Professional Documents
Culture Documents
RMD1609-001 FDR Smart FGXR-S Operation Manual Rev.15-36-169
RMD1609-001 FDR Smart FGXR-S Operation Manual Rev.15-36-169
RMD1609-001 FDR Smart FGXR-S Operation Manual Rev.15-36-169
Introduction
Internal X-ray generator components will heat up during normal use of the generator. This is similar to X-ray
tube heating during normal generator operation. The amount of heat produced is proportional to the product of
kV, mA, and time.
Modern X-ray generators are designed to operate with the majority of X-ray tubes over their rated power ranges.
They are designed for operating duty cycles consistent with practical patient examination routines that
allow for reasonable cooling intervals between X-ray exposures. Insufficient cooling time between
exposures may lead to excessive heat build - up in the generator, which may cause serious generator damage.
Use this procedure when the generator has not been used for several days. This procedure provides for
exposures at medium power before the tube is used at maximum mA or kV values. This will reduce the
possibility of damaging the anode and high voltage components. No test setup is required.
For maximum stability and reliability, use the following techniques at start up:
Select the following:
For a 300 kHU to a 400 kHU tube, use approximately 80 mAs per exposure.
For a 200 kHU to 300 kHU tube, use approximately 64 mAs per exposure.
For a below 200 kHU tube, use approximately 50 mAs per exposure.
Depending on the X-ray tube power rating, select either 100 mA or 200 mA.
of human anatomy. The Diagnostic X-ray System is primarily used in a hospital for diagnosis of diseases in
skeletal, respiratory and urinary systems. Such as the skull, spinal column, chest, abdomen, extremities, and
b) Age: Available all people, but is not intended to use for dedicated pediatric application
e) Nationality: multiple
Maximum • N/A
a) Service engineer
Maximum • N/A
1.6 SPECIFICATIONS
X-ray Generators
FDR Smart FDR Smart FDR Smart FDR Smart FDR Smart
System Model
FGXR-32S FGXR-40S FGXR-52S FGXR-68S FGXR-82S
Generator
GXR-32 GXR-40 GXR-52 GXR-68 GXR-82
Model
Power Rating 32kW 40kW 52kW 68kW 82kW
220-230V~, 400V3~, 480V3~ 400V3~, 480V3~
Line Power
±10% (Frequency: 50/60Hz)
40~125kV,
1kV step
40~125kV, (Option :
kV Range 40~150kV, 1kV step
1kV step 40~150kV for 3
phase line
power)
mA Range 10 to 400mA 10 to 500mA 10 to 640mA 10 to 800mA 10 to 1,000mA
Timer Range 0.001 to 10 sec, 38 steps
AEC Shortest
Irradiation 1ms
Time
mAs Range 0.1 to 500mAs (Optional up to 1000mAs)
Low Speed
Low Speed
(Option:
Rotor Supply (Option: LSS Brake, Dual Speed
LSS Brake,
Dual Speed for 3Ф)
Dual Speed)
Technique
4 point display(kV, mA, Time, mAs)
Selection
Image
2 Bucky + 1 Non-Bucky
Receptors
230VAC, 1A, 230W
External System Power
Auxiliary 110VAC, 1A, 110W
Power Supply Magnetic Lock Power 28VDC, 5A, 140W
Collimator Lamp Power 24VAC, 6.3A, 150W
X-ray Ripple
30kHz
Frequency
Reproducibility Coefficient of Variation: kV < 0.005, Time < 0.005, mAs < 0.01
Accuracy kV < ±(1%+1kV), mA < ±(3%+1mA), Time < ±(1%+0.5ms), mAs < ±(3%+0.1mAs)
2) Elevating table
Model PBT-6
Longitudinal 1,000(±500)mm
Tabletop
Transverse(Lateral) 250(±125)mm
Travel 285(565~850)mm,
Movement Speed 17mm/sec
Vertical
Motorized movement by Foot Switch
Operating
DC-motor (Linear Actuator)
350mm with N-type Bucky
Bucky Longitudinal
310mm with R, RD, RD2, ND, ND2 type Bucky
Laminate : 1.2mmAl at 100kV
Inherent Filtration
Carbon : 0.5mmAL at 100kV
Tabletop
Max. Patient Weight 300kg(660lbs)
Size 2,200(W) x 878(D) x 45(H) mm
3) Mobile table
Model PDT-1
Max. Patient Weight Max. 200kg (441lbs)
Dimension / Weight 2004(W) x 650(D) X 712(H)mm / 62kg (137lb)
Model WBS(Manual)
1,070mm (300~1,370mm from floor to Bucky center)
1,220mm (300~1,520mm from floor to Bucky center)
Vertical Cassette stroke
1,390mm (300~1,690mm from floor to Bucky center)
1,640mm (300~1,940mm from floor to Bucky center)
-N, -R Fixed Grid type / Oscillating Grid type
Bucky Type -RD, -RD2
Removable Grid
-ND, -ND2
Standard tray for CR and Analog cassette
- Tray size: 17”x17” (for 8”x10” to 17”x17” cassette)
-N
- Rotating: Not supported
- Fail safe function (optional)
Rotating tray for DR cassette
- Tray size: 14”x17”(only for 14”x17” cassette)
-R
- Rotating: Center reference
- Fail safe function (optional)
Rotating tray for DR cassette
