MECHANICAL
PARAMETERS TRIAL 1 TRIAL 2 TRIAL 3 MEAN
CHECKS
a) Misalignment in side A (cm, %) 0.41,0.41 0.41,0.41 0.41,0.41 0.41,0.41
b) Misalignment in side B (cm, %) 0.25,0.25 0.25,0.25 0.25,0.25 0.25,0.25
c) Misalignment in side C (cm, %) 0.20,0.20 0.20,0.20 0.20,0.20 0.20,0.20
d) Misalignment in side D (cm, %) 0.42,0.42 0.42,0.42 0.42,0.42 0.42,0.42
1. COLLIMATION e) Total misalignment (cm, %) 1.28,1.28 1.28,1.28 1.28,1.28 1.28,1.28
ASSESSMENT f) The x – ray beam
90ͦ ±1.5ͦ 90ͦ ±1.5ͦ 90ͦ ±1.5ͦ 90ͦ ±1.5ͦ
perpendicularity (degrees)
SETTINGS: SETTINGS:100cm /26 cm x 18 cm
/80kVp/100mA /0.1s
g) Settings
NOTE: Set X-ray Field Size base on Test Tools with
Serial Number
Misalignment in side B
• Page 2 of 5 •
MANUEL BIOMEDICAL
EQUIPMENT TRADING
BLK 38 LOT 45 GOLDEN HILLS SUBD. SAN VICENTE STA. MARIA BULACAN
Tel no. 09063071380 Email : cmgirayle@gmail.com
Misalignment in side C
Misalignment in side D
• Page 3 of 5 •
MANUEL BIOMEDICAL
EQUIPMENT TRADING
BLK 38 LOT 45 GOLDEN HILLS SUBD. SAN VICENTE STA. MARIA BULACAN
Tel no. 09063071380 Email : cmgirayle@gmail.com
• Page 4 of 5 •
MANUEL BIOMEDICAL
EQUIPMENT TRADING
BLK 38 LOT 45 GOLDEN HILLS SUBD. SAN VICENTE STA. MARIA BULACAN
Tel no. 09063071380 Email : cmgirayle@gmail.com
a c
Legends: Radiation dose/exposure Coefficient of variance (COV = SSD / Sample Mean)
b
Sample standard deviation d
Coefficient of linearity [COL = (Xmax – Xmin) / (Xmax + Xmin)]
e
Percent relative error
[% RE = (Set Value – Actual Value)(100 %) / Set Value]
Specifications:
Mechanical Collimation ● Allowable misalignment for each side = ± 1 % of SID
checks assessment ● Allowable total misalignment = ± 2 % of SID
● Allowable x-ray beam perpendicularity = 90° ± 1.5°
X – ray tube and ● The center of x – ray beam and the center of indicator light beam must be at the center
table bucky of image receptor
alignment
X – ray tube and ● The center of x – ray beam and the center of indicator light beam must be at the center
vertical of image receptor
cassette/bucky
alignment
Generator kVp Accuracy ● For < 100 kVp: actual kVp shall be within ± 6 % of the set kVp
checks ● For > 100 kVp: actual kVp shall be within ± 6 % of the set kVp
Exposure time ● For exposure time < 0.1 s, actual time shall be within ± 20 % of the set time
accuracy ● For exposure time > 0.1 s, actual time shall be within ± 10 % of the set time
● The COV shall be within 0.05
mA linearity ● The overall COL shall be ≤ 0.1
Performed by: Checked and attested by: Conformed and accepted by: