Professional Documents
Culture Documents
FLB06 - MU - Calculation - Sheet - Mechanical
FLB06 - MU - Calculation - Sheet - Mechanical
/
Sr. No. Discipline Group Specific test performed Test Method Range of testing MU Value Unit
Material
of test
Stones / Rock
3 Water content 0.01 % to 40 % 0.15 0.32 %
4 Point load index
Unconfined IS 8764 (1998): 2014 1 Mpa to 300 Mpa 0.36 6.11 MPa
5 Compressive Strength IS IS
9143 (1979):
10050 2016
(1981): 2 Mpa to 210 Mpa 1.69 68.84 %
6 Slake durability index 2016 5 % to 100 % 0.43 99.25 %
7 Specific gravity IS 2386 P3 (1963): 1 to 3.5 0.04 2.84
8 Water Analysis)
Absorption 2016 0.05 % to 50 % 0.37 5.75 %
(Sieve -
9 Gravel
Grain Size Analysis 0 %to 100 % 1.48 16.22 %
10 (Size
Silt Analysis) - Sand IS 2720 P4 (1985): 1to 100 1.72 31.03
Mechanical
IS 2720 P6 (1972):
15 Shrinkage Limit IS 27202016
P2 (1973): 5 %to 30 % 1.27 13.48 %
16 Moisture Content 2015 1 %to 75 % 1.18 35.06 %
IS 2720 P3 (1980):
17 Specific Gravity -
HeavyCompaction 2016 1.5to 3.5 0.05 2.44
18 MaximumDry
Content density IS 2720 P8 (1983): 1 g/ccto 3 g/cc 0.03 1.56 g/cc
19 2015 1 %to 40 % 1.29 14.96 %
Light Compaction -
20 MaximumDryDensity IS 2720 P7 (1980): 1 g/ccto 3 g/cc 0.03 1.41 g/cc
OptimumMoisture 2016
21 Content 1 %to 45 % 1.72 25.54 %
IS 2720 P40 (1977):
22 Free SwellIndex 2016 0 %to 200 % 1.87 30.67 %
Year of Calibration Date of last Calibration due
Sl no Name of equipment Department/Section Unique Lab ID Model/Installation ID Make Range Accuracy/LC
Purchase certificate No. calibration on *
1 Pressure Gauge Mechanical - Rock JGG-PG-01 --- --- SKE 10-25kn 0.25kn 20010169/PR/1 2/1/2020 1/31/2021
2 Pressure Gauge Mechanical - Rock JGG-PG-02 --- --- SKE 10-100kn 0.5kn SKE/F/2020/004 9/18/2020 9/17/2021
4 Muffle Furnace Chemical JGG-MF-01 AL-7981 --- Kalika Scientific/KS-18 Amb.-1150˚C 1˚C 2020/39616/01 2/1/2020 1/31/2021
5 Test Sieve 20mm Mechanical-Soil JGG-SS-01 --- --- SKE 20mm --- 2020/39711/01 12.02.20 2/10/2021
6 Test Sieve 19mm Mechanical-Soil JGG-SS-02 --- --- SKE 19mm --- 2020/39711/02 2/10/2020 2/10/2021
7 Test Sieve 10mm Mechanical-Soil JGG-SS-03 --- --- Secor India 10mm --- 2020/39711/03 2/10/2020 2/10/2021
8 Test Sieve 4.75mm Mechanical-Soil JGG-SS-04 --- --- Universal 4.75mm --- 2020/39711/04 2/10/2020 2/10/2021
9 Test Sieve 2mm Mechanical-Soil JGG-SS-05 --- --- SKE 2mm --- 2020/39711/02 2/10/2020 2/10/2021
10 Test Sieve 1.18mm Mechanical-Soil JGG-SS-06 --- --- Universal 1.18mm --- 2020/39711/06 2/10/2020 2/10/2021
11 Test Sieve 600µm Mechanical-Soil JGG-SS-07 --- --- Universal 600µm --- 2020/39711/07 2/10/2020 2/10/2021
12 Test Sieve 450µm Mechanical-Soil JGG-SS-08 --- --- Secor India 425µm --- 2020/39711/08 2/10/2020 2/10/2021
13 Test Sieve 150µm Mechanical-Soil JGG-SS-09 --- --- Universal 150µm --- 2020/39711/09 2/10/2020 2/10/2021
14 Test Sieve 75µm Mechanical-Soil JGG-SS-10 --- --- Universal 75µm --- 2020/39711/10 2/10/2020 2/10/2021
15 Laboratory Oven Mechanical JGG-LO-01 --- --- --- Amb-300˚C --- 2020/39616/02 2/1/2020 1/31/2021
16 Glass Thermometer All JGG-TH-01 JRM/--- --- JRM -10-110˚C 0.5˚C 2020/39711/13 2/7/2020 2/6/2021
17 Digital Stop Watch Mechanical-Soil JGG-SW-01 Racer/--- --- Racer 0-24hrs 0.01 sec 2020/39711/18 2/7/2020 2/6/2021
18 Burette Chemical JGG-BU-01 --- --- J. Sil 0-10ml 0.5ml 2020/39711/15 2/7/2020 2/6/2021
