Professional Documents
Culture Documents
XRF - IITR (Autosaved) (Autosaved)
XRF - IITR (Autosaved) (Autosaved)
XRF - IITR (Autosaved) (Autosaved)
Introduction of X-ray
Fluorescence Analysis
Mr. Ravindra Saini
Institute Instrumentation Centre, IIT Roorkee
Outline
2
INDIAN INSTITUTE OF TECHNOLOGY ROORKEE
Fundamentals of
X-ray Fluorescence Spectrometry
Electromagnetic Spectrum
4
Characteristics of X-rays
5
How Do X-Ray Interact with Matter?
6
Interaction of X-rays with Matter
X-rays Fluorescent
X-rays
Compton scatter
X-rays
Photo-electron
Sample
7
X-ray Fluorescence: how does it work?
8
X-ray Fluorescence: how does it work?
• Fluorescence involves emission of an X-ray photon after ionization of the atom by a primary beam X-ray
9
Source - Sample Interaction
10
Source - Sample Interaction
11
Transitions between energy levels :
12
Principle of X-ray Fluorescence cont.
13
Principle of X-ray Fluorescence
Fluorescent X-rays
M shell Kα 4
X-ray tube Primary X-rays 3
1
Photo
Fluorescent X-rays K shell electrons
Primary 2
X-rays Sample Kβ L shell
Lα
14
What We Can See by XRF
Qualitative analysis
Fe-Kα Cr-Kα Check energy/wavelength of peaks.
Ni-Kα
Primary X-rays
Fluorescent To see what elements are contained.
X-rays
Cr Sample contains Ni,Fe and Cr
Fe Ni
Fe-Kα
X-ray intensity
a specific element is
Cr-Kα
contained.
Ni-Kα
Compare to
reference material
Wavelength
15
Basis of the method of x-ray fluorescence
• X-rays, known as primary (produced by an x-ray tube) excite all the elements present in a specimen.
• Each of these elements under the impact of primary x-rays, emits its own wavelength.
• These wavelengths are then separated with a crystal monochromator and collected by a detector that converts
photons into electrical pulses.
• The output of the detector is processed electronically and then sent to a pulse counter.
• The number of pulses produced per second (cps) is in first approximation proportional to the concentration of
the element characteristic.
• These intensities (cps) are compared with standard intensities, then a calculation allow to determine the
concentration of the element in question.
• In modern equipment, matrix effects correction (inter element effects) are applied before the results are
delivered by the system.
• Importantly, the method itself is not destructive.
16
How do X-Ray interact with Matter?
17
X-Ray Emission Energies
• Fluorescence yield
18
X-Ray Emission Energies
19
WD-XRF and ED-XRF
Crystal
X-ray tube (2d sinθ = n λ) X-ray tube Detector
with high energy
resolution
(Si-PIN, SDD, SSD)
20
X-Ray Fluorescence: Two ways of doing it - EDX and WDX
21
Comparison of Light and Fluorescent X-rays
22
EDXRF vs WDXRF
23
Comparison between WDX and EDX
24
TUBE-ABOVE SEQUENTIAL WAVELENGTH DISPERSIVE
X-RAY FLUORESCENCE SPECTROMETER
Elemental coverage: Be through U
Optics: Wavelength dispersive, sequential, tube
above
X-ray tube: End window, Rh-anode, 4 kW, 60kV
25
XRF capability
26
WDXRF Spectrometer
(8) Scintillation
counter (SC)
(1) X-ray tube
(9) Proportional
counter (PC)
27
How X-rays are Generated?
28
Production of X-RAY
29
The continuous spectrum
• The continuous spectrum comes from the sudden deceleration of electrons on the anode.
• The position of maximum intensity of the spectrum depends on the voltage (kV) applied between the filament
(electron gun) and the anode.
• The higher the voltage increases, the minimum wavelength for excitation (λ0) decreases.
• It is observed that the position of characteristic lines, which are superposed on the spectrum, do not change
when you change the voltage.
30
Output of the X-ray tube
31
End window X-ray tube
32
X-ray tube
Excitation energy
Tube voltage (kV)
X-ray intensity
Increase/decrease of sensitivity
Target material Interference by scattering of tube
target lines
33
Primary beam filter
Function
◆ Eliminates the scattering peaks derived from the x-ray tube (Thomson
and Compton scattering)
◆ Reduces background to improve peak/background ratios
Note) Peak intensities are also reduced.
