Include all Schedule (D) Planned Maintenance

PM Schedule Sys_D

Site: ______________________ Month:____________________


System ID#: ________________ Dispatch/Job #: _____________
Service Engineer: ________________________________________

Test
PM Check Pass Fail Comments
Purpose
Image Quality
Confirm that the site has the latest
Service Pack installed on the system by
referring to MR Service Pack Software
Matrix, DOC1667089, available from IQ
the online documentation library. If
additional information is needed, refer
to Install Service Pack
Alignment Light Check IQ
Patch update (Service contract
Software
customers only)
SPT (PM Mode)
NOTE: For software versions of
IQ
SV25.0 R06 and above, this
task can be omitted.
LV Shim (PM Mode) IQ
PM Check (Service Contract
IQ
Customers Only)
Test
PM Check Pass Fail Comments
Purpose
Magnet Room
Patient Blower Fans & Filter Check
Pneumatic Patient Alert System Safety
Patient Handling
Patient Table Checks Safety
Sentinelle Vanguard Breast MRI Table
Check(Only used in Explorer Lift
system with a Sentinelle Vanguard Safety
Breast MRI Table)

Cooling System
Cabinet FAN & Water Line Check
MCS/LCS/BRM Chiller Coolant Level
Check
Check
Check Noise Level and Leakage Check
Clean MCS Heat Exchanger Clean
Clean BRM chiller Clean
PDU
Emergency Off & Stop Circuits/Indicator
Safety
Lights
Computer
Storelog Check
Clean Dust Filter Clean
Magnet
Inspect Cryogen Vent Check
Test
PM Check Pass Fail Comments
Purpose
Perform quarterly service test for MRU.
Safety
(See MRU manual, 5265188.)
Perform yearly service test for MRU.
Safety
(See MRU manual, 5265188.)
Main Disconnect Panel (MDP)
For GE-supplied MDP, check battery
voltage (refer to MDP Manual and Arc
Flash Safety).
NOTE: GE Field Engineer cannot
install, service, or repair the
MDP. Only a licensed Check
electrician is qualified to
perform those tasks. Refer to
MR Service Safety Manual, PN
5452735, for further safety
information.
Verify that customer performed MDP
Check
Power and Ground Connection Check

Schedule D Planned Maintenance.

Date Completed: ___________________________