Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Biosensors and Bioelectronics 77 (2016) 1078–1085

Contents lists available at ScienceDirect

Biosensors and Bioelectronics


journal homepage: www.elsevier.com/locate/bios

A double signal electrochemical human immunoglobulin G


immunosensor based on gold nanoparticles-polydopamine
functionalized reduced graphene oxide as a sensor platform and
AgNPs/carbon nanocomposite as signal probe and catalytic substrate
Si Zhang, Na Huang, Qiujun Lu, Meiling Liu n, Haitao Li, Youyu Zhang, Shouzhuo Yao
Key Laboratory of Chemical Biology & Traditional Chinese Medicine Research (Ministry of Education, China), College of Chemistry and Chemical Engineering,
Hunan Normal University, Changsha 410081, PR China

art ic l e i nf o a b s t r a c t

Article history: In this paper, a double signal electrochemical Human immunoglobulin G (HIgG) immunosensor based on
Received 5 August 2015 AgNPs/carbon nanocomposite (Ag/C NC) as the signal probe and catalytic substrate was developed for
Received in revised form fast and sensitive detection of HIgG. The as-prepared AuNPs-PDA-rGO nanocomposite and Ag/C NC were
29 October 2015
confirmed by UV–vis, Fourier transform infrared spectroscopy, scanning electron microscopy and
Accepted 30 October 2015
transmission electron microscopy. Electrochemical impedance spectroscopy, cyclic voltammetry and
Available online 31 October 2015
differential pulse voltammetry were used to investigate the electrochemical properties of the proposed
Keywords: immunosensor. The AuNPs-PDA-rGO nanocomposite can improve the electron transfer rate and capture
Double signal electrochemical im- more Ab1. In the sandwich-type immunoassay process, the Ag/C NC functionalized bioconjugates were
munosensor
captured on HIgG/Ab1/AuNPs-PDA-rGO surface and the electrochemical double-signal strategy was
AuNPs-PDA-rGO
employed. These double electrochemical detection signals were directly monitored the oxidation current
Ag/C NC
Sandwich-type immunoassay originated from Ag/C NC and indirectly detected the reduction current of benzoquinone which was
produced from the reaction of H2O2 and HQ by catalysis of Ag/C NC in electrochemical detection of HIgG.
Under the optimized conditions, the current responses were changed with the concentrations of HIgG for
the proposed immunosensor with wide linear ranges of 0.1 to 100 ng mL  1 and 0.01–100 ng mL  1 with
the lowest detection concentration of 0.001 ng mL  1 in the absence and presence of H2O2 and HQ. The
double-signal strategy is used for detection of HIgG, and the results came from the two signals were well
consistent with each other. The proposed immunosensor was successfully applied in analysis of human
IgG in real samples and this strategy may provide a relative simple and effective method for construction
of other immunsensors in detection of other biomarkers in clinical medicine.
& 2015 Elsevier B.V. All rights reserved.

1. Introduction clinical diagnose. Immunoassay, which based on the highly spe-


cific recognition of antigens by antibodies, has attracted much
Human immunoglobulin G (HIgG) is a class of important attention in recent years owing to its high selectivity and low
component mainly existing in human serum, which plays a vital detection limit. Enzyme-linked immunoassays (Zhao et al., 2007),
role in recognizing some diseases (Campanella et al., 2008; Liu electrochemiluminescent immunoassays (Chen et al., 2014; Xu
et al., 2015). The HIgG concentration in normal adults ranges from et al., 2011) and fluorescent immunoassays (Ahn et al., 2015; Ii-
9.4 to 15.1 g/L (Li et al., 2014). High immunoglobulin levels will zumi et al., 2013) as the conventional immunoassays have been
cause rheumatoid arthritis, liver diseases and infectious diseases. commonly used in clinical diagnosis. Although some significant
However, low immunoglobulin levels will also cause humoral researches have been reported, they are often subject to expensive
immunodeficiencies and metabolic diseases (Li et al., 2014). instruments, complicated operations and long assay time. Thus,
Therefore, the development of ultrasensitive and reliable analy- the development of new methods and strategies to enhance the
tical methods for detecting HIgG in human serum is necessary in sensitivity and simplicity of clinical immunoassay is still of great
necessity.
n
Corresponding author. Fax: þ86 731 88872531. Electrochemical immunosensors based on highly specific re-
E-mail address: liumeilingww@126.com (M. Liu). cognition between the antigen and antibody, can give

http://dx.doi.org/10.1016/j.bios.2015.10.089
0956-5663/& 2015 Elsevier B.V. All rights reserved.
S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085 1079

