Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

EQUIPMENT CARE AND MAINTENANCE

Sponsored by the IAPB Standard List • www.iapb.standardlist.org

Figure 1. A systematic approach to


school eye health Electrosurgical units – how they
In the school
As part of the curriculum [using work and how to use them safely
the Healthy Eyes Activity Book]
• Education on how to keep eyes healthy Ismael Cordero both the active electrode and return
• Personal hygiene education, which Biomedical Service Manager: Gradian electrode functions are performed at the
Health Systems, New York, USA. site of surgery. The two tips of the
includes face washing
ismaelcordero@me.com
• Children encouraged to take these forceps perform the active and return
health messages home Electrosurgery is used routinely in eye electrode functions. Only the tissue
• Primary eye care provided by a surgery to cut, coagulate, dissect, grasped in the forceps is included in the
trained school nurse or teacher fulgurate, ablate and shrink tissue. electrical circuit. Because the return
High frequency (100 kilohertz to function is performed by one tip of the
Visit by the eye care team
5 megahertz), alternating electric current forceps, no patient return electrode is
• Screen teachers and alert them to
at various voltages (200–10,000 Volts) needed. Bipolar electrosurgery operates
eye conditions/low vision
is passed through tissue to generate regardless of the medium in which it is
• Train teachers to screen visual acuity
heat. An electrosurgical unit (ESU) used, permitting coagulation in a fluid
at 6/12 level
consists of a generator and a handpiece environment – a great advantage when
After visit by the team with one or more electrodes. The device attempting to coagulate in a wet field.
• Teachers screen children and list is controlled using a switch on the As a result, bipolar electrosurgery is
those who fail handpiece or a foot switch. often referred to as ‘wet field’ cautery.
Second visit by the eye care team Electrosurgical generators can In monopolar electrosurgery
• Refract and dispense spectacles to produce a variety of electrical (Figure 2), the active electrode is placed
children with significant RE waveforms. As these waveforms change, at the surgical site. The patient return
so do the corresponding tissue effects. electrode (also known as a ‘dispersive
In bipolar electrosurgery (Figure 1), Continues overleaf ➤
Refer children with complex
refractive error and other eye Figure 1. Bipolar electrosurgery
conditions Active

In the eye unit Return


• Refract and dispense spectacles to Electrosurgical
children with complex prescriptions generator Handpiece
• Diagnose and manage other eye
conditions
• Low vision assessment. Prescribe
low vision devices if required, and
provide training in their use

In the school
Post-service
• Encourage children to wear their Return Active
spectacles in class
• Support children with low vision
Tissue

Compliance monitoring
by eye care team Figure 2
• Ensure children wear their spectacles

References
1 International Agency for the Prevention of Blindness. Electrosurgical Handpiece
IAPB Briefing Paper: School Health Programme generator Patient return Active
Advocacy Paper. 2011. Available at: http://www.iapb. electrode electrode
org/sites/iapb.org/files/School%20Health%20
Programme%20Advocacy%20Paper%20BP.pdf.
Accessed: February 2015.
2 World Health Organization. Visual impairment and
blindness – Fact Sheet No. 282. 2012; Available from:
http://www.who.int/mediacentre/factsheets/fs282/
en/. Accessed: February 2015.
3 Wu L, Sun X, Zhou X Weng C. Causes and 3-year-
incidence of blindness in Jing-An district, Shanghai,
China 2001–42009, BMC Ophthalmol 2011;11:10.

© The author/s and Community Eye Health Journal 2015.


This is an Open Access article distributed under the COMMUNITY EYE HEALTH JOURNAL | VOLUME 28 ISSUE 89 | 2015 15
Creative Commons Attribution Non-Commercial License.