- Tray size: 14”x17”(only for 14”x17” cassette)
-RD
- Rotating: Center reference
- Automatic connection for D-EVO
- Fail safe function (optional)
Tray Type
Rotating tray for DR cassette
- Tray size: 14”x17”(only for 14”x17” cassette)
-RD2 - Rotating: Center reference
- Automatic connection for D-EVO2
- Fail safe function (optional)
Non-rotating tray for DR cassette
- Tray size: 17”x17”(only for 17”x17” cassette)
-ND
- Automatic connection for D-EVO
- Fail safe function (optional)
Non-rotating tray for DR cassette
- Tray size: 17”x17”(only for 17”x17” cassette)
-ND2
- Automatic connection for D-EVO2
- Fail safe function (optional)
Grid FD 150cm, 103 lpi, ratio 10:1(Optional 215 lpi ration 10:1)
Inherent Filtration 0.5mmAl at 100kV
Lock(Brake) EM Lock, Switch on/off
Balance Counter Weight
Electrical Rating 24VDC, 1A
1,599(H) x 659(W) x 445(D) mm / 110kg(242lbs)
1,749(H) x 659W) x 445(D) mm / 113kg(249lbs)
Dimension / Weight
1,919(H) x 659(W) x 445(D) mm / 116kg(255lbs)
2,169(H) x 659(W) x 445(D) mm / 120kg(264lbs)
* APPLIED PART
Tube stand
1) Floor Mounted
Model TS-FM6(Manual)
Tube Rotation Angle ±135°
2,200mm
Longitudinal
(Optional 3,000mm and 3,700mm)
Tube stroke
Lateral 220mm
Vertical 1,580mm (440~2,020mm from floor to focus)
Lock(Brake) EM Lock, Switch on/off
Balance Counter Weight
Column Rotation 90° step
Electrical Rating 24VDC, 3A
2,317(H) x 1,373(D) mm x 3,006(W)
Dimension
when transverse direction is fully extended
Weight 240kg(529lbs)
Model TS-FC2(Manual)
Tube Rotation Angle ±135°
Longitudinal N/A
Lateral N/A
Tube stroke 1,243mm(294mm~1,537mm from floor to focus)
Vertical 1,413mm(294mm~1,707mm from floor to focus)
1,663mm(294mm~1,957mm from floor to focus)
Vertical Movement Manual
Lock(Brake) EM Lock, Switch on/off
Balance Counter Weight
Electrical Rating 24VDC, 3A
1,749(H) x 659(W) x 790(D) mm / 165kg(363lbs)
Dimension/Weight 1,919(H) x 659(W) x 790(D) mm / 168kg(370lbs)
2,169(H) x 659(W) x 790(D) mm / 171kg(376lbs)
Model TS-FC2(Motorized)
Tube Direction Right-angle or Straight
Tube Rotation N/A
Longitudinal N/A
Lateral N/A
2) Floor-Ceiling Mounted
Model TS-FC6
Tube Rotation Angle ±135°
Longitudinal 2,500mm
Tube stroke Lateral 220mm
Vertical 1,580mm (430~2,010mm from floor to focus)
Lock(Brake) EM Lock, Switch on/off
Balance Counter Weight
Column Rotation 90° step
Electrical Rating 24VDC, 3A
2,465(H) x 1,373(D) mm x 3,600(W)
Dimension
when transverse direction is fully extended
Weight 180kg(397lbs)
3) Ceiling suspended
Model TS-CSA-M
Balance Spring
Model TS-CSA_A
Horizontal axis ±180° (LCD display)
Tube Rotation Angle
Vertical axis ±180° (mechanical detents at every 90°)
Longitudinal 3,280mm(with 4m rail), 4,280mm(with 5m rail)
Tube stroke
Lateral 2,200mm(with 3m rail), 3,200mm(with 4m rail)
(Transverse x Longitudinal)
Vertical 1,600mm
Lock(Brake) EM Lock, Switch on/off
Balance Spring
Vertical Movement Motorized(Motorized supports vertical sync with table and wall stand)
Tube Rotation Manual
Option Auto Collimation, Mechanical detent
SID Indication 7inch Touch Screen LCD with control buttons
Electrical Rating 100–240VAC, 200VA, 50/60Hz
2,830(H) x 3,000(D) mm x 4,000(W) mm
Dimension
when vertical direction is fully extended with 1,600mm stroke and 3x4m rails
Main body: 170kg(375lbs) except tube and collimator,
Weight
Rails: 115kg(254lbs, 3x4m rails)
Model TS-CSA_A(Stitching)
Horizontal axis ±180° (LCD display)
Tube Rotation Angle
Vertical axis ±180° (mechanical detents at every 90°)
Longitudinal 3,280mm(with 4m rail), 4,280mm(with 5m rail)
Tube stroke
Lateral 2,200mm(with 3m rail), 3,200mm(with 4m rail)
(Transverse x Longitudinal)
Vertical 1,600mm
Lock(Brake) EM Lock, Switch on/off
Balance Spring
Vertical Movement Motorized (Motorized supports vertical sync with table and wall stand)
Motorized
Tube Rotation
(Motorized supports the source tilting type image stitching operation)
Option Auto Collimation, Mechanical detent
SID Indication 7inch Touch Screen LCD with control buttons
Electrical Rating 100–240VAC, 200VA, 50/60Hz
2,830(H) x 3,000(D) mm x 4,000(W) mm
Dimension
when vertical direction is fully extended with 1,600mm stroke and 3x4m rails
Main body: 170kg(375lbs) except tube and collimator,
Weight
Rails: 115kg(254lbs, 3x4m rails)
X-ray Tube
* Total filtration including X-ray tube assembly and collimator will be matched by appropriate additional filters to
within the range from 2.9 to 3.3mmAl. eq.
