1

Version :V1.0
Release Date : 2021.07.17
Ultrasound Technical Support Department
www.sonoscape.com 以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Content
2

1 System Overview

2 Software

3 Hardware

4 Troubleshooting

5 Peripherals Connection

6 Preventive Maintenance

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
3

Appearance
System Overview Configuration
Installation

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Appearance
4

21.5 inch monitor


Monitor Arm
插入图片 13.3 inch touch
screen

1
Probe Keyboard
holder

Wheels
Probe connector

插入图片 插入图片

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Appearance
5

DVD Power indicator

USB

ECG
Pencil probe

Space for video printer

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Appearance
6

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Configuration
7

Ultrasound System Configuration


` peripheral remark
1 monitor Standard, 21.5 inches
2 touch scree Standard, 13.3 inches
3 Touch panel Standard, up and down, left and right rotation
4 Probe port Standard, 4 ports
5 Built-in hard drive Standard,1T
6 Built-in DVD burning optical drive components Pre-factory selection
7 Built-in ECG hardware module Standard
8 Built-in WIFI module Standard
9 In-cavity probe holder Standard
10 Keyboard waterproof film Standard

This data is derived from the "P20 Series Ultrasonic Color Doppler Diagnostic System Feature
Configuration Table-A10" and, if there is an updated version, Use the latest version
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Configuration
8

Compatible probe

Models Compatible probe Standard probe

P25 3C-A、C1-6、C613、L741、L742、 3C-A、L741、3P-A、6V1


L752、12L-A、12L-B、3P-A、S1-5、
P20 3C-A、L741、3P-A、6V1
7P-B、MPTEE、6V1、6V3、6V3A、6V7、
P15 VC6-2、VE9-5、BCL10-5、6CI-A、6CT- 3C-A、L741、3P-A、6V1
A、12LI-A、12LT-A、10I2、LAP7

This data is derived from the "P20 Series Ultrasonic Color Doppler Diagnostic System Feature Configuration
Table-A10" and, if there is an updated version, Use the latest version

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Configuration
9

Abdominal shallow heart blood vessels Multi Purpose


Maternity
3C-A BCL10-5
3P-A 6V3
C1-6 10I2
S1-5 6V1
12L-A 12LI-A
7P-B 6V3A
12L-B 12LT-A
C613 6V7
L742 6CI-A
VC6-2
L741 6CT-A
VE9-5

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
10

Physical specifications of the device

 Dimensions: 690mm wide× 945mm deep× 1455mm


 high control panel lift range: 0mm to 150mm
 Weight: Approx. 86kg
 The control panel supports lifting and rotating: lift range 0-
150mm, rotation angle ± 40 degrees

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
11

Power parameters

The supply 110-240V ~


voltage

The power 50/60Hz


frequency

Power supply 500VA


power

Note: IF the power is unstable or frequently powered off, UPS (recommended power ≥ 2KVA) is required

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
12

Environmental requirements

The operating Transport and storage


environment environment

relative humidity ≤90%RH(Non- ≤95%RH(Non-


condensation) condensation)

Ambient temperature 0℃~45℃(When used -20℃~55℃


in conjunction with a 4D
probe, it is 10 to 45 ℃)
atmospheric pressure 700hPa~1060hPa 700hPa~1060hPa

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
13

Unpacking
Unpacking the machine after confirmed the anti tilt tab in good condition.
(Tools:Hexagonal wrench、scissors)

Detailed Process
Cut the packing tape on the upper cover of the outer box.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
14

Unpacking

2. Remove upper cover.

Remove the top cover

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
15

Unpacking

3. Release the buckles,remove the outer box.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
16

Unpacking

4. Checking the accessories grooves and take out the materials.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
17

Unpacking

5. Remove the foam.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
18

Unpacking

6. Remove the baffles for machine.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
19

Unpacking

7. Unlock the wheels.

Lock Unlock Position

Lock Position

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
20

Unpacking

8. Pull out the machine partly until the rear side out from the baffle,then lift the machine
down to the ground.

Don't grab this handle to


lift the machine.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
21

Wires Connection Tips

• Using good quality power strip and power outlet


• Ensure the well grounding condition,voltage between zero line and grounding line less than 1V
• Improper operation may cause:electric shock、fire、machine hanging、hardware failures and
artifacts……

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
22

Adjust the monitor arm

Move the monitor arm up and down to


adjust the height of the display.
Moving the monitor arm left and right
adjusts the horizontal direction of the
display.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
23

Adjust the monitor

Grasp the upper and lower sides of the monitor and


tilt it back and forth Grasp the left and right sides of the monitor and
rotate left and right
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
24

Adjust the control panel

Adjust the height of the control panel


Press and hold the lift switch and the control panel
rises automatically. Release the lift switch when you
are in the right position.
While holding down the lift switch, press the control
panel vertically to reduce its height. Release the lift
switch and stop pressing down when adjusted to the
appropriate position.

Adjust the horizontal position of the control panel


Press and hold the turn switch while moving the
control panel handle left to right to adjust the control
panel in the left and right directions. When adjusted to
the appropriate position, release the turn switch.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
System Overview - Installation
25

The connection of the probe

Connect the probe


As shown, the cable is facing up, turning the switch on the probe connector
counterclockwise, inserting the probe connector into the probe slot and pressing it tightly,
rotating the switch on the probe connector clockwise to lock the probe connector
Disconnect the probe
Rotate the switch on the probe connector counterclockwise to unplug the probe
connector

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Contents
26

1 System Overview

2 Software

3 Hardware

4 Troubleshooting

5 Peripherals Connection

6 Preventive Maintenance

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
27

 The main interface


 Touch screen interface
 Key function
Software system
 Common settings
 Software upgrades
 Software burning changes

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - main interface
28
Number name Number Name

1 Manufacturer Logo 8 Tick bar and focus


2 The information bar 9 Status prompt bar

3 The application site 10 Ultrasound image


icon area
4 The measurement 11 Trackball hint area
result box
5 clipboard 12 Movie playback
progress bar (appears
when frozen)