electrochemical signals by the transducer. Electrochemical im- and low detection limit and it may provide a new mode for pro-
munoassay has been widely studied because of its intrinsic ad- posing other sandwich-type immunosensor for detection of other
vantages, such as low cost, high sensitivity and simple operation biomarkers.
processes compared with other conventional immunoassay tech-
niques. Especially, the electrochemical sandwich-type im-
munosensors have been received striking attention in recent years 2. Experimental section
due to their improved sensitivity by different signal amplified
strategies (Yang et al., 2015; Du et al., 2010). However, these im- 2.1. Reagents and apparatus
munosensors must rely on the use of electroactive substance as
the second antibody's marker (Zhu et al., 2013). Therefore, the Human IgG, rabbit anti-human IgG and goat anti-human IgG
preparation of the bioconjugates must be carefully controlled and were purchased from Bioss (Beijing, China). Dopamine hydro-
the electroactive substance may easily leak out from electrode chloride (DA), chloroauric acid (HAuCl4), bovine serum albumin
surface to the solution, which may result in unstable output sig- (BSA), silver nitrate (AgNO3) and ethylene glycol (C2H6O2) were
nals and unreliable experimental results. Thus, it is necessary to purchased from Sinopharm Chemical Reagent Co., Ltd (Shanghai,
develop new materials to construct novel immunosensors for China). Graphene oxide (GO) was purchased from XFNANO Co.,
sensitive and fast detection of antigen. Ltd. (Nanjing, China) and used without further purification. All
Nanomaterial-based electrochemical immunosensors have other chemicals were of analytical grade and used as received.
been widely studied in recent years owing to their high selectivity Tris–HCl buffer solution (10 mM, pH 8.5) was used for PDA pre-
and sensitivity. Among the various types of nanomaterials, AgNPs paration and functionalization. Phosphate buffer solution (PBS,
have gained significant attention because of their catalytic activity 0.1 M) was prepared with Na2HPO4, NaH2PO4 and 0.1 M Na2SO4.
and good electrical conductivity (Mazloum-Ardakani et al., 2015). All aqueous solutions were prepared with ultrapure water.
However, the relatively poor biocompatibility and stability of The UV–vis spectra (UV) were obtained on a UV-2450 spec-
AgNPs limits their use in immunoassay (Sun et al., 2015). There- trophotometer (Shimadzu, Japan). FT-IR spectra (IR) were collected
fore, the development of novel AgNPs-based nanomaterials with on a Nexus-670 Fourier transform-infrared spectrophotometer
high catalytic activity, good stability and biocompatibility is ne- (Nicolet Instrument Co, USA) with the KBr pressed pellet trans-
cessary. To the best of our knowledge, carbon dots (C-dots), which mission mode. The morphology and composition of nanocompo-
is one of the new members of carbon nanomaterials family, have site were characterized by scanning electron microscopy (SEM)
attracted tremendous attention in nanotechnology field recently and energy dispersive X-ray spectroscopy (EDS) (Zeiss Sigma,
(Zhao et al., 2013; Liu et al., 2012). C-dots show excellent bio- Germany) and the size of the nanoparticles was characterized by
compatibility and low toxicity in biomedical applications. Recently, transmission electron microscopy (TEM) (JEOL, Japan). Electro-
our group has successfully synthesized C-dots by one-pot elec- chemical experiments, including cyclic voltammetry (CV), elec-
trochemical carbonization of ethylene glycol (Lu et al., 2015). The trochemical impedance spectra (EIS), differential pulse voltam-
as synthesized C-dots do not need any further surface functiona- metry (DPV) were carried out with a research electrochemical
lization and possess excellent stability and reducibility. It can act workstation of SP-150 (BioLogic Science Instruments, France)
as the reductant for direct synthesis of AgNPs to obtain silver/ controlled by EC-Lab software. A three-electrode system was used
carbon nanocomposite (Ag/C NC). This method is fast and simple which contained a glassy carbon electrode as working electrode, a
without using any additional reductant and stabilizing agent saturated calomel reference electrode (SCE) and a carbon rod
compared with other conventional synthesis method of AgNPs. electrode as counter electrode. All potentials here were cited
More importantly, the as synthesized Ag/C NC shows excellent versus SCE. All experiments were performed at room temperature.
peroxidase-like activity (Lu et al., 2015) and can give electro-
chemical signal (Gao et al., 2015). Furthermore, it may also afford 2.2. Synthesis of AuNPs-PDA-rGO nanocomposite
abundant active sites for the attachment of biomolecules owing to
the presence of AgNPs. To date, Ag/C NC-based immunoassay that PDA-rGO was synthesized by previous method (Hu et al., 2014).
simultaneously using its peroxidase-like property and employing Firstly, 10 mg of GO and 10 mg of dopamine hydrochloride were
it as a signal label has not been reported. added into 20 mL of 10 mM Tris–HCl solution (pH¼8.5) and dis-
Herein, in this work, a simple and high selectivity im- persed by sonication for 30 min in ice bath. Then, the reaction
munosensor for determination of human IgG was successfully mixture was stirred vigorously at 60 °C for 20 h. After cooling to
constructed. Based on the sandwich-type immunoassay, rabbit room temperature, 600 μL of 2% HAuCl4 was added into the above
anti-human IgG (Ab1) was immobilized onto AuNPs-PDA-rGO mixture under vigorous stirring. 1 mL of 60 mM DA was added
modified GCE, and goat anti-human IgG (Ab2) was integrated with dropwisely to the above solution, and the reaction mixture was
Ag/C NC as a signal label through AgNPs or carboxyl group of stirred 4 h to allow complete reaction and form AuNPs-PDA-rGO
C-dots. Compared with the previously reported approaches, our nanocomposite, followed by centrifugation and washed with wa-
protocol presents several advantages: firstly, more Ab1 can be ter. Finally, the obtained AuNPs-PDA-rGO nanocomposite was re-
immobilized on the electrode surface by using AuNPs-PDA-rGO as dispersed in pH 8.5 Tris–HCl buffer solution before further use.
the immobilizing platform which has many unique properties,
including good biocompatibility, large surface area and excellent 2.3. Synthesis of Ag/C NC/Ab2 composite
conductivity. Secondly, the Ag/C NC/Ab2 biocomposite can result in
double electrochemical signals, one is directly originated from C-dots were electrochemically synthesized based on our pre-
oxidation of Ag/C NC, and the other is indirectly coming from re- vious work (Lu et al., 2015). Briefly, ethylene glycol and sodium
duction of benzoquinone which was produced from the reaction of hydroxide solution were used as the precursor and electrolyte,
H2O2 and HQ by catalysis of Ag/C NC. And both of the currents platinum sheets employed as the positive and negative electrodes.
response changed with the concentration of HIgG. Thirdly, the The C-dots were synthesized by electrolyzing in the electrolyte
proposed electrochemical immunosensor based on the double solution under 30 V for 40 min. After electrolysis, the electrolyte
signal strategy is reliable and the detection results can be sup- solution was adjusted to neutral with hydrochloric acid and then
ported by each other. This double-signal electrochemical im- dialyzed (Mw 1000) against water for two days, followed by
munosensor exhibited excellent sensitivity, wide detection range centrifuging at 12000 rpm for 10 min to remove the precipitate
1080 S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085

