Technical Specifications

INPUT CHARACTERISTICS
Input Current: < 100 pA. AC AMPS (True RMS)
DC NMR: > 60dB at 50/60Hz.
1k Unbalanced > 120dB at DC/50Hz/60Hz, DC ranges; Range Accuracy(±) 45 Hz - 1 kHz Voltage Max Input
CMR: > 60dB at DC/50Hz/60Hz, AC ranges. Burden
200µA 0⋅37% + 100d 300mV 1A, 300V
ACCURACY 2mA 0⋅37% + 100d max fuse pro-
Accuracies apply for 1 year 18oC to 26oC. Temperature coefficient outside tected
these limits is <0⋅1 x quoted range accuracy per oC. Warm-up time to rated
accuracy is 1 hour.
20mA
200mA
0⋅37% + 100d
0⋅37% + 100d
THURLBY THANDAR INSTRUMENTS
DC VOLTS
Range Accuracy(±) Res. Input Impedance Max Input
10A <2000mA
>2000mA
0⋅37% + 100d
0⋅4% + 100d
650mV
max
10A, 300V
fuse pro-
tected
1906
200mV 0⋅017% + 3d 1µV 10MΩ or >1GΩ 1kV DC or Additional error at crest factor=3 typically 0⋅7%. Accuracies apply from
2V 0⋅012% + 3d 10µV (Hi Z mode) AC Peak 10% to 100% of full scale.
any range
20V 0⋅019% + 3d 100µV 10MΩ DISPLAY
200V 0⋅019% + 3d 1mV
Display Type: 13mm LED, 8 digit.
1000V 0⋅019% + 3d 10mV Scale Length: Selectable 5½ digit or 4½ digit.
Reading Rate: 3 readings/sec (5½ digit).
RESISTANCE (OHMS) 10 readings/sec (4½ digit).
Range Accuracy(±) Resol- Max. Measur- Max. Input Overrange: Displays OL if input too great for range.
ution ing Current Overflow: Displays OFLO if calculated result overflows display.
200 0⋅025% + 4d 1mΩ 1⋅5mA 300V DC or Annunciators: LEDs for range, function and program modes.
2k 0⋅019% + 3d 10mΩ 1⋅5mA RMS continu-
ous any COMPUTING FUNCTIONS
20k 0⋅019% + 3d 100mΩ 0⋅4mA Ohms input Null: Operates over full range; values can be stored for
200k 0⋅019% + 3d 1Ω 40µA any range every function.
Digital Filter: 10 options, including optimised default values for
2M 0⋅022% + 3d 10Ω 4µA each range and function.
20M 0⋅07% + 3d 100Ω 400nA Touch & Hold: Reading is held when stable.
% DEV: Displays % deviation from entered reference value.
Max open circuit voltage 3⋅5V. Diode Test on 2k range.
Audible continuity check available on all ranges. dB: Displays measurement in dB relative to 1V, 1mA,
1kΩ or user entered value, or in dBm.
AC VOLTS (True RMS) Ax+B: Linear scaling of results, with offset.
Range Accuracy(±) Limits: Reading displayed with H, L, or P (pass) with respect
to user-defined high and low limits.
45Hz - 5kHz 5kHz - 10kHz 10kHz-20kHz 20kHz-50kHz Min/Max: Minimum and maximum reading stored.
200mV 0⋅2% + 150d 0⋅2% + 250d 3% + 500d - Data Logger: Manual or automatic storage of 100 measurements,
2V 0⋅2% + 150d 0⋅2% + 250d 0⋅5% + 300d 3% + 1500d Storage interval 1s to 9999s.
20V 0⋅2% + 150d 0⋅2% + 250d 0⋅5% + 300d 2% + 1000d INTERFACES
200V 0⋅2% + 150d 0⋅2% + 250d 0⋅5% + 500d - RS232: Baud rates 300, 1200 or 9600. Complies fully with
750V - - - the ARC (Addressable RS232 Chain) interface stan-
0⋅2% + 150d dard. Address selectable from the front panel.
