Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Available online at www.annclinlabsci.

org

372 Annals of Clinical & Laboratory Science, vol. 49, no. 3, 2019
Evaluation of Cobas 8000 Analyzer Series Module e801
Analytical Performance
Sangeun Lim, Kyunghoon Lee, Hee-Yeon Woo, Hyosoon Park, and Min-Jung Kwon

Department of Laboratory Medicine, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine,
Seoul, Korea

Abstract. Background. Immunoassays are important tests that provide essential information for patient
care. The e801, a module of the recently released Cobas 8000 series (Roche Diagnostics, Mannheim,
Germany), is an automated immunoassay system based on streptavidin-biotin interactions. In this study,
we evaluated the analytical performance of the e801. Methods. We evaluated the precision, linearity, assay
comparison, and reference range validation of 16 analytes (AFP, CA19-9, CA125, CEA, CYFRA, ferritin,
NSE, PSA, Vitamin D, E2, fT4, TSH, FSH, insulin, NT-proBNP, and T) according to the guidelines of
the Clinical Laboratory Standards Institute. Results. In precision evaluations, the coefficients of variation
were less than each allowable total error for all analytes. Linearity was observed for all analytes over the
entire tested analytical range (r2≥0.99). Performance comparisons revealed that the two systems are compa-
rable, with correlation coefficients (r)>0.975 for all analytes. The reference range validation was also within
the allowable criteria. Conclusions. In this study, the Cobas 8000 e801 analyzer demonstrated acceptable
performance with respect to precision, linearity, reference range validation, and correlation. Therefore, the
e801 analyzer is expected to be useful for routine immunoassays in clinical laboratories, although educa-
tion and awareness about biotin interference is necessary for successful implementation in clinical practice.

Key words: Immunoassay, Performance, Evaluation.

Introduction The Cobas 8000 e801 analyzer (Roche Diagnostics,


Mannheim, Germany) uses electrochemilumines-
Immunoassays are important tests that provide cence sandwich assays based on biotin and streptavi-
essential information for patient care. Clinical din interactions. The analyte and antibody form a
laboratory automation of immunoassays is inevi- sandwich, and magnetic particles coated with strep-
table in modern medicine. A variety of immuno- tavidin are then added. The free biotin prevents some
assays have been developed since radioimmuno- of the antibody-analyte sandwich from binding to
assay use was first introduced in the 1960s, and the microparticle by binding to streptavidin-coated
antibody sensitivity and marker studies have im- microparticles instead. This method is hampered by
proved over time [1]. Immunoassay analyzers are false decreases in results in sandwich assays and in-
used to test patient samples for a variety of sub- creases the accuracy of the test. It can also be used to
stances, including cardiac markers, tumor mark- examine various analytes, such as tumor markers and
ers, allergy testing, or endocrine hormone testing. hormones. The reagents are ready to use and are
Modern immunoassay systems have greater re- packed in closed cassettes to allow completely auto-
producibility and sensitivity and shorter response mated reagent handling. Up to four e801 modules
times than previous technologies [2]. Despite can be configured in series, yielding throughput of
these improvements, there is a need to analyze up to 1,200 tests per hour, delivering steady, fast re-
larger amounts of data more quickly and cost-ef- sults in a relatively small space and allowing for con-
fectively [3]. In addition, as the results of immu- tinuous analysis loading when reagents are supplied.
noassays are essential for patient care and clinical A recent study evaluated the analytical performance
decisions, accurate test results and quality control of the Cobas 8000 c702 chemistry analyzer and e602
are critical. immunoassay analyzer [4,5]. However, as far as we
know, the e801 immunoassay analyzer has never
Address correspondence to Min-Jung Kwon, MD, PhD, Department
of Laboratory Medicine, Kangbuk Samsung Hospital, Sungkyunkwan been evaluated. Therefore, in this study, we evaluated
University School of Medicine, 29 Saemunan-ro, Jongno-gu, Seoul
03181, Korea; phone: +82 2 2001 5211; fax: +82 2 757 0711; e mail: the overall analytical performance of the Cobas 8000
mjkkmd@gmail.com e801 using 16 immunoassay assays.

0091-7370/19/0300-372. © 2019 by the Association of Clinical Scientists, Inc.


Evaluation of the e801 immunoassay analyzer 373
Table 1. Precision profiles for the 16 analytes.

