________________________6
7. Installation, Operational Qualification & calibration_________________________________________________6
8. Reference___________________________________________________________________________________8
9. Description of Equipment_____________________________________________________________________10
10. Frequency of Validation______________________________________________________________________11
11.Performance Qualification Tests (PQ-Tests)_______________________________________________________11
12.Deviation List______________________________________________________________________________24
13.Deviation Table (Attach Additional page if required)_______________________________________________25
14.Attachment_________________________________________________________________________________26
15. P-Certificate_______________________________________________________________________________27
1. Objective
The objective of this protocol is to verify the performance qualification attributes of sterilization
and to establish sufficient data to assure that the Superheated Water Sterilizer model
HWSM15010 is suitable for sterilizing planned load and is consistently doing sterilization of
products as expected. The objectives of this qualification is to determine the process variables and
acceptable limits (process time, temperature and pressure), and to set up appropriate in process
controls.
To ensure that heat distribution throughout the chamber is uniform and the temperature is
within the limit of 121 to 123 oC.
To ensure that the heat is sufficiently penetrating into the innermost portions of the load
subjected for sterilization to achieve a temperature of 121 to 123 oC during the sterilization
hold period.
To ensure that the superheated water sterilization process is, when challenged with
Geobacillus steriotherophilus biological indicator spore strips having spore population of
10 6 spores strip, should reduce the bacterial load by more than 6 log reduction.
To verify that the computed F0 value is sufficiently greater than the Biological F0.
To prepare master chart of sterilization cycle for reference during normal production cycle.
To establish the Autoclave Cycle, which when operated within specified operating ranges,
helps to achieve a predetermined Sterility Assurance Level (SAL).
2. Scope
This protocol shall be applicable to the Autoclave Superheated Water Sterilizer model
HWSM15010 located at small volume Injection line of Sansheng Pharmaceuticals plc.
The study design also incorporated empty chamber heat distribution, loaded chamber heat distri-
bution &heat penetration, loaded chamber biological challenge test for maximum load.
In this qualification plan all the required performance qualification tests are described, the accept-
ance criteria are mentioned and the test results will be documented.
If test results meet the acceptance criteria, or the explanation about the deviation is acceptable, it
means that the autoclave is qualified. Any deviation shall be explained clearly.
Validation team shall comprise of representatives from the following functional units:
Quality Assurance Department, including the In-Process Quality Assurance
Production Department
Liquid Workshop
Quality Control Department
Engineering Department
R& D Department
The validation team shall be organized by QA Manager or his/her design.
The List of Validation Team Members (Attach additional page if required)
Sr. No Name Department Position Signature
1
2
3
4
5
6
7
8
9
5. Responsibility
5.1.2. Collection of samples as per the sampling plans by In-Process Quality Assurance.
5.1.3. Review and approval of the data compiled, review of deviations (if any), monitoring
the process as per the process parameters and for withdrawal of validation samples.
5.1.4. Preparation of validation summary report, review and approval.
5.1.5. Approval of the qualification report.
5.2. 5.2.1. Executing the batch.
5.2.2. Review of qualification protocol and summary report.
5.2.3. Preparation and approval of operating procedures (SOPs).
5.2.4. Providing Training.
5.3.5.3.1. Testing of samples as per the test procedures.
5.3.2. Review of validation protocol and summary report.
5.3.3. Providing Training.
5.4.
5.4.1. Providing technical support during qualification.
5.4.2. Review of validation protocol and summary report.
5.4.3. Make the machine accessible for qualification and feeding the necessary test parameter.
5.4.4. Calibration and verification of measuring devices and instruments.
6. Training of Personnel
Before validation, the related person should be trained of this validation protocol and related
SOPs. The training record shall be attached behind this validation protocol for reference.
7. Installation, Operational Qualification & calibration
7.1 Confirmation/verification
It is relevant document and other details (Attach Additional pages if required)
Conclusion:
Check by / Date: Reviewed by/Date:
8. Reference
8.1.
SN Reference Document No.
1 Standard Operating Procedure for Superheated Wa-
ter Sterilizer model HWSM15010
2 Standard operating procedure for Validator
3 Standard operating procedure for calibration of
measuring and testing Devices.
4 Standard operating procedure for data loggers
5 Standard operating procedure for handling of Biolo-
gical Indicator
6 Standard operating procedure for Risk Management
7 Standard operating procedure for Deviation man-
agement
Sansheng Pharmaceutical Plc.
Sansheng Pharmaceutical P.L.C
Title: Performance Qualification Protocol For Superheated Water
Sterilizer model HWSM15010
Doc. No.: Equipment ID: Page:
SS-PQ-001/SV W1501A 9 / 28
9. Description of Equipment
10.1.
