Download as pdf or txt
Download as pdf or txt
You are on page 1of 16

Technical Information EN 08/2017

Software 1.02.00, Rev 1.00.00


Dialog iQ Software 1.02.00 Technical Information 08/2017
Copyright
This document is the property of B. Braun Avitum AG with all rights reserved.
Commissioning and Service
Only trained personnel must service the Dialog iQ, i.e. repair, maintenance, software
installation, firmware update, retrofitting and commissioning of the Dialog iQ. Servicing
must only be performed with proper tools, calibration equipment and be in accordance
with the most recent revision of this service manual/technical information, which must be
clearly and thoroughly understood.
Prevent Electrical Shock Hazard
Switch off the Dialog iQ and disconnect unit from mains if you have to open the machine
for servicing. Do not touch any exposed wiring or conductive surfaces while the Dialog iQ
is opened. The voltages present when electrical power is connected to the Dialog iQ can
cause serious injury or death.
ESD Information
Pay attention to ESD information, because electronic components are sensitive to
electrostatic discharges.
High Voltage in TFT Monitor

! High voltage can be present at the backlight inverter board BIB in the TFT monitor, even if
the machine has been disconnected from mains. Pull out the battery compartment in the
base platform and switch off the battery voltage before opening the machine.
Protective Conductor in TFT Housing
If the TFT housing had to be opened during a service job, the tight seat of the protective
conductors in the TFT housing must be checked.
TSM Service Program
Only activate the TSM service program for service activities. It is prohibited to connect a
patient to the Dialog iQ and to run a therapy if the TSM service program is activated in
the Dialog iQ. If the TSM service program is activated the complete alarm system is
disabled. The TSM service program is started in the service mode: digital board, service
switch S1, position 2.
Software
The software is installed in the software mode: digital board, service switch S1, position 3.
Therapy Mode
After completion of all procedures switch back to the therapy mode: digital board, service
switch S1, position 0.
Calibration
Only perform a calibration after the Dialog iQ has reached working temperature and the
machine was disinfected and decalcified. Save the calibration data (CFC) before you exit
the TSM service program: TSM Main Menu, File Operations, Save Calibration Data.
Tubing
Tubing must be replaced only by the same tubing type/length and identical installation
manner. Make sure that the tubings in the machine are not kinked or twisted after
servicing (e.g. if sub-racks are pulled out and inserted again). The tubing must not touch
moving/rotating components (e.g. motors of gear pumps).
Wiring
Wiring must be replaced only by the same cable type/length and identical installation
manner. The cables must not touch moving/rotating components (e.g. motors of gear
pumps).
Prevent Chemical Burns and Scalding
During servicing on running machines: prevent chemical burns and scalding of the skin
due to the penetration of disinfectant or hot liquid.
Contaminated Machines
Protective gear should be worn in case of servicing of assumed contaminated machines.

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 2/16
Dialog iQ Software 1.02.00 Technical Information 08/2017
Cover in Rear Door
Servicing of mechanical assembly groups (components in contact with fluid): the cover for
the switch mode power supply microcontroller SMPS-MC in the rear door must be
assembled during servicing because it serves as a spray protection.
O-Rings
Always check o-rings from disassembled groups/components and replace if necessary.
Figures
The displayed figures can differ slightly from the machines on site, due to different
hardware statuses.
Fuses
If fuses are replaced they must exactly match the type and rating specified by the
manufacturer in the spare parts list/technical information. Where applicable: fuses must
be approved by UL/CSA.
Spare Parts
Only use original spare parts manufactured and sold by B. Braun Avitum AG.
System Configuration
The system configuration saved on a FSU USB stick must be downloaded to another
Dialog iQ machine only if:
 the hardware matches and
 the identical software version number is present.
Function Check
Check the respective function of the assembly group/component after servicing. A
complete function check must be performed after every service, according to the repair
matrix.

