Connect DR to Generator
Service Management Dept.
SHIMADZU Corp.
Contents
1. Required PCB Board
2. Install PCB Board
3. Connect serial communication cable
4. Connect X-ray exposure signal cable
Appendix How to get logs of serial communication
Appendix ARCNET setting when using COMDAR
1. Required PCB Board
Requires 2 cable when connecting DR and Generator
• Serial communication cable
• X-ray exposure signal cable
PCB board that cables are connected to are different depending on DR combined
・ SF-X2 (Edge) with Fuji Film or Serial Exposure (DDR) with Konica Minolta
→ Require COMDAR-2009COMDAR-2009
・ Other cases
→ available to use either of COMDAR-2009 or COM I/F-2002
1. Required PCB Board
X-ray exposure signal cable should be connected to COMDAR-2009
or TERMINAL-2002 / TABLE I/F
・ SF-X2 (Edge) with Fuji Film or Serial Exposure (DDR) with Konica Minolta
→ Require COMDAR-2009
・ Other cases
→ TABLE I/F-2002 (or TERMINAL-2002 if TABLE I/F is not mounted in the cabinet)
2. Install PCB Board
Install COMDAR-2009 In case of SF-X2 system (Edge system),
Should be installed beneath UD CONT 2-3 on JP9 should be short-cut for supplying power
Connect to CH-200
Connect by selecting either cable in the table below
UD cable CH Note
TERMINAL-2002 X-RAY I/F Cable B CHC200/CHC200PA Without TOMOS
COMDAR- CN9 J6 P/N 563-59442 J25
2009
COMDAR-2009 CH-XGEN COMDAR CABLE CHC200/CHC200PA With TOMOS
CN MECA J13 P/N 565-15048-06 J20
COM I/F-2002
In case of SF-X2 (Edge) system,
1-2 on JP4 should be short-cut
for supplying power
Connect to CH-200
UD cabinet Cable CH cabinet
Terminal 2002 J6 J6 J25 J25 CHC-200
422 I/F-2002
3. Connect serial communication cable
Connect to COMDAR-2009
select either way of ① or ②
COM I/F
562-23733, COM conversion cable
9pin DR
9pin D-Sub
J7
3. Connect serial communication cable
[Communication] Setting
In case that system has touch-panel type
3. Connect serial communication cable
[Communication] Setting ・ In case of SF-X2 system ・ Others
In case that system has touch-panel type
No
Yes
TABLE I/F-2002 Connect to TABLE I/F-2002
No
Connect toTERMINAL-2002
4. Connect X-ray exposure signal cable
No
Yes
TABLE I/F-2002 Connect to TABLE I/F-2002
No
Connect toTERMINAL-2002
Refer to Installation manual (M501-E356) 2.9.4 Connecting with Fujifilm SF-X2 system
Table 2 2
Stand 5 3
Free 3 2
Table 2 2
Stand 5 3
Free 3 2
・TABLE I/F out-1 No.9 : [902]… Audio output signal for Auto Voice of Serial Radiography
・DAR I/F out-1 No.1 : [784]… Hand Switch signal of CH-200
・DAR I/F out-2 No.15 : [782]… Table Protocol
・DAR I/F out-2 No.16 : [783]… Stand Protocol
・DAR I/F out-3 No.7 : [808]… RDY1 for Table Serial Radiography
・DAR I/F out-3 No.8 : [611]… RDY2 for Table Serial Radiography