19 Digital Vernier Caliper Mechanical - Rock JGG-VC-01 --- --- Mitutoyo 0-150mm 0.010mm 2019/3539/11 4/17/2019 4/16/2020
20 Pipette Chemical JGG-PP-01 --- --- Avi Scitific 0-5ml 0.5ml 2020/39711/14 2/7/2020 2/6/2021
21 Burette Chemical JGG-BU-02 --- --- J. Sil 0-50ml 1ml 2020/39711/16 2/7/2020 2/6/2021
22 Set of weights E2 Class Mechanical-Soil JGG-WTB-01 --- --- UTC 1mg-100g --- 2020/39711/17 2/8/2020 2/7/2021
23 Digital Weighing Balance Mechanical JGG-WB-01 --- --- Hindustan Scale 5-300g 0.001g 2020/39616/03 2/1/2020 1/31/2021
Company
24 Digital Weighing Balance Mechanical JGG-WB-03 --- --- BPL 20g-10kg 0.5g 2019/35401/04 4/17/2019 4/16/2020
25 Digital Weighing Balance Mechanical JGG-WB-02 Hsco/--- --- Hsco 0.2g-2kg 0.01g 2020/39616/04 2/1/2020 1/31/2021
26 Vernier Caliper Mechanical - Rock JGG-VC-01 --- --- Mitutoyo 0-150mm 0.020mm 2020/39711/11 2/10/2020 2/10/2021
27 ph meter Chemical JGG-PH-01 --- --- Avi Scientific 0-14ph 0.01ph M-200217-5-1 2/19/2020 2/19/2021
28 Density Hydrometer Mechanical-Soil JGG-HY-01 --- --- Liemco 0.995-1.030g/ml 0.0005g/ml M-200217-5-2 2/17/2020 2/17/2021
29 Pressure Gauge Mechanical JGG-PG-03 --- --- ATLAS 0to28kg/cm2 0.4kg/cm2 FR-QT-21 2/13/2020 2/13/2021
30 Dial Indicator (P) Mechanical JGG-DI-01 --- --- Baker 0-25mm 0.01mm M-200211-5-4 2/11/2020 2/11/2021
31 Dial Indicator (P) Mechanical JGG-DI-02 --- --- Baker 0-25mm 0.01mm M-200211-5-2 2/11/2020 2/11/2021
32 Dial Indicator (P) Mechanical JGG-DI-03 --- --- Baker 0-25mm 0.01mm M-200211-5-5 2/11/2020 2/11/2021
33 Dial Indicator (P) Mechanical JGG-DI-04 --- --- Baker 0-25mm 0.01mm M-200211-5-3 2/11/2020 2/11/2021
34 Thermohygrometer Chemical JGG-THG-01 --- --- Avi Scientific 0-50˚C 1˚C 2020/39711/12 2/7/2020 2/6/2021
FRI/
36 Proving ring CBR Mechanical JGG-PV-01 --- --- 0-30kn 0.002mm 10/20/12752 10/22/2020 12/22/2022
FRI/
37 Proving ring DST Mechanical JGG-PV-02 --- --- Aimil 0-2kn 0.001kn 10/20/12750 10/23/2020 12/23/2022
FRI/
38 Proving ring TRI Mechanical JGG-PV-03 --- --- 0-2kn 0.002mm 10/20/12751 10/22/2020 12/22/2022
61 Directother
Shear loading device &
assccessories --- ---
62 Triaxialother
Shear loading device &
accessories --- ---
63 consolidation
loading device &
other accessories --- ---
64 Casagrande
apparatus & other
accessories --- ---
Lawkim 4 µm 25000 2
Lawkim 4 µm 10000 2
Lawkim 4 µm 10000 2
Lawkim 4 µm 10000 2
not req
not req
not req
not req
internal
not calibrated
not calibrated
not req
not req
not req
internal
internal
not req
not req
not req
internal
internal
not req
internal
internal
internal
internal
internal
internal
internal
internal
internal
internal
not req
not req
internal
internal
Jay Gajanan Geotechnics – Soil Mechanics Laboratory Page No. 1 of 2
1. Test Parameter:
Density
Test Method:
IS 13030 (1991):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 2.76 2.87 2.78 2.81 2.79
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Weighing Balance JGG-WB-02 1000 2 0.02 0.0000
Combined uncertainty :
= 0.019
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.05
Result:
1. Test Parameter:
Porosity
Test Method:
IS 13030 (1991):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 0.72 1.15 0.97 1.25 0.41
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Weighing Balance JGG-WB-02 1000 2 0.02 0.0000
Combined uncertainty :
= 0.152
