1 SYSTEM FEATURES
MEDISON’s flagship brand, ACCUVIX, is moving toward perfection with V series by continuing to evolve its
technology. The ACCUIVX V20, which is the new product leading ACCUVIX V brand identity, provides the superb
finest 2D image quality and high performance 3D/4D featuring cutting-edge technologies: Color STIC, Inversion 3D,
Extended Oblique View Imaging (Live Mode & OVIXTM), VCETM as well as 3D XITM, XI STICTM, and XI VOCALTM. Also you
can get a convenient and efficient diagnostic environment with 10.4 Inch Touch Screen and diagnostic simplification
process. ACCUVIX V20 also supports DICOM 3.0 with DICOM Structured Report and is easily connected to PACS
networking.
2 SYSTEM OVERVIEW
The configuration of ACCUVIX V20 is as follows,
Module Description
Ultrasound module Ultrasound Engine Modules: Frontplane, CW, Beamformer, DSP,
S/W DSC module, PC Module, Backplane, ECG
CPU module Main host CPU: CPU card including Pentium Core 2 Duo 1.8GHz processor, max.
2GB main memory, VGA/LAN/Sound functions, and Interfacing function to
Ultrasound module
Key module Key Interface, Key Matrix, Trackball unit
Rear module System Output part: Rear Left/Right
Software module Main control, Measurements, DB engine, SonoView II, 3D view, etc.
Mechanical Design / Rack, Housing, Chassis, etc.
Enclosure module
Power supply module Power supply
Monitor module 17” LCD high resolution monitor / 10.4” Wide Touch Screen
TECHNICAL SPECIFICATION - ACCUVIX V20 | 1
3 MAIN FEATURE LISTS
• Real time 129,024 channel 2D gray-scaled imaging with multi-beam receiving
• Multibeam Beamformers with:
» Dynamic Focus
» Dynamic Aperture
» Dynamic Apodization
» The shape of the acoustic spectrum
• Synthetic Aperture Control
• Motion mode (M-mode)
• Color Motion mode (CM-mode)
• Color Doppler Imaging with Quad-beam receiving
• Power Doppler Imaging with Quad-beam receiving
• Directional Power Doppler
• Pulsed wave (PW) spectral Doppler
• Multi frequency Doppler
• Tissue Doppler Imaging
• Tissue Harmonic Imaging
• Pulse Inversion Harmonic Imaging
• Power Pulse Inversion Harmonic Imaging
• Extreme High Dynamic Range (170dB)
• Full Spectrum Imaging
• Trapezoidal Imaging
• Color Edge Processing
• Combined modes
» B/M, B/PWD, B/C, B/PD, B/DPD, B/PPI, B/TDB, B/CWD
» B/C/PWD, B/PD/PWD, B/DPD/PWD,B/PPI/PWD,B/TD/PWD,
B/C/M, B/C/CWD, B/PD/CWD2D/PD/PWD, 2D/C/PWD
» Dual Live mode
» Dual mode
» Quad mode
• Cine for 5,400 frames and Loop Review for 8,192 lines
• Clip Cine Store
• Volume Cine Save
• Color VGA monitor with non-interlaced display
• Integrated 3D Imaging Package
• Freehand 3D, Static 3D, Live 4D
• VOCAL
• SHELL H
• VCETM (Volume Contrast Enhancement)
• Inversion 3DTM
• Orientation CubeTM
• Multi-Slice View (Optional)
: Multi-Slice View transforms 3D volume data obtained from a regular ultrasound scan into a series of sequential images
captured in intervals of 0.5mm to 5.0mm segments. Enables instantly view, analyze, and understand more in-depth data.
• Oblique View (Optional)
: Oblique View is imaging technology which enables physicians to examine and view 3D volume data in various planes of view
without limitations. The acquired portion of the 3D data allows complete visual examination and better understanding of the
correlation between organs and other areas within the region of interest.