Additional error at crest factor=3 typically 0⋅7%. Input impedance 1MΩ / GPIB (Optional): Fully complies with IEEE-488.1 and IEEE-488.2
<100 pF any range. Max input 750V rms, 1kV peak, any range.
Accuracies apply from 10% to 100% of full scale. GENERAL
DC AMPS AC Input: 220 to 240 volts ±10% or 110 to 120 volts ±10%
50/60Hz, Installation Category II.
Range Accuracy(±) Resol Voltage Max. Input Consumption: 25VA max.
ution Burden
Operating Range: +5oC to +40oC, 20% to 80% RH
200µA 0⋅08% + 12d 1nA 300 mV 1A, 300V Storage Range: -40oC to +60oC
max fuse pro-
2mA 0⋅08% + 12d 10nA Environmental: Indoor use at <2000m, Pollution Degree 1.
tected Safety: Complies with EN61010-1.
20mA 0⋅08% + 12d 100nA
EMC: Complies with EN61326.
200mA 0⋅08% + 12d 1µA Size: 260(W) x 88(H) x 235(D)mm, excl. handle and feet
10A <2000mA 0⋅08% + 12d 100µA 650mV 10A, 300V Weight: 2⋅2kg.
max fused
>2000mA 0⋅12% + 12d 100µA
Thurlby Thandar Instruments Ltd. operates a policy of continuous devel-
5½ digit high performance multimeter
opment and reserves the right to alter specifications without prior notice.
Full autoranging, True RMS, RS-232 & GPIB interfaces
Designed and built in Europe by:

Thurlby Thandar Instruments Ltd.


Glebe Road, Huntingdon. Cambs. PE29 7DX United Kingdom (UK)
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
Email: sales@tti-test.com Web: http://www.tti-test.com

82100-0047 Issue 3
5½ digit performance at a 4½ digit price built-in computing & data logging
higher resolution, accuracy & functionality full bus programmability

Greater resolution and Greater sensitivity and Auto or manual ranging Single key convenience Linear scaling with offset Automatic data logging
accuracy higher input impedance The 1906 offers fully automatic ranging functions for ease of use (Ax+b) The 1906 can store up to 100 readings at
on all functions including current. Alter- any required time interval from 1 reading
The 1906 is a true 51/2 digit meter with a With a maximum resolution of 1µV, 1mΩ Press the “null” key and the present dis- This function enables a reading to be
natively any range can be selected and per second up to 1 reading every 3
scale length of ± 210,000 counts. and 1nA, the 1906 is ten times more sensi-
held manually.
played reading is stored and subtracted multiplied by a scale factor (A) and for an
hours.
tive than a 41/2 digit meter. from all future readings. offset (b) to be added or subtracted.
This gives it not just superior resolution A simple recall sequence allows the
This extra sensitivity enables accurate This powerful feature has a host of uses This means, for example, that the electri-
relative to 41/2 digit meters, but much
measurements to be made in areas previ-
4 terminal Ohms for such as removing test lead resistance, cal output of a transducer can be scaled
readings to be scrolled onto the display
higher effective accuracy as well. whenever required. Alternatively results
ously impossible such as thermocouple improved stability observing deviation levels, or measuring to give a direct readout of the physical
junctions, switch contact resistance or ca- relative to alternative voltage points us- can be downloaded using the RS232 or
Resistance measurements can be made parameter e.g. temperature or weight.
pacitor leakages. ing a fixed ground lead connection. GPIB interfaces.
in either 2 terminal or 4 terminal mode. It is particularly useful with 4-20mA cur-
High impedance sensing ensures that As a result, tedious time related meas-
Industry standard 10MΩ input impedance rent loops.
the meter switches automatically be- Touch and Hold urement sequences can be handled
applies to all dc voltage ranges, but the low-
tween modes. automatically without disrupting your
est two can alternatively be selected at The touch and hold function is invaluable Limits comparison work.
greater than 1000MΩ impedance to elimi- For low resistance measurements, 4 ter- when taking measurements in awkward This function enables the reading to be
nate errors when measuring high imped- minal mode offers much higher accuracy situations where it may be difficult to see Manual storage of readings is also avail-
tested against high and low limits set by able eliminating the need for paper and
ance circuitry. and repeatability. the multimeter display. the user. pencil when making a series of measure-
The resistance function also provides se- With touch and hold selected, the display ments.