Three (3) successful runs for all planned loads
10.2.
One successful run in a year for all loads at defined maximum load pattern.
If the autoclave has undergone major repair or moved to a new location that has different fa-
cility supplies.
Any changes in the load size, load configuration, or container characteristics (volume, geo-
metry, etc).
If replacement of Major components /Instruments
Changing in PLC programme
11.
Operate in accordance with the "HWSM15010 Water Bath Sterilizer Operation Standard
Operating Procedures no………", confirm various parameters and operation reliability of the
water bath sterilizer during operation, and prove that it can achieve the predetermined design
capacity, meet the sterilization requirements, and make the sterilization.
The performance evaluation of the sterilizer includes: no-load (Empty chamber) heat distribution
test, half-load and full-load heat distribution test, half-load and full-load heat penetration test as
part of biological indicator challenge test, each test will be conducted three consecutive times to
confirm the reproducibility of the equipment performance.
11.1. 11.1.1 No-load (Empty chamber) heat distribution test
The intent of this study is to demonstrate the temperature uniformity and stability of the sterilizing
medium throughout the loaded sterilizer.
Procedure
Distribute the calibrated probes evenly in the sterilizer under test, and divide them into 3
layers, upper, middle and lower, with 5 probes in each layer; the top view of the specific
placement position of the probes is shown in the schematic diagram. The probes shall not
touch the wall of the autoclave because walls will have high temperature than the chamber.
Connect the line between the probe and the data logger.
Set the sterilization program at 121°C for 30 minutes, and operate the autoclave as per SOP
…… to collect data for the processes of heating, sterilization, cooling, and drainage from
the start of sterilization.
Turn on the two data logger and start to collect data. The frequency of data collection is
once every 30 seconds.
Acceptance Criteria
Compare the temperature measured by each probe every 2 minutes, and the difference
between the average temperature and the temperature of each probe does not exceed
±1°C.
The temperature at each temperature mapping probe should be within the range of 120
ºC to 123ºC during the sterilization hold period.
The difference between lowest and highest temperature shall not be more than 2°C.
Throughout the dwell time, all temperatures measured in the chamber should not fluctu-
ate by more than 1oC.
Throughout the dwell time, all temperatures measured in the chamber should not differ
from each other by more than 2oC.
The record and analysis of empty chamber heat distribution results are shown in the following ta-
ble:
(A) (B) (C)
(min) Maximum Minimum Average
time temperature temperature temperature A-C C-B A-B
(min) Probe Temperat- Probe Temperat- (C)
No. ure No. ure
0
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
Cold Spot:
Hot Spot:
Statistical analysis
Perform ANOVA to check the presence/absence of significant variation between sterilization tem-
peratures of the three runs at hold temperature.
Procedure
Distribute the calibrated probes evenly in the sterilization cabinet to be tested, and divide
them into 3 upper, middle and lower layers, with 5 probes in each layer; the top view of the
specific placement of the probes is shown in the schematic diagram.
Connect the line between the probe and the verification instrument, turn on the two
verification instruments, calibrate the time of the verification instrument and the sterilization
equipment, and start collecting data. The frequency of collection is once every 30 seconds.
Set the sterilization program at 121°C for 30 minutes, and start the sterilization cabinet to
collect the processes of heating, sterilization, cooling, and drainage from the start of
sterilization.
After the sterilization is over, keep the records. Turn off the instrument and repeat the test
three times.
Product name: ;________________________
Specification: _________________________ ;
Lot number ; __________________________
Quantity: __________plate, each plate is about ______ piece, total ___________ piece
18 19
20
21
22
Prob No. 18 19 20 21 22
Placement
Prob No. 23 24 25 26 27
Placement
Prob No. 28 29 30 31 32
Placement
Prob No. 33 34
Placement
Cold Spot
Hot Spot
Confirmed by Year Month date
Confirmed by Year Month date
Result Analysis
Acceptance Criteria
The minimum Fo value shall not be less than 12 min
11.1.4
Distribute the calibrated probes evenly in the sterilization cabinet to be tested, and divide
them into three layers: upper, middle and lower layers, with 5 probes on each layer; the
top view of the specific placement of the probes is shown in the schematic diagram.
Connect the line between the probe and the verification instrument, turn on the two
verification instruments, calibrate the time of the verification instrument and the
sterilization equipment, and start collecting data, and the collection frequency is once
every 30 seconds.
Set the sterilization program at 121°C for 15 minutes, and start the sterilization cabinet to
collect the processes of heating, sterilization, cooling, and drainage from the start of
sterilization.
After the sterilization is over, keep the records. Turn off the instrument and repeat the test
three times.
Probe placement results:
Sansheng Pharmaceutical Plc.
