Dental Unit Water Lines and their


Disinfection and Management:
A Review
Abstract: The perceived threat to public health from dental unit water line (DUWL) contamination comes from opportunistic and
respiratory pathogens such as Legionella spp, Nontuberculous Mycobacteria (NTM) and pseudomonads. These organisms can grow and
multiply in the DUWL biofilm to reach infective concentrations, with the potential for inhalation leading to respiratory infections or direct
contamination of surgical wounds. In this paper we discuss current legislation and practical methods for delivering water within the DUWL
that meets the standards for safety.
CPD/Clinical Relevance: Understanding the clinical relevance and methods for decontaminating DUWL is essential to create a safe
working environment in dentistry.
Dent Update 2017; 44: 284-292

It was first recognized in the 1960s that Northern Ireland for dental unit water line. Biofilms are defined as an organized
water sampled from the DUWL contained quality of 100−200 cfu/mL of aerobic community of microbes attached to a biotic
large numbers of organisms in the range heterotrophs at 22°C.3 Since then, our or an abiotic surface, and encased in an
104−106 colony forming units (cfu)/mL.1 understanding of the physiology of the extracellular polymeric substance.
Two decades later the American biofilm,4 incidence data on legionellae in The usual source of feed
Dental Association (ADA) set the goal of dental practice5,6and the first proven case water to the dental unit which nourishes
<200 cfu/mL of aerobic heterotrophs as of Legionella transmission and death of the luminal biofilm on DUWL is from the
the standard for dental unit water lines.2 a patient associated with contaminated potable municipal supply or, preferably,
This figure is reflected in the current DUWLs have transformed the management from an independent bottled water supply.3
of DUWLs.7 Challenges caused by Legionella After entering the unit proper, the water
recommendation for England, Wales and
contamination have not just affected passes through a multichannel control box
dentistry but, around the world, there have which distributes the water to dental unit
been major outbreaks of Legionnaires’ water lines (hoses) feeding the high-speed
disease with resultant deaths of many handpiece, the air/water syringe and the
Caroline L Pankhurst, BSc, BDS, ultrasonic scaler. DUWLs have a number
MSc, PhD, MRCPath(Specialist in Oral of those infected.8 Indeed, dentistry is
specifically included in the Approved Code of characteristics that promote biofilm
Microbiology), Senior Specialist Clinical
of Practice for Legionnaires’ disease control growth unless they are regularly cleaned
Teacher (part-time), King’s College
in water systems, which states that DUWL and disinfected. For instance, they have a
London Dental Institute, Crispian Scully,
management is viewed in the context of the very small bore but a high surface area, slow
CBE, DSc, DChD, DMed (HC), Dhc(multi),
dental practice’s overall hot and cold water flow rate and the whole column of water in
MD, PhD, PhD (HC), FMedSci, MDS, MRCS,
the DUWL can be stagnant for long periods
BSc, FDS RCS, FDS RCPS, FFD RCSI, FDS supply systems.9
of time. Biofilms develop rapidly in such
RCSEd, FRCPath, FHEA and Lakshman
systems within about 8 hours,6 forming a
Samaranayake, DSc, DDS, FRCPath, Biofilm formation in DUWLs mature bacterial community enveloped
FHKCPath, FDS RCS(Edin), FRACDS, FDS
The effluent organisms from in a protective matrix within 6 days that is
RCPS, Professor Emeritus, University of
DUWLs essentially originate from the capable of shedding 104−106 cfu/ml if left
Hong Kong, SAR China.
biofilms on the luminal surface of the water untreated.4
284 DentalUpdate April 2017
InfectionControl