True RMS ac functions lectable audible continuity as well as di- is automatically frozen every time a sta-
ode test capabilities. ble reading is achieved. A further press Storage can also be triggered from the
The 1906 provides True RMS ac response
The diagram shows how the effective ac- of the T/HOLD key un-freezes the dis- RS-232 or GPIB interfaces.
which gives accurate measurements regard-
curacy varies with reading after taking less of the waveform shape. Closed case calibration play. The display then shows a code of H
into account the fixed errors. Over a (high), L (low) or P (pass) at the end of
range of voltages, the 1906 is around
The wide bandwidth voltage attenuator
provides high accuracy within the audio
The 1906 stores its calibration constants
in permanent memory (EEPROM). Con- Digital filtering for better the display. Full bus control via
four times more accurate than a typical
41/2 digit meter.
band and gives extended response to sequently it can be recalibrated without noise performance Percentage deviation ( %)
RS-232 or GPIB
avoid distortion when measuring switching ever opening the case. The 1906 is designed for complete bus
The 1906 uses non-linear digital filtering
waveforms. This function enables the meter to dis- control. Every function of the instrument
which is highly effective in eliminating
unwanted noise, but responds rapidly to play the percentage amount by which the can be controlled using either the RS-232
a change in the reading. reading differs from a nominal value cho- interface or the GPIB interface (optional).
sen by the user.
When working with very noisy or unsta-
ble measurements, the characteristics of It is invaluable for measuring tolerance
the digital filter can be adjusted to suit the or stability.
application. The RS-232 interface, which is fitted as
Min-Max storage standard, is compatible with the TTi ARC
The Min-Max function stores the highest system.
Full protection for the and lowest values of a set of readings.
The ARC system (Addressable RS232
instrument and the user This facility has many uses such as Chain) enables up to 32 instruments to
checking for power supply glitches, re- be “daisy-chained” together and to be in-
The 1906 incorporates extensive protec- cording peak temperature excursions
tion against damage from accidental dividually addressed and controlled us-
and avoiding “missed” readings when ing a single RS232 port of a PC.
overloads on all ranges including 10 using the Logger function.
Amps.
In addition the 1906 meets the stringent Compatible with IEEE-488.2
Logarithmic The GPIB interface is an option. When
safety requirements of IEC348 and
IEC1010-1 for measurements up to measurements (dB) installed, a rear panel switch allows the
1kV. This function allows measurements to be user to choose between RS232 or GPIB
displayed in the form 20log10X. Built-in control.
zero reference scaling enables voltages The GPIB interface conforms fully with
● 5 ½ digit scale length ( 210,000 counts) Storage of front-panel to be displayed in dBV or dBm relative to both IEEE-488.1 and IEEE-488.2. The .2
set-ups any required impedance. standard contains many enhancements
● 0.012% basic one year accuracy The 1906 can store up to six front panel which are particularly important when
set-ups in non-volatile memory. programming systems incorporating a
● 1uV, 1m and 1nA resolution, 4 terminal Ohms In addition to the range and function number of instruments.
● Full automatic or manual ranging each set-up defines the active comput-
ing or logging functions along with their Using the null key allows gains and
losses to be measured directly.
● True RMS ac functions, wide ac bandwidth parameters.
This facility is particularly valuable when
● Full set of current ranges from 200µA to 10A similar tests need to be repeated at inter-
vals.