6 Color and gray bars 13 Measurement/comm


ent area
7 The temperature 14 The parameter area
display area

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - touch screen interface
29

Numb Name Function


er
1 touch screen In different states, the touch screen displays different interfaces, including diagnostic selection
interface, imaging mode interface, annotation/body label selection interface, measurement selection
interface, and segment navigation interface
2 Function knob In the imaging selection interface, rotate the knob to change the parameter value for the position at
the bottom of the touch screen. Typically, counterclockwise rotation reduces the value of the
corresponding parameter (or turns off the function for the key), and clockwise rotation increases the
value of the corresponding parameter (or turns on the function corresponding to the soft key)
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - touch screen interface
30

Diagnostic selection interface

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - touch screen interface
31

Imaging mode interface

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - key function
32

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - key function
33

Number symbol Name function


1 Probe Probe key Select the probe and application
2 Patient Patient information Enter the new patient interface
key
3 Review The preview key • When no new patients are established, open the patient checklist interface
• When a new patient is established, go to the View Image interfacre

4 Report The report key Go to the patient reporting interface


5 Body Mark The position marker Add a position marker to the ultrasound image
key
6 End exam End diagnostic key Stop diagnosis of the current patient
7 Angle Angle knob • In PW/CW mode, the cursor is used to adjust the direction of blood flow
• In M mode, it is used to adjust the angle of the anatomical M-type sampling line
• When the arrow comment function is activated, it is used to adjust the direction of
the arrow
• Used to adjust the probe direction after the position marking function is activated
• When performing pediatric measurements, it is used to adjust the measurement
angle of the hip joint

8 Clear Clear Key Clear all measurements, notes, and bodymarks on the screen
9 Arrow arrow key Add arrow marks to the ultrasound image

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - key function
34

Numb Symbol Name Function


er
10 Video Print Video print key Triggers the video printer to print the current screen

11 Print Print key Trigger printer printing

12 Annot Annotation Key Add text comments to the scanned image

13 - The confirmation key Confirm the current operation

14 Pointer Trackball function • Activate the clipboard in 2D (B/CFM/PDI/TDI) and M mode


conversion key • In 3D and 4D mode, activate the trackball cursor
15 - Trackball Used to move the cursor, adjust the position and size of the sampling frame, and more

16 Update Refresh key • Activate the M image in M mode or the PW/CW spectrum image in PW/CW mode
• Used to switch measurement endpoints when making distance and ellipse
measurements
• Used to roll back modifications when performing trace measurements
17 Calc Apply measurement Activate measurements and calculations
key
18 - The confirmation key Confirm the current operation

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - key function
35

Num Symbol Name Function


ber
19 Caliper Underlying Activate the underlying measurements and calculations
measurement key In 2D (B/CFM/PDI/TDI), M, 3D, and 4D modes, distance measurement is
activated
In Spectrum Doppler (PW/CW) mode, speed measurements are activated
20 Freeze key Freeze the image while scanning in real time;

21 Auto Auto-optimize keys Automatically optimize images

22 Cine Movie save key In the frozen state, save the movie

23 Image The image save key • Save pictures in real-time or frozen state
• In the real-time state of the shadow mode, press this key to start the real-
time storage of blood flow perfusion process movies, and then press this key
to complete the movie storage
24 Depth Depth knob • Rotate to adjust the depth of the scan
• Press this knob to restore the default depth of the image
25 Zoom Zoom Key Zoom in on an adjustable area on the ultrasound image

26 TGC TGC Sliding rod Contains an 8-time gain compensation value that increases or decreases at a
specific depth by adjusting the slider

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - key function
36

Num symbol Name Function


ber
27 Dual dual-frame display • Switch to dual-frame display mode
key • Used to switch images in dual-frame display mode

28 B B mode knob • Press this button to activate B mode
• Clockwise rotation increases the B gain, and counterclockwise rotation
reduces the B gain

29 CFM CFM Mode knob • Press this button to enter or exit CFM mode
• Clockwise rotation increases the CFM gain, and counterclockwise
rotation reduces the CFM gain
• In 3D/4D imaging mode, rotate this button to pan the 3D image left
and right

30 PW PW Mode knob • Press this button to enter PW pre-activation mode and then press
Update to enter PW mode
• Clockwise rotation increases the spectrum Doppler gain, and
counterclockwise rotation reduces the spectrum Doppler gain
31 CW CW mode knob • Press this button to enter CW pre-activation mode, and then press
Update to enter CW mode
• Clockwise rotation increases the spectrum Doppler gain, and
counterclockwise rotation reduces the spectrum Doppler gain
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - key function
37

Num Symbol Name function


ber
32 M M Mode knob • Press this button to enter M pre-activated mode, and then press
Update to enter M mode
• Clockwise rotation increases the M gain and rotates counterclockwise
to reduce the M gain
33 3D/4D 3D/4D Key Enter or exit 3D/4D mode

34 Contrast The contrast Enter or exit contrast imaging mode


imaging key
35 Elasto Elastic imaging Enter or exit elastic imaging mode
keys
36 Keyboard keys Press this key to enter the touch screen button interface, and press this key
again to return to the previous interface.
37 - Function knob In the imaging selection interface, rotate the knob to change the
parameter value for the position at the bottom of the touch screen.
Typically, counterclockwise rotation reduces the value of the corresponding
parameter (or turns off the function corresponding to the soft key) and
clockwise rotate increases the value of the corresponding parameter (or
turns on the function corresponding to the soft key)

38 Setup Setup key Press this key to enter the system setup interface

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - number/alphabet keyboard
38

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - number/alphabet keyboard
39

Numb Symbol Name Function


er
1 P3 User-defined keys You can store images, movies, body data, or turn on face navigation.

2 P4

3 Setup Setup key Press this key to enter the system setup interface

4 Help - Reserve

5 Biopsy Biopsy key Turn the biopsy function on or off

6 Pan Widescape imaging key Enter or exit real-time widespectimal imaging mode

7 PDI Energy Doppler mode key Enter or exit PDI mode

8 TDI Tissue Doppler mode key Enter or exit TDI mode

9 Flip up and down Flip the image up and down

10 Flip left and right Flip the image left and right

11 THI Harmonic mode key Enter or exit THI/PHI mode

12 Invert Invert key • In CFM/TDI/PDI mode, turn on blood flow reversal.


• In PW/CW mode, spectrum inversion is turned on.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software system - number/alphabet keyboard
40

Nu Symbol Name Function


mb
er
13 Prtscr - Reserved
14 Del Delete key • When in the comment state, delete the previous
comment or bodymark.
• Delete the last measurement mark and measurement
results during manual tracing or when measuring (not
the measurement data in the measurement report).

