Professional Documents
Culture Documents
PLC1002 - Manual - Rev 06.0
PLC1002 - Manual - Rev 06.0
PLC1002 - Manual - Rev 06.0
Introduction to
High Performance Liquid Chromatography
(HPLC)
Pharmaceutical Liquid Chromatography – PLC1002
Telephone: 416-502-2277
e-mail: info@aaps.ca
Course Objectives:
This course aims to give students the basic understanding in order to perform pharmaceutical HPLC
analysis as per the United States Pharmacopoeia. Common HPLC testing procedures, techniques,
theory, equipment types, and SOP’s are emphasized and accurately covered through step-by-step
instructions and hands-on experiments.
Using interactive exercises, students are guided through compendial HPLC fundamentals. This
course bridges the gap between the theoretical and the practicality of the pharmaceutical workplace.
Mark Allocation
Quizzes 10 %
Experimental and Documentation 40 %
Final Exam 50 %
Class Participation
Due to the practical and participative nature of the learning environment, punctual and regular
attendance is strongly encouraged. Absenteeism will almost guarantee your inability to achieve
satisfactory grades. Some of the progress tests may not be announced in advance and details of
assignment requirements may be explained in class. There is not formal provision for make up tests to
replace tests you miss. If there are any extenuating circumstances concerning a student's absence, the
instructor should be notified as soon as possible. Students are responsible for "catching up" on any info
from missed classes regardless of the reason for absence.
Each student should be aware of the College’s policy regarding this subject. AAPS’s Academic policy
will be strictly enforced.
A+ = 90
A = 85-89
A- = 80-84
B+ = 75-79
B = 70-74
C+ = 65-69
C = 60-64
F = 59
Discrimination/Harassment
All students and employees have the right to study and work in an environment that is free from
discrimination and/or harassment. Language or activities that defeat this objective violate the College
Policy on Discrimination/Harassment and shall not be tolerated. Information and assistance are
available from the admission office.
Prerequisites
Automation – can be designed to run unattended for long periods of time without
interruption or operator intervention.
Raw material testing consists of verifying components before they are released
to production. This includes verification of identity, purity and potency.
Finished product testing consists of verifying drugs that are ready to be shipped
to the consumer. Included in Finished Product testing is verification of potency,
bioavailability, and purity.
Shelf-Life Stability
Quality Control Stability programs verify the shelf life of finished pharmaceutical
products currently residing with pharmacies or consumers. They are used to
verify the continued potency, bioavailability, and purity of a pharmaceutical
product until it reaches its expiry date.
In-Process Testing
Labels and components, which come in direct contact with the finished
pharmaceutical product, are checked for suitability and quality.
Formulation
Quality Control methods must be created and validated before use. Analytical
Method Development is the process of creation. Analytical Method Validation is
the process of verifying that methods work reliably and consistently over time.
Process Validation
Process validation in the R&D laboratory verifies that production can perform the
manufacture of a specific drug product reliably and consistently over time.
The process is based on the varying affinities of the components of the chemical
mixture (in the case of liquid chromatography) dissolved in a solvent system (mobile
phase) to the solid phase. As the solvent percolates through the solid material, the rate
at which the components of the mixture pass around or through the solid varies, thus
leading to a separation of the individual components. Components with a higher affinity
to the solid phase move at a slower rate than those with a lower affinity. Continual
flushing of the solvent will eventually lead to each component being flushed out of the
solid phase.
By passing the eluent through a detection device, which can detect the individual
components, the relative amounts of each component can be measured.
2.1 History
Liquid Chromatography has been used in analytical analysis since its beginnings
in the early 1900’s. By the 1970’s, it was widely used but mostly as an
identification and purification tool. Thin Layer Chromatography is still used today
in the pharmaceutical industry as a means for identification and for degradation
analysis. Liquid chromatography was not used extensively as a quantitative tool,
due to the inability to regulate liquid flow. Initial gravity feed columns required
large volumes of mobile phase and took a long time to achieve the desired
Academy of Applied Pharmaceutical Sciences (AAPS) Page 10
PLC1002 - Manual - Rev 06.0
separations. Pressure was used to increase mobile phase flow rate and thus
decrease analysis time, however, it was difficult to maintain pressure and flow at
a constant rate, thus reproducibility was an issue.
Solid phases initially were silica or alumina gel and were very polar, thus the
procedure was limited to relatively non-polar materials and required non-polar
organic solvents.
The most popular of these are Hexane, Ethyl acetate, Acetone, Chloroform and
Dichloromethane. Unfortunately, most pharmaceutical preparations are water-
soluble or are derivatized to be initially water soluble to allow assimilation in the
body. Thus, to use this type of chromatography required neutralization with acid
or caustic and then extensive extractions into hydrophobic solvents to allow for
chromatographic separation and analysis. This process allows for greater
chances of experimental error and is thus not as reliable as a procedure that
could be developed to analyze the component directly.
Since silica and alumina are fairly reactive, derivatization of the free silanol
groups could be performed to create stable, non-ionizable substrates. By
derivatizing the silica gel with alkyl halides, the solid matrix could then be
rendered hydrophobic, resulting in a solid matrix, which would not be highly
adsorbent to hydrophilic components.
3.1 Development
3.2 Instrumentation
Figure 1, shows a typical component HPLC system. The main functional parts are the
pump, injector, column, detector and data station/processor.
3.3 Pump
The pump system is the heart of any HPLC system. The pump must supply a
constant and reliable flow of mobile phase at all times to ensure the resulting
chromatogram is reproducible. Pumps can be designed to deliver a constant
pressure at the column head, or to generate a constant flow of mobile phase.
The advantages of constant pressure pumps are their simplicity and ease of
maintenance. The pressure on any column is dependent on the viscosity of the
mobile phase and the extent of interaction between the solid and mobile phases.
Viscosity can change dramatically throughout the course of an HPLC run by
changes in temperature of the mobile phase. The viscosity will also change upon
changes in mobile phase composition. Thus gradient elution can cause
significant changes in the flow rate in order to keep a constant head pressure.
This type of pump is thus not used much for either qualitative or quantitative
analysis, since changes in retention time can be significant enough to render
peak identification impossible.
The most common type of pump system used in analytical analysis is the
constant flow pump. A constant flow can be generated by a consistent
displacement of a fixed volume. The most common of these is the reciprocating
pump. This type of pump uses a small bore cylinder that is filled with the mobile
phase either by gravity or by a vacuum generated by the piston action of the
pump. A series of check (one way) valves is used to keep the mobile phase
flowing in one direction. The two main types of these are the single and the dual
piston systems.
The pistons are usually sapphire rods, machined to have a consistent length and
diameter. The advantage of sapphire is the extremely high tensile strength in the
linear direction and the material wets consistently. This allows for a high-
pressure seal around the piston thus eliminating the chance of seepage around
the piston.
Check valves are “ball and seat” type. These are manufactured from ruby, since
when wet, will make an extremely good seal between highly polished surfaces.
Both the single and dual piston systems create a “pulse” effect on the pressure,
which can lead to unsteady baselines. These pumps employ a dampener after
the pump.
3.3.3 Isocratic
The reciprocating pump will deliver a consistent reliable flow rate of any one
mobile phase. This is known as isocratic analysis. Isocratic analysis is a very
versatile and well-used type of HPLC analysis. It has the advantages of not
requiring expensive pump systems and is very reproducible, providing the mobile
phase can be prepared the same each time (slight changes in composition
usually does not drastically effect the chromatographic requirements. This is
known as the “ruggedness” of a procedure). The analyses described in the
experimental section of this session will be isocratic.
3.3.4 Gradient
The system controller controls the flow from each pump, for example increasing
one as the other is decreased, thus maintaining a steady flow rate. The
advantage of this is the gasses normally expelled during the mixing of two
different solvents are not formed due to the elevated pressure and the gradient
can be controlled and reproduced very precisely. The disadvantage is the
requirement of two highly accurate pump systems and a controller for these, all
of which require maintenance and calibration.
There can be multiple solenoid valves, which can be controlled to open and close
at various intervals to change the composition of the final effluent. This generally
The sample is generally prepared in a media in which the analyte is soluble and
which is soluble in the mobile phase it is being injected into. Particulates in the
sample can cause problems in the column, causing the column to plug and thus
increasing the backpressure to the point in which the pump can be damaged.
Thus the samples are usually made particulate free by either centrifugation or
filtering through a membrane filter of 0.45 um or less. The sample must be
introduced to the system after the pump and thus the system is under pressure
when the sample is introduced. This requires the use of a valve switching
system, which will not significantly interfere with the flow.
Figure 7
Figure 8
The process would begin through a controller, which would operate valves
holding compressed gas (usually air but nitrogen and helium were also used).
3.4.3 Auto-sampler
Today, column material is normally Type 316 stainless steel, once again chosen
because it offers the best compromise of cost, workability, and corrosion resistance.
Most commercial columns available today have internal diameters of either 2.6-3 mm or
4.6-5 mm. Most leading manufacturers obtain a great amount of interchangeability
between column sizes without excessive fitting replacement by supplying columns with
a variety of internal diameters, but a uniform ¼ in. outside diameter. Naturally,
preparative columns have larger diameters: usual O.D. values are 3/8 - 5/8 in.,
corresponding to internal diameters of 6.4-12.7 mm, respectively.
Great care should be taken in the choice of end fittings, connectors, and unions
however, since inclusion of rapid diameter changes corners, and other upswept areas in
a system can destroy a good separation. It is best to obtain the above parts from a
reputable liquid chromatograph manufacturer since most of these companies offer
fittings, which are specifically made for LC.
Many reducing unions and column end fittings incorporate special devices designed
to hold packings in place and/or prevent their flowing into the detector or detector
tubing. The most frequently used of these consist of fine porous stainless steel fritted
filter discs, usually with an average opening of 1 m. These may be forced into the ends
of columns or may be in the connector or end fitting. The important point is that the frit
disc must be tightly fitted to avoid occurrence of large-diameter spaces at the edge
where the disc contacts the wall.
An additional benefit of the frit is derived from the fact that it fills what otherwise might
be a large empty area in conversion zones thus, the sample cannot diffuse, cross mix,
or undergo other efficiency-killing effects when passing through these areas.
Reversed phase columns are silica based solid phases derivatized with a
hydrophobic ligand. In the most common case, the free silanol groups are
reacted with an alkyl-silane producing a structure such as:
Si-O-Si-R
Amino – R = -CH2-CH2-CH2-NH2
Cyano – R = CH2-CH2-CN
A cyano-modified surface is very slightly, polar. Columns with this phase are
useful for fast separations of mixtures consisting of very different components.
These mixtures might show very broad range of retention times on the usual
columns.
Phenyl – R = (CH2)3C6H5
Diol – R = CH(OH)2
Diols are a slightly, polar adsorbent for normal-phase separations. These are
useful for separation of complex mixtures of compounds with different polarity,
and which usually show a strong retention on unmodified silica.
4. Detectors
The heart of an efficient HPLC detector is the Flow-Through Cell. The figure
below shows the schematic of a modern follow-through cell. The optics provides
focusing of the light beam in the center of the cell where it is virtually unaffected
by the entry-exit-window-interface disturbance, or drift induced by flow,
temperature, or refractive index changes. The short, wide cell assures that
maximum energy is transmitted, and a post-cell collecting lens focuses all of the
existing light from the cell.
Detectors equipped with the flow-through cell were a major breakthrough in the
development of modern liquid chromatography. The group of Tiselius, in Sweden
in 1940, first applied this by continuously measuring the refractive index of the
column effluent. Current LC detectors have wide dynamic range normally
allowing both analytical and preparative scale runs on the same instrument. They
have high sensitivities often allowing the detection of nanograms of material, and
the better models are very flexible, allowing rapid conversion from one mobile
phase to another and from one mode to another.
The refractive index (RI) detector is the only universal detector in HPLC.
The detection principle involves measuring of the change in refractive index of
the column effluent passing through the flow-cell. The greater the RI difference
between sample and mobile phase, the larger the imbalance will become.
Thus, the sensitivity will be higher for the higher difference in RI between sample
and mobile phase. On the other hand, in complex mixtures, sample components
may cover a wide range of refractive index values and some may closely match
that of the mobile phase, becoming invisible to the detector.
4.3 UV Detector
Any chemical compound may interact with the electromagnetic field. A beam of
the electromagnetic radiation passed through the detector flow-cell will
experience some change in its intensity due to this interaction. Measurement of
this change is the basis of the most, optical HPLC detectors.
The most widely used portion of the electromagnetic spectra for HPLC analysis
is the 190 – 800 nm range, since typical mobile phase solvents are invisible to
much of this range, thus they do not interfere greatly with the detection of
compounds which are detectable in this range. This is the range of the normal
UV-Vis detectors employed, today in HPLC. Four major types of UV-Vis
detectors are available on the market today. Each has applied the UV spectra
absorbance in a similar manner.
The fixed wavelength detector was the first of its kind. A low-pressure mercury
vapor lamp will emit very intense light at 253.7 nm. By filtering out all other
emitted wavelengths, manufacturers have been able to utilize this 254 nm line to
provide stable, highly sensitive detectors. Most components containing an
aromatic ring structure have a strong absorbance in the 254 nm range, and thus
these detectors are capable of measuring sub-nanogram quantities of these
components. The 254 nm was chosen since the most intense line of mercury
lamp is 254 nm, and most UV absorbing compounds have some absorbance at
254 nm.
The major drawback with the fixed wavelength detector is the fact that most
compounds that absorb UV light absorb in the 254 nm region. Thus interference
by other components can be a significant problem. Some compounds are
extremely difficult to separate even with today’s advances in column chemistry.
However, many of these compounds absorb in other regions of the UV-Vis
spectra as well. By employing a detector, which can read at other, unique
wavelengths, complete chromatographic separation is not as necessary. The
sensitivity for a particular component can be enhanced greatly by setting the
wavelength to that in which the component has the maximum absorbance.
The functionality of the variable wavelength detector was enhanced again by the
diode array type source. In diode array, the source light (again from a broad
spectrum source such as a deuterium lamp) is passed through the sample cell
and then is split into its corresponding spectrum. The entire spectrum is then
focused onto an array of photo-diodes. The spacing of the diodes is designed to
measure the intensity at specific wavelengths, usually 1 or 2 nm each. Thus the
electromagnetic spectrum is measured by combining the response from each
individual diode. Since the analyte will absorb the UV radiation whether it is split,
or in combination with other wavelengths, the response at any particular
wavelength will be affected by the sample compared to that of the baseline.
The advantage of this is that the entire spectra can be analyzed instantaneously.
Thus, two, three or many more readings at different wavelengths can be
measured simultaneously. All the measurements can be combined to produce a
three dimensional chromatogram (RT x wavelength x absorbance). This is a very
effective tool for characterization of known and unknown components of a
mixture.
4.4 Fluorescence
Figure 14, below shows the optical schematic of a typical fluorescence detector
for liquid chromatography. The detectors available on the market differ in the
method in which the wavelengths are controlled. Less expensive instruments
4.5 Conductivity
Usually this is a very low volume flow-through capillary equipped with two
electrodes and variations in conductivity of the mobile phase due to the eluted
sample components are continuously recorded. Response is linear with
concentration over a wide range, and quantitation of the output signal is possible
with suitable preliminary calibration. Best use is made of this detector is in
isocratic analysis, since solvent gradients will cause a proportional shift in the
baseline. Such detectors have been used most successfully in ion-exchange
chromatography of anions and cation but generally, they have found only limited
popular acceptance.
4.6 Electrochemical
The areas of application of electrochemical detection are not large, but the
compounds for which it does apply, represent some of the most important drug,
pollutant and natural product classes. For these, the specificity, and sensitivity
make it very useful for monitoring these compounds in complex matrices such as
body fluids and natural products. Sensitivities for compounds such as phenol,
catecholamines, nitrosamines, and organic acids are in the picomole (nanogram)
range. The purity of the eluent is very important, because the presence of
oxygen, metal contamination and halides may cause significant background
current and therefore, noise and drift in the base line.
The output from any detector needs to be collected by a system that can plot the
detector response vs. time. The signal coming from the detector is reading of the
absorbance at regular intervals throughout the run. Detectors will report anywhere from
1 to 20 response readings per second, depending on the settings on the detector. The
greater the number of readings will usually lead to more accurate chromatogram.
However, this can also lead to a huge amount of data points for a chromatogram. The
number of points stored per second is dependent on the type of analysis. With HPLC
runs, tangent peak widths are generally in the 1 to 3 minute range. Thus a “sampling
rate” of 1 to 5 data points per second is usually quite sufficient for most analyses.
In HPLC type and composition of the mobile phase (eluent) is one of the
variables influencing the separation. Despite of the large variety of solvents used
in HPLC, there are several common properties:
Purity - Always use ultra pure, HPLC grade or glass distilled solvents for mobile
phase preparation. In organic solvents, impurities can be retained at the head of
the column, eventually leading to a build-up of material which will reduce the
efficiency of the column and may result in ghost peaks, band broadening, analyte
interactions or clogging of the column. All mobile phases must be filtered through
a membrane filter with a pore size of 0.45 um or smaller.
Solubility of the Sample - If the sample is not soluble in the mobile phase, it will
not elute from the column.
Low Viscosity - High viscosity mobile phases will lead to excessively high back
pressures and can damage the pump, seals and check valves.
Each type HPLC analysis has its own requirements. For normal phase, solvents
are mainly non-polar and for reversed phase, eluents are usually a mixture of
water (sometimes pH adjusted or buffered) with some polar organic solvent such
as acetonitrile, methanol or THF.
Most stationary phases are silica base. These are generally stable to organic
mobile phases but there are some exceptions. Silica has a strong affinity towards
halogens and other strongly electronegative substances. Atoms with lone pairs
such as Nitrogen are also readily attracted to silica. As a result, silica gel tends to
be reactive in low or high pH environments. The best and safest chromatography
with silica gel based columns occurs within the pH range of 2.5 to 7. Mobile
phases that are outside of this range tend to destroy the silica base, thus
reducing particle size and causing loss of bonded substrates.
Hydrochloric acid (HCl) - Hydrochloric acid is highly attracted to silica gel and
the chloride may even bond with the silica, displacing bonded substrates. HCl
should never be used in a mobile phase for a silica-based column. HCl salts are
common among many pharmaceutical compounds. Usually the HCl is in a small
enough concentration to have a negligible effect on the stability of the solid
phase, however, repeated introduction of this type of analyte will have an effect
on the performance of the column. This effect can also be minimized by the
addition of a competing base such as triethylamine in the mobile phase (buffered
to between 3.5 and 6). Here the triethylamine will act as an “HCl scavenger”,
complexing the HCl such that it does not come into contact with the silica base.
Columns are available in today’s market that are designed to handle these types
of problems. Some are so well protected that they are stable at pHs from 2.0 to
10.0 without appreciable loss of performance even after hundreds of hours of
use.
Another type of solid phase that has become popular is the polymer-based
column. These have the advantage that they are stable to strong acids (including
up to 2N HCl) and to strong bases (2N NaOH). However, the polymer backbone
is not stable to strong organic solvents. As much as 25% Methanol or Acetonitrile
in water can cause many of these polymer backbones to swell, rendering the
column useless.
The following are excerpts from the cGMP regulations 21 Code of Federal Regulation,
(CFR) Part 211. These indicate the requirements for records creation and storage. Data
acquired from HPLC analyses fall under this jurisdiction. The requirements are,
basically, that all records must be kept and must be positively identifiable to a specific
batch of the particular product or material. All equipment used must be calibrated such
that the normal mode of operation is within the limits of the calibration. The actual
regulations on Laboratory records are included here for completeness.
(a) Laboratory records shall include complete data derived from all tests necessary
to assure compliance with established specifications and standards, including
examinations and assays, as follows:
Note: important points and points relating to HPLC analysis have been
underlined)
The statement shall indicate the location of data that establish that the
methods used in the testing of the sample meet proper standards of
accuracy and reliability as applied to the product tested. (If the method
employed is in the current revision of the United States Pharmacopeia,
National Formulary, Association of Official Analytical Chemists, Book of
Methods, {2} or in other recognized standard references, or is detailed in
an approved new drug application and the referenced method is not
modified, a statement indicating the method and reference will suffice).
The suitability of all testing methods used shall be verified under actual
conditions of use.
(3) A statement of the weight or measure of sample used for each test, where
appropriate.
(4) A complete record of all data secured in the course of each test, including
all graphs, charts, and spectra from laboratory instrumentation, properly
identified to show the specific component, drug product container, closure,
in-process material, or drug product, and lot tested.
(5) A record of all calculations performed in connection with the test, including
units of measure, conversion factors, and equivalency factors.
(6) A statement of the results of tests and how the results compare with
established standards of identity, strength, quality, and purity for the
component, drug product container, closure, in-process material, or drug
product tested.
(7) The initials or signature of the person who performs each test and the
date(s) the tests were performed.
(8) The initials or signature of a second person showing that the original
records have been reviewed for accuracy, completeness, and compliance
with established standards.
Give a detailed description of a normal phase system. How would increasing the
non-polar(hydrophobic) component of the mobile phase affect analyte retention
time?
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
The mobile phase transports sample components to the detector where the
signal produced by the sample components is plotted against time. This output is
called a "Chromatogram".
Figure 17
The product of retention time and eluent flow rate, so called "Retention Volume,
VR ", is more of a global retention parameter. Retention volume represents the
volume of the eluent passed through the column while eluting a particular
component.
• Dead volume is the volume of the eluent that passed through the column
while the component was moving with the liquid phase.
The second part is equal to the combined volume of the liquid phase in the
column and tubing between the injector and the detector (dead volume, Vo),
and it will be the same for any component eluted on this column and system.
Components that are not retained by the stationary phase but carried through
the column without interaction are said to elute at the dead time "t0" of the
system.
10.1 Capacity
Figure 18
The value of the capacity factor, k' depends on the chemical nature of the
compound, the nature, amount, and surface area of the liquid phase,
the column temperature, and the gas flow rate.
The Capacity factor, k' is a measure of the retention of the sample molecule on
the column. A high k' value means that the sample is highly retained and has
spent a significant amount of time interacting with stationary phase.
• Capacity factor is also equal to the number of moles in the stationary phase
divided by the number of moles in the mobile phase
t1 − t 0
k `=
t0
10.2 Selectivity
The selectivity is the ability of the stationary phase to distinguish between two
separate sample components. Selectivity, is calculated as the ratio of the
capacity factors k'.
Figure 19
k `2
Selectivity =
k `1
The greater separation, the greater selectivity value, however, actual separation
cannot be determined by selectivity alone.
10.3 Efficiency
Figure 20
tR / Wb
The actual equation for column efficiency is:
2
t1
N = 16
W
B1
The efficiency, N, or plate number is a mathematical measure of the broadening
of the peak as it passes through the column. The relationship is an inverse
relationship
Figure 21
2 (t 2 − t1 )
R=
WB1 + WB 2
Where:
R = Resolution
tRB = Retention time of component B
tRA = Retention time of component A
W = Peak width at base
Figure 22
Figure 23
Where:
W = Peak width
Wh/2 = Peak width at 1/2 the peak height
h = Peak height
h/2 = Half the peak height
t1 = Retention time for peak 1
t2 = Retention time for peak 2
The tailing factor, T, is a measure of peak symmetry showing the unity for
perfectly symmetrical peaks. T value increases as tailing becomes more
pronounced.
