DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Important Precautions
TRANSLATION POLICY
i
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
ii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
iii
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
iv -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
v
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
vi -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
vii
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
viii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
ix
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
x -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
xi
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
xii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
Service Documentation
9900 Innovation Drive (RP-2156)
Wauwatosa, WI 53226, USA.
GE employees should use Post-market Quality Management (PQM) to report service documentation
issues. These issues will then be in the internal problem reporting tool and communicated to the writer.
WARNING Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section in
the Service Manual.
xiii
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE.
GE may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
COPYRIGHTS
© 2018 by General Electric Company Inc. All Rights Reserved.
DECLARATION OF CONFORMITY
The radio equipment used in the LOGIQ E10 (if present) conforms to the Radio Equipment
Directive 2014/53/EU (RED). The declaration of conformity can be found at the following link:
https://www.intel.com/content/www/us/en/support/articles/000007443/network-and-i-o/wireless-
networking.html
Search under Current or Legacy Adapters for: Intel® Dual Band Wireless-AC 8260NGW
Figure WLAN Label for Intel 8620NGW Module
02607-15-02198
Complies with
IDA Standard
DB 02914
HDIF15000519
xiv -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Revision History
xv
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
xvi -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Table of Contents
CHAPTER 1
Introduction
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Important conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Conventions used in book . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 6
Standard hazard icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 7
Product icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
Product Labels on the LOGIQ E10 when used in a veterinary environment 1 - 17
Safety considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 18
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 18
Human safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 18
Mechanical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 21
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 24
Battery safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 25
Label locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 26
Customer assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 33
Contact information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 33
Ultrasound system manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 35
xviii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 2
Site Preparations
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
xix
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 3
LOGIQ E10 Setup
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Setup reminders. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Average setup time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Setup warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 35
Purpose of this section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 35
LOGIQ E10 configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 35
xx -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Setup paperwork . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 40
User Manual(s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 40
Product Locator Card/UDI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 40
xxi
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 4
Functional Checks
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Special Equipment required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Functional checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
General procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 4
Power ON/Boot Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 6
Power SHUT DOWN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 9
Top Console checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 10
Brakes and Direction Lock Functional Checks . . . . . . . . . . . . . . . . . . . . . . . 4 - 14
Operation checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 15
Site Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 28
xxii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 5
Components and Functions (Theory)
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Service Desktop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 10
User Level of the Service Desktop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 10
Front End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
Purpose of this section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
Card Rack General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 13
Back End . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 14
Purpose of this section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 14
ECB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 15
SSD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
Backplane. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Purpose of this section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Backplane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 19
Power distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 21
Purpose of this section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 21
xxiii
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Product manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
User documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
Basic Service documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 29
xxiv -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 6
Service Adjustments
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
xxv
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 7
Diagnostics/Troubleshooting
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Service desktop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 10
Purpose of this section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 10
Disruptive mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 11
Color statuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 11
Licenses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 12
Home . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 14
Utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 19
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 31
Agent Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 32
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 34
EMI prevention/abatement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 35
Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 36
Shortcut keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 39
Test of printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 39
xxvi -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 8
Replacement Procedures
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Contents in this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Cable Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 28
xxviii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
xxix
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
xxx -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 9
Renewal Parts
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Purpose of this chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
List of Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 4
Peripherals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 31
Printers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 31
Digital Video Disc (DVD) Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 31
Internal Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 34
LOGIQ E10 Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 35
Main Power Supply cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 37
Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 39
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 44
Hardware Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 49
xxxii -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CHAPTER 10
Care & Maintenance
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Purpose of Chapter 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Periodic maintenance inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Grounding Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Why do Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 3
Keeping Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 3
Quality Assurance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 3
System maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 7
Preliminary checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 7
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Tools, supplies and equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 8
Functional checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 9
System functional checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 9
Peripheral/option checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Mains cable inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Optional diagnostic checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
View the logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 10
Physical inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 11
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
Air filter cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
Probe maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
Using a Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 12
xxxiv -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 1
Introduction
Section 1-1
Overview
1-1-1 Purpose of this chapter
This chapter describes important issues related to safely servicing the LOGIQ E10. The service provider
must read and understand all the information presented here before installing or servicing the
LOGIQ E10.
Section 1-2
Service manual overview
1-2-1 Attention
This manual contains necessary and sufficient information for the Field Service Engineer or Biotech
Engineer to safely maintain and service the LOGIQ E10. Advanced equipment training may be provided
by factory trained Field Service trainers for the agreed-upon time period.
This service manual provides installation and service information for the LOGIQ E10. It is divided into
10 chapters as shown in Table 1-1 "Contents in this service manual" on page 1-3.
In the beginning of the manual, before chapter 1, you will find the Revision overview, the Important
precautions including Translation policy, Damage in transportation, Certified electrical contractor
statement, Omission & errors, Service safety considerations and Legal notes, and the Table of
Contents (TOC).
CHAPTER
NUMBER CHAPTER TITLE DESCRIPTION
1 Introduction Contains a content summary and warnings.
4 Functional Checks Contains functional checks that must be performed as part of the
installation, or as required during servicing and periodic
maintenance.
5 Components and Functions (Theory) Contains block diagrams and functional explanations of the
electronics.
7 Diagnostics/Troubleshooting Provides procedures for running diagnostic or related routines for the
LOGIQ E10.
8 Replacement Procedures Provides disassembly procedures and reassembly procedures for all
changeable FRUs, available option installation instructions, and
upgrade installation instructions.
9 Renewal Parts Contains a complete list of replacement parts for the LOGIQ E10.
10 Care & Maintenance Provides periodic maintenance procedures for the LOGIQ E10.
NOTE: Dates on screenshots are represented in MM/DD/YYYY format throughout the manual.
Information on how to change the LOGIQ E10’s date can be found in Chapter 10, Customizing
Your System, of the LOGIQ E10 User Manual.
The LOGIQ E10 is a high performance digital ultrasound imaging system with total data management.
The fully digital architecture of the LOGIQ E10 allows optimal usage of all scanning modes and probe
types throughout the full spectrum of operating frequencies.
Signal flows from the Probe Connector Panel through the Ultrasound system and finally to the monitor
and peripherals.
All necessary software is loaded from the hard drive on power up.
The user manual should be fully read and understood before operating the LOGIQ E10.
The online versions of the user manual is available through the Help function on the LOGIQ E10’s
operator panel.
User manual documentation is provided through electronic Instructions for Use (eIFU).
Section 1-3
Important conventions
Contents in this section
This manual covers the LOGIQ E10 models listed in 1-2-4 "LOGIQ E10 models covered in this manual"
on page 1-4.
Icons
Pictures, or icons, are used wherever they will reinforce the printed message. The icons, labels, and
conventions used on the product and in the service information are described in this chapter.
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the precautionary
message. Known or potential hazards to personnel are labeled in one of three ways:
• DANGER
• WARNING
• CAUTION
When a hazard is present that can cause property damage, but has absolutely no personal injury risk,
a NOTICE is used.
CAUTION CAUTION IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL OR CAN CAUSE
MINOR PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE IGNORED.
EQUIPMENT DAMAGE POSSIBLE.
NOTICE Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
NOTE: Notes are used to provide important information about an item or a procedure.
NOTE: Be sure to read the notes; the information contained in a note can often save you time or effort.
Important information will always be preceded by the exclamation point contained within a triangle,
as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be
used to make you aware of specific types of hazards that could possibly cause harm. Even if a symbol
is not used in this manual, it may be included for your reference.
Icon Meaning
Biological
Electrical
Moving
Acoustic output
Explosion
Smoke/fire
Heat
Icon Meaning
Pinch point
Table 1-4 Standard Icons that indicate that a special procedure is to be used
Icon Meaning
Signed Date
NOTE: This machine should be used in compliance with law. Some jurisdictions restrict certain uses,
such as gender determination (shown for country specific label).
Monitor Label:
DO NOT push the system. Use the handle to push/pull the system,
e.g., DO NOT use the monitor. Failure to do so may cause serious
injury or system damage. Standard: ISO 7010-P017.
DO NOT place a finger, hand or any object on the joint of the monitor
or monitor arm to avoid injury when moving the monitor and monitor
arm.
Chapter 1 Introduction 1 - 11
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
10. RoHS Label–China systems Indicates the presence of hazardous substance(s) above the
only (shown for country maximum concentration value. Maximum concentration values for
specific label) electronic information products, as set by the People’s Republic of
China Electronic Industry Standard SJ/T11364-2006, include the
hazardous substances of lead, mercury, hexavalent chromium,
cadmium, polybrominated biphenyl (PBB), and polybrominated
diphenyl ether (PBDE). “20” indicates the number of years during
which the hazardous substance(s) will not leak or mutate so that the
use of this product will not result in any severe environmental
pollution, bodily injury, or damage to any assets.
15. Use two people to transport system on inclines. This label also
indicates the system weight.
The equipment weighs approximately 126 kg (278) lbs. To avoid
possible injury and equipment damage when transporting from one
area of use to another:
• Be sure the pathway is clear.
• Limit movement to a slow careful walk.
• Use two or more persons to move the equipment on inclines or long
distance.
16. This WEEE symbol indicates that waste electrical and electronic
equipment must not be disposed of as unsorted municipal waste and
must be collected separately. Please contact an authorized
representative of the manufacturer for information concerning the
decommissioning of your equipment. Standard: EN 50419.
17. Type BF Applied Part (man in the box) symbol is in accordance with
IEC 60417-5333.
19. Symbol indicating that the Instructions for Use are supplied in
electronic form (eIFU). Standard: ISO 7000-1641.
Chapter 1 Introduction 1 - 13
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
UDI Human Readable Label Text: Global Trade Item Number, GTIN,
Manufacturing Date, Serial Number
System Voltage
Monitor DO NOT place a finger, hand or any object on the joint of the monitor
Arm or monitor arm to avoid injury when moving the monitor and monitor
arm.
Op Panel “Mains OFF” indicates the power off position of the mains power
breaker.
Op Panel “Mains ON” indicates the power on position of the mains power
breaker.
Chapter 1 Introduction 1 - 15
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
1-3-4 Product Labels on the LOGIQ E10 when used in a veterinary environment
There are different handling instructions when servicing the LOGIQ E10 used in a veterinary
environment.
Label Location
Chapter 1 Introduction 1 - 17
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 1-4
Safety considerations
Contents in this section
1-4-1 Introduction
The following safety precautions must be observed during all phases of operation, service, and repair
of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual violates safety standards of design, manufacture, and intended use of the equipment.
WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ E10, SOME METAL
SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF
TOUCHED, EVEN WHILE IN SHUTDOWN MODE.
WARNING BECAUSE OF THE LIMITED ACCESS TO CABINETS AND EQUIPMENT IN THE FIELD,
PLACING PEOPLE IN AWKWARD POSITIONS, GE HAS LIMITED THE LIFTING WEIGHT
FOR ONE PERSON IN THE FIELD TO 16 KG (35 LBS). ANYTHING OVER 16 KG (35 LBS)
REQUIRES 2 PEOPLE.
CAUTION HAVE TWO PEOPLE AVAILABLE TO DELIVER AND UNPACK THE LOGIQ E10.
ATTEMPTS TO MOVE THE LOGIQ E10 CONSIDERABLE DISTANCES OR ON AN INCLINE
BY ONE PERSON COULD RESULT IN INJURY OR DAMAGE OR BOTH.
WARNING USE ALL PERSONAL PROTECTION EQUIPMENT (PPE) SUCH AS GLOVES, SAFETY
SHOES, SAFETY GLASSES, AND KNEELING PADS TO REDUCE THE RISK OF
INJURY.
WARNING For console Ultrasound systems and Ultrasound systems mounted on a cart, when
the top console is in its locked position, the gas shock is compressed and stores
mechanical energy. During normal operation the top console, the weight of the main
monitor and the mechanical force of the gas shock are in balance. Take care if/when
you activate this gas shock. Personal injury can occur after the panel is removed and
the shock pressure is released. Take care when you repair the elevation assembly.
WARNING Risk of electrical shock, the LOGIQ E10 must be turned off and disconnected from
power source. Cord must be controlled at all times.
Wait for at least 20 seconds for capacitors to discharge and the ON/OFF button (white
light on the Op Panel) to turn OFF, as there are no test points to verify isolation.
Disconnect the batteries from the backplane.
Beware that the Main Power Supply, the Power Assist batteries or Power Module and
ECB may be energized even if the power is turned OFF if the cord is still plugged into
the AC Outlet.
WARNING Use extreme caution as long as the LOGIQ E10 is un-stable, not resting on all four
casters.
WARNING Tilting the LOGIQ E10 requires two people in order to avoid injury to service
personnel and damage to the equipment.
Chapter 1 Introduction 1 - 19
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
WARNING Wear all PPE including gloves as indicated in the chemical SDS.
WARNING WHEN THE LOGIQ E10 IS RAISED FOR A REPAIR OR MOVED ALONG ANY INCLINE,
USE EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER.
WARNING ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
WARNING NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT
MAY BE DAMAGED.
WARNING REMEMBER: If the front caster swivel lock is engaged for transportation, pressing the
release pedal once disengages the swivel lock. You must depress the release pedal a
second time to engage the brake.
WARNING The LOGIQ E10 should NOT be moved with the Operator I/O Panel extended. Move the
Operator I/O Panel to its centered and locked position. Lower the Operator I/O Panel as
much as possible before moving the LOGIQ E10.
CAUTION BEFORE YOU MOVE OR TRANSPORT THE LOGIQ E10, MAKE SURE TO FIRMLY LOCK THE
MONITOR ARM AND FLIP DOWN THE MAIN MONITOR TO PREVENT DAMAGE TO THE
LOGIQ E10.
CAUTION ALWAYS LOCK THE TOP/UPPER CONSOLE IN ITS PARKING (LOCKED) POSITION BEFORE
MOVING THE LOGIQ E10.
CAUTION TO AVOID INJURY WHEN YOU MOVE THE MAIN MONITOR AND THE MONITOR ARM, DO NOT
PUT YOUR FINGER, HAND, OR OBJECT ON THE JOINT OF THE MAIN MONITOR OR THE
MONITOR ARM.
Chapter 1 Introduction 1 - 21
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CAUTION TO AVOID INJURY OR DAMAGE TO THE MAIN MONITOR, MAKE SURE THERE IS NOTHING
WITHIN RANGE OF THE MAIN MONITOR BEFORE MOVING THE MAIN MONITOR AND
MONITOR ARM. THIS INCLUDES PEOPLE AS WELL AS THINGS.
CAUTION If the Operator Control Panel gas spring is not functioning correctly, there is a chance that the
operator panel could drop down suddenly when you use the manual release handle. To avoid
injury, take care that you are not positioned below the operator panel when using the manual
release handle.
CAUTION Do not move the Ultrasound system if the Operating Panel is in the unlocked position.
CAUTION Ensure that nobody touches the console arm/frogleg when moving the Operator Panel.
CAUTION Use protective glasses during drilling, filing, and during all other work where
eyes protection is needed.
CAUTION Use safety shoes when doing work where there is any chance of foot damage.
CAUTION Use protective gloves when drilling, cutting, and when reaching into
confined or blind spaces.
CAUTION Do not transport LOGIQ E10 in a vehicle without locking the casters (wheels) and securing it as
described in chapter 4.
NOTE: Special care should be taken when transporting the Ultrasound system in a vehicle:
• Before transporting, place the Ultrasound system in its special storage case.
• Ensure that the Ultrasound system is firmly secured while inside the vehicle.
• Secure Ultrasound system with straps or as directed otherwise to prevent motion during transport.
• Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and
erratic stops or starts.
Chapter 1 Introduction 1 - 23
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Follow these guidelines to minimize shock hazards whenever you are using the LOGIQ E10:
• To minimize shock hazard, the equipment chassis must be connected to an electrical ground.
• The LOGIQ E10 is equipped with a three-conductor AC power cable. This must be plugged into an
approved electrical outlet with safety ground. A separate power outlet with a 20 amp circuit breaker
for 120 VAC for 120V area, 7.5 amp circuit breaker for 220-240 VAC for 220/240V area or 15 amp
circuit breaker for 100 VAC for Japan.
• The power outlet used for this equipment should not be shared with other types of equipment.
• Both the LOGIQ E10 power cable and the power connector must meet international electrical
standards.
WARNING Connecting the LOGIQ E10 to the wrong voltage level will most likely destroy it.
Probes
• Inspect the probe prior to each use for damage or degradation to the:
- housing
- cable strain relief
- lens
- seal
- connectors
- locking mechanism
• Do not use a damaged or defective probe.
• Never immerse the probe connector or adapter into any liquid.
• The LOGIQ E10 has more than one type of probe port. Use the appropriate probe port designed for
the probe you are connecting.
Peripherals
• Refer to the “Peripherals used in the patient environment” in Chapter 2 of the User Manual for
peripheral isolation information.
The lithium ion technology is significantly less hazardous to the environment than the lithium metal
technology used in some other batteries (such as watch batteries). Used batteries should not be placed
with common household waste products. Contact local authorities for the location of a chemical waste
collection program nearest you.
NOTE: Regulations vary for different countries. Dispose of used batteries in accordance with local
regulations.
WARNING
• DO NOT short-circuit the batteries by directly connecting the negative terminals with metal objects.
• DO NOT heat batteries or discard them into a fire.
• DO NOT expose batteries to temperatures over 50 °C. Keep batteries away from fire and other heat
sources.
• DO NOT charge batteries near a heat source, such as, a fire or heater.
• DO NOT leave batteries in direct sunlight.
• DO NOT pierce batteries with a sharp object, hit it, or step on it.
• DO NOT use damaged batteries.
• DO NOT apply solder to batteries.
• DO NOT connect batteries to an electrical power outlet.
• DO NOT immerse batteries in water or allow them to get wet.
CAUTION
USE ONLY BATTERIES APPROVED BY GE AS SUITABLE FOR USE WITH THE LOGIQ E10
ULTRASOUND SYSTEM.
CAUTION
NOTE: If you have any questions about the batteries, consult your local GE representative.
Chapter 1 Introduction 1 - 25
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 1-5
Label locations
For information on label locations, refer to the appropriate version of the LOGIQ E10 Basic User Manual
or the latest version of the LOGIQ E10 Release Notes.
Table 1-7 Basic User Manual and Release Notes for the LOGIQ E10
Section 1-6
Dangerous procedure warnings
Warnings, such as the example below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
WARNING IF THE COVERS ARE REMOVED FROM AN OPERATING LOGIQ E10, SOME METAL
SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF
TOUCHED, EVEN WHILE IN SHUT DOWN MODE.
Chapter 1 Introduction 1 - 27
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 1-7
Lockout/tagout (LOTO) requirements
Follow Lockout/Tagout requirements by ensuring you are in total control of the AC power plug at all
times during the service process.
To apply Lockout/Tagout:
7. Verify isolation:
All potentially hazardous stored or residual
energy is relieved.
Lockout/Tagout Procedure
1. TURN OFF the LOGIQ E10 and UNPLUG IT. Maintain control of the mains power plug.
2. Wait for at least 20 seconds for capacitors to discharge and the ON/OFF button (white light on the Op Panel) to
turn OFF, as there are no test points to verify isolation. Capacitors on Ultrasound systems with the Shear Wave
Option can take up to 5 minutes to discharge.
J10
TAG
&
LOCKOUT
J10 J11
Signed Date
J11
J14
J13
Chapter 1 Introduction 1 - 29
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 1-8
Returning/shipping probes and repair parts
Equipment being returned must be clean and free of blood and other infectious substances.
GE policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have
been properly decontaminated prior to shipment. Under no circumstance should a part or equipment
with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an ultrasound
probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purposes and must be
transported as a hazardous material.
For a LOGIQ E10 used in a veterinary environment the same restrictions apply.
NOTE: Return used/unused spare parts from a veterinary environment with the purple recycling label
(regardless of its actual condition) and add a description on the label stating that the items were
removed from a LOGIQ E10 Vet in a veterinary environment. This applies for probes and covers
labeled as Vet used.
NOTE: The USER/SERVICE staff should dispose of all the waste properly, per federal, state, and local
waste disposal regulations.
If the purple recycling label is not used in your region, use the local recycling label.
The LOGIQ E10 is not meant to be used for long-term storage of patient data or images. The user is
responsible for the data on the LOGIQ E10 and a regular backup is highly recommended.
If the LOGIQ E10 is sent for repair, please ensure that any patient information is backed up and erased
from the LOGIQ E10 before shipping. It is always possible during system failure and repair to lose
patient data. GE is not responsible for the loss of this data.
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes,
GE will ascertain agreement from the customer. Patient information shall only be transferred by
approved service processes, tools and devices restricting access, protecting or encrypting data where
required, and providing traceability in the form of paper or electronic documents at each stage of the
procedure while maintaining compliance with cross-border restrictions of patient information transfers.
Section 1-9
Electromagnetic compatibility (EMC)
1-9-1 What is EMC?
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
1-9-2 Compliance
LOGIQ E10 conforms to all applicable conducted and radiated emission limits and to immunity from
electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient
requirements.
NOTE: For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, and clamps
are in good condition and installed tightly without skew or stress. Proper installation following
all comments noted in this service manual is required in order to achieve full EMC performance.
Chapter 1 Introduction 1 - 31
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
ESD Receptacles
Battery Box side Card Cage side
WARNING Risk of electrical shock. The LOGIQ E10 must be turned off. Avoid all contact with
electrical contacts, conductors and components. Always use non-conductive handles
designed for the removal and replacement of ESD sensitive parts. All parts that have
the potential for storing energy must be discharged or isolated before making contact.
WARNING Risk of heat hazard. If the covers are removed from an operating LOGIQ E10 some
metal surfaces (for example, the GPU heat sink on the ECB) may be warm enough to
pose a potential heat hazard even while in shutdown mode.
Section 1-10
Customer assistance
1-10-1 Contact information
If this equipment does not work as indicated in this service manual or in the user manual, or if you
require additional assistance, please contact the local distributor or appropriate support resource as
listed in Table 1-11.
Before you call, identify the following information, and acquire an image (Alt+D) to send to the Customer
Care team:
Chapter 1 Introduction 1 - 33
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
LOCATION NUMBER
TEL: 1-800-425-8025
India Technical Support TEL: 1-800-425-7255
TEL: 1-800-102-7750
MANUFACTURER
AUTHORIZED REPRESENTATIVE
This product complies with the regulatory requirement of Council Directive 93/42/EEC
concerning medical devices:
• The CE label affixed to the product testifies compliance to the Directive.
• The location of the CE marking is shown in the Safety chapter of this manual.
Authorized EU Representative
European registered place of business:
GE Medical Systems SCS
283 rue de la Minière
78530 BUC, France
Chapter 1 Introduction 1 - 35
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 2
Site Preparations
Section 2-1
Overview
2-1-1 Purpose of this chapter
This chapter provides the information required to plan and prepare for the setup of a LOGIQ E10.
Included are descriptions of the facility and electrical needs to be met by the purchaser of the
LOGIQ E10.
Section 2-2
General LOGIQ E10 requirements
Contents in this section
When unpacking the LOGIQ E10, allow the temperature to stabilize before powering up.’
Refer to Table 2-1 for guidelines on reaching operational temperatures from storage or transport
temperatures.
Environmental specifications
The LOGIQ E10 should be operated, stored, and transported according to the listed environmental
requirements.
NOTE: Probe environmental specifications may be different than the LOGIQ E10.
Cooling
The cooling requirement for the LOGIQ E10 with monitor and onboard peripherals is up to 3070 Btu/hr.
This figure does not include cooling needed for lights, people, or other equipment in the room. Each
person in the room places an additional 300 Btu/hr demand on the cooling system.
Lighting
Bright light is needed for LOGIQ E10 set up, updates, and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
NOTE: GE requires a dedicated power and ground for the proper operation of its ultrasound equipment. This
dedicated power shall originate at the last distribution panel before the LOGIQ E10.
The LOGIQ E10 will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using 220
volt power, then a center tapped power source is required.
Sites with a mains power system with defined neutral and live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the LOGIQ E10 outlet.
NOTE: Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility’s incoming power source to the Ultrasound system is only a conduit.
In the table below, the electrical specifications for the LOGIQ E10 include monitor and on board
peripherals.
Inrush current
During power on, an inrush circuit prevents the current from increasing above the stated values.
It is recommended that the branch circuit breaker for the LOGIQ E10 be readily accessible.
A dedicated AC power outlet must be within reach of the LOGIQ E10 without extension cords. Other
outlets adequate for the external peripherals, medical and test equipment needed to support this
LOGIQ E10 must also be present within 1 m (3.2 ft.) of the LOGIQ E10. Electrical installation must meet
all current local, state, and national electrical codes.
If the LOGIQ E10 arrives without a power plug, or with the wrong plug, you must contact your GE dealer
or the installation engineer must supply what is locally required.
Voltage drop-out
Max 10 ms.
Less than 25% of nominal peak voltage for less than 1 ms for any type of transient, including line
frequency, synchronous, asynchronous, or aperiodic transients.
Possible EMI sources should be identified before the LOGIQ E10 is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. These
sources include:
• medical lasers
• scanners
• cauterizing guns
• computers
• monitors
• fans
• gel warmers
• microwave ovens
• light dimmers
• mobile phones
• in-house wireless phones (DECT phones)
• wireless computer keyboard and mouse
• air conditioning system
• High Frequency (HF) surgery equipment
• general AC/DC adapters
The presence of a broadcast station or broadcast van may also cause interference.
Poor grounding is the most likely reason a LOGIQ E10 will have noisy images. Check grounding
Ground the LOGIQ E10
of the power cord and power outlet.
• After you finish repairing or updating the LOGIQ E10, replace all covers and tighten all screws.
Replace all screws, RF • Any cable with an external connection requires a magnet wrap at each end.
gaskets, covers, cores • Install the card rack cover over the card rack. Loose or missing covers or RF gaskets allow radio
frequencies to interfere with the ultrasound signals.
Replace broken RF If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket. Do not
gaskets turn on the LOGIQ E10 until any loose metallic part is removed.
Do not place labels where Never place a label where RF gaskets meet the LOGIQ E10. Otherwise, the gap created will
RF gaskets touch metal permit RF leakage. Or, if a label has been found in such a position, move the label.
Use GE specified The interconnect cables are grounded and require ferrite beads and other shielding. Also, cable
harnesses and peripherals length, material, and routing are all important; do not change from what is specified.
Do not allow cables to lie across the top of the card rack or hang out of the peripheral bays. Loop
Properly dress peripheral
the excess length for peripheral cables inside the peripheral bays. Attach the monitor cables to
cables
the frame.
NOTE: LOGIQ E10 environmental specifications may be different than the probes.
NOTICE STORAGE TEMPERATURES VARY DEPENDING ON THE TYPE OF PROBE. WHEN EXPOSED TO
LARGE TEMPERATURE VARIATIONS, THE LOGIQ E10 AND PROBES SHOULD BE KEPT AT
ROOM TEMPERATURE BEFORE USE.
Refer to the table in "If the LOGIQ E10 is very cold or hot" on page 2-2 to determine the needed
settlement time.
Section 2-3
Facility needs
Contents in this section
NOTE: All electrical installations that are preliminary to the positioning of the equipment at the site
prepared for the equipment must be performed by licensed electrical contractors. Other
connections between pieces of electrical equipment, calibrations, and testing must also be
performed by qualified personnel. The products involved (and the accompanying electrical
installations) are highly sophisticated and special engineering competence is required. All
electrical work on these products must comply with the requirements of applicable electrical
codes. The purchaser of GE equipment must only utilize qualified personnel to perform
electrical servicing on the equipment.
The desire to use a non–listed or customer provided product or to place an approved product further
from the LOGIQ E10 than the interface kit allows, presents challenges to the installation team. To avoid
delays during installation, such variances should be made known to the individuals or group performing
the installation at the earliest possible date (preferably prior to the purchase).
The ultrasound suite must be clean prior to delivery of the machine. Carpet is not recommended
because it collects dust and creates static. Potential sources of EMI (electromagnetic interference)
should also be investigated before delivery. Dirt, static, and EMI can negatively impact LOGIQ E10
reliability.
The LOGIQ E10 will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using 220
volt power, then a center tapped power source is required.
Sites with a mains power system with defined neutral and live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the ultrasound outlet.
• Dedicated single branch power outlet of adequate amperage (see: Table 2-3 "Electrical
specifications for the LOGIQ E10" on page 2-4) meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the LOGIQ E10’s proposed location
• Door opening is at least 76 cm (30 in) wide.
• Proposed location for LOGIQ E10 is at least 0.46 m (1.5 ft) from the wall or objects for cooling.
• Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m (3.2 ft.) of the LOGIQ E10 to connect cables.
• Power outlets for other medical equipment and gel warmer.
• Power outlets for test equipment within 1 m (3.2 ft.) of the LOGIQ E10.
• Clean and protected space to store transducers (in their cases or on a rack).
• Material to safely clean probes (done with a plastic container, never metal).
10 11 15
12
9 1
PW
X Z
CW
M CF
Elasto
Comment
Auto
L R
Clear
16
2 3 4 5 13 14
7
1
17
18
5. Counter top and sink with hot and cold water 6. Overhead light dimmer – dual level lighting (bright/dim)
11. 12. 45,7 cm (1.5 ft) distance of LOGIQ E10 from wall or
Network interface
objects
15. Storage for linens and equipment 16. Examination table – 193 x 61 cm (76 x 24 inches)
17. Lavatory and dressing room 18. Door – at least 76 cm (30 inches)
4 5
2 3
11
10
9
Elasto
1
2
8 12
Comment
Clear
M
CW
X
Body Pattern
Ellipse
PW
Y
e
Zoom
3D/4D
13
CF
Z
L
B
R
7
Auto
6 14
1
3. Counter top, sink with hot and cold water, and supplies 4.
Linen supply
storage
7. Footswitch 8. Stool
13. 45,7 cm (1.5 ft) distance of LOGIQ E10 from wall or 14. Lockable cabinet ordered by GE for its software and
objects proprietary manuals
None.
InSite™ requirements
Need internet access on the outbound port to GE’s remote service platform (Insite), which is only
opened on request by the user and through a secure HTTPS connection on port 443.
Insite requires an Ethernet connection with a 10/100 Mbps or 10/100/1000 Mbps interface.
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Inbound and outbound connections through port 104 are for
DICOM connectivity. Examples of DICOM services include the transfer of images to workstations for
viewing or transferring images to remote printers. As an added benefit, transferring images in this
manner frees up the on-board monitor and peripherals, enabling viewing to be done while scanning
continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower
cost.
To configure the LOGIQ E10 to work with other network connections, the site’s network administrator
must provide information to complete the form in Figure 2-3 "Worksheet for DICOM Network
Information" on page 2-14. Ensure that there are no spaces in any field of the form.
• A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ E10.
• The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
• The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ E10 for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and
the revision of the device, is also included. This information may be useful for troubleshooting.
For connectivity setup information, refer to the current revision of the LOGIQ E10 Basic User Manual.
See: Section 1-5 "Label locations" on page 1-26.
LOGIQ
E10
Host Name IP Address . . . Local Port
DHCP
DICOM APPLICATION
OTHER
NAME MAKE/ AE TITLE IP PORT
CONFIGURATION
Store 2
__Raw Data
. . . __Allow Multiframe
__Structured
Reporting
Compression_______
Store 3
__Raw Data
. . . __Allow Multiframe
__Structured
Reporting
Compression_______
DICOM
Print Vendor: ___________
. . . Print Size: _________
Medium: ___________
Copies: ___________
Orientation: ________
Color ______________
Worklist . . .
Storage Associated
. . .
Commit Storage AE
_____________
DICOM . . .
MPPS
For more information on privacy and security, refer to “Privacy and Security” in chapter 12 of the LOGIQ
E10 Basic User Manual.
2.12.20DV (UL60601-1:2003)
In area in which patients are normally cared for, the patient vicinity is the space with surfaces likely to
be contacted by the patient or attendant who can touch the patient. This encloses a space within the
room 1.83 m (6 ft.) beyond the perimeter of the bed (examination table, dental chair, treatment booth,
and the like) in its intended location, and extending vertically 2.29 m (7.5 ft.) above the floor.
Item Description
1. Patient vicinity
Sub Clause 3.79 and figure A.9 (IEC60601-1:2005 and ANSI AAMI ES60601-1:2005)
Such an area is an environment in which medical diagnosis, monitoring or treatment is carried out. It is
very difficult to attach unique dimensions to the PATIENT ENVIROMENT.
In practice, a distance of 2,5 m (8.2 ft.) above the floor on which the medical personnel stand and a
horizontal distance of 1,5 m (4.9 ft.) have justified themselves as indicative of the dimensions of the
Patient Environment.
The patient environment/vicinity will be depicted as a dashed line in this procedure. See example below.
Item Description
1. Patient environment
Chapter 3
LOGIQ E10 Setup
Section 3-1
Overview
3-1-1 Purpose of this chapter
This chapter contains information needed to set up the LOGIQ E10. Also included in this section are
guidelines for transporting the unit to a new site.
Section 3-2
Setup reminders
Contents in this section
AVERAGE SETUP
DESCRIPTION TIME COMMENTS
UNPACKING THE LOGIQ E10 0.5 HOUR
CAUTION DO NOT WEAR THE ESD WRIST STRAP WHEN YOU WORK ON LIVE CIRCUITS AND MORE
THAN 30 V PEAK IS PRESENT.
CAUTION DO NOT OPERATE THE LOGIQ E10 UNLESS ALL BOARD COVERS AND FRAME PANELS ARE
SECURELY IN PLACE. LOGIQ E10 PERFORMANCE AND COOLING REQUIRE THIS.
NOTICE NEVER REVERSE POLARITY ON ANY METER THAT INTERCEPTS THE POWER CORD WITH
POWER CONNECTED TO THE LOGIQ E10.
EVEN IN THE OFF STATE, REVERSING POLARITY ON THE POWER CAN SERIOUSLY DAMAGE
THE POWER SUPPLY.
CAUTION IF THE LOGIQ E10 IS VERY HOT OR COLD, ALLOW THE TEMPERATURE TO STABILIZE
BEFORE POWERING UP. FOR GUIDELINES ON REACHING OPERATIONAL TEMPERATURES
FROM STORAGE OR TRANSPORT TEMPERATURES, SEE 2-2-1 "LOGIQ E10 environmental
requirements" on page 2-2
Section 3-3
Receiving and unpacking the equipment
Contents in this section
CAUTION TWO PEOPLE ARE NEEDED TO UNPACK THE LOGIQ E10 BECAUSE OF ITS WEIGHT.
ATTEMPTS TO MOVE THE LOGIQ E10 CONSIDERABLE DISTANCES OR ON AN INCLINE
BY ONE PERSON COULD RESULT IN INJURY OR DAMAGE OR BOTH.
TWO PEOPLE ARE REQUIRED WHENEVER A PART WEIGHING 16 KG (35 LBS) OR
MORE MUST BE LIFTED.
Improper handling during transportation may harm the equipment inside the package even if the
package itself is undamaged.
To make it easier to detect if the handling during transportation has been improper, a tilt indicator has
been attached to the transportation box.
NOTE: Before cutting the straps, check the tilt indicator to make sure it has not been triggered. If
triggered, report it to the carrier. If not, then cut the straps around the crate.
The LOGIQ E10 Transportation Box Label is located on the transportation box.
SYMBOL DEFINITION/COMMENTS
RECYCLING
Recyclable Wood
or
RECYCLING
China Specific
KEEP DRY
(protect from moisture)
DO NOT STACK
Improper handling during transportation may harm the equipment inside the package even if the
package itself is undamaged.
Examine package
Damage in transportation
1. Write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed
for“ by a GE representative or hospital receiving agent.
4.
Procedure continues...
11.
12.
Procedure continues...
14.
15.
16. Check the empty pallet to see if the filter covers came off during shipping.
17. For future use, store all of the packing materials inside the box.
See: 2-2-3 "EMI limitations" on page 2-6 for more information about EMI protection.
Section 3-4
Completing the setup
Contents in this section
• Verify that the site meets the requirements listed in (see: Section 2-3 "Facility needs" on page 2-9).
• Verify that the specifications below do not conflict with any on-site conditions.
Physical dimensions
The physical dimensions of the LOGIQ E10 are summarized in Table 3-5.
Table 3-5 Physical dimensions of LOGIQ E10 with monitor and peripherals
* Dimensions given with floating keyboard stowed for transport and the Main Monitor down.
Less than 48 dB(A) at 20 degrees Celsius at a distance of 1 meter at all four sides, 1 meter above the
floor.
WARNING CONNECTING A LOGIQ E10 TO THE WRONG VOLTAGE LEVEL WILL MOST LIKELY
DESTROY IT.
Verification of the LOGIQ E10 voltage setting
Verify that the mains voltage specified for the LOGIQ E10 is available on-site.
The voltage setting for the LOGIQ E10 is found on the rating label on the back of the LOGIQ E10 on
lower rear frame.
Figure 3-5 Rating Label - LOGIQ E10
Connect Ethernet
Connect the Ethernet cable to the Ethernet connector on the external I/O (rear side of LOGIQ E10).
Figure 3-6 I/O Rear Panel
Item Connection
1 USB 3.0 port
3 S-Video
4 HDMI port
Refer to “USB Hard Disk Drive and USB Flash Drive” in Chapter 9 of the LOGIQ E10 Basic User Manual
for more information.
3 - 18 Section 3-4 - Completing the setup
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Refer to “Patient Cardiac and ECG Connections” in Chapter 3 of the LOGIQ E10 Basic User Manual for
more information.
LOGIQ E10s have one type of probe ports: four DLP probe ports (Figure 3-7).
Figure 3-7 Probe connectors, four DLP probe ports
To connect a probe:
NOTE: It is not necessary to turn OFF power to connect or disconnect a probe. However, it is a good
idea to select a different probe or to freeze the image when removing a probe to avoid
disconnecting a live probe.
CAUTION DO NOT CONNECT A PROBE WHILE THE LOGIQ E10 IS POWERING UP.
CAUTION DO NOT ALLOW THE PROBE HEAD TO HANG FREELY. EXCESSIVE IMPACT TO THE PROBE
WILL RESULT IN IRREPARABLE DAMAGE.
To disconnect a probe:
1.) Select a different probe or freeze the image before removing a probe in order to avoid disconnecting
a live probe.
2.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector.
3.) Remove the connector from the port.
4.) Ensure that the probe head is clean before placing the probe in its storage case, see: 10-5-9
"Cleaning" on page 10-12 for cleaning instructions.
DANGER ALWAYS CONNECT THE LOGIQ E10 TO A FIXED POWER SOCKET WHICH HAS
THE PROTECTIVE GROUNDING CONNECTOR.
DANGER ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE
POWER PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
NOTICE Use only power supply cords, cables, and plugs provided by or designated by GE.
NOTE: Do not cycle the circuit breaker ON-OFF-ON in less than five seconds. When turning OFF the
circuit breaker, WAIT until the ON/OFF button is no longer lit. The LOGIQ E10 should de-
energize completely before turning the circuit breaker ON.
Connecting AC Power to the LOGIQ E10 involves preliminary checks of the power cord, voltage level
and compliance with electrical safety requirements.
NOTE: The LOGIQ E10 will function on voltages from 100-240 volts and 50 or 60 Hz. However, if using
220 volt power, then a center tapped power source is required (North America Only).
1.) Ensure that the wall outlet is of appropriate type, and that the circuit breaker is turned off.
2.) Uncoil the power cable, allowing sufficient slack so that the LOGIQ E10 can be moved slightly.
3.) Verify that the:
- power cable is without any visible scratches, any sign of damage or tangled.
- on-site mains voltage is within the limits indicated on the rating label near the circuit breaker
on the rear of the LOGIQ E10.
4.) Connect the female plug of the power cable to the power inlet at the rear of the LOGIQ E10.
5.) Lock the plug in position with the retaining clamp (ACC clamp).
6.) Verify that the mains power circuit breaker is in OFF position, If not, switch it to OFF.
Figure 3-8 Circuit Breaker
7.) Connect the other end (male plug) of the power cable to a hospital grade mains power outlet with
the proper rated voltage, and the LOGIQ E10 is ready for Power ON/Boot Up.
1.) Switch ON the mains power circuit breaker at the rear of the LOGIQ E10. The ON/OFF button turns
amber.
Figure 3-9 ON/OFF button
You should hear a click from the relays in the AC power and the LOGIQ E10 is ready to boot. This
indicates that there is power to the PS, but the LOGIQ E10 is OFF.
2.) Press once on the ON/OFF button on the Operator Panel to boot the LOGIQ E10. The ON/OFF
button turns green when it is pressed.
During a normal boot, you may observe that:
a.) The ventilation fans start on full speed, but slow down after a few seconds (listen to the fan).
b.) Power is distributed to the peripherals, operator panel, monitor, front end and back end.
c.) Back end and rest of the LOGIQ E10 starts with the sequence listed in the next steps:
1.) Back end is turned ON and starts to load the software.
2.) The Start screen is displayed on the monitor.
3.) A start-up bar indicating the time used for software loading, is displayed on the monitor.
4.) The software initiates and sets up the front end electronics and the rest of the instrument.
5.) The backlight in the keyboard is lit.
6.) As soon as the software has been loaded, either a 2D screen is displayed on the screen,
indicating that a probe has been connected, or a No Mode screen is displayed, indicating
that no probe has been connected.
NOTE: Total time used for start-up is typically less than 170 seconds.
After the 2D screen is displayed and any function is touched, the LOGIQ E10 will prompt a password
protected Login.
3.) Under Password Policies, use the drop down selection beside each feature to set the rules for
passwords.
4.) Click Save.
To add users:
NOTE: You will need to edit the ID field from NewUser default. Do not push “Add” again until adding
another user.
DO NOT include the following characters in a user ID: slash (/), dash (-), asterisk (*), question mark (?),
an underscore (_), ampersand (&), or blank spaces. Also, DO NOT set up Users with the same
initials/signifier.
NOTE: The user password is system specific. If the user needs to access multiple systems the
password will need to be entered manually on each system.
The system administrator can specify whether the users account is Active, Blocked or requires a
password change.
The system administrator may also remove a user. Highlight the user Id in the list, select Remove to
mark the user as inactive and then select Remove again to remove user from the list.
To use groups:
You can use predefined groups or create new groups. Assign groups to a user from Users and under
Group Memberships.
Auto LOGON only works when password policies are disabled and if there is no password assigned to
the user.
Bitlocker Drive Encryption uses Advanced Encryption Standard (AES) with configurable key lengths of
256 bits.
NOTE: DO NOT use the USB media for ANY other storage. This should be used for recovery key
storage ONLY.
11.)Select Save and the recovery key will be displayed one more time.
Figure 3-15 Disk Encryption - Recovery Key Window
12.)If you have not already written the key down you can do so now before moving on.
NOTE: It is important to write down or take a print of the password and recovery key. Store these and
the USB key in a secure location.
The displayed recovery key string shown is an example and not a real recovery key. The
generated key will always be different.
13.)Select Ok to continue.
14.)Encryption of data will begin. This may take up to 90 minutes the first time. After the initial encryption
there is no long process for daily use.
Figure 3-16 Disk Encryption - Encrypting Window
15.)If the recovery key needs to be changed, the disk encryption can be performed again. Insert the
USB flash drive and select Change key.
Figure 3-17 Disk Encryption - Change key
After the disk is encrypted, you will be required to unlock the disk at each startup of the system.
The unlock dialog box will appear when the new patient or print button is pushed.
Chapter 3 LOGIQ E10 Setup 3 - 29
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
It is recommended the recovery key is stored in a USB flash drive and printed out.
If both password and recovery key are lost, you will not be able to access archived patient data (images
and measurements included) nor store new patient data on this system.
The only way to recover the system to allow storing patient data is to reset entire disk that deletes all
the archived patient data in the disk.
It is strongly recommended that all the patient data be stored in PACS or backup to external media.
3-4-10 Login
At login, you are notified that “You are accessing a diagnostic medical device that is provided by
authorized usage only. Data stored on this device may be subject to various regulations including but
not limited to regulations which govern disclosure and privacy of this data. By using this device you are
acknowledging that you are authorized to do so and are trained in appropriate use and regulatory
guidelines.”
NOTICE The System Administrator manages system groups, users, and permissions. After you have been
added as a valid user, the System Administrator assigns you a temporary password. When you log into
the system for the first time, you will be prompted to change your password.
After you have established your password, follow these steps to login.
You can change your password when first logging onto the system.
• Logoff
Use this button to log off the current user.
The LOGIQ E10 remains ON and ready for a new user to log on.
If this button is dimmed, it indicates that no user is logged on to the LOGIQ E10 at the moment.
• Shutdown
Use this button to shut down the LOGIQ E10. It is recommended to perform a full shutdown at least
once a week.
If this button is dimmed, press the ON/OFF button or Alt+F10 to shut down the LOGIQ E10.
• Cancel
Use this button to exit from the SYSTEM-EXIT menu and return to the previous operation.
• Exit
Select this button when you want to exit to the Windows Desktop.
1.) Press once on the ON/OFF button on the operator panel to display the SYSTEM - EXIT menu.
Figure 3-22 ON/OFF button
Next, it starts to shut down. The time to turn down the LOGIQ E10, including the back end, may vary
from 10 seconds up to approximately 1 minute.
The last thing that shuts down, is the light on the operator panel, indicating that you can continue
with the next step.
NOTICE Risk of losing data. Be sure to wait with the next step until the LOGIQ E10 has finished its shut-down.
Failing to do so may destroy data on the hard disk drive, making the LOGIQ E10 fail later.
3.) Switch off the Mains Power Circuit Breaker, located on the rear of the LOGIQ E10. This will cut
power distribution within the LOGIQ E10.
Section 3-5
Configuration
Contents in this section
For more information, see: Section 1-5 "Label locations" on page 1-26.
This list covers the internal peripherals available for the LOGIQ E10.
The external printers are connected through Ethernet (TCP/IP Network) as DICOM devices.
• DICOM Printers. Connected through Ethernet (TCP/IP Network or WLAN) as DICOM devices.
• USB Printers, see: 9-14-1 "Printers" on page 9-31.
• USB Footswitch
• Rolling V Nav Stand
• Color Printer
• Report Printer
Additional Information
Printer Checks
2. Press (Print 1), (Print 2), (Print 3) or (Print 4) on Prints image displayed on the screen on a black and white or color
the control panel. printer, depending on the key assignment configuration.
A password (Software Option String) enables a software option or a combination of software options.
This password is specific for each LOGIQ E10.
After you have established your password, follow these steps to login.
4.) From the Touch Panel, select Utility > Admin > System Admin.
5.) Enter the new option key code in the SW Option Key section.
Section 3-6
Setup paperwork
NOTE: During and after setup, the documentation (for example, CDs with documentation, user
manuals, and installation manuals) for the peripheral units must be kept as part of the original
system documentation. This will ensure that all relevant safety and user information is available
during the operation and service of the complete system.
The UDI of the medical device must be captured when any work is performed on that device. Refer to
your local procedures on the actual capturing of the UDI.
Figure 3-26 Product Locator Card (Example)
g
GEHC - Americas Product Locator Cards Asia Service Operation
Mailing Product Locator – W523 XEROX Office No. 1, Yongchang North Road
Address P.O. Box 414 VÀKUUMTECHNOLÒGIAI GÈPGYÀR Beijing Economic and
Milwaukee, WI 53201-0414 FÒTI ÙT 141 Technologic Development Area
1046 Budapest Hungary Beijing 100176 China
DESCRIPTION F DA MOD EL REV SERIAL
OCP BS ORD
DESTINATION
SHIPMENT
NAME AND
ADDRESS
ZIP CODE
46-303268P1 Rev 14
GIB, European Central Admin Product Locator Card
g
GEHC - Americas Product Locator Cards Asia Service Operation
Mailing Product Locator – W523 XEROX Office No. 1, Yongchang North Road
Address P.O. Box 414 VÀKUUMTECHNOLÒGIAI GÈPGYÀR Beijing Economic and
Milwaukee, WI 53201-0414 FÒTI ÙT 141 Technologic Development Area
1046 Budapest Hungary Beijing 100176 China
DESCRIPTION F DA MODEL REV SERIAL
CUSTOMER NO.
INSTALLATION
DESTINATION
NAME AND
ADDRESS
ZIP CODE
46-303268P1 Rev 14
Chapter 4
Functional Checks
Section 4-1
Overview
4-1-1 Purpose of this chapter
This chapter provides procedures for quickly checking major functions of the LOGIQ E10 Ultrasound
system and diagnostics instructions using the built-in service software.
Section 4-2
Functional checks
Contents in this section
Table 4-1 Energy Control and Power Lockout for LOGIQ E10
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/
Signed Date
J10
OFF BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO
J11
4-2-2 Overview
General System Functional Checks
After setting up the Ultrasound system, performing an upgrade, or replacing hardware, perform all
appropriate functional checks before returning a system to the customer. Include the functional check
information in the service dispatch.
• Perform the functional checks and include the debrief script found at the end of the respective
replacement procedure in Chapter 8.
For a Service call with no FRU replacement or when installing the system:
• Perform the following basic functional tests. If all are successful, include the debrief script provided
below.
Some procedures are used more often than others. The intention with this section is to keep the most
used procedures in one place.
NOTE: NOT ALL functional checks will have to be repeated or performed if not all Options are present.
If ONLY a particular Option is present, perform the functional checks for that particular Option
and the Options installed.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
4-2-7 Basic Measurements
10-7-4 Grounding continuity
10-7-5 Chassis leakage current test
Power SHUT DOWN
DO NOT POWER SHUT DOWN until
4-2-4
ALL functional checks have been
performed
LOGIQ E10 Basic Service Manual, Direction 5750007-1EN,
In addition: Perform the following Functional
Rev. 1. Leakage Current measured at (record the value) and
Checks if the Option was or Options were removed
meets allowable limits. Equipment passed all required checks
during Backplane replacement.
and is ready for use.
4D
4-2-7 Probe/Connectors Checks
Perform a 4D sweep
CW
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
4-2-7 Probe/Connectors Checks
Chassis Leakage Current Test
10-8-4 Perform a CW scan with P2D, P6D,
M5S-D probes (plug in and scan)
See:
Section Functional Check Debrief Script
ECG
4-2-7 Probe/Connectors Checks
PW/CW Doppler Mode Checks
4-2-7
use Pencil Probes
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source)
10-7-8
leakage–lead to lead
Isolated patient lead (sink) leakage-
10-7-9
isolation test
ECG Checks
Active ECG:
Utility -> Applications ->
Settings -> ECG -> Show ECG
Tools.
4-2-7 Connect cables. Press Display ECG
Tab on TP. Verify ECG signal
displayed on monitor.
Plug in the ECG cables with nothing
attached so they pick up noise.
LOGIQ E10 Basic Service Manual, Direction 5750007-1EN,
Diagnostics
Rev. 1. Leakage Current measured at (record the value) and
Diagnostics meets allowable limits. Equipment passed all required checks
and is ready for use.
Shear Wave Elastography (Shear
4-2-7
Elasto) Functional Checks
Power Assistant
4-2-7 Probe/Connectors Checks
Verify Power Assistant Option
functions
Shear Wave
4-2-7 Probe/Connectors Checks
Shear Wave Elastography (Shear
4-2-7
Elasto) Functional Checks
Perform a 4D sweep, if 4D MC was
removed.
V Nav
4-2-7 Probe/Connectors Checks
3D/4D and Volume Navigation
4-2-7
Functional Checks
4-2-7 Basic Measurements
diags
NOTICE Use only power supply cords, cables and plugs provided by or designated by GE.
NOTE: Do not cycle the Circuit Breaker ON-OFF-ON in less than five (5) seconds. When turning OFF
the Circuit Breaker, the Ultrasound system should de-energize completely before turning the
circuit breaker ON.
Connecting AC Power to the LOGIQ E10 Ultrasound system involves preliminary checks of the power
cord, voltage level and compliance with electrical safety requirements.
NOTE: The LOGIQ E10 will function on Voltages from 100-240 Volts and 50 or 60 Hz. However, if using
220 volt power, then a center tapped power source is required.
1.) Ensure that the wall outlet is of appropriate type, and that the Circuit Breaker is turned off.
2.) Uncoil the power cable, allowing sufficient slack so that the Ultrasound system can be moved
slightly.
3.) Verify that the power cable is without any visible scratches or any sign of damage.
4.) Verify that the on-site mains voltage is within the limits indicated on the rating label near the Circuit
Breaker on the rear of the Ultrasound system.
5.) Connect the Power Cable’s female plug to the Power Inlet at the rear of the Ultrasound system.
6.) Lock the plug in position with the Retaining Clamp (ACC Clamp).
7.) Verify that the Mains Power Circuit Breaker is in OFF position, if not, switch it OFF.
8.) Connect the other end Power Cable (male plug) to a hospital grade mains power outlet with the
proper rated voltage, and the Ultrasound system is ready for Power ON/Boot Up.
1.) Switch ON the Mains Power Circuit Breaker at the rear of the LOGIQ E10. The ON/OFF button will
become white.
You should hear a “click” from the relays in the AC Power and the LOGIQ E10 is ready to boot. The ON/
OFF button will remain white.
2.) Press once on the ON/OFF button on the Operator Panel to boot the LOGIQ E10. The ON/OFF
button will turn blue when it is pressed.
During a normal boot, you may observe that:
a.) The ventilation fan starts on full speed, but slows down after a few seconds (listen to the fan
sound).
b.) Power is distributed to the peripherals, Operator Panel (Console), Monitor, Front End and Back
End.
c.) Back End and rest of scanner starts with the sequence listed in the next steps:
d.) Back End is turned ON and starts to load the software.
e.) The Start Screen is displayed on the monitor.
f.) A start-up bar indicating the time used for software loading, is displayed on the monitor.
g.) The software initiates and sets up the Front End electronics and the rest of the instrument.
h.) The backlight in the keyboard is lit.
i.) As soon as the software has been loaded, either a 2D screen is displayed on the screen,
indicating that a probe has been connected, or a No Mode screen is displayed, indicating that
no probe has been connected.
NOTE: Total time used for start-up is typically one and a half minutes or less. If starting after a power
loss or a lock-up, the start-up time may be up to four minutes.
Select comments and type different A/N keys to confirm keyboard functionality.
Footswitch
CAUTION To avoid injury or damage, make sure nothing is within the range of motion before moving the
Top Console. This includes both objects and people.
The system Top Console can be freely moved in any horizontal direction (if the power is on or if the
mechanism is not locked in position prior to powering down the system). The vertical displacement of
the Top Console is motor driven. The control buttons are located between the handles.
Table 4-4 Top Console adjustment controls: Lock (1), Up/Down (2)
To avoid injury or damage, make sure nothing is within the range of motion before moving
the Top Console. This includes both objects and people.
To manually release the console’s XY-mechanism (frogleg) if the power is OFF, see: Section 6-4
"Operator Panel movement" on page 6-13.
3
2
2
3
6 4b
4a
5 4c
6
Tilt
a. No Collision
1. Vertical Travel = 150mm 4. b. Approximately 20º
(based on Pan Arm Off-set)
c. 85º-90º
Lift Arm Rotation = 135º (270º 6. Pan Arm Rotation = 45º (90º
3.
total) total)
When the Monitor Arm Lock is unlocked. verify the Monitor can move up/down (vertically) and left/right
(horizontally).
NOTE: The Monitor can move approximately 100 mm (3.94 inches) vertically and
250 mm (9.84 inches) horizontally.
1. Use both hands, one at the top of the Main Monitor and
Tilt Forward/Back one at the bottom, to rotate the screen forward and
back.