Biofilms and risk to health percent of the dentists enrolled in the population, the highest ever observed,
Biofilms on the lumen walls study reported ‘ever having suffered’ from and suggested that this may be due to
of DUWLs act as a reservoir of micro- asthma either in childhood or as an adult.12 favourable meteorological conditions that
organisms. A wide range of species of Endotoxin derived from DUWL bacteria may supported the growth of legionellae.14.
environmental and oral bacteria, fungi, account for this increase in occupational Although, when viewed from a global
amoebae and protozoa have been isolated asthma, as dentists who were exposed in burden of disease perspective, the incidence
from DUWLs. The majority are harmless, their dental surgery to total aerobic counts of Legionellosis is very low, compared
so called saprophytic, Gram negative at 37°C at a concentration of 200 cfu/mL to other major respiratory infectious
environmental bacterial species. However, were more likely to report symptoms of diseases such as tuberculosis, which has
the lipopolysaccharide (LPS) cell wall of all asthma ‘since starting dentistry’.12 an estimated incidence of 10.5 per 100,000
Gram-negative bacteria, a potent source of of the population (5,758 cases) in 2015 in
non-infectious endotoxins, can evoke mild Legionella and respiratory the UK,15 and an estimated 10.4 million TB
or severe inflammatory responses leading to pathogens cases worldwide.16 However, regardless
diseases such as periodontitis and systemic, Legionellae can cause a variety of of the relatively small number of cases of
septic shock. respiratory infections (Figure1). Legionnaires’ Legionellosis, registered providers of dental
A high bacterial load in the disease, or Legionellosis, is a severe infection services have a duty of care to prevent and
DUWL equates with a high endotoxin which typically presents as pneumonia control the risks from contamination of
concentration. Endotoxin has been detected and symptoms may include a high fever, practices' water supplies with legionellae.
in dental unit water up to 50010−2,56011 chills, cough, muscle aches, headaches, The registered provider of dental services is
Endotoxin Units/mL compared to a mean and diarrhoea.13 Although healthy young legally obliged to take account and comply
level of 66 EU/mL found in water samples adults can develop Legionnaires’ disease with the standards as specified within ACOP
collected from adjacent sinks. Putnins et in two thirds of patients, there are L89 for water quality standards in healthcare
al11 have suggested that the endotoxin in other co-morbidities and risk factors.13 premises.
DUWLs might stimulate proinflammatory Legionnaires’ disease is a notifiable disease There is no reported person
cytokines in gingival tissue during surgery in England and Wales and is a relatively to person transmission of legionellae,
and adversely affect healing. Endotoxins are uncommon cause of pneumonia, with and infection is essentially associated
known to trigger occupational asthma in approximately 350−500 cases per annum with exposure to a contaminated water
susceptible people. An association between in the UK, the majority of which are due source. Legionellae are ubiquitous in
the onset of occupational asthma in dentists to Legionella pneumophila serogroup 1.13 warm, stagnant natural waters either free
and the concentration of bacteria in their European Legionnaires’ Disease Surveillance living or growing within amoebae.17 They
DUWLs has been demonstrated in a cross- Network in 2016 reported the notification proliferate in man-made water supplies
sectional multicentre DUWL survey of 265 rate of Legionnaires’ disease in the EU/ with similar properties, eg water cooling
dentists in general dental practice. Fourteen EEA in 2014 was 13.5 cases per million/ towers, humidifiers, hot tubs, birthing

Legionnaires' disease
severe, potentially fatal, atypical
form of pneumonia

Risk factors for legionellosis


Age > 50years
Gender: Male>female
Co-morbidities: smoking,
heart disease, diabetes,
renal and respiratory illness,
immunocompromised
Non-pneumonic
Season: summer months legionellosis
Pontiac fever
non-fatal, mild, self- Geographic location: London, West a urinary legionella antigen
limiting influenza-like Midlands positive case, with no
illness usually resolving in Travel abroad: eg Spain, China, symptoms of Pontiac fever
24−48 hours India, Thailand or pneumonia

Figure 1. The risk factors and three different clinical presentations of Legionellosis.

