Professional Documents
Culture Documents
Inter Laboratory Comparison of SiO2 Anal
Inter Laboratory Comparison of SiO2 Anal
Geothermics
journal homepage: www.elsevier.com/locate/geothermics
a r t i c l e i n f o a b s t r a c t
Article history: The results of an inter-laboratory comparison of silica analysis of seven water samples (coded as
Received 27 July 2011 IIE19–IIE25) are presented. The first six samples (IIE19–IIE24) were prepared by diluting a certain amount
Accepted 18 June 2012 of silicon (Si) standard 1000 ± 0.1% ppm, HYCEL, Mexico, in distilled water, while the seventh sample IIE25
Available online 20 July 2012
is natural geothermal water. The conventional true values for the silica (SiO2 ) concentration of IIE19–IIE24
are 61.1, 213.9, 366.7, 520, 825, and 978 ppm, respectively.
Keywords:
The present analysis database consists of 27 datasets, analyzed by colorimetry (13 sets), atomic
Geothermal water
absorption spectrometry (7 sets) and inductively coupled plasma (7 sets) techniques. The analyses were
Silica
Inter-laboratory comparison
performed with two procedures: (i) individual laboratory procedure (or conventional procedure) and
Atomic absorption spectrometry (AAS) (ii) dilution of samples in the range 50–200 ppm SiO2 . The mean (x̄) and standard deviation (s) calcu-
Colorimetry lation after eliminating extreme values (outliers) with x̄ ± 2s method was used for the statistical data
Inductively coupled plasma (ICP) treatment. Presently, the overall error in the analysis is at least ±10%. A comparison of analytical error
and uncertainty for individual analytical technique suggests that the inductively coupled plasma (ICP)
and colorimetry are relatively better techniques for the analysis of SiO2 in geothermal waters; however,
there are some statistical artifacts on the colorimetric results because their number of measurements
was double that of atomic absorption spectrometry (AAS) and ICP. The analysis quality (i.e., precision and
accuracy) may be improved by using the data control procedure with some high quality standard (e.g.,
NIST standards). One of the factors of systematic error is associated with the preparation of laboratory
standards.
© 2012 Elsevier Ltd. All rights reserved.
Table 1
Characteristics of samples distributed during the inter-laboratory comparison programs (modified after Verma, 2008).
IAEA01 22 Sample 001 from Giggenbach et al. (1992). A clear colorless boiling (97.5 ◦ C) spring
at Waikite. The sample was collected on 31-12-1985.
IAEA02 22 Sample 002 from Giggenbach et al. (1992). Separated water from the weir box of
well WK66 at Wairakei collected on 30-12-1985.
IAEA03 15 Sample RASINT 1 from Gerardo-Abaya et al. (1998). Cold neutral pH bicarbonate,
low Cl− spring water from China; collected by the East China Geological Institute.
IAEA04 15 Sample RASINT 2 from Gerardo-Abaya et al. (1998). High temperature neutral pH,
Cl− geothermal well water from Palinpinon geothermal field, Philippines collected
by PNOC-EDC.
IAEA05 26 Sample WILC-01 from Alvis-Isidro et al. (1999). High salinity water from a boiling
spring (94 ◦ C) at Suoh, Lampung in South Sumatera, Indonesia; collected by
CAIR-BATAN.
IAEA06 26 Sample WILC-02 from Alvis-Isidro et al. (1999). Water from a low salinity warm
spring in San Kamphaeng, Thailand; collected by the Department of Geological
Sciences, ChiangMai University and the Electricity Generating Authority of
Thailand.
IAEA07 35 Sample IGWC-01 from Alvis-Isidro et al. (2000). Low salinity water from a cold
spring (26 ◦ C) situated in a geothermal area in Southern Luzon, Philippines;
collected by PNOC-EDC.
IAEA08 35 Sample IGWC-02 from Alvis-Isidro et al. (2000). High salinity geothermal water
from a high-temperature production well at Leyte, Philippines; collected by
PNOC-EDC.
IAEA09 38 Sample GWIN-01 from Alvis-Isidro et al. (2002). High salinity water form an
injection pipeline in the PGI-Tiwi geothermal field, Philippines; collected by
PNOC-EDC.
IAEA10 38 Sample GWIN-02 from Alvis-Isidro et al. (2002). Low salinity water from a cold
spring (26 ◦ C) in the province of Laguna, Philippines; collected by PNOC-EDC.
