Professional Documents
Culture Documents
Cw213623 Sokaij Qa 5733 00002 Non Destructive Tests Procedure
Cw213623 Sokaij Qa 5733 00002 Non Destructive Tests Procedure
Cw213623 Sokaij Qa 5733 00002 Non Destructive Tests Procedure
PROJECT TITLE:
REPAIR OF SOKU PWD DISPOSAL LOOP LINE TO ACE LINE WORKS
CLIENT (SPDC)
Action Reviewed by Agreed By Approved By
Name Omon Ehizoyanyan Adeshina Chukwunalu
Designation QA/QC Coordinator Project Engineer Contract Holder
Sign/Date
1. Scope
This procedure provides the general requirements, methods, techniques, and acceptance standards
for the radiographic examination of butt welds in plate, formed products and tubular products for
wall thickness range 2.5 mm to 70mm.
2. Reference Documents:
API 1104 Ed.2021
ASME B31.3
ASME SECTION V
RELEVANT DEP
Written Practice: TGRL/WP/001 Rev.01 as per SNT–TC–1A (Ed.2016)
3. Personal Qualification
3.1. Personnel performing examination to this procedure shall be qualified and certified according to
Written Practice for Training, Qualification and Certification of NDE Personnel as per
TGRL/WP/001 Rev.01
4. SAFETY
4.1. and NDT subcontractor Inspectors and Technicians shall ensure environmental conditions are
suitable for the use of inspection, measuring and testing equipment used during radiographic
testing. Equipment shall be safeguarded against damage and malfunction by regular checks and
calibration. Thermo-Illumine scent Dosimeter Badges shall be worn at all time while carrying out
the task.
4.2. For handling or radiation emergency and investigation and medical surveillance see the
following procedures:
i. Local Rules for Site and installation Radiography
ii. Radiation protection procedure.
iii. Safety manual.
5. Surface Preparation
1) Surface shall satisfy the requirements of the applicable materials specification with additional
Conditioning, if necessary, by any suitable process to a degree that surface irregularities cannot
mask or be confused with discontinuities.
2) Weld ripples or weld surface irregularities on both the inside (where accessible) and outside shall
be removed by any suitable process to such a degree that the resulting Radiographic image cannot
mask or be confused with the image of a discontinuity.
3) The Finished surface of all butt-welded joints may be flush with the base material or may have
reasonably uniform crowns with reinforcement not to exceed that specified in the referencing code
section.
5.1. Equipment
The radiographic examination covered by this procedure shall be performed with the following
equipment:
Gamma Ray Projector and related accessories using Iridium 192 radioisotope.
X-Ray generator having maximum of 250kV.
Manual Processing Dark–room equipment and accessories
Suitable viewing apparatus for Radiographic Interpretation
Calibrated Optical Densitometer
7.1.2 Selection shall be based on Apendix-3 for hole type and Wire type.
7.1.3 The thickness based on which IQI is selected is the nominal single wall thickness plus
any estimated weld reinforcement not to exceed maximum permitted by referencing,
Code section, as stipulated in Apendix-1. Backing rings / strips shall not be considered
as a part of thickness during IQI selection.
7.1.4 For weld without reinforcement, IQI is selected based on nominal single wall thickness.
7.2.1 The IQI (s) shall be placed on the source side of the part being examined. Where
inaccessibility prevents hand placing the IQI on the source side, it shall be placed on the
Film side in contact with the part being examined. A lead letter "F" at least as high as
the IQI, shall be placed adjacent to or on the IQI(s), but shall not mask the essential hole
where a hole type IQI’s are used.
7.2.2 For double wall double image or single wall single image technique, except panoramic
exposure, IQI shall be placed on the source side. For double wall single image and
panoramic technique IQI to be placed on the film side.
For welds, a shim of material radiographically similar to the weld metal shall be placed
between the part and the IQI, if needed, so that the radiographic density through the area
of interest is no more than minus 15 % from (lighter than) the radiographic density through
IQI.
The shim dimensions shall exceed the IQI dimensions such that outline of at least three
sides of the IQI image shall be visible in the radiograph.
7.5.2 For segments of a flat or curved (i.e., ellipsoidal, tori spherical, tori conical, elliptical,
etc.) component where the source is placed perpendicular to the center of a length
of weld for a single exposure when using more than three film holders, at least three
IQI’s, one placed at each end of the radiograph span and one in the approximate
center of the span, are required.
