CALIBRATION CERTIFICATE

CERTIFICATE NO: GLT/AR/05/017 DATE OF CALIBRATION:5/18/2024


DATE OF ISSUE: 23/05/2024 CALIBRATION DUE DATE:5/17/2025
NAME AND ADDRESS OF HOSPITAL:
SAL HOSPITAL
OPP. DOORDARSHAN, DRIVE IN
AHMEDABAD-380054
DETAIL OF INSTRUMENT CONDITION OF UNIT: GOOD
NAME: WEIGHT MACHINE SR NO:
COMPANY: ID. NO: N.A
LOCATION:

AMBIENT TEMPERATURE : 27 + 5 ‘C RELATIVE HUMIDITY: 30 % RHTO 75 % RH

 STANDARD USED FOR CALIBRATION


F1 CLLASS S.S KNOB TYPE WEIGHUSED FOR TS USED WITH REFRENCE STANDARD TRACABLE OF NATIONAL /
INTERNATIONAL STANDARDS FOR 2 YEARS.
 RESULTS:
LINEARITY TEST
DENOMATION STANDARD UUC ERROR DENOMATION STANDARD UUC ERROR
MASS VALUE OBSERVED MASS VALUE OBSERVED
READING READING
1 1.0094943 0.89 0.116494 50 50.09314 50.1678858 -0.074746
5 5.0384007 4.88 0.159401 60 60 60.1089667 -0.108967
10 10.04302 10.23 -0.190980 70 70 70.0756638 -0.075664
20 20.084941 20.89 -0.802059 80 80 79.6545452 0.345455
30 30.084891 30.24 -0.152109 100 100 101.362624 -1.362624
40 40.00012 40.63 -0.632280
REPEATABLITY TEST
MASS VALUE WEIGHT INDICATOR BY UUC (KG) AVERAGE ERROR (KG)
VALU (KG)
100 99.9 100 99.8 100.2 99.8 99.94 0.06

CORNER TEST
POSITION (1) (2) (3) (4) (5)
CENTER FROM LEFT BACK LEFT BACK RIGHT FRONT RIGHT
50.0 50.0 50.0 50.0 50.0
ECCENTRICITY 0.0 0.0 0.0 0.0 0.0

CALIBRATION STANDARD USED


NAME & MAKE RANGE SR.NUMBER CERTIFICATE NO CALIBRATION VELID UPTO
S.S KNOB TYPE WEIGHTS- 1 KG TO 100 KG 4002-2 TO 11 WMCL/E/2021-02/1139 04/04/2025
F1 CLASS AND MAKE (F1 CLASS)
WEIGHTRNONICS
STANDARD USED FOR CALIBRATION ARE TRACEBLE TO INTERNATIONAL STANDARD

NOTE:-
THIS CERTIFICATE RELATES ONLY TO THE PARTICULAR ITEM(S) RECIVED FOR CALIBRATION IN GOOD CONDITION AT SAL HOSPITAL, THALTEJ.
THIS CERTIFICATE SHELL NOT BE RE PRODUCE.RECOMMANDAD DUE DATE IS OF 1 YEAR.
THE VALUE HAS BEEN ROUNDED –OFF AS PER IS 2:1960 WHEVER APPLICABLE
EQUIPMENT CALIBRATED & REPORT PREPARED BY REPORT VARIFIED AND AUTHORISED BY