Download as xlsx, pdf, or txt
Download as xlsx, pdf, or txt
You are on page 1of 10

Manufacturer ECRI'S Bioventus LLC Integra

RECOMMENDED SonaStar FS1000 RF LifeSciences Corp


SPECIFICATIONS CUSA Clarity
Ultrasonic Aspirators

WHERE MARKETED Worldwide Worldwide


FDA CLEARANCE Yes Yes
CE MARK Yes Yes
INDICATIONS Gastrointestinal and Surgical procedures
affiliated organ where fragmentation,
surgery, general, emulsification and
gynecologic, aspiration of soft and
laparoscopic, plastic hard (e.g., bone)
and reconstructive, tissue is desirable;
neurosurgery, neurosurgery, plastic
orthopedic, thoracic, and reconstructive
urologic surgery, orthopedic,
gynecologic, thoracic,
general surgery,
gastrointestinal and
affiliated organ
surgery, laparoscopic

CONFIGURATION Handpiece Handpiece

Standalone system No Yes


Components included Console, 2 Console, cart,
handpieces, handpiece
activated carbon filter
Accessories Wireless footswitch 36 kHz sterilization
tray, 36 kHz
sterilization torque
base, 36 kHz torque
wrench, CUSA Quick
Connect cartridge
and tubing set,
cleaning brush,
contamination guard,
specimen trap, 23
kHz sterilization tray,
CUSA Clarity 23k Hz
sterilizable torque
base, CUSA Clarity
23 kHz torque
wrench, CUSA Clarity
23 kHz handpiece
cleaning brush,
CUSA Clarity 23 kHz
CEM nosecone

Wireless functionality Yes NA


TRANSDUCER ASSEMBLY Permanently housed Permanently housed Permanently housed
in handpiece in handpiece in handpiece

ULTRASONIC SYSTEM
Working frequency, kHz 20-80 23 23, 36
Amplitude settings, µm 300 300 210 (36 kHz), 422
(23 kHz)
Oscillation system Piezoelectric Piezoelectric Piezoelectric

FLUIDIC SYSTEM
Aspiration, mm Hg 12.7-635, variable 0-640
Irrigation system Coaxial Coaxial
Fluid Saline Saline
Flow rate, cc/min 2-25, variable 2-20 variable, 25 fast
flush
HANDPIECE CONFIGURATIONS
Straight, kHz 23 Yes
Angled, kHz NA NA
Straight extended, kHz NA NA
Angled extended, kHz 23 NA
TIP
Construction material Titanium, titanium Titanium Titanium
alloy
Available types Straight, curved, Angled, angled
laparoscopic, macro, extended, soft tissue,
notched, bone shaver fibrous tissue, bone
cutting, deep access
Reusable/disposable Disposable Both
STERILIZATION Steam (gravity and Steam, flash,
pre-vac; reusable), autowash (reusable),
EtO (disposable) EtO (disposable)

FOOTSWITCH CONTROL Wireless Yes


Preset mode Preferrred Yes Yes
Linear mode Preferred Yes Yes
CONTROL CONSOLE Power, vacuum,
irrigation levels,
application time
Information displayed Irrigation, aspiration, Laparoscopic mode, Guided set-up
ultrasound aspiration, irrigation, instructions; online
vibration functions, IFU in 20 languages;
bar graphs 0-100, amplitude,
error codes, total TissueSelect,
ultrasound time aspiration, irrigation,
proportional (linear);
system information
(including access to
log file)

Controls Alarm override, alarm Touchscreen


volume, aspiration, controls: amplitude,
audio feedback, TissueSelect,
linear, preset select, aspiration, irrigation,
up/down ramp for proportional (linear)
irrigation, vibration, footswitch mode; on-
aspiration, flush, demand aspiration
coagulation only mode (laparoscopic
mode), fast flush
through footswitch,
prime, test, settings

Alarms and safety features Preferred Clogged/damaged Battery; online help,


tip, handpiece alerts, failure modes
malfunction, result in error code
footswitch/electronics message which is
problem displayed and logged
with the system
memory and can be
exported to USB
drive

POWER REQUIREMENTS
VAC 110-240 100-240
Hz 50/60 50/60
Amp 4/2 4.8/2 (480 VA)
H x W x D, cm (in) 102 x 64 x 48 (40 x 49.3 x 34.9 x 45.7
25 x 19) (19.5 x 13.8 x 18)
console, 133 x 57.2 x
64.8 (52.5 x 22.5 x
25.5) console and
cart