15 - Direction key • When you set up a system menu, select the menu
option
• Adjust the PW mode volume and display brightness
when combined with the Fn key
16 Scr-zoom Zoom key Zoom in on an adjustable area on the ultrasound image
17 Alt GR - Used in combination with other keys, you can enter
multilingual characters

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
41

System settings - General - General settings

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
42

System settings - General - display

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
43

System settings - General - Storage

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
44

System settings - general - keystroke function

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
45

System settings - General - Biospsy

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
46

System settings - peripherals

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
47

System settings - comments

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
48

System settings - position marking

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
49

System settings -measurement-basic measurement

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
50

System settings - measurement - Application settings

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
51

System settings -measurement-list

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
52

System settings - touch screen

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
53

System settings - reporting

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
54

System settings - DICOM - image storage

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
55

System settings - DICOM - commitment storage

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
56

System settings - DICOM - Workslist

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
57

System settings - DICOM-MPPS

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
58

System settings - DICOM-printing

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
59

System settings - DICOM-QR List

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
60

System settings - Load Default - Load

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
61

System settings - load default - export/import

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
62

System settings - User security - User settings

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
63

System Settings - User Security - Administrator Settings

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software systems - common settings
64

System Settings - System Information - Acoustic Output

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
65

Upgrade preparation

Touch screen upgrade

Keyboard upgrade

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
66

Upgrade preparation
Software
CN, software
SonoScape config,
version, SN.
Service Dep. Keypass.

Kernel USB-13.7

Bootable USB
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
67

Upgrade preparation
Copy both files to the update folder.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
68

Touch screen upgrade


Copy the touch screen upgrade file to an empty USB disk, then plug the USB disk into
the USB port at the bottom of the keyboard in the red box, and press the Reset button
In about 5 seconds, you will see the following image. Touch screen upgrades are
required for each software upgrade, including a simple upgrade.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
69

Touch screen upgrade

Press system updates, and then wait a few minutes.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
70

Touch screen upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
71

Touch screen upgrade

When the touch screen upgrade is complete, turn off the machine for
the upgrade.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
72

Touch screen upgrade

Check the upgraded software version here.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
73

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
74

Kernel upgrade
1. Plug the safety mode USB drive and external keyboard into the USB port of the
machine
2. Press the power-on key, press the Delete button on the external keyboard
continuously, and you will be prompted to enter your password: "service" and then press
"Enter"

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
75

Kernel upgrade
3. Set the first startup priority to U disk start
Select the inserted USB drive in Boot-> USB Key Drive BBS Priorities-> Boot option #1,
and set to start from the USB drive

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
76

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
77

Kernel upgrade
4. 【 Save&Exit 】->【 Save Changes and Exit 】->【 Yes 】,SAVE AND EXIT TO
ENTER THE SAFE MODE BRUSH INTERFACE

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
78

Kernel upgrade

After entering this interface, enter forcekill in the


figure and press Enter to run the program.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
79

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
80

Kernel upgrade

When all the procedures


are completed, you will
be prompted to press
any key to shut down. At
this time, you need to
press any key on the
external keyboard. After
shutting down, unplug
the external keyboard
and USB flash drive.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
81

Kernel upgrade

Please wait a moment here.


When the software installation
is complete, the machine will
automatically shut down.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
82

Kernel upgrade

Please wait here for a moment.


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
83

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
84

Kernel upgrade

When the kernel is updated, an alert appears showing that the license
expires within 0 days because there is no profile (config)
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
85

Kernel upgrade

Because
there is no
configuratio
n (config),
the
functionality
in the device
is not
activated

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
86

Kernel upgrade

Press in turn“ shift”,“ E”,“ setup(menu)”

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
87

Kernel upgrade

Enter your annual password here to install the profile.


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
88

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
89

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
90

Kernel upgrade

Restart the device and complete the aconfig file load.


When the profile is loaded, the feature is activated.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
91

Kernel upgrade

Press in turn“ shift”,“ E”,“ setup(menu)”

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
92

Kernel upgrade

Enter your monthly password here to install the


keypass file.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
93

Kernel upgrade

Click “Probe Table In”

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
94

Kernel upgrade

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
95

Kernel upgrade

Press Freeze and enter your monthly password, then enter Engineer mode.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
96

Kernel upgrade

Then press“ Engineer Config”


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
97

Kernel upgrade

Press "i" on the keyboard


and you'll see this
message on the screen.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
98

Easy upgrade

Copy the two files to an empty USB disk and plug the USB into a USB socket

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade
99

Easy upgrade

Press in turn“ shift”,“ E”,“ setup(menu)”

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade 10
0

Easy upgrade

Enter the year/month password here and install the software file.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade 10
1

Easy upgrade

Restart the machine

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Software System - Software Upgrade 10
2

Easy upgrade

After the machine is turned off, the update is complete.