7.5.3 Multiple IQI’s on a single radiograph are required under the following condition:
7.5.4 If the density of the radiograph anywhere through the area of interest varies by
more than - 15% or + 30% from the density through the body of the hole type IQI or
adjacent to the designated wire of a wire type IQI, then an additional pentameter
shall be used for each exceptional area (s). When two (2) IQI’s are used, one shall
represent the lightest area of interest and the other, the darkest area of interest; the
intervening densities on the radiograph shall be considered as having acceptable
density.
8.1.2
Flat components or longitudinal joints in cylindrical or conical components.
Curved or spherical components whose concave side is toward the source and when
source-to-material distance is less than the inside radius of the component. (Curved or
spherical components whose convex side is toward the source.
8.1.6 Mapping
When inaccessibility or other limitations prevent the placement of markers as stipulated
in clause 7.1 and 7.2 above, a dimensional map of the geometric arrangement including
marker locations, shall accompany the radiographs to show that full coverage has been
obtained. It shall be provided with radiograph review form as well.
9. Screens and film holders
Intensifying screens consisting of lead foil (0.1mm thick) shall be used as both front and back
of the film in direct contact with film. Screens shall be free of dirt, oxide, coatings, wrinkles,
pitting and other defects. Fluorescent screens are not under the scope of this work
instruction.
Film holders / Cassettes are of PVC double envelope type. Cassette containing film shall be
in intimate contact to the object being radiographed.
12.1 Location markers which are to appear as radiographic images on the radiograph shall be placed on
the part and not on the film holder / cassette and their location shall be marked permanently
(where accessible) on the surface of the part being radiographed.
12.2 Each radiograph shall be permanently identified providing the following information:
12.7 Any weld / weld segment that has been repaired shall be marked as R1, R2 etc. for 1st and 2nd
repair etc. respectively on the radiograph.
d= Distance from source side of weld or object being radiographed to the film (mm).
D and d shall be measured to the approximate center of the area of interest.
14.1.1 Either a densitometer or a step wedge comparison film shall be used for reading film density.
The density of the step wedge comparison film and densitometer calibration shall be verified
by comparison with a calibrated step wedge film traceable to National Standards step tablet
having at least 5 steps with neutral densities from 1.0 through 4.0. In case of difference of
reading between step wedge comparison film and densitometer affecting acceptability of the
radiograph, then the densitometer reading shall govern. A tolerance of + 0.05 in density is
allowed for variations between densitometer readings.
14.1.2 The transmitted film density through the radiographic image of the body of the appropriate
hole type IQI or adjacent to the designated wire of the wire IQI and the area of interest shall
be 1.8 minimum for single film viewing for radiographs made with gamma ray source The
maximum allowable density shall be 4 for gamma ray source. A tolerance of 0.05 in density is
allowed for variations between densitometer readings.
14.1.3 Densitometer shall be calibrated against national standard step tablet or step wedge
Calibration film traceable to national standard step tablet, at a density closest 1.0, 2.0, 3.0 &
4.0, Every 90 days. The record of such calibration will be documented.
14.1.4 The periodic calibration verification checks shall be performed at least the beginning of each
shift, after 8 hours of continuous use or after change operation, which comes first.
14.2 Density Variation
14.2.1 The transmitted film density of the radiograph anywhere through the area of interest shall
not vary by more than -15% or + 30% from the density through the body of the hole type IQI
or adjacent to the designated wire of a wire type IQI representing that area of interest.
14.2.2 When shims are used, the + 30% density restriction may be exceeded, provided the required
IQI sensitivity is displayed, and the density limitation of 13.1.2 are not exceeded.
14.2.3 When calculating the allowable variation in density, the calculation may be rounded off to
nearest 0.1.
B) For Double-wall viewing, only source side IQI shall be used. Care should be exercised to
ensure that the required geometric un sharpness is not exceeded. If the geometric un-
sharpness requirement is not met; then single wall should be used.
C) In case of extreme difficulty to employ double wall double image technique the weld
may be radiographed with the radiation beam positioned in such a way that the images
of both walls are superimposed as per shooting sketch attached. In this case when
complete coverage is required a minimum of three (3) exposures is taken at either 60 o
or 120 o to each other for each joint.