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 3/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

Table of Contents Page

General Information 4
Software 1.02.00 for Dialog iQ 4
Ordering Numbers 4
1. Hardware and Software Status 5
2 Software Modifications 5
2.1 Software Modifications from SW 1.01.21 to SW 1.02.00 5
3 Software Installation 6
3.1 FSU Stick V2.91 6
3.2 Prior to Installation of Software 6
3.2.1 Document Customer Specific Parameters 6
3.2.2 Document DF/HDF Filter Data 6
3.2.3 Only for Software Upgrade from Version 1.01.21 to Version
6
1.02.00
3.3 After Installation of Software for First Machine 6
3.4 After Installation of Software for Next Machines 7
3.4.1 FSU Stick to Enter Parameters in TSM 7
3.4.2 Set BICLF Ratio 7
3.4.3 Set DF/HDF Filter Data 7
4. New/Modified Menus in TSM Service Program 8
4.1 Dialysate Side Parameters 8
4.2 ABPM Parameters 9
5. Test Run 9
6. Default Table SW 1.02.00 10

General Information
This technical information describes the Dialog iQ software 1.02.00. The field service
utility FSU stick V2.91 has to be used to save and transfer parameters with the FSU stick
(FSU stick V 2.00 and 2.90 can not be used).

Software 1.02.00 for Dialog iQ


The following software sub-versions are included in the software version 1.02.00:
 Dialog iQ TLC SW version 1.02.00  Dialog iQ LLC FW version 1.00.00
 Dialog iQ LLC SW version 1.02.00  Dialog iQ LLS FW version 1.01.00
 Dialog iQ LLS SW version 1.02.00

Ordering Numbers
Art. No. Description
770102 Software 1.02.00 for Dialog iQ
770291 Field Service Utility FSU V2.91

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 4/16
Dialog iQ Software 1.02.00 Technical Information 08/2017
1. Hardware and Software Status
See table for hardware and software modifications depending on the installed software
version.
Dialog iQ Service Hardware/Software Modifications
Software Documentation for Dialog iQ Machines
SW 1.01.21 TI 03/2017 Dialog iQ Machines
 We recommend updating to software 1.01.21 during the next service job.
Note
 All SW 1.01.00 features are included in the SW 1.01.21.
 The software does not require any application training.
 A hardware upgrade is not required.
SW 1.02.00 TI 08/2017 Dialog iQ Machines
 We recommend updating to software 1.02.00 during the next service job.
Note
 All features of previous software versions are included in the SW 1.02.00.
 The software does not require any application training.
 A hardware upgrade is not required.

2. Software Modifications
2.1 Software Modifications from SW 1.01.21 to SW 1.02.00
1. BPA rotates backwards for 0.6 s instead of 4 s during the Optimization
positioning of the PA POD (after a failed BS pressure test, if PA
POD is configured) to prevent cracks of POD.
2. Possible to drain the bicarbonate cartridge during Reinfusion Optimization
3. The bioLogic version number is incremented to 3.1.2 on the Optimization
Service Screen.
4. "1015 - TMP out of alarm limits" alarm reset mode changed to Optimization
use the AQ button.
5. Starting pressure for ABPM measurement during inflation Optimization
corrected
6. Rounding error in displayed UF profile in EOT (End of Therapy) Optimization
fixed
7. Machine remains correctly switched on when “auto-off Optimization
stopped” is activated by the user.
8. The correct dialysate fluid temperature is displayed when Optimization
temperature profile is used
9. Pictures on the GUI during Preparation show the position of Optimization
the dialyzer couplings correctly
10. Rinsing of the CCS lines for 30 seconds during Preparation to Extended function
keep the CCS lines open, if this feature is activated in TSM.
11. Display of ABPM values also in Reinfusion and Dialyzer Extended function
Draining
12. Progress bar displayed during the system boot process Extended function
13. ABPM measurement during inflation possible Extended function
14. Occurrence of data communication error reduced further Increase of availability
15. Cause for alarm “Data Inconsistent” in auto disinfection Increase of availability
corrected
16. Cause for alarm “Data Inconsistent” after modifying UF profile Increase of availability
in therapy corrected
17. Corruption of the CF card prevented in cases where the user Increase of availability
switches the machine off during the boot sequence
18. Various causes for the alarm “Input Data Inconsistent” were Increase of availability
fixed

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 5/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