If the target angle of the applied tube is 12°, the field size (43cm X 43cm) cannot be
realized by SID 100cm when combined with a collimator. To use the field size (43cm X
43cm), please use SID 105cm or 110cm.
Collimator
Model R108 R302A, R302MLP/A, R302MFMLP/A
Manufacturer RALCO RALCO
Control Manual with 30sec. lamp timer
Field Shape Rectangular
Max. Field Size More than 43x43cm(17x17inch) at 100cm SID
Leakage Radiation Less than 100mR/hr
Max. kVp shield 150kV (Optional 125kV) 150kV
Inherent Filtration 2.0mmAl eq. 2.0mmAl eq.
Over 160LUX at 100cm SID (Typ. Over 160LUX at 100cm SID (Typ.
Luminosity
250LUX) 200LUX)
HLX64638 100W 24V
Light source Single LED
/ OSRAM
Auto collimation for R302 MLP/A and R
Tape measure, Line laser
Standard 302 MFMLP/A
rotating flange
Auto filter selection for R 302 MFMLP/A
Tape measure, line laser, rotating flange,
Option -
LED Type
Electrical Rating 20–30Vac, 30VA, 50/60Hz 24Vac, 6.3A, 50/60Hz
223(W) x 246(D) x 140(H) mm / 195.5(W) x 237(D) x 206.5(H) mm /
Dimension / Weight
6.6kg(14.6lb) 9.4kg(20.7lb)
Model DXC-RML/DXC-RMH
Manufacturer DRGEM
Additional filter
Option
Mounting flange mechanical detent
DAP rail
Manufacturer Philips
Stitching specifications
Options
- AEC
Upgrade AEC 3-field ion chamber for table & stand
- DAP
Upgrade DAP kit, Stand-alone Display type
Upgrade DAP kit, for DR system
- Cassette center tracking
Upgrade table cassette tracking for TS-FM6
- Dual Speed Starter (DSS)
Up-grade DSS for the dual speed rotating tube under the 3 phase(32/40/52KW)
- LSS Brake for Low speed rotor supply
- Console stand
- PC X-Ray console option
OPERATING
The main components of FDR Smart FGXR-S System comply with the regulatory requirements and design
standards in this section as follows:
1) SAFETY
■ EN60601-1:2006/A1:2013
Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
IEC60601-1:2005/A1:2012
■ EN60601-1-3:2008/A11:2016
Medical electrical equipment -- Part 1-3: General requirements for basic safety and essential performance -
Collateral Standard: Radiation protection in diagnostic X-ray equipment
IEC60601-1-3:2008/A1:2013
■ EN60601-1-6:2010/A1:2015
Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance -
Collateral standard: Usability
IEC60601-1-6:2010/A1:2013
■ EN60601-2-28:2010
Medical electrical equipment -- Part 2-28: Particular requirements for the basic safety and essential
performance of X-ray tube assemblies for medical diagnosis
IEC60601-2-28:2010
■ EN60601-2-54:2009/A1:2015
Medical electrical equipment -- Part 2-54: Particular requirements for the basic safety and essential
performance of X-ray equipment for radiography and radioscopy
IEC60601-2-54:2009/A1:2015
2) EMC
■ EN60601-1-2:2015
Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic
compatibility - Requirements and tests
■ EN55011:2009/A1:2010
■ CISPR11:2014
■ EN61000-3-2:2006
■ EN61000-3-3:2013
■ EN61000-4-2:2009
■ EN61000-4-3:2006/A1:2008/A2:2010
■ EN61000-4-4:2012
■ EN61000-4-5:2006
■ EN61000-4-6:2014
■ EN61000-4-8:2010
■ EN61000-4-11:2004
The FDR Smart FGXR-S complies with the requirements of IEC 60601-1-2:2014 regarding electromagnetic
compatibility. Surrounding equipment shall follow the standard IEC 60601-1-2:2014.
Mobile telephones or other radiating equipment can interfere with the function of
the FDR Smart FGXR-S and can therefore cause safety hazards.
The FDR Smart FGXR-S is intended for use in the electromagnetic environment specified below. The
customer or the user of the FDR Smart FGXR-S should assure that it is used in such an environment
The FDR Smart FGXR-S uses RF energy only for its internal
RF emissions
Group 1 function. Therefore, its RF emissions are very low and are not
CISPR 11
likely to cause any interference in nearby electronic equipment.
