Professional Documents
Culture Documents
Iso Astm51940
Iso Astm51940
51702 Practice for Dosimetry in a Gamma Irradiation Fa- 3.1.2 absorbed-dose mapping—measurement of absorbed-
cility for Radiation Processing dose within process load using dosimeters placed at specified
51707 Guide for Estimating Uncertainties in Dosimetry for locations to produce a one-, two- or three-dimensional distri-
Radiation Processing bution of absorbed dose, thus rendering a map of absorbed-
51956 Practice for Use of Thermoluminescence-Dosimetry dose values.
(TLD) Systems for Radiation Processing 3.1.3 absorbed-dose rate, Ḋ—absorbed dose in a material
2.3 International Commission on Radiation Units and per incremental time interval, that is, the quotient of dD by dt
Measurements (ICRU) Reports:3 (see ICRU 60).
ICRU 14 Radiation Dosimetry: X-rays and Gamma Rays
Ḋ 5 dD/dt
with Maximum Photon Energies Between 0.6 and 50 MeV
ICRU 17 Radiation Dosimetry: X-rays Generated at Poten- SI Unit: Gy·s−1
tials of 5 to 150 kV 3.1.3.1 Discussion—The absorbed-dose rate can be speci-
ICRU 34 The Dosimetry of Pulsed Radiation fied in terms of average value of Ḋ over long-time intervals, for
ICRU 35 Radiation Dosimetry: Electron Beams with Ener- example, in units of Gy·min−1 or Gy·h−1.
gies Between 1 and 50 MeV 3.1.4 calibration—process whereby the response of a mea-
ICRU 60 Fundamental Quantities and Units for Ionizing suring system or measuring instrument is characterized through
Radiation comparison with an appropriate standard that is traceable to,
2.4 NCRP Publications:4 and consistent with, a national standard.
NCRP Report No. 69, Dosimetry of X-Ray and Gamma- 3.1.5 canister—durable, reusable container, usually an alu-
Ray Beams for Radiation Therapy in the Energy Range 10 minum or steel cylinder, used to house a product (for example,
--``````-`-`,,`,,`,`,,`---
keV to 50 MeV, December 1981 factory-reared insects in packaging containers) during the
radiation process.
3. Terminology 3.1.5.1 Discussion—Canisters are not used in some appli-
3.1 Definitions: cations in which the packaging container is sufficiently rigid
3.1.1 absorbed dose (D)—quantity of ionizing radiation and the design of the irradiator is appropriate.
energy imparted per unit mass of a specified material. The SI 3.1.6 dose uniformity ratio—ratio of maximum to minimum
unit of absorbed dose is the gray (Gy), where 1 gray is absorbed dose within the process load. The concept is also
equivalent to the absorption of 1 joule per kilogram of the referred to as the max/min dose ratio.
specified material (1 Gy = 1 J/kg). The mathematical relation- 3.1.7 dosimeter—device that, when irradiated, exhibits a
ship is the quotient of dē by dm, where dē is the mean quantifiable change in some property of the device which can
incremental energy imparted by ionizing radiation to matter of be related to absorbed dose in a given material using appro-
incremental mass dm (see ICRU 60). priate analytical instrumentation and techniques.
D 5 dē/dm 3.1.8 dosimeter set—one or more dosimeters used to mea-
3.1.1.1 Discussion—The discontinued unit for absorbed sure the absorbed dose at a location and whose average reading
dose is the rad (1 rad = 100 erg/g = 0.01 Gy). Absorbed dose is used to determine absorbed dose at that location.
is sometimes referred to simply as dose. For a photon source 3.1.9 dosimetry system—system used for determining ab-
under conditions of charged particle equilibrium, the absorbed sorbed dose, consisting of dosimeters, measurement instru-
dose, D, may be expressed as follows: ments and their associated reference standards, and procedures
for the system’s use.
D 5 F E µen/r 3.1.10 factory-reared insects—insects that are reared in
where: large quantity in a laboratory or factory setting for use,
F = particle fluence (particles/m2), following reproductive sterilization through irradiation, as live
E = energy of the ionizing radiation (J), and animals in pest management programs.
µen/r = mass energy absorption coefficient (m2/kg). 3.1.11 installation qualification—obtaining and document-
If bremsstrahlung production within the specified material is ing evidence that the irradiator, with all its associated equip-
negligible, the mass energy absorption coefficient (µen/r) is ment and instrumentation, has been provided and installed in
equal to the mass energy transfer coefficient (µtr/r), and accordance with specifications.
absorbed dose is equal to kerma if, in addition, charged particle 3.1.12 irradiator turntable—device used to rotate the can-
equilibrium exists. ister during the radiation process so as to improve dose
uniformity.
3
3.1.12.1 Discussion—An irradiator turntable is often re-
Available from the International Commission on Radiation Units and Measure-
ferred to as a turntable. Some irradiator geometries, for
ments, 7910 Woodmont Ave., Suite 800, Bethesda, MD 20814, USA.
