Equipment information
EQUIPMENT NAME:________________ASSET/INVENTORY#: _________________ Manufacturer: __________________
Model: ___________ Serial number: __________________ Location:____________
Test information
ENGINEER/Technician: ____________________Date Started: __________________
Test type: PPM Post repair
Test equipment needed: ESA615 Electrical Safety Analyzer (or equivalent)
Impulse 7000 Defibrillator Analyzer (or equivalent)
Ohmmeter (can be a multi-meter, such as the Fluke 73 series digital multi-meter)
Cables and connectors to connect defibrillator to analyzer
Test Result
QUANTITATIVE TASK
Pass Fail N/A
Physical condition
< 100 µA NC
Chassis leakage < 500 A SPC
< 100 µA B and BF
Patient leakage current < 10 µA CF
Patient lead leakage current – isolation test (mains on patient < 100 µA BF
applied part) < 10 µA CF
Insulation test (optional) 500 V <2Ω
Verify electrodes, gel and paddles are stored with the defibrillator and are within expiration
dates
Verify proper time and date, correct if necessary
BIOMEDICAL ENGINEERING DEPARTMENT
PLANNED PREVENTIVE MAINTENANCE (PPM)
DEFIBRILLATOR INSPECTION FORM
Test Result
Performance testing
Pass Fail N/A
QUANTITATIVE TASK
Verify unit operates on battery
Output accuracy ± 15 %
Output accuracy, multiple loads ± 15 % (each delivery)
Output energy at maximum setting for 10 charge cycles (battery
± 15 %
power)
Charge time after 10 discharge cycles (battery power) ≤ 15 sec
Energy after 60 sec of full charge (battery power) ≥ 85 %
Internal discharge function
NAME : .
POSITION/DESIGNATION : .
SIGNATURE : .