RF emissions
Class A
CISPR 11
The FDR Smart FGXR-S is suitable for use in all establishments,
Harmonic other than domestic and those directly connected to the public
emissions Not applicable
low-voltage power supply network that supplies buildings used
for domestic purpose. For information purpose the system
Voltage
complies with IEC61000-3-11 and is suitable for connection to
fluctuations/ Flicker
Not applicable public mains network if the impedance is 0,32 Ohm or lower
emissions IEC
61000-3-3
The FDR Smart FGXR-S is intended for use in the electromagnetic environment specified below. The
customer or the user of the FDR Smart FGXR-S should assure that it is used in such an environment.
Electrostatic ± 2 kV, ± 4 kV, ± 6 kV, ± 2 kV, ± 4 kV, ± 6 kV, Floors should be wood, concrete or
discharge (ESD) ± 8 kV contact ± 8 kV contact ceramic tile. If floors are covered with
IEC 61000-4-2 ± 2 kV, ± 4 kV, ± 8 kV, ± 2 kV, ± 4 kV, ± 8 kV, synthetic material, the relative humidity
± 15 kV air ± 15 kV air should be at least 30%.
Electrical fast ± 2 kV for power ± 2 kV for power Mains power quality should be that of a
transient/burst supply lines supply lines typical commercial or hospital
± 2 kV Line to ± 2 kV Line to
Line Line
Voltage dips, short <5 % UT <5 % UT Mains power quality should be that of a
(>95 % dip in UT) for (>95 % dip in UT) for
interruptions and typical commercial or hospital
0,5 cycle at 0, 45, 90, 0,5 cycle at 0, 45, 90,
voltage variations on environment. If the user of the DR- XD
135, 180, 225, 270, 135, 180, 225, 270,
power supply input 315 deg. 315 deg. 200 requires continued operation during
lines. <5 % UT <5 % UT power mains interruptions, it is
(>95 % dip in UT) for (>95 % dip in UT) for
IEC 61000-4-11 recommended that the FDR Smart
1 cycle at 0 deg. 1 cycle at 0 deg.
FGXR-S be powered from an
30 % UT 30 % UT
(70 % dip in UT) for (70 % dip in UT) for uninterruptible power supply or battery.
25/30 cycles at 0 deg. 25/30 cycles at 0 deg.
<5 % UT <5 % UT
(>95 % dip in UT) for (>95 % dip in UT) for
250(50Hz)/300(60Hz) 250(50Hz)/300(60Hz)
cycles at 0 deg. cycles at 0 deg.
NOTE: UT is the A.C. mains voltage prior to application of the test level.
The FDR Smart FGXR-S is intended for use in the electromagnetic environment specified below.
The customer or the user of the FDR Smart FGXR-S should assure that it is used in such an environment.
NOTE 1: At 80 Mhz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation i affected by absorption
and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the FDR Smart FGX-S is
used exceeds the applicable RF compliance level above, the FDR Smart FGX-S should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the FDR Smart FGX-S.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
The FDR Smart FGXR-S is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the FDR Smart FGXR-S can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the FDR Smart FGXR-S as recommended below, according to
the maximum output power of the communications equipment.
3) CLINICAL EVALUATION
4) OTHERS
■ EN ISO 15223-1:2021
Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied -
Part 1: General requirements
ISO 15223-1:2021
■ IEC TR60878:2015
Graphical Symbols for electrical equipment in medical practice
■ IEC60417-1:2002DB
Graphical Symbols for use on equipment-part1 : overview and application
■ EN ISO14971:2012
Medical devices - Application of risk management to medical devices
ISO 14971:2012
■ EN ISO13485:2016
Medical devices - Quality management systems - Requirements for regulatory purposes
ISO13485:2016
■ Regulation (EU) 2017/745
Medical Devices Regulation
■ EN1041:2008/A1:2013
Information supplied by the manufacturer with medical devices
■ EN62304:2006
Medical device software — Software lifecycle processes
IEC62304:2006
■ EN62366:2008
Medical devices - Application of usability engineering to medical devices
IEC62366:2007
2. CONSOLE CONTROLS
LCD display with manual x-ray exposure condition and AEC option selected
Refer to XCONAPP Settings section of the service manual for a list of supported
system types.
Press ON to switch the X-ray generator on. The console will light up
and a brief self - check will be performed.
All front panel LEDs and indicators, including the X-ray exposure indicator, should
light shortly. This is an LED self-test and does NOT indicate the presence of X-
RAYS.
After stand-by indicator lights on, LCD displays logo, model name and firmware version of control console and
HT controller.
The speaker should beep and LCD displays exposure condition at the end of the preparation and power charge.
The prep indicator will light on and the LCD status window will
display 'X-ray exposure ready' message when the system is ready to
make an exposure.
While pressing the prep button, press and hold the EXPOSE button
to make an X-ray exposure.
The X-ray exposure indicator will light on and the LCD status
window will display 'X-ray exposure' message when X-ray exposure is
being made.
Pressing the EXPOSE button only will cycle the generator through prep
and then exposure.
After the exposure, LCD status window displays ‘X-ray exposed’ message
for 1 second.
As for PC interface module, X-ray exposure can be done by PREP and EXP switches as well as the hand
switch.
The ms/mAs button allows you to work in either exposure time (ms) mode or
mAs mode. Pressing this button will toggle between ms and mAs mode.
Xcon_F displays ms with mAs together so this button does not exist.
• In AEC mode, the AEC backup time and calculated Post mAs is displayed after x-ray exposure. The
maximum AEC backup ms and mAs is installer programmable. The generator will then set the backup
ms so as to not exceed the preset ms and mAs limits.