Test item Level Mean Standard CV (%) Total


concentration Deviation Repeatability Within- allow-
labora able
tory error
precision (%)

AFP (ng/mL) Low 10.67 0.16 1.6 1.5 21.9


High 112.40 1.87 1.5 1.7
CA19-9 (U/mL) Low 21.03 0.42 2.0 2.3 46.03
High 81.22 1.91 2.3 2.4
CA125 (U/mL) Low 29.21 0.38 1.1 1.3 35.4
High 84.13 0.87 1.1 1.0
CEA (ng/mL) Low 5.90 0.09 1.3 1.5 24.7
High 51.33 0.76 1.4 1.5
CYFRA (ng/mL) Low 2.64 0.06 1.5 2.2 27.9
High 25.28 0.51 1.6 2.1
Ferritin (ng/mL) Low 25.68 0.47 1.7 1.9 16.9
High 211.64 4.80 2.4 2.2
NSE (ng/mL) Low 12.29 0.15 1.0 1.2 15.7
High 98.06 1.00 0.8 1.1
PSA (ng/mL) Low 4.44 0.06 1.4 1.4 33.6
High 37.40 0.66 1.9 1.7
Vitamin D(ng/mL) Low 11.20 0.77 6.3 6.9 25
High 25.54 1.06 4.4 4.1
E2 (pg/mL) Low 82.69 1.55 1.7 1.9 26.86
High 457.88 4.50 1.0 1.0
fT4 (ng/dL) Low 1.20 0.03 2.3 2.6 8.0
High 3.28 0.09 2.3 2.8
TSH (μIU/mL) Low 1.41 0.04 2.4 2.7 23.7
High 8.95 0.24 2.8 2.7
FSH (mIU/mL) Low 15.43 0.29 1.1 2.0 21.19
High 46.84 0.80 1.2 1.8
Insulin (μU/mL) Low 26.53 0.83 1.2 3.4 32.9
High 81.01 2.20 1.2 2.9
NT-proBNP (pg/mL) Low 140.36 2.23 1.4 1.6 13.0
High 4547.08 65.33 1.2 1.5
Testosterone (ng/mL) Low 2.51 0.08 1.1 1.5 13.61
High 5.67 0.06 1.4 1.4

Abbreviations: AFP, alpha fetoprotein; CA19-9, cancer antigen 19-9; CA125, cancer antigen 125; CEA, carcinoembry-
onic antigen; CYFRA, cytokeratin 19 fragment; CV, coefficient of variation; E2, estradiol; FSH, follicle stimulating
hormone; fT4, free thyroxine; NSE, neuron specific enolase; NT-proBNP, N-terminal prohormone of brain natriuretic
peptide; TSH, thyroid stimulating hormone.

Materials and Methods neuron-specific enolase (NSE), prostate-specific antigen


(PSA), total 25-hydroxyvitamin D (vitamin D), estradiol
This study was performed between December 2017 and (E2), free thyroxine (fT4), thyroid-stimulating hormone
April 2018. The e801 was evaluated for precision, linear- (TSH), follicle-stimulating hormone (FSH), insulin,
ity, comparison, and reference range validation. We N-terminal prohormone of brain natriuretic peptide (NT-
tested 16 representative immunoassay analytes to evalu- proBNP), and testosterone (T). Reagent preparation and
ate its performance in routine laboratory work; alpha- settings were based on the manufacturer’s recommenda-
fetoprotein (AFP), cancer antigen 19-9 (CA19-9), can- tions. The performance of the e801 was compared to that
cer antigen 125 (CA125), carcinoembryonic antigen of a currently used routine immunoassay analyzer with the
(CEA), cytokeratin fragment (CYFRA), ferritin, e602. Quality control (QC) materials of levels 1 and 2
374 Annals of Clinical & Laboratory Science, vol. 49, no. 3, 2019
Table 2. Linearity.