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.42
Result:
1. Test Parameter:
Water Content
Test Method:
IS 13030 (1991):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 0.26 0.40 0.35 0.44 0.15
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Weighing Balance JGG-WB-02 1000 2 0.02 0.0000
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Combined uncertainty :
= 0.053
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.15
Result:
1. Test Parameter:
Point load index
Test Method:
IS 8764 (1998): 2014
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 6.29 6.37 6.03 6.21 5.67
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Pressure Gauge JGG-PG-01 20 2.04 0.20 0.0049
Combined uncertainty :
= 0.128
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.36
Result:
1. Test Parameter:
Unconfined
Compressive Strength
Test Method:
IS 9143 (1979): 2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 67.57 69.09 70.91 68.93 67.70
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Digital Compression
JGG-CTM-02 150 2 0.30 0.0010
Testing Machine
Combined uncertainty :
= 0.607
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.69
Result:
1. Test Parameter:
Slake
IS durability
10050 (1981):index
Test Method:
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 99.39 99.38 99.60 99.16 98.72
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Weighing Balance JGG-WB-03 500 2 0.30 0.0003
Combined uncertainty :
= 0.153
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.426538199053814
Result:
1. Test Parameter:
Specific gravity
Test Method:
IS 2386 P3 (1963):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 2.81 2.89 2.82 2.84 2.82
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Weighing Balance JGG-WB-02 500 2 0.02 0.0000
Laboratory Oven JGG-LO-01 1000 2 0.50 0.0003
Combined uncertainty :
= 0.014
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.039
Result:
1. Test Parameter:
Water Absorption
Test Method:
IS 2386 P3 (1963):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 5.43 5.69 5.52 6.11 5.99
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Weighing Balance JGG-WB-03 500 2 0.30 0.0003
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Combined uncertainty :
= 0.133
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.369459885439035
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 13.39 14.03 15.73 16.11 16.18
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Pressure Gauge JGG-PG-02 40 2 0.30 0.0038
Combined uncertainty :
= 0.579
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.61
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 0.24 0.25 0.23 0.25 0.26
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Digital Compression
JGG-CTM-02 150 2 0.30 0.0010
Testing Machine
ME Indicator JGG-ME-01 0 0 0.00 #DIV/0!
Combined uncertainty :
= #DIV/0!
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= #DIV/0!
U = K x Uc = #DIV/0!
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 1E+05 2E+05 1E+05 1E+05 2E+05
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Vernier Caliper JGG-VC-01 100 2 0.02 0.0001
Digital Compression
JGG-CTM-02 150 2 0.30 0.0010
Testing Machine
ME Indicator JGG-ME-01 0 0 0.00 #DIV/0!
Combined uncertainty :
= #DIV/0!
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= #DIV/0!
U = K x Uc = #DIV/0!
