Val 40 75
Val 40 75
1 2 3 4 5 6 7 8
40.0°C 40.0 40.0 40.1 39.9 39.8 40.1 39.8 39.9
75.0%RH 74.4 74.6 74.7 74.6 74.8 74.8 74.3 75.0
Set value Chamber readings ( Average Readings)
1 2 3 4 5 6 7 8
40.0°C 39.9 40.0 39.9 40.2 40.0 39.9 39.9 40.1
75.0%RH 74.6 74.7 74.8 74.4 74.4 74.6 74.2 74.8
VALIDATION CERTIFICATE
SCOPE
a) Company Name : Graviti Pharmaceuticals Private Limited
b) Equipment : Walk in Humidity Chamber
c) Model No. : NLWH1071SU9
d) Serial No. : NLWH-0219-0137
VALIDATION DETAILS
a) Validation Certificate No : NLEPL/22-23/HYD/VC/658
b) Validation Certificate Date : 03-February-2022
c) Next Validation Due On : 02-February-2023
d) Validated At : Graviti Pharmaceuticals Private Limited
SOURCE OF VALIDATION
Data logger
a) Equipment : Temperature / Humidity Sensor
b) Make and Model No, S.No : Newtronic , NW 70, NW 16051718
c) Calibration
e) Tag No Report No : : NLEPL/CAL/0421/16051718
QCE099
d) Accuracy : Temperature: ± 0.2 °C Humidity: ± 1.5%RH
e) Due Date of Calibration : 13-April-2022
RESULTS :Validation
ACCURACY:
Temperature : Should not vary more than ± 1.0 ° C with Respect to Set value
Humidity : Should not vary more than ± 3.0% RH with Respect to Set value.
CONCLUSION / REMARK
Temperature and Humidity measurements are within the specified acceptance criteria.
APPROVED BY : _____________________________________
Page 1 of 1