16.4 Developing shall be done for a period between 5 and 8 minutes at a temperature of 20 degrees C.
At higher temperatures (not exceeding 24 degrees C) the developing time may be curtailed as
determined by visually observing the extent of film development during the process. For lower
temperatures (not to go below 18 degrees C), the development time may be required to be
increased as appropriate.
16.5 Fixing time shall normally be at least twice the developing time (not to exceed 15 minutes in fresh
fixer).
16.6 Washing after developing and fixing shall be done in running water as appropriate.
16.7 After washing, the films shall be put to dry. A film drying cabinet incorporating circulating heated air
may be used to assist in drying.
17. Viewing Radiographs
17.1 Illuminator used for viewing & interpretation of radiographs shall have sufficient illumination
output and a rheostat control to provide variable light intensity to suitably evaluate the
radiographic sensitivity with respect to the designated hole or wire of the hole type or wire type IQI
respectively.
17.2 Viewing room may not be completely dark but can have subdued background light of an intensity
that will not create troublesome reflections, shadows, or glare on the radiograph to the evaluator.
17.3 The light from around the outer edge of the radiograph or coming through low density portion of
the radiograph shall not interfere with interpretation.
18.1 Artifacts:
Radiographs may have some mechanical, chemical or other blemishes but these shall not mask
or be confused with the image of any discontinuity in the area of interest of the object being
radiographed. Such blemishes include but not limited to the following:
18.1.1 Fogging
18.1.2 Mechanical defects such as scratches, finger marks, crimps, dirtiness, static marks,
smudges, or tears etc.
18.1.3 Processing defects such as streaks, water marks, frilling, puckering, chemical stains.
18.2
If the white image of the lead symbol `B' attached back of the cassette, appears on a darker
background of the radiograph, protection from back scatter is insufficient and the radiograph shall
be considered unacceptable. A dark image of the lead symbol `B' on a lighter background is
accepted.
18.3
TEL shall be responsible for the review, interpretation, evaluation, and acceptance of the
completed radiographs to assure compliance with this procedure. The radiographic technique
details and the radiograph review form documentation shall accompany the radiographs.
Acceptance shall be completed prior to presentation of the radiographs and accompanying
documentation to the Inspector.
b) the aggregate length of indications of IP in any continuous 12-in. (300-mm) length of weld ex
ceeds 1 in. (25 mm);
c) the aggregate length of indications of IP exceeds 8 % of the weld length in any weld less than
12 in. (300 mm) in length
b) the aggregate length of indications of IPD in any continuous 12-in. (300-mm) length of weld e
xceeds 3 in. (75 mm).
b) the aggregate length of indications of IF in any continuous 12-in. (300-mm) length of weld ex
ceeds 1 in. (25 mm);
c) the aggregate length of indications of IF exceeds 8 % of the weld length in any weld less than
12 in. (300 mm) in length.
b) parallel ESI indications separated by approximately the width of the root bead (wagon tracks) s
hall be considered a single indication unless the width of either of them exceeds 1/32 in. (0.8 mm).
In that event, they shall be considered separate indications;
c) the aggregate length of ESI indications in any continuous 12-in. (300-mm) length of weld exceed
s 2 in. (50 mm);
e) the aggregate length of ISI indications in any continuous 12-in. (300-mm) length of weld exceed
s 1/2 in. (13 mm);
g) more than four ISI indications with the maximum width of 1/8 in. (3 mm) are present in any con
tinuous 12-in. (300 mm) length of weld;
h) the aggregate length of ESI and ISI indications exceeds 8 % of the weld length.
a) the length of an ESI indication exceeds three times the thinner of the specified wall thicknesses j
oined;
b) parallel ESI indications separated by approximately the width of the root bead (wagon tracks) s
hall be considered a single indication unless the width of either of them exceeds 1/32 in. (0.8 mm).
In that event, they shall be considered separate indications;
d) the aggregate length of ISI indications exceeds two times the thinner of the specified wall thickn
esses joined and the width exceeds one-half the thinner of the specified wall thicknesses joined;
e) the aggregate length of ESI and ISI indications exceeds 8 % of the weld length.
