SB14‐0030
Date: 4/06/2014
Supersedes Bulletin N°:
Product: Stratos
Author: c.dougados
Compliance Category Corrective Action: Identifies the need for action to correct a
condition that, if left uncorrected, may result in reduced
equipment reliability, efficiency or prevent of material
damages.
Compliance may be required within a specific operating time.
Mandatory / Urgent
ABSTRACT:
The purpose of this instruction is to replace a detector tube SAI.52.022 (ex10100625) for a Stratos
unit.
APPLICABILITY:
DESCRIPTION:
First of all, you need to remove the arm upper cover and shut down the unit.
Remove the data cable and High voltage cable from the old detector and loose 3 HC screws .
Remove the 12 CHC 3*10 screws Remove the 1 CHC 3*10 screw
around the cover which maintain the electronic
3. Remove the 8 CHC 3*10 screws in order to take off the electronic part and column
The 2 column
must be
unscrews
the 8 screws to
removed
4. Loose big screws CHC 6*10and push out the detector from his sheath.
5. Put the detector with is ground braid inside the metallic tube:
6. Tight CARREFULLY the detector (hand pressure) with the 2 CHC 6*10 screws.