Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

The following Contamination Prevention Parameters are specific

to DxC 700 AU software versions 1.07 and previous


DxC 700 AU Reagent Contamination Prevention Parameters OUS^

WATER PREVENT USE


PRECEDINGTEST FOLLOWING REAGENT PROBE WASH CLEANING
NO. NAME TEST NAME CLEANER KIND COUNT EFFECTIVE MIXER CUVETTE

1. TRG1N, TRIG1NP LIP1N, LIP2N Clin Lipase² 3 No Yes No


OSR6x118 OSR6x30
2. HDL1N LIP1N, LIP2N Clin Lipase² 5 No Yes No
OSR6x87 OSR6x30
3. LDL1N LIP1N, LIP2N Clin Lipase² 5 No Yes No
OSR6x83 OSR6x30
4. CRP3G CRP1N, CRP2N Water 1 Yes Yes No
OSR6x47 OSR6x99
5. CRE1N, CRE2N, C-31G, C-31GP Water 1 Yes Yes No
CRE1NP, CRE2NP OSR6x59
OSR6x78
6. CRE1N, CRE2N, C-41G, C-41GP Water 1 Yes Yes No
CRE1NP, CRE2NP OSR6x60
OSR6x78

7. CRE1N, CRE2N, PLB1G Water 1 Yes Yes No


CRE1NP, CRE2NP OSR6x75
OSR6x78
8. CRE1N, CRE2N, ULB1G CLN 100%* 3 No Yes No
CRE1NP, CRE2NP B38858/B46435
OSR6x78
9. TP-1N MG-1N Water 0 No Yes No
OSR6x32 OSR6x89
10. AST1N , AST1NP, LDH2N , LDH2NP Water 1 Yes No No
AST2N , AST2NP, OSR6x26
AST3N, AST3NP
OSR6x09
11. AST1N , AST1NP, LDH1N Water 1 Yes No No
AST2N , AST2NP, OSR6x27
AST3N, AST3NP
OSR6x09

Page 1 of 8 BLCAPDxC700AU.05
2019-08
12. AST1N , AST1NP, LDH3N , LDH3NP Water 1 Yes No No
AST2N , AST2NP, OSR6x28
AST3N, AST3NP
OSR6x09
13. CHO1N, CHO2N PHO1N Water 1 Yes No No
OSR6x16 OSR6x22
14. DBC1N, DBC1NP BUN1N, BUN1NP Water 1 Yes No No
OSR6x11 OSR6x34
15. DBB1N, DBB1NP BUN1N, BUN1NP Water 1 Yes No No
OSR6x11 OSR6x34
16. PHO1N UBC1N Water 0 No No Yes
OSR6x22 OSR6x205
17. CRE1N, CRE2N, FE-1N, FE-1NP Water 0 No No Yes
CRE1NP, CRE2NP OSR6x86
OSR6x78
18. TRG1N, TRG1NP FE-1N, FE-1NP Water 0 No No Yes
OSR6x118 OSR6x86
19. DIG1N ULB1G CLN 100%* 1 No No No
OSR6404 B38858/B46435
20. ULB1G DGT1N
CLN 100%* 1 No No No
B38858/B46435 OSR6403
21. A1C1G
All CLN 100%* 2 No No No
B00389
22. A1C1G A1C1G
CLN 100%* 2 No No Yes
B00389 B00389
23. THB1G UCP1N
CLN 100%* 2 No No No
B00389 OSR6x70

* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed.

² ODR20067 Cleaning Solution (for contamination prevention). Replace weekly or as required.


^ OUS: For Europe, Emerging Markets, Asia

Page 2 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Third Party Product Contamination Prevention Parameters OUS^

WATER PREVENT USE


PRECEDINGTEST FOLLOWING REAGENT PROBE WASH CLEANING
NO. NAME TEST NAME CLEANER KIND COUNT EFFECTIVE MIXER CUVETTE
TRG1N/TRG1P
1. OSR6x118
HDL1N LPS
2. OSR6x87 B82715
https://intranet.sentinel.it/BCI_Applications/
LDL1N
3. OSR6x83
MYO1N
4. OSR6168 ZINC
https://intranet.sentinel.it/sentineldoc/
IGE E9217640 / 17640H Username: beckmanclients
5. 133390 Password: bc12263122
GLDH AMM1G
6. G82100 OSR6x154
AMM1G LDH2N/LDH2NP
7. OSR6x154 OSR6x26
QMS® TAC https://www.thermofisher.com/bci-applications
8. A53727 ULB1G
QMS® EVR B38858/B46435
9. A53716
LIP1N, LIP2N
10. OSR6x30
UBC1N
11. OSR6x205
HCY LDH2N/LDH2NP
12. B08176 OSR6x26
LDH1N http://www.homocysteine.co.uk/BCI/
13. OSR6x27
LDH3N/LDH3NP
14. OSR6x28
UBC1N HCY1G
15. OSR6x205 B08176
GEN1G
16. All CLN-20%* 5 No Yes Yes
OSR4T229
* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed.

