Cunningham 2020

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

European Spine Journal

https://doi.org/10.1007/s00586-020-06440-5

ORIGINAL ARTICLE

Smartphone application technique for localising magnetically


controlled growth rod actuators: the Oxford Magnetic Counter App
Technique (TOMCAT)
Gregory Cunningham1 · Dan Wright1 · Colin Nnadi2

Received: 1 September 2019 / Revised: 2 April 2020 / Accepted: 27 April 2020


© Springer-Verlag GmbH Germany, part of Springer Nature 2020

Abstract
Purpose The purpose of this study is to report a novel smartphone app technique to localise a magnetically controlled grow-
ing rod (MCGR) actuator, along with determining this novel technique accuracy compared to current tactile localisation
techniques through an experimental study.
Methods Five spinal surgery fellows recorded attempts localising the MCGR magnetic actuator using a novel smartphone
app technique, MAGEC Wand and magnetic disc. Three attempts per technique were performed and repeated in both the
average and overweight patient models.
Results In total, 90 separate localisation attempts were recorded. The smartphone app produced less localisation error than
both the MAGEC Wand and magnetic disc. Mean difference was − 0.71 cm (95% CI − 1.24 to − 0.18 cm p = 0.06) and
− 0.58 cm (95% CI − 1.11 to − 0.04 cm p = 0.031), respectively. Mean localisation error for the smartphone app, MAGEC
Wand and magnetic disc, was 0.9 cm, 1.61 cm, 1.47 cm, respectively, for both average and overweight models combined.
Conclusions This novel smartphone app localisation technique is accurate. Current MAGEC Wand and magnetic disc tech-
niques produced more localisation error than the reported tolerance of the external remote control lengthening unit in this
experiment.

Keywords Scoliosis · Early onset scoliosis · Growing rods · Magnetic controlled growing rod · Lengthening · Smartphone

Introduction disparity between true and intended distractions increases


with the implant time in situ [6]. A number of plausible
Magnetically controlled growing rods (MCGR) have causations have been theorised including increasing patient
improved the treatment of early onset scoliosis [1, 2]. One weight, spinal stiffness and spontaneous fusion [3].
of the challenges facing this technology like all growing rod These theories, however, do not completely account for the
systems has been termed ‘the law of diminishing returns’ difference between true and intended distractions observed at
[3]. It has been shown that the amount of true distraction the first distraction before these patient factors have occurred.
achieved when lengthening the MCGR by the external It has been noted that the ideal distance between the internal
remote control (ERC) is less than that intended [4, 5]. This magnet of the MCGR and the external magnet of the ERC is
less than 1 cm to achieve optimal lengthening [7, 8]. There-
fore, it is reasonable to consider that some portion of the
* Gregory Cunningham unattained intended lengthening may be due to an inaccuracy
Greg@GregoryCunningham.com when localising the actuator magnet and the subsequent mis-
Colin Nnadi placement of the ERC when performing a lengthening pro-
colin.nnadi@ouh.nhs.uk cedure. The ERC is unable to determine if successful length-
1 ening is taking place or what true lengthening has occurred
Division of Spinal Surgery, Department of Orthopaedic
and Spinal Surgery, University of Oxford Hospital NHS if inaccurately positioned relative to the magnetic actuator.
Foundation Trust, Oxford, England, UK The technique of actuator localisation described by the
2
Division of Spinal Surgery, University of Oxford Hospital most current magnetic growing rod (NuVasive, CA USA)
NHS Foundation Trust, Oxford, England, UK technique guide is with the utilisation of a ‘MAGEC Wand’