4 SCANHEADS
• Curved Array: C2-6IC, C2-5EL, C3-7IM, EC4-9IS
• Linear Array: L5-13IS, L7-16IS, L4-7EL, L5-12/50EP
• Volume Scanhead: 3D2-6ET, 3D4-8ET, 3D4-9ES, 3D5-9EK
• Phased Array: P2-4AC, P3-5AC
• Pencil Probe: 2.0M CW, 4.0CW
4.1 C2-6IC
• Application : Abdomen, OB/GYN, Vasculoar, Urology, Contrast agent
• Probe Type : Curved Linear Array
• Center Frequency : 3.3[MHz]
• Radius of Curvature : 60[mm]
• Field of View : 58.12[°]
• Number of Elements : 192
• Element Pitch : 0.317[mm]
• Elevation Aperture : 12[mm]
• Geometric Focus : 70[mm]
• Biopsy guide available
• Safety class: BFle
4.2 C2-5EL
• Application : Abdomen, OB/GYN, Vasular, urology
• Probe Type : Curved Linear Array
• Center Frequency : 3.2[MHz]
• Radius of Curvature : 40[mm]
• Field of View : 75.54[°]
• Number of Elements : 128
• Element Pitch : 0.412[mm]
• Elevation Aperture : 13[mm]
• Geometric Focus : 80[mm]
• Biopsy guide available
• Safety class: BF
4.3 C3-7IM
• Application : Abdomen, OB/GYN, Vascular, Urology, Pediatrics, Contrast agent
• Probe Type : Curved Linear Array
• Center Frequency : 5.0[MHz]
• Radius of Curvature : 50[mm]
• Field of View : 77.5[°]
• Number of Elements : 192
• Element Pitch : 0.352[mm]
• Elevation Aperture : 11[mm]
• Geometric Focus : 70[mm]
4.4 EC4-9IS
• Application : OB/GYN, Urology
• Probe Type : Endocavity
• Center Frequency : 6.5[MHz]
• Radius of Curvature : 10[mm]
• Field of View : 148.9[°]
• Number of Elements : 192
• Element Pitch : 0.138[mm]
• Elevation Aperture : 6[mm]
• Geometric Focus : 35.0[mm]
• Biopsy guide available
• Safety class: BF
4.5 L5-13IS
• Application: Breast, Small Parts, Vascular, Musculoskeletal, Pediatrics
• Probe Type: Linear Array
• Center Frequency: 9.0[MHz]
• Radius of Curvature: Flat
• Field of View: 38.4[mm]
• Number of Elements: 192
• Element Pitch: 0.2[mm]
• Elevation Aperture: 4[mm]
• Geometric Focus: 20.0[mm]
• Biopsy guide available
• Safety class: BF
4.6 L7-16IS
• Application: Breast, Small Parts, Vascular, Musculoskeletal, Pediatrics, Superficial
• Probe Type: Linear Array
• Center Frequency: 12.5[MHz]
• Radius of Curvature: Flat
• Field of View: 38.4[mm]
• Number of Elements: 192
• Element Pitch: 0.2[mm]
• Elevation Aperture: 4[mm]
• Geometric Focus: 20.0[mm]
• Biopsy guide available
• Safety class: BF
4.7 L4-7EL
• Application : Breast,Vascular, Small Parts, Musculoskeletal,Pediatrics
• Probe Type : Linear Array
• Center Frequency : 5.0[MHz]
• Radius of Curvature : Flat
• Field of View : 38.4[mm]
TECHNICAL SPECIFICATION - ACCUVIX V20 | 5
• Number of Elements : 128
• Element Pitch : 0.3 [mm]
• Elevation Aperture : 5[mm]
• Geometric Focus : 20.0[mm]
• Biopsy guide available
• Safety class: BF
4.8 L5-12/50EP
• Application : Breast, Small Parts, Vascular, Musculoskeletal, Pediatrics
• Probe Type : Linear Array
• Center Frequency : 7.5[MHz]
• Radius of Curvature : Flat
• Field of View : 50[mm]
• Number of Elements : 128
• Element Pitch : 0.39[mm]
• Elevation Aperture : 6[mm]
• Geometric Focus : 22.0[mm]
• Biopsy guide available
• Safety class: BF
4.9 3D4-8ET
• Application : Abdomen, OB, GYN
• Probe Type : 3D Curved Linear
• Center Frequency : 4.5[MHz]
• Radius of Curvature : 40[mm]
• Volume angle : 70[°]
• Field of view : 84[°]
• Number of Elements : 128
