Professional Documents
Culture Documents
Instruction Manual 2
Instruction Manual 2
PIEZOLITH
3000
from SN 0300
GERMANY USA UK
RICHARD WOLF GmbH RICHARD WOLF RICHARD WOLF UK Ltd.
D--75438 Knittlingen Medical Instruments Corp. Waterside Way
Pforzheimerstr. 32 353 Corporate Woods Parkway Wimbledon
Tel.: (..49)--(0)7043--35--0 Vernon Hills, Illinois 60061 SW 17 0HB
Fax:(..49)--(0)7043--35300 Tel.: 847--913 1113 Tel.: 020--8944 7447
MANUFACTURER Fax: 847--913 14 88 Fax: 020--8944 1311
0 GA--A 175
Contents
1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Device description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.1 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.2 Piezoelectric principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.2.3 Locating systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Indications and usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.5 Warnings and precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5.1 Patient selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5.2 Pretreatment set--up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5.3 Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5.4 Post--treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5.5 Device maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.6 Adverse events . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.6.1 Potential Adverse Events with Extracorporeal Shock Wave Lithotripsy . . . . . . . . . . . . . 7
1.7 Patient selection and treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.8 Safe radiation practices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.9 Positioning examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1.10 Use in gastroenterology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.10.1 Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.10.2 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.10.3 Complications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.10.4 Side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.10.5 Positioning examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1.11 Use in orthopedics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.11.1 Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.11.2 Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.11.3 Complications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.11.4 Side effects . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.11.5 Indication examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
1.12 Combinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
1.12.1 General requirements on products/components of a combination . . . . . . . . . . . . . . . . . 12
1.12.2 Specific requirements on the products/components of a combination . . . . . . . . . . . . . . 13
1.13 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2 Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.1 PIEZOLITH with LITHOARM, X--ray, ultrasound device and treatment table . . . . . . . . 15
2.1.1 Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 PIEZOLITH with stand, X--ray, ultrasound device and treatment table . . . . . . . . . . . . . 16
2.2.1 Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.3 PIEZOLITH with articulated arm and compact ultrasound device . . . . . . . . . . . . . . . . . . 17
2.3.1 Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4 Connections of PIEZOLITH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4.1 Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.5 ”Water treatment” controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.6 Display and control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.6.1 Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.7 Therapy source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
GA--A 175 I
2.7.1 Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1 Parking brakes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.1.1 Parking brakes of Lithotripter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2 Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3 Water treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.1 Removing the coupling membrane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.2 Installing the coupling membrane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.3 Information on water preservative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.3.4 Transducer position for filling/emptying/deaerating the system . . . . . . . . . . . . . . . . . . . . 24
3.3.5 Filling the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.3.6 Emptying the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.3.7 Deaerating the system / evacuating air bubbles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.4 Disassembling and assembling the therapy source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.4.1 Disconnecting and connecting the therapy source connection . . . . . . . . . . . . . . . . . . . . 27
4 Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.1 Daily checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.2 Monthly checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.3 Checks after transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4 Checking the Shock Wave Energy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1 General instructions for safe and effective use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.2 Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2.1 Movable table top . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2.2 Treatment steps with X--ray and ultrasound location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
5.2.3 Treatment steps with ultrasound location and articulated arm . . . . . . . . . . . . . . . . . . . . . 32
5.2.4 Shock wave therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.3 Evaluating the shock wave therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.3.1 Mean intensity value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.3.2 Shock wave statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7 Technical description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
7.1.1 Status/error messages on the ”shock wave counter” display . . . . . . . . . . . . . . . . . . . . . . 37
7.1.2 Device malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7.2 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2.1 Pressure pulse characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.2.2 Locating systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.3 Operating, storage, transport and shipping conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
II GA--A 175
Contents
7.4 Considerations for transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.4.1 General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.4.2 Preparation of the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.4.3 Stowage of the system in a vehicle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.4.4 Set up after Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
7.5 Spare parts and accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
7.6 Replacement of parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.6.1 Replacing device fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7.6.2 Disposal of product, packaging material and accessories . . . . . . . . . . . . . . . . . . . . . . . . 44
9 Literature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
1.1 Symbols
Symbols Meaning
Equipotentiality
”Stop” button
”Function” button
1 GA--A 175
Symbols Meaning
Video In
Video Out
L Signal input lithoarm or laser
Data transfer
Therapy source
Identification in conformity with Medical Devices Directive 93/42/EEC only valid if the product and/or
packaging marked with this symbol. Products of category IIa and above, as well as
sterile products or products with measuring function pertaining to category I, are additionally marked
with the code number of the notified body (0124)
GA--A 175 2
1.2 Device description
3 GA--A 175
1.2.3 Locating systems
The isocentric arrangement of the location systems guarantees that there
is no deviation between the target cross and shock wave focus even
when the direction of the shock wave is changed.