IAEA11 38 Sample GWIN-03 from Alvis-Isidro et al. (2002). Medium salinity water from a
warm spring (60 ◦ C) in Bacon-Manito geothermal field, Philippines; collected by
PNOC-EDC.
IAEA12 31 Sample GW-03-01 from Urbino and Pang (2004). A mixer of 44% GW-03-02 and
56% GW-03-03.
IAEA13 31 Sample GW-03-02 from Urbino and Pang (2004). Medium salinity water from a
geothermal well in the Leyte geothermal field, Philippines; collected by PNOC-EDC.
IIE15–IIE18 4 Samples prepared from commercial standard of SiO2 (M.P. Verma et al., 2002).
IIE19–IIE24 22 Samples prepared from commercial standard of SiO2 (this study).
IIE25 22 A geothermal natural water sample (this study).
σ b=x-µ dure of some laboratories means to highly dilute the samples before
s
the analysis (for more information see Appendix A). For examples,
the dataset 20 was analyzed by diluting first the samples in the
Experimental error (Err)
Random error (b)
range 10–30 ppm Si (see Appendix A). Thus, in the second analyti-
cal procedure, the dilution of samples in the range 50–200 ppm SiO2
was designed to study the systematic error of individual laboratory,
and the comparison of SiO2 concentration among the participat-
ing laboratories when the samples were analyzed with a similar
X
procedure.
Fig. 1. Illustration of the significance of systematic (or bias) and random (or sta- The dilution improves the quality of analysis; but there may
tistical) errors and precision of measurements. The dashed histogram represents a be negative effects on high dilution (M.P. Verma et al., 2002). It
frequency distribution of N measurements with mean x̄ and standard deviation s. If is difficult to define the optimum range. In this work, we consid-
the number of measurements is very large, the frequency distribution results in a
ered 50–200 ppm, but the appropriate dilution range is still being
Gaussian curve with mean and standard deviation . Let the true value be T. The
bias error (a = − T) is associated with faulty calibration of equipment and incorrect sought. To highlight this point, there is need to perform new inter-
procedure of measurement. The statistical error (b = x̄ − ) reduces to zero for the laboratory comparison by diluting each sample in a specific and
population distribution. narrow range in all the participating laboratories.
36 M.P. Verma et al. / Geothermics 44 (2012) 33–42
Table 2
Concentration of SiO2 (ppm) for the samples analyzed with individual laboratory procedures and predefined dilution procedure by participating laboratories.
IIE19 IIE20 IIE21 IIE22 IIE23 IIE24 IIE25 IIE19 IIE20 IIE21 IIE22 IIE23 IIE24 IIE25
1 61 225 387 541 875 966 92 64 226 381 547 853 979 94
2 60 208 303 417b 596 705 74 52 211 319 437 753 823 79
3 64 226 387 546 877 959 96 64 227 390 549 863 973 96
4 67 232 395 568 895 1003 99 71 218 377 564 871 981 103
5 64 225 393 561 876 971 95 67 223 388 558 845 962 98
6 109 233 402 560 888 980 155 66 217 409 530 847 928 104
7 65 225 389 589 830 837 99
8 76 231 426 607 997 1097 114
9 67 233 409 569 907 1006 100
10 65 218 350 522 775 906 96
11 68 236 404 576 912 1035 101
12 47 169 289 425 628 747 69
13 66 258 483 624 838 951 82 65 265 435 603 1062a 930 111
14 122 254 465 797 1065 1118 154 78 230 426 519 863 787 98
15 115 252 465 588 1074 1190 159 95 258 476 581 852 967 116
16 63 233 386 539 864 949 96 64 226 383 544 836 961 98
17 65 233 385 562 946 1031 106 62 203 328 499 872 939 92
18 74 242 369 509 777 701 108 78 252 423 621 1013 950 109
19 60 197 296 708 628 801 88
20 44 205 363 556 871 1079 78 66 189 282 456 688 797 114
21 63 226 382 544 850 962 101 75 231 387 552 876 968 94
22 58 233 396 562 893 1008 72 63 223 373 577 897 1042 76
23 68 242 416 622 966 1059 101 69 243 429 601 952 1056 101
24 81 277 465 638 947 1040 92 72 237 409 610 957 1020 82
25 76 270 447 624 948 1017 76 75 249 415 959 951 80
26 79 277 457 641 947 1033 87 73 250 417 656 879 972 83
27 77 252 437 603 965 1093 99 62 212 358 518 809 880 85
a
Values (marked in bold and underlined) were remove before the statistical analysis.
b
Values (marked in italics and underlined) were remove after the raw data statistics (Verma, in press-a).
c
Sets 1–13 were analyzed by colorimetry, sets 14–20 were analyzed by AAS (atomic absorption spectrometry), and sets 21–27 were analyzed by ICP (inductively coupled
plasma).