W0.05
T=
2 f
Figure 26
System suitability tests are an integral part of gas and liquid chromatographic methods.
They are used to verify that the resolution and reproducibility of the chromatographic
system are adequate for the analysis to be done. The tests are based on the concept
that the equipment, electronics, analytical operations, and samples to be analyzed
constitute an integral system that can be evaluated as such.
Performance measurements, capacity, resolution and tailing are integral parts of system
suitability. Values of these measurements are defined in the analytical procedure to
indicate the analytical system is functioning as required to achieve the desired
separation.
Reproducibility is defined as the ability for the HPLC system, as prepared, to reproduce
the injection volumes desired. This is vital for quantitative analysis. The primary
standard is usually injected 5 times in sequence. The relative standard deviation of the
analyte peak area is determined. The standard in the pharmaceutical industry (USP) is
2.0% for 5 replicate injections.
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
_____________________________________________________________________
Mobile Phase
Preparation of Standard
Preparation of Sample
Equipment Parameters
After allowing the HPLC to come to equilibrium with your mobile phase,
inject your standard and sample. Do you have a positive identification for
your sample. Consider Retention time and peak shape in comparison to
the reference standard.
Most of the HPLC column hardware is constructed of high grade stainless steel, which
is a chromium-nickel-molybdenum steel. The steel tubes are precision drilled and
polished. The columns are closed at both ends with frits that vary in pore size according
to the support particle diameter. The columns are chemically resistant to most HPLC
solvents with exception to chloride ions, which may react with stainless steel surface.
These columns are designed to hold up well under typical HPLC pressures.
Figure 37
FLOW
Detector Pump
The column life may be extended for silica-based columns. Silica dissolves more readily
at extreme pH, high ionic strength, or high mobile phase polarity.
Figure 38 represents the general shape of the chromatogram for this mixture.
Figure 38
Porous silica particles are commonly used as HPLC supports for adsorption
chromatography and forms basis for numerous chemically modified stationary
phases for reversed phase, normal phase, ion exchange, chiral, and size
exclusion chromatography. The chemical composition is SiO2 x H2O. The purity
of the silica can affect the chromatographic reproducibility.
There are two basic shapes are available, irregular and spherical.
Figure 39
Spherical particles have the effect of consistency while mobile phase is passing
around them. The paths through the particles are more regular and thus there is
less variation in the paths the analyte can take through the solid phase.
Figure 40
Figure 41
Number %
Particle Diameter, um
One of the most important adsorbent parameters is the pore size and pore size
distribution. Adsorbent surface area is a major factor directly affecting the analyte
retention. Pore size defines an ability of the analyte molecules to penetrate inside
the particle and interact with its inner surface. This is especially important
because the ratio of the outer particle surface to its inner one is about 1:1000.
The surface molecular interaction mainly occurs on the inner particle surface.
Pore sizes vary from column type to column type, but control of pore size can
greatly affect the capacity and thus the resolution of a chromatogram.
13.3 Dimensions
This can be the most critical part of column performance – the next time the
column is to be used. Analytical columns are not inexpensive – they range
anywhere from $300 to $4000 per column. Although they are generally made of
very strong steel, they are fragile. The hardware will not be damaged if it is
dropped on the floor, but the impact may cause voids to be formed in the packing
material, thus leading to band broadening, loss of efficiency or split peaks. Thus,
they must be handled with care.
Whenever a column is used, it should be washed with a solvent system that will
remove any retained material or any buffer that is used in a mobile phase.
Never store silica based, reversed phase columns in buffer! These can cause
degradation of the silica resulting in loss of bonded phase. The buffer may also
precipitate in the column.
Never wash a reversed phase column with 100% water! This can cause the
hydrophobic “branches” of the bonded phase to collapse.
The best wash and storage solution for a reversed phase column is the mobile
phase without any buffer or organic modifier. The column should be rinsed with
this for about 15 minutes at the method flow rate, followed by a wash with a
higher organic version of the same solution for 15 to 60 minutes, then another 15
minutes with the wash solvent. The column should be stored in the wash
solution. If the wash solution contains less than 25% organic solvent, then a
subsequent wash of 25% organic solvent in water should be performed prior to
column storage to prevent microbial growth.
The column wash procedure should be performed on the HPLC that was used
for the analysis, thus washing the HPLC as well.
Columns can be operated at any flow rate that is consistent with the
backpressure limitations described below. Flow rates should be optimized to
provide the highest efficiency for your sample.
Particle Size Internal Diameter Typical Flow Rate (ml/min) Typical Pressure (psi)
• Flush with at least five column volumes of mobile phase without buffer to
remove any buffers or salts
MeCN:H2O
V=r2L Where:
V = Column volume, ml
r = Column radius, cm
L = Column length, cm
Perform Calculations
Weight of Impurity Standard = 102.21mg
Weight of Sample A= 101.39 mg
Weight of Sample B = 101.18 mg
Purity of Impurity Standard = 99.6%
Standard Dilution Scheme = 200 ml Volumetric flask, 5.0ml to 100ml volumetric
flask
Sample Dilution Scheme = 200 ml Volumetric Flask
Areas as detected by HPLC
1. Standard A - 7791
2. Standard A - 7242
3. Standard A - 7646
4. Standard A - 7697
5. Standard A - 7892
6. Standard A - 7894
7. Sample A – 785
8. Sample B - 774
9. Check Standard – 7783
Does standard "A" meet the criteria of NMT 6.0% for six injections?
Does the sample comply with a specification of NMT 1.0% for this Impurity?
Equipment Parameters
After allowing the HPLC to come to equilibrium with your mobile phase,
inject your System Suitability Solution (Label as Sys Suit 1).
Add one drop of glacial acetic acid to your mobile phase. After allowing
the HPLC to come to equilibrium with your mobile phase, inject your
System Suitability Solution (Label as Sys Suit 2 ).
Add 100ml of methanol. After allowing the HPLC to come to equilibrium
with your mobile phase, inject your System Suitability Solution (Label as
Sys Suit 3).
Compare the effect that mobile phase changes have on your
chromatography. Calculate tailing factor, column efficiency and
resolution for all chromatograms produced.
17.3 Composition
R – SO3-Na+
where “R” is usually an alkyl group of chain length from 1 to 12. These are
useful reagents when the analyte is an ionic molecule that does not have
good retention on a reverse phase column. The emulsifying action of the
ion pair reagent causes the ionic molecule to slow down in the alkyl end
will be attracted to the solid phase while the polar/ionic end will attract the
ionic analyte. These will be used in conjunction with a buffer and quite
often with triethylamine as well.
• Mix solvents
As mentioned previously, all solvents must be high purity with a very low
UV cutoff range. Water for mobile phases must be Type 1 grade, i.e.
purified by reverse osmosis or distillation, ion exchange and filtered to at
least 0.45 µm. This is the USP definition of HPLC grade water:
Absorption Characteristics:
Determine the ultraviolet absorbance in a 1 cm cell, using water as
the blank. The maximum absorbances are 0.005, 0.01, and 0.01 at
400 nm to 250 nm, 200 nm, and 190 nm respectively.
Residue on evaporation:
Mobile phase preparation can be divided into the following major steps:
17.6.3 Filtration
All buffers and prepared solvent mixtures must be filtered through a 0.45
um or smaller membrane filter. This will remove any particulates that are
present in the solution. Vacuum filtration also serves a dual purpose in
degassing the mobile phase. Filters available today are made of many
different types of materials for different purposes. Some filters are not
compatible with different solvents, thus care must be taken in choosing
the correct filter type. One of the most versatile materials is Nylon 66.
These filters will handle most of the aqueous mobile phases available for
reverse phase. They are compatible with buffer systems and with many
organic modifiers in dilute concentrations with water (see membrane
selection chart).
Figure 42 - Typical Mobile Phase Filtration Devices
When a mobile phase on the HPLC system has been idle, there is always
the possibility that air has managed to permeate to the felow. Priming the
HPLC system, which involves pimping each channel at 100% composition
and high flow rate until steady pressure and flow are obtained, is highly
recommended.
Analytes must be soluble in the mobile phase. If there is any doubt, a trial
should be performed to ensure the sample solution does not precipitate
the mobile phase or visa versa. The ideal solvent for the analyte is mobile
phase. If this is not feasible for any reason, then the analyte should be
dissolved in a solvent system that is weaker than the mobile phase - i.e.
for a reversed phase system, choose a solvent that is more aqueous and
All samples should be filtered prior to injection onto an HPLC system. The
best filters to use have a nominal pore size of 0.45 um or less. Disposable
syringe filters are available for this purpose and the membranes discussed
for filtering mobile phases are all available (see section 4.6.3) Filters
should be evaluated prior to use to establish saturation levels of the
analyte. Often 5 to 10 ml of sample solution is required to saturate the
filter. Filter studies are a common part of most analytical method
validations. Syringe filters are available in many sizes – from 3 mm to 47
mm – however, the normal size for pharmaceutical analysis is 25 mm.
Filters should not be re-used. Each sample or standard is unique and
should be treated individually in order to avoid cross-contamination
between samples and standards. Smaller filters have an advantage in that
the amount of analyte solution required for saturation is generally less
than a larger filter. However, small filters are usually more difficult to pass
solution through them, since they can become clogged easily. If there are
any visible particulates in the analyte solution, a pre-filter of larger pore
size should be used prior to the final filter. This filtration can be of a gravity
type using filter paper and a funnel, or a larger disposable filter can be
attached in series with the final filter. A common pre-filter is a glass fibre
filter of 1 to 3 um pore size. Any pre-filtration step must be evaluated to
determine the saturation volume required.
All Glassware used must be labeled for safety reasons and for GMP/GLP.
Your labels should contain:
Details about the testing itself - Is this a stock solution or a working solution? Is
this the third dilution in a series of five? The more you put on a label, the more
organized you’ll be and the less likely you are to make a mistake.
All chromatograms must be labeled for correct identification. An HPLC run will
contain standard chromatograms, system suitability chromatograms, sample
chromatograms (often duplicate samples and multiple lots) and check/control
standards. Each of these needs to be clearly identified to avoid the possibility of
mix-ups. Normally this can be accomplished through the chromatographic
software in the sequence table.
18.3 Documentation
Equipment Parameters
Calculations
PurityStd (%)
Wt Std (mg ) Dil Std (1 / ml )
ASpl 100(%) AvgWt(mg / DU )
% LC = 100(%)
AStd Wt Spl (mg ) Dil Spl (1 / ml ) LC (mg / DU )
This table offers general guidelines for membrane characteristics and applications.
Membrane Type Membrane Characteristics & Application
Regenerated Regenerated Cellulose (RC) is a hydrophilic, solvent resistant, low protein
Cellulose binding membrane. Membrane of choice for low non-specific binding
(biological analysis, protein, peptides). Ideal for removing particulates from
HPLC samples prior to injection. Compatible with all HPLC solvents. Use
for particle removal and degassing of solvents. RC membranes are also
compatible with aqueous solutions in the 3 to 12 pH range. Extractables
with water are less than 1%. When used with a glass pre-filter, RC is ideal
for tissue culture media filtration, as well as general biological sample
filtration.
Nylon A naturally hydrophilic membrane that requires no pre-wetting. Nylon is
extremely low in extractables and is mechnically strong. Can withstand
solution up to 50° C. Can be used with organic or aqueous solutions. Flow
rate is excellent. Not recommended for strong acids or bases.
Highly recommended for general laboratory filtration and filtration of HPLC
samples prior to injection. Nylon binds protein and should not be used
when protein recovery is important.
PTFE PTFE (polytetrafluoroethylene) is hydrophobic and chemically resistant1 to
all solvents, acids, and bases. Membrane is mechnically strong and can
withstand high temperature liquids. Requires prewetting with an alcohol
prior to use with aqueous solutions. Very low in extractables PTFE natural
hydrophobicity blocks water vapor, making it ideal for transducer
protectors. Also ideal for filtering-degassing organic and highly basic
mobile phase solutions for liquid chromatography.
PVDF PVDF (polyvinylidene difluoride) is a hydrophilic, solvent resistant1
membrane that exhibits low levels of UV adsorbing extractables Low non-
specific binding character.
Use for HPLC sample filtration and general biological filtration. PVDF is a
low protein binding membrane.
Polypropylene Polypropylene is a hydrophilic membrane that exhibits a wide range of
chemical compatibility to organic solvents. A good choice for HPLC
sample filtration when performing chromatography protein analysis. Highly
solvent resistant. Low non-specific adsorbing membrane for maximum
protein recovery in critical analysis. Also well suited for biological sample
filtration.
Glass Microfiber (GMF) membranes are commonly used as pre-filters to
Glass Microfiber remove large particulates and to extend the load capacity of the
membrane. Membrane of choice for dissolution testing.
Is a very low protein binding membrane, ideal for aqueous based samples.
CA membranes are an excellent choice when maximum protein recovery
Cellulose Acetate
in filtrate is critical. Laboratory studies show that CA membranes bind less
(CA)
protein than PVDF. CA is ideal for tissue culture media filtration and
sensitive biological samples.
PES (Polyethersulfone) provides high flow rates and good throughput
volume. This membrane is low protein binding and can be used with high
temperature liquids. PES is a mechanically strong membrane and is
Polyethersulfone resistive to a broad range of aqueous solvents. PES is the filter of choice
for tissue culture work‹very low extractables and low binding of protein and
nucleic acids. Good to excellent flow rates. PES is certified for Ion
Chromatography.
Chart Legend
C = Compatible | LC = Limited Compatibility (membrane swells and shrinks)
NC = Not Compatible | ND = No Data Available | PTFE = Polytetrafluoroethylene
PVDF = Polyvinylidene Difluoride | PES = Polyethersulfone | CA = Cellulose Acetate
RC = Regenerated Cellulose | PP = Polypropylene | GMF = Glass MicroFiber
Chemical Nylon PTFE PVDF PES CA RC PP GMF
ACIDS
Acetic, Glacial LC C C C NC C C C
Acetic, 25% C C C C C C C C
Hydrochloric, Concentrated NC C C C NC NC C C
Hydrochloric, 25% NC C C C NC NC C C
Sulfuric, Concentrated NC C NC NC NC NC C C
Sulfuric, 25% NC C C C NC LC C C
Nitric, Concentrated NC C C NC NC NC C LC
Nitric, 25% NC C C C NC NC C LC
Phosphoric, 25% NC C ND ND C LC C ND
Formic, 25% NC C ND ND LC C C C
Trichloroacetic, 10% NC C ND ND C C C ND
ALKALIES
Ammonium Hydroxide,
C C LC C C LC C C
25%
Sodium Hydroxide, 3
C C C C NC LC C NC
Normal
ALCOHOLS
Methanol, 98% C C C C C C C C
Ethanol, 98% C C C C C C C C
Ethanol, 70% LC C C C C C C C
Isopropanol C C C C C C C C
n-Propanol C C C C C C C C
Amyl Alcohol, Butanol C C C C C C C C
Benzyl Alcohol C C C ND LC C C NC
Ethylene Glycol C C C C C C C C
Propylene Glycol C C C C LC C C C
Glycerol C C C C C C C C
Chemical Nylon PTFE PVDF PES CA RC PP GMF
HYDROCARBONS
Hexane, Xylene C C C NC C C NC C
Toluene, Benzene C C C NC C C NC C
Kerosene, Gasoline C C C LC C C LC ND
Tetralin, Decalin ND C C ND C C ND ND
HALOGENATED
HYDROCARBONS
Methylene Chloride LC C C NC NC C LC C
Chloroform C C C NC NC C LC C
Trichloroethylene C C C NC C C C C
Monochlorobenzene C C C LC C C C C
Freon C C C LC C C C C
Carbon Tetrachloride C C C NC LC C LC C
KETONES
Acetone C C NC NC NC C C C
INTRODUCTION
Chromatographic separation techniques are multistage separation methods in which the components of a sample
are distributed between two phases, of which one is stationary and the other mobile. The stationary phase may
be a solid or a liquid supported on a solid or a gel. The stationary phase may be packed in a column, spread as a
layer, distributed as a film, or applied by other techniques. The mobile phase may be gaseous or liquid or
supercritical fluid. The separation may be based on adsorption, mass distribution (partition), or ion exchange; or it
may be based on differences among the physicochemical properties of the molecules, such as size, mass, and
volume. This chapter contains general procedures, definitions, and calculations of common parameters and
describes general requirements for system suitability. The types of chromatography useful in qualitative and
quantitative analysis employed in USP procedures are column, gas, paper, thin-layer (including high-performance
thin-layer chromatography), and pressurized liquid chromatography (commonly called high-pressure or high-
performance liquid chromatography).
GENERAL PROCEDURES
This section describes the basic procedures used when a chromatographic method is described in a monograph.
The following procedures are followed unless otherwise indicated in the individual monograph.
Paper Chromatography
Stationary Phase: The stationary phase is a sheet of paper of suitable texture and thickness. Development may
be ascending, in which the solvent is carried up the paper by capillary forces, or descending, in which the solvent
flow is also assisted by gravitational force. The orientation of paper grain with respect to solvent flow is to be kept
constant in a series of chromatograms. (The machine direction is usually designated by the manufacturer.)
Apparatus: The essential equipment for paper chromatography consists of a vapor-tight chamber with inlets for
addition of solvent and a rack of corrosion-resistant material about 5 cm shorter than the inside height of the
chamber. The rack serves as a support for solvent troughs and for antisiphon rods that, in turn, hold up the
chromatographic sheets. The bottom of the chamber is covered with the prescribed solvent system or mobile
phase. Saturation of the chamber with solvent vapor is facilitated by lining the inside walls with paper wetted with
the prescribed solvent system.
Spotting: The substance or substances analyzed are dissolved in a suitable solvent. Convenient volumes,
delivered from suitable micropipets, of the resulting solution, normally containing 1–20 µg of the compound, are
placed in 6- to 10-mm spots not less than 3 cm apart.
Descending Paper Chromatography Procedure
1. A spotted chromatographic sheet is suspended in the apparatus, using the antisiphon rod to hold the
upper end of the sheet in the solvent trough. [NOTE—Ensure that the portion of the sheet hanging
below the rods is freely suspended in the chamber without touching the rack, the chamber walls, or the
fluid in the chamber. ]
2. The chamber is sealed to allow equilibration (saturation) of the chamber and the paper with the solvent
vapor. Any excess pressure is released as necessary.
3. After equilibration of the chamber, the prepared mobile phase is introduced into the trough through the
inlet.
4. The inlet is closed, and the mobile solvent phase is allowed to travel the desired distance down the
paper.
5. The sheet is removed from the chamber.
6. The location of the solvent front is quickly marked, and the sheet is dried.
7. The chromatogram is observed and measured directly or after suitable development to reveal the location
of the spots of the isolated drug or drugs.
r = tR2 tM/tR1 tM
where tR2 is the retention time measured from the point of injection of the compound of interest; t R1 is the retention
time measured from the point of injection of the compound used as reference; and t M is the retention time of a
nonretained marker defined in the procedure, all determined under identical experimental conditions on the same
column.
Relative Retention Time (RRT): Also known as unadjusted relative retention. Comparisons in USP are normally
made in terms of unadjusted relative retention, unless otherwise indicated.
RRT = tR2/tR1
The symbol rG is also used to designate unadjusted relative retention values.
Relative Standard Deviation in Percentage
or
k = (tR tM)/tM
Retention Time (tR): In liquid chromatography and gas chromatography, the retention time, t R, is defined as the
time elapsed between the injection of the sample and the appearance of the maximum peak response of the
eluted sample zone. tR may be used as a parameter for identification. Chromatographic retention times are
characteristic of the compounds they represent but are not unique. Coincidence of retention times of a sample
and a reference substance can be used as a partial criterion in construction of an identity profile but may not be
sufficient on its own to establish identity. Absolute retention times of a given compound may vary from one
chromatogram to the next.
Retention Volume (VR): The retention volume is the volume of mobile phase required for elution of a component.
It may be calculated from the retention time and the flow rate in mL/min:
VR = tR × F
Resolution (RS): The resolution is the separation of two components in a mixture, calculated by:
Separation Factor ( ): The separation factor is the relative retention calculated for two adjacent peaks (by
convention, the value of the separation factor is always >1):
= k2/k1
SYSTEM SUITABILITY
System suitability tests are an integral part of gas and liquid chromatographic methods. These tests are used to
verify that the chromatographic system is adequate for the intended analysis.
The tests are based on the concept that the equipment, electronics, analytical operations, and samples analyzed
constitute an integral system that can be evaluated as such.
Factors that may affect chromatographic behavior include the following:
• Composition, ionic strength, temperature, and apparent pH of the mobile phase
• Flow rate, column dimensions, column temperature, and pressure
• Stationary phase characteristics, including type of chromatographic support (particle-based or
monolithic), particle or macropore size, porosity, and specific surface area
• Reverse-phase and other surface modification of the stationary phases, the extent of chemical
modification (as expressed by end-capping, carbon loading, etc.)
The resolution, RS, is a function of the number of theoretical plates, N (also referred to as efficiency), the
separation factor, , and the capacity factor, k. [NOTE—All terms and symbols are defined in the preceding
section Definitions and Interpretation of Chromatograms. ] For a given stationary phase and mobile phase, N may
be specified to ensure that closely eluting compounds are resolved from each other, to establish the general
resolving power of the system, and to ensure that internal standards are resolved from the drug. This is a less
reliable means to ensure resolution than is direct measurement. Column efficiency is, in part, a reflection of peak
sharpness, which is important for the detection of trace components.
Replicate injections of a standard preparation or other standard solutions are compared to ascertain whether
requirements for precision are met. Unless otherwise specified in the individual monograph, data from five
%RSD = KB n/t90%, n 1
where K is a constant (0.349), obtained from the expression K = (0.6/ 2) × (t 90%,5/ 6), in which 0.6/ 2
represents the required percentage relative standard deviation after six injections for B = 1.0; B is the upper limit
given in the definition of the individual monograph minus 100%; n is the number of replicate injections of the
reference solution (3 n 6); and t90%, n 1 is the Student’s t at the 90% probability level (double sided) with n
1 degrees of freedom.
Unless otherwise prescribed, the maximum permitted relative standard deviation does not exceed the appropriate
value given in the table of repeatability requirements. This requirement does not apply to tests for related
substances.
Relative Standard Deviation Requirements
Number of Individual Injections
3 4 5 6
where H is the height of the peak measured from the peak apex to a baseline extrapolated over a distance 5
times the peak width at its half-height; and h is the difference between the largest and smallest noise values
observed over a distance 5 times the width at the half-height of the peak and, if possible, situated equally
around the peak of interest (see Figure 5).
in which F1 is the flow rate indicated in the monograph, in mL/min; F2 is the adjusted flow rate, in mL/min; l1 is the
length of the column indicated in the monograph; l2 is the length of the column used; d1 is the column inner
diameter indicated in the monograph; and d2 is the internal diameter of the column used. Additionally, the flow
rate can be adjusted by ±50%.