Ultrasound location
The inline ultrasound locating system allows precise localization of the
indications that can be visualized sonographically as well as permanent
sonographic control during therapy. For this purpose an ultrasound probe
is located in the therapy source and is rotatable inline and movable along
its longitudinal axis.
X--ray location
Fixed assignment of the therapy source to the isocenter of the X--ray
C--arm via LITHOARM or the laser--monitored stand allows X--ray loca-
tion by confocal adjustment of the piezo power source with X--ray C--arm
from AP to oblique view +/-- 30_. X--ray can not be used with articulated
arm.
1.4 Contraindications
Do not use the PIEZOLITH 3000 in patients with:
Z Confirmed or suspected pregnancy.
Z Coagulation abnormalities (as indicated by abnormal prothrombin time,
partial thromboplastin time, or bleeding time) or those currently recei-
ving anticoagulants (including aspirin).
Z Arterial calcification or vascular aneurysm in the lithotripter’s shock
wave path.
Z Urinary tract obstruction distal to the stone.
Z Anatomy which precludes focusing the device at the target stone, such
as severe obesity or excessive spinal curvature.
Z Immobile asymptomatic diverticulum stones or complicated staghorn
calculi in the renal calyx.
GA--A 175 4
1.5 Warnings and precautions
5 GA--A 175
1.5.3 Treatment
WARNING
Bilateral stones: Do not perform bilateral treatment of kidney stones in a
single treatment session, because either bilateral renal injury or total uri-
nary tract obstruction by stone fragments may result. Patients with bilat-
eral kidney stones should be treated using a separate treatment session
for each side. In the event of total urinary obstruction, corrective pro-
cedures may be needed to assure drainage of urine from the kidney.
Air--filled interfaces in shock wave path: Do not apply shock waves to
air--filled areas of the body, i.e., intestines or lungs. Shock waves are
rapidly dispersed by passage through an air--filled interface, which can
cause bleeding and other harmful side effects.
Cardiac arrhythmia during treatment: If a patient experiences cardiac
arrhythmia during treatment reduce the shock wave intensity and the rep-
etition rate. If necessery, the shock wave therapy should be terminated.
As a general practice, patients with a history of cardiac arrhythmia should
undergo extracorporeal shock wave lithotripsy using an ECG--monitoring.
CAUTION
Renal injury: To reduce the risk of injury to the kidney and surrounding
tissues, it is recommended that:
1. the number of shock waves administered during each treatment
session be minimized;
2. retreatment to the same kidney/anatomical site occur no sooner than
1 month after the initial treatment;
3. each kidney/anatomical site be limited to a total of three treatment
sessions.
Use of fluoroscopy: While fluoroscopy must be used during the pro-
cedure, caution should be used to minimize the exposure.
Electromagnetic interference: If electromagnetic interference between
the extracorporeal shock wave lithotripter and near by electronic equip-
ment is suspected (as evidenced by erratic behavior with either device), it
is recommended that their distance be increased until proper operation
resumes. If it is necessary to operate an electronic device in close prox-
imity to the lithotripsy system during treatment, the device and the litho-
tripter should be tested for proper simultaneous operation prior to clinical
use.
1.5.4 Post--treatment
CAUTION
Radiographic follow--up: All patients should be followed radiographi-
cally after treatment until stone--free or there are no remaining stone frag-
ments which are likely to cause silent obstruction and loss of renal func-
tion.
GA--A 175 6
1.6 Adverse events
Potential adverse events associated with the use of extracorporeal shock
wave lithotripsy include those listed below, categorized by frequency and
individually described:
7 GA--A 175
Cardiac arrhythmia: Cardiac arrhythmias, most commonly premature
ventricular contractions, are generally reported during extracorporeal
shock wave lithotripsy at fixed shock wave delivery in < 1% of patients.
These cardiac disturbances rarely pose a serious risk to the healthy pa-
tient, and typically resolve spontaneously by reducing shock wave inten-
sity and the repetition rate or terminating treatment.
Hematoma (perirenal/intrarenal): Clinically significant intrarenal or per-
irenal hematomas occur in < 1% of lithotripsy treatments. These patients
typically present with severe, chronic flank pain. Although clinically signifi-
cant hematomas often resolve with conservative management, severe
hemorrhage and death have been reported. Management of severe renal
hemorrhage includes the administration of blood transfusions, percuta-
neous drainage, or surgical intervention.