4. Statistical data treatment method statistics), (ii) mean (x̄) and standard deviation (s) calculation after
eliminating outliers with x̄ ± 2s, (iii) median and median absolute
The statistical terms involved in the analytical data analysis are difference (MAD), (iv) Huber method, and (v) single-outlier type
well defined in textbooks (Bevington, 1969; Barnett and Lewis, multiple test method involving Dixon, Grubbs, skewness and kur-
1994). Fig. 1 shows a relation between the measured values of a tosis tests at 99% confidence level. The results must be reported
parameter X and its frequency distribution of measurement. The in scientific notation with one significant digit of uncertainty. The
dashed histogram represents a frequency distribution of N mea- statistical data treatment provides a procedure to estimate the pre-
surements of X, with mean x̄ and standard deviation s. If the number cision of measurement.
of measurements is very large, the frequency distribution results The analysis of commercial reference materials (standards) in
in a Gaussian curve with mean and standard deviation , as all the laboratories together with the water samples is fundamen-
shown by solid curve. Additionally, any imperfection of measure- tal to improve the analytical quality (i.e., precision and accuracy of
ment technique can produce some skewness in the distribution measurement) of the participating laboratories (Parr and Clements,
curve, which is not dealt with here. 1991; Araguás and Rozanski, 1995; Lippmann et al., 1999; Verma,
The types of error and uncertainty in the experimental data are 2004). In this study, the data analysis was performed using the
also explained in Fig. 1. Let a true value of X be T. The experimental method, mean (x̄) and standard deviation (s) after eliminating out-
error is the sum of (i) systematic errors or bias and (ii) random liers with x̄ ± 2s. Thus, the confidence interval is 65.2% instead of
or statistical errors. The bias error (a = − T) is associated with 68.3% (Verma, in press-a).
faulty calibration of equipment and incorrect procedure of mea-
surement. If the measurement technique is correct, the systematic 5. Results and discussion
error (a) will be zero. The statistical error is b(= x̄ − ). The range
(x̄ − s) to (x̄ + s) is considered as analytical uncertainty for the sam- Table 2 presents the results of SiO2 measurements in the water
ple distribution (N measurements), and is reported here as ±1 s. samples, distributed during the present inter-laboratory compar-
The analytical uncertainty is associated with the detection limit ison. The datasets 1–13, 14–20 and 21–27 were analyzed by
of equipment and statistical fluctuation in the measurements. The colorimetry (using the Molybdo-silicate method), atomic absorp-
uncertainty is considered as statistical fluctuation, if the value of tion spectrometry (AAS), and inductively coupled plasma (ICP),
detection limit is relatively low and vice versa. respectively. Fig. 2 shows the data together with the compiled
The objective of a statistical data treatment method is to get the data from the literature on the SiO2 inter-laboratory comparisons.
information of population distribution ( and ) from the sample The data correspond to the measured values of the participating
distribution (x̄ and s). Verma (in press-a) presented the statistical laboratories using their own analytical procedure. The data are clas-
data treatments for the detection and elimination of extremums sified according the measurement techniques: colorimetry, atomic
(outliers) using five methods: (i) data visualization (or raw data absorption spectrometry (AAS), inductively coupled plasma (ICP),
M.P. Verma et al. / Geothermics 44 (2012) 33–42 37
2500
IAEA05, 1999
IAEA12, 2004
IAEA08, 2000
IAEA09, 2002
IAEA11, 2002
IAEA07, 2000
IAEA03, 1998
IAEA04, 1998
IAEA10, 2002
IAEA02, 1992
IAEA01, 1992
IAEA06, 1999
IAEA13, 2004
IIE17, 2002
IIE15, 2002
IIE18, 2002
IIE16, 2002
2000
SiO2 (ppm)
1500
1000
500
a
0
0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20
Lab No.
