Sheet Name Sheet Number REV Block 2 4 Input 3 2 HVDC 4 3 Axial Driver 5 2 Lvac 6 2 Customer I/O 7 2

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 7

1 2 3 4 5 6 7 8

A A

B
B

SHEET NAME SHEET NUMBER REV Notes:


BLOCK 2 4 1. This cover sheet controls the revision status of the entire document.
INPUT 3 2 2. Electronic signatures for both the schematic reviewers and approvers can be found
C within my workshop.
HVDC 4 3
AXIAL DRIVER 5 2 C
LVAC 6 2
CUSTOMER I/O 7 2

D D

E E

F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
APP. Yang Xuyong 2326492-61SCH 1/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8
GEHW Medical Systems
1 2 3 4 5 6 7 8

A NGPDU-61 Block Diagram A

Axial Power Control


1 TS3-1
440Y254-A
Input Power and Transformer ASM HVAC(3:1)
HVDC Assy 2 440Y254-B
TS3-2
Axial Driver
1 TS2-1 3 TS3-3
1 474Y273-A 440Y254-C
TS1-L1 474Y273-A TS3-4
2 TS2-2 HVDC Output
Main Input
TS1-L2 474Y273-B
2
3 3
474Y273-B
TS2-3
Sheet 5
TS1-L3 474Y273-C 474Y273-C
B GND
2
KSV-P24 B
1 4
KSV-A1
440Y254-A 3 1
2 PDU-24-F VSS-A 8 LP-CONT-CLSD
KSV-A2
440Y254-B 3
3 VSS-B
440Y254-C 5
Sheet 4 VSS-C

Sheet 3 208Y120-A 1
2 10 1
208Y120-B KXG-A1 CAP-P
5 7
3 11
208Y120-C KXG-A2 CAP-M
4 8
X0 CAP-N
5 6
Y0 KSS-A1
12 LVAC PWR DIST and UPS
1 KSS-A2 4
SHUNT TRIP
XG-CONT-CLSD Interface
2 7 1 120V
1 KSS-CONT-CLSD 208Y120-A
TEMP 1
2 2 0V
TEMP 2 208Y120-B
C 3
4
208Y120-C C
X0
HVAC(3:1) 5 TS5-0
Y0
LVAC(5:1) TS5-1 Console
TS5-2 Power
J3(2:1)
3 TS5-3
J8(2:1) KTG-P24
5 KTG-A1 TS5-4
6 KTG-A2 Sheet 6 TS5-5
Gantry
9 GANTRY-PWR-CLSD TS5-6
Thermal SW Power
TS5-7
TS4-1 TS5-8
TS4-2 TS5-9
To UPS
Control Board TS4-3 TS5-10 AUX Data
System Interface 2334820-X TS4-6 TS5-11 Processing

J7(12:1)
Cabinet Power
D J9 From UPS TS4-7 TS5-12
D
J3-1 TS4-8 TS5-13
J7(12:1)
PDU-24 1 To Gantry Switch NC Shorting Stub J3-2
14 PDU-PGND
PDU-24 2 J2(8:1)
J2(8:1)
Pin16 GANTRY-PWR

15 PDU-PGND
Pin19 XG-CONT

GANTRY-PWR-CLSD 3
16 GANTRY-PWR-1 J8-1
J1(5:1)
ESTP-SRC 4 J8-2 J1(5:1)
17 EXP-INTLK
LP-CONT-CLSD 5
18 CLSE LOOP CUSTMOR
XG-CONT-CLSD 6 CONNECTIONS
19 XG-CONT J6(9:1)
KSS-CONT-CLSD 7 J6(9:1) 1 Customer I/O
20 X-Ray LITE J9(25:1) PDU-24
PHASE LOSS 8 2 TS6-1
21 SYS LITE GANTRY-PWR-CLSD LVAC INPUT
ILK-OPN 9 3 TS6-2
J5(4:1) ESTP-SRC
22 RDY LITE J12-1 TS6-3
Pin21 SYS LITE

BUSS-FLT 10 4 PGND
23 XG-ENBL-CONT J12-2 X-Ray LIGHT
Pin14 PGND

TS6-4
E OVR-TEMP 11
24 TS6-5 E
12
25
J11(8:1) 3
1
CUST24 Sheet 7 TS6-6
SYS-ON LIGHT
13 J4-1 CUST-NEUT
6 TS6-7
J4-2 X-Ray-ON LITE READY LIGHT
To Kdpp Coil Power Supply J10(3:1)
2 TS6-8
J5(4:1) SYS-ON LITE
8 TS6-9
120V ~ 24V+ GEN-RDY LITE DOOR INTLK SW
1 TS6-10
0V ~ 0V EXP-INLTK
J11(8:1) 2
PGND

J10(3:1)

J4(2:1)

F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
4 Update J6 information 2020-Aug GuoHX 2326492-61SCH
APP. Yang Xuyong 2/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8
GEHW Medical Systems
1 2 3 4 5 6 7 8

A A

Input Power and Transformer Assy


B B

T1

X0 X0
1 2 X1 208/120-A
1 2

H1
1 2 3 4 5 6 1 2 X2 208/120-B

TS1 CB1-125A 1 2 X3 208/120-C


C TS1-L1
1 2 1 2 C
3 4 H2
TS1-L2 Y0 Y0
1 2 3 4 5 6
5 6
TS1-L3 Y4 Y1

GND 474Y273-A

SHUNT TRIP 474Y273-B


1 2
Y5 Y2
474Y273-C
H3
1 2 3 4 5 6

440Y254-A
C1 Y6 Y3

C2 440Y254-B
C1 C2 C3 440Y254-C

D D
TEMP1
C1,C2,C3=6.0uF/370VAC
TEMP2
FLOATTING WYE CONFIGURATION
REQUIRED FOR EMC PREVENTION
GE97F9595, OR MALLORY C32FD3706 OR EQUAL