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Contents 10
3

1 System Overview

2 Software

3 Hardware

4 Troubleshooting

5 Peripherals Connection

6 Preventive Maintenance

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
10
4

 System schematic
 I/O module
Hardware system  PC module
 Ultrasound module
 Power supply module

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Hardware System- System schematic 10
5

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Monitor 10
6

1. 21.5-inch LED LCD monitor


2. resolution: 1920*1080
3. The signal type: HDMI
4. Monitor power voltage: 12V

5. Touch screen: 13.1 inches.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Monitor 10
7

Power supply HDMI

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
The current and signal flow of the Monitor 10
8

21.5’’ Monitor

HDMI Invert to
LVDS

Internal IO board Grap


hics
Card current
Power Signal
supply
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
keyboard 10
9

Keyboard control
panel

Drop-down PC
keyboard

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Keyboard board
Connect
the touch Connect the 11
screen 0
speakers
Connect the Connect the knob control
IO board board panel

Connect the
speakers

Connect Connect the


the USB PC keyboard
port

WIFI
module
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
PC keyboard 11
1

Connect to the
keyboard board

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Touch screen control panel 11
Connect the 2
Connect the IO board
Connect the keyboard board touch screen

Connect the
USB port

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Power current and signal flow 11
3

13.1’ LCD
 USB ports (upgrades)
LVDS Signal input port are only used to update
touch screen software
touch
screen LVDS signal output port
USB Upgrade port
Touch screen port

control panel Reset button

current
Signal Keyboard board USB port

mainboard
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
PC module 11
4

mainboard
Internal IO board Mainboard component

Graphics card
memory
HDD hard disk
DVD Driver
External IO Board 1
External IO board External IO Board 2

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
The board unit component
11
5

mainboard +graphics card + Internal IO board


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
mainboard
11
6
SATA power

RS232 USB 3.0 ports

VGA ports

USB ports
LAN port
To HDD&
S-video DVD ROM

Audio 12V
input
USB output
Power switch
& 5V standby
To video card
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Graphics Card
11
7

Video output
The internal IO board
provides 12V voltage

Connect the mainboard

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Internal IO board 1
11
The dongle 8
USB Port interface
PS/2 The power switch port
Connect to
of the mainboard
audio

5V SB to
Mainboard

12Vto the motherboard and To RS232


graphics card
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Internal IO board 2
11
9
Connect video printer (power supply & signal) Connect the monitor (power supply)
Connect the power management board

Connect the
management
board
Connect the touch
screen control panel

Connect the keyboard


ATX power input port board (audio & USB)

Connect the
4D function control keyboard board
interface
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
External IO board 12
0

External IO board 1

External IO board 2

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Ultrasound module 12
1

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
MPC board 12
2
Connect the
Connect the
SBF board
pen probe

Probe
connector back

function:
1. Probe connection
and detection
2. Probe switching
front

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
SBF board
12
3
Connect the Connect
MPC board VHV board

back

function:
1. Probe identification
2. Ultrasonic digital signal
processing
3. Communicate with the main board
front 4. ultrasound signal is generated
USB connected to
the main board
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
VHV board 12
4
function:
Connect the 1. Provide a variety of
ATX power power supplies for
supply ultrasonic systems;
2. Control ECG and 4D.
4D
Connect the SBF board

fron
back t
ECG
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Hardware - Power Module 12
5

Power Flow
Power
Management Battery
Board
110 ~ 220V
AC
5V 5V 12V
3.3V 5V
HV
Isolation 12V 12V
ATX Power VHV Board SBF Board MPC Board
Transformor 110V
5V
110V 12V
5V
5V 12V Touch Panel 5V 12V
Keyboard Touch Panel
12V Control Board

Video Printer Internal IO 5V 12V 5V


Board Main Board HDD 文字
文字
12V
12V
Graphic Card
monitor
Power
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
ATX Power Supply - Transformer 12
6

ATX power supply

transformer

The P20 battery can support the device for approximately 1.5 hours.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
ATX Power Supply - Transformer 12
7

ATX power supply Isolated transformer


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
battery 12
8

Battery monitoring
Battery management board batter
board
y

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Content 12
9

1 System Overview

2 Software

3 Hardware

4 Troubleshooting

5 Peripherals Connection

6 Preventive Maintenance

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
13
0

Power failure
Boot failure
Abnormal display
Troubleshooting Dark lines on the ultrasound image
No initial wave
Keyboard failure
External interference

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Power On Failure 13
1

Troubleshooting process

Power N Main Y Y Y
Power on UPS FUSE
indicator power
failure normal normal
lighted normal

N N N

Main power
UPS Failure FUSE failure
issue

Short-circuit issue Y Main Y ATX Y Isolation


with the load patch board Power transforme
of main board normal normal r

N N N

Isolation
Main board ATX Power
transformer
failure failure
failure

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
故障分析-上电失败 13
2

Phenomenon Analysis and Checking Steps

1. Check the main power condition to confirm it is stable in the range of 110V/220V±10%,
otherwise contact the technician to manage.
2. Check UPS working status also can connect machine to main power directly to eliminate
UPS failure.
Power On Failure
3. Check FUSE.
4. Check ATX power without isolation transformor, to eliminate isolation failure.
5. Check ATX power output with and without load patch to eliminate ATX power failure.
文字
6. Force motherboard to boot to eliminate the Keyboard/power on cables issue.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Power On Failure 13
3

Step 1-Main Power Checking


Check the main power condition to confirm it is stable in the range of 110V/220V±10%.

Step 2-UPS Checking

Check the UPS is in working condition. Can also connect the machine to main power directly to check.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Power On Failure 13
4

Step 3-Fuse Checking

文字

Replace the FUSE to check machine can power on or not.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Power On Failure 13
5

Step 4-Isolation Transformor Checking

110V

Main Power
文字

1. 使用万用表测量隔离变压器给ATX电源的110V电压是否正常
2. 市电直接连到ATX电源开机,如果能够启动则隔离变压器故障
警告:测试交流电路时,使用万用表交流档位测量,注意安全防护,手不能触摸到交流电及与交流电接触的
金属部分,包括万用表表头金属部分
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Power On Failure 13
6
Step 5-ATX Power Checking

Purple:
5VSB Yellow:
12V
Yellow:
12V Black:Grounding

Red:5V Red:5V

Orange:
3.3V
Green:PS_ON Signal

Disconnect the load patch:Connect the ATX power to main power inout,short circuit the 4 PINS of
文字 the
cables connected to the power management board,then short circuit the black and green lines to force
the ATX power to start then use multi-meter to check the outputs of the ATX power.