2500
b
2000
IIE20
IIE19
IIE25
IIE22
IIE23
IIE21
IIE24
SiO2 (ppm)
1500
Colorimetry
AAS
1000 ICP
SP
Not defined
500
0
0 20 0 20 0 20 0 20 0 20 0 20 0 20
Lab No.
Fig. 2. Relations between laboratory number and SiO2 concentration, analyzed as parts of inter-laboratory comparisons. (a) Previous studies. The first 13 samples
(IAEA01–IAEA13) were compiled from the IAEA internal reports (Verma, 2008), and the last four samples (IIE15–IIE19) are from a study conducted among Mexican geo-
chemistry laboratories (M.P. Verma et al., 2002). (b) This study. The results obtained with the analytical procedure of individual laboratory are grouped according to analytical
techniques: colorimetry, atomic absorption spectrometry (AAS), inductively coupled plasma (ICP), UV-spectrophotometry (UV-SP) and undefined method.
UV-spectrophotometry (SP), and undefined method. In the present 213.93 ppm SiO2 (100 ppm Si) was prepared by diluting 0.1 ml of
study (Fig. 2b), there is also high dispersion in the measured val- 10,000 ppm Si standard in 10 ml of distilled water. Any error in
ues for high SiO2 concentration samples as reported in the earlier measuring the standard volume will propagate in the concentra-
studies (Ellis, 1976; Giggenbach et al., 1992). However, there is rel- tion value of laboratory standard. Thus, it is recommended to use
atively less dispersion among the ICP values, even in the earlier in general the 1000 ppm Si commercial standard for the SiO2 anal-
measurements. This suggests that ICP is a better technique for SiO2 ysis of geothermal waters and prepare large volume of laboratory
analysis in geothermal water samples. standard. However, there are some datasets which have both sys-
Fig. 3 shows an XY-plot between the conventional true value tematic and random errors (uncertainties). The reasons are still not
and the measured value of SiO2 . The samples are grouped accord- clear.
ing to the analytical technique and the measurement procedure. For Table 3 shows the results after the statistical data treatment
example, Fig. 3a shows the behavior of values obtained by colorime- of values reported in Table 2. Prior to statistically analyzing the
try using the individual laboratory procedure, while the behavior results, values that were out of trend were removed. Actually, there
of values obtained with the dilution procedure is shown in Fig. 3b. is only one value (1062 ppm). It is marked in bold and underlined
Similarly, Fig. 3c and d correspond to AAS, and Fig. 3e and f are for in Table 2 and shown with a dashed circle in Fig. 3. According to
ICP. the data statistics, the removal of this value will not affect the con-
There are some laboratories which have the same type of fidence limits. These types of values are associated with a variety
systematic error in the measured values for both the analyti- of errors, such as misprints or mistakes made during the handling
cal procedures. For example, the data points corresponding to and analysis of the sample, or passing the information (data) from
set02 are systematically on the lower side of the 45◦ line in person to person.
Fig. 3a and b. If we look at worksheet set02 in the Excel work- The arithmetic mean (x̄) and standard deviation (S.D.) of the
book Interlab2012.xlsx (Appendix A), the standards were prepared remaining values were then computed. The values that fell outside
by dissolving 10,000 ppm Si standard. The laboratory standard the mean ± 2 S.D. were rejected. The values are marked in italics
38 M.P. Verma et al. / Geothermics 44 (2012) 33–42
Fig. 3. Relationship between the “conventional” true value and measured silica concentrations. (a, c, and e) The behavior of measured silica concentration obtained by
colorimetry, AAS and ICP using the individual laboratory procedure, respectively. (b, d, and f) The behavior of respective data obtained by the diluting procedure. The dashed
45◦ line represents the data analytical points which are in agreement with the corresponding true value. The data marked with dashed circles were excluded before conducting
the statistical data treatment analysis.
and underlined in Table 2. Then, the mean and S.D. of the remaining double the number of points and there is a high spread in the
data were recomputed. For example, the colorimetric dataset for values of AAS. Any statistical data treatment will be in favor of
sample IIE19 has the values 61, 60, 64, 67, 64, 109, 65, 76, 67, 65, colorimetric dataset. Thus, it has some artifacts due to statistical
68, 47, and 66 ppm for corresponding laboratories, respectively. data treatment; however, the colorimetric mean value for each
The mean and standard deviation are 68 and 14 ppm, respectively. sample is relatively close to the corresponding conventional true
Clearly the 109 ppm is out of trend (i.e. mean ± 2 S.D.). According to value.