4
Available from the National Council on Radiation Protection and Measure- example, with an annular array of radiation sources surround-
ments, 7910 Woodmont Ave., Bethesda, MD 20814, USA. ing the product, may not need a turntable.
4.4 Specifications for irradiation of factory-reared insects 5.3.1 Typically, accelerators produce a narrow electron
include a lower limit of absorbed dose and may include a beam that is diffused to cover the length and width of a target
central target dose and an upper limit. These values are based area, which may be the location where samples are irradiated
on program requirements and on scientific data on effects of (see 5.3.3) or an X-ray target (see 5.3.2). Diffusion may be
absorbed dose on the sterility, viability, and competitiveness of accomplished using a magnetic scanner (to sweep the beam
the factory-reared insects. back and forth rapidly), magnetic defocusing lens, or scattering
4.5 For each irradiator, absorbed-dose rate at a reference foils.
position within the irradiated volume of insects or simulated 5.3.2 X-rays (bremsstrahlung) are produced by striking an
product is measured using a transfer- or reference-standard X-ray target with an electron beam. The target is made of
dosimetry system. That measurement may be used to calculate tungsten, tantalum, or other metal with a high atomic number,
the timer setting, conveyor speed, or other parameter required high melting temperature, and high thermal conductivity.
to deliver the specified absorbed dose to the insects.
NOTE 2—Insects could be irradiated using a self-contained portable
4.6 Absorbed-dose mapping for establishing magnitudes
X-ray irradiator. The X-rays (bremsstrahlung) are produced in a conven-
and locations of Dmin and Dmax is performed using actual tional manner, but the unit is totally self-contained (requiring no additional
product or simulated product (9). shielding).
4.7 Dosimetry is part of a measurement quality assurance
plan that is applied to ensure that the radiation process meets 5.3.3 For processing, samples are typically carried on a
predetermined specifications (11). moving conveyor through the electron or X-ray beam. Because
of the narrow angular distribution of the radiation, use of
5. Types of facilities and modes of operation continuously moving conveyors (rather than static-irradiation
5.1 Self-Contained Dry-Source Irradiators (see Ref (12))— or shuffle-dwell systems) enhances dose uniformity.
Most insect sterilization is accomplished by using gamma 5.3.4 Additional information on electron accelerator facili-
radiation from either 137Cs or 60Co self-contained irradiators. ties and modes of operation may be found in ISO/ASTM
Practices 51649 (electron beam) and 51608 (X-ray).
--``````-`-`,,`,,`,`,,`---
These devices house the radiation source in a protective shield
of lead (or other appropriate high atomic number material), and
usually have a mechanism to rotate or lower the canister from 6. Radiation source characteristics
the load position to the irradiation position. 6.1 Gamma Irradiators:
5.1.1 A common method is to distribute the source in an 6.1.1 The radiation source used in the gamma facilities
annular array. During irradiation, the sample of factory-reared considered in this guide consists of sealed elements of 60Co
insects is located at the center of the array, where the or 137Cs which are typically linear rods or “pencils” arranged
absorbed-dose distribution is relatively uniform. In this design, in one or more planar or cylindrical arrays.
an irradiator turntable would not normally be necessary.
6.1.2 Cobalt-60 emits photons with energies of approxi-
5.1.2 A second method is to rotate the canister holding the
mately 1.17 and 1.33 MeV in nearly equal proportions (13).
insects on an irradiator turntable within the radiation field to
Cesium-137 emits photons with energies of approximately
achieve a more uniform dose distribution within the process
0.662 MeV (13).
load. The axis of rotation should be parallel to the source
6.1.3 The half-lives for 60Co and 137Cs are 5.2708 years
pencils (see 6.1.1), which are generally vertical.
(14) and 30.07 years (15), respectively.
5.2 Large-Scale Gamma Irradiators—Gamma irradiation
of insects is also carried out in large-scale irradiators, either 6.1.4 For gamma-radiation sources, the only variation in the
pool-type or dry-storage. In these facilities, the source typically source output is the known reduction in the activity caused by
consists of a series of rods (pencils) that contain 60Co and can radioactive decay. The reduction in the source strength, and the
be raised or lowered into a large irradiation room. When required increase in the irradiation time, may be calculated (see
retracted from the irradiation room, the source is shielded by 9.3.4) or obtained from tables provided by the irradiator
water (pool-type), or lead or other appropriate high atomic manufacturer.
number material (dry-storage), or both. 6.2 Electron Accelerator (Electron and X-ray Modes):
5.2.1 Continuous Operation—A common method of use is 6.2.1 For an electron accelerator, the two principal beam
for samples of insects to be carried on a conveyor in one or characteristics are the energy spectrum and the average beam
more revolutions around a central source, resulting in a current. The electron energy spectrum affects the variation of
relatively uniform absorbed-dose distribution. The source is absorbed dose with depth in a given material, and the average
retracted from the irradiation room only when the irradiator is beam current affects the absorbed-dose rate.
not in use. 6.2.1.1 Direct-action electron accelerators employ dc or
5.2.2 Batch Operation—An alternative method of use is to pulsed high-voltage generators typically produce electron en-
place canister(s) of insects into the irradiation room while the ergies up to 5 MeV.
source is shielded, and then raise or lower the source into the 6.2.1.2 Indirect-action electron accelerators use microwave
chamber for the length of time required to achieve the desired or very high frequency (VHF) ac power to produce electron
absorbed dose. energies typically from 5 MeV to 15 MeV.