April 2017 DentalUpdate 285


InfectionControl

pools, hot and cold plumbing systems infection ensued.24 But it is the death of hot and cold water systems.9,25-27
and shower heads, etc.9 Legionellae an elderly lady in Italy following two visits Risk factors for legionella
favour growth temperatures in the to her dentist where she was exposed to contamination of dental water
range 20−45°C. Though clinical cases legionellae in the DUWL that subsequently systems
can occur in outbreaks associated with caused her pneumonia, the first proven
Practitioners should be aware
travel and as hospital acquired infection, case of dental-associated Legionnaires’
of the main risk factors for legionella
most cases are sporadic and community disease, that is the most significant.7 The
contamination as control of these hazards
acquired. Cases arise from exposure of the DUWLs were allegedly managed with a
determines the management of the water
person to inadequately managed water standard biocide (a hydrogen peroxide
supply. The type of questions to consider
supplies, including in the dental surgery. product), which has previously been shown
when formulating a water plan are listed in
When investigating epidemiological risk to be effective. In summary, although
Table 1.
factors associated with notified cases and the number of proven published cases of
If the answer to any of these
outbreaks of Legionnaires’ disease, Health infection resulting from exposure to water
questions in Table 1 is ‘yes’ for either the
Protection England include exposure to from DUWLs is limited, these incidences
practice’s plumbing or DUWL, as is the
‘dental treatment’ during the incubation are noteworthy cautionary tales that need
case in most dental practices, then these
period (2−14 days) as a pertinent risk factor attention to prevent further cases and
risks must be managed according to the
on their case assessment protocols.13 Two improve our management of dental water
regulations.9,3,27 Practices are under the
large studies conducted before the current supplies.
regulations specified in the ACOP-L8
regulations were introduced showed the
incidence of legionellae in DUWLs to be low and are required to employ the services
Legislation and regulations of a suitably qualified water engineer
(0.37% in London and Northern Ireland18
and 1.9% in the South-west of England19), Criterion 2 of The Health and (Competent Person) such as a member
but legionellae was also recovered from the Social Care Act 2008: code of practice on of the Legionella Control Association or
cold water supply of three dental practices.18 the prevention and control of infections similar.9,27 The Competent Person‘s role is to
Where legionellae are present in the water and related guidance (2015)25 sets out the provide professional specialized advice on
supply, raised legionella antibody titres can criteria against which a registered provider’s preparing the risk assessment, writing the
be found in the dental team, indicating compliance with the management of water management plan to control the risk,
occupational exposure.20,21 There is DUWLs and water supplies will be assessed maintaining the cleanliness of the plumbing
insufficient data to determine whether such by the Care Quality Commission (England system and boiler and training the staff
exposures lead to infection such as Pontiac and Wales) for compliance with the on taking sentinel water temperatures.
fever; but a death of a dentist was recorded law. Taking account published national An annual inspection by the engineer is
in the USA following exposure to an unusual guidance outlined in Approved Code of recommended. The dental practice staff
strain of legionella in his DUWL.22 Practice. Legionnaires’ Disease: The Control have the duty of recording and monitoring
Nontuberculous Mycobacteria of Legionella Bacteria in Water Systems, the water temperatures in the hot and cold
(NTM) and pseudomonads are other usually referred to as L8,9 HSG27426 and HTM water systems on a monthly, 6-monthly and
pathogens that pose a transmission risk 01-05,3 the practice should have a policy annual basis, as set out in the water plan.
via DUWLs. NTM found naturally in potable for preventing contamination of dental Records must be kept for 5 years.3,9 A brief
waters are amplified 400-fold in DUWL and unit water lines, including an appropriate résumé of the basic principles on managing
have been linked to two published cases of water supply and maintenance schedules. the risk from legionella in the practice’s hot
serious infection.23 Pseudomonas in dental Dental practices must, with the help of and cold plumbing are outlined in Tables
primary care water lines in the UK can often a Competent Person (Water Engineer3), 2 and 3. Readers are referred to HSG27426
be at high levels in untreated water lines.5 prepare a legionella risk assessment and and HTM 01-053 for a full description of the
In the UK, Walker et al found that the most written control scheme (water plan) for guidance and regulations. For readers in
common species detected in their study was Legionella prevention and control for their Scotland further information is also available
Pseudomonas species, isolated from 16% of
untreated water lines.19
 Is the water temperature in all or some parts of the system between 20−45°C?
Martin24 demonstrated that
 Is water stored or re-circulated as part of your system?
Pseudomonas aeruginosa recovered
 Are there sources of nutrients such as rust, sludge, scale, organic matter and biofilms?
from the dental abscesses in two
 Are the conditions likely to encourage bacteria to multiply?
immunocompromised patients treated in a
 Is it possible for water droplets to be produced and, if so, can they be dispersed over a
dental school were identical to those found
wide area, eg aerosols from handpieces?
in the DUWL of the dental units where they
 Is it likely that any of your employees, patients, visitors, etc are more susceptible to
were treated. In a follow-on prospective
infection due to age, illness and a weakened immune system?
study, 78 non-compromised patients treated
 Could they be exposed to any contaminated water droplets?
in one of six P. aeruginosa contaminated
dental units were transiently colonized
Table 1. Risk assessment questions for practice plumbing and DUWLs.
for 3−5 weeks with P. aeruginosa, but no
286 DentalUpdate April 2017
InfectionControl