34
Diaphragm
Function
◆ Confines detected area from the sample, such as 30 mm in diameter so
that sample area only within the sample mask opening is measured.
Note) The primary x-rays are not confined.
Diaphragm
35
Attenuator
Function
◆ Reduces x-ray intensity to avoid counting loss without change of tube
voltage (kV) or current (mA).
When measured x-ray intensity is higher than the linear counting limit for the
detector (e.g. 1000kcps for SC, 2000kcps for F-PC), the intensity can be reduced
with the attenuator and without changing kV or mA and it ensures stable analysis.
Without attenuator,
•Decrease mA
•Insert a primary x-ray beam filter
•Change to a finer slit
Changing kV and mA can cause instable analysis owing to the power change.
There are many cases that It is not possible to adjust intensity by the other
conditions.
36
Soller Slit
Function
◆ Turns the x-rays from the sample into parallel beams.
37
Analyzing Crystal
Function
◆ Guides the x-rays of interest into the detector by diffraction.
Monochromatic X-rays
38
Analyzing Crystal
Atomic N. 4 5 6 7 8 9 11 12 13 14 15 16 17 19 20 22 23 24 25 26 27 28 29 30 33 - 60
K Line Be B C N O F Na Mg Al Si P S Cl K Ca Ti V Cr Mn Fe Co Ni Cu Zn As - Nd
L Line 48Cd 56Ba 74W 82Pb
LiF(200)
LiF(220)
LiF(420)
PET
Ge
RX25
RX35
RX40
RX45
RX61
RX61F
RX75
RX85
RX4
RX9
Excellent Good
39
Detector
Function
◆ Detects the x-rays entering the detector.
40
Scintillation Counter (SC)
X-rays
To the counting circuit
Electrons
41
Proportional Counter (PC)
Ar → Ar+ + e-
42
PHA (Pulse Height Analyzer)
The Pulse Height Analyzer discriminates the x-ray pulses by the pulse heights to
remove noise of low level and the pulses of higher order lines in order to pick
up the pulses of first order line x-rays.
2nd
- Fe-Kβ1-4th
order
Upper Level
- Si-Kα
Pulse height
1st Window
X-ray intensity
order
Lower Level
- Si-Kα
Time
With PHA
Pulse height
2θ
Time
43
PHA (Pulse Height Analyzer)
n=1 n=1
Detector PHA
Pb-Lβ1
window
Sn-Kα
44
INDIAN INSTITUTE OF TECHNOLOGY ROORKEE
Sample Preparation
Points in Sample Preparation
46
Metal: Grinding/Polishing
Grinder
47
Powder: Grinding and Pressing
Sample
48
Grain-size effect
± 0.45 mass%
10 20 30
Al2O3 mass%
± 1.50 mass%
10 20 30
Al2O3 mass%
49
Powder: Fusion
50
Powder: Loose Powder
• Powder samples are packed into liquid cells with sample film.
• Samples can be measured even without being pressed.
• The repeatability is not as good as that of other sample
preparation methods for powder samples.
51
INDIAN INSTITUTE OF TECHNOLOGY ROORKEE
Applications
Industries where XRF is used
• Production control
• Cement plants
• Foundries (steels, brass, aluminum,…)
• Industries of glass and refractory
• Chemistry
• Petrochemicals
• Polymers and plastics
• Pharmacy and food
• Mines and Mineralogy
• Environment
• Police and customs
53
Strong Points of XRF
• Quality control
• Process control
• Research and development
• Screening
54
Electron Probe Micro Analysis
55
WDS vs EDS
56
EPMA typical results
57
Advantages of X-ray Over Electron Excitation in Microanalysis
58
Advantages of X-ray Over Electron Excitation in Microanalysis
• •Unlike those used in electron excited EDS, XRF samples need not be electrically
conductive. There is no electrical charge being transported in XRF samples.
• Measurement of liquid and organic samples is possible because the electron beam
does not require vacuum. Samples may remain in an air environment during
analysis.
• X-ray excitation does not heat a sample as there is no absorbed kinetic energy of
electrons. This means that temperature sensitive samples, such as thin films, can
be analyzed.
59
INDIAN INSTITUTE OF TECHNOLOGY ROORKEE
Matrix Effect
Matrix Correction
61
Matrix Corrections- Peak Above Background
62
Matrix Correction-Z
63
Matrix Correction-A
64
Matrix Correction- X-ray Absorption effect (A) Cont.
65
Matrix Corrections-F
66
INDIAN INSTITUTE OF TECHNOLOGY ROORKEE
Thank You