Renal injury: Extracorporeal shock wave lithotripsy procedures have
been known to cause damage to the treated kidney. The potential for in-
jury, its long--term significance, and its duration are unknown.
GA--A 175 8
1.9 Positioning examples
Z Kidney stones
' Positioning: Dorsal position
' Therapy source: from underneath
Z Deep ureterolith
' Positioning: Dorsal position
' Therapy source: from above
9 GA--A 175
1.10 Use in gastroenterology
1.10.1 Indications
Extracorporeal shock wave lithotripsy:
D In the gallbladder.
D In the bile duct.
D In the pancreatic duct.
. IMPORTANT!
It is often impossible to determine the actual degree of disintegration im-
mediately after treatment. To determine whether there are still major frag-
ments, sonographic (bile duct calculi) or if necessary endoscopic (stones
in the pancreatic duct) control is necessary.
In many cases multiple treatment is recommend to achieve optimum fine
fragmentation, as this will allow spontaneous discharge of the fragments
or facilitate additional endoscopic measures (bile duct calculi and calculi
in the pancreatic duct) or accompanying drug treatment (gallstones).
1.10.2 Contraindications
Coagulation diseases; administration of anticoagulants; pregnancy; unre-
liable stone location.
1.10.3 Complications
If fragments are discharged into the small intestine colic pain and feverish
inflammations can occur in rare cases. Occasionally fragments can be
jammed in the bile duct causing obstructions or germ transmission into
the blood stream.
. NOTE!
Documented explanations for the patient on ESWL--gastroenterology is available from Perimed
Compliance Verlag, D--91058 Erlangen, Weinstr. 70.
GA--A 175 10
1.11 Use in orthopedics
1.11.1 Indications
Extracorporeal shock wave therapy:
D In soft tissue near bones of the postural and locomotor system.
D Enthesiopathies (tennis or golfer’s elbow, painful stiffness of the
shoulder, plantarfasciitis (heel spur), achillodyna).
D Tendopathies with ectosteal calcareaous deposits (tendinosis
calcarea).
1.11.2 Contraindications
Infections; tumors; clotting disorder (prior coagulation analysis neces-
sary); use of hemodiluting medication; pregnancy; lung tissue in the
shock wave path; children under the age of 14.
1.11.3 Complications
Serious complications have not yet been observed.
1.11.4 Side effects
Occurrence of:
-- Local haematoma, petechiae
-- Local irritation of the skin
-- Local edemas
11 GA--A 175
1.12 Combinations
. IMPORTANT
In addition to this instruction manual follow the manuals of the products used in
combination with this product.
h = 2.5 m
Patient environment
MP MP
-- -- --
µ µ
MP NMP
µ µ
MP NMP
a) additional protective earth connection
** -- -- (to be clarified with manufacturer),
µ
or
MP
b) with additional separating transformer **
NMP
MP NMP
-- --
µ µ
additional separating transformer additional separating device Functional µ power supply grid
according to IEC/ EN 60601--1--1 ** according to IEC/ EN 60601--1--1 connection
MP = medical electrical device according to IEC/ EN 60601--1, UL 60601--1, CSA C22.2 No. 601
NMP = non-- medical electrical device in accordance with the relevant product--specific IEC/ EN/ UL/ IEC standards
* If connected via a joint mains/power cord under normal conditions the earth leakage current of the system must not exceed 500 µA
(300 µA for systems in acc.with UL 60601--1).
** e.g. Richard Wolf Video Trolley with ”separating transformer”.
GA--A 175 12
1.12.2 Specific requirements on the products/components of a combination
. IMPORTANT!
Persons combining products to form a system are responsible for not impairing the system’s compliance with
the performance and safety requirements, and that the technical data and the intended use are adequately
fulfilled.
Electromagnetic interference or other types of interference occurring between this product and other products
can cause failures or malfunctions.
When selecting the system components ensure that they meet the requirements for the medical environment
they are used in, in particular IEC/ EN 60601--1--1. In case of doubt contact the manufacturer(s) of the system
components.
Do not touch connecting devices for electrical connections between the different components (such as signal
input and output connections for video signals, data exchange, control circuits, etc.) and the patient at the
same time.
CAUTION!
Possible parting of protective earth cable in mobile multi- socket outlet or extension.
Danger for patient and user due to inadmissible increase in leakage current.
Connect each system component individually to a wall socket.
Do not connect the components to a mobile multi- socket outlet or extension.
13 GA--A 175
Guidance and manufacturer’s declaration -- electromagnetic immunity for products that are not life--supporting
The product is intended for use in the environment specified below. The user should assure that the product is used in such an environment.
