1 x 1000 mL
DIFF-C5 14050200

Note:
Intended Use
Diff C5 is intended for cleaning the tubing’s of Mispa Count plus -5 part Differential 1. Make sure the tubing’s are clean and dust free before putting into the reagent.
Hematology Analyzer. It is an azide free filtered solution. Ensure that the tubing’s cap is tighten properly with reagent bottle.
Summary 2. Connect the Diff C5 bottle to the blue color tagged tube in the inside the left
This azide-free, formaldehyde-free, biodegradable cleaner contains a proteolytic hand side flap of Mispa count instrument.
enzyme and detergents that prevent and help remove protein buildup.
3. While changing the reagent, clean the tubing’s thoroughly.
Principle 4. Barcode reading details
Plasma or cellular proteins flow through apertures and may accumulate, form a
film that narrows or blocks aperture openings. Various biological debris (proteins, • Connect Diff C5 to Mispa count through the Blue colour tagged tube kept
membranes debris, lipids) also build up and deposit on measuring baths surface inside the instrument.
and throughout fluidic lines. The cleaning reagent dissolves proteins and helps
remove debris build up to clean apertures, baths, and fluidic lines. • Click Reagent icon in the main screen.
Reagent Composition
• Click Cleaner icon and scan two rows of barcode in the label. And ensure for
Boric Acid 0.3 % Beep sound after scanning each row of barcode.
Sodium Chloride 4.5 % • Click icon in the screen.
Sodium Sulphate 6.0 % • If barcode is accepted showing Question “Save changes”? ok or cancel.
EDTA 0.2 % • If barcode is not accepted or repeatation of same barcode, a warning message
CWT 3030 0.005 % “Edition is inconsistent, reagent lot number is already used” will be displayed.
Triton 0.25 %
Proteolytic Enzyme 0.05 - 0.1 %
• Check and ensure that the reagent barcode is smudge free.
• Retry scanning barcode and if not succeeded contact customer support.
Risk & Safety
Precautions and Warnings
Material Safety data sheets (MSDS) will be provided on request
1. To avoid contamination, close the reagent can properly. Avoid direct exposure
Reagent Storage and Stability of reagent to light.
The sealed reagents are stable up to the expiry date stated on the label, when
0 0 2. Ensure the system is installed and performing in a clean dust free environment
stored at 15 C to 30 C.
3. For in vitro diagnostic use only.
Open Vial Stability 4. Do not pipette by mouth.
Once opened, the reagent is stable up to 120 days, if contamination is avoided.
5. Do not inhale the reagent.
Reagent Deterioration 6. If the reagent comes in contact with eye or skin, wash off immediately with
Turbidity or precipitation in the reagent indicates deterioration and the reagent
large amount of water, and seek medical attention immediately.
must be discarded. Sample should be retested using fresh reagent.
7. Do not use the reagent in any procedures other than those described herein.
Direction for Use 8. Do not use containers and other materials in the package for any purpose other
Follow the procedures given in the operation manual of Mispa Count plus 5 part
than those described here in.
differential Hematology Analyzer.
9. When discarding the reagent dispose them off according to local regulations.
10. Take all precautions required for hand ling all laboratory reagents while using
the reagent.

‘Agappe Hills’, Dist. Ernakulam, Kerala, India-683 562.


Tel. +91 484 2867 000 | Customer Support No.: 1800 425 7151(Toll Free) ISO 9001:2015
customersupport@agappe.in | www.agappe.com REV. NO.: ADL/IFU/DC5/HAE/R02 EN ISO 13485:2016