Date : 01/01/2018

Mr. R. K. Das
Principal Scientific Officer
STATE DRUGS TESTING & RESEARCH LABORATOR
NEAR SAINIK SCHOOL, GAJAPATI NAGAR
751017 BHUBANESWAR
ORISSA INDIA

Sub : Pre –requisite for installation of AAS, Model: iCE3500 with accessory Vapour Generator: VP-100 and Graphite
Furnace: GFS35Z.

Ref. : Your Purchase Order No: 12171/PROC.-EQP./OSMC/ER-16231/2017 Dated: 25/11/2017

Dear Sir,

Thank you very much for the aforesaid order awarded to us. You are also aware that for installation of this
sophisticated instrument pre-requisite is very much essential. So, for your ready reference pre-installation
requirement is enclosed herewith. Please go through the detailed site preparation guide.

You are requested to inform us after readiness of the pre-requisite so that we can plan our visit for installation at the
earliest.

We trust that the enclosure will be help to you. However, if you require any further assistance from us, please write
to us without any hesitation.

Thanking you and assuring our best services and sincere attention at all times.

With best regards,

For Thermo Fisher Scientific India Pvt. Ltd.

Krishanu Pal
Customer Care Engineer
Pre-Installation requirements for AAS System

Instrument: Atomic Absorption Spectrometer (AAS): iCE 3000 series AA System

Computer Specification:
Windows XP Professional with service pack 2 or Windows7 professional 32 bit/ 64 bit, 500 GB Hard drive, DVD-ROM,
Dedicated mother board in-build RS232 port, USB connection minimum 4 nos., 2 GB RAM, 19” VGA Monitor.

Room Condition:
AC must require. Lab temperature must be 22-25 ˚C. The temperature should not change by more than 2 ˚C per
hour. Atmospheric humidity should be 20 to 60% Rh for an ambient temperature between 22 and 25°C.
Atmospheric conditions must be non-condensing. Make sure that the working surface is sufficiently rigid to prevent
vibration as this may affect the optical alignment of the spectrometer and accessories.

Power required for AAS system with accessories:


iCE 3500: The spectrometer will require single phase electrical supply of 230V AC, 300VA, 50Hz.
GF 35: The Furnace will required single phase main supply 230V AC, 7.2 kVA, 50Hz.
GF 35Z: If Zeeman Furnace is used. Then required additional single phase main supply 230V AC, 1.5 kVA, 50Hz.
So for GF35Z total power required 8.7kVA power.
VP 100: Vapour Generator will require single phase electrical supply of 230V AC, 125VA, 50Hz.
ID 100: Inline dilutor will require single phase electrical supply of 230V AC, 100 VA, 50Hz.
EC 100: Electrical Heater for atomization cell will require single phase electrical supply of 230V AC, 720 VA, 50Hz.
ASX-520/560: Autosampler will require single phase electrical supply of 230V AC, 200 VA, 50Hz.

UPS Requirement:
For AAS without Graphite Furnace: It is advisable to use UPS system for instrument with accessories. The kVA rating
minimum 3 kVA or above double conversion IGBT based True Online UPS with built-in Isolation Transformer on the
Inverter output (Single phase in put & single phase out put). Double pole 64 amp MCB is required for input to the
UPS. Minimum 30 minute backup.

For AAS with Zeeman Graphite Furnace: It is advisable to use UPS system for instrument with accessories. The kVA
rating minimum 15 kVA of 3 phase double conversion IGBT based True Online UPS with built-in Isolation
Transformer on the Inverter output (three phase input & single phase output). Double pole 64 amp MCB is required
for input to the UPS. Minimum 30 minute backup.
Power distribution point’s requirement (UPS output supply single phase through double pole 64 amp MCB):

For AAS 15/5 A combined 1 no


VP100 15/5 A combined 1 no
GF 35 32A Double pole MCB 1 no
GF 35Z 32A Double pole MCB 1 no
For GF 35Zeeman 15/5 A combined 1 no
Air compressor 15/5 A combined 1 no
Exhaust Hood 15/5 A combined 1 no
Nitrous oxide heater 15/5 A combined 1 no
Computer & printer 15/5 A combined 3 nos
Additional point 15/5 A combined 2 nos
Make sure that each electrical outlet has an effective earth/ground (voltage between neutral & earth
should be less than 2V AC) connection capable of carrying 30A.

Location:
The instruments are designed for use on a normal laboratory bench. Ideally the instrument should be placed on a
moveable bench with 0.5 meters of access behind the instrument is mandatory.