Test item Range Slope (95% CI) Intercept R2 Recovery (%)

AFP (ng/mL) 0.908-1077 0.990 (0.950-1.029) 6.246 0.999 98.1-104.8


CA19-9 (U/mL) 2.215-986.0 0.997 (0.945-1.050) -9.880 0.999 94.9-100.0
CA125 (U/mL) 0.6-3050.5 0.999 (0.957-1.041) -26.64 0.999 95.5-100.0
CEA (ng/mL) 0.3-910.5 0.993 (0.963-1.024) 3.382 0.999 98.8-100.0
CYFRA (ng/mL) 0.1-49.7 0.980 (0.904-1.056) 0.432 0.998 96.0-108.0
Ferritin (ng/mL) 7.125-2819.5 1.012 (0.953-1.070) -26.032 0.999 94.6-102.0
NSE (ng/mL) 0.382-53.85 0.997 (0.979-1.015) 0.008 0.999 98.9-100.0
PSA (ng/mL) 0.0285-91.5 1.004 (0.987-1.021) -0.032 0.999 98.7-101.0
Vitamin D (ng/mL) 1.605-57.7 1.002 (0.952-1.052) -0.650 0.999 94.9-100.0
E2 (pg/mL) 5.0-2631.0 0.950 (0.884-1.016) 38.051 0.999 94.8-106.9
fT4 (ng/dL) 0.545-3.250 0.935 (0.870-1.001) 0.0331 0.997 103.1-107.3
TSH (μIU/mL) 0.1045-122.0 0.980 (0.887-1.074) -0.450 0.997 93.4-100.0
FSH (mIU/mL) 1.12-192.5 0.999 (0.970-1.028) -0.837 0.999 97.0-100.0
Insulin (μU/mL) 1.37-514.0 1.001 (0.919-1.084) 8.150 0.998 100.0-108.5
NT-proBNP (pg/mL) 32.25-4826.5 0.997 (0.911-1.082) -77.488 0.997 91.5-100.0
Testosterone (ng/mL) 0.1195-12.95 0.994 (0.965-1.023) 0.084 0.999 100.0-105.9

Abbreviations: AFP, alpha fetoprotein; CA19-9, cancer antigen 19-9; CA125, cancer antigen 125; CEA, carcinoem-
bryonic antigen; CYFRA, cytokeratin 19 fragment; CI, Confidence interval; E2, estradiol; FSH, follicle stimulating
hormone; fT4, free thyroxine; NSE, neuron specific enolase; NT-proBNP, N-terminal prohormone of brain natri-
uretic peptide; TSH, thyroid stimulating hormone; R2, coefficient of determination.

Table 3. Comparison results between the cobas 8000 e801 ande602 analyzer.

Test item Number Range Slope (95% CI) Intercept R

AFP (ng/mL) 105 0.974-263.5 0.966 (0.952-0.981) 0.178 1.000


CA19-9 (U/mL) 96 3.42-271.5 0.968 (0.958-0.978) -1.462 0.999
CA125 (U/mL) 105 3.37-465 0.996 (0.978-1.013) 0.250 0.996
CEA (ng/mL) 105 0.392-35.5 1.066 (1.056-1.076) 0.123 0.999
CYFRA (ng/mL) 105 0.413-24.300 0.984 (0.974-0.985) -0.051 0.998
Ferritin (ng/mL) 104 3.71-2000 0.998 (0.980-1.017) 16.320 0.996
NSE (ng/mL) 109 3.805-40.05 1.009 (0.984-1.034) 0.795 0.992
PSA (ng/mL) 107 0.1545-33.4 1.045 (1.035-1.054) -0.034 0.999
Vitamin D (ng/mL) 100 6.54-65.54 0.941(0.907-0.976) -0.293 0.984
E2 (pg/mL) 98 5-824 0.955 (0.943-0.967) -6.516 0.998
fT4 (ng/dL) 103 0.387-7.77 0.888 (0.857-0.920) 0.127 0.984
TSH (μIU/mL) 99 0.0053-17.80 0.996 (0.991-1.001) 0.043 1.000
FSH (mIU/mL) 100 1.07-173.5 1.036 (1.030-1.042) 0.006 1.000
Insulin (μU/mL) 100 0.664-179 1.065 (1.060-1.070) -0.189 1.000
NT-proBNP (pg/mL) 100 6.14-35000 0.995 (0.992-0.998) -37.367 1.000
Testosterone (ng/mL) 100 1.54-41.28 1.012 (0.999-1.025) -0.131 0.998

Abbreviations: AFP, alpha fetoprotein; CA19-9, cancer antigen 19-9; CA125, cancer antigen 125; CEA, carcinoem-
bryonic antigen; CYFRA, cytokeratin 19 fragment; E2, estradiol; FSH, follicle stimulating hormone; fT4, free thy-
roxine; NSE, neuron specific enolase; NT-proBNP, N-terminal prohormone of brain natriuretic peptide; R,
Correlation coefficient; TSH, thyroid stimulating hormone.

