Checklist of Radiology and Imaging Department and Its Quality Indicators For NABH Accreditation Preparation
Checklist of Radiology and Imaging Department and Its Quality Indicators For NABH Accreditation Preparation
Checklist of Radiology and Imaging Department and Its Quality Indicators For NABH Accreditation Preparation
Below is the compilation of all requirements that is applicable to an imaging department for
NABH accreditation fulfillment
1. Imaging must be able to provide those tests that are required by various clinical specialities
offered by the hospital. For this hospital must identify various imaging tests required by different
specialities and same should be made available.
2. A list of all imaging tests that can be done in the department should be maintained and
updated from time to time. This list will be the scope of imaging services
3. In case some tests cannot be offered or is not available, an arrangement should be there to
outsource such tests. In this case following things are required to be ensured
1. A list of all tests that can be outsourced in available in the imaging department, specifying
which lab (or labs) the specific test can be outsourced to
2. In each of the imaging lab where the test is being outsourced, identity of the personnel from
that lab has to be written, so that accountability can be fixed
3. A standard process should be there on how to outsource the test. The standard process must
describe, the method of patient identification, test requisition details, communication with
outsourced lab and other relevant details
4. The selection of labs where the tests will be outsourced should be based upon criteria of
quality and safety. The criteria should be documented.
5. A written MoU should be signed with the outsourced lab, specifying the terms and conditions
related to quality assurance requirements
1. Layout approval from AERB of each rooms that houses radiation emitting equipment such as
X-ray and CT-scan
5. Registration of all ultrasonography machines under PNDT act. (Check this post on 'all legal
documents required by a hospital')
5. Standard operating process for carrying out each test procedure should be available
7. Informed consent should be taken from patient before any imaging test
9. A declaration to be taken from pregnant lady undergoing USG, for non-declaration of sex
10. A documented plan of peer review should be available and records related to peer review is
available. In peer review, randomly selected samples of the test results should be interpreted by
peers and findings should be correlated.
11. A randomly selected samples of test results should also be correlated with the clinical
findings and recorded.
12. Minutes of meeting (or any other similar records) should be available to show that peer
review findings and surveillance findings has been communicated, discussed, analysed and
actions has been taken
13. Each equipment in imaging should have a calibration scheduled. This can be documented in
the equipment log. A calibration report / label on equipment should be there as an evidence that
calibration has been done, as per schedule
14. Each equipment in imaging should have a preventive maintenance schedule. This can be
documented in the equipment log. A record of preventive maintenance check carried out should
be available.
15. A register documenting all corrective and preventive action taken should be available
2. The lead aprons should have been checked at-least once in six months to ensure that they have
not formed crevices. This can be done by taking an X-ray image of lead apron and examining it.
These images and the findings should be maintained as a record
3. Each staff working within radiation areas must have his/her own TLD badges, which they are
wearing when on duty. The TLD badged should be worn under lead apron and not over it
5. Each staff should have been trained and are aware of radiation safety practices. Records of
training and evaluation to be maintained
17. For MRI, the safety practices to prevent any hazard from strong magnetic field should be
available and in use. This include measure to prevent any metal item to be taken inside MRI
room.
19. Display and signage should be in place, as per regulatory requirements. These include
21. Radio-active waste disposal, if applicable, in a lead container and as per AERB regulations
5. Occupational health hazard that they are exposed to and how to prevent
7. Other emergency codes used in hospital, such as code red, code pink etc.
10. Their role during disaster situation. (Suggested read - Disaster handling checklist for
hospitals)
12. Turn-around time and other quality indicators being used in imaging
Quality Indicators:
Following quality indicators must be used in imaging as a part of quality assurance programme.
Each doctor and staff should be aware of these indicator
6. % of re-testing required