the data statistics, the removal of values which are out of mean ± 2 Fig. 4 shows histograms to compare the analytical data quality
S.D. (i.e., 109 ppm) will not affect significantly the confidence for the conventional laboratory method and the dilution procedure
interval. The same methodology was applied to all the data sets; implemented in this study. It can be observed that the values
but the colorimetric measurements have only outliers. This may be measured with the dilution procedure have significantly low
associated with the fact that the colorimetric dataset has relatively uncertainty. However, the differences in the uncertainties
M.P. Verma et al. / Geothermics 44 (2012) 33–42 39
Table 3
Results of statistical data treatment of SiO2 analysis (ppm).
are not as pronounced as observed by M.P. Verma et al. curve. The data are shown below each curve. There were only few
(2002), since the conventional analytical procedure of many outliers since the dispersion among values is relatively large.
laboratories consists of the dilution of samples prior to analysis. Table 3 presents the analysis of uncertainty (C.V.) and error
The binomial distribution curve for each sample is shown in for each analytical technique. The ranges of uncertainty (in %)
Fig. 5. The analytical uncertainties for the colorimetry and ICP tech- using the analytical procedure of individual laboratory are 9–21,
niques are approximately the same, while the AAS values have 8–38, and 9–17 for colorimetry, AAS, and ICP, respectively; while
higher uncertainty. All the measured values are shown on the con- the respective ranges for the dilution method are 5–10, 9–18,
centration axis for each sample. The data after the statistical data 6–8. Similarly, the ranges of analytical error (in %) in the same
treatment were used in the derivation of the binomial distribution order are 4–11, 0–27, and 5–25 for the analytical procedure of
40 M.P. Verma et al. / Geothermics 44 (2012) 33–42
Fig. 4. Comparison of SiO2 analysis data quality for the individual conventional laboratory and dilution methods.
individual laboratory and 4–6, 8–21, and 1–14 for the dilution from the NIST commercial standard as samples. If the values are
method. As mentioned above, the effect of dilution on the data anal- within ±0.5%, the calibration is appropriate. Otherwise, both the
ysis quality is not possible to study in the present work; however, steps were repeated. There is discussion among the participants to
there is relatively low uncertainty in the case of dilution proce- initiate a common undertaking to create a standard data control
dure for each analytical technique. In other words, there is better procedure, similar to the SMOW-GISP-SLAP scale used for the
agreement if the samples were analyzed in a similar manner. The isotopic analysis (oxygen-18 and deuterium) of water sample (Parr
analytical accuracy is also better for the dilution method for each and Clements, 1991; Araguás and Rozanski, 1995; Lippmann et al.,
analytical technique. It suggests that the analysis of SiO2 must be 1999). In other words, there is need to establish some common
performed in certain range of dilution. However, the overall ana- standards (e.g., NIST standard) for each chemical parameter in
lytical error in the analysis of SiO2 is at least 10%. The suggestions order to obtain the consistency among the measured values in
presented in this study may provide a mechanism to improve in different laboratories around the world. Indeed, the mechanism
future the quality of SiO2 analysis in water samples. of quality control for the international standards is of primary
The mean value for colorimetry is generally close to the con- concern.
ventional true value. It may be assessed that colorimetry is a better Verma (in press-b) presented an algorithm for quartz geother-
technique for the analysis of SiO2 in water samples. However, there mometry to propagate the analytical uncertainty in the mea-
are two points to be emphasized. First, the number of values for surements of the silica concentration and reservoir enthalpy. He
colorimetry is almost twice as those for AAS and ICP. Statistically, elucidated that there is an uncertainty of ±18 ◦ C in the calcu-
the random error and uncertainty reduce with increasing the lated geothermal reservoir temperature with quartz solubility
number of measurements. Second, there were some laboratories geothermometersin case of Los Azufres geothermal system, if the
using colorimetry (see the datasets, set01, and set03 in Appendix uncertainties in the measured values of SiO2 and enthalpy are
A) which conducted the data control procedure for their calibra- 10% and 2%, respectively. Thus, the prediction of changes in the
tion. Presently, the data control procedure consists of two steps: reservoir characteristics during the exploitation of geothermal sys-
(i) the calibration of equipment with laboratory standards, and (ii) tems demands good quality geochemical analysis of geothermal
measuring the SiO2 concentrations of some standards, prepared waters.
