D-Dimer Neo (Granule DB)

This is not a complete Instruction for use. For more detailed instructions, please refer to IFU. D-Dimer Neo (granule)_Rev00_221109E
Test Components

Ensure that the lot number of the components match.

Matches

Leave test components (if stored in the refrigerator) at


Test cartridge Detector tube (Granule) Detector diluent ID chip room temperature for at least 30 minutes before use. Just right Too little or Bubble In the
middleor near the top of the tip

Test Procedure

Draw 150 μL of Add it into Draw 10 μL. Add it into


1 Detector diluent.
2 Detector tube.
3 (Plasma / Whole blood / Control)
4 Detector tube.
5 Shake 15 times.

15 times
10 μL
10 μL
150 μL 150 μL

6 Draw 75 μL. 7 Load the sample mixture. 8 Wait 12 minutes. 9 Insert the test cartridge. 10 Press ‘Start’. 11 Read the test result .

Result

75 μL
75 μL 12 Min

For technical assistance; please contact: Boditech Med Inc. ‘ Technical Services ’ This reference guide is for the main steps’ procedure brief reminder only. Please check the full descriptions procedure and whether any changes may have been
Tel: +82-33-243-1400 Fax: +82-33-243-9373 E-mail: sales@boditech.co.kr implemented from the product insert. The company declines responsibility in case the instruction for use provided with the regents are not properly followed.
www.boditech.co.kr