CALIBRATION CERTIFICATE

CERTIFICATE NO: GLT/AR/05/0023 DATE OF CALIBRATION:5/20/2024


DATE OF ISSUE: 23/05/204 CALIBRATION DUE DATE:5/19/2025
NAME AND ADDRESS OF HOSPITAL:
SAL HOSPITAL
OPP. DOORDARSHAN, DRIVE IN
AHMEDABAD-380054
DETAIL OF INSTRUMENT: CONDITION OF UNIT: GOOD
NAME: MULTIPARA MONITOR SR NO:SKZ15040002WA
MODEL:B-40 ID. NO:N.A
MAKE:G.E LOCATION:MICU

AMBIENT TEMPERATURE : 27 + 5 ‘C RELATIVE HUMIDITY: 30 % RHTO 75 % RH

 STANDARD USED FOR CALIBRATION


ELECTRICAL SAFETYANALYSER (PTCC-1001) USED FOR CALIBRATION IS CALIBRATED WITH REFRENCE STANDARD TRACABLE OF NATIONAL /
INTERNATIONAL STANDARDS FOR 2 YEARS.
 RESULTS:
VISUAL TEST
SR NO TESTS PARAMETER REMARKS
1 DEVICE IS CLEAN AND DE CONTAMINATED PASS
2 NO PHYSICAL DAMAGE TO CASE, DISPLAY, MOUNTS,CART OR PASS
COMPONENT
3 SWITCH AND CONTROLL OPERATABLE AND CORRECTLY ALLIGNED PASS
4 DISPLAY INTENSITY ADEQUET FOR DAY TIME USE PASS
5 POWER CORD, ACCESSORY CABLE & CHARGER PASS
SAFETY TEST
SR NO TESTS PARAMETER PERMITTED VALUE REMARKS
1 VOLTAGE BETWEEN LIVE AND NETURAL (210 V TO 240V) PASS
2 VOLTAGE BETWEEN LIVE AND EARTH (210 V TO 240V) PASS
3 VOLTAGE BETWEEN EARTH AND LIVE (0 TO 5 V) PASS
4 ENCLOSUR CURRENT (0 TO 10 µ A) PASS
5 LEACKAGE CURRENT (0 TO 500µA) PASS
6 PATIENT LEACKAGE CURRENT BF: 0 TO 100 µA PASS
CF: 0 TO 10 µA
FUNCTIONAL PARAMETER
PARAMETER SET VALUE OBSERVED VALUE SET VALUE OBSERVED VALUE TOLERANCE REMARKS

HEART RATE 90 88 180 179 + 5% PASS


(BPM)-NSR 120 118 240 241 + 5% PASS
RESPIRATION 40 39 100 99 + 5% PASS
TEST(Rbpm) 60 59 80 79 + 5% PASS
NIBP(mmhg) 80/80 78/79 120/80 120/82 + 5% PASS
100/60 99/60 150/100 148/97 + 5% PASS
IBP (mmhg) ARTIAL 120/80 118/79 LEFT VENT 120/00 116 + 5% PASS
RADIAL 120/80 119/82 RIGHT VENT 60/60 58 + 5% PASS
SPO2(%) 80 79 95 93 + 5% PASS
90 89 100 99 + 5% PASS
NOTE:-
THIS CERTIFICATE RELATES ONLY TO THE PARTICULAR ITEM(S) RECIVED FOR CALIBRATION IN GOOD CONDITION AT SAL HOSPITAL, THALTEJ. THIS CERTIFICATE
SHELL NOT BE RE PRODUCE.RECOMMANDAD DUE DATE IS OF 1 YEAR.
THE VALUE HAS BEEN ROUNDED –OFF AS PER IS 2:1960 WHEVER APPLICABLE