Injection Volume (HPLC): The injection volume can be reduced as far as is consistent with accepted precision
and detection limits; no increase is permitted.
Injection Volume and Split Volume (GC): The injection volume and split volume may be adjusted if detection and
repeatability are satisfactory.
Column Temperature (HPLC): The column temperature can be adjusted by as much as ±10 . Column
thermostating is recommended to improve control and reproducibility of retention time.
QUANTITATION
During quantitation, disregard peaks caused by solvents and reagents or arising from the mobile phase or the
sample matrix.
In the linear range, peak areas and peak heights are usually proportional to the quantity of compound eluting. The
peak areas and peak heights are commonly measured by electronic integrators but may be determined by more
classical approaches. Peak areas are generally used but may be less accurate if peak interference occurs. The
components measured are separated from any interfering components. Peak tailing and fronting is minimized,
and the measurement of peaks on tails of other peaks are avoided when possible.
Although comparison of impurity peaks with those in the chromatogram of a standard at a similar concentration is
preferred, impurity tests may be based on the measurement of the peak response due to impurities and
expressed as a percentage of the area of the drug peak. The standard may be the drug itself at a level
corresponding to, for example, 0.5% impurity, assuming similar peak responses. When impurities must be
determined with greater certainty, use a standard of the impurity itself or apply a correction factor based on the
response of the impurity relative to that of the main component.
External Standard Method: The concentration of the component(s) quantified is determined by comparing the
response(s) obtained with the sample solution to the response(s) obtained with a standard solution.
Internal Standard Method: Equal amounts of the internal standard are introduced into the sample solution and a
standard solution. The internal standard is chosen so that it does not react with the test material, is stable, is
resolved from the component(s) quantified (analytes), and does not contain impurities with the same retention
time as that of the analytes. The concentrations of the analytes are determined by comparing the ratios of their
peak areas or peak heights and the internal standard in the sample solution with the ratios of their peak areas or
peak heights and the internal standard in the standard solution.
Normalization Procedure: The percentage content of a component of the test material is calculated by
determining the area of the corresponding peak as a percentage of the total area of all the peaks, excluding those
due to solvents or reagents or arising from the mobile phase or the sample matrix and those at or below the limit
at which they can be disregarded.
Calibration Procedure: The relationship between the measured or evaluated signal y and the quantity (e.g.,
concentration, mass) of substance x is determined, and the calibration function is calculated. The analytical
results are calculated from the measured signal or evaluated signal of the analyte and its position on the
calibration curve.
1 The parameters k, N, r, and rG were developed for isothermal GC separations and isocratic HPLC separations.
Because these terms are thermodynamic parameters, they are valid only for separations made at constant
temperature, mobile phase composition, and flow rate. However, for separations made with a temperature
program or solvent gradient, these parameters may be used simply as comparative means to ensure that
adequate chromatographic conditions exist to perform the methods as intended in the monographs.
2 It is also a common practice to measure the Asymmetry Factor as the ratio of the distance between the vertical
line connecting the peak apex with the interpolated baseline and the peak front, and the distance between that
line and the peak back measured at 10% of the peak height (see Figure 4), would be (W0.10 f0.10)/f0.10. However,
for the purposes of USP, only the formula (As) as presented here is valid.
Auxiliary Information— Please check for your question in the FAQs before contacting USP.
Topic/Question Contact Expert Committee
General Chapter Horacio N. Pappa, Ph.D. (GCCA2010) General Chapters - Chemical Analysis
Principal Scientific Liaison
1-301-816-8319
USP36–NF31 Page 268
Pharmacopeial Forum: Volume No. 38(2)
USP36–NF31
The following list of packings (L), phases (G), and supports (S) is intended to be a convenient reference for the
chromatographer. [NOTE—Particle sizes given in this listing are those generally provided. Where other, usually
finer, sizes are required, the individual monograph specifies the desired particle size. Within any category of
packings or phases listed below, there may be a wide range of columns available. Where it is necessary to define
more specifically the chromatographic conditions, the individual monograph so indicates. ]
Packings
L1 —Octadecyl silane chemically bonded to porous or nonporous silica or ceramic microparticles, 1.5 to 10 µm in
diameter, or a monolithic silica rod.
L2 —Octadecyl silane chemically bonded to silica gel of a controlled surface porosity that has been bonded to a
solid spherical core, 30 to 50 µm in diameter.
L3 —Porous silica particles, 1.5 to 10 µm in diameter, or a monolithic silica rod.
L4 —Silica gel of controlled surface porosity bonded to a solid spherical core, 30 to 50 µm in diameter.
L5 —Alumina of controlled surface porosity bonded to a solid spherical core, 30 to 50 µm in diameter.
L6 —Strong cation-exchange packing–sulfonated fluorocarbon polymer coated on a solid spherical core, 30 to 50
µm in diameter.
L7 —Octylsilane chemically bonded to totally porous or superficially porous silica particles, 1.5–10 µm in
diameter, or a monolithic silica rod.
L8 —An essentially monomolecular layer of aminopropylsilane chemically bonded to totally porous silica gel
support, 1.5 to 10 µm in diameter.
L9 —Irregular or spherical, totally porous silica gel having a chemically bonded, strongly acidic cation-exchange
coating, 3 to 10 µm in diameter.
L10 —Nitrile groups chemically bonded to porous silica particles, 1.5 to 10 µm in diameter.
L11 —Phenyl groups chemically bonded to porous silica particles, 1.5 to 10 µm in diameter.
L12 —A strong anion-exchange packing made by chemically bonding a quaternary amine to a solid silica
spherical core, 30 to 50 µm in diameter.
L13 —Trimethylsilane chemically bonded to porous silica particles, 3 to 10 µm in diameter.
L14 —Silica gel having a chemically bonded, strongly basic quaternary ammonium anion-exchange coating, 5 to
10 µm in diameter.
L15 —Hexylsilane chemically bonded to totally porous silica particles, 3 to 10 µm in diameter.
L16 —Dimethylsilane chemically bonded to porous silica particles, 5 to 10 µm in diameter.
L17 —Strong cation-exchange resin consisting of sulfonated cross-linked styrene–divinylbenzene copolymer in
the hydrogen form, 6 to 12 µm in diameter.
L18 —Amino and cyano groups chemically bonded to porous silica particles, 3 to 10 µm in diameter.
L19 —Strong cation-exchange resin consisting of sulfonated cross-linked styrene-divinylbenzene copolymer in
the calcium form, about 9 µm in diameter.
L20 —Dihydroxypropane groups chemically bonded to porous silica or hybrid particles, 1.5 to 10 µm in diameter.
L21 —A rigid, spherical styrene-divinylbenzene copolymer 3 to 10 µm in diameter.
L22 —A cation-exchange resin made of porous polystyrene gel with sulfonic acid groups, about 10 µm in size.
L23 —An anion-exchange resin made of porous polymethacrylate or polyacrylate gel with quaternary ammonium
groups, 7–12 µm in size.
L28 —A multifunctional support, which consists of a high purity, 100 , spherical silica substrate that has been
bonded with anionic exchanger, amine functionality in addition to a conventional reversed phase C8 functionality.
L29 —Gamma alumina, reverse-phase, low carbon percentage by weight, alumina-based polybutadiene spherical
particles, 5 µm in diameter with a pore volume of 80 .
L30 —Ethyl silane chemically bonded to totally porous silica particles, 3 to 10 µm in diameter.
L31 —A hydroxide-selective, strong anion-exchange resin-quaternary amine bonded on latex particles attached
to a core of 8.5-µm macroporous particles having a pore size of 2000 and consisting of ethylvinylbenzene
cross-linked with 55% divinylbenzene.
L32 —A chiral ligand-exchange packing–L-proline copper complex covalently bonded to irregularly shaped silica
particles, 5 to 10 µm in diameter.
L33 —Packing having the capacity to separate dextrans by molecular size over a range of 4,000 to 500,000 Da. It
is spherical, silica-based, and processed to provide pH stability. [NOTE—Available as TSK-GEL G4000SWxl from
Tosoh Bioscience (www.tosohbioscience.com). ]
L34 —Strong cation-exchange resin consisting of sulfonated cross-linked styrene–divinylbenzene copolymer in
the lead form, 7 to 9 µm in diameter.
L35 —A zirconium-stabilized spherical silica packing with a hydrophilic (diol-type) molecular monolayer bonded
phase having a pore size of 150 .
L36 —A 3,5-dinitrobenzoyl derivative of L-phenylglycine covalently bonded to 5-µm aminopropyl silica.
L37 —Packing having the capacity to separate proteins by molecular size over a range of 2,000 to 40,000 Da. It
is a polymethacrylate gel.
L38 —A methacrylate-based size-exclusion packing for water-soluble samples.
L39 —A hydrophilic polyhydroxymethacrylate gel of totally porous spherical resin.
L40 —Cellulose tris-3,5-dimethylphenylcarbamate coated porous silica particles, 5 to 20 µm in diameter.
L44 —A multifunctional support, which consists of a high purity, 60 , spherical silica substrate that has been
bonded with a cationic exchanger, sulfonic acid functionality in addition to a conventional reversed phase C8
functionality.
L45 —Beta cyclodextrin bonded to porous silica particles, 5 to 10 µm in diameter.
L46 —Polystyrene/divinylbenzene substrate agglomerated with quaternary amine functionalized latex beads,
about 9 to 11 µm in diameter.
L63 —Glycopeptide teicoplanin linked through multiple covalent bonds to a 100- units spherical silica.
[NOTE—Available as Astec Chirobiotic T from Supelco (www.sigmaaldrich.com). ]
S1C —A support prepared from crushed firebrick and calcined or burned with a clay binder above 900 with
subsequent acid-wash. It may be silanized.
S1NS —The siliceous earth is untreated.
S2 —Styrene-divinylbenzene copolymer having a nominal surface area of less than 50 m 2 per g and an average
pore diameter of 0.3 to 0.4 µm.
S3 —Copolymer of ethylvinylbenzene and divinylbenzene having a nominal surface area of 500 to 600 m 2 per g
and an average pore diameter of 0.0075 µm.
S4 —Styrene-divinylbenzene copolymer with aromatic –O and –N groups, having a nominal surface area of 400
to 600 m2 per g and an average pore diameter of 0.0076 µm.
S5 —40- to 60-mesh, high-molecular weight tetrafluorethylene polymer.
S6 —Styrene-divinylbenzene copolymer having a nominal surface area of 250 to 350 m2 per g and an average
pore diameter of 0.0091 µm.
S7 —Graphitized carbon having a nominal surface area of 12 m2 per g.
S8 —Copolymer of 4-vinyl-pyridine and styrene-divinylbenzene.
S9 —A porous polymer based on 2,6-diphenyl-p-phenylene oxide.
S10 —A highly polar cross-linked copolymer of acrylonitrite and divinylbenzene.
S11 —Graphitized carbon having a nominal surface area of 100 m2 per g modified with small amounts of
petrolatum and polyethylene glycol compound.
S12 —Graphitized carbon having a nominal surface area of 100 m2 per g.
Fluvastatin Sodium
(floo'' va stat' in soe' dee um).
C24H25FNNaO4 433.45
6-Heptenoic acid, 7-[3-(4-fluorophenyl)-1-(1-methylethyl)-1H-indol-2-yl]-3,5-dihydroxy-, monosodium salt, [R*,S*-
(E)]-(±)-;
Sodium (±)-(3R*,5S*,6E)-7-[3-(p-fluorophenyl)-1-isopropylindol-2-yl]-3,5-dihydroxy-6-heptenoate [93957-55-
2].
