Professional Documents
Culture Documents
Troubshooting Guide For Optical Box Heater.
Troubshooting Guide For Optical Box Heater.
Open the top cover of the instrument, on the controller PCB find connector J26.
Please check the resistance value (Ohm) from (Brown (1) to Red probe) and (orange (3) to Black
probe) of the meter. Record the reading and input into claim report.
(Ans:_____________2.5____________)
Ohm
Cái mới là vô cùng
Next, please check the resistance value (Ohm) from (Brown (1) to Black probe) and (Orange (3) to
Red probe of the multi meter. Record the reading and input into claim report.
(Ans:_______2.6Ohm_________) 540 Ohm 9.12 - 8.75M Ohm
This is to confirm the condition of the heater.
1) Measure the voltage (Orange to black probe) and (Blue to Red probe) (Ans:_____23.9Vdc ______)
2) Measure the voltage (Yellow to black probe) and (Green to red Probe) (Ans:____23.9Vdc_______)
3) Measure the voltage (Green to black Probe) and (Blue to red probe) (Ans:_____0.0Vdc_______)
From the top cover of the instrument, on the controller PCB find connector J27.
Use a multi meter select Ohm mode. Place the probe on the connector of the temperature sensor.
Record the Ohm value. (Ans:____________9.6KOhm________________)
End of test