19.5 Porosity
The maximum dimension of the indication shall apply to the criteria given in 18.5.1 or 18.5.2.
a) the size of an individual pore exceeds 1/8 in. (3 mm) in diameter or 25 % of the specified wall th
ickness, whichever is less;
b) the size of an individual pore exceeds 25 % of the thinner of the two specified wall thicknesses j
oined, but no more than 1/8 in. (3 mm) in diameter;
19.5.3 For cluster porosity see Fig 19 and Fig 20 of API 1104
19.6 Cracks
Cracks shall be considered a defect should any of the following conditions exists:
a) the crack, of any size or location in the weld, is not a crater crack or star crack;
b) the crack is a crater crack or star crack with a length that exceeds 5/32 in. (4 mm).
19.7 Undercutting
Undercutting shall be considered a defect when either of the following conditions exists:
a) the aggregate length of indications of EU and IU, in any combination, in any continuous 12-in. (3
00-mm) length of weld exceeds 2 in. (50 mm);
b) the aggregate length of indications of EU and IU, in any combination, exceeds one-sixth of the
weld length.
19.8 Accumulation of Imperfections
Excluding inadequate penetration due to high-low and undercutting, any accumulation of imperfe
ctions shall be considered a defect should any of the following conditions exist:
a) the aggregate length of indications in any continuous 12-in. (300-mm) length of weld exceeds 2
in. (50 mm);
b) the aggregate length of indications exceeds 8 % of the weld length.
Examination results shall be reported in Radiography Evaluation form, which shall contain following
information as a minimum.
1) Listing of each radiograph location.
2) Procedure and rev. no.
3) Evaluation & disposition of the welds examined.
4) Identification of the MGVL personnel performing final acceptance.
5) Date of evaluation
22. RECORDS
The result of radiographic examination shall be recorded in Radiograph Examination Report along
with technique sheet.
23. Appendix-1
24. Appendix-2
25. Appendix-3
26. Appendix-4
LOCATION MARKER SKETCHES
27. Appendix-5
28. APPENDIX 6 (RT TECHNIQUE SHEET)
Equipment No/Job No: W.O. No. Client Stage of Exam
Before/After PWHT
X-Ray/Gamma Ray
Radioisotope: - SKETCH
KVp : FOD : mm
SOFD : mm
SOD : mm
Front: mm Back: mm
Shim thk: mm
TECHNIQUE SHEET
ACCEPTANCE STD
NO.
PROCEDURE
DRG. NO. & REV.
&REV.NO.
Accepte
Repair /
Inclusio
surface
Lack of
Others
Fusion
No.
Reject
Check
Crack
Slag
d
SPOT
31. REFERENCES:
API 1104, Ed.2021
ASME Sec-V, Ed.2019
SNT-TC-1A, Ed.2016
Written practice – SUNUP/WP/001 Rev.01
Personnel qualified to Level II/III, in this method and shall be as per Written Practice
SUNUP/WP/001 Rev.01
34. EQUIPMENT:
Magnetization shall be performed using, 220 Volt AC Electromagnetic yokes (Parker and Equivalent
Make) having minimum lifting power of 10 lbs. (4.55 Kg.) at maximum pole spacing of 6.5"(165
mm). This will produce longitudinal magnetic field in examination metal. The magnetizing power of
the yoke shall be verified each day prior to use and also in case if it has been repaired using weight
at max pole spacing. The weight shall be weighed using scale from reputed manufacturer and
stenciled with applicable nominal weight prior to use and need to be verified only in case of
damage, which could lead to potential loss of material.
a) Dry Particles:
The magnetizing current shall remain “on” while the examination medium is being applied.
Accumulations or excess dry particles in examinations shall be removed with a light air stream from
a bulb or syringe or other source of low-pressure dry air. The examination current or power shall be
maintained while removing the excess particles. The Dry magnetic particle examination shall be
carried out below 482 ºF (250ºC).
b) Wet Particles:
The magnetizing current shall be turned on after the particles have been applied. Flow of particles
shall stop with the application of current. Wet particles may be applied before and/or after
magnetizing current is applied. Wet particles may be applied during the application of magnetizing
current if they are not applied directly to the examination area and are allowed to flow over the
examination area or are applied directly to the examination area with low velocities insufficient to
remove accumulated particles. The Wet magnetic particle examination shall be carried out below
122 ºF (50ºC) If an indication is observed, area shall be cleaned and reexamined for conformance.
During reexamination, the yoke legs shall be placed perpendicular to the major direction of
indication (this will give maximum sensitivity and we can determine extent of discontinuity
effectively.)