(PreciControl, Roche Diagnostics, Indianapolis, IN, methods comparisons using the 16 analytes. Only non-
USA) and pooled sera from subjects who visited our in- hemolytic and non-lipemic sera were used. This study
stitution were used for analyses. The comparative study was approved by the Institutional Review Board of
was performed using sera from at least 100 subjects. Kangbuk Samsung Hospital (2017-11-040) and com-
Samples were non-randomly selected to fulfill criteria plied with the World Medical Association Declaration of
including broad value range and proper distribution for Helsinki guidelines.
Evaluation of the e801 immunoassay analyzer 375
376 Annals of Clinical & Laboratory Science, vol. 49, no. 3, 2019

Figure 1. Comparisons between the cobas e801 and e602. (A) Alpha
fetoprotein. (B) Cancer antigen 19-9. (C) Cancer antigen 125. (D)
Carcinoembryonic antigen. (E) Cytokeratin 19 fragment. (F) Ferritin.
(G) Neuron-specific enolase. (H) Prostate-specific antigen. (I) Vitamin
D. (J) Estradiol. (K) Free thyroxine. (L) Thyroid-stimulating hormone.
(M) Follicle-stimulating horm. (N) Insulin. (O) N-terminal prohor-
mone of brain natriuretic peptide. (P) Testosterone.

Repeatability and within-laboratory precision. For Reference range validation. Selection of reference
evaluations of repeatability and within-laboratory preci- specimens and reference range validation were per-
sion, we tested all analytes using both lower and higher formed according to CLSI guideline C28-A3c [13].
levels of QC materials according to Clinical and The reference ranges were validated using sera from 20
Laboratory Standards Institute (CLSI) guideline EP5-A3 healthy subjects aged 20 to 60 years for the following
[6]. The QC materials at each level were analyzed in five test analytes: AFP, CA19-9, CA125, CEA, CYFRA,
replicates per run on five consecutive days. The coeffi- fT4, TSH, insulin, NSE, NT-proBNP, and vitamin D
cients of variation (CV) per assigned analyte were calcu- for 10 males and 10 females; T and PSA for 20 males;
lated from the 25 results generated for each assay. The E2 and FSH for 20 females; ferritin for 20 males and
CV was deemed acceptable for the analysis if equal to or 20 females. If two or fewer of the 20 tested values for
less than desirable for the allowable total error based on each analyte (or 10% of the tested values) fell outside
the biological variation database specification on the of those reference ranges, the result was considered to
Westgard website [7,8] and other journals [9]. be acceptable for reference range validation. If three or
four test results fell outside of reference ranges, another
Linearity. The linearity of the assays was validated ac-
20 reference specimens similar to the first set were ob-
cording to CLSI guideline EP6-A [10]. For each analyte,
tained and retested.
we generated five equally-spaced concentrations of sam-
ples, and three intermediate concentration pools were
Statistics. For repeatability and within-laboratory pre-
prepared by mixing the high (H) and low (L) level pa-
cision, LaboStats (version 1.5, Laboratory Medicine
tient sera as follows: 0.75L+0.25H, 0.50L+0.50H, and
0.25L+0.75H. To determine the best-fit polynomial, Foundation, Seoul, Korea) was used to calculate the
each sample was measured in duplicate, and the mean mean, standard deviation, and CV for each level of QC
values were determined. In the first polynomial, it was material in the 16 analytes. LaboStats was also used to
determined that linearity was maintained. When the sec- validate the best fit model in the linearity test results.
ond- or third-degree polynomial appeared, it was deter- IBM SPSS, version 24.0 (IBM Co., New York, NY,
mined whether it was a clinically meaningful nonlinear- USA) was used to calculate slope, intercept, 95% CI,
ity based on the document [11]. coefficient of determination (R2) at validation of lin-
earity, and correlation coefficient (r) during compari-
Comparisons. Comparisons were performed following sons. All measured data were entered into relation ta-
CLSI guideline EP9-A3 [12]. Specimens from at least bles using Excel 2010 (Microsoft Corporation,
100 subjects with variable values distributed over the Redmond, WA, USA).
analytical measurement range were collected from left-
over serum samples. All test samples were divided into Results
two aliquots and stored at -70°C. A comparison study
was performed between the e801 and e602 on the same Repeatability and Within-Laboratory
day. The concentration of each analyte was measured in
Precision.The e801 yielded repeatability CVs
duplicate using each of the two instruments. The mean
values were used for regression analyses and for calculat- ranging from 0.8% to 6.3% and within-laborato-
ing the correlation coefficient (r). The results were con- ry precision CVs ranged from 1.0% to 6.9% for
sidered acceptable if the correlation coefficient was all analytes. In our evaluation of precision, all ana-
≥0.975. lytes showed acceptable coefficients of variation
Evaluation of the e801 immunoassay analyzer 377
Table 4. Reference Range Validation.

