{InstallationInstructs}{Production}{Carestream Health Inc.

}{Restricted}

Publication No. II4825-2


14OCT09
Supersedes II4825-2
15MAY07

Installation Instructions
for the
New Mini-MCB Board
for the
KODAK DIRECTVIEW CR 825/850 Systems
and
KODAK DIRECTVIEW CR Classic/Elite Systems
Service Codes: 4825, 5285, 5289, 5634
Important
• Qualified service personnel must install this modification.
• See the Service Portal for possible updates before doing this modification.
• When doing the procedures in this document, you must use safe work practices and wear the correct personal
protective equipment (for example, safety eyewear) according to your company’s standard operating
procedures.
• After the installation of the new Mini-MCB board, P6H3377, these procedures correct the following:
– The system displays “Error 14019”
– An “Invalid Software Version message” is displayed
– The Diagnostics screen displays “Model Not Recognized”

Restricted

© CARESTREAM HEALTH, INC. 2009


PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the
subject matter gained by Carestream Health, Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes
no warranty, express or implied, with respect to this information. Carestream Health shall not be
liable for any loss or damage, including consequential or special damages, resulting from any
use of this information, even if loss or damage is caused by Carestream Health’s negligence or
other fault.
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

The installation includes:

Part No. Description Quantity


II4825-2 Installation Instructions 1

2 14OCT09 – II4825-2
Installing the New Mini-MCB Board

Caution
Dangerous Voltage
[1] Does the system have Version 5 software installed?
Yes No
a. Back up the parameters. Click: Go to Step 2.
• Diagnostics
• Config/Backup/Restore
• Copy to PC Backup
b. Continue with Step 2.

[2] De-energize the computer and the CR scanner.


[3] De-energize the system.
[4] Install the new mini-MCB board.

Important
• You must do the following procedures for your system to recognize the new mini-MCB board.
• During this installation, when the Cisco Security Agent (CSA) screen appears, you must click OK.
[5] Is your installation:
CR 825 System CR 850 System CR Classic/Elite System
a. From the main menu, click: Go to Step 6. Go to Step 27.
• Service Functions
• Service Utilities
• Reboot to Desktop
b. Select Start>Run.
c. Type c:\services\CRUnknown.bat
d. Click OK.
e. Select Start>Shutdown>Restart>OK.
f. Continue with Step 6.

[6] From the main menu, click:


• Service Functions
• Diagnostics
[7] At the Model Not Recognized screen, click Service Logs/Software Download.
[8] Click Download to Circuit Boards or Software Download.
[9] Click Change Model.
[10] Select the correct model number for your system:
• CR 825M, or
• CR 850M

II4825-2 – 14OCT09 3
INSTALLATION INSTRUCTIONS

[11] Click:
• Back
• Exit
• Diagnostics
• Service Logs/Software Download
• Download to Circuit Boards

Important
If the list of model numbers only displays one type of CR System, you must run the CRUnknown.bat file again.
[12] Select the correct model number for your system again.
[13] Check that the correct model number displays at the top of the screen.

Important
• The graphic above is an example only. The values on the graphic above might not match what is on your screen.

[14] Click DOWNLOAD MCB BOOT.


[15] Click DOWNLOAD MCB PLD.
[16] Click DOWNLOAD MCB APPLICATION.

Important
After clicking DOWNLOAD MCB APPLICATION, you must click BACK to exit the Software Download screen to
allow the software on the MCB board to restore.
[17] Click:
• BACK
• NEXT

4 14OCT09 – II4825-2
[18] At the Software Download screen, click:
• DOWNLOAD MCB PARAMETERS
• DOWNLOAD MCPU APPLICATION

Caution
Dangerous Voltage
[19] De-energize the system.
[20] Energize the system.
[21] If necessary, at the message “Add Addonit: Creation of hard disk file complete”, click Yes.
[22] From the main menu, click:
• Service Functions
• Diagnostics
• Display Software Versions
[23] Check:
• System model number is correct
• System software version is correct
[24] Is the model number and software version correct?
Yes No
a. Click: a. Click:
• Back • Diagnostics
• Download to Circuit Boards • Download to Circuit Boards
• Slow Scan Profiles • CHANGE MODEL
• Download Slow Scan Profiles b. Select the correct model name for your system.
• Back c. Continue with Step 25.
• Shutdown
• Yes
b. Continue with Checking the Operation.

[25] Go to Step 24.

For Classic/Elite CR Systems


[26] Is the model number and software version correct?
Yes No
a. Click: a. Click:
• Done • Diagnostics
• Back • Software Download
• Main Menu • CHANGE MODEL
• Previous b. Select the correct model name for your system.
b. Continue with Step 27. c. Continue with Step 27.

[27] From the main menu, click:


• Service Functions
• Diagnostics
• Software Download

II4825-2 – 14OCT09 5
INSTALLATION INSTRUCTIONS

Important
• The graphic above is an example only. The values on the graphic above might not match what is on your screen.

[28] Are any of the Available Versions shown on the screen newer than the Current Version?
Yes No
Continue with Step 29. Go to Step 35.

[29] Select the check box for the software that must be changed.
[30] Click:
• START DOWNLOAD
• Current MCB Values - only if the message “Use the current MCB Board Values or use Default Values?”
displays
• DONE after the download is finished

Caution
Dangerous Voltage
[31] De-energize the system.
[32] Energize the system.
[33] Use your USB service jump drive to log on to the service menu.
[34] At the main menu, click:
• Service Functions
• Diagnostics
• Software Download

6 14OCT09 – II4825-2
[35] Make sure that the MCB current version of software is the same as the available version of software.
[36] Click:
• Diagnostics
• Config/Backup/Restore
• Restore from PC Backup
• Update Configuration on MCB BRD

Note
If an the PC Backup results in an error, you must do a recalibration of the MCB parameters.

Checking the Operation


[1] Check that the system operates correctly.

Publication History

Publication
Date Publication No. ECO No. Changed Pages File Name Notes
14OCT09 II4825-2 --- All ii4825_2.fm Revision
15MAY07 II4825-2 --- --- ii4825_2.fm New Publication

DIRECTVIEW is a trademark of Carestream Health, Inc.


The Kodak trademark and trade dress are used under license from Kodak.

Printed in U.S.A. • ii4825_2.fm

Carestream Health, Inc.


150 Verona Street
Rochester, NY 14608