Professional Documents
Culture Documents
Contamination Parameters DXC 700 AU
Contamination Parameters DXC 700 AU
Page 1 of 8 BLCAPDxC700AU.05
2019-08
12. AST1N , AST1NP, LDH3N , LDH3NP Water 1 Yes No No
AST2N , AST2NP, OSR6x28
AST3N, AST3NP
OSR6x09
13. CHO1N, CHO2N PHO1N Water 1 Yes No No
OSR6x16 OSR6x22
14. DBC1N, DBC1NP BUN1N, BUN1NP Water 1 Yes No No
OSR6x11 OSR6x34
15. DBB1N, DBB1NP BUN1N, BUN1NP Water 1 Yes No No
OSR6x11 OSR6x34
16. PHO1N UBC1N Water 0 No No Yes
OSR6x22 OSR6x205
17. CRE1N, CRE2N, FE-1N, FE-1NP Water 0 No No Yes
CRE1NP, CRE2NP OSR6x86
OSR6x78
18. TRG1N, TRG1NP FE-1N, FE-1NP Water 0 No No Yes
OSR6x118 OSR6x86
19. DIG1N ULB1G CLN 100%* 1 No No No
OSR6404 B38858/B46435
20. ULB1G DGT1N
CLN 100%* 1 No No No
B38858/B46435 OSR6403
21. A1C1G
All CLN 100%* 2 No No No
B00389
22. A1C1G A1C1G
CLN 100%* 2 No No Yes
B00389 B00389
23. THB1G UCP1N
CLN 100%* 2 No No No
B00389 OSR6x70
* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed.
Page 2 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Third Party Product Contamination Prevention Parameters OUS^
For carryover to and from third party products except for Siemens GEN1G OSR4T229, where
Beckman Coulter is not the manufacturer, consult the manufacturer’s instructions for appropriate
contamination prevention parameters.
^ OUS: For Europe, Emerging Markets, Asia
Page 3 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Sample Carryover Contamination Prevention Parameters OUS^
CRP1N, CRP2N
2 0 1 0 0 0
OSR6x99
WASH COUNT
NO. COMBINATION DETERGENT -1 DETERGENT -2 WATER
2 Serum to Urine 1□ 0 1□
Page 4 of 8 BLCAPDxC700AU.05
2019-08
The following Contamination Prevention Parameters are specific
to DxC 700 AU software versions 1.08 and higher
DxC 700 AU Reagent Contamination Prevention Parameters OUS^
Page 5 of 8 BLCAPDxC700AU.05
2019-08
24. THB1G UCP1N
CLN 100%* 2 No No No
B00389 OSR6x70
25. DENAT All Yes
B93009
26. A1c All
Yes(CLN-1)
B93009
* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed. CLN-1 = CLN 100%
Page 6 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Third Party Product Contamination Prevention Parameters OUS^
* CLN OSR0001/ODR2000, replace cleaner on a daily basis or as required. Replace the bottle as
needed.
For carryover to and from third party products except for Siemens GEN1G OSR4T229, where
Beckman Coulter is not the manufacturer, consult the manufacturer’s instructions for appropriate
contamination prevention parameters.
^ OUS: For Europe, Emerging Markets, Asia
Page 7 of 8 BLCAPDxC700AU.05
2019-08
DxC 700 AU Sample Carryover Contamination Prevention Parameters OUS^
CRP1N, CRP2N
2 0 1 0 0 0
OSR6x99
WASH COUNT
NO. COMBINATION DETERGENT -1 DETERGENT -2 WATER
2. Serum to Urine 1□ 0 1□
Whole Blood
21. 0 0 1
to Serum
Whole Blood
22. 0 0 1
to Urine
Whole Blood
23. 0 0 1
to Other-1
Whole Blood
24. 0 0 1
to Other-2
Page 8 of 8 BLCAPDxC700AU.05
2019-08