5.3 Electron Accelerator—Accelerator-generated radiation 6.2.2 For an X-ray (bremsstrahlung) facility, besides beam
is in the form of electrons or X-rays (bremsstrahlung). characteristics noted in 6.2.1, X-ray target design is a critical
parameter. X-rays are similar to gamma radiation from radio- 7.2 Calibration of Dosimetry Systems:
active isotopic sources. Although their effects on materials are 7.2.1 Prior to use, the dosimetry system (consisting of a
generally similar, these kinds of radiation differ in their energy specific batch of dosimeters and specific measurement instru-
spectra, angular distributions, and absorbed-dose rates. The ments) shall be calibrated in accordance with the user’s
continuous energy spectrum of the X-rays (bremsstrahlung) documented procedure that specifies details of the calibration
varies from approximately 35 keV up to the maximum energy process and quality assurance requirements. This calibration
of the electrons incident on the X-ray target (see ISO/ASTM procedure shall be repeated at regular intervals to ensure that
Practice 51608). In some X-ray facilities, spectrum filtration is the accuracy of the absorbed dose measurement is maintained
used to reduce the low energy component of the radiation, thus within required limits. Calibration methods are described in
improving dose uniformity. ISO/ASTM Guide 51261.
--``````-`-`,,`,,`,`,,`---
7.2.2 Irradiation is a critical component of the calibration of
7. Dosimetry systems
the dosimetry system.
7.1 Description of Dosimeter Classes: 7.2.3 Calibration of Reference or Transfer Dosimeters—
7.1.1 Dosimeters may be divided into four basic classes Calibration irradiations shall be performed at an accredited
according to their accuracy and areas of application: primary calibration laboratory, or in-house calibration facility meeting
standard, reference standard, transfer standard, and routine the requirements of ISO/ASTM Practice 51400, that provides
dosimeters. ISO/ASTM Guide 51261 provides information an absorbed dose (or absorbed-dose rate) having measurement
about the selection of dosimetry systems for different applica- traceability to nationally or internationally recognized stan-
tions. All classes of dosimeters, except primary standard, dards.
require calibration before their use. 7.2.4 Calibration of Routine Dosimeters—Calibration irra-
7.1.1.1 Primary-Standard Dosimeters—Primary-standard diations may be performed per 7.2.3, or at a production or
dosimeters are established and maintained by national stan- research irradiation facility together with reference- or transfer-
dards laboratories for calibration of radiation environments standard dosimeters that have measurement traceability to
(fields) and other dosimeters. The two most commonly used nationally or internationally recognized standards.
primary standard dosimeters are ionization chambers and 7.3 Measurement Instrument Calibration and Performance
calorimeters (see ISO/ASTM Guide 51261, ICRU Reports 14, Verification—For the calibration of instruments, and for the
17, 34, and 35 and NCRP Report 69). verification of instrument performance between calibrations,
7.1.1.2 Reference-Standard Dosimeters—Reference- see ISO/ASTM Guide 51261, the corresponding ISO/ASTM or
standard dosimeters are used to calibrate radiation environ- ASTM standard for the dosimetry system, and/or instrument-
ments and routine dosimeters. Reference-standard dosimeters specific operating manuals.
may also be used as routine dosimeters. Examples of reference-
standard dosimeters along with their useful dose ranges are 8. Installation qualification
given in a table in ISO/ASTM Guide 51261. For insect
8.1 Objective—The purpose of an installation qualification
irradiators, the following reference-standard dosimeters may
program is to obtain and document evidence that the irradiator
be suitable: ferrous sulfate (Fricke) aqueous solution (ASTM
and measurement instruments have been delivered and in-
Practice E 1026), alanine dosimeters (ISO/ASTM Practice
stalled in accordance with their specifications. Installation
51607), radiochromic solutions (ISO/ASTM Practice 51540
qualification includes documentation of the irradiator equip-
and Ref (16)), and ethanol-chlorobenzene solution (ISO/
ment and measurement instruments; establishment of testing,
ASTM Practice 51538).
operation and calibration procedures for their use; and verifi-
7.1.1.3 Transfer-Standard Dosimeters—Transfer-standard
cation that the installed irradiator equipment and measurement
dosimeters are specially selected dosimeters used for transfer-
instruments operate according to specification.
ring absorbed-dose information from an accredited or national
standards laboratory to an irradiation facility in order to NOTE 3—Table A2.1 gives some recommended steps in the following
establish traceability for the facility. These dosimeters should areas: installation qualification, operational qualification, performance
be used under conditions that are specified by the issuing qualification, and routine product processing. The recommended steps in
Table A2.1 are not meant to be exhaustive.