A minimum light intensity of 100fc (1000 lux) shall be maintained during examination/
evaluation. This can be obtain using 100 W lamp kept at 10" (250 mm) max. from surface. The light
source, technique used, and light level verification is required to be demonstrated one time,
documented and maintained on file. The span covered under light is 16” (400 mm). Light meter
shall be calibrated at least once a year or whenever meter has been repaired. If meter have not
been in use for one year or more calibration shall be done before being used.
38. EVALUATION:
9.1 All indications shall be evaluated in terms of the acceptance standards of the referencing code
section.
9.2 Discontinuities on or near the surface are indicated by retention of the examination medium.
However, localized surface irregularities due to machining or grinding marks or other surface
conditions may produce false indication.
9.3 Broad areas of particle accumulation, which might mask indications form discontinuities, are
prohibited and such area shall be cleaned and reexamined.
A linear indication is one having a length greater than three Times of its width.
A round indication is one of circular or elliptical shape with a length equal to or less than
three times of its width.
a) Any indication with a maximum dimension of 1/16 in. (1.6 mm) or less shall be classified as nonrelevant.
Any larger indication believed to be nonrelevant shall be regarded as relevant until reexamined by
magnetic particle or another NDT method to determine whether or not an actual imperfection exists. The
surface may be ground or otherwise conditioned before reexamination. After an indication is determined
to be nonrelevant, other nonrelevant indications of the same type need not be reexamined.
i) indications evaluated as crater cracks or star cracks exceed 5/32 in. (4 mm) in length;
ii) indications evaluated as cracks other than crater cracks or star cracks;
iii) linear indications evaluated as IF and exceed 1 in. (25 mm) in total length in a continuous
Individual or scattered porosity shall be considered a defect should any of the following conditions exist:
a) the size of an individual pore exceeds 1/8 in. (3 mm) in diameter or 25 % of the specified wall thickness,
whichever is less;
b) the size of an individual pore exceeds 25 % of the thinner of the specified wall thicknesses joined, but no
more than 1/8 in. (3 mm) in diameter;
c) the distribution of scattered porosity exceeds the concentration permitted by Figure 19 or Figure 20 of
API 1104
CP that occurs in any pass except the finish pass shall be considered a defect should any of the following
conditions exist:
b) the aggregate length of CP in any continuous 12 in. (300 mm) length of weld exceeds 1/2 in. (13 mm).
40. repairs:
The unacceptable defects shall be removed and repair when there is a defect thought to be
removed, and prior to making weld repairs, the area shall be examined by suitable method to
ensure it has been removed or reduced to an acceptably Sized imperfection.
43. DEMAGNITIZATION:
Demagnetization is required only if specified in purchase order, drawings or specification. This shall
be performed with AC yoke, by placing the poles on surface, moving them around the area and
slowly withdrawing the yoke while it is still energized. After demagnetization residual magnetic
field should not exceed 3 Gauss anywhere in the piece absolute value, unless otherwise agreed
upon with purchaser.
44. RECORDS:
The results of magnetic particle examination shall be recorded on the magnetic particle Examination
report as per attach format.
PROJECT NAME: REPAIR OF SOKU PWD DISPOSAL LOOP LINE TO ACE LINE
WORKS
CLIENT LOCATION REPORT No.
220 VOLTS
EQUIPMENT MAGNETIC
Procedure & XX/XX/
YOKE DATE
Rev. No. XX
Examination
Wet
Medium
CAPABLE
SURFACE CALIBRATIO OF
CLEANED
CONDITION N LIFTING
Equipment Tag 4.5Kg
Visible
Type of Particle Technique Spray
Fluorescent
METHOD OF
DRAWING N0 MAGNETIZATIO Continuous Thickness
N
ASME SEC
Acceptance VIII DIV 1 Lighting Type of
Day light
Standard APPENDIX Equipment current
6
INSPECTION
BATCH N0
MEDIUM
BACKGROUND
BATCH N0
CONTRAST
N= Non-Rejectable Indication R= Rejectable Indication
LOCATIO
WEL TYPE OF N
WELDE INDICATIO EXTENT OF REMARK
D INDICATION N R (Length,
R NO N INDICAIONS S
NO. S Dia or
Aligned)
APPROVALS
GLOBAL CLIENT
NAME NAME
SIGN SIGN
DATE DATE