• In mA/ms mode, the calculated exposure mAs will be displayed (based on the mA and ms selected).
• In mAs mode, the calculated exposure ms will be displayed (based on the mAs selected).
(XconF)
(XconF)
(XconF)
The rate of scrolling for the kV, mA, and ms display increases if the buttons are
held down for more than 1 second.
The scroll rate changes after one second.
Specially, kV scroll rate increases to turbo speed 3 seconds after buttons are
pressed.
To return to the slowest scroll rate, release the button and start again.
Pressing the button with lighted LED will turn off AEC feature.
This function works only for the system equipped with AEC option.
2.3.7 AEC FIELD SELECT BUTTONS AND INDICATORS (FOR AEC ONLY)
This function works only for the system equipped with AEC option.
Focal spot size indicator shows currently selected x-ray tube focal spot
size.
GXR series generators support auto focal spot selection feature which is
controlled by a reference mA value accessible only to installation/service
engineer.
The selected focus indicator will blink to indicate that X-ray exposure is
unavailable for the following two cases:
• When focal spot size has changed, generator requires filament preheat
time for selected focus. Filament preheat time is about 4 seconds.
• When X-ray exposure exceeds 100mAs, the generator requires a cooling
time for the IGBT in proportion to mA step and mAs.
Indicates the tube anode heat unit for the selected X-ray tube in percentage.
In reality, tube anode heat unit depends on tube housing and ambient temperature
so it may be different from the one calculated by the x-ray generator.
Therefore, for the safety reason, the maximum tube anode HU rating is set as
90% of maximum HU possible.
When ‘HU Warning Level’ message or Anode HU Error message (E18) appears,
tube has to be cooled down before taking next x-ray exposure.
(Membrane Console)
(XconF)
(XconF)
BUTTON DESCRIPTION
Selects no filter.
Warning message is issued to warn the operator of generation of troubles which is not critical for the
system operation except ‘HU Warning Level’.
When the Warning message is issued, warning indicator turns on for 1 second before the
message disappears.
Refer to Section Warning Messages of Service Manual about descriptions for Warning messages.
The GXR console will display error messages during abnormal operation of the generator.
If the error message is not cleared after the following corrective actions are taken, contact a service
representative.
In order to stop the buzzer sound while the error is not cleared, move to APR selection menu or main
menu in which x-ray exposure is not allowed.
Refer to the section Error Messages of the Service Manual for descriptions of the error messages
Image Receptors are depicted as the console is shipped from the factory. The
receptors may have been reprogrammed during installation. As a result, the
receptors may contain installer - applied icons that differ from those shown in this
manual.
For membrane console, the long pressing of the selected Bucky button over 4
seconds will switch the interface between DR and CR.
If CR is needed, this function is needed. If the generator is powered on or Bucky
is changed, DR will be activated automatically.
The following status message indicates that the external device for the selected Bucky is Digital Radiography.
This status is possible only when the generator has DR option.
(Membrane Console)
(XconF)
3. OPERATOR’S UTILITY
Press the button to enter the user mode from the main menu.
There are limits in certain menu for accessibility while in Service mode or User mode.
Service mode is only accessible by the service engineer due to its importance related to the system
setting.
1. SYSTEM SETUP
2. SYSTEM DIAGNOSIS
Operator can execute the following features while in ‘SYSTEM SETUP’ mode.
1. Change Date
2. Change Time
To perform those functions, move to ‘Page 3/4’ of ‘SYSTEM SETUP’ by pressing button.
To change Date, select ‘21. Date Setting’ in the ‘page 3/4’ of ‘SYSTEM SETUP’, and it will show the following
screen.
Select the month, day, year data to change by using buttons, then change the data by using right-
After change, press button to store the modified data. Buzzer sound follows to confirm to change.
Saving Date Setting is available only when ‘21. Date Setting’ is selected.
To change time, select ‘22. Time Setting’ in the ‘page 3/4’ of ‘SYSTEM SETUP’, and it will show the following
screen.
Select the hour, minutes, seconds data to change by using buttons, then change the data by using
After change, press button to store the modified data. Buzzer sound follows to confirm to change.
Saving Time Setting is available only when ‘22. Time Setting’ is selected.
1. Error Log
2. Statistics
3.3.1 STATISTICS
GXR series generator can store up to 50 most recent error messages. Error log page provides detailed
information regarding the errors.
Up to 3 error codes and messages are displayed as well as date, time, x-ray exposure condition (kV, mA, ms),
image receptor setting, AEC, APR setting, heat unit amount, and communication packet numbers regarding the
errors.
To return to ‘SYSTEM DIAGNOSIS’ menu from ‘Error Log’ page, press button.
When the communication packet is displayed in the error log, one can move to Protocol page by pressing
button.
The protocol is not provided for the operator but for service engineer or remote diagnosis software purpose.
Refer to service manual for detailed information.
It provides a convenient way of diagnosing the generator via a PC through an USB connection.
A remote diagnosis report which contains the information for the diagnosis can be sent to the manufacturer via
internet web-site or e-mail to enhance serviceability.