laboratory. Transfer-standard dosimeters may be selected from
either reference-standard dosimeters or routine dosimeters 8.2 Equipment Documentation—Establish and document an
taking into consideration the criteria listed in ISO/ASTM installation qualification program that includes descriptions of
Guide 51261. the instrumentation and equipment and measurement instru-
7.1.1.4 Routine Dosimeters—Routine dosimeters may be ments installed at the facility. This documentation shall be
used for radiation process quality control, dose monitoring, and retained for the life of the facility. At a minimum, it shall
dose mapping. Proper dosimetric techniques, including cali- include:
bration, shall be employed to ensure that measurements are 8.2.1 A description of the irradiator’s specifications, char-
reliable and accurate. Examples of routine dosimeters, along acteristics and parameters, including any modifications made
with their useful dose ranges, are given in ISO/ASTM Guide during or after installation,
51261. Examples of dosimeters that may be useful in routine 8.2.2 A description of the location of the irradiator within
processing or absorbed-dose mapping for insect irradiation the operator’s premises, including its relation to any means
facilities are listed in Annex A1. provided for segregating unirradiated from irradiated products,
8.2.3 Operating instructions and standard operating proce- 9.1.2 Specific information on operational qualification for
dures for the irradiator and associated measurement instru- facilities other than self-contained dry-storage gamma irradia-
ments, tors can be found in ISO/ASTM Practices 51608 (for X-ray
8.2.4 Description of the construction and operation of the facilities), 51649 (for electron beam facilities), and 51702 (for
product handling system, large-scale gamma facilities).
8.2.5 Licensing and safety documents and procedures, in- 9.2 Dosimetry Systems—Calibrate the routine dosimetry
cluding those required by regulatory and occupational health system to be used at the facility as discussed in 7.2.
and safety agencies, 9.3 Irradiator Characterization—The absorbed dose re-
8.2.6 A description of a calibration program to ensure that ceived by any portion of product (in this case, insects) depends
all processing equipment that may influence absorbed-dose on the irradiator parameters (such as the source activity at the
delivery is calibrated periodically (for example, the reset timer time of irradiation, the geometry of the source, the source-to-
mechanism on a gamma irradiator), and product distance, the irradiation geometry) and the processing
8.2.7 Descriptions, operating procedures, and calibration parameters (such as the irradiation time, the product composi-
procedures for associated measurement instruments or systems tion and density, and the loading configuration).
(such as those used for dosimetry). 9.3.1 Absorbed Dose Rate—A reference- or transfer-
8.3 Equipment Testing and Calibration—Test all processing standard dosimetry system, traceable to nationally or interna-
equipment and instrumentation that may influence absorbed tionally recognized standards, shall be used to measure the
dose in order to verify satisfactory operation of the irradiator absorbed-dose rate within product or simulated product at a
within the design specifications. reference position (such as the center of the product or
8.3.1 Implement a documented calibration program (8.2.6) simulated product volume) in near worst-case geometry (such
to ensure that all processing equipment and instrumentation as when the product or simulated product nearly completely
that may influence absorbed-dose delivery are calibrated peri- fills the irradiation volume). The absorbed dose rate at the
odically. reference position shall have a reproducible and documented
8.3.2 If any modification or change is made to the irradiator relationship to the absorbed dose rate at locations of maximum
equipment or measurement instruments during the installation (Dmax) and minimum (Dmin) dose rate (see 9.3.2). This mea-
qualification phase, they shall be re-tested. surement of absorbed dose rate at a reference position may be
8.4 Specific information on installation qualification for used when calculating the timer setting necessary to deliver the
various types of facilities can be found in ASTM Practice specified range of absorbed dose to the insects (see 10.4).
E 2116 (self-contained dry-storage gamma irradiators) and in 9.3.1.1 Most manufacturers of insect irradiators use a
ISO/ASTM Practices 51608 (X-ray [bremsstrahlung]), 51649 reference-standard dosimetry system to measure absorbed dose
(electron beam), and 51702 (large-scale gamma facilities). rate at a reference position within simulated product following
8.4.1 For self-contained irradiators, installation qualification installation of (or, in the case of some self-contained units,
may begin prior to the shipment of the irradiator to the before shipping) the irradiator.
customer’s site. 9.3.1.2 Reference- or transfer-standard measurement of ab-
9. Operational qualification sorbed dose rate at a reference position should be repeated
periodically (for example, every three years for a gamma
9.1 Objective—The purpose of operational qualification of facility) and following any changes to the source, geometry, or
an irradiation facility is to establish baseline data for evaluating other irradiator parameter that could affect absorbed dose rate.
irradiator effectiveness, predictability, and reproducibility for
the range of conditions of operation for key processing NOTE 4—To the degree possible, subsequent re-calibrations of dose rate
parameters that affect absorbed dose in the product. As part of should be performed under similar conditions to allow direct comparison
among test results. Results obtained when re-calibrating an irradiator
this process, dosimetry may, for example, be performed to: (1) should agree with results of previous calibrations, once source decay (if
establish relationships between the absorbed dose for a repro- applicable) or other known factors that may affect dose are taken into
ducible geometry and the operating parameters of the irradia- account. Unexplained discrepancies that are beyond the limit of combined
tor, (2) measure absorbed-dose distributions in insect- uncertainty for the two procedures should be investigated, as they could
equivalent material and other reference materials, (3) indicate problems with the dosimetry or the operation of the irradiator.