WEIGHT
Ultrasonic system, kg (lb) 55 (120) 29.5 (65) console,
81.5 (179.6) console
and cart
Handpieces, g (oz) 91 (3.2) straight with 71 (2.5), 36 kHz
tip, 118 (4.2) curved excluding weight of
extended with tip cable; 102 (3.6), 23
kHz excluding weight
of cable

WORKING CONDITIONS
Temperature 13-30°C (55-86°F) 10-35°C (50-95°F)
Relative humidity 15-90%, non- 30-85%, non-
condensing condensing
STORAGE CONDITIONS
Temperature -20 to +50°C (-4 to -18 to +60°C (-0.4 to
122°F) +140°F)
Relative humidity 15-90%, non- 15-85%, non-
condensing condensing
PURCHASE INFORMATION
List price Not specified $211,000
Handpieces Not specified $53,916
Warranty 1 year, console and 1 year
handpieces
Manufacturer supports user Not specified Not specified
repair
Training Optional field training Set-up and operation
in-service training,
operator's manual,
24-hr customer
support

Delivery time, ARO From stock From stock


Fiscal year July to June Not specified
OTHER SPECIFICATIONS Laparoscopic Tough Tissue
functionality, high Technology including
flow/low vacuum, RF Precision Handpiece
electrocautery (36 kHz), Tough
capable; wireless Tissue Tip with
technology; ShearTip
OsteoSculpt bone- Technology, Secure
shaving tips; can be Suction, and
combined with Adaptive Power;
electrosurgery RF TissueSelect; Quick
using RF interface Connect cartridge
components. Meets and tubing set;
requirements of UL complete system log
2601-1/IEC 601-1. file exportable to
USB drive (including
time-stamped system
performance data,
settings); ergonomic
handpiece designed
with thermal shield.

UMDNS CODE(S) 15756 15756 15756


LAST UPDATED October 2022<1> April 2024
Supplier Footnotes These
recommendations are
the opinions of
ECRI's technology
experts. ECRI
assumes no liability
for decisions made
based on this data.

Model Footnotes
Data Footnotes <1>Specifications
updated using
manufacturer's
website.
Soering GmbH Stryker Instruments
SONOCA 300 Div Stryker Corp
Sonopet iQ

Worldwide Worldwide
Yes Yes
Yes No
General, Gastrointestinal and
gynecologic, affiliated organ
laparoscopic, surgery, general
minimally invasive, surgery, gynecologic,
neurosurgery, laparoscopic,
thoracic neurosurgery
(fragmentation,
emulsification,
aspiration),
orthopedic, plastic
and reconstructive,
thoracic, urological
surgery<1>

Tabletop Console, handpieces

Yes Yes
Integrated pumps, Console, handpieces
console, handpiece
Not specified 3 foot pedal options,
cords, sterilization
tray, torque wrench,
disposable tubing, IV
pole, hand controller,
tips

Not specified Yes


Permanently housed Permanently housed
in handpiece in handpiece

25, 35, 55 25
200 peak 2-300

Electrostrictive Piezoelectric
piezoelectric

0-675 0-515
Central or outside NA
Saline Saline
0-150 3-40

25, 35 NA
25, 35 25
25, 35 NA
25, 35 NA

Titanium alloy Titanium

Not specified Soft tissue, fibrous


tissue, bone cutting,
skull base/deep
access
Reusable Disposable
Autoclavable at Steam
134°C

Yes Yes
Yes Yes
No Yes

Ultrasound, active Power, aspiration,


time, aspiration, irrigation, variable,
frequency, irrigation sync, flush

Aspiration, irrigation, Power, aspiration,


ultrasound irrigation, variable,
sync, flush

Functional error, Various, use only


handpiece Stryker approved
malfunction components

110-120/220-240 240
50/60 50/60
2/1 6.3
15 x 43 x 56 (5.9 x 23.4 x 33 x 44.2 (9.2
16.9 x 22) x 13 x 17.4)

24 (52.9) 18.28 (40.3)

147 (5.2) 320 (11.29) including


straight/short curved, cord
163 (5.8) long
curved, 281 (9.9)
laparoscopic, 114 (4)
micro handpiece

Not specified Not specified


Not specified Not specified

Not specified Not specified

Not specified Not specified

Not specified $189,520<2>


Not specified $46,350
1 year 1 year

Not specified Not specified


Not specified Not specified

3 weeks Not specified


Not specified Not specified
3 default working Longitudinal torsional
frequencies (LT) vibration
available; optional technology, RFID.
cart.

15756 15756
March 2014<1> February 2020

<1>Model is currently <1>Not indicated for


marketed. fragmentation,
emulsification and
aspiration of uterine
fibroids. <2>Console
price; pricing
information derived
from ECRI's
PricePaid database.

You might also like