J-SIL SCIENTIFICINDUSTRIES

ISO/NEC I7025 : 2017 - NABL Accredited Calibration Laboratory, vide Certificate No.CC-3562
alu
48-B, INDUSTRIAL ESTATE, NUNHAI, AGRA(INDIA) 282 006
Tel : (0562) 2281967, 2280186
E-mail : jsilcalibrationlab(@gmail.com, Website : www.j-sil.in jlac-MRA
CC3562
orat tto. . tStteefTa-o1
CALIBRATION CERTIFICATE
Calibration Request Sr. No. J-SIL/126 ULR No. CC35622400000260F
Calibrated For - ABC Certificate No. 2400000260

Item For Calibration |VOLUMETRIC FLASK Equipment & Master Used For Calibration
Make/Catalogue No. J-SIL/256J
1. DIGITAL WEIGHING BALANCE
Capacity 100 + 0.1 ml
Serial /|D No. : A&D/GX-1603A
Least Count NIA
Certificate No. : SRT/24/JM/693
Item Serial/ID No. 100N/AA
Valid Upto : 20/01/2025
Temperature 27 °C
Visual Inspection SATISFACTORY 2. THERMOMETER
Date of Receipt 09/02/2024 Serial /ID No. : RM25/01-25
Date of Calibration 09/02/2024 Certificate No. S-230117
Date of Issue Certificate 12/02/2024 Valid Upto :22/01/2025
Recommended Next Due 08/02/2025
Date of Calibration

Environmental Temperature 254 4 °C,


Location : Volume Calibration Lab
Conditions Humidity. 50 ± 20% RH

C¤LIBRATION RESULTS
Standard Capacity Actual Capacity Delivery Time Expanded Uncertianity
ml ml Sec. @95% CL, at k=2
100 99.993 NIA ±0.008 ml

Conformity Statement : The above product has been tested with compliance to ISO- 1042
Class 'A' tolerances and above results are found within the specified tolerance limits.
Methodology of Calibration : As per Calibration Procedure No. : J-SILJCPM-01
Refernce Document for Calibration : IS0-4787

Note: 1. All the above cquipments are traceable to National/International standards through NABL aceredited Laboratory.
2 The calibration results reported in this certificate are valid at the time of and the stated condition of measurement.
3. This report should not be reproduced except in full without our prior permission in writing.
4. The results stated in this certificate relate only to the item calibrated.
Customer has not asked for any conformity statement.
End of Certificate'

ROHIT SHRIVASTAV

AGRA Phit
Authorised Signatory
282006)
CALIBRATIOC ENGINEER TECHNICAL MANAGER
J-SILSCIENTIFIC INDUSTRIES
Laboratory, vide Certificate No.CC-3562
ISO/TEC 17025: 2017 - NABL Accredited Calibration
(INDIA) - 282 006
48-B, INDUSTRIAL ESTATE, NUNHAI, AGRA
Tel : (0562) 2281967, 2280186 üac-A
E-mail : jsilcalibrationlab(@gmail.com, Website : www.j-sil.in
Format No. : JSI/CC/F31A(7.8-01)

CALIBRATION CERTIFICATE
CC35622400000336F
Calibration Request Sr. No. JS/130 ULR No.
Calibrated For - ABC Certificate No. 2400000336

Item for Calibration VISCOMETER Equipment & Master Used For Calibration
and
01. Standard Viscometer traceable to National
Size BS/U - B international standards.
Serial/ID No. : JS1/822
Make/Catalogue No. JSI /N1432B Certificate No. C1/0000317631
21/02/2025
Item Serial/ID No. JS1NI38 Valid Upto
Visual Inspection Satisfactory 02. PRT wih indicator - With stability 0.01 °C
SetialD No. : 23190/DTC-01
Certificate No. CC288823000001484F
Date of Receipt 12/03/2024
Valid Upto 24/01/2025
Date of Calibration 13/03/2024
03. STOP WATCH
Serial/ID No. : DSW-01
Date of Issue Certificate 15/03/2024
Certificate No. : AT24000000198
Recommended next Due 19/01/2025
Date of Calibration
12/03/2025 Valid Upto

Environmental Temperature 25 ±4 °C, Location: Viscomneter Calibration Lab


Conditions Humidity 50+ 20 %RH
CALIBRATION RESULTS

Constant of Viscometer Temperature Expanded Uncertainty


o @95% CL, at k-2
mm² /s² (cSt/s)
0.010757 40 0.7%

Methodology of Calibration: As per Calibration Procedure No. :J-SiVCPM-02


Reference Document for Calibration: ASTM-D446

Note: 1. All the above equipments are traceable to National/International standards through NABL accredited Laborator.
2. The calibration results reported in this certificate are valid at the time of and the stated condition of measurement.
3. This report should not be reproduced except in full without our prior permission in writing.
4. The results stated in this certificate relate only to the item calibrated.
5. Customer has not asked for any conformity statement.
End of Certificate