particles. 2.5. Electrochemical measurements


The Ag/C NC was synthesized according to our previous work
(Lu et al., 2015). A certain amount of AgNO3 aqueous solution Electrochemical experiments were performed in a conventional
(0.1 mM) was added into 1 mL of the as-synthesized C-dots (NaOH three-electrode system. DPV was used to monitor the current re-
treated) aqueous solution at room temperature for 40 min. sponses after the sandwich reaction to detect HIgG in the absence
Ag/C NC/Ab2 was prepared typically as follows: firstly, 100 μL of or presence of HQ and H2O2. The DPV were carried out in 0.1 M-pH
100 μg mL  1 goat anti-HIgG (Ab2) was added into 1 mL of the as- 7.4 PBS at room temperature. The DPV parameters applied were:
prepared Ag/C NC solution under vigorous stirring at 4 °C for 5 h, pulse amplitude, 2.5 mV; pulse width, 100 ms; pulse period, 0.2 s;
followed by centrifugation and washing with ultrapure water. Fi- scanning potential,  0.2 to 0.5 V and 0.4 to  0.3 V.
nally, the obtained Ag/C NC/Ab2 was redispersed in ultrapure
water before further use.
3. Results and discussion

2.4. Fabrication of the immunosensor 3.1. Characterization of the materials and immunosensor

Glassy carbon electrode (GCE, 3 mm in diameter) was firstly UV spectra of C dots (a), Ag/C NC (b) and Ag/C NC/Ab2 (c) are
polished with emery paper and 50 nm alumina slurries followed displayed in Fig. 1A. The as-synthesized C-dots (a) shows a strong
by rinsing thoroughly with ultrapure water. The electrodes were absorption peak at 257 nm, which can be attributed to the π–π*
successively ultrasonicated in ethanol and water and then washed transition (Jaiswal et al., 2012). After reaction with AgNO3, besides
thoroughly with ultrapure water to obtain fresh and clear surface, the strong absorption at 257 nm, a broad peak at around 410 nm
and dried in the air. 10 μL of 0.1 mg mL  1 AuNPs-PDA-rGO solu- (b) is observed, which is the characteristic absorption peak of AgNPs
(Tan and Chen, 2012). The color of the reaction solution changed
tion was dropped onto the GCE and dried in air. Subsequently,
from pale yellow to bright yellow during the reaction. After goat anti-
10 μL of 100 μg mL  1 primary antibody (Ab1) was attached onto
HIgG conjugation with Ag/C NC, the intensity of the absorption at
the surface of AuNPs-PDA-rGO/GCE for 12 h at 4 °C. Successively,
410 nm decreased. At the same time, the absorption intensity at
the modified immunosensor was blocked with 10 μL of
257 nm increased with conjugation time. All results illustrated that
50 mg mL  1 BSA for 40 min at room temperature. After every step,
Ag/C NC was formed and the goat anti-HIgG was successfully im-
the modified electrode was thoroughly cleaned with ultrapure mobilized on Ag/C NC. The successful preparation and modification
water to remove the physically adsorbed species. Based on the were further confirmed by FTIR. The results are shown in Fig. 1B, the
sandwich-type immunoassay strategy, the immunosensor was characteristic peaks at 3408 and 1074 cm  1 are corresponding to the
incubated with target antigens (HIgG) solution for 35 min at room stretching vibration and bending vibration of the hydroxyl groups of
temperature, and then the as-prepared 10 μL of Ag/C NC/Ab2 was C-dots (Fig. 1B, a) (Lu et al., 2015). Compared with the spectrum of
dropped onto the modified electrode surface for immune reaction C-dots, the relative intensity of hydroxyl groups (–OH) slightly in-
at room temperature. The fabrication procedures of the im- creased and the carbonyl group decreased which was attributed to
munosensor are presented in Scheme 1. the reduction of C-dots during the NaOH-treated process. While

Scheme 1. The fabrication processes for the electrochemical immunosensor.


