Summing Box Guide
Summing Box Guide
If you change the connection like this picture, you can change the load cell type, easily.
As same as above picture, change the “Signal wire”, you can use compression type load cell, as
Tension.
1
#CHAPTER 2
ACCURACY
1. Concept Accuracy
Degree of how sensitive and reliable to use load cell. We can express this accuracy level, with combined
error of nonlinearity and repeatability, expressed as a percent of full scale output.
Normally, we can guarantee 1/3,000 accuracy for only Load cell. But, the main factor to determine
Scale’s accuracy is Digital Weighing Indicator. Digital Weighing Indicator amplifies Load cell output
and converts to Digital signal, so we can guarantee higher accuracy than Load cell itself. If the
Indicator has high Accuracy(resolution), you can make higher accuracy scale with same load cell.
- As above, if your customer has same load cell, but the accuracy level is different. The important
factor to determine accuracy is not load cell, Digital Weighing Indicator. But, normal commercial
Scale, which you can see in super market, is 1/3,000 accuracy one.
2. Accuracy Factors
Example) If you have 100kg capacity load cell and its Rate output is 2.0mV.
Simply, if you press 50kg load on the load cell, load cell gives you 1.0mV
If you press 75kg load, you can get 1.5mV
Normally, Calibration of Digital Weighing Indicator uses rate output of load cell. Digital Weighing
Indicator takes rate output from load cell at each weight point and calculates weight value, and
calibration process is making standard for this weight calculation. But, there is big difference to
use 3mV load cell and 1mV load cell.(The important thing is Digital Weighing Indicator)
2
Creep for 30min. : Accuracy factor of Load Cell
The accuracy factor of measurement. This factor means, after loading weight on the load cell and
observe 30min., then how much degree of error can be occurred. “CS” brand load cell has 0.03%
of loaded weight and it means as follow.
3
#CHAPTER 3
SUMMING BOX GUIDE
Example) plural Load Cells connection : 4pcs of 100kg load cells and rate output is 2.0mV.
Max. capacity calculation : 100kg * 4 = 400kg
Rate Output voltage : 2.0mV (Because the connection is parallel).
※Caution: If you make connection plural pcs of load cell, you have to use all same
Rate output(2mV) models.
Reference Picture
Case 1. Plural load cells connection with “Summing Box”.
Using “Summing Box”, you can connect Max. 8pcs load cells with one Digital Weighing Indicator.
4
#CHAPTER 4
CONNECTION DIAGRAM
Load cell cable specification
Basically, one digital weighing indicator able to connect any Load Cell maker’s load cell. However
each load cell maker they have their own wiring specification.
Example 1) SEWHACNM load cell wiring. There are total 5 wire exist.
Example 2) VISHAY load cell wiring. There are also 5 wire exist. But wire colors are different.
Example 3) PT load cell wiring. There are 7 wire exist. Also wire colors are different.
5
[Summing box]
*Application: There are SI4500 one unit and four SB210 load cell connected.
*Make sure EXC- and EXC+ common terminal for two load cell’s wiring.
Example 1) If you use SEWHACNM’s load cell no need to connect with SEN- and SEN+.
Example 2) If you use PT’s load cell need to connect with SEN- and SEN+ either.
6
[Summing box]
*Application: There are SI4500 one unit and three SS300 load cell connect.
EXC+ EXC- SIG+ SIG- SEN+ SEN- SHIELD EXC+ EXC- SIG+ SIG- SEN+ SEN- SHIELD
SHIELD
SEN-
SEN+
SIG-
SIG+
EXC-
EXC+
EXC+ EXC- SIG+ SIG- SEN+ SEN- SHIELD EXC+ EXC- SIG+ SIG- SEN+ SEN- SHIELD
Example 1) If you use SEWHACNM’s load cell no need to connect with SEN- and SEN+.
Example 2) If you use PT’s load cell need to connect with SEN- and SEN+ either.
7
#CHAPTER 5
How to adjust the weight balance with the Summing box
1. After calibration progress done in Indicator with load cell, double check load the weight(10%
of full capacity) & check the weight value.
If the weight value showing different at every single corner, please turn the POT. in the summing
box. Whenever you turn it, the weight display will be increasing or decreasing.
Example) If the weight is 20kg, each weight value display at single corner 18.9kg,19.1kg,19.0kg,
20.2kg,then make the balance as 19kg than making them 20kg(although the weight is 20.0kg).
(At the last you must process calibration after all the value of corner has same value as 20kg or
19kg.)
Indicator
Summing box
Load cell1 Weight Load cell 4