4. APPARATUS OPERATION
1 Ceiling rail
2 Floor rail
3 Control panel
7 Arm
8 Control box
10 Longitudinal stopper
1 Ceiling rail(optional)
3 Control panel,
8 Arm
9 Control box
12 Longitudinal stopper
1 Control panel
3 Control box
7 Collimator
11 Handle
If the table Bucky tracking (mechanical) option is applied, follow the ways below for its operation.
The longitudinal tube stand travel button will enables the tracking function.
If tracking module at tube stand base and tracking holder at Bucky is not matched, Bucky position will be fixed
by brake, but if the position of tracking module and tracking holder is matched, the brake of Bucky will be
released and move follow the tube column position.
If the tube stand column is moves even though Bucky is reaches at each limit position of Bucky stroke, tracking
will be released and Bucky will be fixed by brake.
If the Bucky position is leaches at the each end of its stroke while tube stand is longitudinally moving with the
tracking function, the tracking will be released.
If the electric column rotation option is applied, rotate the tube stand column while pressing button 4.
1 Vertical stand
3 Control Box
5 Emergency switch
6 Vertical UP Button
8 Tube
9 Collimator
10 Safety Sensor
1 Vertical stand
3 Control Box
5 Emergency switch
6 Vertical UP Button
8 Tube
9 Collimator
10 Safety Sensor
2 Stand console
3 Vertical Stand
4 Arm
6 Safety Sensor
9 Ceiling support(optional)
11 Emergency Switch
15 Line Laser
16 Motorized collimator
If any obstacles are detected by safety sensor, the motorized movement will be
stopped.
1. Auto filter selection & indicator 2. Auto collimation size selection & indicator
3. Auto collimation menu(Option)
Tube head with manual tube rotation and vertical motorized movement
Tube head with motorized tube rotation and vertical motorized movement
System Control menu( Tube head with manual tube rotation Type)
* Vertical synchronization and Table Bucky tracking is available when this option is applied.
1. Auto filter selection & indicator 2. Auto collimation size selection & indicator
3. Auto collimation menu(Option)
Three safety sensors are located at the bottom of tube arm support. If any
stopped.
Patient Info.
Screen transition by touching the Collimator size & Additional Filter Icon
PAGE Fucntion
SYSTE
M PAGE
X-RAY
PAGE
1 Cassette Tray
2 Tray handle
3 Overhead handgrip
4 Chest handgrip
5 Stand column
6 Bucky cover
7 Stand base
9 Control Box
1 Rotating tray
2 Tray handle
4 Detector stretcher
5 Overhead handgrip
6 Chest handgrip
7 Stand column
8 Bucky cover
9 Stand base
11 Control Box
2 Tray handle
4 Overhead handgrip
5 Chest handgrip
6 Stand column
7 Bucky cover
8 Stand base
10 Control Box
3 Imaging Area
4 Mandible rest
5 Vertical stand
8 Chest Handgrip(option)
9 Overhead Handgrip(option)
10 Controls box
11 Safety Sensor
12 Emergency switch
Vertical movement
Regarding the rotating tray, make sure that the tray is set in the correct
position before insert the tray into Bucky.
In case of the R-type (for 14x17inch D-EVO) rotating tray’s operation is shown as below.
Lift up the lever (1) and hold the block (2) and then rotate the tray when it is needed.
The lever (3) is used for fix the panel after inserting it.
In case of the RD-type (for 14x17inch D-EVO) rotating tray’s operation is shown as below.
Do not pull out the tray roughly in the state DR cassette is inserted.
1. Lift up the tray rotating lever and begin to rotate the tray to clockwise direction while
holding the position 2 as above.
2. Release the tray rotating lever.
3. Keep rotate the tray until the tray is locked at horizontal top reference position
1. Push down the tray lever and begin to rotate the tray to counter-clockwise direction
while holding the position 2 as above.
2. Release the tray lever.
3. Keep rotate the tray until the tray is locked at vertical position
1. Push down the tray lever and begin to rotate the tray to clockwise direction while
holding the position 2 as above.
2. Release the tray rotating lever.
3. Keep rotate the tray until the tray is locked at horizontal center reference position
In case of the ND-type (for 17x17inch D-EVO) tray’s operation is shown as below.
In case of the RD2-type (for 14x17inch D-EVO2) rotating tray’s operation is shown as below.
1. Lift up the top and bottom down lever then panel will be slightly drawn out by itself.
2. Draw out the panel while holding position 2 of picture above.
1. Lift up the tray rotating lever and begin to rotate the tray to clockwise direction while
holding the position 2 as above.
2. Once the tray is rotated, release the tray rotating lever.
3. Keep rotate the tray until the tray is locked at horizontal top reference position
1. Get down the tray rotating lever and begin to rotate the tray to counter-clockwise
direction while holding the position circled as above.
2. Once the tray is rotated, release the tray rotating lever.
3. Keep rotate the tray until the tray is locked at vertical position
1. Get down the tray rotating lever and begin to rotate the tray to clockwise direction
while holding the position 2 as above.
2. Once the tray is rotated, release the tray rotating lever.
3. Keep rotate the tray until the tray is locked at horizontal center reference position
In case of the ND2-type (for 17x17inch D-EVO2) tray’s operation is shown as below.
1. Lift up the end of lever then the panel will be drawn out slightly by itself.
2. Draw out the panel.
• In case of the N-type (for all type of cassette or panel) tray’s operation is shown as below.
1. Rotate the lever to CW direction in order to release the holders and then widen the space between
holders.
2. Draw out the panel.
For the vertical movement, switches as below will be provided at the each rear side of the Bucky.
1 Cassette Tray
2 Tray handle
3 Overhead handgrip
4 Chest handgrip
5 Stand column
6 Bucky cover
7 Stand base
1. Vertical movement
1) Hold the chest handgrip
2) Push foot switch 1
3) Move the vertical position
4) Release foot switch 1
2. Tilting
1) Hold the chest handgrip
2) Push foot switch 2
3) Move the tilting position
4) Release foot switch 2
Regarding the tray operation, please refer to the previous section of WBS.
Regarding the rotating tray, make sure that the tray is set in the correct
position before insert the tray into Bucky.
3 Imaging Area
4 Mandible rest
5 Vertical stand
8 Control panel
9 Emergency switch
10 Overhead Handgrip(option)
11 Safety Sensors
12 Remote controller
Regarding the tray operation, please refer to the previous section of WBS.
Regarding the rotating tray, make sure that the tray is set in the correct
position before insert the tray into Bucky.
Control panel
Tilting motorized control buttons will work as follows when the safety sensor has detected
the obstacle.
- If button is pressed, tilting will stop and there will be an alarm sound.
- If button is pressed, tilting will be done and there will be an alarm sound.
Remote controller (Option)
All switches of control panel and remote controller are working by Dead-
man type. So the intended operation will be done only while its switch is
pushed. If not, the operation will be automatically stopped
Drawing out the cassette tray will enable the longitudinal movement of Bucky.
And this will turn on the laser transverse center line if the laser option is applied.
The laser will keep turn on if the tray is drawn out.
Full drawing in the cassette tray will lock the movement of Bucky and turn off
the laser.
While the tray is drawn in the Bucky, the laser can be controlled by the laser
button on the tube head controller.
1 Tabletop
2 Cassette tray
Inserting the foot inside the bottom of the table will release the electromagnetic
brakes of tabletop so the tabletop can be moved by hand.
1 Tabletop
2 Cassette Tray
(Drawing out tray off the tray lock)
3 Emergency switch
5 Elevating up pedal
7 Control box
For the safety, all pedals of PBT-6 will work when pushed twice in 2 seconds.
After table start works by pedals, operator can operate table by only one
pushing pedals within 5 seconds.
If the pedal operation is stopped over 5 seconds, double stepping is required
for pedal operation.
Regarding the rotating tray, make sure that the tray is set in the correct
position before insert the tray into Bucky.
In case of the R-type (for 14x17inch D-EVO) rotating tray’s operation is shown as below.
Lift up the lever (1) and hold the block (2) and then rotate the tray when it is needed.
In case of the RD-type (for 14x17inch D-EVO) rotating tray’s operation is shown as below.
Lift up the rotating lever and hold the lever block, and then rotate the tray.
In case of the ND-type (for 17x17inch D-EVO) tray’s operation is shown as below.
In case of the RD2-type (for 14x17inch D-EVO2) rotating tray’s operation is shown as below.
In case of the ND2-type (for 17x17inch D-EVO2) tray’s operation is shown as below.
In case of the N-type (for all type of cassette or panel) tray’s operation is shown as below.
1. Rotate the lever to CW direction in order to release the holders and then widen the space between
holders.
2. Draw out the panel.
1 Tabletop
2 Caster
4.9 COLLIMATOR
1. R 108
1 Lateral X-ray coverage control knob
3. DXC-RML, DXC-RMH
1 Lateral X-ray coverage control knob
Exposure table may be photocopied as required and placed in a suitable location as per local requirements.
Time mA Selected
(ms) 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
1.0 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5
1.2 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64
1.6 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8
2 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0
2.5 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25
3.2 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6
4 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0
5 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5
6.4 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2
8 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0
10 0.1 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0
12.5 0.125 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4
16 0.16 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8
20 0.2 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10
25 0.25 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5
32 0.32 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16
40 0.4 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20
50 0.5 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25
64 0.64 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32
80 0.8 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40
100 1.0 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40 50
125 1.25 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40 50 64
160 1.6 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40 50 64 80
200 2.0 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40 50 64 80 100
250 2.5 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40 50 64 80 100 125
320 3.2 4.0 5.0 6.4 8 10 12.5 16 20 25 32 40 50 64 80 100 125 160
Time mA Selected
(ms) 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500 640 800 1000
400 4 5 6.4 8 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400
500 5 6.4 8 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
640 6.4 8 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
800 8 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
1000 10 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
1250 12.5 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
1600 16 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
2000 20 25 32 40 50 64 80 100 125 160 200 250 320 400 500
2500 25 32 40 50 64 80 100 125 160 200 250 320 400 500
3200 32 40 50 64 80 100 125 160 200 250 320 400 500
4000 40 50 64 80 100 125 160 200 250 320 400 500
5000 50 64 80 100 125 160 200 250 320 400 500
6400 64 80 100 125 160 200 250 320 400 500
8000 80 100 125 160 200 250 320 400 500
10000 100 125 160 200 250 320 400 500
500mA is only available for output rating of x-ray generator from 40kW.
640mA is only available for output rating of x-ray generator from 52kW.