13
Vol.:(0123456789)
European Spine Journal

magnet device held above the skin surface [8]. Previous tech-
nical guides had described the actuator localisation using a
magnetic disc before the ‘MAGEC Wand’ became available.
Both of these tactile techniques rely upon a practitioner sens-
ing the attraction force generated by either the Wand or disc.
We have identified a novel technique of magnetic actua-
tor localisation utilising a smartphone (iPhone Apple, CA
USA) magnet detector application (abbreviation: app). This
is made possible by modern smartphones that contain a mag-
netometer chip capable of detecting the magnetic field of the
actuator. Such apps do not rely upon a practitioner’s tactile Fig. 1  Experimental setup
perception of force but rather an objective numeric display
of magnetic field strength upon a smartphone.
This novel technique of localisation has not been previ- true actuator position was then calculated to yield the localisa-
ously described, and currently no other published study is tion error.
available reporting the accuracy of the two current tactile
techniques to localise the actuator magnet. Localisation techniques
The aim of our study is to report this novel smartphone
app-based localisation technique and compare its accuracy Participants were given simple instructions on how to use
to the existing tactile techniques. The secondary aim of this the ‘MAGEC Wand’ (Fig. 2a) and magnetic disc (Fig. 2b) to
study is to report and compare the accuracy of the two tactile localise the magnetic actuator. They were also provided with
localisation techniques. a separate MCGR to understand the tactile feeling generated
by the disc and wand when the magnets converged. To use the
smartphone magnet finder app (Fig. 2c), they were instructed
Materials and methods to rest the smartphone on the model’s surface, moving the
smartphone only in a craniocaudal direction until a maximal
Experimental model reading was generated. Once at this craniocaudal position, they
were then to move the smartphone in a mediolateral direction
An experimental model was designed to test the accuracy only until the maximal magnetometer reading was generated.
of the localisation methods. A plastic phantom thoracic and For clarity, participants were instructed that the exact location
lumbar human model was fitted with a MCGR to the dorsal of the magnetometer e-Compass within the smartphone, as
surface. The posterior surface of the model was covered with marked at 1.5 cm diagonally from the top left screen corner,
a contoured opaque silicone mat. This completely obscured would correlate with the location of the underlying magnetic
the growing rod to visualisation simulating the clinical expe- actuator. This e-Compass hardware location may vary between
rience of rod lengthening. Two different thicknesses of soft smartphone models, but the technical information specifying
tissue covering were replicated with either 4-cm or 7-cm- the exact locations is readily available [9].
thick silicone. A laminated 0.5-cm lined graph paper sheet
was applied to allow marking for recording of the participant Statistical analysis and clinical significance
localisation attempts (Fig. 1).
Five spinal surgery fellows took part in the experiment. Analysis of descriptive statistics, Students T Test and ANOVA
Each participant made three separate attempts per device were performed using SPSS 23 for Mac (IBM, Armonk, New
(Wand, disc and iPhone) to localise the magnetic actuator. York). A p value of less than 0.05 was considered statistically
All participants made attempts with each of the different significant. A localisation error of more than 1 cm, based on
devices in random order. The markings on the laminated published manufacturer recommendations, was considered
graph paper were cleaned away after every single localisa- clinically significant.
tion attempt, and each participant performed the localisation
test independently. There was no communication of results
between participants. The experiment was conducted firstly Results
using a 4-cm silicone covering to imitate average patient
adiposity and then a second time on a different occasion with A total of 90 separate localisation attempts were recorded
a 7-cm silicone covering to imitate an overweight patient. within the experiment; 45 attempts were recorded for each
The position of every localisation attempt was recorded as x of the average (4-cm silicone covering) and overweight
and y coordinates from the graph paper. The distance from the (7-cm silicone covering) models. A mean localisation error