CEHJ89_OA_Revise.indd 15 01/06/2015 16:15


EQUIPMENT CARE AND MAINTENANCE
Sponsored by the IAPB Standard List: a great platform to source and compare eye health products www.iapb.standardlist.org

pad’ is placed somewhere else on the

Heiko Philippin
patient’s body. The current passes
through the patient as it completes the
circuit from the active electrode to the
patient return electrode. The function of
the patient return electrode is to remove
current from the patient safely. A return
electrode burn will occur if the heat
produced, over time, is not safely dissi-
pated by the size or conductivity of the
patient return electrode.
Modern electrosurgical machines have
built-in safety features to prevent burns
from occurring due to poor contact between
the patient and the return electrode when
using the monopolar mode.
Often, the term ‘electrocautery’ is
incorrectly used to describe electro-
surgery. Electrocautery refers to direct
current (electrons flowing in one direction)
Electrosurgery. TANZANIA
whereas electrosurgery uses alternating
current. In electrosurgery, the patient is eschar (dead tissue from burning) builds a metal clip creates an electrical
included in the circuit and current enters up on the tip, electrical impedance transformer that can cause a hazard
the patient’s body. During electrocautery, increases and this can cause arcing, and may ignite drapes.
current does not enter the patient’s body. sparking or ignition and flaming of the • Never operate electrosurgical
Instead, current flows through a heating eschar. When cleaning the electrode, equipment with wet hands or wet
element, which burns the tissue by direct the eschar should be wiped away using gloves. If sterile gloves have holes
transfer of heat. Electrocautery or, more a sponge rather than the common in them, electrical current can pass
precisely, thermocautery units (Figure 3) scratch pad, because these pads will through. Be sure that all team members
are usually portable battery powered scratch grooves into the electrode tip, at the surgical field have intact gloves.
devices that can be either disposable increasing eschar build-up. • Never operate electrosurgical
or reusable. equipment while standing on a wet
Don’ts
surface. Keep the foot pedal dry. Protect
Using the ESU safely • ESUs should not be used in the
it from fluid spillage by covering it with a
presence of flammable agents or in
ESUs produce very high current that can clear, waterproof cover.
oxygen-enriched environments.
injure both patient and operator if not
• Avoid using flammable substances
properly used and maintained. Many Monopolar electrosurgery
that can be ignited by sparks, such
problems have been associated with the • Determine whether the patient has
as alcohol and skin degreasers. If you
use of ESUs, such as burns at the return any metal implants, including cardiac
must use alcohol-based skin preps,
electrode site and surgical fires. Some of pacemakers. There is potential for injury
do not allow them to pool near the
these safety problems can be avoided by if a patient return electrode is placed
dispersive pad; be sure prep solutions
taking simple precautions. on the skin over a metal orthopaedic
are thoroughly dry and fumes have
implant.
Dos dissipated before ESU activation.
• For optimum safety, have the patient
• The hand piece should always be placed • Rubber catheters or other materials
remove any jewellery to avoid complications
in the nonconductive holster when not should not be used as a sheath on
from possible current leakage.
in use. active electrode tips.
• Position and insulate the patient so that
• Always use the lowest possible • Cables should never be wrapped around
she or he is not touching any grounded
generator setting that will achieve the metal instruments, as the current
metal objects.
desired surgical effect. When higher running through them can pass into the
• Choose a location for the return
than necessary voltages are used, the metal instrument, causing burns.
electrode/dispersive pad that is as
chances of arcing are increased. If the • Do not use sharp towel clips or metal
close to the operative site as possible,
surgeon continues to ask for a higher instruments to attach cables to drapes.
clean and dry, well vascularised, and
setting, this could be a signal that the Sharp metal clips can damage electrical
over a large muscle mass. Avoid bony
integrity of the skin/dispersive pad cables or provide an unwanted point
prominences, adipose tissue, scar
interface is compromised. of contact with the patient’s skin.
tissue, skin over implanted metal
• Clean the electrode tip frequently. As Overlapping electrical wire around
prostheses, hairy surfaces, and pressure
Figure 3. Forceps for electrocautery/thermocautery
points. If necessary, shave very hairy skin
at the dispersive pad site. Make sure that
Activation
button conductive gel is moist and uniformly
spread all over the contact area and that
Batteries the dispersive pad achieves uniform
contact with the patient’s skin.
Handle • Position ECG electrodes away from
Heated
the electrosurgery site and the current
tip
pathway through the body.

16 COMMUNITY EYE HEALTH JOURNAL | VOLUME 28 ISSUE 89 | 2015 © The author/s and Community Eye Health Journal 2015. This is an Open Access
article distributed under the Creative Commons Attribution Non-Commercial License.

CEHJ89_OA_Revise.indd 16 01/06/2015 16:15

You might also like