Connect the load patch:Connect the ATX power to main power inout,short circuit the 4 PINS of the
cables connected to the power management board,then short circuit the black and green lines to force
the ATX power to start then use multi-meter to check the outputs of the ATX power.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Power On Failure 13
7

Step 6-Main Board Checking

Short circuit the left-right 3rd and


4th pins of J23 to force boot the
main board.
文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 13
8

Phenomenon Analysis and Checking Steps

Main monitor stay black,


touch panel stay in 1. Main board didn't boot try force boot.
SonoScape page

Show Boot media error


1. Check HDD SATA connection, check the 5V power to HDD;
message,touch panel stay in
2. Enter BIOS,check boot media detected or not;
SonoScape page

Can enter scan mode but


touch panel has no display 1. Touch panel display monitor issue.
but can got command 文字

Main monitor and touch panel


1. Try to enter touch panel safe mode, if can't enter suspect
stay in SonoScape page, can't
touch panel and control board failure.
enter scan mode

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 13
9
Phenomenon Analysis and Checking Steps

Can't enter scan mode show


orange screen with error
1. Try to enter touch panel safe mode, if can't enter suspect touch panel
code“0x5453 0x0001”
control board failure;
(Touch panel communication
2. Reinstall touch panel software;
failed)
3. Check touch panel signal circuit cables connection;

Can't enter scan mode show


orange screen with error code
文字
“0x5553 0x0001”,“0x5553 1. Check the USB cable between main board and SBF board;
0x0002”,“0x5453 0x0002” 2. Check the ATX power outputs to VHV 5V/12V;
(SBF Board communication 3. Check VHV test points voltage;
failed)

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
0

Boot failure

Reinstall
Software N Touch Y
HDD Problem
panel
Replacement solved
Safemode
N

Problem Y N
solved

Y
Touch Panel Main Board
HDD Issue
Issue Issue
Software
Issue

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
1

Shutdown during
booting

Ultrasoun
d and PC SBF
Before modules Reconn Board Power Y
Solved
Step 4 communi ect USB or normal
cation SMPS
issue
Y Y

USB Power SBF板或


软件或主板问题
Connection Issue 主板问题

①:详见149页
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
2

Error code“0x5453 0x0001”


Enter BIOS directly when booting

HDD not detected Error code“0x5553 0x0001


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
3

①Booting BIOS ③Start service programs ⑤ Start touch panel softare

②Load kernel ④Start Application Softare

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
4

Software

Every
Cables
time
Orange
Hardware
Screen
Occassio
Cables
nally
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
5

Check Ethernet route

Cable1 Cable2
Touch Cable3

Main Internal Contro Keyboar


Board IO l Board d
Board
文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
6

Check Ethernet route Cable1

Main Internal
Board IO
Board

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
7

Check Ethernet route Cable2


Touch
Internal Contro
IO l Board
Board

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Boot Failure 14
8

Check Ethernet route Cable3


Touch
Contro Keyboard
l Board

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Monitor Display Abnormal 14
9

Phenomenon Analysis and testing

1. 1. Unplug the signal cable of the monitor and connect it directly to the
external monitor to test whether the graphics card is normal
2. 2. Use a notebook computer to output HDMI signals and connect to the
machine monitor to test whether the machine monitor is normal
3. 3. Check whether the 1-2 section HDMI connection part is crimped;use
Black Screen or
laptop directly connected to the HDMI connector at the monitor arm.
Colorful Screen
4. 4. Plug and unplug the HDMI port of the monitor, and loosen the HDMI
bolt(Troubleshoot poor contact on the display port)
5. The troubleshooting sequence is for reference only, and can be handled 文字
flexibly according to the on-site situation

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Monitor Display Abnormal 15
0

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Black Band in Image 15
1

Troubleshooting process

Y Check
All
Black Band Ultrasound
Probes
Module

Check Probe
文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Dark Lines on Image 15
2
Phenomenon Analysis and Checking Steps

1. Clean the pins on the probe matching box and probe base with a brush;

2. Check the pins of each probe base. If the pins are oxidized or damaged,
replace the probe board;

3. Unplug and insert the probe again and check whether the connection is
loose. If the fault is eliminated by gently pressing by hand, there is a problem
of looseness. Use different probes to conduct cross test on each probe base;
Dark Lines on
Image
4. Reinsert the probe without looseness. Use different probes to conduct cross
test on each probe base. If only one probe shows the dark lines on the
ultrasound image, replace the probe; 文字

5. Use different probes to conduct cross test on each probe base. If all probe
bases have black lines or missing images, prepare the probe board and EBF
board and replace them according to the situation on site.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Dark Lines on Image 15
3

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-No Initial Wave 15
4

Troubleshooting process

Y Check
All
No Initial Wave Ultrasound
Probes
Module

Check Probe
文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-No Initial Wave 15
5

Phenomenon Analysis and testing

1. Shutdown the machine for 5 mins to eliminate the voltage protect status and
restart the machine. Enter engineer mode do HV test to check the HV status.
2. Check whether there is dust and abnormal PINS in probe socket and connector.
3. Check probe seperately if there is problem in some single probe then replace it.
4. Open ultrasound module keep all cables connected and check the indicators
No Initial Wave
status on SBF board to eliminate the SBF board failure.
5. Check the power inout form ATX power to VHV board is normal or not.
6. Open ultrasound module keep all cables connected and check the outputs of 文字
VHV board to eliminate the VHV board failure.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-No Initial Wave 15
6

文字

探头无始波

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-No Initial Wave 15
7

Self Checking in Engineer Mode


In the system information(About) interface, press the freeze key, enter the monthly password / yearly password, press
enter to get into the engineering mode, check "ProbeID, HV&Temp" in equipment diagnosis, click PLAY, and then view the test
results.

文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-No Initial Wave 15
8

SBF Board communication abnormal-Check the input of VHV board

Use multi-meter to check whether


the input 5V/12V voltage is normal.
文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-No Initial Wave 15
9

SBF Board communication abnormal-Check the output of VHV board

Use multi-meter to check the


voltages of the test points.