from Scottish Health Technical Memorandum better control over the microbial quality recommended by the biocide (disinfectant)
04-0128 and the websites for Health of the water used in patient care by the manufacturer. Biocides employed to prevent
Protection Scotland and Scottish Water. A addition of biocides and other methods to the growth of legionellae in DUWLs must
discussion of the methods used to manage control contamination. In addition, they comply with bacteriocidal activity specified
the DUWL is outlined in the next section. act as a type A air-gap, a physical gap that in the standard BS EN 13623:2010 and
prevents back siphonage of contaminated before purchasing a DUWL biocide check in
Managing risk factors for water into the mains supply. Reservoir which the manufacturer states compliance
legionellae in DUWLs water bottles are recommended to be filled with the standard on the accompanying
with freshly produced (less than 12 hours data sheet.
Dental unit water is both
old9) reverse osmosis or distilled water. Furthermore, HSG274 (part 3)26
swallowed and inhaled by the patient.
These purified waters are not sterile but are recommends that ‘independent water storage
Therefore, the microbial quality of the water
unlikely to contain legionella, NTMs and bottles should be cleaned, rinsed with reverse
must take these factors into account. Health
pseudomonads found in potable tap water. osmosis or distilled eater, dried and stored dry
Technical Memorandum (HTM) 01-053
and inverted overnight’.
recommends that the bacterial count in
water sampled from the DUWLs should Recommendations on care of
have a bacterial total viable count (TVC) of water lines Microbiological measurements
≤100−200 cfu/mL, which is comparable with The dental unit is classified as a These are not mandatory under
drinking water standards in the European medical device under the Medical Devices the L8 ACOP9 or HTM 01-053 but are dictated
Union (≤100 cfu/mL) and USA (≤500 cfu/ Directive 2002 and equipment used in by the individual dental practice’s risk
mL). Generally, the water entering the DUWL practices must carry the European CE mark. assessment, highlighting a possible hazard.
contains very few organisms: 0−100 colony- It is well recognized that a combination Examples of possible hazards that would
forming units (CFU)/ml. However, water of methods are required to prevent trigger microbiological measurements
exiting the handpieces may contain up to contamination and maintain water quality include, amongst others for example,
10 000−1,000,000 cfu/mL. This is mainly in the DUWL.3 This goal can currently be technical factors causing an inability to
as a result of the organisms that are shed achieved using antiretraction and check comply with the regulations; after disruption
from the bacterial biofilms growing within valves, flushing, biocides and purified water in the service or physical changes noticed
untreated lines; from the independent in the reservoir bottle. Other methods are in the DUWL quality, such as foul taste or
bottled water supply and from suck-back also commercially available. odour and cloudiness of the water.3.Dip-stick
from the mouth due to negative pressure in methods are not suitable for estimating
the handpieces when drilling is stopped. The technical guidance HSG274 (part 3)26 of the legionella and water samples should be
Although water can be taken L8 ACOP9 processed by a UKAS accredited water
from the mains supply (potable water), to This recommends dental and environmental testing laboratory.3,18
achieve the specified water quality, dental practices to ‘Drain down, clean, flush and The Laboratory will advise on sample
units are usually fitted with a separate water disinfect all system components, pipework and collection methods but typically 1L of water
reservoir that is independent of the public bottles TWICE daily, eg at the start and finish is collected in a thiosulphate-coated bottle
water supply. This allows dentists to have of each day’. Disinfection contact time is as and transported directly to the laboratory
for legionella testing, and enumeration
of water-borne indicator organisms and
Regulations under Identify and assess sources of risk
aerobic colony counts obtained as specified
ACOP – L8 (risk assessment)
in national standard operating procedures.
Prepare a written scheme for preventing or controlling the Usually, the Laboratory will provide result
risk interpretation guidance.
(water plan)
Daily flushing to reduce stagnation
Implement, manage and monitor precautions
According to HTM 01-05, all
Keep records of the precautions DUWLs should be flushed for two minutes
(retain records for 5 years) at the beginning of each day, prior to
commencing treatment, and at the end of
Appoint a competent person with sufficient authority and the day.3 Staff should be alert to malodour,
knowledge of the installation to help take the measures cloudiness and bad taste imparted to the
needed to comply with the law water by microbial contamination, which
(water engineer to help in applying health and safety law, eg are particularly noticeable after periods of
HSG274; COSHH; provide staff training) stagnation.3 They signal that conditions
are appropriate to support the growth
Table 2. Legionella control in the dental practice − requirements that have legal status under the
of legionella. The practice should seek
Approved Code of Practice L8.
advice on microbial sampling for legionella
April 2017 DentalUpdate 289
InfectionControl

