Professional Documents
Culture Documents
Ass ThermoMETER CTL en
Ass ThermoMETER CTL en
Ass ThermoMETER CTL en
33)
92 (3.62) 22 (.87)
The thermoMETER CTL is designed for use in industrial and laboratory appli- The thermoMETER CTL sensor works with a double laser sight at a wave- The CTL features a metric M48x1.5 thread and can be directly installed into
4 (.16)
4 (.16) 13 (.51)
cations. It is used for non-contact temperature measurement. length of 635 nm (visible/red). existing mounting devices by using this thread or by using the hexagonal nut
The system must only be operated within the limits specified in the technical The warning sign below is attached to the controller: (default) and fixed mounting bracket (default). M12x1.5
data, see operating instructions, Chap 2. The system must be used in such a
Ø
Avoid rough mechanical force on the sensor.
4.
way that no persons are endangered or machines are damaged in the event
70 (2.76)
LASER
LASER LIGHT
LASER LIGHT
LIGHT
>> Destruction of the system
(.94)
DODO
NOT
NOTSTARE
NOT STARE INTOBEAM
STARE INTO BEAM
of malfunction or total failure of the system. Take additional precautions for
DO INTO BEAM
47 (1.85)
24
CLASS
CLASS 222 LASER
CLASS LASER
LASER
23 (.91)
< 1< mW
1 mW/ 500
/ 500--650
650 nm
nm
Do not look deliberately into the laser beam! Close 100 (3.94)
(1.97 dia.)
ENEN60825
608251:1:2014
(.16)
2014
(.91)
Warnings
23
your eyes, or immediately turn away if the laser beam 8 (.31)
4
THIS PRODUCT COMPLIES WITH
COMPLIES21 CFR
COMPLIES WITH
WITH 1040.10AND
21 CFR 1040.11
CFR1040.10
1040.10
1040.10 AND
AND
COMPLIES WITH 21 AND
Ø50
EXCEPT FORFORCONFORMANCE
Connect the power supply and the display/output device according to the
1040.11
1040.11
1040.11 EXCEPT
EXCEPT
EXCEPT
WITH
FOR
FOR
IEC
CONFORMANCE
CONFORMANCE
CONFORMANCE
60825-1
ED.ED. 3., hits the eye.
52
WITHWITH
WITH IEC IEC
IEC 60825-1
60825-1
60825-1 ED.
ED. 3.,3.,
3.,AS
ASDESCRIBED
AS
DESCRIBED
DESCRIBED
AS DESCRIBED ININ
DATED
DATED
DATED
DATED
IN
LASER
IN LASER
LASER
LASER
MAY
MAY
MAY
MAY
NOTICE
NOTICE
NOTICE
NOTICE
8, 2019.
8,2019.
8, 2019.
2019.
NO.NO.
NO.
NO. 56,56,
56,
56, 13 4 30
safety regulations for electrical equipment. 89 (3.50)
WS
(.51) (.16) (1.18)
(1.94 dia.)
(2.17 dia.)
M48x1.5
>> Risk of injury, damage to or destruction of the sensor and/or the controller max. 120 (4.72)
Notes on CE Marking
Ø49.3
Avoid shocks and impacts to the sensor and the controller.
Ø55
The following apply to the thermoMETER CTL measuring system: Fig. 2 Dimensional drawing of controller, dimensions in mm, not to scale
>> Damage to or destruction of the sensor and/or the controller
-- EU Directive 2014/30/EU The mounting bracket is included in the scope of delivery.
Avoid rough mechanical force on the sensor. -- EU Directive 2011/65/EU, “RoHS” Category 9 M 12 x 1.5
>> Damage to or destruction of the sensor
(.54 dia.)
The sensor satisfies the requirements if the guidelines in the operating instruc- Cable gland
46.27
13.73
tions are maintained in installation and operation. 30 4 (.16) 57
.)
The supply voltage must not exceed the specified limits.
dia
R3.
(1.82)
( .5
85.5 (1.18) (2.24) 8.5 ia.)
Proper Environment
Ø6
>> Damage to or destruction of the sensor and/or controller. Ø4.91 d
50 .26
3
R5
(3.37) (1
-- Protection class: 135°
(1.97)
Protect the sensor cable against damage.
60°
87 (3.43)
Sensor: IP 65 (NEMA 4) Fig. 1 Dimensional drawing of CTL sensor, dimensions in mm, not to scale R28
>> Destruction of the sensor, failure of the measuring device
ia.)
Controller: IP 65 (NEMA 4) The optical path of the beam must be free of any obstructions. °
6d
Never fold the sensor cable and do not bend it in tight radii. The minimum -- Ambient temperature: 49 (1.93) 90
0
>> Deviation in measurement, inaccurate measured result
Ø6
(2.3
bending radius is 14 mm (static). Dynamic movement is not permitted. Sensor: -20 °C ... +85 °C (-4 °F ... +185 °F) 60 (2.36)
>> Damage to sensor cable, failure of the measuring device (+50 °C (+122 °F) when laser is on) Please enable the integrated double laser to precisely align the sensor with
Controller: 0 ... +85 °C (+32 °F ... +185 °F) 1 the object; see operating instructions. Fig. 3 Dimensional drawing of mounting bracket, dimensions in mm, not to
Avoid exposure of sensor (both optics and housing) to cleaning agents that -- Storage temperature:
contain solvents. scale
Sensor: -40 °C ... +85 °C (-40 °F ... +185 °F)
>> Damage to or destruction of the sensor Controller: -40 °C ... +85 °C (-40 °F ... +185 °F) The sensor can be adjusted on 2 axes by using the adjustable mounting
-- Humidity: 10 ... 95 %, non-condensing bracket; see also operating instructions, Optional Accessories.
Avoid abrupt changes in ambient temperature.
>> Inaccurate or incorrect measurements Unpacking/Included in Delivery
-- 1 thermoMETER CTL sensor and sensor cable
MICRO-EPSILON MESSTECHNIK GmbH & Co. KG
-- 1 Controller You can download a PDF of the detailed operating instructions from our Koenigbacher Str. 15
-- 1 Mounting nut and fixed mounting bracket website: 94496 Ortenburg / Germany
-- 1 Assembly instructions Tel. +49 8542 / 168-0 / Fax +49 8542 / 168-90
http://www.micro-epsilon.de/download/manuals/man--thermoMETER-CTL--en.pdf X9771197-A022030HDR
e-mail info@micro-epsilon.com
1) With temperatures < 0 °C, display function is not guaranteed any more. www.micro-epsilon.com *X9771197-A02*
Pin
+8 - 36 VDC
GND
Explanation
Power supply
Power supply ground (0 V)
GND
3V SW
GND
Ground (0 V)
PINK/power supply for laser (+)
PINK/power supply for laser (-)
i When using the thermocouple output, separation of the ground con-
nection GND - housing is recommended in principle.
i
Pressing screw bles! The sensor must F2, F3 = low level
Metal washer Please note that, depending on the output used, different connection High level: ≥ +3 V- +36 V GREEN No alarm active
Rubber washer be grounded!
pins (OUT-mV/mA or OUT-TC) are used. Low level: ≤ +0.4 V - -36 V For advanced settings, such as defining them as low or high alarm
(by changing normally open/closed) or selecting the signal source [TObject,
034J - PQKE - 0JKH
Output channel 2 (only for CTL, CTLG models) THead, TBox], a digital interface (e.g., USB, RS232) and the CompactConnect
The sensor temperature [-20 ... 180 °C as 0 ... 5 V or 0 ... 10 V signal] is output software are required.
to the OUT AMB connection pin. Output channel 2 can also be programmed
Shield
as an alarm output by using the software. Here the object temperature
TObject or controller temperature TBox can also be used as an alarm source Ratio D = Distance From Device Front Edge to Measured Object/S = Spot Size
Fig. 10 Cable installation
instead of the sensor temperature THead. The size of the object to be measured and the optical resolution of the IR thermometer determine the maximum distance between sensor and object. To avoid
measuring errors, the measured object should completely fill the field of vision of the sensor's optical system. This means, the spot must always be at least as
large as or smaller than the measured object.
D 0 150 300 450 600 750 900 1050 1200 1350 1500 1800 2100 2400
Operation Display Mode (example) Settings range Display Mode (example) Settings range CompactConnect Software
After the supply voltage is applied, the sensor starts an initialization routine 142CA Current object temperature Cannot be changed M 01 Multidrop address [1] 01 ... 32 Place the CompactConnect installation CD into the corresponding drive
and shows INIT on the display for a few seconds. Next, the object tempera- (only with RS485 interface) on your PC or download the software from our website at: https://www.
MV5 Signal output in output channel 1 0 - 20 = 0 - 20 mA/
ture is displayed. The color of the display lighting changes depending on the B 9.6 Baud rate in kBaud [9.6] 9.6/19.2/38.4/57.6/115.2 kBaud micro-epsilon.de/download/software/thermoMETER-CompactConnect/.
[0 - 5 V] 4 - 20 = 4 - 20 mA/
alarm settings. If Autorun has been enabled on your computer, the installation wizard starts
MV5 = 0 - 5 V/ Error Messages automatically. Otherwise, please start CDsetup.exe on the CD-ROM.
Configuring the Sensor MV10 = 0 - 10 V/
TCJ = thermocouple output The error messages below may appear on the thermoMETER CTL display: Please follow the instructions in the wizard until the installation has been
The programming keys , and allow the configuration of the sensor completed.
type J/ CTL, CTLF, CTLC-4, CTLC-2, CTLC-6, CTLG models
on site. The display shows the current measurement or the selected func-
TCK = thermocouple output OVER Object temperature too high After installation, the CompactConnect software is available on your desktop
tion. The key moves you to the desired function, and change the
type K UNDER Object temperature too low (as a program icon) and in the start menu.
function parameters - a settings change is applied immediately. If no key
is pressed for more than 10 seconds, the display automatically switches to E0.970 Emissivity [0.970] 0.100 ... 1.100 If you want to uninstall the software, please use Uninstall in the start menu.
^^^CH Sensor temperature too high
showing the calculated object temperature (according to the selected signal T1.000 Transmission [1.000] 0.100 ... 1.100 vvvCH Sensor temperature too low A detailed description of the software is available on the CompactConnect
processing). A 0.2 Mean signal output [0.2 s] A---- = inactive/ software CD.
0.1 ... 999.9 s CTLM-1, CTLM-2, CTLM-3, CTLM-5 models
When the key is pressed, you automatical- System Requirements
1st digit -- Windows XP, Windows Vista, Windows 7, 8 and 10
ly reach the last function called. P---- Maximum signal output [inactive] P---- = inactive/ 0.1 ... 999.9 0x No error -- At least 128 MByte RAM
The “maximum search” and “minimum s/P oo oo oo oo = infinite -- USB interface
1x Sensor probe short-circuited after ground (bn)
search” signal processing functions cannot be V---- Minimum signal output [inactive] V---- = inactive/ 0.1 ... 999.9 s/V -- CD-ROM drive
oo oo oo oo = infinite 2x Box temperature too low
selected concurrently. -- Hard drive with at least 30 MByte storage space
u 0.0 Temperature range lower limit Depending on model/inactive 4x Box temperature too high
[0 °C] for TCJ and TCK output 6x Box temperature probe interrupted Main Functions
Mode Up Down Fig. 11 Display and programming keys n 500.0 Temperature range lower limit 8x Box temperature probe short-circuited after ground -- Visual depiction and
[500 °C] recording of temperature
Restoring Factory Setting [ 0.00 Output signal lower limit [0 V] According to the range of the 2nd digit measurements for later
To reset the thermoMETER CTL to factory-set parameters, first press the selected output x0 No error analysis and documenta-
] 5.00 Output signal upper limit [5 V]
key and then the key and hold both for 3 seconds. x2 Object temperature too high tion
U °C Temperature unit [° C] °C/°F -- Setting all sensor param-
x4 Sensor temperature too low
For confirmation, RESET appears on the display. eters and remote sensor
/ 30.0 Lower alarm limit [30 °C] Depending on model x8 Sensor temperature too high monitoring
Display Mode (example) Settings range // 100.0 Upper alarm limit [100 °C] Depending on model -- Programming signal pro-
xC Sensor temperature probe interrupted (bn)
S ON Laser sighting [On] ON/OFF XHEAD Ambient temperature compen- XHEAD = sensor temperature/ cessing functions
142.3C Object temperature (after signal Cannot be changed sation -40.0 - 900.0 °C (for LT) as fixed -- Scaling outputs and setting
processing) [142.3 °C] [sensor temperature] value for compensation/press- parameters for function
ing and at the same inputs
127CH Sensor temperature [127 °C] Cannot be changed
25CB Box temperature [25 °C] Cannot be changed
time switches back to XHEAD
(sensor temperature) i A detailed description of the commands is available on the Compact-
Connect software CD in the folder: \Commands.