● Wide range of computing and data logger functions
● Fully programmable by RS-232 or GPIB interfaces
5½ digit performance at a 4½ digit price built-in computing & data logging
higher resolution, accuracy & functionality full bus programmability

Greater resolution and Greater sensitivity and Auto or manual ranging Single key convenience Linear scaling with offset Automatic data logging
accuracy higher input impedance The 1906 offers fully automatic ranging functions for ease of use (Ax+b) The 1906 can store up to 100 readings at
on all functions including current. Alter- any required time interval from 1 reading
The 1906 is a true 51/2 digit meter with a With a maximum resolution of 1µV, 1mΩ Press the “null” key and the present dis- This function enables a reading to be
natively any range can be selected and per second up to 1 reading every 3
scale length of ± 210,000 counts. and 1nA, the 1906 is ten times more sensi-
held manually.
played reading is stored and subtracted multiplied by a scale factor (A) and for an
hours.
tive than a 41/2 digit meter. from all future readings. offset (b) to be added or subtracted.
This gives it not just superior resolution A simple recall sequence allows the
This extra sensitivity enables accurate This powerful feature has a host of uses This means, for example, that the electri-
relative to 41/2 digit meters, but much
measurements to be made in areas previ-
4 terminal Ohms for such as removing test lead resistance, cal output of a transducer can be scaled
readings to be scrolled onto the display
higher effective accuracy as well. whenever required. Alternatively results
ously impossible such as thermocouple improved stability observing deviation levels, or measuring to give a direct readout of the physical
junctions, switch contact resistance or ca- relative to alternative voltage points us- can be downloaded using the RS232 or
Resistance measurements can be made parameter e.g. temperature or weight.
pacitor leakages. ing a fixed ground lead connection. GPIB interfaces.
in either 2 terminal or 4 terminal mode. It is particularly useful with 4-20mA cur-
High impedance sensing ensures that As a result, tedious time related meas-
Industry standard 10MΩ input impedance rent loops.
the meter switches automatically be- Touch and Hold urement sequences can be handled
applies to all dc voltage ranges, but the low-
tween modes. automatically without disrupting your
est two can alternatively be selected at The touch and hold function is invaluable Limits comparison work.
greater than 1000MΩ impedance to elimi- For low resistance measurements, 4 ter- when taking measurements in awkward This function enables the reading to be
nate errors when measuring high imped- minal mode offers much higher accuracy situations where it may be difficult to see Manual storage of readings is also avail-
tested against high and low limits set by able eliminating the need for paper and
ance circuitry. and repeatability. the multimeter display. the user. pencil when making a series of measure-
The resistance function also provides se- With touch and hold selected, the display ments.
True RMS ac functions lectable audible continuity as well as di- is automatically frozen every time a sta-
ode test capabilities. ble reading is achieved. A further press Storage can also be triggered from the
The 1906 provides True RMS ac response
The diagram shows how the effective ac- of the T/HOLD key un-freezes the dis- RS-232 or GPIB interfaces.
which gives accurate measurements regard-
curacy varies with reading after taking less of the waveform shape. Closed case calibration play. The display then shows a code of H
into account the fixed errors. Over a (high), L (low) or P (pass) at the end of
range of voltages, the 1906 is around
The wide bandwidth voltage attenuator
provides high accuracy within the audio
The 1906 stores its calibration constants
in permanent memory (EEPROM). Con- Digital filtering for better the display. Full bus control via
four times more accurate than a typical
41/2 digit meter.
band and gives extended response to sequently it can be recalibrated without noise performance Percentage deviation ( %)
RS-232 or GPIB
avoid distortion when measuring switching ever opening the case. The 1906 is designed for complete bus
The 1906 uses non-linear digital filtering
waveforms. This function enables the meter to dis- control. Every function of the instrument
which is highly effective in eliminating
unwanted noise, but responds rapidly to play the percentage amount by which the can be controlled using either the RS-232
a change in the reading. reading differs from a nominal value cho- interface or the GPIB interface (optional).
sen by the user.