detection.3 Care should be taken by the a scaffold for biofilm formation and are peracetic acid formulations, Tetrasodium
operator to avoid splatter and aerosol very inefficient unless they are regularly EDTA, chlorhexidine formulations, iodine,
exposure during DUWL flushing and masks maintained and replaced periodically.30 quarternary ammoniums and chlorine
and eyewear should be donned. Irregularly Antiretraction valves are also located dioxide.31,32,33,34 Hypochlorite, popular
used dental units should be flushed on a in the dental handpieces. They prevent for disinfecting DUWLs in the past, led
routine basis (at least weekly) to prevent the re-aspiration of fluid contaminated to corrosion of handpieces and its use is
stagnation.3 with oral flora of the patient into the mainly confined to ‘shock treatment’ for
water line. Routine decontamination and the eradication of microbiologically proven
Flushing to reduce the risk of cross periodic handpiece maintenance helps in legionella contamination.7 For a detailed
contamination maintaining the integrity of the handpiece review of the adverse reactions reported for
This is a simple and efficient anti-retraction valves. a range of different biocides see Coleman
means of reducing the planktonic et al.31
(suspended) bacterial load in the water Biocides and disinfectant systems Continuously infused biocides
line due to oral suck-back. Thereby helping Thermal disinfection and used in the water line must be non-toxic
to prevent cross contamination from avoidance of temperature between and safe if imbibed or inhaled by the
a previous patient. DUWLs supplying 20−45°C, though highly effective as patient and staff. Commercial products for
handpieces, ultrasonic scalers and air/water a control measure against legionella continuous infusion are less concentrated
syringes should all be flushed. Although in the water distribution system, than those sold for intermittent purging of
flushing can reduce the numbers of bacteria is not applicable to DUWLs. The the system.24
in expelled water, the effect is transient and ambient temperature in the DUWL is
has no impact on the water line biofilm and approximately 18−23°C.5 Sterile water for surgical irrigation
may even increase the bacterial shedding Therefore, in DUWLs, we are Water from DUWLs should
momentarily as it disturbs the biofilm.29 reliant on biocides and disinfectants to never be used as an irrigant in procedures
Flushing also has the advantage of drawing remove, inactivate and prevent formation involving breaches of the mucosa and
up fresh biocide into the DUWL and thereby of biofilm. The disinfectant must be active bone exposure − instead use a sterile water
facilitating disinfection of the line. against the range of micro-organisms supply and delivery system or sterile saline.
found in the DUWL, including legionella
Antiretraction valves (check valves) spp. Biocides can either be continuously Point of use filters
All DUWLs should be fitted with infused into, or intermittently added Microbial point of use filters may
check valves (non-retractable devices), to, the dental unit water by varying be installed, for instance between the water
to prevent suck-back (backflow/back- technologies. A huge variety of products line and the dental handpiece. These have
siphonage) of contaminants. If the dental are available for use as intermittent purges no effect on the biofilm in the water lines,
unit takes water directly from the mains and continuous disinfection. A range of but will remove micro-organisms as the
supply, then the unit should be fitted with different disinfectants are deployed by water is delivered to the patient. Disposable
a type A air gap. However, it is now known manufacturers, eg alkaline or hydrogen filters must be replaced daily, which can be
that the antiretraction valves can become peroxide, hydrogen peroxide/silver ions, expensive.3