DEFINITION
Fluvastatin Sodium contains NLT 98.0% and NMT 102.0% of C24H25FNNaO4, calculated on the anhydrous basis.
IDENTIFICATION
• B. IDENTIFICATION TESTS—GENERAL, Sodium 191 : Meets the requirements for the pyroantimonate precipitation
test
ASSAY
• PROCEDURE
Solution A: Add 20 mL of 25% aqueous tetramethylammonium hydroxide solution to 880 mL of water. Adjust with
about 2.3 mL of phosphoric acid to a pH of 7.2 ± 0.2. Add 100 mL of a mixture of methanol and acetonitrile (3:2).
Solution B: Add 20 mL of 25% aqueous tetramethylammonium hydroxide solution and 80 mL of water to 900 mL of
a mixture of methanol and acetonitrile (3:2). Adjust with about 2.3 mL of phosphoric acid to a pH of 7.2 ± 0.2.
0 60 40
6 60 40
20 18 82
20.1 60 40
25.1 60 40
System suitability solution: 0.5 mg/mL of fluvastatin sodium from USP Fluvastatin for System Suitability RS,
dissolved first in Solution B, using 40% of the final volume, then diluted with Solution A to volume
Standard solution: 0.5 mg/mL of USP Fluvastatin Sodium RS, dissolved first in Solution B, using 40% of the final
volume, then diluted with Solution A to volume
Sample solution: 0.5 mg/mL of Fluvastatin Sodium, dissolved first in Solution B, using 40% of the final volume,
then diluted with Solution A to volume
Chromatographic system
Column temperature: 35
Flow rate: 3 mL/min
Injection size: 20 µL
[NOTE—Adjust the start time of the gradient step and the equilibration time for each instrument. ]
System suitability
Samples: System suitability solution and Standard solution
[NOTE—The relative retention times for fluvastatin and fluvastatin anti-isomer are about 1.0 and 1.2, respectively. ]
Suitability requirements
Resolution: NLT 1.6 between fluvastatin anti-isomer and fluvastatin, System suitability solution
Column efficiency: NLT 700 theoretical plates for the fluvastatin peak, System suitability solution
Tailing factor: NMT 3.0 for the fluvastatin peak, System suitability solution
Relative standard deviation: NMT 1.0%, Standard solution
Analysis
Samples: Standard solution and Sample solution
Calculate the percentage of C24H25FNNaO4 in the portion of Fluvastatin Sodium taken:
Result = (rU/rS) × (CS/CU) × 100
rU = = peak response from the Sample solution
Fluvastatin
hydroxydienee 2.0 0.92 0.1
Fluvastatin t-butyl ester (fluvastatin related compound B)g 3.4 0.94 0.2
Any other
individual
impurity — 1.0 0.1
a [R*,S*-E]-(±)-7-[3-(4-Fluorophenyl)-1-ethyl-1H-indol-2-yl]-3,5-dihydroxy-6-heptenoic acid monosodium salt.
b [R*,R*-E]-(±)-7-[3-(4-Fluorophenyl)-1-(methylethyl)-1H-indol-2-yl]-3,5-dihydroxy-6-heptenoic acid monosodium
salt.
SPECIFIC TESTS
• PH 791 : 8.0–10.0, in a 10-mg/mL solution. Perform the test immediately after preparation.
• WATER DETERMINATION, Method I 921 : NMT 4.0%; if labeled as a hydrated form, NMT 12.0%. [NOTE—For this
monograph, the term “hydrated form” refers to several known forms of fluvastatin sodium that are not
stoichiometric hydrates. ]
ADDITIONAL REQUIREMENTS
• PACKAGING AND STORAGE: Preserve in tight, light-resistant containers, protected from moisture. Store at a
temperature not exceeding 40 .
• LABELING: Where it is a hydrated form, the label so indicates.
Ampicillin
Academy of Applied Pharmaceutical Sciences (AAPS) Page 115
PLC1002 - Manual - Rev 06.0
(am'' pi sil' in).
• INFRARED ABSORPTION 197K : Except that where the specimen under test is the trihydrate, both it and the USP
Ampicillin Trihydrate RS are undried.
ASSAY
• PROCEDURE
[NOTE—The Standard solution and the Sample solution should be analyzed immediately after preparation. ]
Solution A: 6.54 mg/mL of monobasic potassium phosphate and 0.34 mg/mL of dibasic potassium phosphate,
adjusted with 1 N sodium hydroxide or 1 N phosphoric acid to a pH of 5.5 before final dilution
Solution B: Acetonitrile and Solution A (2:23)
Solution C: Acetonitrile and Solution A (3:7)
Mobile phase: See gradient table below.
0 100 0
6 100 0
15 0 100
16 0 100
18 100 0
20 100 0
Solution D: 46.3 mg/mL of monobasic potassium phosphate and 27.8 mg/mL of dibasic potassium phosphate,
adjusted with 1 N sodium hydroxide or 1 N phosphoric acid to a pH of 6.5 before final dilution
Relative Acceptance
Retention Criteria,
Name Time NMT (%)
Ampicillin 1.0 —
• STERILITY TESTS 71 : Where the label states that Ampicillin is sterile, it meets the requirements when tested as
directed for Test for Sterility of the Product to be Examined, Membrane Filtration, except to dissolve 6 g in 800 mL
of Fluid D containing sufficient sterile penicillinase to inactivate the ampicillin and to swirl the vessel until solution
is complete before filtering.
• PH 791 : 3.5–6.0
Sample solution: 10 mg/mL
• WATER DETERMINATION, Method I 921 : NMT 2.0% where it is labeled as Ampicillin (anhydrous); between 12.0%
and 15.0% where it is labeled as Ampicillin (trihydrate)
• BACTERIAL ENDOTOXINS TEST 85 : Where the label states that Ampicillin is sterile or must be subjected to further
processing during the preparation of injectable dosage forms, it contains NMT 0.15 USP Endotoxin Unit/mg of
ampicillin.
USP Amoxicillin RS
USP Ampicillin RS
Packaging and storage— Preserve in Containers for Sterile Solids as described under Injections 1 .
USP Ampicillin RS
USP Endotoxin RS
USP Sulbactam RS
Constituted solution— At the time of use, it meets the requirements for Constituted Solutions under Injections
1 .
Identification— The retention times of the major peaks in the chromatogram of the Assay preparation
correspond to those in the chromatogram of the Standard preparation, as obtained in the Assay.
BACTERIAL ENDOTOXINS 85 — It contains not more than 0.17 USP Endotoxin Unit in a portion equivalent to 1
mg of a mixture of ampicillin and sulbactam (0.67 and 0.33 mg, respectively).
STERILITY 71 — It meets the requirements when tested as directed for Membrane Filtration under Test for
Sterility of the Product to be Examined.
PH 791 : between 8.0 and 10.0, in a solution containing 10 mg of ampicillin and 5 mg of sulbactam per mL.
Other requirements— It meets the requirements for Uniformity of Dosage Units 905 and for Labeling under
Injections 1 .
Assay—
0.005 M Tetrabutylammonium hydroxide— Dilute 6.6 mL of a 40% solution of tetrabutylammonium hydroxide with
water to obtain 1800 mL of solution. Adjust with 1 M phosphoric acid to a pH of 5.0 ± 0.1, dilute with water to 2000
mL, and mix.
Mobile phase— Prepare a filtered and degassed mixture of 0.005 M Tetrabutylammonium hydroxide and
acetonitrile (1650:350). Make adjustments if necessary (see System Suitability under Chromatography 621 ).
Standard preparation— Quantitatively dissolve accurately weighed quantities of USP Ampicillin RS and USP
Sulbactam RS in Mobile phase to obtain a solution having known concentrations of about 0.6 mg of ampicillin per
mL and 0.3 mg of sulbactam per mL. [NOTE—Inject this solution promptly. ]
Resolution solution— Prepare a solution of USP Sulbactam RS in 0.01 N sodium hydroxide containing 0.3 mg per
mL, and allow to stand for 30 minutes. Adjust with phosphoric acid to a pH of 5.0 ± 0.1. Transfer 5 mL of the
solution to a 25-mL volumetric flask, add 4.25 mL of acetonitrile, dilute with 0.005 M Tetrabutylammonium
hydroxide to volume, and mix. Transfer 1 mL of this solution to a second 25-mL volumetric flask, add 15 mg of
USP Ampicillin RS, dilute with Mobile phase to volume, and mix. [NOTE—Inject this solution promptly. ]
Chromatographic system (see CHROMATOGRAPHY 621 )— The liquid chromatograph is equipped with a 230-
nm detector and a 4-mm × 30-cm column containing packing L1. The flow rate is about 2 mL per minute.
Chromatograph the Resolution solution, and record the responses as directed for Procedure: the relative
retention times are about 0.7 for ampicillin and 1.0 for sulbactam alkaline degradation product; and the resolution,
R, between ampicillin and sulbactam alkaline degradation product is not less than 4.0. Chromatograph the
Standard preparation, and record the responses as directed for Procedure: the relative retention times are about
0.35 for ampicillin and 1.0 for sulbactam; the column efficiency determined from the sulbactam peak is not less
than 3500 theoretical plates; the tailing factor is not more than 1.5; and the relative standard deviation for
replicate injections is not more than 2.0%.
Procedure— Separately inject equal volumes (about 10 µL) of the Standard preparation and the appropriate
Assay preparation into the chromatograph, record the chromatograms, and measure the areas for the major
peaks. Calculate the quantities, in µg, of ampicillin (C16H19N3O4S) and of sulbactam (C8H11NO5S) in the portion of
Ampicillin and Sulbactam for Injection taken by the same formula:
(CSP / CU)(rU / rS)
in which CS is the concentration, in mg per mL, of the appropriate USP Reference Standard in the Standard
preparation; P is the assigned content, in µg per mg, of the appropriate USP Reference Standard; CU is the
concentration, in mg per mL, of Ampicillin and Sulbactam for Injection in Assay preparation 1, based on the
weight, in mg, of powder removed from the container and the extent of dilution; and rU and rS are the peak areas
for the appropriate analyte obtained from Assay preparation 1 and the Standard preparation, respectively.
Calculate the quantities of ampicillin (C16H19N3O4S) and of sulbactam (C8H11NO5S) withdrawn from the container,
or in the volume of constituted solution taken by the same formula:
(L / D)(CSP)(rU / rS)
in which L is the labeled quantity, in mg, of ampicillin or sulbactam, as appropriate, in the container or in the
volume of constituted solution taken; D is the concentration, in mg per mL, of ampicillin or sulbactam in Assay
preparation 2 or Assay preparation 3, on the basis of the labeled quantity, in mg, of ampicillin or sulbactam, as
appropriate, in the container and the extent of dilution; rU and rS are the peak areas for the appropriate analyte
obtained from Assay preparation 2 or Assay preparation 3 and the Standard preparation, respectively; and the
other terms are as defined above.
C62H111N11O12 1202.61
Cyclo[[(E)-(2S,3R,4R)-3-hydroxy-4-methyl-2-(methylamino)-6-octenoyl]-L-2-aminobutyryl-N-methylglycyl-N-
methyl-L-leucyl-L-valyl-N-methyl-L-leucyl-L-alanyl-D-alanyl-N-methyl-L-leucyl-N-methyl-L-leucyl-N-methyl-L-valyl].
[R-[R*,R*-(E)]]-Cyclic(L-alanyl-D-alanyl-N-methyl-L-leucyl-N-methyl-L-leucyl-N-methyl-L-valyl-3-hydroxy-N,4-
dimethyl-L-2-amino-6-octenoyl-L- -aminobutyryl-N-methylglycyl-N-methyl-L-leucyl-L-valyl-N-methyl-L-leucyl)
[59865-13-3].
