Professional Documents
Culture Documents
PMT Hps 34 SL 33 01
PMT Hps 34 SL 33 01
All local electrical codes, and relevant service and repair regulations
must be observed.
Applicability
All transmitters with a wire probe are supported and can be identified with SW as the first two
characters shown in Table 2 of the model key number.
For example: SLG72_-SW_ _ _ _ _ _ _ _ _ _ _-_ _ _ _-_-_ _ _.
Part Replacement Probes (reference)
This Installation Instruction applies to the following Parts Replacement Probes:
SLGP2_-_-SW_ _ _ _ _ _ _ _ _ _ _-_ _-_ _.
Required Publications
Publication # Publication Title
M5 set screw for wire probe end weight AF 2.5 mm Allen key
(SWA, SWB)
Probe nut and swage (SLG720) AF 8 mm wrench (2 required)
Probe nut and swage (SLG726) AF 14 mm wrench (2 required)
Centering disk bolt (wire probe) AF 17 mm wrench
SLG720 wire swage AF 7 mm wrench
SLG726 wire swage AF 14 mm wrench
Procedures
Step – 1: Determine the wire type of the new wire probe, multi-twist or single twist, see Figure 1.
Multi-twist
Single-twist
Firmware versions:
Step – 3: Remove existing probe using appropriate wrenches specified in Tools required
section (varies depending on the model key). Remove the existing end-of-probe hardware if it is
needed for the new probe.
Step Action
1 Fully thread the nut onto the central conductor (note the correct orientation). Place
Lock Washer on central conductor. Thread the wire swage on to the central
conductor. Torque the nut against probe and lock washer to secure the connection.
Note: Tighten the wire stud and nut to the following torque:
- SLG720 6.0 Nm (4.4 ft-lbs)
- SLG726 15 Nm (11 ft-lbs)
For flanged SLG726 models, ensure the nut does not intrude into the bore of the
process connector. See figure below for acceptable spacing.
2 If applicable, insert wire probe into end weight. Tighten the 3 set screws to secure
end weight to wire probe. Note: Torque set screws to 6 Nm (4.4 ft-lbs)
If the probe length has changed, measure the length from the reference plane to the end of the
probe. Refer to the transmitter dimensions section of the SLG700 Transmitter User’s manual
(34-SL-25-11), for the definition of the reference plane. Keep in mind that for a wire probe
with an end weight, the probe length is measured from the transmitter reference plane to
the top of the end weight. If the wire probe is untrimmed, it is possible to determine the correct
probe length from the factory by following these steps:
Refer to the underlined characters in the spare probe model number. For example:
SLGP2A-C-SW_ _ _ _ _ _ X X X X X-_ _-_ _. The probe length is used by the transmitter
software shown in Table 2. Find the matching model number pattern, calculate the probe length
accordingly, and enter into transmitter using user the interface.
Step – 5: Configure the transmitter. Follow the instructions pertaining to the firmware version.
a) If a field background is being used, disable it. Refer to SLG700 Transmitter User’s
manual 34-SL-25-11.
b) Ensure the probe type is WIRE.
c) If the probe length was changed, ensure that the correct probe length is correctly
configured and reset the transmitter (see above for instructions on determining the
correct probe length).
d) An adjustment will have to be made to the propagation factor and vapor attenuation
depending on the wire type. See Table 3.
c) If the probe length was changed, ensure that the correct probe length is correctly
configured and reset the transmitter (see above for instructions on determining the
correct probe length).
d) If a field background is desired, a new field background can be taken. To enable, refer to
SLG700 Transmitter User’s manuals 34-SL-25-11.
Singapore
Honeywell Pte Ltd.
Phone: +(65) 6580 3278
Fax: +(65) 6445-3033
South Korea
Honeywell Korea Co Ltd
Phone: +(822) 799 6114
Fax: +(822) 792 9015
Process Solutions
Honeywell
1250 W Sam Houston Pkwy S
Houston, USA, TX 77042