• Element Pitch : 0.375[mm]
• Elevation Aperture : 13[mm]
• Geometric Focus : 60[mm]
• Biopsy guide available
• Safety class: BF
4.10 3D2-6ET
• Application : Abdomen, OB/GYN
• Probe Type : 3D Curved Linear
• Center Frequency : 3.5[MHz]
• Radius of Curvature : 40[mm]
• Volume angle : 70[°]
• Field of view :84[°]
• Number of Elements : 128
• Element Pitch : 0.468[mm]
• Elevation Aperture : 13[mm]
• Geometric Focus : 80[mm]
• Biopsy guide available
• Safety class: BF
4.12 3D5-9EK
• Application: General, OB, Gyn, Urology
• Probe Type: 3D Curved Linear
• Center Frequency: 6.5[MHz]
• Radius of Curvature: 11.65[mm]
• Field of View: 146[degree]
• Number of Elements: 128
• Element Pitch: 0.1964[mm]
• Elevation Aperture: 6[mm]
• Geometric Focus: 30.0[mm]
• Biopsy guide available
• Safety class: BF
4.13 P2-4AC
• Application : Adult cardiac, Pediatric cardiac, Renal, Aorta, TCD
• Probe Type : Phased Array
• Center Frequency : 2.5[MHz]
• Radius of Curvature : Flat
• Field of View : 19.2[mm]
• Number of Elements : 64
• Element Pitch : 0.3[mm]
• Elevation Aperture : 14[mm]
• Geometric Focus : 85.0[mm]
• Biopsy guide available
• Safety class: BF
4.14 P3-5AC
• Application : Adult cardiac, Pediatric cardiac, Renal, Aorta, TCD
• Probe Type : Phased Array
• Center Frequency : 4.0[MHz]
• Radius of Curvature : Flat
• Field of View : 14.1[mm]
• Number of Elements : 64
TECHNICAL SPECIFICATION - ACCUVIX V20 | 7
• Element Pitch : 0.22[mm]
• Elevation Aperture : 12[mm]
• Geometric Focus : 70.0[mm]
• Biopsy guide available
• Safety class: BF
4.15 CW2.0
• Application : Cardiology, TCD
• Probe Type : Pencil type
• Center Frequency : 2.0[MHz]
• Acoustic aperture: 13.8[mm]
• Elevation focus depth : 55[mm]
• Safety class: BF
4.16 CW4.0
• Application : Cardiology, TCD
• Probe Type : Pencil type
• Center Frequency : 4.0 [MHz]
• Acoustic aperture: 7.0[mm]
• Elevation focus depth : 45[mm]
• Safety class: BF
6.1 Applications
• Abdominal
• Obstetrical
• Obstetrical Early
• Neonatal
• Peripheral vascular
• Gynecological and fertility
• Infertility
• Small parts (breast, thyroid, parathyroid, penis, testes)
• Abdominal surgery
6.2 ERGONOMICS
• Advanced ergonomic key grouping
• Durable design
• Prevention of noise of the system
• Tilt and swivel articulation arm monitor
• 5 active transducer ports for simultaneous transducer connection(include pencil probe)
• High quality stereo audio speaker system
• Input and output connections on the rear panel
• Rear compartment for storage of accessories
• Front/Rear handle
• Attachable key panel
• Lighted alphanumeric keyboard
• Rotate control panel
• 6 Back USB ports (for digital connection of peripherals)
• 4 Front USB ports(for used memory stick and external hard drive connection)
6.4 MONITOR
• 17” high resolution LCD non-interlace color monitor (19” upgradable)
• Resolution: 1280x1024x24bit
• High brightness & contrast
TECHNICAL SPECIFICATION - ACCUVIX V20 | 9
• Monitor Moving
» Lift
» Tilt
» Swivel
6.25 2D MODE
• Dynamic Range: 50~170dB, 1dB steps
• Gray scale: max 256 level(8bit)
• Gray scale map: type1~8,User1~2 (10 steps)
• Scan line: max 1024line
• Image format: Convex, Linear, Phased array
• Transmit Focus
» Predetermined points (max.8)
» Multi-zone Focal point (max.4)