» Cyclosporine contains not less than 98.5 percent and not more than 101.5 percent of cyclosporine A
(C62H111N11O12), calculated on the dried basis.
Packaging and storage— Preserve in tight, light-resistant containers.
USP Cyclosporine RS
USP Cyclosporine Resolution Mixture RS
This material is a 100:1 mixture of cyclosporine and cyclosporine U.
Identification— The chromatogram of the Assay preparation obtained as directed in the Assay exhibits a major
peak for cyclosporine, the retention time of which corresponds to that exhibited in the chromatogram of the
Standard preparation, as obtained in the Assay.
LOSS ON DRYING 731 — Dry about 100 mg, accurately weighed, in a capillary-stoppered bottle in vacuum at a
pressure not exceeding 5 mm of mercury at 60 for 3 hours: it loses not more than 2.0% of its weight.
Chromatographic system (see CHROMATOGRAPHY 621 )— The liquid chromatograph is equipped with a 210-
nm detector, a 0.25-mm × 1-m stainless steel tube connected to a 4-mm × 25-cm column that contains 3- to 5-µm
packing L1. The tube and column are maintained at 80 . The flow rate is about 1.2 mL per minute.
Chromatograph the Resolution solution, and record the responses as directed for Procedure: the cyclosporine U
peak and the main cyclosporine peak are resolved from each other. Chromatograph Standard preparation 1, and
record the responses as directed for Procedure: the relative standard deviation for replicate injections is not more
than 1.0%. Chromatograph Standard preparation 2, and record the responses as directed for Procedure: the
relative standard deviation for replicate injections is not more than 10%.
Procedure— [NOTE—Use peak areas where peak responses are indicated. ] Separately inject equal volumes
(about 20 µL) of Standard preparation 1, Standard preparation 2, and the Assay preparation into the
chromatograph, record the chromatograms, and measure the peak responses. Calculate the percentage of
cyclosporine A (C62H111N11O12) in the Cyclosporine taken by the formula:
(CP/10U)(rU / rS)
in which C is the concentration, in mg per mL, of USP Cyclosporine RS in Standard preparation 1; P is the
specified purity, in µg per mg, of USP Cyclosporine RS; U is the concentration, in mg per mL, of specimen in the
Assay preparation; and rU and rS are the main cyclosporine peak responses obtained from the Assay preparation
and Standard preparation 1, respectively.
Auxiliary Information— Please check for your question in the FAQs before contacting USP.
Topic/Question Contact Expert Committee
DISSOLUTION 711 —
WHERE CAPSULES CONTAIN LIQUID—
Medium: water; 500 mL.
Apparatus 2: 50 rpm.
Time: 15 minutes.
Procedure— Place 1 Capsule in each vessel, and allow the Capsule to sink to the bottom of the vessel before
starting rotation of the blade. Observe the Capsules, and record the time taken for each Capsule shell to rupture.
Tolerances— The requirements are met if all of the Capsules tested rupture in not more than 15 minutes. If 1 or 2
of the Capsules rupture in more than 15 but not more than 30 minutes, repeat the test on 12 additional Capsules.
Not more than 2 of the total of 18 Capsules tested rupture in more than 15 but not more than 30 minutes.
WHERE CAPSULES CONTAIN POWDER—
Medium: 0.1 N hydrochloric acid containing 0.5% of sodium lauryl sulfate; 1000 mL.
Apparatus 1: 150 rpm.
Time: 90 minutes.
Determine the amount of C62H111N11O12 dissolved by employing the following method.
Mobile phase— Prepare a filtered and degassed mixture of acetonitrile, water, methanol, and phosphoric acid
(900:450:50:0.5). Make adjustments if necessary (see System Suitability under Chromatography 621 ).
Standard solution— Quantitatively dissolve an accurately weighed quantity of USP Cyclosporine RS in
Dissolution Medium to obtain a solution having a known concentration of about 0.001L mg per mL, L being the
labeled quantity, in mg, of cyclosporine in each Capsule. Transfer 25.0 mL of this solution to a 50-mL volumetric
flask, dilute with acetonitrile to volume, and mix. This solution contains about 0.0005L mg of USP Cyclosporine
RS per mL.
Test solution— Filter a portion of the solution under test. Transfer 5.0 mL of the filtrate to a 10-mL volumetric
flask, dilute with acetonitrile to volume, and mix.
Chromatographic system (see CHROMATOGRAPHY 621 )— The liquid chromatograph is equipped with a 210-
nm detector and a 4.6-mm × 25-cm column that contains packing L1 and is maintained at a constant temperature
of about 80 . The flow rate is about 2 mL per minute. Chromatograph the Standard solution, and record the peak
areas as directed for Procedure: the column efficiency is not less than 700 theoretical plates; and the relative
standard deviation for replicate injections is not more than 2.0%.
Procedure— Separately inject equal volumes (about 10 µL) of the solution estimated to contain 0.1 mg of
cyclosporine per mL, or 40 µL of the solution estimated to contain 0.025 mg of cyclosporine per mL) of the
Standard solution and the Test solution into the chromatograph, record the chromatograms, and measure the
areas for the major peaks. Calculate the quantity, in mg, of C62H111N11O12 dissolved by the formula:
2000C(rU / rS)
in which C is the concentration, in mg per mL, of USP Cyclosporine RS in the Standard solution; and rU and rS are
the cyclosporine peak areas obtained from the Test solution and the Standard solution, respectively.
Tolerances— Not less than 80% (Q) of the labeled amount of C62H111N11O12 is dissolved in 90 minutes.
Academy of Applied Pharmaceutical Sciences (AAPS) Page 126
PLC1002 - Manual - Rev 06.0
UNIFORMITY OF DOSAGE UNITS 905 : meet the requirements.
WATER, Method I 921 — For Capsules that contain powder, not more than 3.5% is found, using finely ground
Capsule contents.
Assay—
WHERE CAPSULES CONTAIN LIQUID—
Mobile phase and Chromatographic system— Proceed as directed in the Assay under Cyclosporine Injection.
Standard preparation— Dissolve an accurately weighed quantity of USP Cyclosporine RS in dehydrated alcohol
to obtain a solution having a known concentration of about 1 mg per mL. Use this solution promptly after
preparation.
Assay preparation— Using a sharp blade, carefully cut open not fewer than 20 Capsules, and with the aid of
dehydrated alcohol transfer the contents of the Capsules to a suitable volumetric flask. Wash the blade with
dehydrated alcohol, and transfer the washings to the volumetric flask. Dilute the contents of the volumetric flask
with dehydrated alcohol to volume, and mix. Quantitatively dilute an accurately measured volume of this solution
with dehydrated alcohol to obtain a solution having a concentration of about 1 mg of cyclosporine per mL.
Procedure— Separately inject equal volumes (about 20 µL) of the Standard preparation and the Assay
preparation into the chromatograph, record the chromatograms, and measure the areas for the major peaks.
Calculate the quantity, in mg, of cyclosporine (C62H111N11O12) in each Capsule taken by the formula:
(L/D)(CP/1000)(rU / rS)
in which L is the labeled quantity, in mg, of cyclosporine in each Capsule taken; D is the concentration, in mg per
mL, of the Assay preparation, based on the labeled quantity of cyclosporine in the Capsules taken and the extent
of dilution; C is the concentration, in mg per mL, of USP Cyclosporine RS in the Standard preparation; P is the
purity, in µg per mg, of USP Cyclosporine RS; and rU and rS are the peak areas obtained from the Assay
preparation and the Standard preparation, respectively.
WHERE CAPSULES CONTAIN POWDER—
Mobile phase— Prepare a filtered and degassed mixture of acetonitrile, water, methanol, and phosphoric acid
(605:400:50:0.5). Make adjustments if necessary (see System Suitability under Chromatography 621 ).
Diluting solvent— Prepare a mixture of acetonitrile, tetrahydrofuran, and dehydrated alcohol (9:5:4).
Standard preparation— Transfer about 25 mg of USP Cyclosporine RS, accurately weighed, to a 25-mL
volumetric flask. Add 2.5 mL of water, and sonicate for 10 minutes. Add about 10 mL of Diluting solvent, sonicate
for 5 minutes, dilute with Diluting solvent to volume, and mix.
Assay stock preparation— Transfer the contents of 20 Capsules to a volumetric flask of such capacity, V, in mL,
to make a final concentration of 10 mg of cyclosporine per mL. Add 0.1V mL of water to the flask, and sonicate for
10 minutes. Add 0.4V mL of Diluting solvent to the flask, and sonicate for 5 minutes. Dilute with Diluting solvent to
volume, and mix.
Assay preparation— Transfer 5.0 mL of Assay stock preparation to a 50-mL volumetric flask, add 5 mL of water,
dilute with Diluting solvent to volume, and mix.
Chromatographic system (see CHROMATOGRAPHY 621 )— The liquid chromatograph is equipped with a 210-
nm detector and a 4.6-mm × 25-cm column that contains packing L13 and is maintained at a constant
temperature of about 70 . The flow rate is about 2 mL per minute. Chromatograph the Standard preparation, and
record the peak areas as directed for Procedure: the column efficiency is not less than 700 theoretical plates; the
tailing factor is not more than 1.5; and the relative standard deviation for replicate injections is not more than
2.0%.
Procedure— Separately inject equal volumes (about 20 µL) of the Standard preparation and the Assay
preparation into the chromatograph, record the chromatograms, and measure the areas for the major peaks.
Calculate the quantity, in mg, of cyclosporine (C62H111N11O12) in each Capsule taken by the formula:
The relative polarity of a solvent is a useful guide to solvent selection in partition chromatography. The relative
polarities of the listed solvents may differ slightly depending on the literature source, since the scale used to
measure polarity may be different. The following should suffice as a general reference for relative solvent polarity.
Polarity is increasing from top to bottom.
• Fluoroalkanes (least polar)
• Isooctane
• Hexane
• Carbon tetrachloride
• Toluene
• Chloroform
• Methylene chloride
• Diethyl ether (ether)
• Ethyl acetate
• Dioxane
• Acetone
• Tetrahydrofuran (THF)
• Isopropanol
• Ethanol
• Acetonitrile
• Methanol
• Acetic acid
• Water
• Buffer Solutions (most polar)
The various analytes to be separated may also be arranged based on the polarities of their functional groups. A
general guide to relative solute polarity going from non-polar to the most polar group is as follows:
• Hydocarbons (least polar)
• Amides
• Ethers
• Esters
• Ketones
• Aldehydes
• Amines
• Alcohols
• Carboxylic Acids
A
Activity
In adsorption chromatography, this is the relative strength of the surface of the packing. For
silica gel, the more exposed the silanol groups, the more active the surface. Activity can be
controlled by adding water or another polar modifier, which is hydrogen bonded to the active
sites, thereby reducing the surface activity.
Adsorbent
Packing used in adsorption chromatography. Silica gel and alumina are the most frequently used
adsorbents in HPLC.
Adsorption
The process of interaction between the solute and the surface of an adsorbent. The forces
involved can be strong such as hydrogen bonds, or weak such as van der Waals forces. For silica
gel, the silanol group is the driving force for adsorption, and any solute functional group that can
interact with this group can be retained by liquid-solid chromatography on silica.
Adsorption chromatography
One of the basic LC modes which relies on the adsorption process to effect the separation. Silica
gel and alumina are the most frequently used supports. The molecules are retained by the
interaction of their polar functional groups with the surface functional groups such as silanols of
silica.