Sansheng Pharmaceutical P.L.C
Title: Performance Qualification Protocol For Superheated Water
Sterilizer model HWSM15010
Doc. No.: Equipment ID: Page:
SS-PQ-001/SV W1501A 19 / 28
The top layer of the sterilizer is 35/36/37/38/39, the middle layer is 40/41/42/43/44, the
bottom layer is 45/46/47/48/49, and there is one probe for the inlet and outlet of high
temperature circulating water. There are 15 probes in the sterilizer, 17 in total.
Top view of probe distribution
35
36
37
38
39
Probe No. 35 36 37 38 39
Placement
Probe No. 40 41 42 43 44
Placement
Probe No. 45 46 47 48 49
Placement
Probe No. 50 51
Placement
Throughout the dwell time, all temperatures measured in the chamber should not fluctuate
by more than 1oC.
Throughout the dwell time, all temperatures measured in the chamber should not differ
from each other by more than 2oC.
Minimum F0 during sterilization hold period shall be more than 12 min (Which is F 0 value
of biological indicator).
The record and analysis of no-load heat distribution results are shown in the following table:
(A) (B) (C)
(min) Maximum Minimum Average
time temperature temperature temperature A-C C-B A-B
(min) Probe Temperat- Probe Temperat- (C)
No. ure No. ure
0
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
Result Analysis
Statistical analysis
Perform ANOVA to check the presence/absence of significant variation between sterilization tem-
peratures of the three runs at hold temperature
11.1.5.
This test is carried out along with the full load heat distribution test, repeated three times.
This challenge test serves as heat penetration study confirmation test.
Place 15 Bacillus stearothermophilus biological indicators (packed in vials/Ampules along
with the product as per layout of probe distribution and put them into different positions in
the sterilization cabinet to ensure that a bacterial tube is placed in the cold spot area of the
sterilizer. After the sterilization is completed, take out the biological indicator, and operate
together with a control tube according to the incubation temperature (56-60°C) and
incubation time (24-48 hours) specified in the instruction manual of the biological indicator.
After incubation, if the sterilization tube does not change color (purple), it means that the
biological indicator test has passed, and the sterilization tube changes color (yellow),
indicating that the biological indicator test fails, and the cultured control tube should be
positive, that is, it will change color (yellow). Otherwise, the test is invalid. The above test
should be carried out three times in a row to confirm the reproducibility of the sterilization
process.
Biological indicator name: Bacillus stearothermophilus ATCC7953
Acceptable Criteria
6-log reduction should be observed in any exposed biological indicators on complete incubation.
Biological Indicators and Heat Penetration Testing record
Validation results Control bacteria
label Location Remark
Negative Positive Negative Positive
Sansheng Pharmaceutical Plc.
Sansheng Pharmaceutical P.L.C
Title: Performance Qualification Protocol For Superheated Water
Sterilizer model HWSM15010
Doc. No.: Equipment ID: Page:
SS-PQ-001/SV W1501A 23 / 28
12.Deviation List
SN Deviation name Deviation No.
14.Attachment
In the table all the attachments which are used for the Qualification, have to be mentioned.
Of the attachments, which are not attached to the Qualification protocol, from where it
comes has to be mentioned clearly.
Attachment 01: Common for all tests
Calibration Report of Thermocouples
Verification Report of Thermocouples
Attachment 02: For Heat Distribution Study
Standard Format: for logging data
Yokogawa Graph
Sansheng Pharmaceutical Plc.
Sansheng Pharmaceutical P.L.C
Title: Performance Qualification Protocol For Superheated Water
Sterilizer model HWSM15010
Doc. No.: Equipment ID: Page:
SS-PQ-001/SV W1501A 27 / 28
PLC print
Attachment 03: For Heat Penetration Study /Biological Challenge test
Pictures showing Color of Biological indicator before and after incubation
Standard Format: for logging data
Yokogawa Graph
PLC print
15. P-Certificate
The persons, who have signed the document, confirm the completeness of the test documentation
and the execution of the tests.
The execution and the results fulfil the requirements. No deviation is applicable. The equipment
has been released for the PQ activities and contract requirement.
The execution and the results do not fulfil the requirements. Critical deviations are not noticed.
The equipment has been released for the PQ activities and contract requirement.
The execution and the results do not fulfil the requirements. Critical deviations are noticed. The
equipment has not been released for the PQ activities and contract requirement. Further measures
have to be done. Corrections have to be documented separately.
Sansheng Pharmaceutical Plc.
Sansheng Pharmaceutical P.L.C
Title: Performance Qualification Protocol For Superheated Water
Sterilizer model HWSM15010
Doc. No.: Equipment ID: Page:
SS-PQ-001/SV W1501A 28 / 28
Conclusion
Final Approval
QA Director Signature /Date