M.P. Verma et al. / Geothermics 44 (2012) 33–42 41
0.10
0.09 IIE19 IIE19 IIE23 IIE23
0.01
0.08
0.07
0.06
0.05 Lab method Dilution method Lab method Dilution method
0.04
0.03
0.02
0.01
0.00 0.00
0 50 100 150 0 50 100 150 500 700 900 1100 1300 500 700 900 1100 1300
0.07
IIE20 IIE20 IIE24 IIE24
0.06 0.01
0.05
Frequency
0.04
0.03
0.02
0.01
Frequency
0.00 0.00
100 150 200 250 300 350 100 150 200 250 300 350 500 700 900 1100 1300 1500 500 700 900 1100 1300 1500
0.02 0.06
IIE21 IIE21 IIE25 IIE25
0.05
0.04
0.01 0.03 True value
Colorimetry
0.02 AAS
0.01 ICP
0.00 0.00
200 300 400 500 600 200 300 400 500 600 0 100 200 0 100 200
IIE22 IIE22 SiO2 (ppm)
0.01
True value
Colorimetry
AAS
ICP
0.00
300 500 700 900 300 500 700 900
SiO2 (ppm)
Fig. 5. Binomial distribution of the measured SiO2 concentration values for each sample. The raw values are shown on the
concentration axis. The values used to draw the binomial distribution are shown below the curve. There were only few extreme values (outliers). The conventional
true value for each sample is shown by the solid vertical line. The analytical error is the difference between the mean (vertical line) for each distribution and corresponding
true value.
6. Conclusions Acknowledgements
From the results of the present inter-laboratory comparison, it Authors appreciate the IIE authorities for the financial support
can be concluded that there is, in general, high analytical uncer- and the authorities of each participating laboratory. Victor Arellano
tainty (error) in the measurement of SiO2 in water samples. The G. encouraged for conducting this project. Many people, especially
overall analytical error (uncertainty) is presently of ±10%. However, A. Aguayo, O. Cruz, and N. Ceniceros helped in the analytical work.
some laboratories have good quality control for each analytical The paper is substantially improved in its content and presentation
technique (e.g., for colorimetry set01 and set03, for AAS set16, and with the constructive comments and suggestions of the anonymous
for ICP set 21). Colorimetry and ICP are relatively better techniques reviewers.
for the analysis of SiO2 . In the present study, the number of values
for colorimetry is almost twice those for AAS and ICP. This makes Appendix A. Supplementary data
it difficult to define the quality of colorimetric measurements with
any certainty. However, analytical error in the colorimetric mea- A.1. Analytical results
surements is relatively small.
Using commercial standards as water samples in this inter- The analytical data of each participating laboratory are pre-
laboratory comparison has highlighted some clues to improve the sented in a worksheet of excel workbook, Interlab2012.xlsx. The
analytical quality. First, there is a need to take special care dur- results of each laboratory are divided in four groups: general infor-
ing the preparation of laboratory standards. Second, the calibration mation, laboratory method, dilution method, and comparison of data.
curve must be verified with the data control procedure (i.e., run- The data of some laboratory are reported as the final results only.
ning some quality standards as water samples). In other words, Supplementary data associated with this article can be
there is need to establish good quality standards for each chemical found, in the online version, at http://dx.doi.org/10.1016/
parameter worldwide. j.geothermics.2012.06.003.
The inter-laboratory comparison program should be continued
to ensure a high analytical quality laboratories worldwide. Sim- References
ilarly, the geochemical modeling of hydrothermal systems must
Alvis-Isidro, R., Urbino, G.A., Gerardo-Abaya, J., 1999. 1999 Inter-laboratory Compar-
incorporate the propagation of analytical uncertainty (error) in
ison of Geothermal Water Chemistry Under IAEA Regional Project RAS/8/075.
order to provide reliable conclusions from each particular study. Report. IAEA, Vienna, 39 pp.
42 M.P. Verma et al. / Geothermics 44 (2012) 33–42
Alvis-Isidro, R., Urbino, G.A., Pang, Z., 2000. Results of the 2000 IAEA Inter-laboratory Lim, P.G., 2005. Total silica analysis using a double beam atomic absorption spec-
Comparison of Geothermal Water Chemistry. Report. IAEA, Vienna, 40 pp. trophotometer. In: Proceedings of World Geothermal Congress 2005, Antalya,
Alvis-Isidro, R., Urbino, G.A., Pang, Z., 2002. 2001 Inter-laboratory Comparison of Turkey.