characterize absorbed-dose variations when irradiator and NOTE 5—When an irradiator’s absorbed dose rate is measured per
processing parameters fluctuate statistically through normal 9.3.1, it is convenient to calibrate the facility’s routine dosimetry system
operations, and (4) measure the absorbed-dose rate at a concurrently per 7.2.4. ISO/ASTM Guide 51261 and Ref (17) provide
guidelines on procedures and numbers of sets of dosimeters needed.
reference position within the canister filled with insects or
simulated product. 9.3.2 Dose Mapping—Ideally, the irradiation process is
9.1.1 For self-contained irradiators, operational qualifica- designed to irradiate insects uniformly throughout the irradi-
tion may begin prior to the shipment of the irradiator to the ated volume; in reality, a certain variation in absorbed-dose
customer’s site. As part of release-for-shipment criteria, the through the product will exist. The irradiator characterization
irradiator manufacturer may perform absorbed-dose mapping process includes mapping the absorbed-dose distributions for
to establish baseline data. After the unit is installed at the user’s samples of insects or simulated product, and identifying the
site, operational qualification is performed as part of the user’s magnitudes and locations Dmax and Dmin within the samples.
quality assurance plan (see ASTM Practice E 2116). Dosimetry data from previously characterized irradiators of the
--``````-`-`,,`,,`,`,,`---
same design or theoretical calculations may provide useful increased dose. Understanding the dose distribution throughout
information for determining the number and locations of the process load is critical for ensuring program security and
dosimeter sets needed for this characterization process. sterile insect quality. This is accomplished by absorbed-dose
9.3.2.1 Map the absorbed-dose distribution by placing do- mapping (see 10.3) of the specific product (or simulated
simeter sets throughout the volume of actual or simulated product) and specific product loading configuration to deter-
product (see ASTM Guide E 2303). Select placement patterns mine the magnitude and location of Dmax and Dmin, and to
that can identify the locations of Dmax and Dmin. establish the appropriate values for the irradiator timer setting,
NOTE 6—In the case of static irradiations (such as when the product is
conveyor speed, or other parameter(s) necessary to achieve the
located at the center of an annular source array of a self-contained absorbed doses within the set requirements.
dry-storage irradiator), the dose mapping should be done in three NOTE 7—In sterile insect programs, tests of absorbed dose versus
dimensions. When product is irradiated on turntables, the dose mapping sterility and viability of irradiated insects are critical (11). Thus, tests to
may be done in two dimensions, such as on an arbitrary vertical plane evaluate the sterility and competitiveness of irradiated insects are con-
through the axis of rotation. In this case, the result is a three-dimensional ducted periodically to provide evidence that the dose limit and other
mapping due to the axial symmetry that results from rotation. aspects of sterilization processing (for example, induction of hypoxia prior
9.3.2.2 Changes in the product handling system (for ex- to irradiation) are still valid.
ample, irradiator turntable) or radiation source characteristics 10.2 Product Loading Configuration—A loading configura-
require a new absorbed-dose mapping. tion for the irradiation should be established for each product
9.3.3 In gamma-radiation facilities that operate in batch type. The documentation for this loading configuration shall
mode, the transit dose should be small relative to the total dose include specifications for parameters that influence the ab-
delivered to the insects (for example, <1 %) in order to sorbed dose distribution. For irradiation of insects, these
facilitate reproducible absorbed-dose delivery. The transit dose parameters could include species, mass or volume of the
and its relation to total absorbed dose should be considered and insects, size and shape of the packaging container, position of
quantified, if necessary. Procedures for measuring and correct- the packaging container(s) within the canister, and position and
ing for transit dose in terms of transit time are given in composition of any plugs of simulated product that are used
ISO/ASTM Guide 51261. routinely to improve dose uniformity by excluding insects from
9.3.4 An important calculation in the use of gamma- portions of the canister.
radiation sources is the correction for radioactivity decay. For
NOTE 8—The canister shall not be loaded beyond its designed maxi-
a pure radionuclide source, the exponential loss of activity, A,
mum volume.
is given by:
10.3 Product or Simulated Product Absorbed-Dose
At 5 A0e 2lt (1)
Mapping—For each type of insect treated in the irradiation
--``````-`-`,,`,,`,`,,`---
Where At is the activity at time t, A0 is the known activity at facility, there is a minimum dose to achieve the required level
some earlier time (t = 0), and l is the decay constant for the of sterility (or other desired effect) and a maximum dose that
given radionuclide. Daily decay constants (ld) for radionu- the animal can tolerate without unacceptable degradation in
clides commonly used in gamma-radiation sources are: quality. Establish the locations of the regions of Dmax and Dmin
For 60Co, ld 5 3.6005 3 1024 day 21 (2) for each selected product loading configuration by placing
dosimeter sets throughout the volume of insects or simulated
product within the process load (see ASTM Guide E 2303).