Result:
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 16.27 17.485 16.315 14.29 16.75
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Test Sieve 10mm JGG-SS-03 10000 2 17.10 0.0009
Test Sieve 4.75mm JGG-SS-05 4750 2 17.10 0.0018
Combined uncertainty :
= 0.533
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4.1
U = K x Uc = 1.48
Result:
1. Test Parameter:
Grain Size Analysis
(Size Analysis) - Sand
Test Method:
IS 2720 P4 (1985):
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
Jay Gajanan Geotechnics – Soil Mechanics Laboratory Page No. 2 of 2
Measurement uncertainty Calculation Sheet FLB06 / 01 / 01.12.2020
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 29.615 33.075 31.145 30.645 30.645
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Test Sieve 2mm JGG-SS-05 2000 2 2.00 0.0005
Test Sieve 1.18mm JGG-SS-06 1180 2 2.00 0.0008
Test Sieve 600µm JGG-SS-07 600 2 2.00 0.0017
Test Sieve 450µm JGG-SS-08 425 2 2.00 0.0024
Test Sieve 150µm JGG-SS-09 150 2 2.00 0.0067
Test Sieve 75µm JGG-SS-10 75 2 2.00 0.0133
Combined uncertainty :
= 0.744
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 12
U = K x Uc = 1.72
Result:
Test Method:
IS 2720 P4 (1985):
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 25.19 23.02 24.46 25.61 24.49
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Test Sieve 75µm JGG-SS-10 75 2 2.00 0.0133
Density Hydrometer JGG-HY-01 1.011 2 0.00 0.0006
Combined uncertainty :
= 0.554
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 10
U = K x Uc = 1.42
Result:
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 28.92 26.42 28.08 29.45 28.11
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Test Sieve 75µm JGG-SS-10 75 2 2.00 0.0133
Density Hydrometer JGG-HY-01 1.011 2 0.00 0.0006
Combined uncertainty :
= 0.640
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 10
U = K x Uc = 1.65
Result:
1. Test Parameter:
Liquid limit
IS 2720 P5 (1985):
Test Method:
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 61.25 58.69 62.25 60.03 61.5
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.644
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.79
Result:
1. Test Parameter:
Plastic
IS 2720Limit
P5 (1985):
Test Method:
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.9597116921089 26.2680088052174 27.66045548654 26.41 25.22
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.486
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.35
Result:
1. Test Parameter:
Shrinkage Limit
Test Method:
IS 2720 P6 (1972):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 13.14 14.83 13.32 14.01 12.1
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.457
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.27
Result:
1. Test Parameter:
Moisture
IS 2720 P2Content
(1973):
Test Method:
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 35.63 35.01 33.6 35.02 36.04
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.423
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.18
Result:
1. Test Parameter:
Specific Gravity
Test Method:
IS 2720 P3 (1980):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 2.41 2.42 2.47 2.4 2.48
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 10 2 0.02 0.0010
Balance
Combined uncertainty :
= 0.02
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 5
U = K x Uc = 0.05
Result:
1. Test Parameter:
HeavyCompaction -
MaximumDry density
Test Method:
IS 2720 P8 (1983):
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 1.57 1.56 1.59 1.56 1.53
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.01
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 5
U = K x Uc = 0.03
Result:
Test Method:
IS 2720 P8 (1983):
2015
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 15.33 14.21 13.63 16.22 15.41
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.46
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.29
Result:
1. Test Parameter:
Light Compaction -
MaximumDryDensity
IS 2720 P7 (1980):
Test Method:
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 1.42 1.38 1.43 1.39 1.43
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.01
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 5
U = K x Uc = 0.03
Result:
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 0.2567565 0.2766555 0.24643575 0.26336475 0.27272025
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.01
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.02
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
Free SwellIndex
Test Method:
IS 2720 P40 (1977):
2016
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 30 33.33 30 30 30
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 10 2 0.02 0.0010
Balance
Combined uncertainty :
= 0.67
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.87
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 6 5 6 7 6
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.32
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.88
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 0.322208180545162 0.347207452054365 0.317191736865 0.311347457111 0.329476104815
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.01
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.02
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 24.15 25.62 26.41 24.22 27.32
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Laboratory Oven JGG-LO-01 100 2 0.50 0.0025
Digital Weighing
JGG-WB-02 1000 2 0.02 0.0000
Balance
Combined uncertainty :
= 0.62
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 1.72
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 2.92 3.05 3.82 2.83 4.1
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Dial Indicator (P) JGG-DI-01 25000 2 4.00 0.0001
Digital Stop Watch JGG-SW-01 60 2 2.20 0.0183
Combined uncertainty :
= 0.26
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.74
Result:
1. Test Parameter:
#NAME?
Test Method:
#NAME?
Uncertainty
CRM/SRM Name Identification Certificate No. Source
Value Provided
5. Evaluation:
Type A Uncertainty
Obs. No. 1 2 3 4 5
Measured Value 0.019 0.018 0.018 0.017 0.015
Type B Uncertainty
Equipment/CRM/ Degrees of Relative std.
Equipment/CRM/SRM Values Uncertainty
SRM ID freedom Uncertainty
Combined uncertainty :
= 0.00
νeff = [ Uc4 ] / [N ∑i=1 {Ui4 (y / νi )} (Note : If any ‘ν’ value is ∞ effective degrees of freedom will be ∞)
= 4
U = K x Uc = 0.00
Result: