1. PURPOSE:
This instruction describes a method to measure colour of wort, beer and cider, using the
Spectrophotometer.
5. BACKGROUND
Absorbance of wort, beer and Cider is measured at wavelength of 430 nm except for Absorbance of
Strong bow Red Berries and Dark Fruit are measured at wavelength 500nm. The colour in EBC unit
is obtained by multiplying by the factor of 25.
Temperature of sample should be 20-30oC
6. REFERENCE DOCUMENTS:
7. DEFINITION:
E.B.C: European Brewery Convention.
8. EQUIPMENT
8.1. Hach spectrophotometer or Spectrophotometer (UV/VIS) to determine absorbance at 430
nm (accuracy at least ± 0.5 nm; this can be checked using a holmium oxide filter)
HVBVT use Hach DR3900 to check color
8.2. Cuvette Glass 10*10*45 mm.
8.3 Erlenmeyer 250 ml.
8.4. Funnel.
8.5. Pleated filter paper, Schleicher & Schuell No 597½.
8.6. Spoon.
8.7. Membrane filtration unit
8.8. Membrane filters, spore size 0.45 µm
8.9. Tissue paper
9. REAGENT
Standard Kieselguhr (pink powder)
10. DESCRIPTION/ PROCEDURE:
Doc No.: HVB_VT.TEC.WI.04.01.02 UNCONTROLED COPY FOR
Author: TKLong_LA PRINTING
Issue: 5
Approver: DTNTuyen_QCTL Page: 1/3
Eff. Date: 6-Aug-2021
WORK INSTRUCTION OF COLOR
MEASUREMENT OF WORT, BEER AND CIDER
10.1.2. For green beer/cider, turbid beer/cider samples (EOS samples, Fermentation
samples) and BBT samples/ finished product sample of Tiger Platinum
- Remove excess carbon dioxide by shaking about 200 ml of beer or Cider in a 500 ml round
bottom flask (medium neck) at a temperature of 20 °C for one minute.
- Clarify turbid and/or green beer/ or Cider samples, after shaking, by filtering through a pleated
filter paper filled with about 10 gr Kieselguhr (~ 2 big spoons) into a second flask. Discard the
first part (about 50 ml) of the filtrate.
- Filter the sample through a 0.45μm membrane filter. Filtration can be omitted if the turbidity of
the (diluted) sample is less than 1 EBC unit.
- The clarity of the sample can also be checked by measuring its absorbance at 700 nm and at 430
nm.
- Let the degassed sample to about 20 °C. If necessary, dilute the sample so that the absorbance at
430 nm is within the linearity of the spectrophotometer (i.e.: abs between 0.100 and 1.000).
NOTE: Vial for storage filtered sample must be rinsed by this sample to avoid the significant
error.
10.1.3. For Bright beer (BBT samples and finished product sample):
- Pre-treatment sample same as OE measurement ( i.e. degassed and filter through tissue paper 2
times)
10.2. Measurement.
- Check that the wavelength setting is 430 nm ± 0.5 nm (for SB Red Berries wavelength is setting at
500nm)
- Select “User Programs”
10.2.1 For Wort/ Beer/ Cider (not SB Red Berries and Dark Fruit):
10.2.1.1. Select “colour 430-700”
10.2.1.2. Fill the cuvette with distilled water and zero the absorbance to 0.000.
10.2.1.3. Rinse the cuvette and fill with the sample.
10.2.1.4. Abs 700nm/abs 430nm can be read directly on screen of Hach meter
10.2.1.5. Read the absorbance at 430 nm to calculate colour as formula in part 11
11. CALCULATION
Color of sample is calculated on extinction of sample recorded in step 8.
Color of sample (EBC) = 25 * absorbance of sample
For Amber stout beer: color = 25* absorbance of sample*100