2. Use both hands, one at the lower left handle hold of the
Move Left/Right Main Monitor and one at the lower right handle hold, to
adjust the screen left and right.
- Turn the knob clockwise to raise the lock and move the Monitor into the parked position.
Different functions can be assigned to the Trackball depending of the current active mode. The Trackball
area consist of:
• The Trackball: used as a cursor control in acquisition mode, scrolling control in freeze and as a
selection tool (like a mouse cursor) in post-processing mode.
• Three SELECT keys (identical): Perform the selected control or highlighted menu item.
• The TRACKBALL key: Toggles between the available Trackball function assignments displayed in
the Status bar.
• The update menu key: enables quick access to image related functions from a pop-up menu.
Other Controls
Other controls are located on separate keys (knobs and buttons) or groups of keys on the Operator
Panel.
GE recommends the RMI 430GS phantom is preferred, but it is not required. It is the most current
phantom recommended to our field service personnel and provides the necessary targets and extended
life necessary for consistent Ultrasound system testing.
Figure 4-7 Performance tests
Lateral Distance
Measurement
Penetration
2
4
Axial Distance
Contrast Measurement
Resolution
& 6
Gray Scale
Photography 4 3 2 1
8 Functional
Resolution
10
Lateral
Resolution 12
14
Axial 16
Resolution
RMI 403GS
Probe/Connectors Checks
NOTE: Probes can be connected at any time, whether the LOGIQ E10 is ON or OFF.
To connect a probe
1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable
damage.
4.) Turn the connector locking handle counterclockwise.
5.) Align the connector with the probe port and carefully push into place.
6.) Turn the connector locking handle clockwise to secure the probe connector.
7.) Carefully position the probe cable in the probe cord holder spot so it is free to move, but not
resting on the floor.
If the probe requires a Leakage Test, see: Section 10-7 "Electrical Safety Tests" on page 10-12.
B-Mode Checks
For information on the Ultrasound system’s different modes as well as expected results, refer to the
appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5, or the appropriate LOGIQ E10
Release Notes. See: Section 1-5 "Label locations" on page 1-26.
Introduction
Depending on probe availability, choose the preset application shown below. The LOGIQ E10
probes are ONLY approved for use in the applications listed in the LOGIQ E10 Basic User Manual,
Chapter 11, or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on
page 1-26.
Image Quality Check (IQC) is intended to facilitate Image Quality checks during Quality Assurance
Evaluations. Quality Assurance tests are used to determine whether a scanner is providing the same
level of performance year after year.
By using the same settings year after year, this ensures that the data collection is consistent,
independently of who performs the test.
1.) Activate IQC via Utility -> Imaging Preset Manager -> Category (select the Category first).
2.) Click on the plus sign in front of IQC for Service.
3.) Assign IQC to a Touch Panel key by using the right arrow key.
4.) Map the IQC to the location you want it to appear on the Touch Panel.
5.) Select Model -> IQC.
See On-line Help or refer to the appropriate version of the LOGIQ E10 Basic User Manual, Chapter 11.
See: Section 1-5 "Label locations" on page 1-26.
Preparations
1.) Connect one of the probes, to the scanner’s active probe connector.
- see: Section 3-4-6 "Connecting Probes" on page 3-19 for info about connecting the probes.
- For available probes, see: Section 9-17 "Probes" on page 9-39.
2.) Turn ON the scanner.
The B-Mode window is displayed (default mode).
3.) If needed, adjust the Display’s Brightness and Contrast setting (see: Section 6-2 "Monitor
adjustments" on page 6-2).
WARNING ALWAYS USE THE MINIMUM POWER REQUIRED TO OBTAIN ACCEPTABLE IMAGES
IN ACCORDANCE WITH APPLICABLE GUIDELINES AND POLICIES.
1.) Press B-MODE on the Operator Panel to access B-Mode.
2.) These Image Controls are used to optimize the B-Mode picture:
- Use Gain and TGC controls to optimize the overall image together with the Power control.
- Use Depth to adjust the range to be imaged.
- Use Focus to center the focal point(s) around the region of interest.
- Use Frequency (move to higher frequencies) or Frame rate (move to lower frame rate) to
increase resolution in image.
- Use Frequency (move to lower frequency) to increase penetration.
- Use the control to optimize imaging in the blood flow regions and make a cleaner, less noisy
image.
- Use Reject controls to reduce noise in the image.
Checks
M-Mode Checks
For information on the Ultrasound system’s different modes as well as expected results, the Basic User
Manual or User Guide will familiarize you with image optimization for B-Mode, M-Mode, Color Flow,
and Doppler.
Control Description
Power Output (Acoustic Optimizes image quality and allows user to reduce beam intensity. 10% increments between
Power) 0-100%. Values greater than 0.1 are displayed
Controls how echo intensities are converted to shades of gray, thereby increasing the
Dynamic Range
adjustable range of contrast.
Increases the number of transmit focal zones or moves the focal zone(s) so that you can
Focus Number and
tighten up the beam for a specific area. A graphic caret corresponding to the focal zone
Position
position(s) appears on the right edge of the image.
Selects a level below which echoes will not be amplified (an echo must have a certain
Rejection
minimum amplitude before it will be processed).
Temporal filter that averages frames together. This has the effect of presenting a smoother,
Frame Average
softer image.
Colorize Enables gray scale image colorization. To deactivate, reselect a Gray Map.
Gray Map Determines how the echo intensity levels received are presented as shades of gray.
Rotation (Up/Down) Rotates the image by selecting the value from the pop up menu.
Multi Frequency mode lets you downshift to probe's next lower frequency or shift up to a
Frequency
higher frequency.
Frame Rate/Resolution Optimizes B-Mode frame rate or spatial resolution for the best possible image.
Preparations
1.) Connect one of the probes, to the scanner’s active probe connector.
- see: Section 3-4-6 "Connecting Probes" on page 3-19 for info about connecting the probes.
- For available probes, see: On-line Help or refer to the appropriate version of the LOGIQ E10
Basic User Manual. See: Section 1-5 "Label locations" on page 1-26.
2.) Turn ON the scanner.
Checks
Introduction
For information on the Ultrasound system’s different modes as well as expected results, the Basic
User Manual or User Guide will familiarize you with image optimization for B-Mode, M-Mode, Color
Flow, and Doppler.
For complete information, refer to the appropriate version of the LOGIQ E10 Basic User Manual,
Chapter 5, or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on
page 1-26.
Preparations
1.) Connect one of the probes, to the scanner’s active probe connector.
- see: Section 3-4-6 "Connecting Probes" on page 3-19 for info about connecting the probes.
- For available probes, see: Section 9-17 "Probes" on page 9-39.
2.) Turn ON the scanner.
Color Flow screens are B or M-Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from B-Mode, or from M mode, or a combination of these.
For information, refer to the appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5,
or the appropriate LOGIQ E10 Release Notes.
For information, refer to the appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5,
or the appropriate LOGIQ E10 Release Notes.
For information, refer to the appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5,
or the appropriate LOGIQ E10 Release Notes.
For information on the Ultrasound system’s different modes, the Basic User Manual or User Guide
will familiarize you with image optimization for B-Mode, M-Mode, Color Flow, and Doppler.
Doppler is used to measure velocity (most often in blood). Doppler mode can be done with a special
pencil probe or with an ordinary probe. By using an ordinary probe, you can first bring up a B-Mode
picture for navigation purpose and then add Doppler.
For information, refer to the appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5,
or the appropriate LOGIQ E10 Release Notes.
Preparations
Adjust the Active mode gain to set the gain in the spectral Doppler area.
- Adjust Low velocity reject to reduce unwanted low velocity blood flow and tissue movement.
- In PW mode, adjust Sample volume to low setting for better resolution, or higher setting to
more easily locate the disturbed flows.
- Adjust the Compress setting to balance the effect of stronger and weaker echoes and obtain
the desired intensity display.
- Adjust Frequency to optimize flow display. Higher setting will improve resolution and the
lower setting will increase the depth penetration.
- Adjust Frame rate to a higher setting to improve motion detection, or to a lower setting to
improve resolution.
NOTE: Frequency and Frame rate settings may affect the Low Velocity Reject.
- Adjust Power to obtain an acceptable image using the lowest setting possible. This is
particularly important in CW mode, as the energy duty cycle is 100% (constant).
NOTE: The Doppler Power setting affects only Doppler operating modes.
- Adjust the following settings to further optimize the display of the image.
- Use the Horizontal sweep to optimize the sweep speed.
- To view signal detail, adjust Scale to enlarge the vertical spectral Doppler trace.
- Use Invert to reverse the vertical component of the spectral Doppler area of the display.
- Use Angle correction to steer the ultrasound beam to the blood flow to be measured.
Cineloop Checks
For complete information, refer to the appropriate version of the LOGIQ E10 Basic User Manual,
Chapter 5, or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on page
1-26.
To Activate CINE
To Omit Images
- Roll the Trackball to the frame you want to delete and press Omit Image.
To Restore Images
- To undo deleting an image from the CINE Loop, press Restore Image.
1.) Turn the Start Frame dial to the left to move to the beginning of the CINE Loop. Turn the dial
to the right to move forward through the CINE Loop.
2.) Turn the End Frame dial to the right to move to the end of the CINE Loop. Turn the dial to the
left to move backward through the CINE Loop.
- Turn the Loop Speed dial right/left to increase/decrease the CINE Loop playback speed.
- Turn Frame by Frame to move through CINE memory one frame at a time.
Checks
- Check Left Marker, Right Marker, Cycle Select and Number of Cycles
- Check First, Last Cycle, Select all
- Adjust Scroll and Cine Speed
Removable Media
Refer to the appropriate version of the LOGIQ E10 Basic User Manual, or the appropriate LOGIQ E10
Release Notes. See: Section 1-5 "Label locations" on page 1-26 to perform the following tasks:
For information, refer to the appropriate version of the LOGIQ E10 Basic User Manual, or the
appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on page 1-26.
See
Section Functional Test Debrief Script
LOGIQ E10 Basic Service Manual, Direction
7-4-4 Test of printers 5750007-1EN, Rev. 1. Equipment passed all
required tests and is ready for use.
Basic Measurements
NOTE: The following instructions assume that you first scan the patient and then press Freeze.
After you complete the measurement, to erase all data that has been measured to this point, but not
data entered onto worksheets, press Clear.
NOTE: Be careful not to press the Ellipse control as this activates the Body Pattern.
7.) Move the Trackball to position the ellipse and to size the measured axes (the calipers).
8.) To increase the size, turn the Ellipse control in a clockwise direction. To decrease the size, turn
the Ellipse control in a counterclockwise direction.
9.) To toggle between active calipers, press Measure.
10.)Press Set to complete the measurement.
11.)The Ultrasound system displays the circumference and area in the measurement results
window.
- To erase the ellipse and the current data measured, press Clear once. The original caliper is
displayed to restart the measurement.
- To exit the measurement function without completing the measurement, press Clear again.
Worksheets
Report Pages
Measurements/Calculations that are included on the worksheet can also be displayed on Report
Pages. Report Pages can be customized to meet the appropriate needs of the user.
For information on the Ultrasound system’s different modes as well as expected results, refer to the
appropriate version of the LOGIQ E10 Basic User Manual, Chapter 5, or the appropriate LOGIQ E10
Release Notes. See: Section 1-5 "Label locations" on page 1-26.
ECG Checks
Introduction
The ECG capability on this LOGIQ E10 is intended as a trigger for measurements, but can also be
viewed on the screen.
Parts needed
- ECG simulator
Preparations
None
ECG Checks
For complete information, refer to the appropriate version of the LOGIQ E10 Basic User Manual,
Chapters 5 and 6, or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on
page 1-26.
If the Capacitor Pack is not present or the Power Supply to Capacitor Pack Cable is not connected, or
connected properly, the option to select the “Shear Elasto” soft button after pressing the Elasto button,
the Touch Panel key “Shear Elasto/strain Elasto” button WILL NOT be present.
Section 4-3
Site Log
Table 4-11 Site Log
Chapter 5
Components and Functions (Theory)
Section 5-1
Overview
5-1-1 Purpose of this chapter
This chapter explains LOGIQ E10’s Ultrasound system concepts, component arrangement, and
subsystem functions. It also describes the Power Distribution System. Refer to the Basic User Manual
for more information.
Section 5-2
LOGIQ E10 models
Table 5-1 LOGIQ E10 Models
MODEL SOFTWARE
DESCRIPTION
NUMBER VERSION
LOGIQ E10,
5935000E10 R1.x.x
100-240 VAC
Section 5-3
LOGIQ E10 description
5-3-1 Purpose of this section
The purpose of this section is to give you an overview of LOGIQ E10 and how it functions.
5-3-2 Introduction
The LOGIQ E10 Ultrasound system is a high performance digital Ultrasound imaging System with total
data management.
The LOGIQ E10 provides image generation in B-Mode, Color Doppler, Power Doppler, M-Mode, PW,
4D, Harmonic Imaging, and Contrast imaging applications. The fully digital architecture of the
LOGIQ E10 Ultrasound system allows optimal usage of all scanning modes and probe types throughout
the full spectrum of operating frequencies.
System configuration is stored on the hard drive and all necessary software is loaded from the hard drive
on power up.
• Front End
• Back End
• OP Panel
• Peripherals
• Options
• Power Distribution
• Mechanical Parts
• Software
Section 5-4
Software overview
5-4-1 Purpose of this section
The intention with this section is to give you a brief overview of the software used on LOGIQ E10. You
can also refer to the Basic User Manual for more information.
Partition GB
Drive Type Usage
Letter (approximately)
C: Windows 10/application 79
SSD
Z: Software Repository 31
D: User settings/configuration 42
HDD E: Database/data 879
V: Future Use 10
• Software Installation:
- Software can be loaded/reloaded with USB media.
- C partition (Base+Application) can be reloaded from the Z partition.
- C partition (Base+Application) can be updated through a remote connection to the Service
Back Office.
• Network Capabilities:
- LAN is IPv6 ready.
- System can only be setup for IPv4 or IPv6. They cannot be mixed.
NOTE: If your PC does not have Adobe Reader, a free download is available on the Adobe website at
http://www.adobe.com.
1.) SysConfig.csv – Contains serial number, hardware info, software version, etc.
2.) log/ExceptionTrace – subdirectory contains text file with call stack of system crashes
3.) log/DL – (Deep Learning) subdirectory contains anonymized low resolution jpg images. Can be
used for troubleshooting. Every time you press the print key. Very low resolution images, may be
useful to get examples of image issues visible in low resolution. Used for data analytics. User can
disable them via UI.
4.) log/Diag – subdirectory contains diagnostics results
5.) log/Crashes – subdirectory may contain Windows/application crash info
6.) log/CrashDump – subdirectory may contain Windows/application crash info
7.) Log/ScLogsDatabase – subdirectory containing tab-separated text files (Data Logging)
8.) Log/ScLogsDatabase_PI – subdirectory with Patient Proprietary Info (keyboard shadow)
9.) WinEvt – subdirectory with Windows Event logs
Logs are text files and can be open in Excel. Fields are separated by tabs in order to make it easier to
be opened in Excel.
Logs will not be present there until they are collected through Alt+D or Gather Logs.
Via the Remote Control App, you will be able to adjust the following controls:
• Depth
• PW Gain
• CF Gain
• B-Mode Gain
• Left/Right
• Measure
• Comment
• Clear
• Trackball
• Freeze
• Store
• Color Flow Scale
• Cursor
Camera App
You can use the camera to capture images that will go to the patient’s exam.
Section 5-5
Service Desktop
The Service Desktop is an interface that provides access to system information, status and diagnostics.
The Service Desktop has different content or views depending on the access level. The access level is
determined by the user profile as well as the service options enabled on the LOGIQ E10.
• Basic view is the standard view, restricted only by the user through the user profile settings.
Administrator default user has access to the Service Desktop. Any user with "local Service access"
in their user profile can have access to this view.
• Class C view is the view enabled by the service options purchased.
- Service Advanced
- Service Expert (requires Service Advanced)
- Service Pro (requires Service Advanced)
Contents in this section
Diag History X X X
DICOM X X X
DICOM
X X X
Service Status
DICOM
X X X
Spooler
Utilities X X X X
Change
X X X X
Password
Disk
X X X
Defragment
Gather Logs X X X X
Delete Files X X X X
Reset Patient
X X
Database
Clean
X X
Userdefs
Network
X X X X
Capture
Third Party
X X X X
Licences
Checkpoints X X X
Disruptive
X X X
Mode Utility
Data Transfer X X X X
SSA License X X X X
Options X X X X
Agent
X X X X
Configuration
Section 5-6
Front End
5-6-1 Purpose of this section
The intention with this section is to give you a brief overview of the Front End configuration used on
LOGIQ E10.
Section 5-7
Back End
5-7-2 ECB
General Description
The LOGIQ E10 Carrier Board (ECB) is a single board computer built from a COM Express module fitted
into customized rack. Its main function is to be the central processing unit of the Ultrasound system,
controlling all the functionality, user interface and connectivity of the scanner.
Backplane connectors give the connectivity to the rack Backplane to interface to the other components/
boards installed in the rack/system.
Front connectors give the connectivity to interface to user-IO and components installed in the system.
The ECB also host a single slot GPU card, a standard SATA Solid State Drive (SSD) and a mini-PCI
Express module.
Figure 5-1 ECB/Back End Block Diagram
SMA
mini PCI-e Ant.
module
(WiFi) SMA
Ant.
Optional HDD
USB USB 2.0 ECG
On- board COMe USB 2.0
2.0
Skylake USB 2.0 VGU
USB 2.0 hub
USB 2.0
Main
CP U Memory USB 2.0
USB USB 3.0
SATA HDD
USB 3.0 USB 3.0 BW printer
On- board
PCIe S - VIDEO
HDMI Monitor
USB 3.0
SATA
SATA
ECB
w/ Power
On- Board
SATA
SATA
USB
InterConnection
FPANEL
OPIO
/ Fan
SATA Power
The ECB is located on the right side of the Card Cage, behind the I/O Board and is shielded from the
other Cards.
ECB LEDs
Figure 5-2 ECB LEDs
J5
J6
J7
A) J8
B)
C)
D)
J9
J10
5-7-3 SSD
The SSD (Solid State Drive) is connected to the ECB to further host the C (Windows 10 operating
system and application and utilities) and Z (Software Repository and additional files for Software
Reload) partitions of the HDD. No patient data is stored on the SSD.
Section 5-8
Backplane
5-8-2 Backplane
The LOGIQ E10 Backplane is part of an Ultrasound system, capable of processing 192 ultrasound
channels concurrently. It functions as interconnect for the different boards and Options in the
LOGIQ E10 Hard Ware platform.
Section 5-9
Top Console with Main Monitor and Operator Panel
5-9-1 Purpose of this section
The purpose of this section is to give you an overview of the top console and to tell you how it functions.
When the console is not locked (floating), press the lock button, this will only release the brakes in the
froglegs, to allow easy movement. When the lock button is pressed and the console is pushed back into
the locked position, the U-bolt mechanically engages the park lock (similar to a car door). Pushing the
lock button does not affect the locking action of the park lock, it only releases the brakes, to allow the
OP to get pushed into the park lock easier. But, the lock button must be pushed to get the OP to the
locked position. Once the OP is in the lock position and the lock button is released, you can hear the
sound of the park lock rotating to engage the U-bolt.
At shutdown, a message will appear ONLY if the Operator Panel is not parked. This message tells the
user that the Operator Panel is not parked.
The user can proceed to park the Operator Panel or to leave it un-parked.
Notice that ONCE the user selects shutdown in the Monitor, the 2 buttons (Z-motion and XY-motion)
are disabled. The brakes on the XY mechanism are released the XY movement is free but the Z-motion
is disabled. If the user wants to adjust the Z-Mechanism after shutdown, they will need to use the handle
in the back of the Ultrasound system.
NOTE: Remember, if the device brakes are not locked when trying to lock the OP, this can act as a
“shock absorber” and make it more difficult for the park lock to lock.
5 - 20 Section 5-9 - Top Console with Main Monitor and Operator Panel
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 5-10
Power distribution
5-10-1 Purpose of this section
The power distribution within the LOGIQ E10 is described in this section.
The Main Power Supply’s main task is to galvanically isolate the scanner from the on-site Mains Power
System and to supply the various internal subsystems with DC power.
Power from the wall outlet (100 to 240 VAC, 50/60 Hz) is connected to the Main Power Supply. The
Main Power Supply delivers the needed voltages to the rest of the LOGIQ E10:
The mains cord has plugs in both ends. A female plug connects to the scanner and a male plug to the
wall outlet.
Fuses are located inside the power supply (only to be replaced by the manufacturer).
General description
Figure 5-4 Main Power Supply location (rear of LOGIQ E10)
Temperature Control
The Main Power Supply is equipped with an internal fan with variable speed for temperature control.
Both the temperature of the air entering the power supply and leaving the power supply are measured.
Fan speed is controlled by the power supply.
Input
Section 5-11
Air Flow control
5-11-1 General description
The Air Flow Control includes the following components: Air cooling fans located at the bottom of the
LOGIQ E10 Card Cage and air cooling fans in the power supply that are self-managed by the power
supply itself.
Cooling air is drawn in, evenly into the filters located in the bottom of the Side Covers/Air Intake and the
Rear Bumper/Air Intake. The warmed air is exhausted from the Rear Cover/Air Exhaust.
Section 5-12
Casters and Brakes description
The front brake is used for everyday use; the back brake is ONLY used to transport the LOGIQ E10.
Section 5-13
Options and Peripherals descriptions
5-13-1 Purpose of this section
The Options and Peripherals within the LOGIQ E10 is described in this section.
NOTE: Only CD-R and DVD-R (write and read) are supported. CD-R/W and DVD R/W disks are not
supported.
• 4D Option
• CW Option
• Power Assistant Option
• Printer (Digital Graphic) Option
• Probe Holder Option
• Shear Wave Option
• EGG (Patient I/O) Option
• S-Video Option
• Volume Navigation Option
• Wireless LAN Option Option
• Bin(s) Option
• USB Flash Drives
• Color Inkjet Printer
• USB Hard Disk Drives
• Footswitch
• Dicom Printers
• Rolling V Nav Stand
• Service Options
Section 5-14
Product manuals
The information needed to use and service the LOGIQ E10 scanner is collected in the documents
described in this section.
NOTE: Dates on screenshots are represented in MM/DD/YYYY format throughout the manual.
Information on how to change the LOGIQ E10’s date can be found in Customizing Your System,
Chapter 10 in the LOGIQ E10 User Manual/User Guide.
Chapter 6
Service Adjustments
Section 6-1
Overview
6-1-1 Purpose of this chapter
This section describes how to adjust the LOGIQ E10.
Section 6-2
Monitor adjustments
6-2-1 Purpose of this section
This section describes how to adjust the Monitor for optimal performance.
All Monitor settings have been optimized at the factory, so normally, there should be no need for any
further adjustments.
Under special lighting conditions, you may want to adjust the Monitor screen’s backlight intensity.
• Brightness 50%
• Contrast 40%
• Room Setting: Dark
For complete information, refer to the appropriate version of the LOGIQ E10 Basic User Manual
Chapter 3, or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on page
1-26.
Brightness/Contrast
Room Profile 22 inch OLED
Dark 50/40
Semi-Dark 50/65
Light 50/100
Brightness
Touch panel encoder labeled as Touch Panel Light, rotate for brightness.
Push the mode button once to display the popup and select Unlock or wait for the timeout.
For complete information, refer to the appropriate version of the LOGIQ E10 Basic User Manual,
Chapter 3 or the appropriate LOGIQ E10 Release Notes. See: Section 1-5 "Label locations" on page 1-
26.
Gamma settings can be adjusted in two ways, with or without DICOM Gray Scale Standard Display
Function (GSDF) enabled.GSDF-compliant with all contrast settings, but not all brightness settings.
Because of differences in display technology, the 22-inch OLED monitor is GSDF-compliant with all
contrast settings, but not compliant with all brightness settings.
To enable the GSDF mode, in the touch panel, go to Utilities > System > System Display tab and
select Enable DICOM grayscale display mode (GSDF).
Figure 6-2 Check to enable GSDF Mode
If a site’s PACS workstations are calibrated to the GSDF, this setting may help to make the image
appearance more uniform between the Ultrasound system and PACS.
• With GSDF disabled, the monitor still uses a gamma curve that may be selected on the Touch Panel
(for backwards compatibility with sites that are happy with their PACS or may have a mix of older
systems).
• With GSDF enabled on Widescreen Monitors, the gamma button on the touch panel affects the
image, emulating the appearance of that gamma, but does not affect the monitor.
Set Use Wide Screen for… > Single Image to Off.
Test patterns are available to aid in evaluating the display visually and quantitatively.
Test
Pattern Description Corresponding Graphic
Confirm that you can clearly see all of the
different gradations of the grayscale in
each gray cell.
Gray Bars
Color Bars
Resolution
Test
Pattern Description Corresponding Graphic
Luminance for different grayscale levels. LN-1
Usually, the facility’s Biomed Engineer
tests each gray level map on the monitor
from LN-1 to LN-18 with a light meter and
then graphs these on a scale to ensure that
the curve falls within the GSDF’s standard.
This can be performed with all three
LOGIQ E9 monitors with GSDF enabled.
Generally, this is performed yearly.
LN1 to LN18
LN-18
QC
Test
Pattern Description Corresponding Graphic
Contrast. Confirm that you can detect the
circles within the square grayscale boxes.
CT
Used for
Luminance
response
For complete information, Refer to the appropriate version of the LOGIQ E10 Basic User Manual,
Chapter 3. See: Section 1-5 "Label locations" on page 1-26.
Section 6-3
Monitor Arm and/or Monitor range of motion Adjustment procedure
Contents in this section
Ergotron Arm
Adjustments available
Monitor Range of motion and
Monitor Arm vertical adjustment.
6 - 10 Section 6-3 - Monitor Arm and/or Monitor range of motion Adjustment procedure
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
6-3-3 Main Monitor Arm and Monitor Friction Adjustment - Ergotron Arm
Monitor Arm Vertical adjustment
1. Tilt the Monitor down to access the Monitor Arm Vertical Adjustment Screw
Vertical Adjustment Screw.
If the Monitor is too loose or too tight when it is tilted, perform a Tilt adjustment.
6 - 12 Section 6-3 - Monitor Arm and/or Monitor range of motion Adjustment procedure
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 6-4
Operator Panel movement
6-4-1 Purpose of this section
The Operator Panel movement within the LOGIQ E10 is described in this section.
WARNING
Operator console can drop unexpectedly when the mechanism is released.
WARNING If the Operator Control Panel gas spring is not functioning correctly, there is a chance
that the operator panel could drop down suddenly when you use the manual release
handle. To avoid injury, take care that you are not positioned below the operator panel
when using the manual release handle.
To avoid injury or damage, make sure nothing is within the range of motion before moving
the Top Console. This includes both objects and people.
Section 6-5
Adjust time-out for DICOM servers
If you are experiencing problems with slow responses from DICOM servers, increase the time-out in the
DICOM server properties dialog. (Utility -> Connectivity -> Service -> Properties -> Maximum
Retries and Timeout.)
Problems with slow responses may result in images being re-sent automatically and low transfer rates.
The retry settings can be used to make jobs retry on bad networks.
When portable (off-line), use minimum time-out and no retries or it will affect shutdown speed.
Section 6-6
Volume Navigation Calibration Procedure
There is no calibration available in the field. If the registration seems to be out of calibration, reloading
the software will restore the factory settings. If still out of calibration, replace hardware.
Chapter 7
Diagnostics/Troubleshooting
Section 7-1
Overview
7-1-1 Purpose of this chapter
This chapter describes how to setup and run the tools and software that help maintain image quality and
system operation. Basic host-, system-, and board-level diagnostics are run whenever power is applied.
Some Service Tools may be run at the application level.