Hot water plumbing:*


 Avoid storage of water between 20−45°C
 Hot water should be stored at ≥60°C
 Aim is to achieve a hot water temperature of 55°C at all points of use within one minute of turning on the hot tap
 Maintain cleanliness of system, eg reduce scale and deposits
 Integral thermostatic mixer valves for hand wash basins set at 38−44°C to avoid scalding
 Flush infrequently used taps for several minutes on a weekly basis to reduce risks associated with stagnation and low water usage

Cold water plumbing:*


 Site cold water tanks in a cool place with thermal insulation
 Aim is to achieve cold water temperature of below 20°C after running the taps for two minutes
 Introduce measures to reduce stagnation (eg remove or shorten length of dead leg pipes, remove redundant taps)
 Maintain cleanliness of the system, eg hard water areas, softening of the cold water supply to the hot water distribution system should
be considered to reduce the risk of scale being deposited at the base of the calorifier and heating coils
 Keep the amount of water stored to a minimum, ie equivalent to no more than 12 hours of usage

Table 3. Basic principles for controlling growth of legionella in hot and cold water systems in the dental practice. *Refer to HSG274 (part 2) and HTM 01-05
(2013) for detailed guidance and regulations.

290 DentalUpdate April 2017


InfectionControl

Water purification system practices in the UK and around the world 12. Pankhurst CL, Coulter WA, Philpott-
Water purification systems have responded to the challenge of the Howard JN, Surman-Lee S, Warburton
treat the water coming into the dental unit DUWL goal set by the ADA in 1996 and are F, Challacombe S. Evaluation of the
(source water). They kill or remove micro- routinely treating their patients in a safe and potential risk of occupational asthma in
organisms by methods such as filtration, hygienic environment. dentists exposed to contaminated dental
electrolyzed water or ultraviolet light. One unit waterlines. Prim Dental Care 2005; 12:
advantage of these methods is that they 53−59.
may delay biofilm formation on water lines
References
13. Public Health England. Guidance on the
or synergize other treatment methods. 1. Blake G. The incidence and control of
Control and Prevention of Legionnaires’
As they are not discharged in the effluent infection in dental spray reservoirs.
Disease in England. Technical Paper
water, they are non-polluting and less likely Br Dent J 1963; 115: 412−416.
1 − Disease Surveillance. 2015 gateway
to select for antimicrobial resistant species 2. Anonymous. Infection control
number: 2015226.
than biocides used in DUWLs.35 recommendations for the dental office 14. European Centre for Disease Prevention
and the dental laboratory. J Am Dent and Control (ECDC). Surveillance Report:
Assoc 1996; 127: 672−680. Legionnaires’ Disease in Europe 2014.
Miscellaneous
3. Department of Health. Health ecdc.europa.eu/en/publications/
Other, rather expensive methods,
Technical Memorandum HTM 01-05: publications/legionnares-disease-
for delivery of quality water include the use
Decontamination in Primary Care europe-2014.pdf
of sterile water and autoclavable DUWL
Dental Practices 2013. www.gov. 15. Public Health England. Tuberculosis in
systems.
uk/government/publications/ England 2016 report. www.gov.uk/
decontamination-in-primary-care- government/uploads/system/uploads/
Boil-water advisories dental-practices attachment_data/file/564656/TB_
A ‘boil-water’ advisory notice is 4. Walker JT, Marsh PD. Microbial biofilm annual_report_2016.pdf