For carryover to and from third party products except for Siemens GEN1G OSR4T229, where
Beckman Coulter is not the manufacturer, consult the manufacturer’s instructions for appropriate
contamination prevention parameters.
^ OUS: For Europe, Emerging Markets, Asia

Page 3 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Sample Carryover Contamination Prevention Parameters OUS^

Contamination Parameters → Carry-Over Prevention (Test)

PRE-DISPENSE WASH COUNT POST- DISPENSE WASH COUNT

TEST NAME DETERGENT -1 DETERGENT -2 WATER DETERGENT -1 DETERGENT -2 WATER


ULB1G
1 0 1 0 0 0
B38858/B46435

CRP1N, CRP2N
2 0 1 0 0 0
OSR6x99

DxC 700 AU Sample Carryover Contamination Prevention Parameters OUS^

Contamination Parameters →Carry-Over Prevention (Type Changes)

WASH COUNT
NO. COMBINATION DETERGENT -1 DETERGENT -2 WATER
2 Serum to Urine 1□ 0 1□

Detergent 1 = 2% Wash Solution (OSR0001/ODR2000), replace cleaner on a daily basis or as


required. Replace the bottle as needed.

□ Recommended sample contamination prevention parameters for Urine/CSF Albumin


B38858/B46435.Test or type change parameters can be used for Urine/CSF Albumin
B38858/B46435 depending on lab usage.
^ OUS: For Europe, Emerging Markets, Asia

Page 4 of 8 BLCAPDxC700AU.05
2019-08
The following Contamination Prevention Parameters are specific
to DxC 700 AU software versions 1.08 and higher
DxC 700 AU Reagent Contamination Prevention Parameters OUS^

WATER PREVENT USE


PRECEDING FOLLOWING REAGENT PROBE WASH CLEANING
NO. TEST NAME TEST NAME CLEANER KIND COUNT EFFECTIVE MIXER CUVETTE
1. TRG1N, TRIG1NP LIP1N, LIP2N Clin Lipase² 3 No Yes No
OSR6x118 OSR6x30
2. HDL1N LIP1N, LIP2N Clin Lipase² 5 No Yes No
OSR6x87 OSR6x30
3. LDL1N LIP1N, LIP2N Clin Lipase² 5 No Yes No
OSR6x83 OSR6x30
4. CRP3G CRP1N, CRP2N Water 1 Yes Yes No
OSR6x47 OSR6x99
7. CRE1N, CRE2N, C-31G, C-31GP Water 1 Yes Yes No
CRE1NP, CRE2NP OSR6x59
OSR6x78
8. CRE1N, CRE2N, C-41G, C-41GP Water 1 Yes Yes No
CRE1NP, CRE2NP OSR6x60
OSR6x78
9. CRE1N, CRE2N, PLB1G Water 1 Yes Yes No
CRE1NP, CRE2NP OSR6x75
OSR6x78
10. CRE1N, CRE2N, ULB1G CLN 100%* 3 No Yes No
CRE1NP, CRE2NP B38858/B46435
OSR6x78
11. TP-1N MG-1N Water 0 No Yes No
OSR6x32 OSR6x89
12. AST1N , AST1NP, LDH2N , LDH2NP Water 1 Yes No No
AST2N , AST2NP, OSR6x26
AST3N, AST3NP
OSR6x09
13. AST1N , AST1NP, LDH1N Water 1 Yes No No
AST2N , AST2NP, OSR6x27
AST3N, AST3NP
OSR6x09
14. AST1N , AST1NP, LDH3N , LDH3NP Water 1 Yes No No
AST2N , AST2NP, OSR6x28
AST3N, AST3NP
OSR6x09
15. CHO1N, CHO2N PHO1N Water 1 Yes No No
OSR6x16 OSR6x22
16. DBC1N, DBC1NP BUN1N, BUN1NP Water 1 Yes No No
OSR6x11 OSR6x34
17. DBB1N, DBB1NP BUN1N, BUN1NP Water 1 Yes No No
OSR6x11 OSR6x34
18. PHO1N UBC1N Yes
OSR6x22 OSR6x205
19. CRE1N, CRE2N, FE-1N, FE-1NP Yes
CRE1NP, CRE2NP OSR6x86
OSR6x78
20. TRG1N, TRG1NP FE-1N, FE-1NP Yes
OSR6x118 OSR6x86
21. DIG1N ULB1G CLN 100%* 1 No No No
OSR6404 B38858/B46435
22. ULB1G DGT1N
CLN 100%* 1 No No No
B38858/B46435 OSR6403
23 A1C1G All
Yes(CLN-1)
B00389