S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085 1081

0.8 a
a

Transmittance(%)
b
0.6

Absorbance 0.4 c

1571
1674
0.2

867
985
b
0.0

1126
200 300 400 500 600 700 800 4000 3500 3000 2500 2000 1500 1000 500
Wavelength(nm) Wavenumber(cm ) -1

a
1.5
b

Transmittance(%)
Absorbance

1.0

3200
b

0.5 c
c

1729
0.0 a

1613
1511
200 300 400 500 600 700 800 4000 3500 3000 2500 2000 1500 1000 500
Wavelength(nm) Wavenumber(cm )
-1

Fig. 1. The UV (A) and FTIR spectra (B) of C dots (a), Ag/C NC (b) and Ag/C NC/Ab2 (c); the UV (C) and FTIR spectra (D) of PDA (a), PDA-AuNPs (b), AuNPs-PDA-rGO (c).

Fig. 2. TEM images of C-dots (A), Ag/C NC (B) and SEM image (C) and EDS (D) of AuNPs-PDA-rGO.
1082 S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085

comparing with the C-dots, the intensity of carbonyl groups (C¼O) for C ¼O in GO obviously decreased, and the characteristic peak of
(1721 cm  1) slightly decreased after reaction with AgNO3 and this PDA was almost remained in the resultant composite. Thus, the
was because the C-dots was only partially oxidized in the AgNO3 above results indicated that the AuNPs-PDA-rGO nanocomposite
solution to produce AgNPs. After goat anti-HIgG was conjugated with was successfully prepared.
Ag/C, the new characteristic peaks at 1674 cm  1 for amide I and The morphologies and nanostructures of C-dots and Ag/C NC
1571 cm  1 for amide II absorption bands were found (Wu et al., were investigated by transmission electron microscopy (TEM). As
2007). Therefore, this indicated that goat anti-HIgG was successfully shown in Fig. 2A, the C-dots were of low multiple dispersity with
conjugated with Ag/C NC. the diameters ranging from 2 to 4 nm (Lu et al., 2015). After re-
UV spectra have been employed to monitor the preparation of action with AgNO3, the as-synthesized Ag/C NC diameters ranged
AuNPs-PDA-rGO nanocomposite. As shown in Fig. 1C, the spec- from 10 to 12 nm in diameters (Fig. 2B). The Ag/C NC shows good
trum of PDA shows two strong characteristic absorption peaks at dispersion without obvious aggregation. This may result from the
around 215 nm, 278 nm (Fig. 1C, a) and a small peak at 410 nm C-dots that acted as the stabilizing agents to attach to the Ag
(which can be seen clearly in Fig. S1A, a). After addition of HAuCl4, surface through hydrogen bonding to protect AgNPs from
there was a new absorption peak around 500 nm (Fig. 1C, b), aggregation.
which was the characteristic absorption peak of AuNPs (Hu et al., Fig. 2 shows the SEM (C) and EDS (D) of the AuNPs-PDA-rGO. It
2014). When the AuNPs-PDA-rGO nanocomposite was synthe- was found that gold nanostructures distributed on the PDA-rGO
sized, the characteristic absorption peaks at 215 nm and 278 nm nanosheets on SEM image. EDS was used to analyze the chemical
shifted to 208 nm and 280 nm (Fig. 1C, c). Meanwhile, a new ab- composition of AuNPs-PDA-rGO, and it was found that the peaks
sorption peak around 510 nm was observed, which can be seen of C, N, O and Au were obviously found as shown in Fig. 2D.
clearly in Fig. S1A, (b). Thus, all results illustrated that the AuNPs- Therefore, the above results showed that AuNPs-PDA-rGO nano-
PDA-rGO nanocomposite was synthesized successfully. composite was successfully prepared.
The as-prepared AuNPs-PDA-rGO nanocomposite can also be CVs were performed to characterize the surface features of the
characterized by FTIR. As shown in Fig. 1D (b), the characteristic electrodes in 2 mM Fe(CN)63  /4  containing 0.1 M K2SO4 solution.
peaks at 1519, 1316 and 3200 cm  1 originating from the scissoring As shown in Fig. 3A, a pair of well-defined redox peaks for
vibration of N-H, bending vibration of –CH2 and –NH group of PDA Fe(CN)63  /4  was obtained at bare GCE (Fig. 3A, a). It is obvious
(Hu et al., 2014) increased significantly after DA reaction with that the peak current was the highest on AuNPs-PDA-rGO mod-
HAuCl4. The results showed that DA was self-polymerized and ified electrode (Fig. 3A, b). This may result from the excellent
HAuCl4 was reduced. No obvious peak for C ¼O around 1729 cm  1 conductivity and larger specific surface area of AuNPs-PDA-rGO
was observed in the spectrum of PDA-rGO (Fig. S1B, b), indicating nanocomposite. After the immobilization of anti-HIgG, BSA and
that the carboxyl groups on GO were disappeared and the GO was HIgG layer by layer in Fig. 3A (c, d and e), the peak currents de-
reduced during the self-polymeriztion of DA (Kang et al., 2011). creased gradually. After incubation of Ag/C NC/Ab2 bioconjugates
Besides, as shown in Fig. 1D (c), the peak intensity at 1729 cm  1 (Fig. 3A, f), the peak current decreased obviously, this may result