13
European Spine Journal

Fig. 2  a MAGEC Wand. b


Magnetic disc. c Smartphone
app

Table 1  Mean measurement error by device disc; mean difference was − 0.71 cm (95% CI − 1.24 to
Average Overweight Combined
− 0.18 cm p = 0.006) and − 0.58 cm (95% CI − 1.11 to
patient model patient model patient models − 0.04 cm p = 0.031), respectively. There was no significant
(cm) (cm) (cm) difference in the accuracy of localisation when using the
MAGEC Wand and magnetic disc p = 0.824 (Table 2).
MAGEC Wand 1.29 (0.90) 1.93 (0.90) 1.61 (0.94)
In the overweight patient model, the smartphone app
Magnetic disc 0.88 (0.50) 2.07 (1.11) 1.47 (1.04)
produced significantly less localisation error than both
Smartphone app 0.76 (0.55) 1.03 (0.53) 0.90 (0.55)
the MAGEC Wand and magnetic disc; mean difference
Total 0.98 (0.70) 1.68 (0.98) 1.32 (0.91)
was -0.89 cm (95% CI − 1.67 to − 0.11 cm p = 0.021) and
cm centimetres. SD standard deviation − 1.03 cm (95% CI − 1.81 to − 0.25 cm p = 0.007), respec-
tively. There was no significant difference in the accuracy
of localisation when using the MAGEC Wand and magnetic
of 0.98 cm and 1.68 cm occurred in the average and over- disc p = 0.90 (Table 3).
weight models, respectively (Table 1). In the average adiposity patient model, there was no sig-
The combined data of localisation accuracy for each of nificant difference in accuracy of localisation between the
the devices in both the average and overweight models were smartphone app, MAGEC Wand and magnetic disc tech-
analysed. The smartphone app produced significantly less niques. The average error in localising the position of the
localisation error than both the MAGEC Wand and magnetic magnetic actuator was 0.98 cm in this model (Table 1).

Table 2  Measurement error in both overweight and average adiposity models

Technique Mean difference P 95% confidence interval


(cm)
Lower bound Upper bound

Smartphone App MAGEC Wand − 0.72* 0.006 − 1.24 − 0.18


Magnetic disc − 0.58* 0.031 − 1.11 − 0.04
MAGEC Wand Magnetic disc 0.13 0.824 − 0.40 0.67

*Indicates statistical significance (p ≤ 0.05)

Table 3  Measurement error in overweight adiposity model


Technique Mean difference P 95% confidence interval
(cm)
Lower bound Upper bound

Smartphone App MAGEC Wand − 0.89* 0.021 − 1.67 − 0.11


Magnetic disc − 1.03* 0.007 − 1.81 − 0.25
MAGEC Wand Magnetic disc − 0.14 0.900 − 0.92 0.64

*Indicates statistical significance (p ≤ 0.05)

13
European Spine Journal

When taking into account all measurement techniques, Compliance with ethical standards
there was significantly less error in localisation on the
average patient model compared to the overweight patient Conflict of interest No conflicts of interest to declare.
model (a mean of 0.98 cm on the average patient model and
Ethical standards Ethics board approval not required.
1.68 cm on the overweight model, p = 0.007).