文字

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Control Panel Failure 16
0

Phenomenon Analysis and testing

Keyboard test first to confirm the abnormal button,check the button keys physical
PC keyboard keys damage or not and clean the dust. If this step not working then disconnect the PC
sticky keyboard do keyboard test, if works then PC keyboard issue otherwise keyboard PCB
failure.
Keyboard test first to confirm the abnormal button,check the button keys physical
No response from damage or not and clean the dust. If this step not working then disconnect the PC
some button keyboard do keyboard test, if works then PC keyboard issue otherwise keyboard PCB
failure.
Button stuck or
1. Check silicon pad bounce or not;2. Check whether the cover moved to block the
doesn't bounce
buttons;
back
文字
Trackball cursor
Open trackball ring and clean it.
jumps

Trackball stuck Open trackball ring and clean it.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Control Panel Failure 16
1

Phenomenon Analysis and testing

Touch panel light


Replace touch panel.
leaking

No response from 1. Clean touch panel;2. Check touch panel test;3. Check touch panel cables
touch panel connections;

Machine doesn't
response after Touch function is working, machine system doesn't receive signal.
touch
Touch panel display1. Cehck the software version same with machine or not;2. Try reinstall 文字
abnormal software;
1. Click touch panel to check machine system can response or not,if it respones
Machine booted
then try reinstall software otherwise cross check touch panel and touch control
but touch panel
board;
has no display
2. Check the touch panel related cables connection;
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Control Panel Failure 16
2
Control panel
failure

插入图片

Button N Engineer Mode


Backlight
Sticky? Keyboard Test
Y
Y

Engineer Mode
Keyboard Test N

N
Normal

Button Power to
Sticky? Keyboard Issue
Y
Y

Keyboard/PC Clean keyboard, Keyboard/PC


keyboard Issue check connection keyboard Issue

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Control Panel Failure 16
3

Troubleshooting Process Steps

Keyboard test first to confirm the abnormal


Keyboard button,check the button keys physical
damage or not and clean the dust. If this step
Only Several Keys Abnormal not working then disconnect the PC keyboard
do keyboard test, if works then PC keyboard
PC Keyboard issue otherwise keyboard PCB failure.

Check machine time runing or not, if it hanging


then may be software issue, if it runing then
Software check the cables connection, do cables
Whole control panel Hardware reconnect to check, if problem still here then
suspect keyboard PCB issue.
Cables

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-Control Panel Failure 16
4

Keyboard Test

文字

1. Press the reset button to enter input interface.


2. Click【Edit Enable】,Input command “ts.test”,click【OK】to enter keyboard engineer mode,click
【Keyboard Test】to enter keyboard test, test key/buttons and check the results.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-External interference 16
5

文字

Image interference

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-External interference 16
6

Troubleshooting process
Image
interference

Check the mains Is there a No


new type
supply and of probe?
ground the
machine Yes

Slight external Yes Problem Replaced with the


interference solved? new shielded process
probe
No

Install anti- No Try changing the


Problem
interference cable solved? operating
文字 environment
Yes
Yes
External Medium Problem Strong external
interference solved?
interference

No

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-External interference 16
7

Phenomenon Analysis and testing

1. Understand the time period of interference, whether it is necessary or accidental;

2. Confirm which probes have interference and which modes (two-dimensional, blood flow and other
modes);

3. Observe the surrounding environment of the machine and whether there are large equipment or
signal transmitting tower nearby. You can try to move the machine to another place for testing;

4. Use a multimeter to measure whether the mains voltage and zero ground voltage are normal (mains
voltage 220V ± 10%, 0V < zero ground voltage ≤ 2V);
External interference
5. Check the peripherals connected to the equipment, turn off all other electrical equipment such as UPS,
voltage regulator or workstation, disconnect the power supply, and eliminate the interference sources
one by one; 文字
6. If it is the interference caused by peripheral workstation, add anti-interference power line at the
computer end, or add magnetic ring on DVI and network transmission line. If it is video / S-Video
video output, its anti-interference line can be used;

7. Use a new shielded process probe;

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-External interference 16
8

Supplementary notes

1. External interference can have weak and strong types, and corresponding measures shall be added
according to the strength

2. If there is strong interference, try to use the same power plug to supply power to peripheral workstation
equipment and ultrasonic host, or use high-quality ups

3. For peripheral workstations (including PC motherboard, workstation, display, electric bed, etc.), add AC
power filter box or magnetic ring under the power supply line according to the severity of interference

4. For the signal line between the peripheral working end class and the ultrasonic host (such as S-Video /
CVBS) and the signal line connected with the ultrasonic, add an AC filter box, or add a low-frequency
specified magnetic ring filter box (for the interference of the peripheral host itself) 文字

5. If the above conditions cannot meet the requirements, use a new process probe (fully shielded process
probe)

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-External interference 16
9

Common scenarios (4)


外设工作站类 1. For the peripheral workstation (4), the low-
AC电源 电源线 (工作站、PC主机、 frequency interference of itself shall be good. For
滤波盒
example, the interference generated by the
显示外设,电动床
peripheral workstation itself, the treatment of S-
等) Video / video / VGA / HDMI / DVI / network cable
(2) shall be strengthened (5). If the measure (5) is
used, the measure (2) can be weakened or not
used according to the situation

(1)
AC电源
插排
(5)
S-VIDEO/VIDEO线(滤波盒)或
2. If there is an AC filter board (6) in the ultrasound
main system, the measure (3) may not be used
网线,HDMI,DVI,VGA(增加绕制磁环 according to the situation

3. If the new process probes can be introduced and


the process is reliable, the measures at positions
文字
(2), (3), (4) and (5) can be weakened or not added

(3) (6)
4. Use the grounding column conductor specified
by the whole machine as appropriate
AC电源
滤波盒
电源线
超声主机

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Troubleshooting-External interference 17
0

CFM Mode Interference B Mode Interference

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Contents 17
1

1 System Overview

2 Software

3 Hardware

4 Troubleshooting

5 Peripherals Connection

6 Preventive Maintenance

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
17
2

 Connect external monitor


 Connect printer
Peripheral connection
Connect DICOM
Connect PC
Plug-and-play peripherals

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
introduce 17
3

This PPT is designed to inform our customers about the external connections of the P20.
As we all know, doctors always like to connect some external equipment, such as external
monitors, printers, USB disks, foot switches, PACS systems, etc.
Now, we'll cover the connection process for these projects above.