Test item Proposed reference Total Outside %


interval count count Outside

AFP (ng/mL) 0-7 20 0 0


CA19-9 (U/mL) 0-26.6 20 0 0
CA125 (U/mL) 0-39 20 0 0
CEA (ng/mL) 0-4.7 20 0 0
CYFRA (ng/mL) 0-3.3 20 0 0
Ferritin (male, ng/mL) 59-507 20 0 0
Ferritin (female, ng/mL) 6-174 20 0 0
NSE (ng/mL) 0-16.3 20 0 0
PSA (ng/mL) 0-2.99 20 0 0
Vitamin D (ng/mL) 20-100 20 2 10
E2 (female, pg/mL) 5-498 20 0 0
fT4 (ng/dL) 0.84-1.74 20 0 0
TSH (μIU/mL) 0.27-4.2 20 2 10
FSH (female, mIU/mL) 1.7-21.5 20 0 0
Insulin (μU/mL) 1.5-14.5 20 0 0
NT-proBNP <115 (20 ~ 45 years) 20 0 0
(male, pg/mL) <172 (45 ~ 54 years) 20 0 0
<263 (> 55 years) 20 0 0
Testosterone 2.49-8.36 (20 ~ 50 years) 20 1 5
(male, ng/mL) 1.93-7.40 (> 51 years) 20 0 0

Abbreviations: AFP, alpha fetoprotein; CA19-9, cancer antigen 19-9; CA125, cancer antigen 125; CEA, carcinoem-
bryonic antigen; CYFRA, cytokeratin 19 fragment; E2, estradiol; FSH, follicle stimulating hormone; fT4, free thy-
roxine; NSE, neuron specific enolase; NT-proBNP, N-terminal prohormone of brain natriuretic peptide; TSH, thy-
roid stimulating hormone.

within each allowable total error. A statistical Comparisons. We compared the performances of
summary of precision evaluated using low- and Cobas 8000 e801 and e602. The slopes ranged be-
high-concentration QC materials is shown in tween 0.888 and 1.066, intercepts between -37.37
Table 1. and 16.32, and correlation coefficients (r) between
0.984 and 1.000 (Table 3&Figure 1). The correla-
Linearity. The results of the linearity evaluation of tion coefficients for all 16 analytes were >0.975,
the five concentrations of serum samples prepared suggesting that the two instruments produced com-
by mixing sera at high and low concentrations are parably acceptable results.
shown in Table 2. Coef¬ficients of determination
(R2)≥0.99 were observed for all analytes evaluated. Reference Range Validation. In the validation
The first polynomial was noted for 8 analytes study, we determined that one measured result for
(CA19-9, CEA, ferritin, FSH, NSE, PSA, T, and T and two results for THS and vitamin D were be-
TSH), and it was thus determined that linearity yond the reference ranges. However, as less than
was maintained. For the CA125, insulin and vita- 10% of the values were outside the reference ranges,
min D analytes, a second-degree polynomial was the data were deemed acceptable (Table 4).
observed, while a third-degree polynomial was not-
ed for the remaining analytes (AFP, CYFRA, E2, Discussion
NT-proBNP, and fT4). We next evaluated the data
set to determine whether it showed any clinically Routine immunoassays are routine procedures in
significant differences, and determined that there any clinical laboratory. Therefore, the utility of a
was no clinically meaningful nonlinearity. routine clinical immunoassay analyzer should be
Consequently, the linearity of the data was deemed carefully determined by evaluating its performance.
acceptable. In this study, we used CLSI guidelines to assess the
378 Annals of Clinical & Laboratory Science, vol. 49, no. 3, 2019