• Angle of the sector/convex: optimized by all probes
• Zoom: Read zoom/Write zoom
• Frequency: Pen, Gen, Res
TM
• Full Spectrum Imaging : 1/2/3 step
• DynamicMRTM : 1/2/3/4/5 step
• Speckle Reduction FilteringTM : 1/2/3 step
• Spatial Compound ImagingTM : 1/2/3 step (available all linear probes)
• Orientation control: 0°/90°/180°/270°
• Panning: Positioning X, Positioning Y
• Dual/Quad control
TM
• Quickscan
• Biopsy Guideline contol
6.27 M MODE
• Dynamic Range: 50~170dB, 1dB steps
• Gray Scale: max.256level(8bit)
• Sweep speed: 60/120/180/200/240/300/360Hz, 6steps
• M gray scale map : 8steps
• M edge enhancement: 7steps(-3~3)
• M colorization: 9 chroma map
• M-color flow mode
• Loop Format: Top/Bottom,Left/Right
• Free Angle M mode: Fixed M Line
• Loop cine memory: max. 8192line
• ECG trace
7 ULTRASOUND WORKSTATION
• Pentium VI Processor 2GHz • Integrated ODD: DVD Multi Recordable Driver
• Main Memory: DDR2 SDRAM 2GB • LAN: 10BASE-T
• Integrated Hard drive: SATA HDD • USB2.0
(Capacity: 250 GB) • Windows Embedded OS
8 MEASUREMENT PACKAGE
Function Description
Measurement B mode: distance, line trace, angle, area, ellipse, circumference, volume
D mode: velocity, pressure, acceleration, deceleration
M mode: time, slope, distance
OB measurements Mode Menu Group Items
B Fetal Biometry GS, CRL, YS, BPD, OFD, HC(BPD, OFD)*, APD, TAD,
AC(APD, TAD)*, FTA(APD, TADD)*, ThC(APTD, TTD)*,
FL, SL, TTD, APTD
Calculation: MAD(APD,TAD)*, APTDxTTD(APTD, TTD)*
Fetal Long Bones HUM, ULNA, TIB, RAD, FIB, CLAV, Vertebral, NB
Fetal Cranium CEREB, OOD, IOD, CM, NF, NT, Lat Vent.
Fetal Others Foot, Ear, MP, Lt. Kidney, Rt Kidney, Pelvis
AFI Q1, Q2, Q3, Q4, MVP(Maximum Vertical Pocket)
CTAR ThD ap, ThD trans, HrtD ap, HrtD trans, ThA, HrtA
Calculation: CTAR(D), CTAR(A)
D PLI Sys Flow, Dias Flow, A. Rev Flow
Umb. Artery Auto Trace, Limited Trace, Manual Trace, PSV, EDV,
(Umbilical Artery) %StA, %StD, Vesl. Area, Vesl. Dist
MCA Auto Trace, Limited Trace, Manual Trace, PSV, EDV,
(Mid Cerebral Artery) %StA, %StD, Vesl. Area, Vesl. Dist
9 DOCUMENTATION CAPABILITIES
• On-board VCR controls
• On-board printing device control
• Selective printing on two connected printers
• SonoView II
• Image Filing Package
• Export Media:CD/DVD+R/-R/RW,USB Flash, External USB HDD
• Exprot Format: JPEG,BMP,TIFF,DICOM
• Print Function
• Capacity:10BASE-T(Min.4000frame)
• Patient list and data search
• Report save available
• Post image processing available
• Caliper measurement available
• 3D View
• DICOM 3.0
• Service: Storage/Printer/Worklist
• DICOM Structured Reports
10 OPTIONAL DEVICES
Device Description
Video Cassette Recorder (VCR) Panasonic MD835 SVHS (NTSC & PAL)
Analog
Video Cassette Recorder (VCR) Sony DVO-1000MD
Digital
Video Page Printer(BW) Mitsubishi P-93WM Monochrome Video Printer
Video Page Printer(Color) 1) Mitsubishi CP-910U Color Video printer
2) Sony UP20 Digital Photo Thermal Printer
Foot Switch The functions of Left &Right Foot Pedals can be selected in Setup Mode.
Freeze, Update, Record, Print, Store, 3D, ECG Trigger On/Off
11 PERIPHERAL SIGNALS
14 OPERATING ENVIRONMENT
• Ambient temperature: 10°C-35°C (50°F-104°F)
• Relative humidity: Up to 75% non-condensing
• Pressure: 700~1060hPa
• Audible noise: 37dB
• Safety class: B or BF