issued by authorities when the public water formation in DUWS and their control 16. World Health Organization. Tuberculosis
supply is likely to be contaminated with using disinfectants. J Dent 2007; 35: (TB): Global Tuberculosis Report 2016.
pathogenic organisms, eg Cryptosporidium, 721−730. www.who.int/tb/publications/global_
or the numbers of microbes in the public 5. Pankhurst CL, Coulter WA. Do report/en/
potable water supply system are above that contaminated dental unit water lines 17. Singh T, Coogan MM. Isolation of
which is compatible with health. A recent pose a risk of infection? J Dent 2007; 35: pathogenic Legionella species and
example occurred in Lancashire, UK during 212−220. Legionella-laden amoebae in dental
the summer of 2015. During such periods 6. Barbot V, Robert A, Rodier M, Imbert C. unit waterlines. J Hosp Infect 2005; 61:
the following apply: Update on infectious risks associated with 257−262.
 Do not use potable to feed the DUWL dental unit waterlines. FEMS Immunol 18. Pankhurst CL, Coulter WA, Philpott-
or other dental equipment. Only reverse Med Microbiol 2012; 65: 196−204. Howard JJ, Harrison T, Warburton F,
osmosis or distilled water should be 7. Ricci ML, Fontana S, Pinci F, Fiumana Platt S et al. Prevalence of Legionellae
used.3 E, Pedna MF, Farolfi P et al. Pneumonia waterline contamination and Legionella
 Do not use potable water for dental associated with a dental unit waterline. pneumophila antibodies in General
treatment, patient-rinsing or hand- Lancet 2012; 379(9816): 684. Dental Practitioners in London and Rural
washing. For the latter purpose, 8. Greig JE, Carnie JA, Tallis GF, Ryan NJ, Northern Ireland. Br Dent J 2003; 195:
antimicrobial-containing products Tan AG, Gordon IR et al. An outbreak of 591−594.
that do not require water can be used Legionnaires’ disease at the Melbourne 19. Walker JT, Bradshaw DJ, Bennett
(eg alcohol-based hand rubs). If hands Aquarium, April 2000: investigation and AM, Fulford MR, Martin MV, Marsh
are visibly contaminated, use bottled case–control studies. Med J Aust 2004; PD. Microbial biofilm formation and
water and soap for hand-washing or an 180: 566−572. contamination of dental- unit water
antiseptic hand towel. 9. Legionnaires’ Disease: The Control of systems in general practice. Appl Environ
 Once the boil notice is cancelled, follow Legionella Bacteria in Water Systems. Microbiol 2000; 66: 3363−3367.
the detailed guidance specified in HTM Approved Code of Practice and Guidance 20. Fotos PG, Westfall HN, Snyder IS,
01-053 for DUWLs and other advice given L8 (4th edn). HSE Books, 2013. www.hse. Miller RW, Mutchler BM. Prevalence of
by the local water utility. If no guidance gov.uk Legionella-specfic IgG and IgM antibody
is provided, flush dental water lines and 10. Puttaiah R, Cederberg R. Assessment of in a dental clinic population. J Dent Res
faucets for at least three minutes before endotoxin levels in dental effluent water. 1985; 64: 1382−1385.
resuming patient care. Then disinfect J Dent Res 1998; 77: 263. 21. Reinthaler FF, Mascher F, Stϋnzner D.
dental water lines and in line filters as 11. Putnins EE, Di Giovanni D, Bhullar AS. Serological examinations for antibodies
recommended by the manufacturer. Dental unit waterline contamination against Legionella species in dental
In the last two decades in and its possible implications during personnel. J Dent Res 1988; 67: 942−943.
response to the evidence base and periodontal surgery. J Periodont 2001; 72: 22. Atlas RM, Williams JF, Huntington MK.
publication of new regulations, dental 393−400. Legionella contamination of dental-unit
April 2017 DentalUpdate 291
InfectionControl