Compliance
Immunity test IEC 60601 test levels Electromagnetic environment -- guidance
level
REMARKS: At 80 MHz and 800 MHz the higher frequency range applies.
These guidelines may no apply in all situations, as the propagation of electromagnetic waves is affected by absorption and
reflexion from buildings, objects and people.
1 = The field strength of fixed transmitters (e.g. base stations for radio telephones, land mobile radios, amateur radio, radio broadcast and
TV broadcast, ...), cannot be predicted theoretically with accuracy. To assess the EMC environment due to fixed transmitters an
electromagnetic site survey should be conducted. If the measured field strength in the location in which the product is used exceeds
the applicable compliance level above, the product should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be required, such as reorienting or relocating the product.
2 = Over the frequency range between 150 kHz and 80 MHz the field strength should be below 3 V/m.
The recommended separation distances between portable and mobile HF telecommunication devices and
devices which are not life--supporting
The product is intended for use in an electromagnetic environment with HF disturbances are controlled.
The user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
HF telecommunications equipment and the product.
Separation distance as a function of transmitter frequency (m)
Rated nominal output power of the
transmitter (Watts) 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
d = 1.2 p P d = 1.2 p P d = 2.3 p P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a nominal output power not listed in the table above, the recommended separation distance (d) in meters (m) can
be determined using the applicable equation (observe frequency). P = nominal power of the transmitter in Watts (W).
REMARKS: At 80 MHz and 800 MHz the higher frequency range applies.
These guidelines may no apply in all situations. Electromagnetic propagation is affected by absorption and reflexion
from buildings, objects and people.
GA--A 175 14
2 Illustration
2.1 PIEZOLITH with LITHOARM, X--ray, ultrasound device and treatment table
5
9 4 6
10.10 7
10
2.1.1 Legend
15 GA--A 175
2.2 PIEZOLITH with stand, X-- ray, ultrasound device and treatment table
9 5
4 6
10.10 7
10
2.2.1 Legend
1 Therapy source 6 Remote control for shock wave triggering
2 Stand for therapy source 7 Ultrasound device
3 Display and control panel 9 X--ray device
4 Power on/off switch of lithotripter 10 Treatment table
5 Lithotripter 10.10 Extension for treatment table
GA--A 175 16
2.3 PIEZOLITH with articulated arm and compact ultrasound device
8 3
1
6
2.3.1 Legend
1 Therapy source 6 Remote control for shock wave triggering
3 Display and control panel 7 Ultrasound device
4 Power on/off switch of lithotripter 8 Articulated arm
5 Lithotripter
17 GA--A 175
2.4 Connections of PIEZOLITH
19
Service
14
15
MODEM
L
13
18
17 20 16
2.4.1 Legend
13 Equipotential connector 17 Power input socket with fuse holder
14 BNC “Video Out“ socket (option) 18 Identification plate
15 Connection for LITHOARM or Laser 19 Service connector
16 Interface for ultrasound unit 20 Modem connector (option)
12.2
GA--A 175 18
2.6 Display and control panel
3.2
3.3
3.1
3.5
3.4 3.6
3.8
3.7 3.9
3.11
3.12
3.10
3.13
3.14
2.6.1 Legend
3.1 ”Shock wave counter” display 3.8 ”Reduce pulse rate” button
3.2 ”Reset shock wave counter” button 3.9 ”Increase pulse rate” button
3.3 ”Mean intensity” button 3.10 ”Membrane pressure” display
3.4 ”Shock wave intensity” display 3.11 ”Reduce membrane pressure” button
3.5 ”Reduce shock wave intensity” button 3.12 ”Increase membrane pressure” button
3.6 ”Increase shock wave intensity” button 3.13 ”Deaerate system” button
3.7 ”Pulse rate” display 3.14 ”Function” button
19 GA--A 175
2.7 Therapy source
1.6
1.7
1.5
2.7.1 Legend
”Therapy source” control field ”Articulated arm” control field
1.1 ”Retract ultrasound probe” button 8.1 ”Enable articulated arm” button
1.2 ”Extend ultrasound probe” button 8.2 ”Enable joint 1” button
1.3 ”Reduce membrane pressure” button 8.3 ”Enable joint 2” button
1.4 ”Increase membrane pressure” button 8.4 ”Enable joint 3” button
GA--A 175 20
3 Setup
. NOTE!
First instructions are given to the person responsible for the device either
by the manufacturer or a person authorized by him.
WARNING!
The device is not protected against explosions.
Explosion hazard!
Do not operate this device in areas where there is a danger of
explosion.
. NOTE!
Check that the line voltage is the same as the voltage specified on the
identification plate. Connect the device only with the supplied power
cable or a power cable meeting the same specifications.
CAUTION!