Page 5 of 8 BLCAPDxC700AU.05
2019-08
24. THB1G UCP1N
CLN 100%* 2 No No No
B00389 OSR6x70
25. DENAT All Yes
B93009
26. A1c All
Yes(CLN-1)
B93009

* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed. CLN-1 = CLN 100%

² ODR20067 Cleaning Solution (for contamination prevention). Replace weekly or as required.


^ OUS: For Europe, Emerging Markets, Asia

Page 6 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Third Party Product Contamination Prevention Parameters OUS^

WATER PREVENT USE


PRECEDING FOLLOWING REAGENT PROBE WASH CLEANING
NO. TEST NAME TEST NAME CLEANER KIND COUNT EFFECTIVE MIXER CUVETTE
TRG1N/TRG1P
1. OSR6x118
HDL1N LPS
2. OSR6x87 B82715
https://intranet.sentinel.it/BCI_Applications/
LDL1N
3. OSR6x83
MYO1N
4. OSR6168 ZINC
https://intranet.sentinel.it/sentineldoc/
IGE E9217640 / 17640H Username: beckmanclients
5. 133390 Password: bc12263122
GLDH AMM1G
6. G82100 OSR6x154
AMM1G LDH2N/LDH2NP
7. OSR6x154 OSR6x26
QMS® TAC https://www.thermofisher.com/bci-applications
8. A53727 ULB1G
QMS® EVR B38858/B46435
9. A53716
LIP1N, LIP2N
10. OSR6x30
UBC1N
11. OSR6x205
HCY LDH2N/LDH2NP
12. B08176 OSR6x26
LDH1N http://www.homocysteine.co.uk/BCI/
13. OSR6x27
LDH3N/LDH3NP
14. OSR6x28
UBC1N HCY1G
15. OSR6x205 B08176
GEN1G
16. All CLN-20%* 5 No Yes No
OSR4T229
All GEN1G Yes
17. OSR4T229

* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed.

For carryover to and from third party products except for Siemens GEN1G OSR4T229, where
Beckman Coulter is not the manufacturer, consult the manufacturer’s instructions for appropriate
contamination prevention parameters.
^ OUS: For Europe, Emerging Markets, Asia

Page 7 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Sample Carryover Contamination Prevention Parameters OUS^

Contamination Parameters → Carry-Over Prevention (Test)

PRE-DISPENSE WASH COUNT POST- DISPENSE WASH COUNT

TEST NAME DETERGENT -1 DETERGENT -2 WATER DETERGENT -1 DETERGENT -2 WATER


ULB1G
1 0 1 0 0 0
B38858/B46435

CRP1N, CRP2N
2 0 1 0 0 0
OSR6x99

DxC 700 AU Sample Carryover Contamination Prevention Parameters OUS^

Contamination Parameters →Carry-Over Prevention (Type Changes)

WASH COUNT
NO. COMBINATION DETERGENT -1 DETERGENT -2 WATER
2. Serum to Urine 1□ 0 1□
Whole Blood
21. 0 0 1
to Serum
Whole Blood
22. 0 0 1
to Urine
Whole Blood
23. 0 0 1
to Other-1
Whole Blood
24. 0 0 1
to Other-2

Detergent 1 = 2% Wash Solution (OSR0001/ODR2000), replace cleaner on a daily basis or as


required. Replace the bottle as needed.
□ Recommended sample contamination prevention parameters for Urine/CSF Albumin

B38858/B46435.Test or type change parameters can be used for Urine/CSF Albumin


B38858/B46435 depending on lab usage.
^ OUS: For Europe, Emerging Markets, Asia