For 137Cs, ld 5 6.311 3 1025 day21 (3)
Concentrate the dosimeters in regions of Dmax and Dmin with
These constants are based on half lives of 5.2708 years fewer dosimeters placed in areas likely to receive intermediate
for 60Co (14) and 30.07 years for 137Cs (15). In practice, absorbed dose. In many insect irradiation applications, the
absorbed dose rate can be substituted for activity in Eq 1. The product is relatively close to the radiation sources, resulting in
absorbed-dose rate that was initially established during opera- pronounced absorbed-dose gradients near the periphery of the
tional qualification or during a subsequent calibration of the volume of the sample. It is important, therefore, to choose a
irradiator with reference- or transfer-standard dosimeters (see dosimeter that is small enough to detect these gradients.
9.3.1) provides the value at t = 0. The absorbed-dose rate at t Dosimeter film in strips or sheets may be employed to obtain
days later can then be computed from Eq 1 using ld for the useful information (9).
appropriate radionuclide from Eq 2 or Eq 3. 10.3.1 Results of absorbed-dose mapping will be used to
determine the degree of dose uniformity. Because the quality
10. Performance qualification and viability of insects tends to decline rapidly as radiation
10.1 Objective—The purpose of dosimetry in performance dose is increased, a small dose-uniformity ratio can be critical
qualification is to ensure that the absorbed-dose requirements to ensuring the successful deployment of sterile insects. In
for a particular product can be satisfied. In sterile insect release some cases, irradiator or processing parameters can be adjusted
programs, the end user or a regulatory agency typically to improve dose uniformity (for example, installing an irradia-
specifies the minimum absorbed dose necessary to produce the tor turntable or using plugs of simulated product to exclude
desired level of reproductive sterility. Although a maximum insects from areas with low or high dose rates).
absorbed dose is not usually specified, the ability of sterile 10.3.2 If any changes that could affect the magnitude or
insects to successfully compete for mates will decline with location of the absorbed dose extremes are made to the
irradiator or mode of operation (for example, those noted in NOTE 10—The absorbed-dose distribution in the product is already
10.3.1), repeat the absorbed-dose mapping to the extent nec- known from operational and performance qualification, and from the most
essary to establish the effect. recent dose mapping. However, strategic placement of a sufficient number
of dosimeter sets as part of routine dosimetry can serve to confirm that the
10.3.3 Reference Dose Location—Identify a reference dose absorbed dose delivered is within specification.
location for each loading configuration. This may be, for
example, the location of Dmin or Dmax, or an alternate location 11.2.2 Dosimeter Placement Frequency—Select a sufficient
in or on the canister or packaging container. Dosimeter sets number of process loads in which to place dosimeter sets in
should be placed at the reference dose location during routine order to verify that the measurements of absorbed dose are
dosimetry (see 11.2), so accessibility should be considered. statistically meaningful and absorbed dose received by the
The absorbed dose at this location shall have a reproducible insects falls within specified limits.
and documented relationship to the absorbed dose at the 11.2.2.1 Gamma-radiation Facilities—Routine dosimetry
should be performed at specified periodic intervals, which may
--``````-`-`,,`,,`,`,,`---
locations of Dmin and Dmax within the process load.
10.4 Establishing Operating Parameters—To ensure that be daily, weekly, or monthly.
the absorbed dose received by processed insects is within NOTE 11—Although source output in a gamma irradiator is affected
specified limits, values of operating parameters should be only by radioactive decay, frequent routine dosimetry can detect otherwise
established for each combination of insect species, loading unnoticed processing problems which have, at times, resulted in the
configuration, and dose specification. Value(s) of all param- release of large numbers of fertile insect pests.
eters that affect absorbed dose are established based on results 11.2.2.2 Electron and X-ray (bremsstrahlung) Facilities—
of the absorbed-dose mapping described in 10.3 in conjunction Always place dosimeter sets at the start of a production run.
with results of reference- or transfer-standard measurements of For long runs, in addition to this, place dosimeter sets near the
absorbed dose rate at a reference position (see 9.3.1). For most middle of the run, at the end of the run, and at other intervals
insect irradiation facilities, the absorbed dose delivered to the as appropriate.
insects is controlled by adjusting a single operating parameter
NOTE 12—For production runs with contiguous process loads, the first
such as timer setting or conveyor speed. The value that is and last process loads may experience dose distributions different from the
established for that parameter shall result in an absorbed dose other units. If prior dosimetry data indicate that an unacceptable dose
distribution that is within specified limits throughout the distribution exists for these two end loads, place compensating dummies
volume of irradiated insects. adjacent to these units so as to make their dose distributions acceptable.
11.4 Environmental Effects—The response of a dosimeter or appropriate locations in the irradiation volume (so that the near
radiation-sensitive indicator may be affected by exposure to worst-case is approximated), and by placing the sample of
such environmental conditions as heat, high humidity, ultra- insects in the center of the canister.
violet radiation, or gases produced during the radiation process.