Danger of faults and malfunctions.
To guarantee the safety of the user, the patient and others use only
accessories and spare parts specified by the manufacturer of this
product.
Other accessories or spare parts can cause the emission of
increased electromagnetic radiation or reduced immunity against
interference.
. IMPORTANT!
Medical devices are subject to special precautions with regard to electro-
magnetic compatibility (EMC).
Make sure you observe the notes on EMC for installation and operation.
Medical electrical devices can be influenced by mobile HF communica-
tion devices.
If it is necessary to stack the devices or place them next to each other
and HF interference is observed, make sure you observe the intended
use of the devices.
. IMPORTANT!
Never pull, pinch, squeeze or overbend the cables and flexible tubes.
This may lead to failure of the device.
Failure to follow this requirement voids the warranty for the product.
21 GA--A 175
3.1 Parking brakes
The two parking brakes of the front wheels serve to lock the device in
place.
. NOTE!
During treatment both brakes must be locked (engaged).
3.2 Preparation
Z Insert plug of cable--linked remote control button for shock wave
triggering in socket and tighten collar.
Z Connect the locating system(s) to the lithotripter as shown in the
connection diagrams, chapter 8.
Z Connect the power supply/mains plug of the lithotripter.
Z Switch on the power switch.
' The lamp in the power switch is lit.
Z Equipotentiality of the device and the equipotential bonding point of the
room can be achieved via a potential equalization cable.
GA--A 175 22
3.3 Water treatment
1.5 1.5
1.6
1.6
M
1.6 B
1.7
23 GA--A 175
3.3.4 Transducer position for filling/emptying/deaerating the system
GA--A 175 24
3.3.6 Emptying the system
Z Position the therapy source in accordance with section 3.3.4.
Z Place the drain hose (D) without suction filter in a empty container of at
least 10 liters capacity.
D Z Actuate the ”empty system” button.
' During the emptying procedure the ”shock wave counter” display will
read “EEEE”.
' After approx. 20 minutes the system is empty.
' The ”empty system” procedure is completed when the ”shock wave
counter” display switches from letters to numbers only.
' If you want to stop the ”empty system” routine prematurely, press the
”stop” button.
. IMPORTANT!
Empty the water system every 4 weeks at the latest, refill with fresh
water and add the required water preservative.
. NOTE!
There must be no visible air bubbles in the therapy system.
25 GA--A 175
3.4 Disassembling and assembling the therapy source
. NOTE!
Make sure that you do not damage the cables and plugs when disassem-
bling and assembling the therapy source.
1.7 1.6
1.6 1.7
GA--A 175 26
Connecting the therapy source to the LITHOARM or stand:
Z The holding device should be in the lowest therapy position.
' When placing the therapy source onto the holding device, mind the
position of the guide pins (1.8) and actuate both locking buttons
(1.6).
Z After assembly check the therapy source for secure connection and
fully lock the safety lever (1.7) (it must engage in the locked position).
Z Connect the plug for the therapy source to the PIEZOLITH (see chap-
ter 3.4.1) and the scanner plug to the ultrasound device.
Z Check X--ray target cross (see chapter 8).
WARNING!
Residual water can get onto the contacts.
Danger of electrical short- circuit.
Always ensure that the electrical contacts are dry.
. IMPORTANT!
Always secure the plug with a steel cable loop when disconnected.
Never let it trail or fall down.
5.1
Disconnecting the connection:
Z Before you disconnect the connection make sure that the PIEZOLITH
is switched off and disconnected from the power supply/mains.
Z Disconnect the connection by pulling the lever (5.1) upward, then fully
5.1 disconnect the plug.
' After disconnecting hold the plug downward to allow any residual
water to be drained, then dry again using a cloth.
IMPORTANT!
If the PIEZOLITH is ON and the plug is not connected, status message “A500” is
displayed (see 7.1.1). If you disconnect the connection during the filling or empty-
ing process it is necessary to empty the system and fill it once more.
27 GA--A 175
4 Checks
. NOTE!
For technical safety checks see chapter 6.
GA--A 175 28
4.4 Checking the Shock Wave Energy
. IMPORTANT!
The shock wave energy can only be determined correctly if the system water is
free of air bubbles and clean. Air bubbles and microparticles can cause a reduc-
tion in performance. If necessary carry out a new water treatment cycle or deaer-
ate the system.
partment.
Z Upon completion of the test, remove the water from the test cylinder.
' For this purpose hold a container under the test cylinder and care-
fully return the therapy source to horizontal position.
Z Fully retract the membrane bulge using the ”reduce membrane pres-
sure” button, then remove the test cylinder.
29 GA--A 175
5 Application
CAUTION!