60 5000
a
a
b
40 c 4000 b
c
d
d
e 3000 e
20 f
-Z'' /Ω

f
2000
0 C
1000
R
-20
0
R Z
-40
-0.2 -0.1 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0 2000 4000 6000
E /V (vs.SCE) Z' /Ω

14 0
a B
12 b
c
10 d -10
e
8 f
-20
6
4 -30 a
d
2
c
0 -40 b

-0.2 0.0 0.2 0.4 0.6 -0.4 -0.2 0.0 0.2 0.4
E /V (vs.SCE) E /V (vs.SCE)

Fig. 3. CV (A) and EIS (B) of bare GCE (a), AuNPs-PDA-rGO/GCE (b), Ab1/AuNPs-PDA-rGO/GCE (c), BSA/Ab1/AuNPs-PDA-rGO/GCE (d), HIgG/BSA/Ab1/AuNPs-PDA-rGO/GCE
(e) and Ag/C NC/Ab2/HIgG/BSA/Ab1/AuNPs-PDA-rGO/GCE (f) in 2 mM Fe(CN)63  /4  . (C) DPVs of bare GCE (a) Ag/C NC/Ab2/HIgG/GCE (b), HIgG/BSA/Ab1/AuNPs-PDA-rGO/GCE
(c) and Ag/C NC/Ab2/HIgG/BSA/Ab1/AuNPs-PDA-rGO/GCE with different concentrations of HIgG (d: 1 ng mL  1, e:10 ng mL  1, f: 100 ng mL  1) and (D) Ag/C NC/Ab2/HIgG/GCE
(a), Ag/C NC/Ab2/HIgG/BSA/Ab1/AuNPs-PDA-rGO/GCE with different concentrations of HIgG (b: 1 ng mL  1, c: 10 ng mL  1, d: 100 ng mL  1) in 0.1 M-pH 7.4 PBS (C: in absence
of HQ and H2O2, D: in presence of HQ and H2O2) at a scan rate of 50 mVs  1 and 100 mVs  1, respectively.
S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085 1083

