Detection Limit
Detection Limit
Detection Limit
The detection limit was determined by spiking negative urine specimens with
intact hCG standard at 0, 25, 100mIU/mL and 1000mIU/mL. The hCG
standards were randomized and coded. Abbott AxSYM confirmed the
results. A total of 10 replicates for each standard were tested according to the
package insert. The specimens were tested with visual interpretations 5
minutes after specimen application. The results are presented in Table
below.