performances of the Cobas 8000 analyzer series and that the overall linearity was acceptable for sec-
module e801. We determined it to be a stable ana- ond- or third-degree polynomial analytes.
lytical system for routine laboratory work based on
the results of precision, linearity, comparison, and In this study, we determined that the Cobas 8000
reference range validation. e801 and e602 produced comparable data
(r>0.975) for all tested analytes. The reference
The CV values for repeatability and within-labora- range validation for all analytes was also acceptable,
tory precision were <5%, excluding low-level vita- with equal or less than 10% of the values falling
min D. For vitamin D, the repeatability and with- outside of the proposed reference range. The man-
in-laboratory precision values were 6.3% and 6.9% ufacturer notes that the reference ranges for E2 and
for QC material at lower concentrations of 11.20 FSH vary throughout the menstruation cycle and
ng/mL. According to the Westgard recommenda- with age. However, we were not able to consider
tions [8], the allowable total error for vitamin D the menstruation cycle when performing reference
measurement is traditionally set for analytical per- range validation for E2 and FSH.
formance and can be traditionally calculated using
a total error of 25%. Moreover, in a previous study Many automated immunoassays incorporating bi-
of the repeatability and within-laboratory precision otinylated antibodies and streptavidin-coated mag-
values for vitamin D, the Cobas 8000 e602 assay netic beads are widely used for routine clinical lab-
produced repeatability and precision CVs of 3.9% oratory tests because this approach offers amplified
and 6.0% for low concentrations of 21.1 ng/mL signals and relatively high sensitivity [17]. These
[5], and the DiaSorin LIAISON assay showed a assays, however, may be vulnerable to interference
within-run precision range of 2.9% to 5.5% and a from high biotin concentrations within the sample,
between-run precision range of 6.3% to 12.9% for because excess biotin competes with biotinylated
the vitamin D assay [14]. Both studies evaluated antibody. This competition may yield erroneously
the quality of the data using CVs and concluded low results in sandwich immunoassays and errone-
that the systems provided acceptable precision for ously high results in competitive immunoassays
vitamin D measurement. Likewise, we demonstrat- [18,19]. The biotin tolerance of any given assay is
ed that repeatability and within-laboratory preci- ultimately a function of assay design and the par-
sion for vitamin D measurements were within ac- ticular assay microenvironment. The problem of
ceptable limits. unwanted and unexpected interference has yet to
be overcome, so laboratory and clinical staff remain
For linearity evaluation, we performed data pro- the best strategy to avoid clinical mismanagement
cessing using linear regression analysis to determine due to unsuspected interference. For clinical use of
whether the regression equation best fit a model of Cobas 8000 e801 including patients taking more
a straight line or a nonlinear curve in the form of a than 5 mg/day of biotin supplement, we recom-
second or third-degree polynomial, according to mend waiting for 8 hours after supplementation
CLSI guideline EP6-A [10]. The goal of maintain- before blood collection [20]. American Thyroid
ing linearity is a fundamental characteristic of good Association Guidelines recommend that patients
analytic measurement methods and should follow do not take biotin for at least 2 days before thyroid
the first-degree model, which produces a straight- function tests [21].
line graph. However, in actual diagnostic tests, if
the deviation of the data set due to linearity is not In conclusion, the Cobas 8000 e801 analyzer
clinically important, such clinical results are con- showed excellent performance in terms of preci-
sidered acceptable according to clinically relevant sion, linearity, comparisons, and reference range
criteria for assessing linearity [11,15,16]. In this validation for 16 analytes tested according to the
study, the extremely low levels of eight analytes CLSI guidelines. Therefore, we conclude that the
seemed to represent second- or third-degree poly- e801 is a reliable component of the Cobas 8000