from the weak electron transfer speed of Ag/C NC/Ab2. Therefore, benzoquinone (Q) increased with concentrations of HIgG in the
the above results indicated that each layer was modified on the same condition (Fig. 3D, b, c, d). This may ascribe to the catalytic
electrode successfully. property of Ag/C NC towards the reaction of H2O2 and HQ, the
The electron transfer properties of the modified electrodes reaction mechanism may be as follows (Ionescu et al., 2007):
after each modification process were further characterized by EIS. Ag/C
As shown in Fig. 3B, the Nyquist plots were fitted by the Randles HQ + H2 O2 → Q + H2 O (1)
equivalent circuit (Fig. 3B, inset), where Rs is the solution re-
sistance, Cd is the capacitance, Rct is the charge transfer resistance Q + H+ + e− → HQ (2)
and ZW is the Warburg impedance. The Rct of the electrode is equal
In this process, Ag/C NC catalyzed the reduction of H2O2 which
to the semicircle diameter of the Nyquist diagram. Well-defined
was transferred into H2O, and meanwhile, HQ was oxidized into
semicircles with Rct of 307 Ω and 199.6 Ω at higher frequencies
benzoquinone (Q) (1). And then Q got an electron and was reduced
were obtained for the bare GCE (Fig. 3B, a) and AuNPs-PDA-rGO/
to HQ on the electrode surface (2). With different amount detec-
GCE (Fig. 3B, b), indicating that the bare GCE had relatively larger
tion objects, the different amount of Ag/C NC conjugates will be
interface electron transfer resistance than that of AuNPs-PDA-rGO/
immobilized on the electrode, which may show different current
GCE. It suggested that AuNPs-PDA-rGO/GCE could improve the
conductivity and accelerate the electron transfer. With the fol- responses for Q (Zhao et al., 2014). Therefore, the quantitative
lowing immobilization, blocking and incubation steps, the Rct va- detection of HIgG can also be realized. Thus, this may indicated
lues increased gradually. Especially, after Ag/C NC/Ab2 was grafted that method B may also can be used to quantitatively detect of
onto the HIgG/BSA/Ab1/AuNPs-PDA-rGO/GCE, a remarkable in- HIgG.
crease in the interfacial resistance (Fig. 3B, f) was found. These
results indicated that the immunocomplex layers were modified 3.3. Optimization of experimental conditions
on the electrode surface successfully. This maybe result from that
the immunocomplex layers on the electrode can act as the elec- To get the optimal experimental results, the optimization of
tron transfer blocking layers to hinder the electron transfer. The experimental condition is necessary. The volume of PDA-rGO-
above results were agreed with the CV results. AuNPs nanocomposite will affect the binding sites for HIgG and
the current of the immunosensor. As shown in Fig. S3A, the peak
3.2. Analytical feasibility of the electrochemical immunosensor current increases with the volume of PDA-rGO-AuNPs from 4 to
10 μL and then decreases when the volume is over 10 μL. In fact,
In this study, we choose the double-signal strategy to in- the PDA-rGO-AuNPs nanocomposite can capture antibodies and
vestigate the feasibility of the electrochemical immunosensor. improve the performance of the immunosensor. But excess PDA-
Method A is directly using Ag/C NC/Ab2 composite to produce rGO-AuNPs nanocomposite may hinder interface electron transfer
electrochemical signal to detect HIgG. In this process, the experi- between nanocomposite and electrode. Therefore, 10 μL was the
ment was controlled by monitoring the oxidation currents of Ag/C optimal volume for immunosensor.
NC with different HIgG concentrations without introduction of any The incubation time for the immune-reaction is also studied as
other electroactive substance in PBS. Method B is using Ag/C NC to shown in Fig. S3B. Upon the addition of HIgG solution, the peak
catalyze the reaction of H2O2 and HQ for indirectly electrochemical current obviously increased within 35 min, and then decreased
detection of HIgG in the presence of HQ and H2O2. To prove the over 35 min. So the optimal incubation time was about 35 min for
catalysis properties of Ag/C NC, the catalysis of the reduction of next investigation.
H2O2 to oxidize HQ by Ag/C NC was conducted. The reduction of The properties of biomolecules were affected by pH, so the pH
benzoquinone (the oxidation product of HQ) in the presence of values of PBS ranging from 6.0 to 8.0 were studied. As shown in
H2O2 was studied by DPV on the Ag/C NC modified electrode. As Fig. S3C, the current value increased with pH from 6.0 to 7.4, and
shown in Fig. S2, the peak currents of benzoquinone increased then decreased in the range of 7.4 to 8.0. Therefore, pH 7.4 was the
with concentrations of HQ at the Ag/C NC modified electrode in pH optimal pH value for the immunosensor, which was also close to
7.4 PBS containing 1.25 mM H2O2. The results illustrated that Ag/C the pH of the human body.
NC can catalyze the reduction of H2O2 to oxidize HQ. In addition, the ratio of the Ab2 and Ag/C NC in preparation of
The electrochemical immunosensor was then conducted on the the Ag/C NC and the concentration of H2O2 and HQ in the detec-
basis of Ag/C NC and AuNPs-PDA-rGO. As shown in Fig. 3C, the tion process were also conducted. As shown in Fig. S3D–F, the
experiment was performed by comparing their current responses optimal ratio of the Ab2 (100 μg mL  1) and Ag/C NC was 1:10
of different modified electrodes under the same conditions. It was (V1:V2), the optimal concentration of HQ was 1 mM, the optimal
found that there was not any obvious peak for the oxidation of Ag concentration of H2O2 was 1.25 mM.
was observed on bare GCE (a), Ag/C NC/Ab2/HIgG (b) and
HIgG/BSA/Ab1/AuNPs-PDA-rGO (c) modified electrodes. The rea- 3.4. Electrochemical detection of HIgG based on the double signal
sons may be that the Ag/C NC was not stably immobilized on the strategy
electrodes or there were not form sandwich structures on those
electrodes (Li et al., 2015). On the other hand, the oxidation peak Under the optimal experimental conditions, the relationship
currents of the Ag (Fig. 3C, d, e, f) increased with the concentra- between peak current values and HIgG concentrations was studied
tions of HIgG in PBS when Ag/C NC/Ab2/HIgG/BSA/Ab1/AuNPs- with DPV using the double-signal strategy for detecting of HIgG.
PDA-rGO was used. This may indicated that this method may be The proposed immunosensor based on method A was eval-
used to quantitatively detect of HIgG. uated by measuring the current generated from the oxidation of
Secondly, the indirectly detect of HIgG is shown in Fig. 3D, and AgNPs in Ag/C NC. As shown in Fig. 4A, the oxidation peak currents
the current responses for different electrodes in the presence of of Ag increased with concentrations of HIgG in pH 7.4 PBS. This can
1 mM HQ and 1.25 mM H2O2 were investigated. The reduction be attributed the increasing amount of Ag/C NC on the platform
current of benzoquinone (the oxidation product of HQ) on the Ag/ after the sandwich reaction. It was indicated that the adsorbed
C NC/Ab2/HIgG/BSA/Ab1/AuNPs-PDA-rGO modified electrode was HIgG is directly related to the amount of the immobilized Ag/C
obviously improved in comparison with the Ag/C NC/Ab2/HIgG NC/Ab2 on electrode. Therefore, through the monitor the oxidation
(Fig. 3D, a) modified electrode. And the reduction peak currents of current of Ag/C NC, we can evaluate the amount of the detected
1084 S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085

10 7.5

7.0
8
6.5

6 6.0

5.5
4
5.0

2 4.5

4.0
-0.2 0.0 0.2 0.4 0.6 -2 -1 0 1 2 3
E /V (vs.SCE) logc/ng mL-1

0 -16

-18
-10
-20

-22
-20
-24

-26
-30
-28

-40 -30
-0.4 -0.2 0.0 0.2 0.4 -3 -2 -1 0 1 2 3
E /V (vs.SCE) logc/ng mL-1

Fig. 4. DPVs of immunosensor to different HIgG concentrations (A, C) and the calibration curves (B, D) for HIgG determination in 0.1 M PBS ( pH 7.4) in the absence (A, B) and
in the presence of (C, D) 1.0 mM HQ and 1.25 mM H2O2.