Jump Shown In 480V Connection

Voltage Connection
380V 2-6
400V 2-5
E 420V 2-4 E
440V 3-6
460V 3-5
480V 3-4

Note: All Input Max Power 150KVA

F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
APP. Yang Xuyong 2326492-61SCH 3/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8
GEHW Medical Systems
1 2 3 4 5 6 7 8

A A

HVDC Assembly

B
VF3
B
VF2

VF1

C4,C5,C6 = 6.0uF/370VAC
GROUNDED WYE CONFIGURATION
KXG REQUIRED FOR EMC PREVENTION
GE97F9595, OR MALLORY C32FD3706 BR1
13/17 14/18
OR EQUAL TS2
474Y273-A F1 5 6 + C7 1
AC
= 4700uF
474Y273-B F2 3 4 2
AC
474Y273-C 1 2 3Ph C8
F3 - 3
C AC
KXG-A1 Choke 4700uF
A1 A2

KXG-A2 C4 C5 C6 C
6uF 6uF 6uF

PDU-24

XG-CONT-CLSD

KSS-CONT-CLSD

CAP-P
KSS
13 14 CAP-M

5 6 R1 CAP-N

3 4 R2
D D
1 2 R3
KSS-A1 A1 A2
R1/R2/R3=60W
KSS-A2

E E

F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
3 Update KXG auxiliary contact number 2018-10
APP. Yang Xuyong 2326492-61SCH 4/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8
GEHW Medical Systems
1 2 3 4 5 6 7 8

A A

B
B

Axial Driver Power Control

C
C
LP-CONT-CLSD

KSV
13 14
CB2 16A/D TS3
440Y254-A 1 2 5 6 1

440Y254-B 3 4 3 4 2
440VAC to Gantry
440Y254-C 5 6 1 2 3 Axial Driver
KSV-A1 A1 A2 4
D D
KSV-A2

PDU-24

E E

F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
APP. Yang Xuyong 2326492-61SCH 5/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8
GEHW Medical Systems
1 2 3 4 5 6 7 8

A 1 2 CB10 6A/C A
UPS Power Interface
================================================================
PDU shipped with jumpers A,B,C on TS4 as show
A1 G 15 A1 G 15

To connect three phase UPS: U1 U2 U3 U1 U2 U3


1. Remove jumpers A,B,C
VMR2 VMR2
2. Connect UPS input power to TS4-1,2,3
3. Connect UPS output power to TS4-6,7,8 25 26 28 25 26 28
4. Connect input & output neutral & ground wires to TS4-4,9 & 5,10
16 18 A2 16 18 A2
----------------------------------------------------------------
B To connect two phase UPS: B
1. Remove jumpers A,C
120V
2. Connect UPS input power to TS4-1,3
3. Connect UPS output power to TS4-6,8 KCN 0V
To PDU 24Vdc PS
4. Connect input & output neutral & ground wires to TS4-4,9 & 5,10 13 14

1 2 CB9 6A/C 5 6 TS5


1 2 3 4 0

3 4 1 2 1
CB7 20A/D A1 A2
Console Power
2

3
GANTRY-POW CLSD KTG
KTG-P24 13 14
C TS5 C
1 2 CB5A 40A/D 5 6 4

1 2 CB5B 40A/D 3 4 5

1 2 CB5C 40A/D 1 2 6
Gantry Power
KTG-A1 A1 A2 7

KTG-A2 8

KDP
13 14
TS5
1 2 5 6 9

3 4 3 4 10

D 5 6 1 2 11 AUX Data Processing D


CB3 50A/K Cabinet Power
CB8 20A/D A1 A2 12
208Y120-A 1 2

208Y120-B 3 4 13

A1

5
208Y120-C 5 6
KDPP
SYSYLITE

A2

9
PDU PGND

E E
X0

Y0

TS4 TS4
1 2 3 4 5 6 7 8 9 10
Neutral Ground

LVAC Power Distribution


F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
APP. Yang Xuyong
2326492-61SCH 6/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8
GEHW Medical Systems
1 2 3 4 5 6 7 8

A A

Customer I/O

B B
PDU-24 Gantry-PWR
J5-1
Gantry-PWR-CLSD
J5-2

Gantry
E-Stop SYS-ON Gantry Power Switch
Power

ESTP-SRC
C J5-3 C
PGND Gantry-PWR-1
J5-4

TS6
CUST24 1
LVAC INPUT
CUST-NEUT 2

X-Ray-ON-LITE 3
X-RAY LIGHT
D 4
D
SYS-ON-LITE 5
SYS-ON LIGHT
6

GEN-RDY-LITE 7
READY LIGHT
8

EXP-INLK 9
DOOR INTLK SW
PGND 10

Customer Room Interface

E E

F F
DR. Guo Hexin NGPDU-61 Schematic Diagram
CH.
PAGE
APP. Yang Xuyong
2326492-61SCH 7/7
SIZE REV. ZONE DATE REV.BY

A3 MICRO 2 3 4 5 6 7 8