Use of the PIEZOLITH 3000 for the treatment of urinary stones in the middle
and lower ureter should be performed using x- ray localization.
. IMPORTANT!
Should the coupling membrane be damaged and system water leak from the ther-
apy source, stop the treatment immediately, switch off the device and install a new
coupling membrane. Then refill the system.
. IMPORTANT!
Before any treatment disinfect the coupling membrane with surface disinfectant
which has been approved for patients.
Use only ultrasound gels certified as suitable for the use on patients.
Follow the gel manufacturer’s instructions.
. IMPORTANT!
During lithotripsy fine cavitation movements in the target area are visible during
ultrasound location indicating that the therapy focus is the same as the ultrasound
target cross.
. IMPORTANT!
If shock wave triggering is inhibited and the display reads “A400” immediately
check the position of the X--ray target cross.
. IMPORTANT!
Check stone position during treatment as often as necessary to ensure correct
treatment.
. NOTE!
Cushioning of the pressure pulse when it passes tissue and additional energy
absorption in bones.
. NOTE!
Avoid trapped air between the coupling membrane and the patient’s body.
Make sure that you choose an anatomically suitable entry window for the shock
waves. Move the therapy source to the required therapy position. If this is not ob-
served energy losses may occur.
During treatment the parking brakes of all system components used must be
locked.
GA--A 175 30
5.2 Treatment
31 GA--A 175
Z If the ultrasound image is poor apply further ultrasound gel to the cou-
pling membrane or the patient. Increase the membrane coupling pres-
sure until good contact is established between the membrane and the
patient. Do not move the patient by increasing the membrane pressure
excessively.
' If the patient has moved, stop the shock wave therapy and reposition
under ultrasound and X--ray control until the therapy field again coin-
cides with the target cross.
Z After 1000 shock waves at the latest perform fluoroscopy and reposi-
tion the treatment table if necessary.
Good ultrasound image = good transmission of the shock wave energy =
highly successful treatment!
Therapy zone visible in the ultrasound targeting cross = high accuracy =
quick success of treatment
Z In the case of deep therapy zones set the penetration depth of the
therapy focus into the body to maximum focal depth by pressing the
”reduce membrane pressure” button (fully retract the membrane).
' The bulging of the coupling membrane determines the depth of pen-
etration of the therapy focus.
GA--A 175 32
. NOTE!
Rotate the ultrasound probe to adjust the optimum image angle.
For location, the ultrasound probe can be moved axially in direction of the therapy
focus to achieve direct coupling of the ultrasound probe via the coupling mem-
brane with the patient’s skin. This prevents multiple reflections and an optimum
ultrasound image.
To avoid energy losses during shock wave therapy the ultrasound probe should
be retracted as far as possible.
33 GA--A 175
5.3 Evaluating the shock wave therapy
5.3.1 Mean intensity value
Z The mean (average) shock wave intensity value per treatment is dis-
played when you press the ”mean intensity” button on the shock wave
counter display field (e.g. ”C 10.5”).
' Pressing the ”reset shock wave counter” button deletes the ”mean
intensity value”
GA--A 175 34
6 Reprocessing and maintenance
Clean the device with a soft cloth moistened with surface disinfectant,
alcohol or spirit.
Follow the disinfectant manufacturer’s instructions.
. IMPORTANT!
Make sure that no humidity enters the device. Do not use any cleaning
agents, scouring agents or solvents on this device.
35 GA--A 175
6.4 Maintenance
. IMPORTANT!
In your inquiries or correspondence please always indicate the model and
serial number on the identification plate. Further documentation is avail-
able from the manufacturer on request.
. IMPORTANT!
Safety checks and tests may only be carried out by the manufacturer or
persons with specialized technical knowledge.
Document the test results and include them in the device record.
The device must not be used if the specified values are not displayed or
functions are not fulfilled.
Test intervals:
Every 12 months the safety check must be carried out as follows
Z Visual check for completeness, soiling and aging.
' Lettering and labels.
' All electrical lines and connections for damage and proper position/
firm connection.
' Leakage test of the water system.
Z Check of electrical safety to EN / IEC 60601--1
Z Function check as described in the instruction manual.
Z Check of the attached locating device(s) in accordance with chapter 8.
GA--A 175 36
7 Technical description
7.1 Troubleshooting
. IMPORTANT!
If the fault or errors cannot be eliminated with the help of this table,
please contact our service department or send in the device for repair.
' Do not attempt to do any repairs yourself.
