MANUAL
R E V. 2 . 1
Caution!
This manual describes how to install and operate the G210 ColorChex 60. Read this manual
completely before attempting to install or use the instrument. Important information is included in
G210 ColorChex 60 wiring diagram, which must be referred to for electrical installation.
The G210 ColorChex 60 is a professional laboratory instrument used for assessment of fabric
color. You, as the Operator, must be authorized and should have received appropriate training. It is
the user’s responsibility to ensure the safe operation of this instrument. SDL Atlas Ltd cannot
accept any responsibility whatsoever for accident resulting from failure to observe these precautions
or safety instructions.
1. This apparatus utilizes high voltage. Any unauthorized modifications or damage to the
apparatus may result in electrical shock. Handle all components with care. Inspect
regularly for damage to components and cabling. Take the unit to a qualified service
person for inspection and necessary repairs.
2. Ultraviolet source inside, UV radiation protection for eyes and skin is recommended
during operation. During use the lamp tray will become hot, do not touch.
3. The mains plug of the power supply cable is the only way to completely switch off the
unit. In order to isolate the apparatus from the mains supply, the plug of the supply
cable should be removed from the mains outlet.
4. Connect the instrument to the power supply outlet using the supplied power supply
cable. Do not modify the supplied cable. The mains outlet shall be located in a user
accessible position for easy unplugging.
5. Replacement of lamps shall be carried out by qualified service personnel. The apparatus
shall be isolated from the mains supply prior to replacement.
6. Please follow the procedure laid out in this manual. Improper installation or modification
of the apparatus shall void all warranties.
7. Safety Symbols:
1
G210 ColorChex 60
1 x Back plate
2 x Side plate
Left side plate Right side plate
1 x Canopy Base plate
2
G210 ColorChex 60
groove).
IMPORTANT
As a safety precaution this cabinet should be properly earthed.
3
G210 ColorChex 60
Light
Lamp Information
Source
Correlated color temperature is 6500K. Within the
tolerances prescribed in BS 950: Part 1, Illuminant for
Artificial color matching appraisal. Specified for most
D65
Daylight D65 applications where there is a need to maintain color
consistency and quality. Conforming to the CIE
specifications, for accurate color matching.
CABINET LOCATION
For best results, glare, extraneous light and anything which can affect the visual response must
be avoided. Any windows near the cabinet should be fitted with grey blinds. No direct light should
be in the field of view and the general lighting in the vicinity should also be “Artificial Daylight” to
a level of 200-300 lux. The wall area behind the cabinet should be finished in light grey emulsion.
These measures will ensure comfortable and accurate color appraisal.
Specification
Dimensions : Outer, 710mm(W) x 565mm(H) x407mm(D)
Voltage : 230V/50-60Hz / 110V/60Hz
Rated power : 250W
4
G210 ColorChex 60
TL84
D65
CWF
A H H H H A
TL84
D65
FRONT
LAMPS
Note:In order to get the accuracy visual assessment, it is proposed all lamps should be
changed every 2,000 hours (except A&H only 1,000 hours) (RE.: Menu & Setting
they should be repainted using SDL Atlas supplied paint to ensure correct color
5
G210 ColorChex 60
Basic Operation:
V ie w D a ta MENU
Figure 1
Connect ColorChex to power supply using the power supply cable provided. Press any key on
the panel to continue. The LCD screen displays the company logo and then proceeds to the
operation screen (figure 1). In the operation screen the top line of the LCD display indicates the
various light sources available. To illuminate any light source press buttons numbered 1-5 as
required. To switch off the light source, press the same numbered button. With no lamps
illuminated, press and hold ESC for 1 second to power off the ColorChex.
6
G210 ColorChex 60
Figure 2
The Remote Control can be used for basic operation of the ColorChex up to range of 10 meters. Connect
ColorChex to power supply using the power supply cable provided. Switch on remote control (figure 2) by sliding
key lock to the left position. Press power button to switch ColorChex on. To switch on any light source press
buttons numbered 1-5 as required and hold for two seconds. To switch off light source press the same button.
Press power button to switch off ColorChex. Slide key lock to the right position when not in use.
7
G210 ColorChex 60
8
G210 ColorChex 60
9
G210 ColorChex 60
instrument will carry out program once and then 1. Run once
stop (figure 10). Select Run continuously and
2. Run continuously
3. Reset
instrument will repeat program until ESC is
pressed (figure 11). Select Reset to enter the
Figure10
teach mode where the sequence can be
re-programmed (figure 12). To do this press the
lamp No. you wish to start with, after the required RUN ONCE
viewing duration press the next lamp No. and so D65 0 0 0 TL84 0 0 0 A 000
on for a maximum of 9 selections each with a H000 CWF0 0 0
maximum time of 90 seconds. When the final lamp
viewing duration is complete, press Enter, And
Figure11
then select Ok to save. Run the program from the
screen shown in figure 9.
RUN CONTINUOUSLY...
TL84 0 0 0 D65 0 0 0 A 000
H000 CWF0 0 0
10
G210 ColorChex 60
Figure12
11
G210 ColorChex 60
Maintenance
Though your apparatus requires little care, you can still maintain its excellent conditions and
• Keep the apparatus away from excessive moisture, extreme heat or cold and the risk of fire or
electric shock.
• To avoid damage, do not place external devices or other objects on the apparatus.
• Do not excessively scratch the inner surface. Clean the inner surface using a soft, dry cloth or a
vacuum cleaner. To remove stains, use a soft cloth and warm soapy water.
• Fuse Replacement:
a) PCB: F1(F2.5A/250V); F2(T500mA/250V)
12