・DAR I/F out-3 No.9 : [612]… RDY2 for Stand Serial Radiography
・DAR I/F out-3 No.5 : [809]… RDY1 for Stand Serial Radiography (when COMDAR-2009
used)
4. Connect X-ray exposure signal cable
No
Yes
TABLE I/F-2002 Connect to TABLE I/F-2002
No
Connect toTERMINAL-2002
4. Connect X-ray exposure signal cable
Connect to TABLE I/F-2002 (In case of Fujifilm DR)
Connect When using Tech System 6
DR system
TABLE I/F-2002
1ch 2ch SHOT GO B
SHOT GO A
OUT7A SHOT1 RDY 2 SHOT2 RDY 2
OUT7B SHOT1 #RDY 2 SHOT2 #RDY 2
IN4A SHOT1 GO A SHOT2 GO A
IN4B SHOT1 GO B SHOT2 GO B
OUT6A SHOT1 RDY 1 SHOT2 RDY 1 SHOT #RDY 2
SHOT RDY 2
OUT6B SHOT1 #RDY 1 SHOT2 #RDY 1 SHOT #RDY 1
SHOT RDY 1
・ [Signal] setting (no need to change anything)
4. Connect X-ray exposure signal cable
Connect to TABLE I/F-2002 (In case of Fujifilm DR)
Connect When using Tech System 5
DR system
TABLE I/F-2002
1ch 2ch SHOT GO B
SHOT GO A
OUT1A SHOT1 RDY 2 SHOT2 RDY 2
OUT1B SHOT1 #RDY 2 SHOT2 #RDY 2
IN0A SHOT1 GO A SHOT2 GO A
IN0B SHOT1 GO B SHOT2 GO B
OUT0A SHOT1 RDY 1 SHOT2 RDY 1 SHOT #RDY 2
SHOT RDY 2
OUT0B SHOT1 #RDY 1 SHOT2 #RDY 1 SHOT #RDY 1
SHOT RDY 1
・ [Signal] setting (no need to change anything)
4. Connect X-ray exposure signal cable
Connect to TABLE I/F-2002 (In case of Fujifilm DR)
Connect When using Tech System 4
DR system
TABLE I/F-2002
1ch 2ch SHOT GO B
SHOT GO A
OUT3A SHOT1 RDY 2 SHOT2 RDY 2
OUT3B SHOT1 #RDY 2 SHOT2 #RDY 2
IN1A SHOT1 GO A SHOT2 GO A
IN1B SHOT1 GO B SHOT2 GO B
OUT2A SHOT1 RDY 1 SHOT2 RDY 1 SHOT #RDY 2
SHOT RDY 2
OUT2B SHOT1 #RDY 1 SHOT2 #RDY 1 SHOT #RDY 1
SHOT RDY 1
・ [Signal] setting (no need to change anything)
4. Connect X-ray exposure signal cable
Connect to TABLE I/F-2002 (In case of Canon DR)
Connect When using Tech System 6
No
Yes
TABLE I/F-2002 Connect to TABLE I/F-2002
No
Connect toTERMINAL-2002
4. Connect X-ray exposure signal cable
Connect to TERMINAL-200 (In case of Fujifilm DR)
Connect When using Tech System 2
TERMINAL I/F-2002 DR system
SHOT #RDY 2
X1-1B2 SHOT1 RDY 2
SHOT RDY 2
X1-1B2 SHOT1 #RDY 2
X2-0V SHOT1 GO A SHOT #RDY 1
SHOT RDY 1
X2-1B1 SHOT1 GO B
X1-KC2 SHOT1 RDY 1 SHOT GO B
X1-KC2 SHOT1 #RDY 1 SHOT GO A
・ TERMINAL-2002
SW1-1 ON
SW2-1 OFF
・ [Signal] setting (no need to change anything)
4. Connect X-ray exposure signal cable
Connect to TERMINAL-200 (In case of Canon DR)
Connect When using Tech System 3
TERMINAL I/F-2002 DR system
#1 WHT
X1-2B2 #1 WHT #2 YEL
X1-2B2 #2 YEL
X2-0V #3 ORG
X2-2B1 #4 GRY
X1-KC - #4 GRY
X1-KC - #3 ORG
・ TERMINAL-2002
SW1-2 ON
SW2-2 OFF
・ [Signal] setting (no need to change anything)
4. Connect X-ray exposure signal cable
Connect to TERMINAL-2002 (In case of Konica Minolta DR)
Connect when using Tech System 2
When to use XG cable set DC(A6DE) or XG cable set 2(A7R7) ※(A7R7)
Ex.)