• External monitor (TV, VGA, HDMI display);
• Printer (USB, video, Windows);
• DICOM connection (sharing files in ultrasound devices);
• PC connection;
• Foot switch connection;
• USB disk, keyboard and mouse connection.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
17
4

External display (TV, VGA, HDMI)

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
TV (BNC and S video ports) 17
5

For TV monitors, the TV signal is an analog signal; The port we can use is the S video port when we
want to connect to the TV via S video;
We need to do some setup on the P20, we need to prepare the S video cable, then follow below
settings:

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
TV (BNC and S video ports) 17
6

Note that if the external monitor has an S-video port,


simply plug the cable into it; If the external display
does not have an S-video port, for example a Philips
display, the male connector must be inserted into the
CVBS port instead of the "L" or "R" port.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
TV (BNC and S video ports) 17
7

Then we need to click Save and Exit, and then


restart the device.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
VGA Display (VGA Port) 17
8

When connecting the unit to a VGA display, the following settings are required:

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
VGA Display (VGA Port) 17
9

Select the “VGA option” and “Save and Exit” (the resolution is determined by the resolution of
the display, and the resolution you select on your device must not exceed the resolution of the
external monitor), and then restart the device to complete the setup.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
HDMI display (HDMI port) 18
0

When we want to connect this machine to the HDMI display;


We need to prepare an HDMI cable first.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
HDMI display (HDMI port) 18
1

Select the desired resolution in the HDMI option, click “Save and Exit”, and then restart the
device.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
18
2

Printers (USB, Video, Windows)

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
18
3

note:
The printer brands that machines can support are as follows:

• HP (with PCL3, PCL5, PCL6 printing language)


• Sony-UPD Series (UPD897,898)
• Mitsubishi (P95D);
• Canon Selphy 900 and 1200;

USB printer connection


• Get an empty USB disk;
• Printer model;
• Driver (obtained from sooscape Services)

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
USB printer connection 18
4

Connection steps (Sony UPX898MD, for example)


Copy the driver to an empty USB disk and change the name of the driver: deskjet;

 Turn on P20 and printer;


 Connect the printer to the USB port of the P20;
 Plug the USB disk with the driver into the USB port;

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
USB printer connection 18
5

Select “Settings” and the“ Peripheral option”, “USB Printer” and“ Add Printer”;

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
USB printer connection 18
6

 Select Save and Exit.


 Wait for the system to add a printer
 Then when the installation is complete, you can print

the pictures and reports you want.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Video printer connection 18
7

 First, we need to prepare the printer, remote control cable, S-video cable, and then connect the
printer and machine through the remote control cable and S-video cable, as shown in the
following image;

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Video printer connection 18
8

 Then, in the Peripherals interface, make the


settings as shown in the image above. Once
you've completed these steps, select Save and Exit.
Then you can print the picture as needed.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Windows printer connection 18
9
• Part 1. Printer connection

Connecting the printer to the computer via a local area


network cable (or WIFI);

• Part 2. Install the printer driver

First, you can apply from our engineers for a package


like this,

Then install the printer driver on your computer, run


the printer driver installer, and select a location to
create a shortcut.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Windows printer connection 19
0

As you'll see from the prompts, you'll see this interface, and remember to select Create Desktop
Icons, as shown below:
After you install the printer driver, you'll find a new icon

On the desktop, click the icon to run the program.

The program runs in the background, so when you want to


print some pictures, your PC issues commands to its default
printer to print pictures as needed, and more information about
how to set the printer as the default printer for your PC is as
follows.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Windows printer connection 19
1

 Part 3: Connect the SonoScape color Doppler


to your computer.

Check the local static IP address on the PC,


which will be used during transmission.
You can then go to the ultrasound's system
setup interface and fill in the IP address, network
mask and default gateway. Make sure that the IP
address of the ultrasound belongs to the same
network segment as the PC; The network mask
and default gateway should be the same as your
PC.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Windows printer connection 19
2

After you complete these steps, you can click Ping


and wait a few seconds for the screen to display Ping
Success. If it does not appear, check the connection
between the machines and reset the settings.
Then click Save and Exit.
Press the print key on the machine, and if you do not
operate the computer, the printer agent will
automatically process the order and command the
default printer to print the picture.
Note: The default printer here is the printer on your
PC that has the first priority, not the default printer
for the ultrasound.
For example, we have a PC with a Windows 7 system
where we have a driver for an HP laser printer
installed and set the printer as the default printer
that you can use when you press the Print key on
your computer.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Windows printer connection 19
3

 Part 4: Print

After completing the above procedure, your printer


can print Freeze Images, Stored Images, and
Diagnostic Reports. Pressing the print key on the
machine and not operating the computer, the printer
agent may automatically process the order and
command the default printer to print.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 19
4

Before you connect, you need to make sure that the DICOM feature is turned on.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 19
5

• Here are the steps:

• Install ONIS on your PC;


• Connect the P20 to the DICOM server via a LAN
cable, or the P20 and DICOM server via WIFI;
• Once the connection is complete, we need to do
some setup on the DICOM server and P20.
Follow the two screenshots below to set up P20.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 19
6

In the image above, note that the IP address of the


machine should be in the same IP segment as your PC,
and the network mask and default gateway should be
the same as your PC. After entering this information, you
can ping your PC's IP address and then wait a few
seconds, it will show "Ping successfully", then you can
click "Save and exit" and then need to make some
settings in ONIS;

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 19
7

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 19
8

Then in the MACHINE's "DICOM" interface, you need


to fill in some information,

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 19
9

In the fourth step, if all


settings are correct, "echo
success" will appear on the
screen. Then click "save-exit"
to send the picture to the
DICOM server.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 20
0

Then in the DICOM server interface, you can click


Refresh the server, and if you want to see the picture clearly, double-click the folder
in Step 2, and you'll see the following picture:
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM connection 20
1

In addition, this guide is only for DICOM server ONIS 2.5, for other DICOM servers, they only
use a similar principle, if you have any questions, please contact us.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM MPPS connection 20
2

 When the DICOM local setup is complete, select


DICOM.
 Select MPPS in the DICOM setup interface;
 Ask the hospital for the IP address, port number,
and AE title of the DICOM server.
 When you're done with all of these settings, click
Save and Exit.
 Click Ping and Echo and if the communication
with the DICOM server is successful, the message
"SUCCESS" is notified.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
DICOM Worklist Connection 20
3