Geothermal Water Chemistry. Report. IAEA, Vienna, 55 pp. Lippmann, J., Gröning, M., Rozanski, K., 1999. 2nd Interlaboratory Comparison for
André, L., Spycher, N., Xu, T., Vuataz, F.D., Pruess, K., 2006. Modeling brine–rock Deuterium and Oxygen-18 Analysis of Water Samples. Report. IAEA, Vienna, 32
interactions in an enhanced geothermal system deep fractured reservoir at pp.
Soultzsous-Forets (France). A joint approach using two geochemical codes Mahon, W.A.J., 1966. Silica in hot water discharged from drillholes at Wairakei, New
Frachem and ToughReact. Report LBNL-62357, 71 pp. Zealand. New Zealand Journal of Science 9, 135–144.
Araguás, L., Rozanski, K., 1995. IAEA Inter-laboratory Comparison for Deuterium and Mazor, E., Cioni, R., Carrazza, E., Fratta, M., Magro, G., Matsuo, S., Hirabayashi, J.,
Oxygen-18 Analysis of Precipitation Samples. Report. IAEA, Vienna, 33 pp. Shinohara, H., Martín, M., Piccardi, G., Gellini Legittimo, P., 1988. Evolution of
Arnórsson, S., 1970. Underground temperatures in hydrothermal areas in Iceland as fumarolic gases-boundary conditions set by measured parameters: case study
deduced from the silica content of the thermal water. Geothermics 2, 536–541 at Vulcano, Italy. Bulletin of Volcanology 50, 71–85.
(spec. issue). Morey, G.W., Fournier, R.O., Rowe, J.J., 1962. The solubility of quartz in water in the
Arnórsson, S., 2000. The quartz and Na/K geothermometers. I. New thermodynamic temperature interval from 29 to 300 ◦ C. Geochimica et Cosmochimica Acta 26,
calibration. In: Proceedings of World Geothermal Congress 2000, pp. 929–934. 1029–1043.
Barnett, V., Lewis, T., 1994. Outliers in Statistical Data, third edition. John Wiley, Parr, R.M., Clements, S.A., 1991. Inter-comparison of Enriched Stable Isotope Ref-
Chichester, 584 pp. erence Materials for Medical and Biological Studies. Report. IAEA, Vienna, 31
Bevington, P.R., 1969. Data Reduction and Error Analysis for the Physical Sciences. pp.
McGraw Hill, New York, 336 pp. Reed, M., Mariner, R.H.,1991. Quality control of chemical and isotopic analysis of
Bödvarsson, G., Pálmason, G., 1961. Exploration of subsurface temperatures in geothermal waters. In: Proc. XVI Workshop on Geothermal Reservoir Engineer-
Iceland. In: Proceedings of U.N. Conf. New Sources Energy, G/24, Rome, pp. ing. Stanford University, Stanford, CA, USA, pp. 9–13.
82–90. Sugita, H., Matsunaga, I., Yamaguchi, T., 2000. Silica scale prevention method
Brown, K.L., Bacon, L.G., 2000. Manufacture of silica sols from separated geothermal using seed made from geothermal brine. In: Proceedings of World Geothermal
water. In: Proceedings of World Geothermal Congress 2000, pp. 533–537. Congress 2000, pp. 2925–2930.
Corazza, E., 1986. Field workshop on volcanic gases, Vulcano (Italy), 1982. Geother- Thompson, M., Potts, P.J., Kane, J.S., Webb, P.C., Watson, J.S., 2000. GeoPT5. Interna-
mics 15, 197–200 (general report). tional proficiency test for analytical geochemistry laboratories—report on Round
Crandall, J.R., Blaine, R.L., 1959. Statistical evaluation of inter-laboratory cement 5 (August 1999). Journal of Geostandards and Geoanalysis 24, E1–E28.
tests. In: Proceedings of the Amer. Soc. Testing Materials, vol. 59, pp. 1129– Urbino, G.A., Pang, Z., 2004. 2003 Inter-laboratory Comparison of Geothermal Water
1154. Chemistry. Report. PNOC, Philippines, 51 pp.
De Bièvre, P., 1997. Measurement results without statements of reliability (uncer- Verma, M.P., 2002. Geochemical techniques in geothermal development. In:
tainty) should not be taken seriously. Accreditation Quality Assurance 2, 1–269. Chandrasekharam, D., Bundschuh, J. (Eds.), Geothermal Energy Resources for
Ellis, A.J., 1976. The IAGC inter-laboratory water analysis comparison programme. Developing Countries. Swets & Zeitlinger Publishers, The Netherlands, pp.