A020 Voltage of high voltage generator too high Contact service department
A030 Voltage of high voltage generator too low Contact service department
A040 Error in high voltage generator Contact service department
A080 Parting of pressure sensor cable in therapy source Contact service department
A100 Weight monitoring Contact service department
Perform locating check and acknowl-
A200 Shock wave stop after 1000 shock waves
edge with ”F” key
No shock wave transmission possible,
A300 Empty / fill device
system not ready for use
Check position monitoring of X--ray
A400 Shock wave triggering inhibited arc, if adjustment not possible
contact service department
A500 Plug for therapy source not connected Connect plug to socket
A700 Water change deadline exceeded by more than 2 weeks Carry out a water change
Check suction filter, carry out new
A800 Suction filter clogged or insufficient water supply
water treatment cycle
A900 Error in high voltage control to therapy source Contact service department
37 GA--A 175
7.1.2 Device malfunctions
GA--A 175 38
7.2 Technical data
39 GA--A 175
7.2.1 Pressure pulse characteristics
fx = fy
proximal zg distal
x g , yg
z axis
toward shock za target location away from
wave source shock wave source
acoustic focus
geometric (max. p+)
focus
--6 dB tolerance between geometric
contour x/y axis focus and target location
Definitions:
Target Location is the location in space where the manufacturer intends the operator to locate the calculi.
Acoustic focus is the location in the pressure pulse field of the maximum peak--positive acoustic pressure.
Focal volume is the volume in space contained within the surface defined by the --6 dB (relative to the value at the
focus) peak--positive acoustic pressure contour measured around the focus.
fz is the focal extent, defined as the shortest distance along the z axis that connects points on the --6 dB contour of
peak--positive acoustic pressure in the x--z plane on either side of the focus.
fx is the maximum focal width, defined as the maximum width of the --6 dB contour of peak--positive acoustic pressure
around the focus in the x--y plane that contains the focus.
fy is the orthogonal focal width, defined as the width of the --6 dB contour of peak--positive acoustic pressure around the
focus, in the x--y plane that contains the focus, in the direction perpendicular to fx
GA--A 175 40
7.2.2 Locating systems
. NOTE!
To prevent damage during transport or shipment of the products we recommend using the original packaging
material.
41 GA--A 175
7.4 Considerations for transport
GA--A 175 42
7.5 Spare parts and accessories
43 GA--A 175
7.6 Replacement of parts
4 4
5
3 2
Z Switch off the device and disconnect the power cable from the wall
socket and from the power input connector of the device.
Z Use a screwdriver [6] to take out the fuse housing [1].
Z Push the locking ledge [2] upward until the fuse holder is free and pull
out the fuse holder [3].
Z Remove the fuses [4] from their seats [5] and replace.
Z Reinsert the fuse holder into the fuse housing until it clicks into place.
Z Reinsert the fuse housing into the device and push until it clicks into
place.
GA--A 175 44
8 Locating systems and attachments
8.3
8
8.4
8.5 GN
RD
8.7
BU 8.6
Legend
1 Therapy source 8.4 Ultrasound probe
8 Ultrasound device 8.5 Connection cable between device trolley and ultrasound device
8.3 Probe connector 8.6 Jumper cable (ultrasound operation without PIEZOLITH)
. NOTE!
For additional instructions for use please refer to the instruction manual of
the ultrasound device used.
45 GA--A 175
8.1.2 Connection of compact ultrasound unit (Aloka SSD--900)
8.3
8.2
8.1
1
8.4
Legend
1 Therapy source 8.2 Probe plug
8 Ultrasound device 8.3 Power supply for ultrasound device
8.1 Connecting cable between Ultrasound device -- PIEZOLITH 8.4 Ultrasound probe
. NOTE!
For additional instructions for use please refer to the instruction manual of
the ultrasound device used.
GA--A 175 46
8.1.3 Connection ultrasound device (Aloka SSD--1000 / SSD--3500)
8.3
8.4
8.5
Legend
1 Therapy source 8.4 Ultrasound probe
8 Ultrasound device 8.5 Connecting cable between Ultrasound device -- PIEZOLITH
8.3 Probe connector
. NOTE!
For additional instructions for use please refer to the instruction manual of
the ultrasound device used.
47 GA--A 175
8.1.4 Checking the ultrasound targeting cross
. IMPORTANT!
Proper checking of the ultrasound targeting cross is only possible with air--
bubble--free and clean system water. Air bubbles and microparticles can reduce
the power output. Run new water treatment cycle if necessary or deaerate the
system.
Z Move the therapy source to horizontal.
Z Retract coupling membrane (reduce membrane pressure) and ultra-
sound probe.
Z Place test device for targeting cross onto therapy source and fasten
with the 3 clamps (D).
' Close the clamps clearly over dead center to ensure that they are
locked.
