Manual of CO2 Laser【en】
Manual of CO2 Laser【en】
This manual contains important information on the use and operation of this machine.
Please read all the information carefully for optimal performance to prevent any damage or misuse of
the machine. Any change or modification not expressly approved in this manual could void the
warranty for this equipment.
Basic Introduction
Theory
Applications
Contraindications
Advantages
Specifications
Installation Instruction
Accessories
Installation steps
Operation Instruction
Before operation
During operation
After operation
Recommended parameters
Other Information
Precautions
Warranty
Troubleshooting
Following these instructions will help you use the equipment efficiently and safely and will prolong
the machine life.
www.bestviewmedical.com -2-
Fractional technology is a benchmark for efficiency and comfort in laser resurfacing procedures. Fractional
resurfacing consists in the emission of tiny laser energy beams which penetrate into the skin in the form of
thermal columns all the way to the dermis while leaving untreated tissue in-between them. Laser energy is
deposited into the tissue with a pattern of random shots which alternate areas treated by laser with those of
intact skin. This provides for a faster recovery and tissue regeneration and is a virtually painless treatment.
These thermal micro-columns treat the dermis superficially and also in depth, reaching the underlying layers
of the dermis and stimulating collagen formation. In addition, this mechanism allows for individualized
treatments, adapting them to the needs of each patient.
This CO2 fractional laser also applies for vaginal treatment and has a good effect.
-3- www.bestviewmedical.com
Smooth scars, such as Skin renewal and Eliminate wrinkles;
repaving, especially
surgical scars, burned Skin tightening.
recovery from sun
scars, acne scars, etc. damage.
Eliminate pigments,
Acne treatment, Vaginal stretching;
such as intractable
chloasmas, age spots, Vaginal whitening;
Restore your beauty
etc. Urinary incontinence
www.bestviewmedical.com -4-
Fractional mode (aesthetic), pulse mode (surgery), Gynae mode (gynecology), 3 treatment modes can meet all
the needs of beautician, plastic surgeon, gynecologist.
South Korea imported 7-articulated arms with good quality lenses.
Germany imported Mini-Stepper-Motors with the best quality makes the laser shoot faster and the energy more
uniform and stable.
High-quality constant current and constant voltage power supply, ensuring a stable and continuous power
supply for the machine.
Variety of scan graphics optional, applying for different skin types and treatment areas
Only 10~15 minutes is needed for the whole treatment process, it is very fast.
No downtime, don’t worry about the postoperative impact.
Wavelength 10600nm
Power 1 ~ 60W
-5- www.bestviewmedical.com
Picture Description Quantity
Articulated arm 1
Fractional head 1
Clips 5
Foot pedal 1
Protective glasses 1
Protective goggles 1
Power line/Keys 3
www.bestviewmedical.com -6-
Install articulated arm:
1 2 3 4
P1 P2 P3 P4
F1 F2 F3 F4
-7- www.bestviewmedical.com
F5 F6 F7
F8
F8 G1 G2
G3
G4 G5 G6
www.bestviewmedical.com -8-
Connect foot pedal:
Insert one end of the foot pedal into the preset slot on the back of machine and then place the other end on a flat
ground.
Add water:
Get some purified water (can be drink water) around 3.5L, and change it about 1~2 months;
Don’t stop injecting until there is water flowing out from “SPILT WATER”.
-9- www.bestviewmedical.com
It is necessary that users should read the manual carefully prior to the application of this machine;
Consult whether the client has the above contraindications;
It is suggested that this machine is operated in a separated room.
No makeup on the day of surgery. If clients does this, please remove it to keep the treatment area clean.
Please do a physical examination to ensure treatment effect and prevent allergic reactions.
Disinfect the treatment heads and other therapeutic devices.
Make sure the handle and power line are properly connected into the machine.
2. Ensure good water flow. Switch on and off the machine for two minutes respectively. Then repeat this cycle for 3
times to normalize the inside water circulation.
3. Then you will see the main interface on the screen. Click “Fractional mode”, “Gynae mode” or “Pulse mode” to enter
into the corresponding treatment interfaces:
www.bestviewmedical.com - 10 -
4. Click “Fractional mode” to enter into the corresponding treatment interface:
It would be better to do treatment 1~2 sections and the interval is about 2~3 months.
- 11 - www.bestviewmedical.com
5. Click “Gynae mode” to enter into the corresponding treatment interface:
Vaginal probe 1 for inner vagina Vaginal probe 2 for uroclepsia Vaginal probe 3 for vulva
Put on protective goggles for operator and patient to protect their eyes from the laser.
Press “READY”, the machine will start.
Place the treatment head firmly on private parts and then trample the foot pedal. After scanning, release the pedals at
that time and end treatment. If you still need to continue scanning, do it again.
Treatment sessions: for the inner vagina scan 2 times, the energy of the second time is lower; for the vulva scan
2~3 times. And the interval is about 2~3 months.
www.bestviewmedical.com - 12 -
6. Click “Pulse mode” to enter into the corresponding treatment interface:
2 Pulse mode
3 Continuous mode
8 Click it to treatment
Install treatment head of pulse mode and choose suitable cutter heads.
Choose “Pulse” (point) or “Continuous” (line).
Adjust parameters from low to high depending on size and shape of treatment area and the personal feeling of
patient.
NOTE: Please refer to the next part for recommended parameters.