References
Discussion
1. Pkm-Cc MD, Jp-Yc MBBS, DSc DS et al (2012) Magnetically
controlled growing rods for severe spinal curvature in young chil-
This study is the first to identify this novel MCGR magnetic dren: a prospective case series. Lancet 379:1967–1974. https​://doi.
actuator localisation technique. In this trial, the smartphone org/10.1016/S0140​-6736(12)60112​-3
app localisation technique has proved equivalent to existing 2. Doany ME, Olgun ZD, Kinikli GI et al (2018) Health-related
quality of life in early-onset scoliosis patients treated surgically:
techniques in the average patient adiposity model and supe-
EOSQ scores in traditional growing rod versus magnetically con-
rior in accuracy in the overweight patient model. trolled growing rods. Spine 43:148–153. https​://doi.org/10.1097/
MCGR may avoid the psychological stress associ- BRS.00000​00000​00227​4
ated with conventional growing rods by reducing hospital 3. Sankar WN, Skaggs DL, Yazici M et al (2011) Lengthening of dual
growing rods and the law of diminishing returns. Spine 36:806–809.
admissions and repeated anaesthetics [10]. The non-invasive
https​://doi.org/10.1097/BRS.0b013​e3182​14d78​f
lengthening procedure can still be a stressful experience 4. Lebon J, Batailler C, Wargny M et al (2017) Magnetically controlled
for a child in an unfamiliar environment. The possibility growing rod in early onset scoliosis: a 30-case multicenter study. Eur
of using the ubiquitous smartphone that most children are Spine J 26:1567–1576. https​://doi.org/10.1007/s0058​6-016-4929-y
5. Rolton D, Thakar C, Wilson-MacDonald J, Nnadi C (2016) Radi-
familiar with, instead of an unfamiliar device, may prove
ological and clinical assessment of the distraction achieved with
useful in reducing anxiety levels. The use of various smart- remotely expandable growing rods in early onset scoliosis. Eur Spine
phone applications in the spine surgery has already been J 25:3371–3376. https​://doi.org/10.1007/s0058​6-015-4223-4
validated for measuring range of neck movements, Cobb 6. Ahmad A, Subramanian T, Panteliadis P et al (2017) Quantifying
the “law of diminishing returns” in magnetically controlled growing
angles and thoracic rotation in scoliosis patients [11–13].
rods. Bone Joint J 99-B:1658–1664. https​://doi.org/10.1302/0301-
This study has also demonstrated the increased error 620x.99b12​.bjj-2017-0402.r2
in localisation of the magnetic actuator in the overweight 7. Cheung JPY, Cahill P, Yaszay B et al (2015) Special article:
patient model. This newly confirmed source of error may be Update on the magnetically controlled growing rod: tips and
pitfalls. J Orthop Surg (Hong Kong) 23:383–390. https​://doi.
partially responsible for the increased difference in intended
org/10.1177/23094​99015​02300​327
and true rod lengthening observed in patients with higher 8. (No Authors Listed) (2016) MAGEC Technique Guide, pp 1–20
BMI [6]. 9. Yang D, Wegner S, Fontaine R Apple iPhone 8 Plus Teardown. In:
One drawback from the smartphone-based technique is https​://techi​nsigh​ts.com/blog/apple​-iphon​e-8-plus-teard​own. https​
://techi​nsigh​ts.com/blog/apple​-iphon​e-8-plus-teard​own. Accessed
the need to know the exact location of the magnetometer
26 Apr 2019
e-compass which will vary between smartphone models. 10. Kain ZN, Wang SM, Mayes LC et al (1999) Distress during the
Fortunately, this information is readily available from tech- induction of anesthesia and postoperative behavioral outcomes.
nical instructions relating to smartphone repairs. The precise Anesth Analg 88:1042–1047
11. Pourahmadi MR, Bagheri R, Taghipour M et al (2018) A new
location of this internal hardware would need to be displayed
iPhone application for measuring active craniocervical range of
on the smartphone screen, specific to each phone model, for motion in patients with non-specific neck pain: a reliability and
accurate clinical use. validity study. Spine J 18:447–457. https​://doi.org/10.1016/j.spine​
This study is of course limited by its experimental nature. e.2017.08.229
12. Qiao J, Liu Z, Xu L et al (2012) Reliability analysis of a smartphone-
The phantom model was useful in creating a known soft tissue
aided measurement method for the cobb angle of scoliosis. Clin
thickness, not usually known in routine clinical practice. This Spine Surg 25:E88–E92. https​://doi.org/10.1097/BSD.0b013​e3182​
allowed the variable to be accurately investigated for its influ- 46396​4
ence on localisation accuracy. A dedicated smartphone app 13. Furness J, Schram Ben, Cox AJ et al (2018) Reliability and concur-
rent validity of the i­Phone® Compass application to measure tho-
specifically designed and tested for this purpose is required
racic rotation range of motion (ROM) in healthy participants. PeerJ
to ensure accuracy and reproducibility before wider clinical 6:e4431. https​://doi.org/10.7717/peerj​.4431
use. Alternatively, a dedicated device utilising a similar mag-
netometer could be considered. Following development, fur- Publisher’s Note Springer Nature remains neutral with regard to
ther clinical testing with confirmation of successful lengthen- jurisdictional claims in published maps and institutional affiliations.
ing on imaging would be needed before widespread adoption.

Funding No funding was received for this study.

13

You might also like