3. Software Installation
The standard installation of the software and the required tests and measures are
described in detail in the Dialog iQ service manual, chapter 5. The following information
describes additional steps which have to be performed during this software installation.
The factory default parameters are available after the software installation.
3.1 FSU Stick V2.91
The FSU stick version 2.91 has to be used to transfer the configuration data from one
machine of software version 1.02.00 to any other machine of the same software version
and hardware status.
The FSU stick version 2.90 is obsolete and shall be replaced by the new version 2.91. Some
functions cannot be used on machines of software version 1.02.00 and higher.
3.2 Prior to Installation of Software
3.2.1 Document Customer Specific Parameters
Use the default table and document all customer specific parameters in TSM/Treatment
Support, because these have to be entered manually on the first machine after software
installation. Subsequent machines with the same software and hardware status can be
updated with the FSU stick.
3.2.2 Document DF/HDF Filter Data
The filter data might get lost during the software update and have to be set manually
after the update.
1. Switch on machine in therapy mode.
2. Select Disinfection/Filter tab and record the Remaining Time and Remaining
Treatments for the current DF/HDF filters.
3.2.3 Only for Software Upgrade from Version 1.01.21 to Version 1.02.00
Download all Trend files and Error Log files to the FSU stick prior to software installation
and load those files to the existing B.DoCS folder in the Windows Explorer under
Enterprise Connect into B.DoCS – PRD:
B.Braun\Avitum\Division\Project Management Avitum\03 Projects\iQ Trends

3.3 After Installation of Software for First Machine


Enter Parameters in TSM
1. Use default table and check all configuration parameters manually in TSM service
program.
2. Enter all customer configuration parameters documented from the default table, if
any are different from the factory default.
3. Save the configuration parameters to the FSU stick version 2.91.

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 6/16
Dialog iQ Software 1.02.00 Technical Information 08/2017
3.4 After Installation of Software for next Machines
3.4.1 FSU Stick to Enter Parameters in TSM
Use FSU stick to transfer the configuration parameters and the customer specific
parameters documented in the default table for the first machine to the next machines.
Note: The new FSU stick version 2.91 has to be used to transfer the system configuration
from one machine of SW version 1.02.00 to another machine.
3.4.2 Set BICLF Ratio
1. Switch on machine in TSM mode.
2. Select Treatment Support, System Configuration and open Dialysate Side Parameters
menu.
3. Change the BICLF Ratio for bicarbonate canisters manually from 25 to 27.
3.4.3 Set DF/HDF Filter Data
4. Select Treatment Support, System Configuration and open Filter Parameters menu.
5. Enter the Therapy No. to the value previously recorded as Remaining Treatments
prior to software installation.
6. Enter the Filter Operation Time to the value previously documented as Remaining
time prior to software installation.
7. Save settings.
8. Switch off machine and switch on machine in therapy mode.
9. Select Disinfection/Filter tab and start Filter Draining.
10. Press OK in all pop-up windows.
11. Check if filter data match the documented data.
12. Switch off machine and switch on in TSM mode.
13. Select Treatment Support, System Configuration and open Filter Parameters menu.
14. Enter the default values, i.e. 150 treatments for Treatment No. and 900 hours for
Filter Operation Time.
15. Save settings.
16. Switch off machine and switch to therapy mode.

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 7/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

4. New/Modified Menus in TSM Service Program


4.1 Dialysate Side Parameters (from SW 1.02.00)
Central Concentrate Supply CCS

The parameters can be selected if Central Concentrate Supply is activated in Low Level
Options.
CSS Rinsing in Preparation
The new parameter can be selected with the Yes button.
Both CCS lines are rinsed for 30 seconds during Preparation to keep the CCS lines open, if
Yes is selected.
Mixing Ratios

BICLF Ratio
Change the mixing ratio setting for the BIC canisters manually from 25 to 27.

