Quick Reference Guide

Elix® Essential 3, 5 TPK

© 2012 Millipore SAS


FTPF13692 - V1.0, 12/2012
Display and keypad

System status Warning icons

Tank Level Keypad

Values area Units Help indicators

Consumables status
From Standby, press twice to go to Ready.
From Ready, press once to go to Standby.

Operating status

Normal operation Alert message Alarm message


Press on the arrow keys to The display turns yellow to indicate The display turns red
display real time values of that a maintenance action is to indicate that an
the product water quality. required or you need to contact urgent action is required.
Service Representative.

System status Keypad

Maintenance tool This button is used to switch


The system is in Standby or between different system modes.
maintenance can be started. Press when indicated or required.

The arrow keys are used to navigate


Configuration or gears
in the user interface software.
Configuration parameters such
They are also used to change
as set points can be changed.
configuration or set point values.

Process or hour glass This button is used to confirm a change or


The system is in busy. an action required by the water system.
No action can be done. Press when indicated or required.

2
Consumables status

CL2 sanitization or pH cleaning


should be performed

Progard® pack must


be ordered or replaced

Contact Service Representative

The following symbole is always associated with an ERROR CODE


(for example ER:22).
For more information, please refer to the User Manual or contact
Service Representative.

When the key help icon is blinking, the keypad button next to the icon should
be pressed. See the User Manual for more information.

Warning icons

Product water resistivity


CL2 sanitization required
below set point

System sanitization required Low feed water pressure

Water detected

3
Maintenance schedule

Progard® pack

Order new Progard® pack. Replace Progard® pack.


It will be exhausted in 15 days. See the following pages
for how to replace the
consumables

Cl2 cleaning
1
Perform CL2 cleaning. Using the tool, unscrew
the sanitation port cap.

2 3 4

Remove the plug and insert Screw the plug back: Press for 1 second to
the tablet in the port. make surethe o-ring is in place. start CL2 sanitization.

5 6
The sanitization System goes in Ready
lasts 19 min. or Tank filling.

Press during the cycle to abort sanitization.

pH cleaning Perform a pH - acid or base cleaning depending on the type of feed


water.

Tank vent filter Replace tank vent filter when Progard® pack is replaced.

4
Replacing Progard® pack

Removing exhausted pack


Verify or put the system in Standby by pressing - MODE key.
1- Unlock the locking handle by pulling up.
2- Pull out the pack.
Note: when the Progard® Pack is removed, a red color
display appears. This is normal. This will go away
when a new pack is installed.

Placing new pack


Remove the new pack from its shipping box. Remove the 2 protective caps
from the ports at the top of the pack.
Locate the o-ring inside each port. Make sure the o-rings are
pushed inside against the inner edge of the ports.
1- Mount the bottom of the Pack onto the small hole in the cabinet.
Swing the rest of the Pack so that the top of the Pack aligns
with pack locking handle and the cabinet ports.
2- Push the top of the Pack in-place.
3- Push the Pack locking handle down to secure
the Pack.
• The new pack is detected.
• The display prompts you to start a Progard® flush:
press for 1 second.
• Wait 15 minutes for the Progard® flush to finish.