waters. Appl Environ Microbiol 1995; 61: Water in Healthcare Premises. (HTM water systems and its practical control.
1208−1213. 04-01): Guidance on Design, Installation, J Appl Microbiol 2009; 106: 1424−1437.
23. Schulze-Robbecke R, Feldman C, Fischeder Commissioning, Testing, Monitoring and 32. Schel AJ, Marsh PD, Bradshaw DJ,
R, Janning B, Exner M, Wahl G. Dental Operation of Water Supply Systems in Finney M, Fulford MR, Frandsen E et
units: an environmental study of sources Healthcare Premises. July 2016. www. al. Comparison of the efficacies of
of potentially pathogenic mycobacteria. gov.uk/government/publications/hot-
disinfectants to control microbial
Tuber Lung Dis 1995; 76: 318−323. and-cold-water-supply-storage-and-
contamination in dental unit water
24. Martin MV The significance of the bacterial distribution-systems-for-healthcare-
contamination of dental unit water premises systems in general dental practices
systems. Br Dent J 1987; 163: 152−154. 28. Health Facilities Scotland. Scottish Health across the European Union. Appl Environ
25. Department of Health. The Health and Technical Memorandum 04-01: Water Microbiol 2006; 72: 1380−1387.
Social Care Act 2008: Code of Practice on safety for healthcare premises Part A: 33. Percival RS, Devine DA, Nattress B,
the Prevention and Control of Infections Design, installation and testing. Kite P, Marsh PD. Control of microbial
and Related Guidance. 2015. www. 29. Cobb CM, Martel CR, McKnight SA, contamination in dental unit water
gov.uk/government/publications/ Pasley-Mowray C, Ferguson BL, Williams systems using tetra-sodium EDTA.
the-health-and-social-care-act-2008- K. How does time dependent dental J Appl Microbiol 2009; 107: 1081−1088.
code-of-practice-on-the-prevention- unit waterline flushing affect planktonic 34. Smith AJ, McHugh S, Aitken I, Hood
and-control-of-infections-and-related- bacteria? J Dent Educ 2002; 66: 549−555.
J. Evaluation of the efficacy of Alpron
guidance 30. Berlutti F, Testarelli L, Vaia F, De Luca M,
disinfectant for dental unit water lines.
26. Health and Safety Executive. Legionnaires’ Dolci G. Efficacy of anti-retraction devices
Disease: Technical Guidance Part 3: The in preventing bacterial contamination of Br Dent J 2002; 193: 593−596.
Control of Legionella Bacteria in Other Risk dental unit water lines. J Dent 2003; 31: 35. O’Donnell MJ, Boyle MA, Swan J, Russell
Systems. HSG274, Part 3 2013. www.hse. 105−100. RJ, Coleman DC. Management of dental
gov.uk/pubns/books/hsg274.htm 31. Coleman DC, O’Donnell MJ, Shore AC, unit waterline biofilms in the 21st
27. Department of Health (England). Safe Russell RJ. Biofilm problems in dental unit century. Fut Microbiol 2011; 6: 1209−1226.

MJDF: the best step


in a dentist’s career,
with unrivalled support
No other examination is as relevant to your career
The MJDF is the only examination of its type to test you on the skills and
knowledge you use in practice. In addition, skills that bridge the divide
between primary and secondary care are tested, making the MJDF a
good step to specialism.

Two Faculties for less than the price of one


Run by the two Dental Faculties at The Royal College of Surgeons, upon
successfully passing the MJDF you can be a member of both Faculties for
less than the price of a single membership. This means dedicated support
on primary care dentistry from the Faculty of General Dental Practice
(FGDP) and in secondary care from the Faculty of Dental Surgery (FDS)