Caffeine is most commonly used primary standard. Preparation of stock solution In this application it is used to verify: Caffeine standard 1000 ppm (mg/L): 1. Wavelength accuracy in UV region at 275 nm Weigh 100 mg of caffeine standard in 100 mL 2. Micro volume techniques using two different volumetric flask. Dissolve the standard by taking sample volumes (1.0 L, and 3.0 L) sufficient amount of deionized water. Fill up to the accuracy and reproducibility volume of 100 mL with deionized water. 3. Path length accuracy This is achieved by using 3 different path lengths Preparation of standard solution (1 cm, long and short path length) for 3 different Prepare 3 different concentration ranges of standards concentration range (low, mid and high) standard as mentioned in table1, 2 and 3 using the 1000 ppm solution. Here a series of standard concentration standard stock solution. ranging from 2 – 1000 mg/L are prepared. They Standard Volume stock Water to final are intended to be used to create a calibration curve concentration solution (mL) volume (mL) of absorbance versus concentration of standard (mg/L) solution. It is found to be linear and having a 0 0.0 coefficient of determination greater than or equal to 2 0.1 0.999. 4 0.2 50 6 0.3 8 0.4 Sample 10 0.5 Caffeine standard, Sigma Aldrich, No. C 6035 Table1: Low concentration range caffeine standard
Instruments and Accessories Water to final concentration solution (mL) volume (mL) UV5Nano Spectrophotometer (ME-30254729) (mg/L) XP 205 Analytical balance (ME-11106027) 0 0.0 Rainin pipettes 100 1.0 - 1000 L (ME-17011782) 200 2.0 - 10 mL (ME-17011790) 400 4.0 10 12 x 50 mL volumetric flask 600 6.0 6 x 10 mL volumetric flask 800 8.0 2 x 100 mL volumetric flask 1000 10.0 Table3: High concentration range caffeine standard
METTLER TOLEDO Page 1 of 2
M9102 V1.0 Procedure Results Fill a cuvette of 1 cm path length with low UV/VIS Application Note
concentration of caffeine standards. Blank
measurement is performed using deionized water. Sample volume used for mid concentration solution is 3.0 L (long pathlength) and for high concentration 1.0 L sample volume (short pathlength) is used. Blank measurement is performed using deionized water. Note: To prevent sample carryover, clean the micro volume platform after every measurement using a lint free tissue wipe. A final cleaning is recommended Standard solution Coeffient of determination of 2 with distilled water. Do not use organic solvents or of caffeine R abrasive cleaning agents. Low Concentration 0.9997
Measurement Mid Concentration 0.9990
Method parameters High
0.9996 Concentration Methods: Quant Wavelength: 273 nm Conclusion Background correction: 1-point Wavelength accuracy is possible with a Quant Reference wavelength: 400 nm measurement on UV/VIS spectroscopy. Caffeine No of standards: 6 standard solutions were found to be linear in the Unit: mg/L concentration range of 2 - 1000 mg/L. Specified Fit type 1: Linear acceptance criteria are met: the coefficient of Measurement cell and path length (depending on the determination meets the requirements and working solution concentrations) measurements are repeatable. Comparable results were obtained at different sample volumes as well as Low concentration: None different path lengths. Calibration curve shows that it 1.0 cm obeys Lambert-Beer law within the concentration Mid Concentration: Micro volume platform range. Long path length High Concentration: Micro volume platform Reference Short path length S.Sethuraman et al., Analytical Method Development and Validation Of Caffeine In Tablet Dosage Form By Using UV/VIS Spectroscopy. Acceptance Criteria Coefficient of determination of Further information 2 0.999 (R ) www.mt.com/uv-vis