Adsorption isotherm
In adsorption chromatography, this is a plot of the equilibrium concentration of sample in the
mobile phase per unit volume verses the concentration in the stationary phase per unit weight.
The shape of the adsorption isotherm can determine the chromatographic behavior of the solute
such as tailing, fronting, and sample overload.
Affinity chromatography
A technique in which a biospecific adsorbent is prepared by coupling a specific ligand (such as
an enzyme, antigen, or hormone) for the macromolecule of interest to a solid support (or
carrier). This immobilized ligand will interact only with molecules that can selectively bind to it.
Molecules that will not bind elute unretained. The retained compound can later be released in a
purified state. Affinity chromatography is not a chromatographic technique but selective
filtration.
Alumina
An adsorbent sometimes used in adsorption chromatography. Aluminum oxide (AI203) is a
porous adsorbent that is available with a slightly basic surface. For this reason, it can have
advantages over silica, which is considered to have an acidic surface.
Amino phase
A propylamino stationary phase used mostly in normal bonded phase chromatography. It is
somewhat reactive for any solute molecule or mobile phase additive that can react with amines.
The amino phase has found some applications as a weak anion exchanger.
Asymmetry
A factor describing the shape of a chromatographic peak. Theory assumes a Gaussian shape
peak that is symmetrical. The peak asymmetry factor is the ratio (at 10 percent of the peak
height) of the distance between the peak apex and the back side of the chromatographic curve to
the distance between the peak apex and the front side of the chromatographic curve. A value >1
is a tailing peak, while a value <1 is a fronting peak.
E
Eddy diffusion term
This is represented by the A term in the van Deemter equation. It is the contribution to plate
height that is due to molecules traveling along different paths through the column and it
depends on the particle size and geometry of the packing. It is also called the multipath term.
See van Deemter equation.
Effective theoretical plates (Neff)
The true number of plates in a column that corrects the number of theoretical plates for dead
volume. , where tm is the void time.
Efficiency (N or H)
A measure determined by the number of theoretical plates calculated from the equation:
This can be visually represented by the following figure.
Measure peak width at 4.4% of peak height
Electrochemical detector
Al detector which monitors an oxidation or reduction reaction between the detector’s working
electrode and the sample. Electrochemical detectors are characterized by high sensitivity, with
detection limits in the low ppb range common for electroactive compounds.
Eluate
Combination of mobile phase and solute exiting column.
Eluent
Mobile phase used to carry out a separation.
Eluotropic series
A series of solvents with an increasing degree of polarity, generally used to explain solvent
strength in liquid-solid or adsorption chromatography. A nonpolar solvent such as pentane
would be at one end of the scale; dichloromethane would be an intermediate solvent; a strongly
polar solvent, such as water, would be at the other end of the scale. Thus, when developing a
method or running a gradient, an eluotropic series is useful for selecting solvents.
Elution
Guard column
A small column placed between the injector and the analytical column. It protects the analytical
column against contamination by sample particulates, and by strongly retained species. The
guard column is usually packed with the same material as the analytical column and is often of
the same i.d. It is much shorter, costs less, and is usually discarded when it becomes
contaminated.
H
HETP
The height equivalent of a theoretical plate. It is a carryover from distillation theory which is a
measure of a column’s efficiency. For a typical HPLC column packed with 5 mm particles,
HETP (or H) values are usually between 0.01 and 0.03 mm. HETP = L/N, where L is column
length, and N is the number of theoretical plates.
Hydrophilic
It is often referred to as water loving. It adverts both to water compatible stationary phases, and
to water soluble molecules. Most columns used to separate proteins are hydrophilic in nature
and should not sorb or denature protein in the aqueous environment.
Hydrophobic
It is often referred to as water hating. It adverts both to stationary phases not compatible with
water and molecules with little affinity for water. Hydrophobic molecules have few polar
functional groups and are mostly hydrocarbons or have high hydrocarbon content.
Hydrophobic interaction chromatography
A technique in which reversed-phase packings are used to separate molecules by the
interactions between their hydrophobic moieties and the hydrophobic sites on the surface. High
salt concentrations are used in the mobile phase and separations are effected by changing the
salt concentration. The technique is analogous to "salting out" molecules from solution.
Gradients are run by decreasing the salt concentration over time.
I
Injector
A mechanism for accurately injecting a predetermined amount of sample into the mobile phase
stream. The injector can be a simple manual device, or a sophisticated autosampler that permits
automated injections of many different samples for unattended operation.
In-line filter
A device that prevents particulate matter from damaging the column by filtration. Modem in-
line filters can be placed between the injector and the column without contributing to band
broadening. A filter in this position is used to prevent sample particulates from entering the
packed bed or the inlet frit.
Inlet
The initial part of the column, where the solvent and sample enter. There is usually an inlet frit
that holds the packing in place and, in some cases, protects the packed bed.
Internal standards
Internal standards consist of a specific quantity of a compound that is known not to be in the
sample, but that exhibits the same characteristics under the separation conditions as the sample
components. Internal standards are used primarily to calibrate quantitation, especially for
Isocratic
A constant composition mobile phase used in liquid chromatography.
K
k or k'
The capacity factor. It can be calculated from the equation, where tR is retention time for the
sample peak, and to is the retention time for an unretained peak.
L
Ligand
In ligand-exchange chromatography, it is the molecule added to the mobile phase that acts as
the chelating agent. In affinity chromatography, it is the biospecific material (enzyme, antigen,
or hormone) coupled to the support (carrier) to form the affinity column.
Ligand-exchange chromatography
A technique in which chelating ligands are added to the mobile phase and undergo sorption
onto a packing. These sorbed molecules, can act as chelating agents with solutes. For example,
using of copper amine chelates for the separation of amino acids. Chelating resins function in a
similar manner. The chelating groups are chemically bonded to the polystyrene backbone.
Linearity
A measurement process which ensures accurate quantitation. In chromatography it is important
that the detector provide a linear response over the range of sample concentrations it
encounters.
Linear elution adsorption chromatography (LEAC)
A term coined by Lloyd Snyder, which refers to adsorption chromatography carried out in the
linear portion of an adsorption isotherm.
Linear velocity(u)
A measure of the speed with which an unretained compound moves through the column. Linear
velocity is represented by the letter u and is reported in cm/min or mm/sec. Linear velocity is
used to calculate flow rate based on the cross sectional area of a column. Linear velocity is
useful for adapting methods from one column diameter to another.
Liquid-liquid chromatography (LLC)
This is the same as Partition chromatography. It was the earliest form of HPLC, which was
replaced with chemically bonded phases in the early 1970s.
Liquid-solid chromatography (LSC)
Same as Adsorption chromatography.
Loading
The amount of stationary phase coated or bonded onto a solid support. In liquid-liquid
chromatography, it is the milligram amount of liquid phase per gram of packing. In BPC, the
loading may be expressed in mmol/m2 or in percent C. See Coverage.
M
Macroporous resin
Cross-linked ion-exchange resins that have both micropores of molecular dimensions and
macropores several hundred Å wide. These are highly porous resins with large internal surface
areas accessible to large molecules.
Quantitation
An analysis process which is designed to determine the amounts or proportion of the
components of a substance.
R
Radial compression
The use of radial pressure applied to a flexible wall column to lessen wall effects.
Radial diffusion
Diffusion across the LC column in a radial direction. If the sample is injected into the exact
center of a column, it will spread not only in a longitudinal direction as it flows, but radially as
well.
Recovery
The amount of solute (sample) that elutes from a column relative to the amount injected. This is
most often used with protein separations in which proteins "hang up" on active sites of the
packing in certain columns.
Recycling
A technique in which the column effluent is recirculated onto the head of the column in an
attempt to gain the advantage of an extended column length. It can be carried out on a single
column by passing the effluent back through the pump. An alternative technique uses two
columns connected by a switching valve. This technique is very seldom used in HPLC, and
only in size-exclusion chromatography.
Reduced plate height (h)
A calculated value used to measure efficiencies of columns. A BPLC column with an h value
<2 is considered to be well packed.
Reduced velocity (v)
A calculated value used to compare different chromatographic columns. It relates the solute
diffusion coefficient (Dm) in the mobile phase to the particle size of the column packing (dp).
Regeneration
A process of restoring the packing in the column to its initial state after a gradient elution. The
mobile phase is passed through the column step-wise or in a gradient solvating the stationary
phase to its original condition. In ion-exchange chromatography, regeneration involves
replacing ions taken up in the exchange process with the original ions that occupied the
exchange sites. Regeneration can also refer to bringing back any column to its original state by
removing impurities with a strong solvent.
Relative retention
Relative retention is a measure of the difference of affinities of two compounds for the
stationary phase. Mathematically it is the ratio of the retention factors of two compounds, one
of which is usually a standard. It is represented by the letter r, and is reported in dimensionless
units. Relative retention is used in quality control, reproducibility, and method validation
calculations.
Relative retention ratio
Same as Separation factor or Selectivity.
Residual silanols
The silanol (SiOH) groups that remain on the surface of a packing after a phase is chemically
S
Sample capacity
The amount of sample that can be injected onto an LC column without overload. It is often
expressed as grams of sample per gram of packing. Overload is defined as the sample mass
injected at which the column efficiency falls to 90 percent of its normal value.
Stationary phase
The immobile phase involved in the chromatographic process. The stationary phase in liquid
chromatography can be a solid, a bonded or coated phase on a solid support, or a wall coated
phase. The stationary phase used often characterizes the LC mode. For example, silica gel is
used in adsorption chromatography, an octadecylsilane bonded phase in reversed-phase
chromatography, etc.
Stepwise elution
This process uses eluents of different compositions during the chromatographic run. These
eluents are added in a stepwise manner with a pump, or by a selector valve. Gradient elution
incorporates continuous changing of solvent composition.
Steric exclusion chromatography (SEC)
A major LC mode in which samples are separated by virtue of their size in solution. Also
known as size-exclusion, gel permeation, gel filtration, or gel chromatography.
Straight-phase chromatography
Same as Normal-phase chromatography.
Supercritical fluid chromatography (SFC)
A technique that uses a supercritical fluid as the mobile phase. The technique has been applied
to the separation of substances that cannot be handled effectively by liquid chromatography
(because of detection problems) or gas chromatography (because of the lack of volatility).
Examples are separations of triglycerides, hydrocarbons, and fatty acids. GC detectors and
HPLC pumps have been used together in SFC.
Support
The solid particles in a column. The support can be naked, coated, or can have a chemically
bonded phase in HPLC.
Suppressed IC
A second column is used to remove the buffer ions so that sample ions can be more easily
detected; membrane separator is sometimes used.
Suppressor column
In ion chromatography, it is the column placed after the ion-exchange column. Its purpose is to
remove or suppress the ionization of buffer ions so that sample ions can be observed in a
weakly conducting background with a conductivity detector.
Surface area
In an adsorbent, it is the total area of the solid surface as determined by an accepted
measurement technique such as the BET method using nitrogen adsorption. The surface area of
a typical porous adsorbent such as silica gel can vary from 100 to 600 m2/g.
Surface coverage
The mass of stationary phase per unit area of an LC support. It is often expressed in mmol/m2
of surface. Sometimes the percent C is given as an indicator of surface coverage.
Swelling
A process in which resins and gels increase their volume because of their solvent environment.
mmddyy = the date the project was created in the format of Month
(mm - 01) Day (dd - 12) and Year (yy - 07) = 011207 = January 12,
2007.
Enter “New” for the Audit Trail Comment and click “Finish”
Click on the blue triangle & Tap in the lower centre of the screen.
** Note – The “samples” tab allows for a sample set…. a group of injects from
different vials. Your instructor will assist you with sample sets in later sessions. **