Section 7-2
Gathering troubleshooting data
Contents in this section
- Software Version
- Software Revision
- Software Part Number
- Build Name
- Build View
- Build Date
Under System Image, record:
• To include a duration, check Time Duration and select a duration from the dropdown.
• Under Destination, select a storage media or Service directory for remote viewing.
• Click the Store button.
NOTE: To save to a CD/DVD or USB Flash Drive, you MUST select CD/DVD Recordable or USB drive
as the destination device, otherwise the data is written to the default Export/Service directory
on the hard drive. The Export/Service directory is only used for remote service and is not
intended for images or report storage use.
For a CD/DVD, the LOGIQ E10 automatically formats an un formatted disk, gathers logs and writes it
out to the disk.
The subsequent file is compressed and time stamped. The screen capture is a bitmap which eliminates
the possibility of artifacts from compression.
Figure 7-3 Contents of the Zipped Trouble Image and Logs File
If a customer is experiencing issues during operation, mark the issue by pressing Alt+1 or Alt+2 when
the event occurs. When Alt+1 or Alt+2 is pressed, a marker is placed in the log to aid analysis.
To collect a trouble image you need to enable Transfer Images (captured without patient information)
to GE under ADMIN > System Admin.
To collect the image, press the P1 key (copy to hard drive). This places an unidentified .jpeg image in
the D:\Log \DL folder. This folder will be transferred to the back office if a transfer is scheduled.
To collect logs using Gather Logs, the image or images will be captured in the log collection Zip file and
will be deleted from the D:\Log \DL folder.
1.) On the LOGIQ E10, run the Network Capture utility to generate .etl and .cab files for analysis. For
more information, see "Network Capture" on page 7-29.
2.) Locate the .etl file at D:\Service.
3.) Transfer these files to a laptop.
4.) On the laptop, download Microsoft Message Analyzer (if not already done).
5.) In Microsoft Message Analyzer, open the .etl file.
If the DICOM connection is encrypted, you will not be able to see the contents of the packets.
Microsoft Message Analyzer does not have a plug-in to decode DICOM packets.
6.) To export the .etl file to .cab, select File > Save As and then select Save As or Export.
If the network capture was of a wireless connection, Microsoft Message Analyzer creates a .pcap
file that does not have any of the packets and is probably useless.
7.) Use Wireshark or some other third-party network sniffer to examine the .cab file.
Section 7-3
Service desktop
Contents in this section
• Service Basic access (Class A) – a user locally logged into the machine with Local Service Access
privilege. This level provides limited access to Service desktop widgets and utilities.
• Service Expert, Pro, and Advanced access (Class C) Local – Depending on the purchase level,
includes an option string to control access.
• GE Service access (Class M) and an SSA key. For users with local Service Access privileges, this
level provides unrestricted access to all Service desktop widgets and utilities.
• Remote access – a user remotely accessing the LOGIQ E10. This level provides unrestricted
access to all Service desktop widgets and utilities. Disruptive mode is limited to the user access
privileges to Remote Service Access.
Specifically, Disruptive mode is required to run diagnostics, clean presets, and reset the patient
database, and turn on Virtual Console Observation (VCO).
• When Disruptive mode is On, all service functionality on the Service desktop is allowed but user
operation of the LOGIQ E10 may be limited.
• When Disruptive mode is Off, some service functionality on the Service desktop is not available and
user operation of the LOGIQ E10 is normal.
Additionally, the ability to enable Disruptive mode depends on the logged in user.
• Local user – a user locally logged into the machine will be able to set the LOGIQ E10 to Disruptive
mode or allow a Disruptive mode request from a remote user through the Service desktop. The local
user must have Authorize Remote Service Access to allow Disruptive mode. If the local user does
not have this right, the remote user’s request will be automatically denied.
• Remote user – a user remotely accessing the LOGIQ E10 will not be able to automatically switch
Disruptive mode to On. The logged in user (user actually logged on to the LOGIQ E10) needs to
have the ability to grant remote access. The logged in user will be notified through a dialog box and
asked to allow Disruptive mode.
NOTE: Change Password and Disk Defragment are not available for the remote user whether
Disruptive mode is On or Off.
Chapter 7 Diagnostics/Troubleshooting 7 - 11
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
7-3-4 Licenses
With Service Basic Access (Class A), these are the available options:
• HOME
• Utilities
- Change Password
- Data Transfer
- Delete Files
- Gather Logs
- Network Capture
- SSA License
- Third Party Licenses
• Options
• Agent Configuration
With Service Advanced (Class C), these are the available options:
NOTE: With a Class C license, options display according to these purchased level of access.
• HOME
• Diags
• DICOM
• Utilities
- Change Password
- Checkpoints
- Data Transfer
- Delete Files
- Disk Defragment
- Disruptive Mode Utility
- Gather Logs
- Network Capture
- SSA License
- System Shutdown
- Third Party Licenses
- Virtual Console Observation
• Options
• Agent Configuration
With Service Advanced plus Service Expert (Class C), the Clean Userdefs, Reset Patient Database,
and Software Reload utilities are added to the Service Advanced options listed.
With Service Advanced and Service Expert plus Service PRO (Class C), the probe assessment tool (e-
PAT) diagnostic is added to the Service Advanced and Service Expert options listed.
With GE Service access (Class M) and an SSA key, these are the available options:
• HOME
• Diags
• DICOM
• Utilities
- Change Password (not available through a remote connection)
- Checkpoints
- Clean Userdefs
- Data Transfer
- Delete Files
- Disk Defragment (not available through a remote connection)
- Disruptive Mode Utility
- Gather Logs
- Network Capture
- Reset Patient Database
- Software Reload
- SSA License
- SSH
- System Shutdown
- Third Party Licenses
- Virtual Console Observation
• Options
• Agent Configuration
For more information, see Section 5-5 "Service Desktop" on page 5-10.
Chapter 7 Diagnostics/Troubleshooting 7 - 13
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
7-3-5 Home
Home configurations vary depending upon the purchased service level.
System Information
System Information displays general information about the LOGIQ E10. When the LOGIQ E10 has
been successfully configured with the back office, these elements will have the corresponding values:
Chapter 7 Diagnostics/Troubleshooting 7 - 15
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
This table shows all the elements available on System Information with descriptions.
Element DESCRIPTION
Model Number GE part number for the LOGIQ E10. The same number as listed on the rating plate.
Serial Number Serial number of the LOGIQ E10. The same number as listed on the rating plate.
Facility Name of the hospital or facility where the LOGIQ E10 is installed.
Software Status
Use Software Status to view general information about the software installed on the LOGIQ E10.
This table shows all the elements available on Software Status with descriptions.
Element DESCRIPTION
System Date Current date in the format <day>, <month> <date> <year>.
Local time based on the last time the system desktop was refreshed in the format
System Time
<hh:mm:ss>.
Date the application software was installed. The application software includes the
Application Installation Date
LOGIQ E10 product-specific software.
Date the base image software was installed. The base image software includes the
Base Image Installation Date
Windows operating system and other supporting software.
Chapter 7 Diagnostics/Troubleshooting 7 - 17
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Connected Probes
Connected Probes shows probes connected to the LOGIQ E10. The order on the user interface is top
down matching the left-to-right order on the LOGIQ E10.
This table shows all the elements available on Connected Probes with descriptions.
Element DESCRIPTION
Serial number of the probe connected to the LOGIQ E10. If the serial number of the
Serial Number
probe is not available, then N/A displays.
Statuses of the probe connected to the LOGIQ E10. Valid values are:
Status • Active
• Non Active
7-3-6 Utilities
Utilities configurations vary depending upon the service class.
Chapter 7 Diagnostics/Troubleshooting 7 - 19
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Gather Logs
Gather Logs provides a way to collect system logs and place the log files in the D:\Service directory for
retrieval by the online center. These log files do not include protected data such as crash dumps and
keyboard shadow logs. The customer can collect logs (including protected data) using Alt+D when
Protected Data is checked. Log files are compressed into a .zip file and the file path and name display.
If the application software is not running, use the Gather Logs shortcut on the Windows desktop.
To access Gather Logs, select Utility (second page) > Service > Utilities > Gather Logs.
This table shows all the elements available on Gather Logs with descriptions.
Element DESCRIPTION
1 Day Logs When selected, gathers log files for one day.
1 Week Logs When selected, gathers log files for one week.
Gather Logs Select to gather the log files for the selected time period.
1.) Navigate to Utility (second page) > Service > Utilities > Gather Logs.
2.) Select one of the following:
- 1 Day Logs
- 1 Week Logs
- All Logs
3.) Click Gather Logs. In the resulting dialog box, record the location of the log files and click OK.
4.) When the gather log operation is complete, click the notification icon in the banner to view the
location of the log files.
Delete Files
Delete Files displays all the files and folders present in the D:\Service folder and allows for their
deletion. Deleting unneeded files improves performance and reduces the need to defragment the disk
drive.
To access Delete Files, select Utility (second page) > Service > Utilities > Delete Files.
This table shows all the elements available on Delete Files with descriptions.
Element DESCRIPTION
Delete Files Displays the files that are available for deletion.
To delete files:
1.) Navigate to select Utility (second page) > Service > Utilities > Delete Files.
2.) Under Delete Files, select the available folders and files that you want to delete.
3.) Click Delete.
4.) In the resulting dialog box, click Delete and then click OK.
For more information, see:
Chapter 7 Diagnostics/Troubleshooting 7 - 21
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Change Password
Change Password allows you to change the password for a specified user type.
To access Change Password, select Utility (second page) > Service > Utilities > Change
Password.
This table shows all the elements available on Change Password with descriptions.
Element DESCRIPTION
1.) Navigate to Utility (second page) > Service > Utilities > Change Password.
2.) Under User Type, select the user.
NOTE: Before changing the GEService password (the default is SvcForward123$), make sure the
LOGIQ E10 is connected to the network and the agent is configured. The GEService password
is used to perform portions of remote service. If the password is changed and the system
information is not updated, it may slow down remote service. Both file transfer and SSH depend
on the GEService password.
3.) In New Password and Confirm Password, enter the new password.
Chapter 7 Diagnostics/Troubleshooting 7 - 23
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Third Party Software Licenses displays the third-party software licenses used as part of the service
platform.
The information on Third Party Software Licenses is available to all service class licenses.
To access Third Party Software Licenses, select Utility (second page) > Service > Utilities > Third
Party Software Licenses.
SSA License
To access SSA License, select Utility (second page) > Service > Utilities > SSA License.
This table shows all the elements available on SSA License with descriptions.
Element DESCRIPTION
Drive Letter Drive where the SSA key is plugged into the LOGIQ E10.
Key Counter Value Number of times the SSA key has been used.
Max Key Counter Value Number of remaining times the SSA key can be used.
Status of the access to the associated service class option. Valid values are:
Status • True
• False
Chapter 7 Diagnostics/Troubleshooting 7 - 25
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Data Transfer
To access Data Transfer, select Utility (second page) > Service > Utilities > Data Transfer.
This table shows all the elements available on Data Transfer with descriptions.
Element DESCRIPTION
Type of log file. For example, Incremental Logs or Full Logs. Monitoring Logs,
System Logs, and Windows Logs are incrementally transferred when automatic
Type of Upload
transfer is enabled. To enable automatic transfer, navigate to System Admin and,
under Service, check Enable Automatic Request for Service.
Upload Permission Whether the permission to upload the log file is allowed or not.
Last Upload Status Whether the last log file upload was successful or not.
Last Upload Attempt Date and time the last log file upload was attempted.
Last Successful Upload Date and time the last log file was successfully uploaded.
When selected, enables the related day selections. For example, All Days,
Scheduler
Monday, and Tuesday.
Send All Manually send the selected log files to the server.
Chapter 7 Diagnostics/Troubleshooting 7 - 27
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Network Capture
Network Capture displays network traffic between the LOGIQ E10 and configured devices. A network
capture outputs two log files: one for main logging with no protected information and another including
protected information. These log files are useful when debugging connectivity issues. Because these
log files can be large, they are only kept for one week.
To access Network Capture, select Utility > Service > Utilities > Network Capture.
This table shows all the elements available on Network Capture with descriptions.
Element DESCRIPTION
Select to start the process. This causes the network capture to start, enables the
Start Network Capture Stop button, and updates the Network Capture Status pane and changes the
Status to Running.
Displays information about the status of the network capture. The language setting
for this information is set in Windows and not through the Service desktop or LOGIQ
E10 application software.
Displays the current status of the network capture. Valid values are:
• Not Running
• Running
Chapter 7 Diagnostics/Troubleshooting 7 - 29
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
1.) Navigate to Utility > Service > Utilities > Network Capture.
4.) Click the Stop button to end data collection. Stopping is a two-step process:
- Stops the data collection and immediately closes the .etl file.
- Collects additional diagnostic data that may help diagnose network issues. When the file is
closed, you see “There is no trace session currently in progress”. When the remaining data is
collected and the .cab file is closed, you are notified in the banner.
7-3-7 Options
Use Options to:
This table shows all the elements available on Options with descriptions.
Element DESCRIPTION
Software Options
Number for the hardware. The hardware number is the hash of the serial number
Hardware Number
that is used to generate the option key.
Available Keys List of the option keys installed on the LOGIQ E10.
Chapter 7 Diagnostics/Troubleshooting 7 - 31
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
To access Agent Configuration, navigate to Utility > Service > Agent Configuration.
This table shows all the elements available on Agent Configuration with descriptions.
Element DESCRIPTION
Agent Configuration
Phone number for the person at the customer site a GE remote service engineer
Contact Details would contact. The phone number is entered during installation and reviewed at
every service call to make sure the information is correct.
Element DESCRIPTION
Agent Model Number GE part number for the LOGIQ E10. The same number as listed on the rating plate.
Serial number of the agent (read-only). If the agent is not registered with a serial
Serial Number number, this field is populated with the serial number of the LOGIQ E10. The serial
number of the agent is tied to the serial number of the LOGIQ E10.
Advanced Configuration
Proxy Configuration
Proxy Server When Enable Proxy is selected, name of the proxy server IP.
Proxy Port When Enable Proxy is selected, number of the proxy server port.
Credentials
Password When Enable Proxy Credentials is selected, password for the user.
Chapter 7 Diagnostics/Troubleshooting 7 - 33
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 7-4
Troubleshooting
Contents in this section
Connect the LOGIQ E10 to another power outlet and check to see if the noise changes or disappears.
NOTE: Image artifacts can occur if, at any time within the facility, the ground from the main facility's incoming
power source to the LOGIQ E10 is only a conduit. See: 2-3-4 "Minimal floor plan suggestions" on page
2-11.
Different system
Try another LOGIQ E10 at the same location and look for the same noise. If the noise is present on the
new system too, the noise is most likely from an external source/equipment.
Different location
Move the LOGIQ E10 to another location and check to see if the noise changes or disappears. This may
help to locate an external noise source.
Disconnect all external cables (network, all unused probes, ECG leads, etc.) and check to see if the
noise disappears.
Chapter 7 Diagnostics/Troubleshooting 7 - 35
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
7-4-2 Noise
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. Possible EMI sources
should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Some of
these sources include:
• medical lasers
• scanners
• cauterizing guns
• computers
• monitors
• fans
• gel warmers
• microwave ovens
• light dimmers
• portable phones
The presence of a broadcast station or broadcast van may also cause interference.
When talking to the customer, try to gather as much information as possible about the conditions when
the noise appears.
At all locations in the hospital n/a See "Noise picked up from the air" on page 7-37.
See "Noise picked up from the air" on page 7-37 and "Noise
Only in one room or area The room or area
received through the external cables" on page 7-37.
From time to time with no special See "Noise picked up from the air" on page 7-37 and "Noise
n/a
pattern of time is observed received through the external cables" on page 7-37.
Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the wiring.
The noise can enter the system through the mains power cable, probe cable(s) or any other external
connected cable(s).
Disconnect cables that are not needed for the basic use of the LOGIQ E10:
• Network cable
• Cables to any external peripherals
• ECG cables and other cables connected to the Patient I/O
Verify whether the noise changed or disappeared when the cables were removed.
Often, this type of noise is due to grounding problems in the mains power system or when the LOGIQ
E10 is sharing a power line with other equipment.
Doppler noise
Intermittent noise
Symptom Solution
Intermittent noise Check equipment that is turned on and off near the scanner.
Chapter 7 Diagnostics/Troubleshooting 7 - 37
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Noise on a board
Symptom Solution
Isolate the board to see if you can determine the specific board
where the noise originates.
There is noise on a board
For example, unplug the receive area of the board and see if
the noise persists.
Key Description
Alt+1 or Alt+2 Marks events in the log file. Used for the flagging issues during operation.
Displays SW Shut Down. If the LOGIQ E10 can be shut-down by using
Alt+F10 these keys, it means that SW is working and the problem could be the
ON/OFF switch or the wire that carries the shut-down signal.
For the echo simulator (keyboard simulation), brings up or removes the
Alt+F5 or F11 front panel simulator. It may be used with VCO to press buttons on the
front panel that are not standard keys on your PC.
Alt+A Displays a cursor (Alt+A stands for arrow).
Alt+D Captures scanner logs.
With the SSA key connected, displays the Debug command window to
Alt+Ctrl+D
run Sclogs.
With the SSA key connected, switches between screens. Access
Alt+Tab
Windows and relaunch the application to use this functionality.
1.) Open the Printers folder, either from Start > Settings > Printers or Utilities > System > Printers.
2.) Right click on a printer and select Preferences.
3.) Select Print Test Page (this sends a print to the printer bypassing all of the scanner software).
4.) Observe the page. If the page prints out, the problem is probably a Windows configuration issue or
a configuration issue with Connectivity. If the page does not print out, there probably is a cabling
issue or a printer configuration issue in Windows.
NOTE: For the Sony small format printers you will see the test page clipped. This is normal.
Chapter 7 Diagnostics/Troubleshooting 7 - 39
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Chapter 8
Replacement Procedures
Section 8-1
Overview
8-1-1 Purpose of this chapter
This chapter describes how to remove and install, or replace, modules and subsystems in the
LOGIQ E10. It also includes instructions for installing and re-installing the software.
Section 8-2
Warnings and important information
8-2-1 Purpose of this section
This section includes important information. Read it before doing any of the procedures in this chapter.
8-2-2 Warnings
Table 8-1 Preparations and Preparation Links
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
J10
TAG
&
LOCKOUT
J10 J11
Signed Date
J11
J14
J13
WARNING AT LEAST TWO PERSONS ARE NEEDED WHEN REPLACING CASTERS (WHEELS) OR
ADJUSTING BRAKES.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO
ONE OF THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR
THAT PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
ESD Receptacles
Battery Box side Card Cage side
GE policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have
been properly decontaminated prior to shipment. Under no circumstance should a part or equipment
with visible body fluids be taken or shipped from a clinic or site (for example, body coils or an ultrasound
Probe). The purpose of the regulation is to protect employees in the transportation industry, as well as
the people who will receive or open this package.
NOTE: The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purposes and must be
transported as a hazardous material.
• Casters Replacement
• Z-Mechanism replacement
The rest of the replacement procedures can be carried out by one person.
Section 8-3
Definitions of Left, Rear / Back, Right and Front
Figure 8-1 illustrates what is the Left Side (A), Rear / Back (B), Right Side (C), and Front (D) of the
LOGIQ E10.
Figure 8-1 Definitions of Left, Rear / Back, Right and Front
Section 8-4
Loading / Reloading / Upgrading the Software
8-4-1 Purpose of this section
This section describes how to install and/or re-install software on LOGIQ E10.
MODEL SOFTWARE
DESCRIPTION
NUMBER VERSION
LOGIQ E10,
5935000E10 R1.x.x
100-240 VAC
Refer to the latest revision of the LOGIQ E10 Basic User Manual, Chapter 9 Recording Images.
Speak with the personnel at the site to determine which patient images need to be backed up prior to
starting.
CAUTION Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!
NOTICE An error, or a power loss may occur during the software loading.
Always backup the Patient Images, Patient Archive and the System Configurations before starting a
software loading!
In order to complete a successful restore of the Patient Database, if needed, the images must be
moved away from the LOGIQ E10 before doing backup of Patient Database. Depending on the location
set-up, either move the images to a remote server or to removable media like DVD or CD discs.
If the backup procedure is not completed correctly, the images and database information could be lost.
Manpower /
Tools
Time
One person /
Software Thumb Drive.
2 Hours
Overview
The Hard Disk Drive on LOGIQ E10 is divided into five partitions.
NOTE: The Hard Disk Drive on LOGIQ E10 has three visible partitions (two partitions are hidden).
C: SYSTEM
D: USER
E: ARCHIVE
When installing (and re-installing) software, you get the choice to either replace all the contents on the
Hard Disk Drive, or only install the software on the C: partition. If you chose to replace all the contents
on the Hard Disk Drive, all stored data, including all set-up data will be lost. It is considered to be the
best practice to always move images away from the Hard Disk Drive, do backup of databases etc. and
record important configuration data on paper before starting a software load or reload, as described
earlier in this section.
The software is delivered on one Thumb Drive with the Base Load Image and LOGIQ E10 Application
Software. In some cases, a second Thumb Drive with a software patch, may also be included. A
software patch is used when a few new files should be installed or changed, but the changes do not
require a complete software load.
When installing the software, start with the system software, then the application software and if
included, install the patch software at the end. This is described in detail over the next pages.
NOTE: System Software may also be referred to as the Base Image or Ghost.
Before the System Software Load, you must perform the following steps:
• contact the customer to allow enough lead time to back up the system.
• be sure no images are in the clipboard or opened exams.
• review troubleshooting section to collect information for future analysis if you are troubleshooting
loss of image or patient data.
• disconnect the system from the network and remove all transducers from the system.
• be sure all USB devices (printers only) are connected and turned on (except for USB Flash Drives).
If a printer is plugged in at this time, it is automatically installed.
• save all Service Presets to restore remote connectivity after software installation is completed.
• save all User Presets.
NOTE: After a full or partial system software load, confirm the Windows automatic Daylight Saving
Time feature is turned off (Utility -> System -> General -> Date/Time -> Time Zone tab ->
Automatically Adjust Clock checkbox). Be sure the system and DVR date, time and time zone
are set correctly.
NOTE: System and Application software disables the Run window normally accessed from Start ->
Run, and the mouse right-click (left set key) feature.
CAUTION Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!
1.) From the Touch Panel, select Utility -> System -> About.
2.) If upgrading the Software, verify software version matches the version in the software used during
installation, version received with Kit.
NOTE: Software Option Keys are software version dependent. If a new software version has been
installed, please follow installation instructions for that software. Usually, with new software
Option Strings are needed. When re-installing the same software, the Software Option Strings
should be reused.
1.) From the Touch Panel, select Utility -> Admin -> System Admin.
2.) Verify that the options previously installed in the system are still activated after the software
installation.
3.) If any option is missing, request new Option Strings from the On Line Support. Enter any new or
missing option key string(s) in S/W Option Key field and select Add, or by importing it from USB or
DVD drive.
1.) From the Touch Panel, select Utility -> Connectivity -> TCP/IP.
2.) Verify TCP/IP settings.
3.) Select “Save settings.”
4.) Select OK to continue.
5.) Reboot system.
1.) From the Touch Panel, select Utility -> Connectivity -> Device.
2.) Verify Device settings.
3.) Select “Save settings.”
4.) Select OK to continue.
Restore the Patient Archive or Report archive using the Utility -> System -> Backup/Restore Function
or import the Patient Data if previously exported. If the patient archive is encrypted, you will need the
Encryption Key or password to unlock.
1.) From the Touch Panel, select Utility -> System -> Backup/Restore.
2.) On the Restore screen, select System configuration to restore System settings/User Defined
Configuration.
3.) Select the appropriate Source Device.
CAUTION The Restore procedure will OVERWRITE the existing data on the local harddrive. Make sure to select
the correct source device.
4.) If restore is done from a backup on a removable media, insert the media in the drive.
5.) Select Restore Now.
Depending on the selection, one or two Restore confirmation windows are displayed.
6.) Ensure that the correct source is selected, and select OK.
The selected items are copied to the LOGIQ E10. If items from the System Configuration are
restored, the LOGIQ E10 needs to be rebooted. The Reboot system window is displayed.
7.) Select OK to reboot the system, and verify the Customer Presets are working properly.
8.) Import scan assistant programs previously exported.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
CAUTION Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!
1.) From the Touch Panel, select Utility -> System -> About.
2.) Verify software version matches the version in the software used during installation, version
received with Kit.
1.) From the Touch Panel, select Utility -> Connectivity -> TCP/IP.
2.) Verify TCP/IP settings.
3.) Select “Save settings.”
4.) Select OK to continue.
5.) Reboot system.
1.) From the Touch Panel, select Utility -> Connectivity -> Device.
2.) Verify Device settings.
3.) Select “Save settings.”
4.) Select OK to continue.
Confirm that the Patient Archive is available. If the patient archive is encrypted, you will need the
Encryption Key or password to unlock.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools
Time
One person /
Software Thumb Drive.
1 Hour
1.) From the Touch Panel, select Utility -> System -> About.
2.) Verify software version matches the version in the software used during installation, version
received with Kit.
1.) From the Touch Panel, select Utility -> Connectivity -> TCP/IP.
2.) Verify TCP/IP settings.
3.) Select “Save settings.”
4.) Select OK to continue.
5.) Reboot system.
1.) From the Touch Panel, select Utility -> Connectivity -> Device.
2.) Verify Device settings.
3.) Select “Save settings.”
4.) Select OK to continue.
Confirm that the Patient Archive is available. If the patient archive is encrypted, you will need the
Encryption Key or password to unlock.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools
Time
One person /
No tools required.
1 Hour
Use this procedure to reload Base Load and Application Software from the repository partition in the
harddrive.
CAUTION Before loading Software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR, if the
patient archive has been encrypted and if the customer has the password or the Key to unlock the data.
Failure to obtain those items, will cause permanent DATA LOSS!
1.) From the Touch Panel, select Utility -> System -> About.
2.) Verify software version matches the version in the software used during installation, version
received with Kit.
Check Patient Archive
Confirm that the Patient Archive is available. If the patient archive is encrypted, you will need the
Encryption Key or password to unlock.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Section 8-5
Cable Replacement
When replacing cables, see the Section in which the cable or cables are used. When replacing the Main
Cable Harness, see:8-8-4 "Z-Mechanism replacement" on page 8-199.
Section 8-6
Replacing Covers and Bumpers
8-6-1 Purpose of this section
This section describes how to replace the Covers and Bumpers on the LOGIQ E10.
The Main Console has a Front Cover, two Side Covers with a Foot Rest (front), Bumper (rear), a Rear
Cover with a top trim and a Top Cover.
The Side and Front Covers, Foot Rest, Rear Bumper and Rear Cover Trim are positioned and secured
to the LOGIQ E10 with custom designed, self-clinching studs (A) and a catch (B). The catch (B) is
mounted loosely to help guide the stud into place, to ease removal/attaching and provide a better fit for
the covers.
B
A
In addition to the self-clinching studs and catch, the Side Covers are positioned and secured with a
series of tabs (C) at the top, inside of the Cover and slots (D) in the Top Cover. The Front Cover is
positioned and secured with guides (tabs E) to the Top Front Cover.
D E
Contains:
1. Pan Arm Down
Cover
Main Monitor Arm 2. Lift Arm Cover -
Right Side Cover Right
Plastic
Assembly/Air 3. Lift Arm Cover -
Covers (for,
Intake Top Cover Left
Ergotron Arm
See: 8-6-3 "Side See: 8-6-10 "Top
1 1 Assembly). See: 4. Joint Cover
Cover(s)/Side Air Cover 5. Extension Arm 1
8-6-13 "Main
Intake (Filter) replacement" on Cover
Monitor Arm
inspection / page 8-55.
Assembly Covers 6. Rotation Limit
replacement" on
replacement" on Screw
page 8-33. 3
page 8-64. 1
4
2
6
5
Caster Hubcap
Foot Rest Bumper Rear Handle
(cover)
See: 8-6-5 "Foot See: 8-6-12 "Rear
See: 8-10-4
Rest Bumper 1 Handle 1 4
"Caster Hubcap
replacement" on replacement" on
replacement" on
page 8-41. page 8-62.
page 8-236.
Pedal Mechanism
Printer Filler Covers:
Storage Parking Brake (1) 1
Front Cover
Swivel Lock (2)
See: 8-6-4 "Front See: 8-6-14
See: 8-10-7
Cover 1 "Printer Filler 1 1
"Pedal
replacement" on Storage 2
Mechanism
page 8-38. replacement" on Covers
page 8-66. replacement" on
page 8-250.
The Side Covers must be removed before any lower console cover(s) can be removed.
Manpower /
Tools / PPE
Time
One person /
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
5 minutes per
8-2-6 "PPE Required During Service" on page 8-6.
Side Cover
The Side Covers are attached to the LOGIQ E10 with tabs at the top, inside of the Cover and slots in
the Top Cover, and three self-clinching studs. One is located at the front, center and two are located at
the bottom of the cover. An M4 screw, is also located at the bottom, center of the cover.
NOTE: When the Side Cover(s) are removed, always inspect the filter for debris, blockage or damage.
If the Filter is blocked, clean it. If the Filter is damaged, replace it. If the Side Covers are
replaced, the Filters must be transferred to the replacement cover.
5. When replacing Side Covers for a LOGIQ E10 used in veterinary environment, the LOGIQ E10 Spare Parts
Handling for Veterinary Systems MUST BE re-installed before the LOGIQ E10 returned for use.
See: "Label Placement for the LOGIQ E10 used in a Veterinary Environment" on page 8-58.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
The Front Cover is attached to the LOGIQ E10 with tabs at the top, inside of the Cover and slots in the
Top Cover, and four self-clinching studs. Two studs are located above the probe connectors and two
are located at the bottom of the cover.
2. With both hands, firmly press in, evenly in the center, above the probe connections, to place the
top two self-clinching studs. Then, with both hands, firmly press in, evenly at the bottom to com-
plete installation.
3. Re-install Side Covers.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 System CFM and PWD Checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person /
No tools are required.
15 minutes
The Foot Rest Bumper is attached to the LOGIQ E10 with two self-clinching studs.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Brakes and Direction Lock Functional
4-2-6
Checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
2. Install replacement hook and re-install screws. Torque: 9.0 Nm (6.6 lbf-ft {79.2 lbf-in}).
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
The Rear Cover is attached to the LOGIQ E10 with four self-tapping screws, two on each side. These
mounting screws are screwed into plastic. Use care during installation, DO NOT overtighten.
NOTE: When installing the Rear Cover place the Mains Power Cable clamp in so it will clear the cover
installation. The cover should be flush with the Circuit breaker and ON/OFF panel and Rear
I/O panel when installed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
4-2-4 Power SHUT DOWN 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Make sure the Z-Mech manual release
lever is accessible and functions.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
The Rear Bumper is attached to the LOGIQ E10 with two self-clinching studs.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
The Rear Trim Cover is attached to the LOGIQ E10 with two self-clinching studs.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
The Top Cover is attached to the LOGIQ E10 with two Bin Pivot Brackets, at the rear. One Pivot Bracket
on each side. The Pivot Brackets are secured to the Cover Supports.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 System CFM and PWD Checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person /
No tools are required.
15 minutes
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN Required torque was applied.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
Table 8-72 Main Monitor Arm Assembly Covers removal and installation
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-73 Main Monitor Arm Assembly Covers replacement Functional Checks
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this Printer Filler
Storage replacement:
1.) Verify that all screws that removed earlier has been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 B-Mode Checks required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time Total
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
One person /
8-2-6 "PPE Required During Service" on page 8-6,
15 minutes
Flexible shaft bit driver extension (optional).
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks
5750007-1EN, Rev. 1. Equipment passed all
Operator I/O Movement - XY and Z- required checks and is ready for use.
Mechanism
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time Total
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Operator I/O Movement - XY and Z-
Mechanism
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time Total
One person /
No tools are required.
15 minutes
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
2. Under the XY / Frogleg mechanism, Right side frog leg, from underneath
identify the four separate covers. There are
two types of covers, one type has two
screws (1) and the other type has one
screw (2).
Remove the screws from the cover(s) to
replace.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-5 Top Console checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time Total
One person /
No tools are required.
15 minutes
Bulkhead Cover
2. Use thumbs to press upper lock tabs Upper lock tabs (placement on Bulkhead Cover)
toward the front of the Ultrasound System
to release the top of the Bulkhead Cover.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 System CFM and PWD Checks
4-2-4 Power SHUT DOWN
Section 8-7
Replacing Top Console Parts
8-7-1 Purpose of this section
This section describes how to replace the Top Console parts.
Manpower /
Tools / PPE
Time Total
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
40 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
3. Re-connect the HDMI (4), the USB (5) and Cable connectors
the Power (6) Cables from the Monitor.
See: Section 6-2 "Monitor adjustments" on page 6-2 for Main Monitor calibration instructions and/or
Monitor Arm and Monitor Friction Adjustment.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
NOTE: The Wide Screen Monitor may need a second reboot for the system to fully identify the
monitor’s USB device. To confirm, verify that the Monitor Controls for the Wide Screen Monitors
are available on the Touch Panel Utility pages.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
10-7-4 Grounding continuity allowable limits. Equipment passed all required
10-7-5 Chassis leakage current test checks and is ready for use.
Manpower /
Tools / PPE
Total Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
This procedure assumes that the Monitor and Bulkhead Cover have been removed.
NOTICE
DO NOT connect the Monitor power cable to the power (P3) connector on the Bulkhead when
the LOGIQ E10 is powered up. Damage to the Monitor Power Cable and/or the Bulkhead Board
can occur. See images below of damage that can occur.
48VDC pin in the P3 connector (center pin), shows damage.
The outside two pins are Ground. Cable left, Bulkhead Connector right.
See: Section 6-2 "Monitor adjustments" on page 6-2 for Monitor calibration instructions and/or Monitor
Arm and Monitor Friction Adjustment.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Move the Monitor Arm from side to side and ensure that it moves as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-5 Main Monitor Positions and Lock
4-2-5 Operator I/O Movement - Main Monitor LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
4-2-7 System CFM and PWD Checks allowable limits. Equipment passed all required
10-7-4 Grounding continuity checks and is ready for use.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
1. WARNING
WHEN REMOVING OR INSTALLING THE
MONITOR ARM, KEEP THE ARM
LOCKED, AS SHOWN TO ENSURE THE
MONITOR ARM ASSEMBLY IS IN THE
LOCKED POSITION UNTILL THE ARM
AND MONITOR ARE INSTALLED. THE
SPRINGS TO SUPPORT THE MONITOR
CAN CAUSE THE ARM TO SPRING
OPEN CAUSING SEVERE PERSONAL
INJURY AND PROPERTY DAMAGE.
NOTICE
DO NOT connect the Monitor power cable to the power (P2) or (P3) connector on the Bulkhead
when the LOGIQ E10 is powered up. Damage to the Monitor Power Cable and/or the Bulkhead
Board can occur. See: images below of damage that can occur.
48VDC pin in the P3 connector (center pin), shows damage.
The outside two pins are Ground. Cable left, Bulkhead Connector right.
5. Re-install Monitor and Arm Assembly and connect the Monitor Cables.
6. Re-install all Covers removed.
See: Section 6-2 "Monitor adjustments" on page 6-2 for Main Monitor calibration instructions.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the LOGIQ E10 to verify that it operates as intended.
Functional Checks
Perform the following functional tests to confirm the LOGIQ E10 is operational before returning the
LOGIQ E10 to the customer.
See:
Section Functional Checks Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
10-7-4 Grounding continuity allowable limits. Equipment passed all required
10-7-5 Chassis leakage current test checks and is ready for use.
Introduction
The Probe Holder Inserts are soft rubber inserts, used to protect the Probes from scratches,
when stored on the LOGIQ E10. You can place the Probe Holder Inserts in any of the desired
places on the edge of the Operator Panel. The following types of inserts are available:
• Probe Holder Insert STD,
• Probe Holder Insert 3D,
• Probe Holder Softinsert Doppler.
Manpower /
Tools / PPE
Time
One person /
No tools are required.
15 minutes
1. Install the flexible Probe Holder Insert into Probe Holder placement
the Lower UI Frame. Be sure the hooks (1)
and tabs (2) fit properly and the Probe
Holder fits snugly.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 Basic Measurements required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person /
No tools are required.
15 minutes
Preparations
OP Knobs removal
Step
OP Knobs installation
Esc
~ ?
F1 F2 F3 F4 F5 F6 F7
1
F8 F9 F10
CHI
Set Home 2 Word Prt Sc
` Help Arrow Eject Spooler Macro Macro Home Copy Move Delete
! @ # $ % ^ & * ( ) _ +
B
1 2 3 4 5 6 7 8 9 0 - = Abc
Auto
Q W E R T Y U I O P { } |
[ ] \
A S D F G H J K L : L R
"
; ' Clear P4
Z X C V B N M < > ?
, . / Pg Up
P1 P2 P3
Ctrl Fn Alt Alt Gr Ins Del Freeze
Home End Pg Dn
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 Basic Measurements required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
One person /
8-2-6 "PPE Required During Service" on page 8-6.
15 minutes
A thin, flexible plastic card, such as a credit card or an EHS card.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE WRIST STRAP SHOULD BE CONNECTED TO ONE OF
THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
1. Make sure the Upper OP is in its uppermost position, with the Main Monitor out of the way.
NOTICE
Failure to remove the six Upper OP Knobs first, could cause damage to the knob shafts.
B J25
5V D
P4 J21
USB USB
J17 J2
HU B
J23 J16 J1
A - ECB
J15 3.3V
12V 5V 3.3V
J6 +48V uP
REG
5V TOUCH E
J22 12w"
C
22
TEST CONNECTOR
J6 J5 J4
U2 C115
J7 1
C105
SP16 2 6
13 13 R15
A6P5 25
25 FP22 P1 3
R11
FP24 4 8
FL3
FL1
11
FL2
11
FL4
C113
R20
5 9
R23
U3
FL8
R25
FL7
FL6
FL5
J12 J8
D C EB A E D C A B E D C B A
25 25 25 25 25 25
1 1 B C D
A E FZ
D C EB A 19 19
11 11
C2
C1
14 14 A4P4
A3P4 A2P4
A6P4
15 15 15 15 15 15
11 11
C3
A5P4
C9
C8
1 E D C B A 11
1 11 11 11 11 11
C15
C7
C6
C5
J12
C10
D C EB A
C4
C11
C12
C13
C14
19 19
C19
C17
C16
C18
C21
C20
C22
C24
C26
C27
C23
C25
1 1
C28
Z A B C D E F
C29
C30
C31
C40
1 1
C41
1 D C EB A 1
C36
C42
C37
C39
C38
C44
C43
FP21
C33
C32
C34
C35
11 11
19
A6P3
C47
C50
C45
C46
C48
C49
C51
C52
C53
C54
C55
C56
C59
C61 C62
1 1 9 9 C60
C63
E D C B A
C68
C67
DG BG
22 22 1 DC BA
1
C66
C147
C65
C64
C148
C149
DG FG HG C150
C151 A2P3
E F G H 10 C152 A3P3
10 SP15 C153
C138
C179 1
C180 C155 1 1 1
C156 A4P3
C181 C157 22 22 E D C B A
C182 C158
C183 C159
C184 C160
C185 C161
C186 C162
A5P3 C187 FP8
11 11 C188
C189
A6P2 C190 10 10 FP10
C191
C192 FP12
C193
C194 R30 A3P2
DG
BG
FP25 SP14 1 DC BA
1 C163 R31
C195 1
Upper OP cables connected, see Figure 8-2 "Upper OP/Touch Panel assembly
cable placement" on page 8-119.
4. The Knobs are keyed, notice that the joystick has a flat side to match the Knobs.
Re-install the six OP Knobs removed.
Before continuing, test the Encoders and Op Panel controls to ensure they operate properly. If
the Encoder seals are too close to the Encoder, the seals can cause the Encoder operation to
bind. If the seals are to high, the can limit the joystick operation. If the joystick operation is limited,
remove the Knob and gently press the seal(s) in.
Re-install the six OP Knobs and all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 Basic Measurements required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Lower OP removal
1. Keep XY-Mech in the released state and the console extended forward to gain access to the
Lower OP screws. The screws are accessible from the bottom of the OP Lower Frame.
Lower OP installation
2. Re-install the screws to secure the Lower OP to the Lower OP Frame five longer (rear) and the
three shorter (front).
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE WRIST STRAP SHOULD BE CONNECTED TO ONE OF
THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
2. Disconnect the cables from the Bulkhead Board at the front of the console:
• Power P4 - slide locking latch to release,
• Speakers P5 (right) and P6 (left), filters may not be present,
• Video (HDMI) P10 - remove 5 mm screw (from back of Bulkhead Plate) to release cable
retainer,
• USB P8
PWB
5482677 REV 3
P6 P3 L9 P1 P5
L2
P8
L4 L3 L6
2
L5 L7
BULKHEAD BOARD
PWA
5482676 VER B
P4
ASSEMBLED
IN
e1
3. NOTICE
NOTICE
DO NOT connect the Monitor power cable to the power P2 or P3 connector on the
Bulkhead when the LOGIQ E10 is powered up. Damage to the Monitor Power Cable and/
or the Bulkhead Board can occur. See images below of damage that can occur.
48VDC pin in the P3 connector (center pin), shows damage.
The outside two pins are Ground. Cable left, Bulkhead Connector right.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
45 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
1. Position the Bulkhead Plate so it aligns with the holes for the fixing screws.
re-install the two upper shorter screws and the remaining three. Torque: 10 Nm (7.4 lbf-ft).
3. Re-install:
• Bulkhead Board,
• Speaker cables,
• Lower OP,
• Upper OP,
• OP Knobs,
• all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 Color Mode Checks required checks and is ready for use.
4-2-7 PW/CW Doppler Mode Checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
2. Re-install the five screws to secure the Cover. Torque: 3.0 +/- Nm.
3. Re-install:
• Monitor Arm Adapter,
• Monitor Arm and Monitor,
• Upper OP,
• all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 System CFM and PWD Checks
LOGIQ E10 Basic Service Manual, Direction
Monitor Arm and/or Monitor range of
Section 6-3 5750007-1EN, Rev. 1. Equipment passed all
motion Adjustment procedure
required checks and is ready for use.
Audio Check
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE WRIST STRAP SHOULD BE CONNECTED TO ONE OF
THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
1. Re-install the screws to secure the Trackball the Trackball Mounting Bracket.
Remember the orientation of the Cable in the Bracket.
2. Re-install the screws to secure the Trackball Mounting Bracket to the Bezel.
3. Connect the signal cable connector to the Lower Switch and Encoder Board.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Verify that the Trackball operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
PW
TVI
X Z
CW PDI
M CF
3D/4D Elasto V Nav LOGIQ Contrast
View
B
Flow
Esc
~ ?
F1 F2 F3 F4 F5 F6 F7
1
F8 F9 F10
CHI
Set Home 2 Word Prt Sc
` Help Arrow Eject Spooler Macro Macro Home Copy Move Delete
! @ # $ % ^ & * ( ) _ +
B
1 2 3 4 5 6 7 8 9 0 - = Abc
Auto
Q W E R T Y U I O P { } |
[ ] \
A S D F G H J K L : L R
"
; ' Clear P4
Z X C V B N M < > ?
, . / Pg Up
P1 P2 P3
Ctrl Fn Alt Alt Gr Ins Del Freeze
Home End Pg Dn
5. Remove the 30 screws securing the boards to the Bezel, using a T10 Torx driver. KEEP the old
Bezel facing down.
There are 22 screws on Trackball board and 8 on the other board. (Complete board shown for
reference.)
6. Remove the boards and transfer the Op Panel controls the new Bezel, one at a time.
3. Re-install the Lower OP, Upper OP/Touch Panel Assembly (verify Encoder operation) and all
covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 Basic Measurements
Joysticks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Upper Op Encoders/Joysticks
If replacing the Upper Op Encoders, ONLY remove the Upper Op and Lower Op, then replace the
Encoder(s). Follow Step in 8-7-15 "Lower Bezel replacement" on page 8-144 and Perform Functional
Checks.
Lower Op Encoders/Joysticks
The four Mode Select (Concentric Shaft) Encoders adjust the M-mode, PW-mode, CF-Mode, and B-
Mode scans.
NOTE: This procedure you will essentially perform the same procedure as replacing the Lower Op
Bezel, except, DO NOT remove the:
1. Fold back the elastomer to expose the nuts that secure the Encoders/Joysticks to the Lower
Switch and Encoder Board.
Mode Encoders circled - Use a 7/16 inch, deep socket wrench or 7/16 inch open-ended wrench
to loosen the nut(s).
Encoders/Joysticks (A) - use a 24 mm socket or an adjustable wrench and remove the nut(s).
Remove Encoder(s)/Joystick(s) as necessary.
Continued
2. Re-install the elastomer. MAKE SURE the elastomer is completely seated on the board.
4. Re-assemble Lower Op, re-install Lower Op, Upper Op and all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 Probe/Connectors Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
2. Install the Upper Frame with the screws and washers removed.
3. Support the two Frames and remove the temporary screws. Continue to support the Frames to
install the Bulkhead Plate.
Re-install the Bulkhead Bracket Housing (behind the Bulkhead Plate) and install the Plate with
the two 20 mm at the top and the two 25 mm at the bottom. Torque: 4.3 Nm (3.2 lbf-ft).
4. Re-install the:
• Main Cable to the Lower OP,
• Grounding Strap Clamp. Torque: 4.3 Nm (3.2 lbf-ft).
• P-clamp screw and lock washer, using a #1 Phillips screwdriver.
• Bulkhead Board,
• Lower OP and Upper OP,
• Monitor Adapter, Monitor Arm and Monitor,
• Re-install all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Basic Measurements
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
OP I/O Movement - XY and Z-Mechanism
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
5. Remove the two screws temporarily installed to hold the Lower UI Frame, while supporting the
Lower UI Frame and remove the Frame.
1. Position the Lower UI Frame and temporally install two screws to hold the Lower UI Frame when
attaching the Main Cable Assembly.
Feed the cables into the Lower UI Frame and re-install the two screws to secure the Main Cable
Assembly to the Lower Frame.
4. Install the:
• Main Cable to the Lower OP,
• Bulkhead Board,
• Lower OP and Upper OP,
• Monitor Adapter, Monitor Arm and Monitor,
• re-install all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Basic Measurements
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
OP I/O Movement - XY and Z-Mechanism
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTICE
Failure to remove the six OP Knobs first, could cause damage to the knob shafts.
4. Make sure the XYZ Buttons cable runs along the Lower OP cable channel, as shown so the cable
is not pinched when the Palm Rest is replaced.
Install the replacement cover and screws.
These screws are screwed into plastic. Use care during installation, DO NOT overtighten.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Operate the XY and Z Controls to verify operation as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 Basic Measurements required checks and is ready for use.
Operator I/O Movement - XY and
Z-Mechanism
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTICE
Failure to remove the six OP Knobs first, could cause damage to the knob shafts.
2. Remove the:
• Upper OP,
• Lower OP,
• Palm Rest.
3. Release the:
• lock mechanisms (1) from the top Left Handle Cover,
• lock mechanisms (2) from the bottom Left Handle Cover.
1. Position:
• bottom Left Handle Cover and secure with locking mechanisms.
• top Left Handle Cover and secure with locking mechanisms.
2. Re-install:
• Palm Rest.
• Lower OP.
• Upper OP,
• all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO
ONE OF THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY
FOR THAT PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Energy Control and Power Lockout for LOGIQ E10:
When servicing parts of the Ultrasound System where there is exposure to voltage greater than 30
Volts:
1. TURN OFF THE SCANNER.
TAG
2. UNPLUG THE SYSTEM.
&
LOCKOUT
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/OFF
Signed Date
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTICE
Failure to remove the six OP Knobs first, could cause damage to the knob shafts.
2. Remove the:
• Upper OP.
• Lower OP.
• Palm Rest.
3. Release:
• the lock mechanisms (1) from the top Right Handle Cover.
• the lock mechanisms (2) from the bottom Right Handle Cover.
1. Position:
• bottom Right Handle Cover and secure with locking mechanisms.
• top Right Handle Cover and secure with locking mechanisms.
2. Re-install:
• Palm Rest.
• Lower OP.
• Upper OP.
• all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTE: If this Options Holder will be supporting a Gel Warmer, place the Gel Warmer in the Options
Holder first, then install the Gel Warmer to the Options Holder.
Make sure the Gel Warmer cable fits in the cable channels and is not pinched.
2. Re-install the three screws to secure the Options Holder. Torque: 1.5 Nm (1.1 lbf-ft).
3. Re-install the Storage Tray or Transvaginal Probe Holder to the Options Holder, if applicable.
Re-install all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) If a Gel Warmer is installed in the Options Holder, see: additional Verification and Functional
Checks for Gel Warmer (see: 8-7-23 "Gel Warmer replacement" on page 8-173).
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-203 Options Holder / Left or Right Support replacement Functional Checks
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
1. Disconnect the Gel Warmer cable from (P1) on the back of the bulkhead.
The Gel Warmer cable is routed through the cable channel (1) and the Lower Frame support (2).
Pull the Gel Warmer cable through the Lower Frame cable support to free the cable.
2. Remove the Options Holder (see: 8-7-22 "Options Holder / Left or Right Support replacement"
on page 8-169).
Options Holder removed, with Gel Warmer in place
1. Place the Gel Warmer in the Options Holder, routing the cable through the cutout (1) between
the Options Holder and Gel Warmer and through the cable channel (2).
BE CAREFUL not to pinch the wires.
2. Install the two screws at the Gel Warmer base, securing the Gel Warmer to the Options Holder.
Torque: 1.5 Nm (1.1 lbf-ft).
4. Route Gel Warmer cable through the Lower Frame cable support.
Temperature is preset to its highest level: 38 degrees C (+/-2 degrees C). It is recommended to retain
default setting.
User can turn dial to adjust the temperature to its lowest level: 32 degrees C (+/-2 degrees C).
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Place a closed bottle of gel, applicator end first, in the gel warmer. Verify warmer is operational.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-7 System CFM and PWD Checks required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Follow this procedure for replacement, or replacing a tray with a Probe Holder or a Probe Holder with a
tray.
1. Seat the tray or Probe Holder tab into the options holder base and snap into place.
8-7-25 Replacing the Gel Warmer with a Storage Tray or TV/TR Probe Holder
Follow these instructions when removing the Gel Warmer and replacing it with a Storage Tray or a TV/
TR Tray.
Before starting this procedure, see: 8-7-23 "Gel Warmer replacement" on page 8-173.
Push the left button, and then reposition the XY (horizontal) location of the Upper Console. Three
seconds after pushing the left button, the four motors in the XY frog leg lock to stabilize the console.
Toggle the right button to reposition the Z (vertical) location of the Upper Console. Toggle up to raise,
or down to lower, the console.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTACLES INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTICE
Failure to remove the six OP Knobs first, could cause damage to the knob shafts.
3. Gently pop the XYZ Buttons out of the Palm Rest section of the Lower OP Frame handle.
2. Re-install the Palm Rest, Lower OP, Upper OP and all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Operate the XY and Z Controls to verify operation as intended.
5.) Select comments and type different A/N keys to confirm keyboard functionality.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 Basic Measurements 5750007-1EN, Rev. 1. Equipment passed all
Operator I/O Movement - XY and required checks and is ready for use.
Z-Mechanism
4-2-4 Power SHUT DOWN
Brakes and Direction Lock Functional
4-2-6
Checks
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
One person / • LOCTITE 242 (LOCTITE P/N 24221). See:
15 minutes "Material Safety Data Sheet link for LOCTITE 242:" on page 8-183 to access data
sheet.
LOCTITE 242 can be obtained from Grainger.
• Grainger item number 5A228.
2. Position replacement Probe Cable Hooks and fix it in place by using the M5x20 screw and M6
washer removed. Apply LOCTITE 242.
Torque: 50 Ncm (0.37 lbf-ft or 4.4 lbf-in) (fingertight).
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 Color Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 PW/CW Doppler Mode Checks
4-2-5 Top Console checks
4-2-4 Power SHUT DOWN
Section 8-8
Replacing XYZ Parts
8-8-1 Purpose of this section
This section describes how to replace the replaceable parts in the Main Console.
Manpower /
Tools / PPE
Time
One Person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
5. After the system is powered down, unplug the power cord from the wall outlet.
6. Identify which lock motor needs to be replaced and
remove the corresponding cover, using a T-10
driver:
One TORX-head screw (6) to remove the forward
cover and two TORX-head screws (6a) to remove
the back cover.
If it is difficult to lock the XY-Mechanism in parked position, the locking mechanism need adjustments.
For Operator Panel adjustments, see: Section 6-4 "Operator Panel movement" on page 6-13.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-5 Top Console checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One Person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
2 Hours 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
XY-Mechanism removal
WARNING
Operator console can drop unexpectedly when the mechanism is released.
2. Lock the Caster Brakes.
3. Remove the rear Brake Motor Covers, using a T-10 driver.
Carefully disconnect the rear Brake Motors and the ground cables, using a 4 mm hex wrench.
XY-Mechanism installation
If it is difficult to lock the XY-Mechanism in parked position, the locking mechanism need adjustments.
For Operator Panel adjustments, see: Section 6-4 "Operator Panel movement" on page 6-13.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
Gel Warmer Check
Operator Functional Check LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
Run ECB I/O Diagnostics to Check
required checks and is ready for use.
Audio Speakers
4-2-5 Top Console checks
4-2-4 Power SHUT DOWN
Brakes and Direction Lock Functional
4-2-6
Checks
Manpower /
Tools / PPE
Time
Two Persons / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
4-6 Hours 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Z-Mechanism removal
WARNING
Operator console can drop unexpectedly when the mechanism is released.
2. Remove the four screws securing the
Main Cable Cover and the Column Cover,
using a #2 Phillips screw driver.
Support the Column Cover when the
screws are removed so the Cover does
not fall.
The right side, rear screw can be
accessed through the hole in the Drive
Gear housing.
3. Take note on how the Cables are routed, they MUST BE returned to the same location.
Remove the screws securing the Ground P-clamp, (A) Cable Clamps (B) and the
screw (C) securing the XYZ Motor Controller, using a #2 Phillips screw driver.
4. Disconnect all of the Main Cable Harness connections from the ECB - ECB J5, ECB J6,
ECB J15 and MOTOR CTRL (Button Interface), and power from BACKPLANE J12.
5. Disconnect the remaining cabling to the XYZ Motor Controller, USB Interface, Lock Motor
Z motor, Brakes and remove the grounds cables.
If the Brake Cable are not identified (left / right), mark them to ensure they are connected
correctly later.
6. Remove the Main Cable Harness from the
top of the Z-Mech, using a 5 mm hex
wrench.
Note how the XY-Mech Brake cables are
routed.
Feed all the cables through the Z-Mech.
7. Remove the screws (D) securing the right and left Cable Harness Brackets, using a 4 mm hex
wrench.
11. With both hands and possibly some help, lift the Z-Mech straight up off the Frame.
Z-Mechanism installation
3. The Cables MUST BE returned to the same location and channeled under the Cable Cover.
Re-install the screws to secure the Ground P-clamp, (A) Cable Clamps (B) and the
screw (C) securing the XYZ Motor Controller, DO NOT tighten the Clamps until the Main Cable
Cover and the Column Cover are installed, to ensure there is enough slack for the Covers to be
installed. Install Covers and tighten Clamps.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
Gel Warmer Check LOGIQ E10 Basic Service Manual, Direction
4-2-5 Top Console checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Operator Functional Check
Run ECB I/O Diagnostics to Check
Audio Speakers
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
Two People / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30-45 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTE: There is need to disconnect the Probes, they are probably safest if kept on the LOGIQ E10. The External I/O
Cabling does not need to be removed unless it is in the way.
1. While the LOGIQ E10 is powered up, move the Operator Interface up and towards the user.
2. Power down the system.
3. Disconnect the mains power cable from the wall outlet.
4. Remove the Side Covers, the Battery Cover, Top Cover and the Rear Cover.
NOTE: Two people are recommended for this procedure because it is very difficult for one person to
position the motor and wiggle the Top Console assembly at the same time; the second person
does not need to be a trained FE or do any lifting.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes if removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-5 Top Console checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
Two People / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Energy Control and Power Lockout for LOGIQ E10:
When servicing parts of the Ultrasound System where there is exposure to voltage greater than 30
Volts:
1. TURN OFF THE SCANNER.
TAG
2. UNPLUG THE SYSTEM.
&
LOCKOUT
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/OFF
Signed Date
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTE: There is a USB cable connected to the XYZ Controller, from J11 of the ECB. The cable must
also be removed from the USB port labeled USB Interface of the controller.
2. Pull the top of the XYZ Control Assembly forwards, away and up from the bracket.
5438170 R1
Prog Lock
Motor
Z motor
48 V
3. Install the five countersunk screws at the top, rear, if they have been removed as part of
Z-Mechanism replacement. Torque: 9.8 Nm (7.2 lbf-ft).
Install the two screws to secure the cable on the top. (M4x8, sunken. Torque: 4.9 Nm {3.6 lbf-
ft}).
4. Re-install all Covers removed.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-5 Top Console checks required checks and is ready for use.
4-2-4 Power SHUT DOWN
Section 8-9
Main Console parts replacement
8-9-1 Purpose of this section
This section describes how to replace the replaceable parts in the Main Console.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Subwoofer removal
CAUTION
During this procedure, wear cut resistant gloves to avoid scraping your hands when
reaching into the Main Console or blind spaces.
A A A
Rear I/O USB Cable
NOTE: Be aware, the Subwoofer magnet
may draw the Option Box to it, as it is
removed.
Lift the Option Box up slightly and turn it
away from the Main Power Supply to
access the USB Cable from the bottom of
the Rear I/O. Remove the USB Cable and
continue to remove the Option Box away
from the Card Cage.
Subwoofer installation
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Turn the volume control on at the OP.
5.) Select PW mode and press on the selected Probe to ensure the Subwoofer produces sound.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
NOTE: NOT ALL functional checks will have to be repeated, if the functional check was performed
when the previous Option was re-installed. If ONLY a particular Option was removed, perform
the functional checks for the Subwoofer replacement and that particular Option.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 Probe/Connectors Checks
4-2-7 System CFM and PWD Checks
Power SHUT DOWN LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
4-2-4 DO NOT POWER SHUT DOWN until ALL
required checks and is ready for use.
functional checks have been performed.
Verify Options re-installed are functioning.
In addition: Perform the following functional checks if
the Option was or Options were removed during
Subwoofer replacement.
4D
4-2-7 Probe/Connectors Checks
Perform a 4D sweep
CW LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
PW/CW Doppler Mode Checks measured at (record the value) and meets
4-2-7
use pencil probes allowable limits. Equipment passed all required
checks and is ready for use.
4-2-7 Probe/Connectors Checks
Chassis Leakage Current Test
10-8-4 Perform a CW scan with P2D, P6D,
M5S-D probes (plug in and scan)
Continued
See:
Section Functional Check Debrief Script
ECG
4-2-7 Probe/Connectors Checks
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source) leakage–
10-7-8
lead to lead
Isolated patient lead (sink) leakage-
10-7-9
isolation test
ECG Checks
Active ECG:
Utility -> Applications ->
Settings -> ECG -> Show ECG Tools.
Connect cables. Press Display ECG Tab LOGIQ E10 Basic Service Manual, Direction
4-2-7
on TP. Verify ECG signal displayed on 5750007-1EN, Rev. 1. Leakage Current
monitor. measured at (record the value) and meets
Plug in the ECG cables with nothing allowable limits. Equipment passed all required
attached so they pick up noise. checks and is ready for use.
Diagnostics
Diagnostics
Shear Wave Elastography (Shear Elasto)
4-2-7
Functional Checks
Power Assistant
4-2-7 Probe/Connectors Checks
Verify Power Assistant Option functions
Shear Wave
4-2-7 Probe/Connectors Checks
Shear Wave Elastography (Shear Elasto)
4-2-7
Functional Checks
Perform a 4D sweep, if 4D MC was
removed.
Section 8-10
Casters and Brakes replacement
8-10-1 Purpose of this section
This section describes how to replace the replaceable parts for the Casters and Brakes.
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
Two people / This procedure requires:
30 minutes • Wheel Change Kit, P/N FC200829.
• an extended hex wrench (UNBRAKO / 10 mm hex wrench or the torque wrench
used to torque) due to the high torque: 130 Nm {95.9 lbf-ft}.
WARNING Prior to elevating scanner, verify that the floating Operator Panel is locked in its lowest,
parking position.
Use extreme caution as long as LOGIQ E10 is un-stable, not resting on all four Casters.
CAUTION
Use extreme caution as when the LOGIQ E10 is un-stable, NOT RESTING on all four Casters.
D
C
C
L
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-5 Top Console checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Required torque was applied.
Brakes and Direction Lock Functional
4-2-6
Checks
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
Two people /
8-2-6 "PPE Required During Service" on page 8-6.
30 minutes
This procedure requires the Wheel Change Kit, P/N FC200829.
WARNING Prior to elevating scanner, verify that the floating Operator Panel is locked in its lowest,
parking position.
Use extreme caution as long as LOGIQ E10 is un-stable, not resting on all four Casters.
CAUTION
Use extreme caution as when the LOGIQ E10 is un-stable, NOT RESTING on all four Casters.
Use extreme caution as when the LOGIQ E10 is un-stable, NOT RESTING on all four Casters.
C
L KEEP the Caster as centered as possible on the Bevel Edged Board when resting.
NOTE: Prior to elevating LOGIQ E10, verify the Operator Panel is locked in its lowest, parked
position.
The Caster not being replaced; pull the Ultrasound System backwards on that Caster, onto the
Bevel Edged Board (A). This will lift the other Front Caster off the floor.
The Front Caster on the Bevel Edged Board will be facing forward, in the direction as shown
(away {C} from the Base Casting (D).
Wedge the Wedge (B) from the side of the Base Casting (D), on the side in which the Caster is
being replaced, to keep the LOGIQ E10 solidly supported.
D
C
L
4. The Ultrasound System is now nearly balanced between one Front and both Rear Casters.
5. MAKE SURE the LOGIQ E10 is at rest, on both Rear Casters.
Lift the Front Caster and put the Wooden Wedge (B), upright as shown, under the Base Casting
to stabilize the LOGIQ E10 with one Front Caster off the floor. This Front Caster can now be
removed.
8 mm
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-5 Top Console checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Required torque was applied.
Brakes and Direction Lock Functional
4-2-6
Checks
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
5 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING
Make sure to wear safety glasses or safety goggles for personal protection, as the hubcap
OR may splinter and/or shatter during removal.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-4 Power SHUT DOWN
5750007-1EN, Rev. 1. Equipment passed all
Brakes and Direction Lock Functional required checks and is ready for use.
4-2-6
Checks
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
60 minutes 8-2-6 "PPE Required During Service" on page 8-6.
2
1 3
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-5 Top Console checks LOGIQ E10 Basic Service Manual, Direction
4-2-4 Power SHUT DOWN 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Brakes and Direction Lock Functional
4-2-6
Checks
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
(Tools):
• Torx driver - T30. Optional Torx L-wrench or Torx socket.
• Open end Wrench, 10 mm.
• Wrench, Box, 10 mm. Optional deep socket.
• Adjustable wrench, 152 mm (6 inches) long (if a longer adjustable
One person / wrench is used, it will contact the floor when trying to turn it.
30 minutes • Two M5 X 0.8 X 16 mm screws (P/N 46-312358P52) from Box 1 of
the LOGIQ E10 Hardware Kit (P/N 5310019-2).
• LOCTITE 290 (LOCTITE P/N 29031). See:
"Material Safety Data Sheet link for LOCTITE 290:" on page 8-244 to access data
sheet.
LOCTITE 290 can be obtained from Grainger.
• Grainger item number 5E217.
NOTE: An adjustable wrench is deeper than an open end wrench and will secure
the Cam Plate better in this procedure.
2
1 3
6
4
5
14mm
(0.55 inches) 4
@ 100%
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-269 Pedal Mechanism replacement Functional Tests
See:
Section Functional Tests Debrief Script
4-2-3 Power ON/Boot Up
4-2-5 Top Console checks LOGIQ E10 Basic Service Manual, Direction
4-2-4 Power SHUT DOWN 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Brakes and Direction Lock Functional
4-2-6
Checks
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
A small ratchet with 1/4 inch socket and a #2 Phillips, six inch extended driver and
a mirror will be very helpful during this procedure.
One person /
5 minutes
per cover
1a
2a
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-5 Top Console checks LOGIQ E10 Basic Service Manual, Direction
4-2-4 Power SHUT DOWN 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
Brakes and Direction Lock Functional
4-2-6
Checks
Section 8-11
Front End Acquisition / Card Cage parts replacement
8-11-1 Purpose of this section
This section describes how to replace the parts in the Front End / Card Cage.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO
ONE OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit,
to indicate boards are NOT powered.
8 - 254 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTE: If the V Nav Option is present, the Hi-Pass connection from the GRLY Board to the DriveBay2+
must be disconnected.
If the WLAN Option is present, note the orientation of the antennas, they MUST BE returned to the same
position. Reposition the antennas when removing the Cover.
2. If the V Nav Option is present, remove the Hi-Pass connection - V Nav Option
Hi-Pass EMI Cover by removing the screw
and the screw holding the cable P-clamp,
using a #2 Phillips screwdriver.
Disconnect the Hi-Pass cable from GRLY
Board.
8 - 256 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
1. Re-install the Card Cage Cover and secure it with the eight screws removed. Install all the
screws before tightening.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-279 Front End Card Cage Cover replacement Functional Checks
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
4-2-7 Basic Measurements allowable limits. Equipment passed all required
10-7-4 Grounding continuity checks and is ready for use.
8 - 258 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
NOTE: Read through the removal and installation steps completely before performing. Perform the
steps carefully to avoid damage to the Front Plane connectors.
Front Plane Boards
8 - 260 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
1. Holding the upper and lower edges of the board with both hands, carefully install the Front Plane
Board. Ensure that you do not bend any of the connector pins during the installation.
Apply even pressure across the board and to apply gentle, even pressure at the four corners of
the Front Plane Board to make full contact with the other boards.
2. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
8 - 262 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Energy Control and Power Lockout for LOGIQ E10:
When servicing parts of the Ultrasound System where there is exposure to voltage greater than 30
Volts:
1. TURN OFF THE SCANNER.
TAG
2. UNPLUG THE SYSTEM.
&
LOCKOUT
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/OFF
Signed Date
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
1. Using the board ejector levers, remove the Relay Board. Place it on an ESD safe place.
1. Carefully align the Relay Board with the rails and push it in until starts to seat in the Backplane
connectors and close the board ejector levers.
3. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY and re-install the cover, if present.
8 - 264 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Energy Control and Power Lockout for LOGIQ E10:
When servicing parts of the Ultrasound System where there is exposure to voltage greater than 30
Volts:
1. TURN OFF THE SCANNER.
TAG
2. UNPLUG THE SYSTEM.
&
LOCKOUT
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/OFF
Signed Date
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
8 - 266 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
1. Using the board ejector levers, remove the ETX Board(s). Place on an ESD safe place.
1. Carefully align the ETX Board(s) with the rails and push in until it starts to seat in the Backplane
connectors and close the board ejector levers.
3. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
8 - 268 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
CRX removal
1. Using the board ejector levers, remove the CRX Board. Place on an ESD safe place.
8 - 270 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CRX installation
1. Carefully align the CRX Board with the rails and push in until it starts to seat in the Backplane
connectors and close the board ejector levers.
3. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
NOTE: If only the ECB is to be replaced, transfer the GPU and the WLAN Option, if present to the
replacement ECB.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6. /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY
ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
ECB removal
8 - 274 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
ECB installation
1. Carefully align the ECB with the rails and push in until it starts to seat in the Backplane
connectors and close the board ejector levers.
2. Re-install the two screws to secure the ECB to the Card Cage.
3. Re-connect all cabling to the ECB. See: "ECB cable connections" on page 8-276.
4. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
8 - 276 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
A small, close quarters ratchet set will be very useful for this procedure.
If a small, close quarters ratchet set is not available, see: Figure 8-9 "Small, close
One person / quarters ratchet set Alternate" on page 8-279.
30 minutes Close Quarters Ratchet:
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
8 - 278 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
B
GPU removal
NOTE: If a small, close quarters ratchet set is not available, plug (A) can be removed to access screws
with a T-10 Torx driver. Re-install plug after replacement. Plug (A) removed and access (B)
shown.
B
8 - 280 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
GPU installation
8 - 282 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
9. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
10. Re-install all Covers removed.
8 - 284 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional tests to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
One person / 8-2-6 "PPE Required During Service" on page 8-6.
30 minutes A magnifying glass to connect the WLAN antenna cables to the WLAN Card will be
very useful for this procedure.
8 - 286 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
3. Place the lockwashers onto the threaded shaft of the Cables and install the new Cables. Torque
the nuts on the face of the ECB to 0.7 ±10% Nm.
Install the new Clamps, as shown. Torque to 0.15 ±10% Nm. Route the Cables as shown.
Secure the replacement WLAN Card to the ECB. Torque to 0.15 ±10% Nm.
Connect the Cables to the WLAN Card.
WLAN Set-up
For configuring the WLAN, refer to the latest revision of the LOGIQ E10 Basic User Manual, Chapter 10.
NOTE: This step is mandatory in order to insure proper functionality of services and communications
security settings.
Run the batch file “EnableWLAN.bat”. A screen will flash “updating network configuration for
4.
WLAN.“
8 - 288 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verifying Connectivity
Testing WLAN
8 - 290 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that the screw that you removed earlier has been installed.
2.) If finished, connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Check
Perform the following functional tests to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all required tests and is ready for use.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Energy Control and Power Lockout for LOGIQ E10:
When servicing parts of the Ultrasound System where there is exposure to voltage greater than 30
Volts:
1. TURN OFF THE SCANNER.
TAG
2. UNPLUG THE SYSTEM.
&
LOCKOUT
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/OFF
Signed Date
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
8 - 292 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
3. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Leakage Current Limits
4-2-4 Power SHUT DOWN
8 - 294 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
EPM removal
8 - 296 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
EPM installation
1. Carefully align the EPM with the rails and push in until it starts to seat in the Backplane
connectors and close the board ejector levers.
3. Install the Card Rack Cover and fasten it with the fixing screws.
Connect the Hi-Pass cable to the GRLY Board and re-install the cover, if present.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
4-2-7 Basic Measurements measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-4 Grounding continuity
checks and is ready for use.
10-7-5 Chassis leakage current test
4-2-4 Power SHUT DOWN
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts, conductors
and components. Always use non-conductive handles designed for the removal and replacement of ESD
sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before
making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR PACK ARE
OFF BEFORE REMOVING THE CAPACITOR PACK.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to indicate
boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD
PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE OF THE
TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11
J11
LOCKOUT
Signed Date (black and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
J10
CAUTION To avoid injury, handle the Backplane by the edges. There are two guides to guide the ECB onto the
Backplane, which protrude and are VERY SHARP.
8 - 298 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Backplane removal
NOTE: The Option covers are secured to the Option Bracket with M4 threaded studs, as part of the
Option Bracket and M7 flange nuts. All Options may not be present.
All the Option covers, can be removed at once, if desired. The Shear Wave Capacitor Pack is
the only Option in the Option Box that does not have a removable cover.
It is not necessary to remove the ECG (Patient I/O Module) or the Rear I/O, but the USB Cable
at the bottom of the Rear I/O, and all other cabling will have to be removed.
CAUTION
During this procedure, wear cut resistant gloves to avoid scraping your hands when
reaching into the Main Console or blind spaces.
8 - 300 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 302 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 304 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
A A A
Rear I/O USB Cable
8 - 306 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 308 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Backplane installation
CAUTION To avoid injury, handle the Backplane by the edges. There are two guides to guide the ECB onto the
Backplane, which protrude and are VERY SHARP.
1. Grasping the outer edges of the Backplane, carefully place the Backplane toward the Card Cage,
guiding the connectors to the Main Power Supply into the Card Cage.
Install the 21 Phillips screws to secure and ground the Backplane to the Card Cage. Tighten
screws alternately.
3. Re-install the Option Bracket. BE SURE to support the Bracket as it is being installed, it has
mass.
Re-install the five M6 screws that secure the Option Bracket to the Base Casting, loosely.
4. Re-install the four hex screwlocks (on the right) and two locking posts (on the left) to the
Connector Panel Torque to 0.6 +/- 0.1 Nm (0.4 +/- 0.07 lbf-ft).
Starting at the top, re-install the 7 Phillips screws that secure the Option Bracket to the Card
Cage. Tighten alternately.
Torque the M6 screws to 3.7 +/- 0.6 Nm (2.7 +/- 0.44 lbf-ft), ONLY the M6 screws are torqued to
this value.
Re-connect Subwoofer to J9, on the Backplane.
5. Position the Option Box up slightly and turn it toward the Main Power Supply to access the
Network Cable at the bottom of the Rear I/O. Re-connect the Network Cable and continue to
position the Option Box onto the studs that secure the Option Box to the Card Cage. Three are
underneath the Option Box, in slots.
Re-install the flange nuts. Tighten all the nuts, alternately, using a 7 mm driver.
6. Place the Subwoofer cable and grounding clamp into the Option Box.
Re-install the screw to secure the clamp to the Option Box. MAKE SURE the clamp is grounding
the cable.
7. Re-install the Battery Bracket.
Install the flange nuts and secure the Battery Bracket to the Option Bracket.
12. Re-install the Battery Bracket, re-connect the Batteries and install the Battery Cover.
1.) Reboot the system into Windows desktop and double-click the VPD Editor icon on the desktop. You
need Service SSA Key to access Windows.
2.) Select the Backplane in the device grid (BP).
3.) Locate the System serial field and select.
4.) Click the write dirty fields button and click in the Value column, in front of system serial number field
and type the serial number of the console. For example: 123456US7.
5.) Click the write dirty fields button, wait cursor (hourglass) may come up while the data is written to
the component. When the data is written to the component, it will be shown in the Previous Value
column confirming the value has been updated, Both columns should display the updated serial
number.
6.) Click the close button (small x) in the upper right corner to exit.
7.) Reboot the system and proceed to Functional Checks.
8 - 310 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
NOTE: NOT ALL functional checks will have to be repeated, if the functional check was performed
when the previous Option was re-installed. If ONLY a particular Option was removed, perform
the functional checks for the Backplane replacement and that particular Option.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
4-2-7 Basic Measurements
10-7-4 Grounding continuity
10-7-5 Chassis leakage current test
Power SHUT DOWN
4-2-4 DO NOT POWER SHUT DOWN until ALL
functional checks have been performed
LOGIQ E10 Basic Service Manual, Direction
In addition: Perform the following Functional Checks if
5750007-1EN, Rev. 1. Leakage Current
the Option was or Options were removed during
measured at (record the value) and meets
Backplane replacement.
allowable limits. Equipment passed all required
4D checks and is ready for use.
4-2-7 Probe/Connectors Checks
Perform a 4D sweep
CW
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
4-2-7 Probe/Connectors Checks
Chassis Leakage Current Test
10-8-4 Perform a CW scan with P2D, P6D,
M5S-D probes (plug in and scan)
Continued
8 - 312 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
See:
Section Functional Check Debrief Script
ECG
4-2-7 Probe/Connectors Checks
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source) leakage–
10-7-8
lead to lead
Isolated patient lead (sink) leakage-
10-7-9
isolation test
ECG Checks
Active ECG:
Utility -> Applications ->
Settings -> ECG -> Show ECG Tools.
Connect cables. Press Display ECG Tab
4-2-7
on TP. Verify ECG signal displayed on
monitor.
Plug in the ECG cables with nothing
attached so they pick up noise. LOGIQ E10 Basic Service Manual, Direction
Diagnostics 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
Diagnostics
allowable limits. Equipment passed all required
Shear Wave Elastography (Shear Elasto) checks and is ready for use.
4-2-7
Functional Checks
Power Assistant
4-2-7 Probe/Connectors Checks
Verify Power Assistant Option functions
Shear Wave
4-2-7 Probe/Connectors Checks
Shear Wave Elastography (Shear Elasto)
4-2-7
Functional Checks
Perform a 4D sweep, if 4D MC was
removed.
V Nav
4-2-7 Probe/Connectors Checks
3D/4D and Volume Navigation Functional
4-2-7
Checks
4-2-7 Basic Measurements
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts, conductors
and components. Always use non-conductive handles designed for the removal and replacement of ESD
sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before
making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR PACK ARE
OFF BEFORE REMOVING THE CAPACITOR PACK.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to indicate
boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD
PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE OF THE
TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11
J11
LOCKOUT
Signed Date (black and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
J10
8 - 314 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
NOTE: The Option covers are secured to the Option Bracket with M4 threaded studs, as part of the
Option Bracket and M7 flange nuts. All Options may not be present.
All the Option covers, can be removed at once, if desired. The Shear Wave Capacitor Pack is
the only Option in the Option Box that does not have a removable cover.
It is not necessary to remove the ECG (Patient I/O Module) or the Rear I/O, but the USB Cable
at the bottom of the Rear I/O, and all other cabling will have to be removed.
CAUTION
During this procedure, wear cut resistant gloves to avoid scraping your hands when
reaching into the Main Console or blind spaces.
8 - 316 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 318 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
A A A
Rear I/O USB Cable
8 - 320 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 322 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
15. Transfer the four hex screwlocks and the two locking posts the Connector Panel.
Torque to 0.6 +/- 0.1 Nm (0.4 +/- 0.07 lbf-ft).
16. Re-install the Option Bracket. BE SURE to support the Bracket as it is being installed, it has
mass.
Re-install the five M6 screws that secure the Option Bracket to the Base Casting, loosely.
17. Starting at the top, re-install the 7 Phillips screws that secure the Option Bracket to the Card
Cage. Tighten alternately.
Torque the M6 screws to 3.7 +/- 0.6 Nm (2.7 +/- 0.44 lbf-ft), ONLY the M6 screws are torqued to
this value.
18. Position the Option Box up slightly and turn it toward the Main Power Supply to access the
Network Cable at the bottom of the Rear I/O. Re-connect the Network Cable and continue to
position the Option Box onto the studs that secure the Option Box to the Card Cage. Three are
underneath the Option Box, in slots.
Re-install the flange nuts. Tighten all the nuts, alternately, using a 7 mm driver.
19. Place the Subwoofer cable and grounding clamp into the Option Box.
Re-install the screw to secure the clamp to the Option Box. MAKE SURE the clamp is grounding
the cable.
20. Re-install the Battery Bracket.
Install the flange nut and secure the Battery Bracket to the Option Bracket.
8 - 324 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
NOTE: NOT ALL functional checks will have to be repeated, if the functional check was performed
when the previous Option was re-installed. If ONLY a particular Option was removed, perform
the functional checks for the Connector Panel replacement and that particular Option.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
4-2-7 Basic Measurements
10-7-4 Grounding continuity
10-7-5 Chassis leakage current test LOGIQ E10 Basic Service Manual, Direction
Power SHUT DOWN 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
4-2-4 DO NOT POWER SHUT DOWN until ALL
allowable limits. Equipment passed all required
functional checks have been performed
checks and is ready for use.
In addition: Perform the following Functional Checks if
the Option was or Options were removed during
Backplane replacement.
4D
4-2-7 Probe/Connectors Checks
Perform a 4D sweep
Continued
See:
Section Functional Check Debrief Script
ECG
4-2-7 Probe/Connectors Checks
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source) leakage–
10-7-8
lead to lead
Isolated patient lead (sink) leakage-
10-7-9
isolation test
ECG Checks
Active ECG:
Utility -> Applications ->
Settings -> ECG -> Show ECG Tools.
Connect cables. Press Display ECG Tab
4-2-7
on TP. Verify ECG signal displayed on
monitor.
Plug in the ECG cables with nothing
attached so they pick up noise. LOGIQ E10 Basic Service Manual, Direction
Diagnostics 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
Diagnostics
allowable limits. Equipment passed all required
Shear Wave Elastography (Shear Elasto) checks and is ready for use.
4-2-7
Functional Checks
Power Assistant
4-2-7 Probe/Connectors Checks
Verify Power Assistant Option functions
Shear Wave
4-2-7 Probe/Connectors Checks
Shear Wave Elastography (Shear Elasto)
4-2-7
Functional Checks
Perform a 4D sweep, if 4D MC was
removed.
V Nav
4-2-7 Probe/Connectors Checks
3D/4D and Volume Navigation Functional
4-2-7
Checks
4-2-7 Basic Measurements
8 - 326 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
3. Slide the replacement Rear I/O straight onto the Option Box and re-connect the Network and
Power Cables.
4. Re-install all Covers removed.
8 - 328 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 Probe/Connectors Checks 5750007-1EN, (latest revision). Equipment
passed all required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
8 - 330 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Batteries replacement
8 - 332 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
NOTE: Diagnostics can be performed on a Battery when the battery condition is over 15%. The Battery
FRU is shipped at 20~30%. You do not need to check the capacity as a functional test. If the
capacity is less than 15% (normally this does not happen), you will have to wait for it to charge,
it takes around 5 minutes to increase 1%.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
10-8-3 Grounding continuity 5750007-1EN, (latest revision). Leakage
10-8-4 Chassis Leakage Current Test Current measured at (record the value) and
meets allowable limits. Equipment passed all
Verify Power Assistant Option functions, required checks and is ready for use.
if present.
4-2-4 Power SHUT DOWN
8 - 334 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person /
60 minutes, includes Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
Software Load and 8-2-6 "PPE Required During Service" on page 8-6.
testing.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
HDD removal
If the HDD has patient data, leave the HDD at the site.
8 - 336 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
HDD installation
2. Re-connect the HDD cable from J1 to the HDD and tighten the connector screws.
3. Re-install the nut that secures the HDD housings to the DVD bracket. Torque to 1.5 Nm
(1.1 lbf-ft).
4. Re-install all Covers removed.
5. Power up the LOGIQ E10.
6. Re-install the Base Load and Application Software.
7. Enter the TCPIP settings and Option strings for the LOGIQ E10.
8. Restore the Patient Archive images, Report Archive, User-Defined Configuration (Customer
Presets), and Service settings.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
Refer to Confirm DVR Operation required checks and is ready for use.
Basic User
(if installed)
Manual
4-2-4 Power SHUT DOWN
8 - 338 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
CAUTION
During this procedure, wear cut resistant gloves to avoid scraping your hands when
reaching into the Main Console or blind spaces.
3. Re-install the replacement Fan Tray. Make sure the Fan Tray seats completely into the Card
Cage. Re-install the two screws.
8 - 340 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Verify that there is no unusual noise from the Fan Assembly.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
Let the LOGIQ E10 run for 10 minutes and required checks and is ready for use.
observe that the Fans are running and
there are no errors reported by the
LOGIQ E10.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
8 - 342 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
3. Install the replacement Module and the foam seal. Tighten the screws securely, they will bottom
out when they are installed.
The seal will not be tight against the Module. It is intentional for the seal to be loose. The seal
will be pressed against the Card Cage and Front cover to capture the light activity and direct it
through the Front Cover.
4. Route the Module cable along the Card
Cage, as shown and around the back of the
Card Cage to J21 on the ECB.