Instrument models and Accessory dimensions:


Spectrometer Width Depth Height Weight
iCE 3500 780 mm 577 mm 520 mm 54 kg
(30.1 in) (22.7 in) (20.5 in) (119 lbs)
iCE 3400 780 mm 577 mm 520 mm 54 kg
(30.1 in) (22.7 in) (20.5 in) (119 lbs)
iCE 3300 567 mm 589 mm 510 mm 45 kg
(22.3 in) (23.2 in) (20.1 in) (99 lbs)
iCE 3300GF 567 mm 589 mm 510 mm 38 kg
(22.3 in) (23.2 in) (20.1 in) (84 lbs)
Accessory:
Accessory Width Depth Height Weight
Furnace 265 mm 495 mm 390 mm 50 kg
(GF33 or GF35) (10.4 in) (19.5 in) (15.4 in) (110.2 lbs)
Zeeman Furnace 265 mm 495 mm 390 mm 66 kg
(GF 35Z) (10.4 in) (19.5 in) (15.4 in) (145.5 lbs)
Furnace 269 mm 320 mm 380 mm 7.3 kg
Autosampler (11.7 in) (12.6 in) (15 in) (16.1 lbs)
EC100 305 mm 500 mm 215 mm 16 kg
(12 in) (19.7 in) (8.5 in) (35.5 lbs)
VP100 340 mm 360 mm 280 mm 8.5 kg
(13.4 in) (14.2 in) (11 in) (18.7 lbs)
ID100 200 mm 200 mm 170 mm 3.1 kg
(7.9 in) (7.9 in) (6.7 in) (6.8 lbs)
Autosampler 330 mm 508 mm 610 mm 8.4 kg
CETAC (13 in) (20 in) (24 in) (18.5 lbs)
ASX-260
Autosampler 520 mm 482 mm 610 mm 10.5 kg
CETAC (20.5 in) (19 in) (24 in) (23 lbs)
ASX-520
Autosampler 520 mm 482 mm 610 mm 10.5 kg
CETAC (20.5 in) (19 in) (24 in) (23 lbs)
ASX-560

Table Size:
Standard Table is required for installation. Length: 6 foot, Width: 3 foot, Height: 2.5 foot. If Furnace
accessory is used need Length: 8 foot rest are same.
Gas Requirements:
1. Acetylene Gas: The gas purity at least 99.95%. Impurities for Sulphur and phosphorus of less than 15ppm,
and water of less than 100ppm. The balance of the gas should be inert gas. The instrument will be supplied
with acetylene gas tubing terminating in 3/8in BSP Left Hand tighten thread connections. Double stage
acetylene regulator with suitable fittings.
2. Nitrous Oxide Gas: Gas purity minimum 99.95%. Oxygen impurities of less than 150ppm; nitrogen less than
400ppm; carbon dioxide of less than 50ppm; oxides of nitrogen less than 2ppm; carbon monoxide less than
1ppm and water less than 15ppm. The instrument will be supplied with N2O gas tubing terminating in 3/8in
BSP. Required two stage regulator with pre heater.
3. Argon Gas: Need two stage Argon regulator. The gas purity at least 99.999%. The instrument will be supplied
adopter fitting terminating in 3/8in BSP. For Argon Gas line gas paneling consist of Moisture trap and 5
micron filter is required.
4. Air Compressor: Clean, dry and oil free from a compressor providing at least 20l/min at a pressure up to 4.41
bar (60psi) and passed through a filter/regulator unit incorporating a 5 micron filter.

Fume Extraction System:


It is recommended that the hood of the extraction system and ducting be made of stainless steel with a circular cross
section of at least 150mm (6in) for flame compartment and 80mm (3.2in) for furnace compartment. A centrifugal fan,
preferably with plastic coated metal blades, extracting at least 2800lmin-1 (100ft3min-1) must be fitted into ducting
connected to the hood over the flame. An air gap of 128mm (5in) to 230mm (9in) must be made between Extraction
system and Flame chamber.
Reagents requirement:
Standard solution required for AAS of 100ml 1000ppm as per lamps ordered.
Millique or triple distilled water should be used for standards and samples preparation. All the necessary standards
for analysis should be freshly prepared by the customer. For Vapour generation system concentrated hydrochloric
acid, sodium tetrahydroborate (III) (sodium borohydride, NaBH4), sodium hydroxide (NaOH) are required for
analysis.

Tools required:

Gas cylinder opener for acetylene gas as well as nitrous oxide gas 1no.
Adjustable spanner 0-30mm 1no.
Teflon tap 2nos.