When working with very noisy or unsta-
ble measurements, the characteristics of It is invaluable for measuring tolerance
the digital filter can be adjusted to suit the or stability.
application. The RS-232 interface, which is fitted as
Min-Max storage standard, is compatible with the TTi ARC
The Min-Max function stores the highest system.
Full protection for the and lowest values of a set of readings.
The ARC system (Addressable RS232
instrument and the user This facility has many uses such as Chain) enables up to 32 instruments to
checking for power supply glitches, re- be “daisy-chained” together and to be in-
The 1906 incorporates extensive protec- cording peak temperature excursions
tion against damage from accidental dividually addressed and controlled us-
and avoiding “missed” readings when ing a single RS232 port of a PC.
overloads on all ranges including 10 using the Logger function.
Amps.
In addition the 1906 meets the stringent Compatible with IEEE-488.2
Logarithmic The GPIB interface is an option. When
safety requirements of IEC348 and
IEC1010-1 for measurements up to measurements (dB) installed, a rear panel switch allows the
1kV. This function allows measurements to be user to choose between RS232 or GPIB
displayed in the form 20log10X. Built-in control.
zero reference scaling enables voltages The GPIB interface conforms fully with
● 5 ½ digit scale length ( 210,000 counts) Storage of front-panel to be displayed in dBV or dBm relative to both IEEE-488.1 and IEEE-488.2. The .2
set-ups any required impedance. standard contains many enhancements
● 0.012% basic one year accuracy The 1906 can store up to six front panel which are particularly important when
set-ups in non-volatile memory. programming systems incorporating a
● 1uV, 1m and 1nA resolution, 4 terminal Ohms In addition to the range and function number of instruments.
● Full automatic or manual ranging each set-up defines the active comput-
ing or logging functions along with their Using the null key allows gains and
losses to be measured directly.
● True RMS ac functions, wide ac bandwidth parameters.
This facility is particularly valuable when
● Full set of current ranges from 200µA to 10A similar tests need to be repeated at inter-
vals.
● Wide range of computing and data logger functions
● Fully programmable by RS-232 or GPIB interfaces
Technical Specifications

INPUT CHARACTERISTICS
Input Current: < 100 pA. AC AMPS (True RMS)
DC NMR: > 60dB at 50/60Hz.
1k Unbalanced > 120dB at DC/50Hz/60Hz, DC ranges; Range Accuracy(±) 45 Hz - 1 kHz Voltage Max Input
CMR: > 60dB at DC/50Hz/60Hz, AC ranges. Burden
200µA 0⋅37% + 100d 300mV 1A, 300V
ACCURACY 2mA 0⋅37% + 100d max fuse pro-
Accuracies apply for 1 year 18oC to 26oC. Temperature coefficient outside tected
these limits is <0⋅1 x quoted range accuracy per oC. Warm-up time to rated
accuracy is 1 hour.
20mA
200mA
0⋅37% + 100d
0⋅37% + 100d
THURLBY THANDAR INSTRUMENTS
DC VOLTS
Range Accuracy(±) Res. Input Impedance Max Input
10A <2000mA
>2000mA
0⋅37% + 100d
0⋅4% + 100d
650mV
max
10A, 300V
fuse pro-
tected
1906
200mV 0⋅017% + 3d 1µV 10MΩ or >1GΩ 1kV DC or Additional error at crest factor=3 typically 0⋅7%. Accuracies apply from
2V 0⋅012% + 3d 10µV (Hi Z mode) AC Peak 10% to 100% of full scale.
any range
20V 0⋅019% + 3d 100µV 10MΩ DISPLAY
200V 0⋅019% + 3d 1mV
Display Type: 13mm LED, 8 digit.
1000V 0⋅019% + 3d 10mV Scale Length: Selectable 5½ digit or 4½ digit.
Reading Rate: 3 readings/sec (5½ digit).
RESISTANCE (OHMS) 10 readings/sec (4½ digit).
