INGM NAN - 6430-722 - 20230412 - TME-0-22759942 - Megger S1-5001

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

AC-1342

Certificate of Calibration
Manufacturer: Megger Report #: TME-0-22759942
Description: Insulation Tester P.O. #: N/A

Model #: S1-5001 Procedure: CPE053 (A)


Serial #: 6430-722 Asset #: NAN

Calibration Lab.: Mexicali, BC


Customer: INGENIUM Ingenieria y Servicios Industriales
Av. Montes Kenia #3857,
Mexicali, BC 21376

This certifies that the results presented relate only to the above instrument which was calibrated in compliance with the
Calibration Requirements of ISO /IEC 17025:2017 in accordance with referenced procedures. Standards used in this
calibration are traceable to SI units; their traceability derives from a National Metrology Institute such as NIST, CENAM,
NPL, DIN, from natural physical constants, consensus standards or derived by the ratio type of calibrations. Collective
uncertainties are determined as required with a distribution that corresponds to a probability of approximately 95% (k=2).
Unless otherwise noted calibrations are performed to manufacturer's specifications. Compliance statements are in
conformance with ILAC-G8:2019 simple acceptance decision rule. This form shall not be reproduced, except in full, without
the expressed written consent of Techmaster.

Test Result: PASSED Calibration Cycle: 24 Months


Temperature: 20ºC Calibration Date: Apr/12/2023
Relative Humidity: 38% Next Due Date: Apr/12/2025
Calibration Location: In-Laboratory Technician: Francisco Paredes

Standard # Manufacturer Model # Due Date Test Report #


137-0532 Cropico RH9A Oct 13, 2024 TDM-0-22636426
TJ-HVM-001 Kikusui 149-10A Oct 10, 2023 TDM-0-22635053

Special Remarks or Requirements:

A. Lopez
Ernesto Matamoros Ing. Mauricio Garayzar Abraham Lopez
Quality Assurance Technical Manager Certified By

Tijuana, B.C. Calle Seminario No. 8610-11, Col. Niños Héroes Este, CP 22120, Tijuana, B.C. México Tels.: (664) 624-4444 / 624-3128 / 624-4445 / 624-0440

Mexicali, B.C. Blvd. Lázaro Cárdenas 543 F03, Ex ejido Coahuila, C.P. 21360, Mexicali, B.C. Tels.:(686) 555-1660 / 557-6117 / 555-1766

Cd. Juárez,Chih. Ave. Antonio J. Bermudez No 2151-5 Fracc. Parque Ind J.Bermudez, C.P32470 Cd. Juarez Chih. Tels.:(656) 687-2471 / 648-1181

Monterrey, N.L. Ave. Morones Prieto No. 914, Oriente Int. 112, Col. La Huerta, C.P. 67144. Tels.:(81)1334-0701 / 1334-0702

Queretaro, QRO. Ctra. Federal 57,Mex-Qro Lateral Norte Km 210+100,Advance Business Park Conin, int.26 CP.76240,Col. Cumbres de Conin, Tels.:(442)241 51 04
Issued on: 2023-04-13 07:52:28.1970000 -07:00 540.1 2304

Page 1 of 2
TME-0-22759942 4
Calibration Results
Manufacturer: Megger Asset No. NAN

Description: Insulation Tester Serial No. 6430-722

Model #: S1-5001 Report No. TME-0-22759942

Step # Function Tested / Nominal Value As Found ± MU* In Tol As Left ± MU* In Tol Calibration Tolerance
1 Volts DC 510.0 V 506.0 ± 6.4 YES 506.0 ± 6.4 YES 489.6 to 530.4 V
2 1022.0 V 1010.0 ± 9.4 YES 1010.0 ± 9.4 YES 981.2 to 1062.8 V
3 2563.0 V 2523.00 ± 18 YES 2523.00 ± 18 YES 2460.5 to 2665.5 V
4 5052.0 V 5042.00 ± 33 YES 5042.00 ± 33 YES 4850.0 to 5254.0 V
5 Resistencia 500V 1.000 MOhm 1.0000 ± 0.0013 YES 1.0000 ± 0.0013 YES 0.950 to 1.050 MOhm
6 10.00 MOhm 9.950 ± 0.013 YES 9.950 ± 0.013 YES 9.50 to 10.50 MOhm
7 100.0 MOhm 99.0 ± 1.2 YES 99.0 ± 1.2 YES 95.0 to 105.0 MOhm
8 1.000 GOhm 0.985 ± 0.012 YES 0.985 ± 0.012 YES 0.950 to 1.050 GOhm
9 Resistencia 1000V 1.000 MOhm 0.9800 ± 0.0013 YES 0.9800 ± 0.0013 YES 0.950 to 1.050 MOhm
10 10.00 MOhm 9.950 ± 0.013 YES 9.950 ± 0.013 YES 9.50 to 10.50 MOhm
11 100.0 MOhm 99.0 ± 1.2 YES 99.0 ± 1.2 YES 95.0 to 105.0 MOhm
12 1.000 GOhm 0.985 ± 0.012 YES 0.985 ± 0.012 YES 0.950 to 1.050 GOhm
13 10.00 GOhm 9.85 ± 0.12 YES 9.85 ± 0.12 YES 9.50 to 10.50 GOhm
14 Resistencia 2.5 kV 100.0 MOhm 99.0 ± 1.2 YES 99.0 ± 1.2 YES 95.0 to 105.0 MOhm
15 1.000 GOhm 0.985 ± 0.012 YES 0.985 ± 0.012 YES 0.950 to 1.050 GOhm
16 10.00 GOhm 9.85 ± 0.12 YES 9.85 ± 0.12 YES 9.50 to 10.50 GOhm
17 100.0 GOhm 101.0 ± 1.2 YES 101.0 ± 1.2 YES 95.0 to 105.0 GOhm
18 Resistencia 5 kV 100.0 MOhm 99.0 ± 1.2 YES 99.0 ± 1.2 YES 95.0 to 105.0 MOhm
19 1.000 GOhm 0.990 ± 0.012 YES 0.990 ± 0.012 YES 0.950 to 1.050 GOhm
20 10.00 GOhm 9.85 ± 0.12 YES 9.85 ± 0.12 YES 9.50 to 10.50 GOhm
21 100.0 GOhm 101.0 ± 1.2 YES 101.0 ± 1.2 YES 95.0 to 105.0 GOhm

*MU stands for Measurement Uncertainty at a 95% confidence level.

*********************************End of Report***********************************

CPE053 (A)
540.6 1503

Page 2 of 2
TME-0-22759942 4