HIgG. The calibration plots showed a good linear relationship be- HIgG quantitatively in real samples.
tween the currents and the logarithmic values of the HIgG con-
centrations from 0.1 to 100 ng mL  1, as shown in Fig. 4B. The re- 3.5. Selectivity and stability of the immunosensor
gression equation was I (μA)¼5.439 þ 0.7461 log c (ng mL  1) with
the correlation coefficient of 0.9908 and a detection limit of 0.0075 Possible interfering substances were used to evaluate the se-
ng mL  1 (signal/noise ¼3). lectivity of the proposed immunosensor. As shown in Fig. S4A–B,
The proposed immunosensor based on method B was in- the immunosensors were incubated with 1.0 ng mL  1 HIgG so-
vestigated by measuring the current generated from the reduction lution containing 100 ng mL  1 of the interference of BSA, L-cy-
of benzoquinone. As shown in Fig. 4C, the peak current of ben- steine, tyrosinase (TYR) and prostate specific antigen (PSA), re-
zoquinone increased with concentrations of HIgG in the pH 7.4 PBS spectively. There is no remarkable change in DPV peak current
in the presence of 1 mM HQ and 1.25 mM H2O2. It was indicated based on the two methods for the mixture in comparison with
that the amount of the immobilized Ag/C NC/Ab2 will indirectly individual HIgG solution. Therefore, it can be concluded that the
affected the reduction of the benzoquinone which was an oxida- selectivity of the proposed immunosensors was acceptable. The
tion product of HQ catalyzed by Ag/C NC in presence of H2O2. Si- stability of the immunosensor is a key factor in developing of good
milarly, the amount of the HIgG will directly lead to change of Ag/C immunosensor. As shown in Fig. S4C–D, the immunosensor was
NC and then affected the response of benzoquinone. Through the stored at 4 °C when it was not in use, and it retained 94.53%
monitor the reduction current of benzoquinone, the amount of the (method A) and 96.63% (method B) of its initial response when
detected HIgG can be obtained. It showed a good linear relation- detecion of 10 ng mL  1 HIgG after 15 days of storage. The fact
ship between currents and the logarithm concentrations of HIgG indicates the good stability of the proposed immunosensor.
from 0.01 to 100 ng mL  1 (Fig. 4D). The regression equation was I
(μA) ¼ 22.89 to 2.211 log c (ng mL  1) with the correlation coef- 3.6. Application of the immunosensor
ficient of 0.9921 and the lowest detection concentration of
0.001 ng mL  1. In order to further investigate the applicability of the proposed
Comparing of the results of the two methods, we can know that immunosensor for real samples. The three serum samples were
the detection ranges for HIgG were overlapped from 0.1 to diluted to proper concentrations with PBS (pH 7.4, 0.1 M), then the
100 ng mL  1 when detecting of HIgG. Based on the double-signal immunosensor was used to detect of the concentration of HIgG.
strategy, according to the two linear equations, quantitative de- The standard addition method was employed to validate the ac-
tection of HIgG could be realized. Meanwhile, the detection results curacy and precision of the proposed immunosensor (Wang et al.,
can be supported by each other simply and conveniently. Fur- 2015). By adding different standard concentrations of HIgG to the
thermore, the linear ranges and detection limits of the proposed diluted human serum samples (Zhao et al., 2015; Mazloum-Ar-
immunosensor were compared with previous reports, and the dakani et al., 2015) and then dropped on the modified electrode to
results are listed in Table S1. It was found that the immunosensor form the sandwich-type structure followed by the conjugation
exhibited good performance. The results demonstrated that the with the Ab2 complex, the samples were detected based on the
proposed double-signal immunosensor could be used to detect double signals strategy. The detection results come from the two
S. Zhang et al. / Biosensors and Bioelectronics 77 (2016) 1078–1085 1085

Table 1 and Technology Department (14JJ4030, 13JJ2020), the Opening


Comparison of the responses obtained by the proposed immunosensor (containing Fund of state key Laboratory of Chemo/Biosensing and Chemo-
method A and method B) towards HIgG in serum samples.
metrics (Hunan University, 2013017), the Construct Program of the
Sample Method Added Found: mean 7SD Recovery RSD Key Discipline and Aid Program for Science and Technology In-
(ng mL  1) (ng mL  1 n ¼6) (%) (%, n ¼6) novative Research Team in Higher Educational Institutions of Hu-
nan Province.
Human serum A 0.00 9.39 7 0.28 – 2.98
B 0.00 9.42 7 0.21 – 2.23
A 0.10 9.50 7 0.31 110.0 3.26
B 0.10 9.517 0.34 90.0 3.58
Appendix A. Supplementary material
A 50.0 58.34 7 1.31 97.9 2.25
B 50.0 60.81 7 1.59 102.3 2.61
A 100.0 112.687 3.76 103.3 3.34 Supplementary data associated with this article can be found in
B 100.0 107.83 7 2.14 98.41 1.98 the online version at http://dx.doi.org/0.1016/j.bios.2015.10.089.