Z Apply ultrasound gel to the test device in the area of the targeting
mark.
Z Press the ”increase membrane pressure” button to couple the mem-
brane to the targeting mark.
Z Press the “extend ultrasound probe” button to move the probe forward
as far as it will go.
' The targeting mark must lie within the ultrasound targeting cross.
' Rotate the ultrasound probe for better image adjustment.
. IMPORTANT!
If the targeting mark is not aligned with the targeting cross do not use the
device for therapy. ' Contact the service department.
D D
GA--A 175 48
8.2 Attaching the X--ray C--arc to the stand
. NOTE!
For additional instructions for use please refer to the instruction manual of
the X--ray device used.
8.2.1 Connecting the stand for the therapy source to the X--ray C--arc
Z Move the X--ray C--arc to the intended treatment position and apply the
parking brakes.
' Ensure that the front wheel (9.3) is in transversal position or is facing
towards the rear.
Z Slide stand (2) completely onto the 2 locating cones (2.1).
Z Push clamping mechanism (2.2) upward and tighten hand wheel.
' Ensure that the nose of the clamping mechanism engages behind
the metal foot of the C--arc.
Z Adjust feet (2.3) of stand until they rest firmly against the floor.
Z Connect the laser connection cable (for monitoring the connection be-
tween the therapy source and the X--ray C--arc to the lithotripter).
CAUTION!
Laser radiation.
Do not look into the laser beam.
Class 2 laser.
2.1 2.2
2.3
9.3
49 GA--A 175
8.2.2 Adjusting the X--ray C--arc to the stand
9.1
9.2
Legend
9.1 Laser reflection mark 9.2 Laser beam
CAUTION!
Laser radiation.
Do not look into the laser beam.
Class 2 laser.
. NOTE!
If the laser position control does not match laser beam adjustment may
only be performed by adequately trained technical personnel.
GA--A 175 50
8.3 Using the X--ray C--arc with LITHOARM
. NOTE!
For additional notes and instructions for use please refer to the manual of
the corresponding X--ray device.
To lock / unlock the LITHOARM, pull out / push in the locking button
(11.3) while pushing the LITHOARM briefly towards the C--arc.
If the therapy source is not in therapy position (the LITHOARM must be
locked in the therapy position) the error message „A 400“ will appear and
a brief alarm is sounded. Shock wave triggering is inhibited.
To check the X--arc targeting cross, proceed as described in section 8.4
11
11.3
11.3
11.1 11.2
11.2
B A
Caution:
Retract the therapy source and reduce the
membrane pressure before you swivel the LITHOARM!
Legend
11 X--ray C arc A Therapy position
11.1 LITHOARM B X--ray position
11.2 Unlocking button for sliding unit C Diagnosis position
11.3 Locking mechanism for LITHOARM
51 GA--A 175
8.4 Checking the X--ray targeting cross
. NOTE!
The functions image inversion, image magnification (zoom), and image
rotation must not be selected or activated.
. IMPORTANT!
If the X--ray targeting cross does not match (i.e. if there is a deviation be-
tween the targeting mark and the X--ray targeting cross) do not use the
device for therapy. ' Contact the service department.
D D
GA--A 175 52
8.4.2 Version Stand
. NOTE!
After checking or adjusting the X--ray targeting cross all levers except the
lever for lateral rotation on the C--arc, must be closed/locked.
The functions image inversion, image magnification (zoom), image rota-
tion and height adjustment must not be selected or activated.
. IMPORTANT!
If the X--ray targeting cross cannot be adjusted (i.e. if there is a deviation
between the targeting mark and the X--ray targeting cross) do not use the
device for therapy. ' Contact the service department.
D D
53 GA--A 175
8.5 Using the X--ray U--arc with LITHOARM
. NOTE!
For additional notes and instructions for use please refer to the manual of
the corresponding X--ray device.
When using an X--ray U--arc the therapy unit is located directly on the
X--ray device.
The error message “A 400” is displayed if the therapy source is not in
therapy position (therapy unit must be locked in therapy position) and a
brief alarm is sounded. Shock wave triggering is inhibited.
The position of the X--ray targeting cross is checked as described in
section 8.4
11
B C 11.3
A 11.4
11.2
11.1
Legend
11 X--ray U--arm A Therapy position
11.1 LITHOARM for therapy source B X--ray position
11.2 Release knob for sliding unit C Diagnosis position
11.3 Locking mechanism for diagnosis position
11.4 Locking mechanism for therapy position
GA--A 175 54
9 Literature
. IMPORTANT!
We do not guarantee that this literature index is complete. Users should
therefore always ensure that they keep themselves informed about the
latest findings in this field of use.
55 GA--A 175