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 8/16
Dialog iQ Software 1.02.00 Technical Information 08/2017
4.2 ABPM Parameters (from SW 1.02.00)

ABPM Reading During Cuff Inflation


The selection is only enabled if the installed ABPM is able to use this feature, i.e. from
firmware 6.0 onwards.
If Yes (default) is selected: The ABPM measures the systolic/diastolic blood pressure already
during inflation and stops the inflation when the measurement is successful.
If No is selected: The ABPM inflates the cuff to the set pressure and measures the
systolic/diastolic blood pressure during the deflation.
5. Test Run
The test run should be performed under normal environmental condition, i.e. at a room
temperature of 25 °C ±5 °C.
1. Press the Connect Patient icon to switch to therapy mode. This icon is enabled after
all self tests were performed successfully.
2. Turn service switch S1 to position 0 - Therapy and close both doors.
3. Start machine (fully equipped) in dialysis mode.
4. Insert line system with dialyzer and prime and rinse with fluid – see instructions for
use, i.e. perform preparation (insert heparin syringe and select 2 ml/h and use BIC
cartridge if present).
5. Perform self test according to instructions for use.
6. Set dialyzing fluid flow to 800 ml/min.
The machine must run without errors up to the message Connect Patient.
7. Press the Connect Patient icon to switch to therapy mode. This icon is enabled after
all self tests were performed successfully.
8. Immerge arterial and venous lines in container with water and make sure that no air
is drawn in.
9. Set the blood pump so that no blood side alarms are activated.
10. Start and run test therapy for at least 15 minutes without alarms.
11. Terminate therapy.
12. Remove disposables.
13. Perform a disinfection/decalcification of machine.
14. Switch off Dialog iQ.

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 9/16
Dialog iQ Software 1.02.00 Technical Information 08/2017
7. Default Table SW 1.02.00
S ys t e m C on f i gu r at io n
Unit HD HDF Online Modified
Dialysate Side Parameters
Na Concentration/Conductivity Measurement Unit - mS/cm mS/cm
Acetate Mode - YES YES
Default Conductivity Mode - BIC BIC
Stand-by after Self Test/Rinsing - NO NO
Maximum Stand-by Time h:min 1:00 1:00
Selected Acetate - Acetate Acetate
Selected Bicarbonate - BIC without NaCl BIC without NaCl
Selected Acid - Acid 1 Acid 1
Automatic Bypass Deactivation at Patient Connection - NO -
Automatic Drain Dialyzer/Bic Cartridge - NO -
Bicarbonate Cartridge Change with Draining - NO -
Temperature Measurement Unit - °C °C
Dialysate Flow ml/min 500 500
Dialysate Temperature °C 36.0 37.0
Concentrate Source - Concentrate -
Central 1 Name - Central 1 -
Central 2 Name - Central 2 -
CCS Rinsing in Preparation - No No
BICLF Ratio - *27 *27
* to be changed manually from 25 to 27.
ENDL Ratio - 29 29
BICLF Cart Ration - 35 35

Selected Acetate
1st to 10th Concentrate Name - Acetate 1 – 10
Conversion Factor mS/cm per mmol/l 0.10500
Default Value mS/cm 14.3

Selected Bicarbonate
Bicarbonate without NaCl - BIC without NaCl
Conversion Factor mS/cm per mmol/l 0.0950
Default Value mS/cm 3.0
Acid Selection - Acid 1

Selected Acid
1st to 20th Concentrate Name - Acid 1 – 20
Conversion Factor mS/cm per mmol/l 0.10500
Default Value mS/cm 14.3

Blood Side Parameters


Blood Flow for Connection Patient ml/min 100 100
Default Blood Flow at Reinfusion ml/min 100 100
Bloodline Configuration for Art. Pressure Measurement - T Piece T Piece
Bloodline Configuration for PBE Measurement - Chamber Chamber

Min-Max Parameters
PA
PA Upper Delta mmHg 70 70
PA Lower Delta mmHg 70 70
PA Min. Limit mmHg -300 -300
PV
PV Max. Limit mmHg 250 250
PV Upper Delta mmHg 70 70
PV Lower Delta mmHg 35 35
PV Min. Limit mmHg 20 20

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 10/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

Unit HD HDF Online Modified


PBE
PBE Max. Limit mmHg 700 700
PBE Upper Delta mmHg 150 150
TMP
TMP Max. Limit mmHg 350 350
TMP Lower Delta mmHg 50 50
TMP Upper Delta mmHg 50 50
TMP Min. Limit mmHg 0 0
Monitoring Limits Low-High TMP - NO NO