Geochimica et Cosmochimica Acta 40, 1359–1374. 225–251.
Fournier, R.O., Rowe, J.J., 1966. Estimation of underground temperatures from silica Verma, M.P., 2004. A revised analytical method for HCO3 − and CO3 2− determina-
content of water from hot springs and wet-steam wells. American Journal of tions in geothermal waters: an assessment of IAGC and IAEA inter-laboratory
Science 264, 685–697. comparisons. Geostandards and Geoanalytical Research 28, 391–409.
Fournier, R.O., Truesdell, A.H., 1970. Geochemical indicators of subsurface tempera- Verma, M.P., 2008. Qrtzgeotherm: an ActiveX component for the quartz solubility
ture applied to hot spring waters of Yellowstone National Park, Wyoming, USA. geothermometer. Computers & Geosciences 34, 1918–1925.
Geothermics 2, 529–535 (spec. issue). Verma, M.P., 2008b. IAGC and IAEA inter-laboratory comparisons of geothermal
Gerardo-Abaya, J., Schueszler, C., Groening, M., 1998. Results of the Inter-laboratory water chemistry: the propagation of uncertainty in the reservoir pH calculation.
Comparison for Water Chemistry in Natural Geothermal Samples Under Geostandards and Geoanalytical Research 32, 317–330.
RAS/8/075. Report. IAEA, Vienna, 18 pp. Verma, M.P. IAEA inter-laboratory comparisons of geothermal water chemistry:
Giggenbach, W.F., Goguel, R.L., Humphries, W.A., 1992. IAEA inter-laboratory critiques on analytical uncertainty and accuracy, and geothermal reservoir mod-
comparative geothermal water analysis program. Geothermal Investigations eling. Journal of Iberian Geology, in press-a.
with Isotope and Geothermal Techniques in Latin America, IAEA-TECDOC-641, Verma, M.P. QrtzGeotherm: a revised algorithm for quartz solubility geother-
Vienna, pp. 439–456. mometry to estimate geothermal reservoir temperature and vapor fraction
Giggenbach, W.F., Matsuo, S., 1991. Evaluation of results from second and third IAVCI with multivariate analytical uncertainty propagation. Computers & Geosciences,
field workshop on volcanic gases, Mt. Usu, Japan, and White Island, New Zealand. http://dx.doi.org/10.1016/j.cageo.2012.01.008, in press-b.
Applied Geochemistry 6, 125–141. Verma, M.P., Tello, E., Sandoval, F., Tovar, R., Martinez, J.L., 2002. An inter-lab cali-
Gigganbach, W.F., Tedesco, D., Sulistiyo, Y., Caprai, A., Cioni, R., Favara, R., Fisher, bration of silica for geothermal water chemistry. Geothermics 31, 677–686.
T.P., Hirabayashi, J.-I., Korzhinsky, M., Martini, M., Menyailov, I., Shinohara, H., Verma, S.P., Santoyo, E., 1997. New improved equations for Na/K, Na/Li and SiO2
2001. Evaluation of results from the fourth and fifth IAVCEI field workshop on geothermometers by outlier detection and rejection. Journal of Volcanology and
volcanic gases, Vulcano Island, Italy and Java, Indonesia. Journal of Volcanology Geothermal Research 79, 9–23.
and Geothermal Research 108, 157–172. Verma, S.P., Santoyo, E., Velasco-Tapia, F., 2002. Statistical evaluation of analytical
Iler, R., 1979. The Chemistry of Silica: Solubility, Polymerization, Colloid and Surface methods for the determination of rare-earth elements in geological materi-
Properties and Biochemistry. John Wiley & Sons, 866 pp. als and implications for detection limits. International Geology Reviews 44,
Kane, J.S., Potts, P.J., 1997. ISO guides for reference material certification and use: 287–335.
application to geochemical reference materials. Journal of Geostandards and White, D.E., Brannock, W.W., Murata, K.J., 1956. Silica in hot-spring waters. Geochim-
Geoanalysis 21, 51–58. ica et Cosmo-chimica Acta 10, 27–59.