Range Accuracy(±) Resol- Max. Measur- Max. Input Overrange: Displays OL if input too great for range.
ution ing Current Overflow: Displays OFLO if calculated result overflows display.
200 0⋅025% + 4d 1mΩ 1⋅5mA 300V DC or Annunciators: LEDs for range, function and program modes.
2k 0⋅019% + 3d 10mΩ 1⋅5mA RMS continu-
ous any COMPUTING FUNCTIONS
20k 0⋅019% + 3d 100mΩ 0⋅4mA Ohms input Null: Operates over full range; values can be stored for
200k 0⋅019% + 3d 1Ω 40µA any range every function.
Digital Filter: 10 options, including optimised default values for
2M 0⋅022% + 3d 10Ω 4µA each range and function.
20M 0⋅07% + 3d 100Ω 400nA Touch & Hold: Reading is held when stable.
% DEV: Displays % deviation from entered reference value.
Max open circuit voltage 3⋅5V. Diode Test on 2k range.
Audible continuity check available on all ranges. dB: Displays measurement in dB relative to 1V, 1mA,
1kΩ or user entered value, or in dBm.
AC VOLTS (True RMS) Ax+B: Linear scaling of results, with offset.
Range Accuracy(±) Limits: Reading displayed with H, L, or P (pass) with respect
to user-defined high and low limits.
45Hz - 5kHz 5kHz - 10kHz 10kHz-20kHz 20kHz-50kHz Min/Max: Minimum and maximum reading stored.
200mV 0⋅2% + 150d 0⋅2% + 250d 3% + 500d - Data Logger: Manual or automatic storage of 100 measurements,
2V 0⋅2% + 150d 0⋅2% + 250d 0⋅5% + 300d 3% + 1500d Storage interval 1s to 9999s.
20V 0⋅2% + 150d 0⋅2% + 250d 0⋅5% + 300d 2% + 1000d INTERFACES
200V 0⋅2% + 150d 0⋅2% + 250d 0⋅5% + 500d - RS232: Baud rates 300, 1200 or 9600. Complies fully with
750V - - - the ARC (Addressable RS232 Chain) interface stan-
0⋅2% + 150d dard. Address selectable from the front panel.
Additional error at crest factor=3 typically 0⋅7%. Input impedance 1MΩ / GPIB (Optional): Fully complies with IEEE-488.1 and IEEE-488.2
<100 pF any range. Max input 750V rms, 1kV peak, any range.
Accuracies apply from 10% to 100% of full scale. GENERAL
DC AMPS AC Input: 220 to 240 volts ±10% or 110 to 120 volts ±10%
50/60Hz, Installation Category II.
Range Accuracy(±) Resol Voltage Max. Input Consumption: 25VA max.
ution Burden
Operating Range: +5oC to +40oC, 20% to 80% RH
200µA 0⋅08% + 12d 1nA 300 mV 1A, 300V Storage Range: -40oC to +60oC
max fuse pro-
2mA 0⋅08% + 12d 10nA Environmental: Indoor use at <2000m, Pollution Degree 1.
tected Safety: Complies with EN61010-1.
20mA 0⋅08% + 12d 100nA
EMC: Complies with EN61326.
200mA 0⋅08% + 12d 1µA Size: 260(W) x 88(H) x 235(D)mm, excl. handle and feet
10A <2000mA 0⋅08% + 12d 100µA 650mV 10A, 300V Weight: 2⋅2kg.
max fused
>2000mA 0⋅12% + 12d 100µA
Thurlby Thandar Instruments Ltd. operates a policy of continuous devel-
5½ digit high performance multimeter
opment and reserves the right to alter specifications without prior notice.
Full autoranging, True RMS, RS-232 & GPIB interfaces
Designed and built in Europe by:

Thurlby Thandar Instruments Ltd.


Glebe Road, Huntingdon. Cambs. PE29 7DX United Kingdom (UK)
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
Email: sales@tti-test.com Web: http://www.tti-test.com