800mA is only available for output rating of x-ray generator from 68kW.
1000mA is only available for output rating of x-ray generator from 82kW.
APPENDIX B. XCONF
B1. PURPOSE
This section contains instructions on how to use XconF software.
B2. INTRODUCTION
XconF is an x-ray console software for operating FGXR x-ray systems. XconF in conjunction with Console
Advance software provides the operator an integrated console for radiography (refer to the figure below).
- When operator selects a menu in a study, the x-ray console automatically sets the exposure condition
(APR function)
- When operator exposes an image, exposure results (kV, mA, mSec, mAs, DAP) returned by the x-ray
system are displayed in the image annotations.
- FGXR Basic
patient
size
button
enabled
patient
size
button
disabled
patient
size
button
enabled
patient
size
button
disabled
- FGXR TS-CSA
patient
size
button
enabled
patient
size
button
disabled
- FGXR TS-CSP
patient
size
button
enabled
patient
size
button
disabled
Power Indicator
Generator is off.
Generator is booting.
Error Indicator
There is an error.
(flashing)
Warning Indicator
There is a warning.
(flashing)
X-ray Indicator
X-ray is off.
X-ray is on.
HU Level Indicator
Fluoro mode
Status Information
X-ray is on.
Warning indication
Error indication
Charge level.
Bucky Buttons
Non Bucky
Bucky 1
Bucky 2
Reset Button
Clears error.
DR Button
Shot synchronization is enabled. Generator will x-ray expose when when the
detector gives Shot Go signal.
Selects no filter.
This button opens the stitch setting window for setting the long view exposure
range.
Press and hold for 1 second for auto positioning according to menu selection.
Note: The button image reflects the exam type of selected menu.
(normal / stitch)
kV Setting
Control KV setting.
mA Setting
Control mA setting.
Time Setting
Density Setting
mAs Setting
Patient Size
Info Button
(4) Modify the exposure parameters (kV, mA, mSec, and AEC) if adjustment is necessary.
If the x-ray irradiation time set in the x-ray high voltage generator is changed,
the change is not reflected in the mode setting of the detector as the change
is not fed back to the detector before exposure. If the X-ray irradiation time
is changed to be longer than the charge accumulation time specified in the
MC, the irradiation is terminated upon completion of accumulation, and
images with insufficient dose may result. Therefore, do not set x-ray
irradiation time to be longer than the charge accumulation time specified in
MC.
(5) Move tube to appropriate position for the exam type. If the system type is TS-CSP, press and hold
button for 1 second to move the tube to preset position by motorized movement.
(6) Position the patient at the wall-stand or table depending on the exam type.
(9) Repeat steps 3~9 until all images for the study are acquired.
(2) Modify the exposure parameters (kV, mA, mSec, and AEC) if adjustment is necessary.
The same exposure parameters (kV, mA, mSec, and AEC) will be used for
all exposures in the series.
If the x-ray irradiation time set in the x-ray high voltage generator is changed,
the change is not reflected in the mode setting of the detector as the change
is not fed back to the detector before exposure. If the X-ray irradiation time
is changed to be longer than the charge accumulation time specified in the
MC, the irradiation is terminated upon completion of accumulation, and
images with insufficient dose may result. Therefore, do not set x-ray
irradiation time to be longer than the charge accumulation time specified in
MC.
(3) Move the tube to an appropriate position for long-view exposure at the wall-stand. If the system
type is TS-CSP, press and hold button for 1 second to move the tube to preset position by
motorized movement.
(4) Stitch setting window should open in both the PC monitor and THU touchscreen:
#1 If the WBS is tilted away from 0 degree, rotate it to 0 degree detent angle using the tilt control
Even when the display shows 0 degree, if the Bucky pitching angle slightly
deviates from the reference position, an error may occur at the time of
exposure. If an error that the angle of WBS is incorrect is displayed, please
rotate WBS to 0 degree again.
#1 Position the stitch-stand in front of WBS and lock the wheels in place.
If this is a long bone exam, place also a step platform in front of the stitch stand for the patient to
stand on.
#1 Adjust collimation in the cross dimension to desired size using the collimator control knob.
#1 Rotate the tube so that the beam center is at the upper bound and then press the [SET] button.
#2 Rotate the tube so that the beam center is at the lower bound and then press the [SET] button.
※ In case of TS-CSA, set the long-view range by moving the tube vertically up and down.
If either the upper bound or lower bound is outside of valid range, this will be
indicated by change of value’s color from green to red.
(12) The tube and WBS will move to the first position of exposure (~5 seconds).
Wait until the status bar shows READY and then press down on the exposure switch 1 and 2 to begin
serial exposure.
Observe the patient while performing the exam. If the exam needs to be stopped
at any time, release the hand switch. Releasing the hand switch will immediately
stop both the serial exposure and equipment movement.
The irradiation field is aligned to the top edge and the bottom edge of the imager
area for the first shot and the last shot, respectively, so as to maximize the long-
view coverage given the range of movement of the bucky apparatus. Refer to the
diagrams below for illustration.
(14) When auto-stitch exam is complete, the ready indicator in x-ray console will turn off.
(16) Tube and WBS move back to default position (~5 seconds). While the equipment is in motion, it
will sound beeps periodically as a caution.