Dosimeter response should be corrected for any effects that 12. Measurement uncertainty
arise from changes in the environment of the dosimeter during 12.1 To be meaningful, a measurement of absorbed dose
the radiation process or pre- or post-irradiation storage. A shall be accompanied by an estimate of uncertainty.
radiation-sensitive indicator’s response cannot be corrected for 12.2 Components of uncertainty shall be identified as be-
such conditions, and they should not be used in inappropriate longing to one of two groups:
environments. Care should also be taken in handling and 12.2.1 Type A—Those which are evaluated by statistical
storage of dosimeters and indicators before and after irradiation methods, or
(see ISO/ASTM Guides 51261 and 51539, and practices for 12.2.2 Type B—Those which are evaluated by other means.
individual dosimetry systems listed in Annex A1). 12.3 Other ways of categorizing uncertainty have been
11.5 Chilled Product—Absorbed dose is not a function of widely used and may be useful for reporting uncertainty. For
the temperature of the insects being irradiated. The response of example, the terms precision and bias or random and system-
a dosimeter or radiation-sensitive indicator, however, may be a atic (non-random) are used to describe different categories of
function of their temperature. Dose-mapping information for uncertainty.
simulated product (representing the actual product geometry or
near-worst case geometry) at ambient temperature can be NOTE 15—The identification of Type A and Type B uncertainties is
applied to chilled insects. Determine the temperature of the based on methodology for estimating uncertainties published in 1995 by
the International Organization for Standardization (ISO) in the Guide to
dosimeter during irradiation of chilled insects and apply the the Expression of Uncertainty in Measurement (18). The purpose of using
appropriate temperature correction. Dosimeters that exhibit a this type of characterization is to promote an understanding of how
highly temperature-dependent response should not be placed in uncertainty statements are developed and to provide a basis for the
locations with large temperature gradients (see ISO/ASTM international comparison of measurement results.
Guide 51261 and the practices for individual dosimetry sys- NOTE 16—ISO/ASTM Guide 51707 defines possible sources of uncer-
tems listed in Annex A1). Radiation-sensitive indicators should tainty in dosimetry performed in radiation processing facilities, and offers
not be used on chilled insects unless they have been evaluated procedures for estimating the resulting magnitude of the uncertainties in
and proven to perform adequately at the temperatures used for the measurement of absorbed dose using a dosimetry system. The
document defines and discusses basic concepts of measurement, including
processing samples. the estimation of the measured value of a quantity, “true” value, error and
11.6 Partially Loaded Canisters—Irradiations may be per- uncertainty. Components of uncertainty are discussed and methods are
formed using less product than was used in dose mapping that provided for estimating their values. Methods are also provided for
was performed with the near-worst case geometry (for ex- calculating the combined standard uncertainty and estimating expanded
ample, when the canister is nearly filled with product or (overall) uncertainty.
simulated product). In that case, the Dmax received by the
insects may be greater than the Dmax measured during dose 13. Keywords
mapping for operational qualification (10.3). Care should be 13.1 absorbed dose; biological control; cesium-137; cobalt-
taken, therefore, to ensure that the specified limits for Dmax are 60; electron accelerator; dosimeter; dosimetry; gamma; insect
not exceeded during routine use. Changes to the absorbed-dose irradiation; insect rearing; ionizing radiation; irradiation; irra-
distribution arising from partially loaded canisters may be diator characterization; parasitoid rearing; radiation; sterile
minimized by the use of simulated product placed at the insect technique; X-ray; ICS 17.240
ANNEXES
(informative)
A1.1 Thermoluminescence dosimeter (TLD) or suspended in plastics. After irradiation, crystalline material
−2 4 is subjected to a carefully controlled heating cycle, when the
A1.1.1 Applicable Dose Range—10 to 10 Gy.
freed electrons and hole traps recombine with the emission of
A1.1.2 Applicable Dose Rate—10−2 to 1010 Gy/s. characteristic light. This heating cycle erases the dose infor-
A1.1.3 Use—Electron/gamma radiation/X-ray. mation in the TLD.
A1.1.4 Physical Characteristics—The most commonly A1.1.5 Instrumentation Characteristics—TL reader com-
used materials for TLD are LiF, CaF2, CaSO4 and Li2B4O7. prising a heating element, photomultiplier tube measurement
The dosimeter is small, and the material is used in the form of system to measure light output and convert to absorbed dose.
powder, pellets, single crystals, or in sealed glass tubes or bulbs Reader requires skilled operator.
--``````-`-`,,`,,`,`,,`---
--``````-`-`,,`,,`,`,,`---
A1.3 Alanine EPR dosimetry system A1.4.6.2 Humidity—Radiation response typically not af-
A1.3.1 Applicable Dose Range—1 to 105 Gy. fected by humidity.
A1.3.2 Applicable Dose Rate—<108 Gy/s. A1.4.6.3 Ambient Light—Dosimeters are sensitive to light,
A1.3.3 Use—Electron/gamma radiation/X-ray. especially during the irradiation and stabilization periods.