Infusion Bolus Parameters


Infusion Bolus Volume ml 100 100

Single Needle Parameters


Single Needle Valve
Min. Limit PA for SN Valve mmHg -180 -180
Max. Control Press. Min. PV mmHg 150 150
Min. Control Press. PV mmHg 100 100
Max. Control Press. PV mmHg 350 350
Single Needle Cross Over
Control Press. PA mmHg -200 -200
Control Press. PV mmHg 260 260
Access Type - Cannula Cannula
Default Phase Volume Cannula ml 20 20
Default Phase Volume Catheter ml 30 30
Mode - Manual -

Heparin Parameters
Treatment with Heparin - YES YES
Heparin Measurement Unit - ml/h ml/h
Conversion Factor Heparin (IU/ml) - 1000 1000
Max. Heparin Rate ml/h 10.0 10.0
Heparin Rate ml/h 3.0 3.0
Heparin Bolus Volume ml 2.0 2.0
Heparin Stop Time h:min 00:30 00:30
Warning at Heparin Stop Time - NO NO
Automatic at Start of Therapy - NO NO
Selected Syringe Type - Omnifix 30 ml Omnifix 30 ml
Syringe Table Inner Diameter [mm]
Braun OPS 20 ml 19.00
Terumo 20/25 ml 20.10
Terumo 30/35 ml 22.57
B-D 20 ml 18.90
B-D 30 ml 21.53
Monoject 20 ml 19.83
Dispomed 30 ml 22.17
Omnifix 10 ml 16.05
Omnifix 20 ml 19.80
Omnifix 30 ml 22.00
Syringe 11 – 20 99.99

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 11/16
Dialog iQ Software 1.02.00 Technical Information 08/2017
Unit HD HDF Online Modified
Rinsing Parameters
Priming with Circulation - NO NO
NaCl Bag: Priming Rate ml/min 150 150
NaCl Bag: Priming Volume ml 700 700
Online Priming Rate ml/min 150 150
Online Priming Volume ml 700 700
Automatic Priming - YES YES
Automatic Rinse Program after Self Tests - NO NO
UF Pump Rinsing Time h:min 00:59 00:59
UF Pump Rinsing Rate ml/h 203 203
UF Pump Rinsing Volume ml 200 200
Blood Pump Rinsing Time h:min 00:05 00:05
Blood Pump Rinsing Rate ml/h 6000 6000
Blood Pump Rinsing Volume ml 500 500

UF Parameters
Minimal UF Rate ml/h 50 50
UF Rate Compensation - YES YES
Max. Value Upper Limit UF Rate ml/h 3000 3000
Default Value Upper Limit UF Rate ml/h 2000 2000
Therapy Time h:min 04:00 04:00
Absolute End of Therapy Time setting - NO NO
Sequential Time Compensation - NO NO
Sequential Time h:mm 00:30 00:30
Gross UF Rate vs. Blood Flow
Alarm Limit - NO NO
Alarm Limit Ratio % 40 40
Warning Limit - YES YES
Warning Limit Ratio % 30 30
PFV Test cyclical interval h:mm 01:00 01:00
Gross UF Rate vs. Blood Flow - active active
UF Profile Editor % 30 30

Disinfection Parameters
Disinfection after each Therapy - NO YES
Automatic Preparation Start after Disinfection - NO NO
Auto Off during Idle Rinsing - NO NO
Max Idle Time h:mm 00:30 00:30
Auto Disinfection Start - NO NO
Disinfection Method - Citric Acid 50 % Citric Acid 50 %
Thermal Disinfection - NO NO
Citric Thermal Disinfection - YES YES
Chemical 1 - NO NO
Chemical 2 - NO NO
Rinsing Water Inlet - YES YES
Descaling - NO NO
Central Manual Chemical Disinfection - NO NO
Central Automatic Chemical Disinfection - NO NO
Central Thermal Disinfection - NO NO
Central Rinsing - NO NO
Maximum Out of Action Time Hours 48 48
Monitoring - YES YES
Maximum Interval Between Chemical Disinfection - YES YES
Maximum Interval Time Hours 12 12

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 12/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

Unit HD HDF Online Modified


Disinfection Configuration Data (Thermal, Citric Thermal)
o
Thermal C 86.0 86.0
Citric Thermal Disinfection
Disinfectant - Citric Acid 50 % Citric Acid 50 %
Volume ml 120 120
Reaction Time h:mm 00:15 00:15
o
Oper. Temp. C 83 83
Cond. Max. mS/cm 4.5 4.5
Cond. Min. mS/cm 2.5 2.5
Rinsing Volume liter 4 4
Disinfection Configuration Data (Chemical 1, Chemical 2)
Chemical 1
Disinfectant -
Volume ml 0 0
Reaction Time h:mm 00:00 00:00
o
Oper. Temp. C 10 10
Cond. Max. mS/cm 0 0
Cond. Min. mS/cm 0 0
Rinsing Volume liter 0 0
Chemical 2
Disinfectant -
Volume ml 0 0
Reaction Time h:mm 00:00 00:00
o
Oper. Temp. C 10 10
Cond. Max. mS/cm 0 0
Cond. Min. mS/cm 0 0
Rinsing Volume liter 0 0
Disinfection Configuration Data (Rinsing Descaling)
Rinsing
Inlet ml/min 700 700
Time h:mm 00:10 00:10
Descaling
Descaling Time h:mm 00:05 00:05
o
Descaling Temperature C 40 40
Disinfection Configuration Data (Central Thermal, Central Chemical, Rinsing)
Central Thermal
Inlet Flow ml/min 250 250
Time h:min 00:30 00:30
o
Temperature C 86 86
Central Manual Chemical
Inlet Flow ml/min 250 250
Inlet Time h:min 00:05 00:05
Rinsing Flow ml/min 700 700
Rinsing Time h:min 00:30 00:30
Central Automatic Chemical
Inlet Volume ml 260 260
Switch off for Retention without Auto-Switch On - NO NO
Retention Time h:min 00:20 00:20
Rinsing Flow ml/min 800 800
Rinsing Time h:min 00:30 00:30
Central Rinsing
Inlet Flow ml/min 700 700
Time h:min 00:02 00:02

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 13/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

Unit HD HDF Online Modified


Filter Parameters
Filter - Diacap-Ultra Diacap-Ultra
Therapy Numbers - 150 150
Filter Operation Time h 900 900
Online Features Blocked due to Expired DF-Filter - NO NO

Adimea Parameters
Default Target Kt/V - 1.20 1.20
Kt/V Correction Mode - Single Pool Single Pool
Default Setting for Treatment Mode
Warning of Kt/V Target NO NO
Warning for Saving Kt/V Table - NO NO

HDF/HF Online Parameters


HDF Substitution Flow ml/min - 60
HF Substitution Flow ml/min - 100
HDF Substitution Volume liter - 14.4
HF Substitution Volume liter - 24
Dialysate Flow for Postdilution ml/min - 600
Dialysate Flow for Predilution ml/min - 700
Post-/Predilution Selection - - Post

Language Selection
English - active active
German - inactive inactive

ABPM
Auto Start of Cyclic Reading (from Patient - NO NO
Card/Nexadia)
Stop Cycle Reading at Begin of Disinfection - YES YES
Taking Over Limits from Patient’s Parameters - YES YES
ABPM Reading During Cuff Inflation - YES YES
ABPM Measurement Interval h:mm 00:30 00:30

HCT Parameters
pO2 Sat. Min. Limit % 80 80
Delta BV slope Limit %/h -10 -10
HCT Alarms/Warnings - YES YES

Summer Time Setting


Begin of Summer Time D/M/Y 25.03.2018 25.03.2018
D/M/Y 31.03.2019 31.03.2019
D/M/Y 29.03.2020 29.03.2020
D/M/Y 28.03.2021 28.03.2021
D/M/Y 27.03.2022 27.03.2022
h:min 2:00 2:00
End of Summer Time D/M/Y 28.10.2018 28.10.2018
D/M/Y 27.10.2019 27.10.2019
D/M/Y 25.10.2020 25.10.2020
D/M/Y 31.10.2021 31.10.2021
D/M/Y 30.10.2022 30.10.2022
h:min 3:00 3:00
Time Difference h:min 1:00 1:00
(set the time difference to 0:00 if daylight saving time
is not required)

bioLogic Fusion
Suggested SYS Lower Limit - NO NO
bioLogic mode selection - OFF OFF
TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 14/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