Put on goggles for operator and patient to protect their eyes from the laser.
Press “READY”, the machine will start.
Hold the treatment head close to the skin tissue needed to be cut, trample foot pedal and move along the
organization. After cutting, release the foot pedal.
Pulse mode is very effective, most of sarcomas can be completely removed in one operation.
7. During the treatment, the client will feel heat and a little pain.
8. After the treatment, click “Back” to return to the previous interface. And then turn off the power, clean probe and working
area, etc.
NOTE: Any abnormal phenomenon during operation, please press the urgent stop button down.
- 13 - www.bestviewmedical.com
Fractional / Pulse Mode:
After treatment, there may appear slight bleeding and scabbing which will fall off naturally in several days and
shouldn’t be torn up by hand.
It is normal and temporary that slight swelling and redness occur, with which cold compress can help. If severe
situation takes place, ask for medical advice.
Do not clean the treatment area with water within 24 hours after operation or use cosmetics that are stimulating
to skin.
Keep your hands off your treatment area in case of infection. The wound is probably bacterially infected if
there is honey-colored suppuration or scab along with painful feelings.
Patients are unable to take dark-colored food and drinks like black soybeans, chocolate, coffee, cola, etc.
Since they are not in favor of skin recovery, photosensitive food like lemon, white radish, celery, lettuce, rape,
eggplant, nori, spinach, adzuki beans and mango are prohibited.
Tomatoes, cabbage as well as strawberry can help restore the targeted area so you can use them in moderation.
Gynae Mode:
Slight redness and swelling will happen to vulva, which is normal and temporary.
Protect the targeted area from water within 12 hours after treatment.
Avoid strenuous exercises like running, rope skipping, riding a bicycle, etc.
Irritated food, dark-colored food, cigarettes and drinks should be kept away within a week.
Please keep underclothes clean and wear cotton underwear as well as skirt rather than skinny pants.
Private products that are able to restore the targeted area can be in use.
www.bestviewmedical.com - 14 -
Fractional mode
Power 15~20W
Height 20mm
Width 20mm
Density 0.1~1.0mm
Pulse duration 1.0~2.0us
Interval time 1000~2000us
Scan mode Normal
Times 1th
Circle, Triangle, Hexagon,
Shapes
Rectangle, Ellipse, Straight liner
It may take 1 or up to 2
sessions of fractional lasers to
substantially improve them.
The interval is about 2~3
months.
It may take 1 or up to 2
sessions of fractional lasers to
substantially improve them.
The interval is about 2~3
months.
- 15 - www.bestviewmedical.com
Pulse mode
Working Energy Pulse Interval
Power
Modes Density Duration Time
0~6000
Pulse 1~60W 0.1~100us 0.1~100us
J/cm2
Gynae mode
Density 0.1~1mm
Duration 5.0us
Interval 3~5us
Power 1~10W
Times 1th
Point energy 1~50mj
These parameters and pictures above are ONLY FOR REFERENCE! Those applied in practical operation should be
adjusted according to personal feeling and actual physical conditions.
www.bestviewmedical.com - 16 -
Safety about Laser
Ignitable & Explosive: Keep ignitable and explosive liquid or gas (such as alcohol, nitrous oxide and oxygen)
away from treatment room.
Ornament: Before treatment, ornaments that reflect light should always be taken off from both treatment
receiver and operator.
Warning sign: It is recommended that a warning sign be placed at the entrance to treatment room whenever the
system is in use.
Temperature: Laser has a high request for treatment room temperature, especially in summer time when it’s
best to be controlled in 25 or 26 degree.
- 17 - www.bestviewmedical.com
Put the machine on the flat and stable surface.
Check the main machine, accessory and the power cable are intact well before use.
The machine is fragile. Do not drop it.
Clean the equipment with a wet cloth but make sure no liquid penetrates.
Do not operate the machine with wet hands.
Do not bend or turn the power cable and the wires.
Before you clean or transport the machine, make sure that it is switched off and the power cable is
unplugged.
Put the equipment in a ventilated area. Do not cover the fan or the machine nor put any objects on it.
Do not use the machine in the toilet or near a microwave or oven.
Make sure that the power cable is not worn or broken before you plug it.
The operator must be trained to be familiar with the structure and performance of the equipment and
master the knowledge of the operating system.
Users under 18 cannot use the equipment.
The machine cannot be used on animals.
Do not use the machine for other purposes than those described in this manual.
The operating temperature should be in good control.
The treatment button should always be pressed by hand.
When there is an error, the system will show a message to the operator, at which time please stop any
operation.
The continuous working time of this machine shouldn’t exceed 2 hours, 8 hours/day at most.
Turn off the power after the operation. When not in use, be sure to unplug the power line from the power
outlet.
Please take good care of the equipment. The power cord and accessories should be carefully inspected
before operation. If there are any traces of damage, stop using them and replace them with a new one.
All products are strictly tested before sale. The guarantee of free repair is provided starting from the purchasing day,
during which time the maintenance for any breakdown due to non-artificial factors will be performed free of charge.
Accessories are not included in the warranty, so please contact us if they are needed.
The standard of the operation process and the environment should be in accordance with all the instructions above.
www.bestviewmedical.com - 18 -
Problems Possible causes Solutions
The buttons do not work Defective buttons or circuit damaged Contact the technical service
Nancy Lei
Nancylei@bestviewmedical.com
- 19 - www.bestviewmedical.com