A1.3.4 Physical Characteristics—This dosimeter is used in A1.4.6.4 Time—Signal increases with time following irra-
the form of tablets, small rods, rope of 3-mm to 5-mm diameter diation and requires a signal development process, which can
and various lengths, or on bar-coded film strips, consisting be accelerated and stabilized by heat-treating the dosimeters
primarily of a-alanine and a small amount of paraffin or other after irradiation and before readout.
binder. A1.4.7 For more information, see ASTM Guide E 2304.
Bibliography
(1) Knipling, E. F., “Possibilities of Insect Control or Eradication Through 521A”, Isomedix, Inc., Whippany, NJ, 1984.
the Use of Sexually Sterilized Males,” Journal of Economic Entomol- (9) Walker, M. L., McLaughlin, W. L., Puhl, J. M., and Gomes, P.,
ogy, Vol 48, 1955, pp. 459-466. “Radiation Field Mapping of Insect Irradiation Canisters,” Applied
(2) Declaration of Emergency because of Mediterranean, Mexican, and Radiation and Isotopes, Vol. 48, 1997, pp. 117-125.
Oriental Fruit Flies, Federal Register, Vol 57, No. 13, Jan. 21, 1992, (10) Zavala, J. L., Fierro, M. M., Schwarz, A. J., Orozco, D. H., and
p. 2249. Guerra, M., “Dosimetry Practice for the Irradiation of the Mediter-
(3) Bushland, E. B. and Hopkins, D. E., “Sterilization of Screw-Worm ranean Fruit Fly Ceratitis capitata (Wied),” In High Dose Dosimetry,
Flies Sterilized by X-rays,” Journal of Economic Entomology, Vol. 46, Proceedings of the International Symposium, IAEA STI/PUB/671,
1953, pp. 648-656. International Atomic Energy Agency, Vienna, 1985, pp. 23-30.
(4) McLaughlin, W. L., Hussman, E. K., Eisenlohr, H. H. and Chalkley, L., (11) “Manual for Product Quality Control and Shipping Procedures for
“A Chemical Dosimeter for Monitoring Gamma Radiation Doses of 1 Mass-Reared Tephritid Fruit Flies,” FAO/IAEA/USDA Report, Ver-
- 100 krad,” International Journal of Applied Radiation and Isotopes, sion 5.0, International Atomic Energy Agency, Vienna, Austria, 2003.
Vol. 22, 1971, pp. 135-140. (12) ANSI N-433.1, “Safe Design and Use of Self-Contained, Dry-
(5) Sivinski, J. and Smittle, B., “Effects of Gamma Radiation on the Storage Gamma Irradiators,” American National Standards Institute,
Development of the Caribbean Fruit Fly ( Anastrepha suspensa) and 25 W. 43rd St., 4th Floor, New York, NY, USA, 10036, 1978.
the subsequent development of its parasite Diachasmimorpha longi- (13) Handbook of Chemistry and Physics, 71st ed., Lide, D. R., ed., CRC
caudata,” Entomologia experimentalis et applicata, Vol. 55, 1990, pp. Press, Boca Raton, FL, 1990.
295-297. (14) Unterweger, M.P., Hoppes, D. D., and Shima, F. J.,“ New and
(6) Darden, E. B. Jr., Maeyens, E., and Bushland, R. C., “A Gamma-Ray Revised Half-Life Measurement Results,” Nuclear Instruments and
Source for Sterilizing Insects,” Nucleonics, Vol. 12, No. 10, 1954, pp. Methods, Physics A, Vol 312, 1992, pp. 349-352.
60-62. (15) Tuli, J. K., “Nuclear Data Sheets Update for A = 137,” Nuclear Data
(7) Seo, S. T., “Husman Irradiator: Calibration Procedure,” Internal Sheets, Vol 72, 1994, p. 136.
Report, USDA-ARS, Washington, D.C. 1993. (16) Hjortenburg, P. E. and McLaughlin, W. L., “Use of Radiochromic
(8) “Descriptive and Operating Manual for Husman Irradiator Model Dye Systems for Dosimetry,” Proceedings of Regional Conference on
Radiation Protection, Jerusalem, March 1973. Published by Israel (18) “Guide to the Expression of Uncertainty in Measurement,” Interna-
Atomic Energy Commission, Soreq Nuclear Research Centre, Yavne, tional Organization for Standardization, 1995 ISBN 92-67-10188-9.
1973, pp. 122-140. Available from the International Organization for Standardization, 1
(17) “Gafchromic dosimetry system for SIT: Standard Operating Proce- rue de Varembé, Case Postale 56, CH-1211, Geneva 20, Switzerland.
dure,” v. 1.0, IAEA/FAO Report, International Atomic Energy
Agency, Vienna, Austria, 2001.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ISO, Case postale 56, CH-1211, Geneva 20, Switzerland, and ASTM International, 100 Barr Harbor
Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States. Individual reprints (single or multiple copies) of this
standard may be obtained by contacting ASTM at the above address or at 610-832-9585 (phone), 610-832-9555 (fax), or
service@astm.org (e-mail); or through the ASTM website (www.astm.org).
--``````-`-`,,`,,`,`,,`---