Connect the cable to J21 on the ECB.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks
5750007-1EN, Rev. 1. Equipment passed all
Verify Network Activity lights emit through required checks and is ready for use.
the Front Cover
4-2-4 Power SHUT DOWN
8 - 344 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
3. Push the Printer Tray in, approximately 1/2 inch to free the three tabs retaining the Tray to the
Option Box.
Remove the Tray.
8 - 346 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 System CFM and PWD Checks
4-2-4 Power SHUT DOWN
8-11-21 Cable Holders - ECB, Main Power Supply and Rear I/O replacement
NOTE: If a Cable is being replaced, perform all of the Functional Checks for that component
replacment.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
45 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
8 - 348 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Table 8-364 Cable Holders - ECB, MPS and Rear I/O replacement
Table 8-364 Cable Holders - ECB, MPS and Rear I/O replacement
2 3
1
J8 (USB2) U
J9 (USB2)
J10 (Network)
M
J11 (USB2)
J12 (USB2) L
J13 (USB2)
J14 (USB2) K
4
5
6
7
To:
G) USB Cable - MPS
1) MPS USB2
J) Rear I/O USB3
2) XYZ Controller USB2
K) Rear I/O USB2
3) BW Printer USB2
L) Rear I/O Network
4) Rear I/O HDMI
M) Rear I/O HDMI
5) Rear I/O Network
N) USB Cable - XYZ Controller
6) Rear I/O USB2
U) USB Cable - BW Printer
7) Rear I/O USB3
4. Re-install all Covers removed.
8 - 350 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Perform a test print, if Printer is present.
Functional Checks
Table 8-365 Cable Holders - ECB, MPS and Rear I/O replacement Functional
Checks
See:
Section Functional Check Debrief Script
If the MPS or Rear I/O Cable Holder was replaced,
perform the following Functional Checks
4-2-4 Power ON/Boot Up
Probe/Connectors Checks
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
10-7-4 Grounding continuity
LOGIQ E10 Basic Service Manual, Direction
10-7-5 Chassis leakage current test
5750007-1EN, Rev. 1. Leakage Current
4-2-5 Top Console checks measured at (record the value) and meets
allowable limits. Equipment passed all required
Perform a 4D sweep
checks and is ready for use.
Power SHUT DOWN
DO NOT POWER SHUT DOWN until ALL
4-2-4
functional checks have been performed, if
required
In addition: Perform the following Functional Checks if
the ECB Cable Holder was replaced
4-2-7 Basic Measurements
4-2-4 Power SHUT DOWN
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal and
replacement of ESD sensitive parts. All parts that have the potential for storing energy must be
discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR PACK ARE
OFF BEFORE REMOVING THE CAPACITOR PACK.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to indicate
boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY
ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE OF THE
TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11
J11
LOCKOUT
Signed Date (black and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/
J10
8 - 352 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CAUTION
During this procedure, wear cut resistant gloves to avoid scraping your hands when
reaching into the Main Console or blind spaces.
8 - 354 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
8 - 356 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
A A A
1. Position the replacement Option Box up, slightly and turn it toward the Main Power Supply and
continue to position the Option Box onto the studs to secure the Option Box to the Option
Bracket. Three are underneath the Option Box, in slots.
Re-install the flange nuts. Tighten all the nuts, alternately, using a 7 mm driver.
2. Transfer the Rear I/O, Patient I/O Filler or Patient I/O Option, if present.
3. Place the Subwoofer cable and grounding clamp into the Option Box.
Re-install the screw to secure the clamp to the Option Box. MAKE SURE the clamp is grounding
the cable.
4. Re-install the Battery Bracket Assembly.
Install the flange nut and secure the Battery Bracket Assembly to the Option Bracket.
6. Re-install the Printer Tray, Printer Filler Storage or the Printer, if present
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Perform a test print, if Printer is present.
8 - 358 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
NOTE: NOT ALL functional checks will have to be repeated, if the functional check was performed
when the previous Option was re-installed. If ONLY a particular Option was removed, perform
the functional checks for the Backplane replacement and that particular Option.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
4-2-7 Basic Measurements LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
Power SHUT DOWN required checks and is ready for use.
4-2-4 DO NOT POWER SHUT DOWN until ALL
functional checks have been performed
In addition: Perform the following Functional Checks if
the Option was disconnected from the Back Plane.
Continued
See:
Section Functional Check Debrief Script
ECG
4-2-7 Probe/Connectors Checks
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source) leakage–
10-7-8
lead to lead
Isolated patient lead (sink) leakage-
10-7-9
isolation test
LOGIQ E10 Basic Service Manual, Direction
ECG Checks
5750007-1EN, Rev. 1. Leakage Current
Active ECG:
measured at (record the value) and meets
Utility -> Applications ->
allowable limits. Equipment passed all required
Settings -> ECG -> Show ECG Tools.
checks and is ready for use.
Connect cables. Press Display ECG Tab
4-2-7
on TP. Verify ECG signal displayed on
monitor.
Plug in the ECG cables with nothing
attached so they pick up noise.
Diagnostics
Shear Wave Elastography (Shear Elasto)
4-2-7
Functional Checks
Power Assistant
4-2-7 Probe/Connectors Checks
Verify Power Assistant Option functions
8 - 360 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
6 hours 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Continued
8 - 362 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
CAUTION
During this procedure, wear cut resistant gloves to avoid scraping your hands when
reaching into the Main Console or blind spaces.
NOTE: When removing the HDD, it is not necessary to remove the HDD from HDD Mounting Bracket
or the Rear I/O from the Option Box.
4. Slide the Card Cage out of the console from the right side.
5. Slide the replacement Card Cage into the console from the right side. Secure the Card Cage to
the Base Casting and Card Rack Mount. Torque to 10 +/- 10% Nm (7.4 +/- 0.74 lbf-ft).
8 - 364 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Perform a test print, if Printer is present.
Functional Checks
NOTE: NOT ALL functional checks will have to be repeated, if the functional check was performed
when the previous Option was re-installed. If ONLY a particular Option was removed, perform
the functional checks for the Card Cage replacement and that particular Option.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
4-2-7 Basic Measurements
10-7-4 Grounding continuity
10-7-5 Chassis leakage current test
Power SHUT DOWN
4-2-4 DO NOT POWER SHUT DOWN until ALL
functional checks have been performed
LOGIQ E10 Basic Service Manual, Direction
In addition: Perform the following Functional Checks if
5750007-1EN, Rev. 1. Leakage Current
the Option was or Options were removed during
measured at (record the value) and meets
Backplane replacement.
allowable limits. Equipment passed all required
4D checks and is ready for use.
4-2-7 Probe/Connectors Checks
Perform a 4D sweep
CW
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
4-2-7 Probe/Connectors Checks
Chassis Leakage Current Test
10-8-4 Perform a CW scan with P2D, P6D,
M5S-D probes (plug in and scan)
Continued
8 - 366 Section 8-11 - Front End Acquisition / Card Cage parts replacement
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
See:
Section Functional Check Debrief Script
ECG
4-2-7 Probe/Connectors Checks
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source) leakage–
10-7-8
lead to lead
Isolated patient lead (sink) leakage-
10-7-9
isolation test
ECG Checks
Active ECG:
Utility -> Applications ->
Settings -> ECG -> Show ECG Tools.
Connect cables. Press Display ECG Tab
4-2-7
on TP. Verify ECG signal displayed on
monitor.
Plug in the ECG cables with nothing
attached so they pick up noise. LOGIQ E10 Basic Service Manual, Direction
Diagnostics 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
Diagnostics
allowable limits. Equipment passed all required
Shear Wave Elastography (Shear Elasto) checks and is ready for use.
4-2-7
Functional Checks
Power Assistant
4-2-7 Probe/Connectors Checks
Verify Power Assistant Option functions
Shear Wave
4-2-7 Probe/Connectors Checks
Shear Wave Elastography (Shear Elasto)
4-2-7
Functional Checks
Perform a 4D sweep, if 4D MC was
removed.
V Nav
4-2-7 Probe/Connectors Checks
3D/4D and Volume Navigation Functional
4-2-7
Checks
4-2-7 Basic Measurements
Section 8-12
MPS (Main Power Supply) replacement
8-12-1 Purpose of this section
This section describes how to replace the Main Power Supply and components and to service parts.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
MPS removal
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Leakage Current
10-7-4 Grounding continuity measured at (record the value) and meets
allowable limits. Equipment passed all required
10-7-5 Chassis leakage current test
checks and is ready for use.
Perform a 4D sweep
4-2-5 Top Console checks
4-2-4 Power SHUT DOWN
Section 8-13
Peripherals replacement
8-13-1 Purpose of this section
Follow the instructions in this section to replace peripherals.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up LOGIQ E10 Basic Service Manual, Direction
4-2-7 Probe/Connectors Checks 5750007-1EN, Rev. 1. Leakage Current
measured at (record the value) and meets
10-7-4 Grounding continuity allowable limits. Equipment passed all required
4-2-4 Power SHUT DOWN checks and is ready for use.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-388 Drive Bay Storage Drawer (DVD Storage) replacement Functional
Checks
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 B-Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 System CFM and PWD Checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks LOGIQ E10 Basic Service Manual, Direction
Shear Wave Elastography (Shear Elasto) 5750007-1EN, Rev. 1. Equipment passed all
4-2-7 required checks and is ready for use.
Functional Checks
4-2-4 Power SHUT DOWN
Section 8-14
Cables replacement
8-14-1 Purpose of this section
This section describes how to replace the cables used by LOGIQ E10.
Tools / PPE
For tools needed, refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
8-2-6 "PPE Required During Service" on page 8-6.
Preparations
When preparing for the replacement, you must perform the following steps:
1.) The cable with the D-SUB connector was disconnected from the Controller Board when you
removed the Operator Panel, Upper.
2.) Disconnect the HDMI cable from the Bulkhead.
3.) Disconnect the cable from the Z switch (Up/Down control).
4.) If not already done, unscrew the fixing screws on the Ground (GND) Clamp on the Bulkhead
Bracket, so you can release the cables from the clamp.
Two cables are connected to the XY (Frog) brakes. They are routed to connectors inside the rear XY
(Frog) legs. To disconnect these cables, follow this procedure:
1.) Remove two screws from the Cable Chain’s anchor point at the lower UI.
2.) Remove one screw on the other end of the Cable Chain.
3.) Remove the Main Cable, including the Cable Chain away from the LOGIQ E10.
Main Cable (Harness) installation
1.) Route the Main Cable, including the Cable Chain into its position.
2.) Install the two screws used to attach the Cable Chain to the UI (Torque: 2.2 Nm {1.6 lbf-ft}).
3.) Install the screw at the other end of the Cable Chain (Torque: 6 Nm {4.4 lbf-ft}).
Route the Main Cable following the steps in
Table 8-393 "Main Cable (Harness) Routing" on page 8-388.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 Color Mode Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-7 PW/CW Doppler Mode Checks
4-2-5 Top Console checks
4-2-4 Power SHUT DOWN
Section 8-15
V Nav Roll Stand Assembling or Replacement
The purpose of this section is to describe assembly or replacement of the Roll Stand.
NOTE: Be sure the wheels on the Roll Stand are not locked and move the Roll Stand slowly, with
caution to avoid tip over. DO NOT move with the arm extended.
NOTE: The Roll Stand basket is designed to hold the transmitter, which weighs about five pounds
(2.27 kg). DO NOT overload the basket.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes 8-2-6 "PPE Required During Service" on page 8-6.
LOCKOUT
& 3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
Signed Date
Figure 8-11 Volume Navigation Roll Stand Installation Kit - See: Table 8-397 "Volume Navigation Roll Stand
Installation Kit - Parts Reference" on page 8-396 for item description
Table 8-397 Volume Navigation Roll Stand Installation Kit - Parts Reference
1.) Insert the post in the base. The locator pin ensures correct alignment.
Figure 8-12 Post into Base - (1) 5/16 fLAT Washer, (2) 5/16 Lock Washer,
(3) 5/16-18 x 1 inch Screw - Roll Stand only
1
2
3
2.) Using the 1/2 inch socket wrench provided, fasten the post to the base with the hardware shown.
4.) Using the hex wrenches provided, tighten all of the set screws in each accessory to secure the po-
sition in the track.
a.) Tighten the top four set screws with the 3/32 inch hex wrench.
b.) Tighten the center four set screws with the 1/8 inch hex wrench.
c.) Tighten the bottom two set screws with 3/32 inch hex wrench.
NOTE: Make sure the Transformer Support Platform (larger view) in Figure 8-14 "Loosening Knob" on
page 8-399 is installed in the UP position.
NOTE: The channel fits between the front and rear slide plates.
8.) While supporting the arm (loaded with the transmitter), slowly loosen the knob which allows the arm
to gradually move down the channel.
NOTE: The arm must move down the channel very slowly when the knob is loosened, and it must move
easily upward when the lifting arm is in the channel. If the arm free-falls or will not move, follow
the adjustment procedure, see:
8-15-3 "Adjusting Tension on Sliding Mechanism" on page 8-403.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the LOGIQ E10 is operational before returning the
LOGIQ E10 to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
Brakes and Direction Lock Functional
4-2-6 5750007-1EN, Rev. 1. Equipment passed all
Checks
required checks and is ready for use.
4-2-4 Power SHUT DOWN
CAUTION THE COMPANY MAKES NO CLAIMS REGARDING THE EFFICACY OF THE LISTED CHEMICALS
OR PROCESSES AS A MEANS FOR CONTROLLING INFECTION. CONSULT YOUR HOSPITAL’S
INFECTION CONTROL OFFICER OR EPIDEMIOLOGIST. TO CLEAN OR STERILIZE MOUNTED
INSTRUMENTS OR ACCESSORY EQUIPMENT, REFER TO THE SPECIFIC INSTRUCTIONS
DELIVERED WITH THOSE PRODUCTS.
Section 8-16
Options replacement
8-16-1 Purpose of this section
This section describes how to install LOGIQ E10 options.
Contents in this section
OPTION NAME
4D Option
CW Option
DC Printer Option
External Monitor Option
Patient I/O ECG Option
Power Assistant Option
Shear Wave Option
Patient I O ECG Option
V Nav Option
Wireless LAN Option
4D, see: 8-16-4 "4D Option replacement" on Shear Wave, see: 8-16-9 "Shear Wave Option
1. 6.
page 8-409. replacement" on page 8-431.
DC Printer, see: 8-16-8 "Printer (Digital Graphic) V Nav, see: 8-16-10 "V Nav Option
3. 8.
replacement" on page 8-427. replacement" on page 8-434.
Patient I/O ECG, see: 8-16-6 "Patient I/O ECG Wireless LAN Option (on ECB), see: 8-16-11
4. 9.
Option replacement" on page 8-417. "Wireless LAN Option" on page 8-437.
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
4D Option removal
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
4D Option installation
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional tests to confirm the system is operational before returning the system
to the customer.
See
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN. Equipment passed all required
4-2-7 System CFM and PWD Checks tests and is ready for use.
Perform a 4D sweep
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
20 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
CW Option removal
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
CW Option installation
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws that you removed earlier have been installed.
2.) Connect cables and probes removed earlier.
3.) Power up the system to verify that it operates as intended.
Functional Checks
Perform the following functional tests to confirm the system is operational before returning the system
to the customer.
See
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 B-Mode Checks
4-2-7 System CFM and PWD Checks
PW/CW Doppler Mode Checks
4-2-7 LOGIQ E10 Basic Service Manual, Direction
use pencil probes
5750007-1EN, (latest version). Leakage
4-2-7 Probe/Connectors Checks Current measured at (record the value) and
meets allowable limits. Equipment passed all
10-8-3 Grounding continuity
required tests and is ready for use.
10-8-4 Chassis Leakage Current Test
Perform a CW scan with P2D, P6D,
M5S-D probes (plug in and scan)
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person /
Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
30 minutes,
8-2-6 "PPE Required During Service" on page 8-6.
excludes travel
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF BEFORE REMOVING THE CAPACITOR PACK.
WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE
CONNECTED TO ONE OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME,
SPECIFICALLY FOR THAT PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
WARNING Energy Control and Power Lockout for LOGIQ E10:
When servicing parts of the Ultrasound System where there is exposure to voltage greater than 30
Volts:
1. TURN OFF THE SCANNER.
TAG
2. UNPLUG THE SYSTEM.
&
LOCKOUT
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AND THE ON/OFF
Signed Date
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
Verification
Perform the following steps to verify that the product is functioning as intended after this installation:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Perform and record a scan, and play back to confirm proper ECG operation.
Functional Checks
Perform the following functional tests to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
PW/CW Doppler Mode Checks
4-2-7
use pencil probes
10-7-4 Grounding continuity
10-7-5 Chassis leakage current test
Isolated patient lead (source) leakage-
10-7-7
lead to ground
Isolated patient lead (source) leakage–
10-7-8
lead to lead
LOGIQ E10 Basic Service Manual, Direction
Isolated patient lead (sink) leakage- 5750007-1EN, Rev. 1. Leakage Current
10-7-9
isolation test measured at (record the value) and meets
allowable limits. Equipment passed all required
ECG Checks
tests and is ready for use.
Active ECG:
Utility -> Applications ->
Settings -> ECG -> Show ECG Tools.
4-2-7 Connect cables. Press Display ECG Tab
on TP. Verify ECG signal displayed on
monitor.
Plug in the ECG cables with nothing
attached so they pick up noise.
Diagnostics
Shear Wave Elastography (Shear Elasto)
4-2-7 Functional Checks, if the Shear Wave
Option was present.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with
electrical contacts, conductors and components. Always use non-conductive
handles designed for the removal and replacement of ESD sensitive parts. All
parts that have the potential for storing energy must be discharged or isolated
before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE
CAPACITOR PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board
should be unlit, to indicate boards are NOT powered.
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave
Capacitor Pack may be energized, even if the power is turned OFF if the cord is still
plugged into the AC Outlet.
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
NOTE: Diagnostics can be performed on a Power Assistant Battery when the battery condition is over
15%. The Battery FRU is shipped at 20~30%. You do not need to check the capacity as a
functional test. If the capacity is less than 15% (normally this does not happen), you will have
to wait for it to charge, it takes around 5 minutes to increase 1%.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks LOGIQ E10 Basic Service Manual, Direction
5750007-1EN, Rev. 1. Equipment passed all
Verify Power Assistant Option functions required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
20 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
3. Re-connect the USB Printer Cable from J12, of the ECB and the Power Cable from the MPS
(P8, 24VDC) to (PWR) on the Printer.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
LOGIQ E10 Basic Service Manual, Direction
4-2-7 Probe/Connectors Checks 5750007-1EN, Rev. 1. Equipment passed all
required checks and is ready for use.
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
20 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF BEFORE REMOVING THE CAPACITOR PACK.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
LOGIQ E10 Basic Service Manual, Direction
Shear Wave Elastography (Shear Elasto) 5750007-1EN, (latest revision). Leakage
4-2-7
Functional Checks Current measured at (record the value) and
10-8-3 Grounding continuity meets allowable limits. Equipment passed all
required checks and is ready for use.
10-8-4 Chassis Leakage Current Test
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
15 minutes 8-2-6 "PPE Required During Service" on page 8-6.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
For location of the Option in the LOGIQ E10, see: Table 8-401 "LOGIQ E10 Option Location" on page
8-408.
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) Connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) Perform scans requiring V Nav module.
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
See:
Section Functional Check Debrief Script
4-2-3 Power ON/Boot Up
4-2-7 Probe/Connectors Checks
4-2-7 B-Mode Checks
LOGIQ E10 Basic Service Manual, Direction
4-2-7 System CFM and PWD Checks
5750006-1EN, Rev. 1. Equipment passed all
4-2-7 Basic Measurements required checks and is ready for use.
3D/4D and Volume Navigation Functional
4-2-7
Checks
4-2-4 Power SHUT DOWN
Manpower /
Tools / PPE
Time
One person / Refer to: 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6 /
20 minutes 8-2-6 "PPE Required During Service" on page 8-6.
WARNING Risk of electrical shock, system must be turned off. Avoid all contact with electrical contacts,
conductors and components. Always use non-conductive handles designed for the removal
and replacement of ESD sensitive parts. All parts that have the potential for storing energy
must be discharged or isolated before making contact.
IF THE SHEAR WAVE OPTION IS PRESENT, MAKE SURE THE LEDS ON THE CAPACITOR
PACK ARE OFF.
Prior to removing the Card Rack Boards, the LEDs on the end of each board should be unlit, to
indicate boards are NOT powered.
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF. THE ESD WRIST STRAP SHOULD BE CONNECTED TO ONE
OF THE TWO RECEPTICALS INTEGRATED ON THE FRAME, SPECIFICALLY FOR THAT
PURPOSE.
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
BUTTON (WHITE LIGHT ON THE OP PANEL) TO TURN OFF, AS THERE ARE NO TEST POINTS
J10 TO VERIFY ISOLATION.
TAG
&
5. REMOVE THE BATTERY COVER AND DISCONNECT THE POWER CABLES J10 and J11 (black
J11 and red) FROM THE BACKPLANE. See: 8-2-2 "Warnings" on page 8-3 and 8-11-15 "Rear I/O
LOCKOUT
Signed Date
J10
Beware that the Main Power Supply, the Batteries, Power Module, ECB Shear Wave Capacitor Pack
may be energized, even if the power is turned OFF if the cord is still plugged into the AC Outlet.
Chapter 9
Renewal Parts
Section 9-1
Overview
9-1-1 Purpose of this chapter
This chapter gives you an overview of the available Spare Parts for LOGIQ E10 and probes that may
be used on LOGIQ E10.
Section 9-2
List of Abbreviations
Table 9-1 List of Abbreviations
ABBREVIATIONS DESCRIPTION
3D THREE DIMENSIONAL (See: RT3D and 4D)
ACT AC TRANSFORMER
ASSY ASSEMBLY
CTRL CONTROL
EXT. EXTERNAL
I/O INPUT/OUTPUT
INT INTERNAL
OP OPERATOR PANEL
PS POWER SUPPLY
RX RECEIVER
TX TRANSMITTER
XFRMR TRANSFORMER
CW CONTINOUS DOPPLER
Section 9-3
Parts list groups
Table 9-2 Parts List Groups
Table 9-21 Product Labels on LOGIQ E10 and Consoles used in a Veterinary Environment 9-48
Table 9-22 Box 1 of P/N 5310019-2 Hardware Kit Compartment Contents 9-49
Table 9-23 Box 2 of P/N 5310019-2 Hardware Kit Compartment Contents 9-54
Section 9-4
LOGIQ E10 Model Software Configuration and Hardware
Table 9-3 LOGIQ E10 Software Configuration and
Hardware
MODEL SOFTWARE
DESCRIPTION
NUMBER VERSION
LOGIQ E10,
5935000E10 R1.x.x
100-240 VAC
Section 9-5
Software
Table 9-4 LOGIQ E10 Software
1.
9-4 Section 9-4 - LOGIQ E10 Model Software Configuration and Hardware
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 9-6
Covers and Bumpers
Table 9-5 Covers and Bumpers - LOGIQ E10 1 of 6
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
15.
1
16. Contains:
1. Pan Arm Down Cover
2. Lift Arm Cover - Right
3. Lift Arm Cover - Left
4. Joint Cover
5. Extension Arm Cover
6. Rotation Limit Screw
LCD Arm Plastic Covers
(Ergotron Arm)
5957000-1 3 4 1 Y
1
2
6
5
17.
18.
19.
20.
21.
22.
24.
25.
26.
27.
Section 9-7
Top Console parts
Table 9-6 Top Console parts 1 of 5
! _ +
3. Norwegian Silicone Keyboard Cover 6208000-52 1
@
2
#
3
$
4
%
5
^
6
&
7
*
8
(
9
)
0 - = 1 Y
Q W E R T Y U I O P { } |
4. Swedish Silicone Keyboard Cover 6208000-53 [ ] \ 1 Y
A S D F G H J K L : "
; '
5. Russian Silicone Keyboard Cover 6208000-54 Z X C V B N M < > ?
1 Y
, . / Pg Up
6. Greek Silicone Keyboard Cover 6208000-55 Ctrl Alt Alt Gr Ins Del
Home End Pg Dn
1 Y
9.
10.
11.
13.
15.
Y TVI
PW
X Z
CW PDI
M CF
3D/4D Elasto V Nav LOGIQ Contrast
View
6208000-1 1 Y
Esc
~ ?
F1 F2 F3 F4 F5 F6 F7
1
F8 F9 F10
CHI
Set Home 2 Word Prt Sc
` Help Arrow Eject Spooler Macro Macro Home Copy Move Delete
Q
@
2
W
#
3
E
$
4
R
%
5
T
^
6
Y
&
7
U
*
8
I
(
9
O
)
0
P
_
-
{
+
=
} |
Abc
B
Auto
[ ] \
A S D F G H J K L : L R
"
; ' Clear P4
Z X C V B N M < > ?
, . / Pg Up
P1 P2 P3
Ctrl Fn Alt Alt Gr Ins Del Freeze
Home End Pg Dn
LOGIQ Contrast
V Nav View
16.
19.
20.
21.
22.
23.
25.
26.
27.
28.
29. Cable FF
Section 9-8
XYZ Mechanism parts
Table 9-7 XYZ Mechanism parts
2.
3.
XY Mechanism GC200461-2
5.
6.
7.
Section 9-9
Main Console parts
Table 9-8 Main Console parts 1 of 4
2.
3.
4.
7.
8.
9.
11.
12.
13.
14.
15.
16.
18.
19.
20.
21.
Section 9-10
Casters (Wheels) parts
Table 9-9 Casters (Wheels) parts 1 of 2
1.
2.
3.
4.
5.
6.
Section 9-11
Card Rack parts
Table 9-10 Card Rack parts 1 of 4
1.
2.
3.
4.
5.
7.
F E DC B A Z 1 F E DC B A Z 1 F E DC B A Z F E DC B A Z
1 1
11 11 11 11 11 11 11 11
P5 P9 P8 P1
15 15 15 15 15 15 15 15
P10 P7 P2
P6
22 22 22 22
22 22 22 22
F E D C B A Z F E DC B A Z F E DC B A Z E D FC B A Z
8.
9.
10.
Section 9-12
Back End parts
Table 9-11 Back End parts
2.
3.
4.
5.
6.
7.
Complies with
IDA Standard
DB 02914
HDIF15000519
8.
BIOS Battery
2404028-7 1 Y
Type BR2032
Section 9-13
Main Power Supply
Table 9-12 Main Power Supply
1.
Section 9-14
Peripherals
9-14-1 Printers
Table 9-13 Printers
1.
2.
1.
Section 9-15
Mains Power Cables
Table 9-15 Mains Power Cables 1 of 2
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
1.
2.
3.
4.
5.
6.
7.
8.
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
9.
10.
11.
12.
13.
14.
Section 9-16
Internal Cables
Contents in this section
USB
GEL WARMER
TEMP
CONTROL F
BB
EXT PORTS
POWER UPPER OP R7
5V
SUB USB3
J25
WOOFER R
HUB
BULKHEAD
P T BOARD 5V
J21
P1 P5
USB USB3 J16
4D SHEAR WAVE HUB
PATIENT IO (OPTIONAL) USB
OPTION) MOTORCONTROL P8 J23 J17
(OPTIONAL) AUDIO
P3 48V
3.3V 3.3V
P4 12V 5V 3.3V
Touch Bklight +
P6 +48V
REG Ctrl VPD Ctrl
AA
P10
L 5V TOUCH
4D ADAPTER 12w"
VIDEO
CTRL
CW PROBE
BACKPLANE J22
CONNECTION E1
(BT2010 Ver D
OPTION) BULKHEAD
UP/DN/LOCK
E2 PLATE
BUTTONS
W
OP PANEL R
METAL SHIELD
METAL SHIELD
D
D1
VPD
V
BW PWR
PRINTER USB
(OPTIONAL)
U CC
J
I K
+3.3V
+5V
CTRL
GRAPHICS
+ VPD
CARD
ISOL
NETWORK
ATE
HDMI OUT
SCAN
SVIDEO OUT
L CONVERT
REAR IO
M
24vdc
48vdc
CONTROL
O G
PWR
HARD
DRIVE
SATA
Z MOTOR
X-Y
H LOCK
O
XYZ CONTROLLER
MOTOR
PWR
Q FROG
DVD
BRAKES
SATA
INNER OUTER
LEFT BRAKE LEFT BRAKE
B C
Z
VNAV
INSIDE
Y
BAYBIRD B
(VGU USB N
OPTION)
POWER
INNER
C OUTER
RIGHT BRAKE RIGHT BRAKE
X
1. Cable T
1. Cable N
(750 mm)
Cable - ECB USB to XYZ Controller 5194492-22 1 Y
XYZ MOTOR CONTROL 5194492-22 ECB J11
2. Cable Q
MOTOR CTRL 5310448.... XY LOCK
1. Cable G
2. Cable H
Cable - XYZ Controller Power 5731039 MAIN PS P9 5731039 REV X XYZ CONTROLLER J9
1 Y
3. Cable BB
Cable - PWR-USB, Backplane to
5746119 1 Y
Patient I/O BACKPLANE J2 5746119 ...... PATIENT IO
4. Cable O
5. Cable J
6. Cable M
7. Cable L
8. Cable I
Cable - Power Backplane to Rear I/O 5746118 1 Y
BACKPLANE 5746118 ...... REAR IO J5
9. Cable D
ECB
J5
ECB
J15 BACK
ECB PLAN
J6 E J12
UPPE
R OP
J21
10. Cable U
11. Cable X
12. Cable W
13. Cable Y
ECB J7
14. Cable Z
Cable V Nav 5439524 1 Y
DRIVE BAY
15. Cable R
16. Cable P
17. Cable CC
Cable - Power, ECB to GPU 5761667 5761667 REV X
1 Y
#NAME# #ORIGIN#
#DATE#
18. Cable K
(860 mm)
Cable - USB ECB to Rear IO 5194492-23 1 Y
19. Cable F
Section 9-17
Probes
Table 9-19 Supported Probes 1 of 5
MINIMUM
PROBE CATALOG NUMBER PART NUMBER COMMENTS SOFTWARE
REQUIRED
MINIMUM
PROBE CATALOG NUMBER PART NUMBER COMMENTS SOFTWARE
REQUIRED
MINIMUM
PROBE CATALOG NUMBER PART NUMBER COMMENTS SOFTWARE
REQUIRED
MINIMUM
PROBE CATALOG NUMBER PART NUMBER COMMENTS SOFTWARE
REQUIRED
MINIMUM
PROBE CATALOG NUMBER PART NUMBER COMMENTS SOFTWARE
REQUIRED
Section 9-18
Options
Table 9-20 Options 1 of 4
2.
3.
4.
5.
7.
8.
9.
11.
12.
13. Cable V
Cable - 24V Power Main PS to BW Printer 5758058 1 Y
BW PRINTER 5758058 REV X #NAME# #ORIGIN# #DATE# MAIN PS
14.
18.
Section 9-19
Product Labels on LOGIQ E10 and Consoles used in a Veterinary Environment
Table 9-21 Product Labels on LOGIQ E10 and Consoles used in a Veterinary Environment
PART
ITEM PART NAME NUMBER DESCRIPTION QTY FRU
1.
LOGIQ E10
GE Medical Systems Ultrasound
and Primary Care Diagnostic, LLC
9900 Innovation Drive
Wauwatosa, WI, USA, 53226
5789698REV 2
3.
4.
5.
9 - 48 Section 9-19 - Product Labels on LOGIQ E10 and Consoles used in a Veterinary Environment
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Section 9-20
Hardware Kit
Hardware Kit for LOGIQ E10 (includes two boxes with hardware) See: Figure 9-2 and Figure 9-3 "Box
2 of P/N 5310019-2 Hardware Kit" on page 9-54. See: Table 9-22 and Table 9-23 "Box 2 of P/N
5310019-2 Hardware Kit Compartment Contents" on page 9-54 to identify and keep the kit organized.
External Toothed Lock Washer, DIN 6797A, M4, Arm adapter; Z-Mech; trackball; subwoofer onBT09/BT10
11.)
Spring Steel, Zinc Plated, RoHS Fastener consoles, Removable trackball ground lead
Curved Lock Washer, DIN 128A, M5, Spring Steel, Daeil arm; kortek monitor to ergotron arm (+ if LCD is updated in
15.)
Zinc Plated, RoHS Fastener the field)
17.) Flat Washer, ISO7089, M4, Nylon, RoHS Fastener BEP5 EPS
65.) Hex Standoff M3 x 5, with 6.5 long M3 Male Thread Bulkhead plate
Freedriv Cross Recessed Pan Head Machine Filler bracket; Column cover assy; Z-Mech; Op panel lower;
33.) Screw_M4-0.7_ X10mm Long Zinc Plated removable trackball screws; LCD mon to Ergotron arm; OLED
Steel_RoHS Fastener and 23" mon to arm
75.) Custom Divider Series StowAway Utility Box Hardware kit box
Chapter 10
Care & Maintenance
Section 10-1
Overview
10-1-1 Purpose of Chapter 10
This chapter describes Care and Maintenance on the Ultrasound system and peripherals. These
procedures are intended to maintain the quality of the Ultrasound system’s performance. Read this
chapter completely and familiarize yourself with the procedures before performing a task.
However, some customers’ Quality Assurance Programs may require additional tasks and or
inspections at a different frequency than listed in this manual.
Table 10-1 Contents in Chapter 10
CAUTION DO NOT OPERATE THIS SYSTEM UNLESS ALL BOARD COVERS AND FRAME PANELS ARE
SECURELY IN PLACE, TO ENSURE OPTIMAL SYSTEM PERFORMANCE AND COOLING.
WHEN COVERS ARE REMOVED, EMI MAY BE PRESENT.
Any unpainted or coated metal surface that is grounded to the Base Casting (chassis):
Section 10-2
Why do Maintenance
10-2-1 Keeping Records
It is good business practice that ultrasound facilities maintain records of periodic and corrective
maintenance. The Ultrasound Equipment Quality Check (EQC and IQC), see: 10-8-6 "Ultrasound
Equipment Quality Check (EQC and IQC)" on page 10-30, provides the customer with documentation
that the Ultrasound system is maintained on a periodic basis.
A copy of the Ultrasound Equipment Quality Check form should be kept in the same room or near the
Ultrasound system.
Routine quality control testing must occur regularly. The same tests are performed during each period
so that changes can be monitored over time and effective corrective action can be taken.
Testing results, corrective action and the effects of corrective action must be documented and
maintained on the site.
Your GE service representative can help you with establishing, performing and maintaining records for
a quality assurance program. Contact GE for coverage and/or price for service.
Section 10-3
Maintenance Task Schedule
10-3-1 How often should care & maintenance tasks be performed?
The Care and Maintenance task schedule specifies how often your Ultrasound System should be
serviced and outlines items requiring special attention.
NOTE: It is the customer’s responsibility to ensure the Ultrasound System care and maintenance is
performed as scheduled in order to retain its high level of safety, dependability and
performance.
Your GE Service Representative has an in-depth knowledge of your Ultrasound System and can best
provide competent, efficient service. Contact GE for coverage information and/or price for service.
The service procedures and recommended intervals shown in the Care and Maintenance Task
Schedule assumes that you use your Ultrasound System for an average patient load and not use it as
a primary mobile Ultrasound system which is transported between diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the periodic maintenance frequencies.
Refer to the Customer Care Schedule in the service manual for the Ultrasound System for the
recommended maintenance care schedule. See: Table 10-2 "Customer Care Schedule" on page 10-5.
10-3-1 How often should care & maintenance tasks be performed? (cont’d)
Abbreviations used in the Customer Care Schedule Table 10-2:
Air Filter Grid Remove the filter grid and clean the air filter. • Or at least bi-weekly
Clean LCD Monitor surface and Probe Be careful not to get the cloth
LCD Monitor, Touch
Panel and Probe holder
holder with a fluid detergent in warm water • too wet so that moisture does
on a soft, damp cloth. not enter the entire system.
Functional Checks of all Check general functions and image Also after corrective
probes section 10-5-3 appearance at human body with all • maintenance or as required by
on page 10-9 available Modes. your facilities QA program.
Section 10-4
System maintenance
10-4-1 Preliminary checks
The preliminary checks take about 15 minutes to perform. Refer to the Ultrasound system user
documentation whenever necessary.
Table 10-3 System preliminary checks
2. Paperwork Fill in the top of the EQC inspection form. Record all probes and Ultrasound system options.
3. • Turn the Ultrasound system power on and verify that all fans and peripherals turn on.
• Watch the displays during power up to verify that no warning or error messages are displayed.
Power up
• Where applicable, confirm that the battery is charged. If no AC Input present, use the internal
battery.
4. Probes Verify that the Ultrasound system properly recognizes all probes.
5. Displays Verify proper display on the monitor and touch panel (where present).
Section 10-5
Tools Required
10-5-1 Tools, supplies and equipment
Table 10-4 lists the suggested tools, supplies and equipment that can be used during the maintenance
tasks.
See: Table 10-4 and 8-2-5 "Tools needed for servicing the LOGIQ E10" on page 8-6.
Table 10-4 Tools Supplies and Equipment
Item Comments
Digital Volt Meter (DVM)
Anti Static Kit Typically includes antistatic-mat, wrist strap and ground cable.
CD-R/DVD-R Media
(minimum quad speed)
B/W Printer Cleaning Sheet see printer user manual for requirements.
Color Printer Cleaning Sheet see printer user manual for requirements.
Disposable Gloves
Verify basic B-Mode (2D) operation. Check the basic Ultrasound system controls that affect this
1 B-Mode
mode of operation.
Verify basic CF-Mode (Color Flow Mode) operation. Check the basic Ultrasound system controls
2 CF-Mode
that affect this mode of operation.
3 Doppler Modes Where applicable, verify basic Doppler operation (PW and CW if available). Check the basic
Ultrasound system controls that affect this mode of operation.
Verify basic M-Mode operation. Check the basic Ultrasound system controls that affect this mode
4 M-Mode
of operation.
Where applicable, verify basic 3D Mode operation. Check the basic system controls that affect this
5 3D Mode
mode of operation.
RealTime 4D Where applicable, verify basic RealTime 4D Mode operation. Check the basic system controls that
6
Mode affect this mode of operation.
Basic
7 Check Distance and Tissue Depth Measurement.
Measurements
Perform an Element Test on each probe to verify that all the probe elements and system channels
8 Probe Elements
are functional.
Applicable
Verify the basic operation of all optional modes. Check the basic system controls that affect each
9 Software
options operation.
Options
System
10 Perform the Automatic Tests.
Diagnostic
Transmit/ Use the Visual Channel Utility on the 12L probe to verify that all system xmit/recv channels are
11
Receive functional.
Operator Panel
12 Perform the Operator Panel Test Procedure.
test
14 Touch Panel Where applicable, verify basic Touch Panel display functions.
2 B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3 Color Printer Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4 DICOM Verify that DICOM is functioning properly. Send an image to a DICOM device.
3 Verify Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
Scratches &
2 Inspect the exterior for dents, scratches or cracks.
Dents
3 Covers Where applicable, verify all covers are secured in place and are properly aligned with other covers.
Replace any covers that are damaged.
4 Input Power Refer to: 10-5-5 "Mains cable inspection" on page 10-10.
• Where applicable, check all wheels and casters for wear and verify operation of foot brake, to stop
Wheels and
6 the Ultrasound system from moving, and release mechanism.
Brakes
• Where applicable, check all wheel locks and wheel swivel locks for proper operation.
• Where applicable, verify ease of Operator Panel (Operator Control Panel) movement in all
Control Panel
7 acceptable directions.
Movement
• Where applicable, ensure that the Control panel latches in position as required.
Control Panel
8 Check for proper operation of all operator panel and TGC lights.
Lighting
10 Monitor Light Check for proper operation of any monitor lighting, if available.
Cables and Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
11
Connectors special attention to probe strain or bend reliefs.
Shielding and Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
12
Covers Missing covers and hardware could cause EMI/RFI problems while scanning.
13 Control Panel Inspect alphanumeric keyboard and Operator Panel. Record any damaged or missing items.
14 Probe Holders Where applicable, inspect the Probe Holders for cracks or damage.
Power and
15 System Status Check for proper operation of all Power and System Status Indicators.
Indicators
Where applicable, check that the battery is not damaged, does not leak, does not emit an odor, and
16 Battery is not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal,
10-5-9 Cleaning
Refer to the User Manual for the Ultrasound System for instructions.
Section 10-6
Using a Phantom
The use of a Phantom is not required during Preventive Maintenance. Customer may use it as part of
their Quality Assurance Program tests. Refer to the Phantom’ s User Manual for information on how to
use it.
Section 10-7
Electrical Safety Tests
10-7-1 Overview
NOTE: For all instructions in the “Electrical safety tests” section in case of using a UPS (un-interruptible
power supply) the terms outlet, wall outlet, AC wall outlet and power outlet refer to the AC power
outlet of the UPS. In case of further available AC (or DC) power outlets at the same used UPS,
these must remain unused i.e.not connected to any other devices.
These tests may refer to particular safety analyzer equipment as an example. Always consult the
manufacturer’s user manual of the Safety Analyzer that will be used to perform the tests.
Prior to initiating any electrical test, the Ultrasound system must be visually inspected. Perform the
following visual checks:
• Check for missing or loose enclosure covers that could allow access to internal live parts.
• Examine the mains cord, mains plug and appliance inlet for damaged insulation and adequacy of
strain relief and cable clamps.
• Locate and examine all associated transducers. Inspect the cables and strain relief at each end.
Inspect the transducer enclosure and lens for cracks, holes and similar defects.
Equipment users must ensure that safety inspections are performed whenever damage is
suspected and at least every 12 months in accordance with local authorities and facility
procedures. Do not use the Ultrasound system or individual probes which fail any portion of the
safety test.
NOTE: For all instructions in the “Electrical safety tests” section in case of using a UPS (un-interruptible
power supply) the terms outlet, wall outlet, AC wall outlet and power outlet refer to the AC power
outlet of the UPS. In case of further available AC (or DC) power outlets at the same used UPS,
these must remain unused i.e. not connected to any other devices.
WARNING To minimize risk of electric shock, only trained persons are allowed to perform the
electrical safety inspections and tests.
CAUTION To avoid electrical shock, the Ultrasound system under test MUST NOT be connected to other
electrical equipment. Remove all interconnecting cables and wires. The Ultrasound system
under test must not be contacted by users or patients while performing these tests.
CAUTION Possible risk of infection. Do not handle soiled or contaminated probes and other components
that have been in patient contact. Follow appropriate cleaning and disinfecting procedures
before handling the equipment.
Single fault Condition for Medical Equipment (ME EQUIPMENT) shall be so designed and
manufactured that it remains SINGLE FAULT SAFE, or the RISK remains acceptable as determined
through application of 4.2.
CAUTION Compare all safety-test results with safety-test results of previously performed safety tests (e.g.
last year etc.). In case of unexplainable abrupt changes of safety-test results consult
experienced authorized service personnel or GE for further analysis.
Table 10-9 Leakage current limits for Ultrasound system operation on 100-120 Volt mains
(US/Canada/Japan)
System Grounding/PE
Leakage Current Test
Power Conductor Limit in mA (1)
Closed 0.1
Type BF Applied Parts On (transmit)
Open 0.5
Closed 0.01
Type CF Applied Parts On (transmit)
Open 0.05
(1) UL standard
Table 10-10 Leakage current limits for Ultrasound system operation on 230-240 Volt mains
Closed 0.01 10
Type CF Applied Parts On (transmit)
Open 0.05 50
NOTE: Leakage current limits for Ultrasound system operation on 230-240 Volt mains
Table 10-11 ISO and Mains Applied (On a calibrated Electrical Safety Analyzer compliant with
AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551.) Limits*
*ISO and Mains Applied (On a calibrated Electrical Safety Analyzer compliant with AAMI/ESI 1993 or
IEC 60601 or AS/NZS 3551.) refer to the sink leakage test where mains (supply) voltage is applied to
the part to determine the amount of current that will pass (or sink) to ground if a patient contacted mains
voltage.
Table 10-12 Equipment Type and Test Definitions
Applied Parts Parts or accessories that contact the patient to perform their function. For ultrasound
(AP) equipment, this includes transducers, ECG leads and e-TRAX Needle Sensor.
Sink The current resulting from the application of mains voltage to the applied part. This test
Leakage is required test for Type CF applied parts.
Measure the resistance from the third pin of the attachment plug to the exposed protectively - earthed
metal parts of the case. The ground wire resistance should be less than 0.2 ohms.
Reference the procedure in the IEC60601-1.
Figure 10-2 Ground Continuity Test
2 1
3
Ω 4
CAUTION Equipment damage possibility. Never switch the Polarity and the status of Neutral when the
Ultrasound system is powered ON. Be sure to turn the Ultrasound system power OFF before
switching them using the POLARITY switch and/or the NEUTRAL switch. Otherwise, the
Ultrasound system may be damaged.
Definition
Also known as Enclosure Leakage current test, this test measures the current that would flow through
a grounded person who touches the accessible conductive parts of the equipment during normal and
fault conditions.
The test verifies the isolation of the power line from the chassis.The testing meter is connected to parts
of the equipment, easily contacted by the user or patient.
Generic Procedure
The test verifies the isolation of the power line from the chassis.
The testing meter is connected from accessible metal parts of the case to ground.
5.) Test opening and closing the ground with the scanner on and off as indicated in Table 10-9 or
Table 10-10 as applicable.
NOTE: Consult the manufacturer’s user manual of the Safety Analyzer that will be used to perform the tests.
The maximum allowable limit for chassis source leakage is shown in:
Measurements should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by Ultrasound system.
Table 10-13 Typical data format for recording enclosure/chassis leakage
Off Open
On Closed
On Open
Definition
This test measures the current which would flow to ground from any of the isolated ECG leads. The
meter simulates a patient who is connected to the monitoring equipment and is grounded by touching
some other grounded surface.
Generic Procedure
5.) Test opening and closing the ground with the scanner on and off as indicated in Table 10-9 or
Table 10-10 as applicable.
NOTE: Consult the manufacturer’s user manual of the Safety Analyzer that will be used to perform the tests.
H3 2 1
G
4
5
A
Table 10-16 on page 10-27 shows a typical format for recording the patient lead to lead leakage current.
Measurements should be recorded from each lead combination under each set of test conditions
specified in:
• Test opening and closing the ground with the scanner on and off as indicated in Table 10-9 or
Table 10-10 Typical data format for recording patient lead to lead leakage
Table 10-14 Typical data format for recording isolated lead (sink) leakage
System Lower
Power Grounding/PE RA-LA Frame LA-LL LL-RA
System Off Closed
System On
Open
(Transmit)
Table 10-15 on page 10-24 shows a typical format for recording the isolated patient lead sink leakage
current.
Measurements should be recorded for the full lead combination under each set of test conditions
specified in:
Off Closed
• Test opening and closing the ground with the scanner on and off as indicated in Table 10-9 or
Table 10-10 as applicable.
This test measures the current that would flow to ground from any of the probes through a patient who
is being scanned and becomes grounded by touching some other grounded surface.
NOTE: Some leakage current is expected on each probe, depending on its design. Small variations in
probe leakage currents are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement. It is abnormal if no leakage current is measured. If no leakage
current is detected, check the configuration of the test equipment.
Generic Procedure on Probe Leakage Current
The most common method of measuring probe leakage is to partly immerse the probe into a saline bath
while the probe is connected to the Ultrasound system and active. This method measures the actual
leakage current resulting from the transducer RF drive.
2 1
9
8
5
3
4 7
A6
NOTE: Each probe will have some amount of leakage current, dependent on its design. Small variations in
probe leakage currents are normal from probe to probe. Other variations will result from differences in
line voltage and test lead placement.
The ultrasound probe’s imaging area is immersed in the Saline solution along with a grounding probe
from the test meter to complete the current path.
NOTE: The Saline solution is a mixture of water and salt. The salt adds free ions to the water, making
it conductive. Normal saline solution is 0.9% salt and 99.1% water. If ready-mixed saline
solution is not available, a mixture of 1 quart or 1 liter water with 9 or more grams of table salt,
mixed thoroughly, will substitute.
CAUTION To avoid probe damage and possible electric shock, do not immerse probes into any liquid
beyond the level indicated in the probe users manual. Do not touch the probe, conductive liquid
or any part of the unit under test while doing the test.
CAUTION Equipment damage possibility. Never switch the Polarity or the status of the Neutral when the
Ultrasound system is powered on. Power off the Ultrasound system, allow the stored energy to
bleed down, and turn the circuit breaker off BEFORE switching the POLARITY switch and/or the
NEUTRAL switch on the leakage meter to avoid possible power supply damage
Table 10-16 on page 10-27 shows a typical format for recording ultrasound probe source leakage
current.
Table 10-16 Typical data format for recording probe (source) leakage
System
Power Grounding/PE 4C i12L TS E8C
Off Closed
Off Open
On Closed
On Open
Mains on applied part is one of the described leakage current tests applicable for probes (Ref:
IEC60601-1). This is to be performed with the probe disconnected from the Ultrasound system. Apply
mains voltage over the insulation barrier. (Between protective earth on the probe connector, and an
electrical anode in saline solution. The patient applied part of the probe is immersed into the saline
solution.) Measure current flowing in the circuit. = leakage current.
As a minimum, tests according to IEC606601-1 must be performed once a year. The requirements for
Body Floating (BF) have to be applied for TEE and Trans thorax probes bearing the symbol for safety
class BF.
Where applicable, a typical test setup of non-TEE Probes can be as illustrated in: "Set Up for Probe
Leakage Current" on page 10-25.
The test passes when the reading measure less than the values in: Table 10-11 on page 10-15.
Section 10-8
When There’s Too Much Leakage Current...
10-8-1 Chassis Fails
Check the ground on the power cord and plug for continuity.Ensure the ground is not broken, frayed, or
intermittent. Replace any defective part.
Where applicable, tighten all grounds. Ensure star washers are under all ground studs.
Test the wall outlet; verify it is grounded and is free of other wiring abnormalities. Notify the user or
owner to correct any deviations. As a work around, check the other outlets to see if they could be used
instead.
NOTE: No outlet tester can detect the condition where the white neutral wire and the green grounding
wire are reversed. If later tests indicate high leakage currents, this should be suspected as a
possible cause and the outlet wiring should be visually inspected.
NOTE: Each probe will have some amount of leakage, dependent on its design. Small variations in
probe leakage currents are normal from probe to probe. Other variations will result from
differences in line voltage and test lead placement. The maximum allowable leakage current for
body surface contact probe differs from an inter-cavity probe. Be sure to enter the correct probe
type in the appropriate space on the check list.
If excessive leakage current is slot dependent, inspect the Ultrasound system connector for bent pins,
poor connections, and ground continuity.
Where applicable, in the case of using a UPS (un-interruptible power supply), perform the tests in the
“Electrical Safety tests” section without using the UPS (i.e. directly connect the Ultrasound system to
the AC wall outlet). If this leads to a pass result, the specific UPS must no longer be used.
New Unit
If the leakage current measurement tests fail on a new Ultrasound system and if situation can not be
corrected, submit a Safety Failure Report to document the Ultrasound system problem. Remove
Ultrasound system from operation.
INDEX
A F
Acclimate Time 2-2 Facility Needs 2-9
Acoustic Noise Output 3-16 Functional Checks 4-1
Archiving and Loading Presets 4-24 2D Mode (B mode) Checks 4-18
Average Installation Time 3-2 Basic Controls 4-15
Cineloop Check 4-23
Color Mode Checks 4-21
C Doppler Mode Checks 4-22
Color Mode ECG Check 4-27
Overview 4-22 Locks 4-14
Compatibility M Mode Checks 4-20
Hardware/Software Compatibility 9-4 Multi Image Checks 4-26
Completing the Installation 3-16
Components and Functions 5-1
Configuration 3-35 G
Console Gathering Trouble Data 7-2
Mechanical Checks 4-10
Contact Information 1-33
Conventions H
Conventions Used in Book 1-6 Hardware/Software Compatibility 9-4
Cooling 2-3 Hazard Icons 1-7
Customer Assistance 1-33 Height 3-16
Human Safety 1-18
Humidity 2-2
D
Dangerous Procedure Warnings 1-27
Depth 3-16 I
DICOM Installation
DICOM Option Pre-Installation Requirements 2-13 Warnings 3-2
Worksheet for DICOM Network Information 2-14 Internal Peripherals 3-36
Disconnect Probe 3-20 Introduction to this manual 1-1
Display Platform Mechanical Checks 4-10
K
E Keyboard Mechanical Checks 4-10
Electrical
Needs 2-4
Safety 1-24 L
Specifications 3-17 Labels Locations 1-26
Electromagnetic Compatibility (EMC) 1-31 LCD (Monitor) Adjustments 6-2
Electrostatic Discharge Warning 1-32 Lighting 2-3
EMC 1-31 List of Abbrevations 9-2
EMI 1-31 Loading and Archiving Presets 4-24
Limitations 2-6 Loading Presets from a Removable Disk 4-24
Protection 3-15 LOGIQ E10 Models and Hardware/Software
Environmental Conditions Compatibility 5-2, 9-4
Probes 2-8 LOTO 1-28
ESD 1-32
External Peripherals 3-36
Description 5-28 M
Index Index - 1
GE
DIRECTION 5750007-1EN, REVISION REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
N
Networking Pre-installation Requirements 2-13 S
Safety
Considerations 1-18
O Precaution Messages 1-6
Off-board Peripherals 3-36 Screen Captures 7-10
On-board Peripherals 3-36, 5-27 Service Adjustments 6-1
Operating Temperature 2-2 Site
Optional Peripherals/Peripheral Connection 3-36 Circuit Breaker 2-5
Options (Software) Configuration 3-37 Power Outlets 2-5
Requirements
Desirable Features 2-11
P Software
Packing Materials Software Option Installation 3-37
Recycling Information 3-11 Software Options Configuration 3-37
Peripherals 5-27 Software/Hardware Compatibility 9-4
Checks 3-36 Special Tools, Supplies and Equipment 10-8
Connection 3-36 Standard Hazard Icons 1-7
External Peripherals 3-36 Storage temperature 2-2
Internal Peripherals 3-36 System
Physical Configuration 3-35
Dimensions 3-16 Manufacture 1-35
Inspection 3-12 Overview 1-5
Power Stability Requirements 2-5 Systems Covered by this Manual 1-4
Power Transients 2-5
Voltage Dop-out 2-5
Power Transients 2-5 T
Power-up Procedures Theory 5-1
Voltage Settings 3-17 Time and Manpower Requirements 2-8
Pre-Installation 2-1 Top Console 5-20
Requirements Transporting the scanner 5-20
DICOM 2-13 Trouble Image with Logs 7-4
Preparing Unit for Installation 3-12 Troubleshooting
Probe Gathering Trouble Data 7-2
Disconnect Probe 3-20 Screen Captures 7-10
Probes Trouble Image with Logs 7-4
Disconnecting 4-27 Vital System Information 7-3
Environmental Requirements 2-8
Probes/Connector Usage
Functional Checks 4-17 U
Product Icons 1-10, 1-17 Unpacking the Equipment 3-4
Purchaser Responsibilities 2-9
V
R Voltage
Receiving and Unpacking the Equipment 3-4 Drop-out 2-5
Recycling Information Settings 3-17
Index - 2 -
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
W
Warnings and Cautions 1-18
Width of LOGIQ E10 3-16
Worksheet for DICOM Network Information 2-14
Index - 3
DIRECTION 5750007-1EN, REV. 1 LOGIQ E10 BASIC SERVICE MANUAL
Index - 4 -
© 2018, General Electric Company.