nomial results in formal statistical tests. Therefore, modular analyzer series for use in clinical laborato-
we further evaluated clinical relevance for data sets ries, although education and awareness about bio-
with nonlinear results. We found that most of the tin interference is necessary for successful imple-
deviation from linearity was not clinically relevant mentation in clinical practice.
Evaluation of the e801 immunoassay analyzer 379
Acknowledgements 11. Kroll MH, Præstgaard J, Michaliszyn E, Styer PE. Evaluation
of the extent of nonlinearity in reportable range studies.
This work was supported by the National Research Archives of pathology & laboratory medicine
Foundation of Korea(NRF) grant funded by the Korea 2000;124:1331-1338.
government(MSIT) (No. 2017R1D1A1B03036425). 12. Budd J, Durham A, Gwise T, Iriarte B, Kallner A, Linnet K,
et al. Measurement procedure comparison and bias estimation
References using patient samples: approved guideline. CLSI document
EP09-A3. Wayne (PA): Clinical and Laboratory Standards
1. Yalow RS, Berson SA. Immunoassay of endogenous plasma in- Institute 2013.
sulin in man. J Clin Invest 1960;39:1157-75. 13. Horowitz GL, Altaie S, Boyd JC. Defining, establishing, and
2. Hendriks H, Kortlandt W, Verweij W. Analytical performance verifying reference intervals in the clinical laboratory; ap-
comparison of five new generation immunoassay analyzers. proved guideline: CLSI, 2010
Nederlands Tijdschrift voor Klinische Chemie 14. Wagner D, Hanwell HE, Vieth R. An evaluation of automat-
2000;25:170-177. ed methods for measurement of serum 25-hydroxyvitamin D.
3. Markin RS, Whalen SA. Laboratory automation: trajectory, Clinical biochemistry 2009;42:1549-1556.
technology, and tactics. Clin Chem 2000;46:764-71. 15. Killeen AA, Long T, Souers R, Styer P, Ventura CB, Klee GG.
4. Kim S-Y, Jeong T-D, Lee W, Chun S, Min W-K. Performance Verifying performance characteristics of quantitative analyti-
evaluation of the Roche-Hitachi Cobas 8000 c702 chemistry cal systems: calibration verification, linearity, and analytical
autoanalyzer. Laboratory Medicine Online 2014;4:132-139. measurement range. Archives of pathology & laboratory med-
5. Ko K, Kwon M-J, Woo H-Y, Park H. Evaluation of Analytical icine 2014;138:1173-1181.
Performance of the Cobas 8000 Analyzer Series Module e602. 16. Jhang JS, Chang C-C, Fink DJ, Kroll MH. Evaluation of lin-
The Journal of Laboratory Medicine and Quality Assurance earity in the clinical laboratory. Archives of pathology & labo-
2015;37:101-109. ratory medicine 2004;128:44-48.
6. McEnroe R, Durham A, Goldford M, Kondratovich M, 17. Diamandis EP and Christopoulos TK. The biotin-(strept)avi-
Lababidi S, Magari R, et al. Evaluation of precision of quantita- din system: principles and applications in biotechnology. Clin
tive measurement procedures: approved guideline. Wayne: Chem 1991; 37: 625–636.
Clinical & Laboratory Standards Institute 2014. 18. Piketty ML, Polak M, Flechtner I, Gonzales-Briceno L,
7. Quality Requirements. Desirable Biological Variation Souberbielle JC. False biochemical diagnosis of hyperthyroid-
Database specifications. Available at: https://www.westgard. ism in streptavidin-biotin-based immunoassays: the problem
com/biodatabase1.htm. Updated on 2014 of biotin intake and related interferences. Clin Chem Lab
8. QC Design. What's the TEa for D? Allowable Error for vita- Med 2017;55:780-8.
min D. Available at: https://www.westgard.com/vitamin-d-al- 19. Li J, Wagar EA, Meng QH. Comprehensive assessment of bio-
lowable-error.htm. Updated on 2014 tin interference in immunoassays. Clin Chim Acta
9. Braga F, Ferraro S, Mozzi R, Dolci A, Panteghini M. Biological 2018;487:293-8.
variation of neuroendocrine tumor markers chromogranin A 20. http://dlslab.com/wp-content/uploads/bulletins/Biotin_
and neuron-specific enolase. Clinical biochemistry FDA_Safety_ Communication_ 011918.pdf
2013;46:148-151. 21. Peyro Saint Paul L, Debruyne D, Bernard D, Mock DM,
10. Tholen DW, Kroll M, Astles JR, Caffo AL, Happe T, Krouwer Defer GL. Pharmacokinetics and pharmacodynamics of
J, et al. Evaluation of the linearity of quantitative measurement MD1003 (high-dose biotin) in the treatment of progressive
procedures: a statistical approach; approved guideline: CLSI, multiple sclerosis. Expert Opin Drug Metab Toxicol
2003 2016;12:327-44.

You might also like