CALIBRATION CERTIFICATE

ULTRASONIC FLAW DETECTOR


Certificate No: NBTC-TSD-NDT-UT-133 Calibration Date: 24-Apr-23 Recommended Due Date: 23-Oct-23

UNIT UNDER TEST (UUT) DETAILS


Naser M.AL-Baddah & Partner General Trading &Contracting
Customer Details: Co.W.l.l
Asset Number: 260085
Serial No: 170972501
Description: Ultrasonic Flaw Detector
Manufacturer: Olympus
Model: Epoch 600
Physical Condition of item when received: Normal / Good

ENVIRONMENTAL CONDITIONS
Calibration Venue: NBTC TSD CALIBRATION LABORATORY
Temperature: 24°C
Relative Humidity: 18%

REFERENCE
PROCEDURE: NBTC-TS/OP-CA07

CALIBRATION VERIFICATION INFORMATION


TIME BASE LINEARITY
PROBE TYPE: MSEB4(4MHz,Dual Crystal,10 mm)
BLOCK
Back Wall Echo Readings (mm) Acceptance Limit RESULT
TEST RANGE (mm) THICKNESS
(mm) 1 2 3 4 5
Devaiation Shall be within ±2%
25-250 25 25.00 50.00 149.47 198.92 248.85 PASS
of test range
Deviation: 0.00 +0.00 -0.53 -1.08 -1.15
SCREEN HEIGHT LINEARITY
PROBE TYPE: DA 501 (5MHz, Dual Crystal, Ø12mm)
STANDARD Indication set at Measured Percentage of Acceptance Limit
STANDARD RESULT
DESCRIPTION % FSH % FSH Error % FSH %
100 % 51 1% 5% PASS
80 % 39 1% 5% PASS
V1 Block 25mm 60 % 30 0% 5% PASS
40 % 20 0% 5% PASS
20 % 9 1% 5% PASS
AMPLITUDE CONTROL LINEARITY
Indication set at Indication Indication
dB control
% FSH Limits observed RESULT
change
% FSH % FSH
V1 Block 25mm 80 % -6 dB 35% to 45% 40% PASS
80 % -12 dB 15% to 25% 21% PASS
40 % +6 dB 65% to 95% 80% PASS
20 % +12 dB 65% to 95% 79% PASS
The calibration was verified using Certified Reference Materials, This instrument has been calibrated in accordance NBTC Calibration
Procedure with the manufacture's instrument User's Handbook using standards that are traceable to the international System of units
(SI). This report applies only to the item identified and shall not be reproduced, except in full, unless written permission for an approved
abstract is obtained from NBTC TSD Calibration Laboratory. Calibration reports without authorizing signature(s) are invalid.

Calibrated By Authorizing Signature


SAIFER R.SURESH KUMAR
ASNT UT LEVEL-II ASNT UT LEVEL-III

NBTC-TS/ND/FO35 Rev.2