Professional Documents
Culture Documents
12 Series-30
12 Series-30
MAINTENANCE
Figure 6-5 Pressure Switch (P/N 040694) clockwise to increase or counterclockwise to de-
crease the setting.
Above 100 PSIG (689kPa), the regulator should al-
low pressure to flow into the control chamber of the
Sullicon Control. The Sullicon Control lever should
start to move at this time.
Cycle the control system several times and recheck
all pressure settings.
MINIMUM PRESSURE/CHECK VALVE MAINTE-
NANCE
Refer to Figure 6-6. Minimum pressure/check valve
(P/N 241581) maintenance is quite minimal. The only
part which normally requires replacement is the O-
ring on the piston. To replace this ring, order seal re-
pair kit No. 250020-344. and follow the procedure ex-
plained below.
! WARNING
Before performing maintenance on the valve, be
sure that all pressure has been relieved in the com-
pressor sump, and all downstream pressure has
been vented to the atmosphere. Also be sure that
the components of the compressor are cool to the
touch.
! DANGER
DO NOT touch the electrical contacts, terminal or
leads with any metallic object. Severe electrical
shock may occur.
24