signals were consistent with each other when detection of the


same sample, as shown in Table 1. The recovery range was from References
97.9 to 110.0% and the relative standard deviations range from 2.25
to 3.34% by method A. While for method B, the recovery rate was Campanella, L., Martini, E., Tomassetti, M., 2008. Sens. Actuators B 130, 520–530.
Liu, L., Li, Y., Tian, L., Guo, T., Cao, W., Wei, Q., 2015. Sens. Actuators B 211, 170–176.
90.0–102.3% and the relative standard deviations ranged from 1.98 Li, R., Wu, K., Liu, C., Huang, Y., Wang, Y., Fang, H., Zhang, H., Li, C., 2014. Anal. Chem.
to 3.58%. Meanwhile, the results based on the proposed method 86, 5300–5307.
were compared with that of the method in hospital, which was Zhao, C., Liu, W., Ling, H., Lu, S., Zhang, Y., Liu, J., Xi, R., 2007. J. Agric. Food Chem. 55,
6879–6884.
shown in Table S2 in the supporting information. It can be con-
Chen, L., Zeng, X., Si, P., Chen, Y., Chi, Y., Kim, D.H., Chen, G., 2014. Anal. Chem. 86,
cluded that results originated from the proposed strategy is sa- 4188–4195.
tisfactory. Thus, we can conclude that the proposed immunosensor Xu, S., Liu, Y., Wang, T., Li, J., 2011. Anal. Chem. 83, 3817–3823.
could be used for clinical application. Ahn, J., Shin, Y.B., Lee, J., Kim, M.G., 2015. Biosens. Bioelectron. 71, 115–120.
Iizumi, Y., Okazaki, T., Ikehara, Y., Ogura, M., Fukata, S., Yudasaka, M., 2013. ACS
Appl. Mater. Interfaces 5, 7665–7670.
Yang, Z.H., Zhuo, Y., Yuan, R., Chai, Y.Q., 2015. Biosens. Bioelectron. 69, 321–327.
4. Conclusions Du, D., Zou, Z., Shin, Y., Wang, J., Wu, H., Engelhard, M.H., Liu, J., Aksay, I.A., Lin, Y.,
Chem., A., 2010. Anal. Chem. 82, 2989–2995.
Zhu, Q., Chai, Y., Yuan, R., Zhuo, Y., Han, J., Li, Y., Liao, N., 2013. Biosens. Bioelectron.
In summary, a relatively simple, sensitive and convenient 43, 440–445.
electrochemical double-signal immunosensor for the detection of Mazloum-Ardakani, M., Hosseinzadeh, L., Taleat, Z., 2015. Biosens. Bioelectron. 74,
30–36.
HIgG based on PDA-rGO-AuNPs and Ag/C nanocomposite was
Sun, G., Zhang, L., Zhang, Y., Yang, H., Ma, C., Ge, S., Yan, M., Yu, J., Song, X., 2015.
proposed for the first time. In this study, Ag/C NC acting as a signal Biosens. Bioelectron. 71, 30–36.
label and antibodies carrier, could not only give electrochemical Zhao, L., Di, F., Wang, D., Guo, L.-H., Yang, Y., Wan, B., Zhang, H., 2013. Nanoscale 5,
signal directly from Ag/C NC but also catalyze the reduction of 2655.
Liu, C., Zhang, P., Zhai, X., Tian, F., Li, W., Yang, J., Liu, Y., Wang, H., Wang, W., Liu, W.,
H2O2 to oxidize HQ for detection of HIgG. From the experiment 2012. Biomaterials 33, 3604–3613.
results, we can conclude that the proposed immunosensor possess Lu, Q., Deng, J., Hou, Y., Wang, H., Li, H., Zhang, Y., Yao, S., 2015. Chem. Commun. 51,
excellent selectivity, acceptable stability and good reliability. The 7164–7167.
Gao, T., Liu, F., Yang, D., Yu, Y., Wang, Z., Li, G., 2015. Anal. Chem. 87, 5683–5689.
double signals strategy based electrochemical immunosensor can Hu, W., He, G., Zhang, H., Wu, X., Li, J., Zhao, Z., Qiao, Y., Lu, Z., Liu, Y., Li, C.M., 2014.
be used to evaluate the HIgG in real samples and the results were Anal. Chem. 86, 4488–4493.
satisfactory. Therefore, this double signals strategy might provide a Jaiswal, A., Ghosh, S.S., Chattopadhyay, A., 2012. Chem. Commun. 48, 407–409.
Tan, H., Chen, Y., 2012. Sens. Actuators B 173, 262–267.
promising and reliable method for construction other im-
Wu, S., Ju, H.X., Liu, Y., 2007. Adv. Funct. Mater. 17, 585–592.
munosensor of biomarkers in clinical medicine in the future. Kang, S.M., Park, S., Kim, D., Park, S.Y., Ruoff, R.S., Lee, H., 2011. Adv. Funct. Mater. 21,
108–112.
Li, L., Zhang, L., Yu, J., Ge, S., Song, X., 2015. Biosens. Bioelectron. 71, 108–114.
Ionescu, R.E., Cosnier, S., Herrmann, S., Marks, R.S., 2007. Anal. Chem. 79,
Acknowledgment 8662–8668.
Zhao, Y., Liu, L., Kong, D., Kuang, H., Wang, L., Xu, C., 2014. ACS Appl. Mater. Inter-
This work was supported by the National Natural Science faces 6, 21178–21183.
Wang, Y., Ma, H., Wang, X., Pang, X., Wu, D., Du, B., Wei, Q., 2015. Biosens. Bioe-
Foundation of China (21305042, 21375037), Scientific Research lectron. 74, 59–65.
Fund of Hunan Provincial Education Department (14B116), Science Zhao, Y., Zheng, Y., Zhao, C., You, J., Qu, F., 2015. Biosens. Bioelectron. 71, 200–206.

You might also like