 When the DICOM local setup is complete, select


DICOM.
 Select To-Do List in the DICOM setup interface.
 Ask the hospital for the IP address, port number,
and AE title of the DICOM server.
 After completing all of the above settings, click
Save and exit;
 Click Ping and Echo, and if the communication
with the DICOM server is successful, the message
"SUCCESS" pops up, and you can process the data
to the server through DICOM.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
PC connection 20
4
Before you connecting your PC to an ultrasound, you need to prepare a LAN cable, and after you connect both
ends, you need to set up some settings in the ultrasound.
Follow these steps:

In this illustration, you need to set the native IP


address to the same network segment as your PC, the
network mask and default gateway should be the same
as your PC, and then select Save and Exit.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
PC connection 20
5

As a first step, you need to click the Start menu icon in your PC and enter the IP address of P20. For
example:

The second step is to enter the IP address of P20 directly into your computer's file browser.

In the third step, press win and R, and then type the IP address in the blank space.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
PC connection 20
6

After you connect your computer to an


ultrasound. It will display this interface:

User name: samba; Password: Vista.


After filling in your username and
password, you'll see this interface:

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
PC connection 20
7

Open this file and you'll see patient data in the ultrasound unit, as follows:

If you still have questions, please contact us.

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Foot switch 20
8

The foot switch is usually shown in the following image:

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Content 20
9

1 System Overview

2 Software

3 Hardware

4 Troubleshooting

5 Peripherals Connection

6 Preventive Maintenance

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
21
0

 Maintenance of the appearance of the system


 Trackball Cleaning
Preventive Maintenance
 Maintenance of dust filter
 Maintenance of the probe
 Precautions for disinfection of the probe

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Preventive maintenance
- maintenance of the appearance of the system 21
1

• It is recommended to clean the system once a month: After cutting off the power of the
machine, use a soft, lint-free towel (do not use a very damp cloth to prevent liquid from
dripping inside the color Doppler ultrasound) to clean the system cover, keyboard, display,
and touch screen.

• The parts that are more difficult to clean can be cleaned locally with medical alcohol. Do
not wipe the screen and touch screen with chemical liquids (including alcohol) to avoid
corrosion of the coating. The probe cup can be removed for cleaning, wiped dry and put
back

• In daily use, avoid the couplant dripping on the keyboard panel to prevent it from
penetrating into the internal circuit board and causing short circuits or insensitive key
functions. When accidentally dripping, wipe it off as soon as possible

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Preventive maintenance - Trackball Cleaning 21
2

It is recommended to clean the trackball once a month

The cleaning steps of the trackball are as follows:

• Turn off the device and press your finger on the convex mark of the
trackball fixing ring
• Rotate the retaining ring counterclockwise until it can be released from
the keyboard panel
• Take out the fixed ring and trackball ball
• Use a soft cloth dipped in medical alcohol to clean the stains on the
trackball ball and the fixing ring
• Use a cotton swab dipped in medical alcohol to clean the notch of the
trackball
• After cleaning, put the ball and fixing ring back into the notch of the
trackball, and rotate clockwise to fix

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Preventive Maintenance - Maintenance of dust filter 21
3
Make sure cleaning is done every three months
The cleaning steps of the air filter are as follows:

• Turn off color Doppler ultrasound


• Take out the air filter as shown below
• Use a small brush and running clean water to remove the dust on the air filter, and wash the
sponge with clean water
• After cleaning, make sure to air dry and reinstall the color Doppler ultrasound

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Preventive Maintenance - Maintenance of the probe 21
4
• Check the surface lens, cables and etc. of the probe before use. If you find any damage, please
stop using it to avoid injury to the patient or the operator.

• Use couplant that meets national standards and does not contain oily and corrosive ingredients

• Clean the probe surface with a soft tissue after each patient examination

• Remember to Freeze the image during operation gaps and when not in use for a long time

The most economical and effective


way to extend the lifespan of the probe

Replace the probe when the power is disconnected or under Freeze condition

The probe is a precision instrument, so we should avoid dropping or hitting it; be careful not to scratch
the surface of the probe when using it, and try to avoid pulling or bending the probe cable tightly.
以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Preventive maintenance –
precautions for disinfection of the probe 21
5

Disinfection category Definition Applicable probe type

The device enters the human body's sterile


tissues, organs, vasculature, or has sterile body Probes used for puncture
Sterilize
fluids flowing through it, or touches broken or contact with blood
skin or broken mucous membranes.

The device touches the intact mucosa, but


does not enter the human body's sterile tissues,
High-level disinfection Intracavity probe
organs and bloodstream, nor does it touch
damaged skin or damaged mucosa.

Low or medium level The device touches intact skin and does not
Body surface probe
disinfection touch mucous membranes.

For reference only, please refer to the instructions

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
Preventive maintenance –
precautions for disinfection of the probe 21
6

• The probe disinfection method is mainly glutaraldehyde and hydrogen peroxide soaking, and the biopsy guide
disinfection method is mainly high temperature, high pressure and glutaraldehyde disinfection and sterilization.
• Transesophageal TEE: high-level disinfection, 50 minutes for glutaraldehyde / 5 minutes for phthalaldehyde
• Intraoperative probe (uterine cavity probe/puncture probe/laparoscopic probe): sterilization, 10 hours for
glutaraldehyde / 40 minutes for plasma (please refer to the instructions for specific requirements)
• The cleaning, disinfection and maintenance of the probe must strictly follow the recommended methods in the
manual.
• Liquid is strictly forbidden to enter the matching box of the probe, and the immersion disinfectant liquid must
not exceed the junction line of the probe head

For reference only, please refer to the instructions


以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
21
7

Version information

Version Creator Confirm Date Comment

V1.0 李昊旸 张腾 20210615 初版发布

V1.0.1 Gordon.Gao 20210914 English Version

V1.0.2 李昊旸 20211013 English Version

以上所有信息均为深圳开立生物医疗科技有限公司所有,未经授权不得外传。
深圳开立生物医疗科技股份有限公司 All Rights reserved,NO Spreading abroad without Permission of SonoScape.
21
8

Thanks!