Unit HD HDF Online Modified


Additional Parameters
Click Sound for Confirming Keys - NO NO
Warning for Saving Patient Data - YES YES
Start Automatic Preparation - NO NO
If YES: HD-DN or HD-SN or HDF-DN - - -
End of Therapy Sound Duration - 100 % 100 %
Clock Screen Appearance - NO NO
Activation After min - -
Cube Rotation Length ms 400 400
Customize prescription screen
Conductivity - Activated Activated
Bic. Conductivity - - -
Concentrate Type - - -
Dialysate Flow - Activated Activated
Dial. Fluid. Temp. - Activated Activated
Heparin Stop Time - Activated Activated
UF Rate - Activated Activated
HDF Subst. Flow - - -
Subst. Volume - - -
Heparin Rate - Activated Activated
Customize Protocol Screen
Total Blood Volume - Activated Activated
Total Subst. Volume - Activated Activated
PBE - Activated Activated
Kt/V - Activated Activated
Total Inf. Volume - Activated Activated
Elapsed Ther. Time - Activated Activated
UF Volume - Activated Activated
UF Rate - Activated Activated
Heparin Rate - Activated Activated
pO2 Saturation - Activated Activated
Heparin Volume - Activated Activated
Conductivity - Activated Activated
BIC Conductivity - Activated Activated
∆ BV - Activated Activated
Dial. Fluid Temp. - Activated Activated
Nurse Timer Free Text Editing
Check Venous Pressure - Activated Activated
Check Arterial Pressure - Activated Activated
Check Temperature - Activated Activated
Check Filter Quality - Activated Activated
Set Bolus Volume Value - Activated Activated
- - Not activated Not activated
Show PBE instead of TMP as bar - YES YES
User Setup logout time min 15 15
PIN Change for User Setup -
New PIN - **** ****
Repeat PIN - **** ****
- Confirm PIN Confirm PIN
Skip Self Tests (permitted for servicing only!) - NO NO
Manual Bypass Confirmation - NO NO
Intensive Care Unit Mode - YES YES

Preventive Maintenance
Warning for Preventive Maintenance - NO NO
None - active active
Load Interval - inactive inactive
Elapsed Month - inactive inactive
Elapsed Working Time Counter WTC - inactive inactive
TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 15/16
Dialog iQ Software 1.02.00 Technical Information 08/2017

Unit HD HDF Online Modified


DNI Parameters
Pre-SN String is Needed d YES YES
SN String is Needed 123456 YES YES
Post-SN String is Needed - NO NO
Network Name (max. 15 Chars.) - d123456 d123456
Ethernet Interface Enabled - NO NO
DHCP enabled - YES YES
Network Speed Limit to 10 Mbit/sec - YES YES
Server IP Address - 192.168.1.101 192.168.1.101

R es t o re M an u f ac t u r er ' s Se t t i n gs
Warning: All configuration data – including saved data – will be lost, if restore is chosen. Every parameter has to be set up again.

L o w L e ve l A la r m L im it s
The values in the table have to be checked and confirmed after Low Level settings were changed
Actual Value Desired
Controller Supervisor Settings
PV Min. Limit 20 20 20
PA Min. Limit -300 -300 -300
BICLF Ratio 27 27 27
ENDLF Ratio 29 29 29
BICLF Cart Ratio 35 35 35
Intensive Care Unit Mode - YES YES
CPIT Time Interval - 01:00 01:00

P r o d u ct io n R e p o r t
Features/Options
ABPM Active Active
Kt/V inactive inactive
Adimea Active Active
HCT Hematocrit 1) 1)
Low Level Options:
Waste Port 1) 1)
HDF Online 1) 1)
Battery Active Active
Central Concentrate Supply 1) 1)
DNI-WAN 1) 1)

DNI-Nexadia 1) 1)
1) = selected if feature or option is installed

TI_2017_08_Dialog iQ_Software 1.02.00_EN.doc/pdf Art. No. 38910428 Rev. 1.00.00/November 2017 16/16

You might also like